Wafer Rf characterisation
Gavin Fisher
Cascade Microtech Europe Ltd
Absorber Block
Absorber
Metal CPW Tip Low-Loss Cable
Thin-Film
Coax
0.25
Conventional Tungsten Probe
Contact resistance
0.2 on un-patterned
aluminum averages
0.15 about 30 mΩ over
Rc (Ω )
0.05
Infinity Probe
0
0 1000 2000 3000 4000 5000
Contact Cycle
Repeatability
Company Confidential See it. Touch it. Measure it. ®
New Infinity Waveguide Probe
Waveguide/flange
WR15 – 50-75GHz
WR12 – 67-90GHz
WR10 – 75-110GHz
WR8 – 90-140GHz
WR6 – 110-170GHz
SSMC bias T
connector WR5 – 140-220GHz
WR3 - 220-325GHz
(end 2005)
Membrane
coupon
Non-symmetrical
grounds can
cause resonance
loops even at
frequencies
<10GHz
DUT
Coplanar probe tips do not shield from the DUT
DUT
Microstrip structure shields signal line better
Company Confidential See it. Touch it. Measure it. ®
Probe Station Essentials
Completely Integrated
Measurement Environment
• FULL access to Positioners, Stage
and Microscope
• Roll-out stage – Complete chuck,
not just top layer
• Easy, fast & safe wafer loading
Triax connection
panels
• Easy power
supply
connections
• Cable strain relief
Goretm RF cables
• Low Loss
• Phase stable
• Flexible
Contact Substrate
• PN 005-018
• Dull gold finish
• Bright contact marks
Adjust planarity until
equal marks from all
probe contacts
Important to initialise instrument settings paying attention to power, number of points, Start and stop
and particularly IF bandwidth
It is important that only 50 ohm loads are used for this part
of the calibration
Following calibration a validation is carried out against a known standard. Typically this
is an open whose capacitance is known by the probe pitch, but can be a golden dut
whose characteristics are pre-measured and stored. For lrrm the open is the raw open
measured during the cal and corrected by the calibration (post corrected)
Various lengths of
transmission lines
0.4
0.3
Reflect
(probes in air) 0.2
0.1
[dB]
0.0
-0.1
-0.2
-0.3
40.0
10 GHz-0.4
-0.5
0 5 10 15 20 25 30 35 40
20 [GHz]
0.1
50 GHz
10 2050 • Unity Gain
-50
• Negative Capacitance
-20
-10
-Due to wave propagation being faster
in air than on the Alumna substrate
Company Confidential See it. Touch it. Measure it. ®
Independent Verification Standards
1
0.8
1.5
0.6
2
Open 0.4
3
stub 4
0.2 5
GROUND
10
SIGNAL 20
50
0.2 0.4 0.6 0.8 1 1.5 2 3 4 5 10 2050
0.1 GHz
GROUND -50
-20
-10
-0.2 -5
-4
-2
-0.5 -0.6
-1.5
-0.8
-1
[dB]
-1.0
• Linear Phase Lag
-1.5
• 50ohm to stub Z miss-match
0 5 10 15 20 25 30 35 40
• Fringing C at stub open end
[GHz]
RF Performance
• Pad configuration (GS Vs GSG)
• Probe pitch
Ability to Physically Probe
• Pad size
• Pad height
• Distance between probes
• Number of contacts per side
Calibration
• Paths
• Best calibration methods
• De-embedding devices
Company Confidential See it. Touch it. Measure it. ®
Pad Sizes
200um
>500um
RF probes should have more than 200um separation to avoid cross-
talk
All pads must be on top surface
All grounds should be connected together
Adjacent devices should be >500um away for mm-wave
measurements (>250um for Infinity)
Company Confidential See it. Touch it. Measure it. ®
Maximum Probe Contacts
calibration is
VERY
repeatable
Frequency (GHz)
Company Confidential See it. Touch it. Measure it. ®
Pad Parasitic Removal
G G
S S
G G