Escolar Documentos
Profissional Documentos
Cultura Documentos
Model YC-1800
SERVICE MANUAL
XYC16*RDA001A/E
Total pages: 86
2005.9
NIDEK CO., LTD. : 34-14, Maehama, Hiroishi-cho, Gamagori, Aichi 443-0038, Japan
(Manufacturer) Telephone: (0533) 67-6611
Facsimile: (0533) 67-6610
NIDEK CO., LTD : 6th Floor, Takahashi Bldg., No.2, 3-chome, Kanda-jinboucho
(Tokyo Office) Chiyoda-ku, Tokyo 101-0051, Japan
Telephone: (03) 3288-0571
Facsimile: (03) 3288-0570
Telex: 2226647 NIDEK J
NIDEK INCORPORATED : 47651 Westinghouse Drive, Fremont, California 94539, U. S. A.
(United States Agent) Telephone: (510) 226-5700
Facsimile: (510) 226-5750
NIDEK SOCIETE ANONYME : Europarc 13, rue Auguste Perret, 94042 CRETEIL, France
(Authorized Representative) Telephone: (01) 49 80 97 97
Facsimile: (01) 49 80 32 08
Table of Contents
§1 Page
INTRODUCTIONS ...................................................................................................... 1-1
§1 INTRODUCTION
• This service manual is for the OPHTHALMIC YAG LASER SYSTEM YC-1800.
• Be sure to use this manual after understanding the operator’s manual for the OPHTHALMIC
YAG LASER SYSTEM YC-1800 thoroughly.
• Refer to the OPHTHALMIC YAG LASER SYSTEM YC-1800 Operator’s Manual and Parts
List.
• Specifications are subject to change without notice for improvement. Refer to the Technical
Bulletin for important changes.
• In case the system can not be repaired according to the procedures described in this manual,
please tell us the serial number of the system and details of the symptom or symptoms.
1-1
XYC16*RDA001A/E
1-2
XYC16*RDA001A/E
§2 CAUTIONS
2.1 General
• Perform all servicing according to the procedures described in this manual. Failure to do so
may cause unexpected accident or malfunction.
• Be sure to turn the power off during servicing unless there is an instruction to turn it on.
• If malfunction occurs, check the symptom and turn the power off.
• Apply a threadlocking adhesive to the screws which have been loosened and tightened again.
• If the YC-1800 produces smoke or strange odors, immediately turn off the system and disconnect
the electrical plug from the outlet.
Usage of the system under such abnormal conditions may cause fire or electric shock.
2-1
XYC16*RDA001A/E
2.5 Adjustment
• Do not use adjustment jigs for other than the intended purpose.
• Since the Nd:YAG laser used in the YC-1800 is an invisible infrared ray, improper operation
may cause the hazardous exposure to radiation.
• Never face to the laser beam aperture in the laser beam path.
• Never gaze directly the Nd:YAG laser beam even when wearing the safety goggles.
• Do not touch optical parts such as mirror or lens with bare hands.
• Take care not to receive electric shock when opening the cover to check or repair the electrical
system.
• Depending on the energy meter to be used, the necessity of the dispersion lens, which is attached
to the objective lens at the time of energy measurement, is varied.
Refer to the table below when measuring the energy.
Energy meter Dispersion lens
MODEL-365 Necessary
MODEL PHDX50 Unnecessary
• The detector of the energy meter may be damaged if the excessive energy density is applied to
the photoreceptor of the detector.
Maximize the spot size of the aiming beam on the photoreceptor as possible so that the YAG
laser beam never come into focus on the photoreceptor.
2-2
XYC16*RDA001A/E
§3 TROUBLESHOOTING
No
Does the pilot lamp light up? 4.1 The pilot lamp does not light up.
Yes
Yes
Does an error display appear? 4.4 The error display appears.
No
No
Is the illumination condition normal? 4.9 There are some spots in the illumination light.
Yes
3-1
XYC16*RDA001A/E
No
Is the aiming beam visible? 4.13 The aiming beam is not emitted.
Yes
3-2
XYC16*RDA001A/E
Does the pressing the emergency stop switch stop No 4.20 The YC-1800 does not stop with the emergency
the YC-1800? stop switch.
Yes
End
3-3
XYC16*RDA001A/E
3-4
XYC16*RDA001A/E
§4 SUB TROUBLESHOOTING
Yes
Is the voltage between the P10 (J10) 1st and 2nd pins on No
Replace the switching power supply.
the switching power supply (13704-E020) DC + 12 V?
Yes
Is the voltage between the J20 1st and 3rd and 2nd and 4th No Replace the base plate assy.
pins on the base plate assy. (13706-EA17) DC + 12 V? (Refer to 6.2.10.)
Yes
Is the voltage between the P401 (J401) 1st and 3rd and
No
2nd and 4th pins on the relay board (13706-BA04) DC Replace the cable assy. (13706-EA16).
+ 12 V?
Yes
Is the voltage between the J403 4th and 5th pins on the No Replace the relay board.
relay board DC + 2.0 V? (Refer to 6.1.2.)
Yes
4-1
XYC16*RDA001A/E
Is the voltage between the J40 1st and 2nd pins on the No Replace the base plate assy.
base plate assy. (13706-EA17) DC + 12 V? (Refer to 6.2.10.)
Yes
Yes
Yes
Is there a break in the flexible flat cable (13706-E020)? Replace the flexible flat cable.
No
4-2
XYC16*RDA001A/E
Does the solenoid of the YAG shutter function when No Replace the YAG shutter assy. (13706-EA03).
the YC-1800 is in READY? (Refer to 6.2.1.)
Yes
Is the voltage between the P103 (J103) 3rd and 7th and Yes
Replace the YAG shutter assy. (Refer to 6.2.1.)
4th and 7th pins on the master board DC + 0 V or 5 V?
No
Yes
4-3
XYC16*RDA001A/E
Does the solenoid of the aiming shutter function when No Replace the aiming shutter (13706-EA04).
the YC-1800 is in READY? (Refer to 6.2.2.)
Yes
Is the voltage between the P104 (J104) 3rd and 5th and Yes
Replace the aiming shutter. (Refer to 6.2.2.)
6th and 5th pins on the master board DC + 0 V or 5 V?
No
Yes
4-4
XYC16*RDA001A/E
Is 4 (number of pulses) displayed on the COUNTER Yes Replace the master board (13706-BA01).
display in eight or more of ten shots? (Refer to 6.1.1.)
No
End
Use the YC-1800 away from the instruments which may make noises.
4-5
XYC16*RDA001A/E
No
Is the PFN voltage normal? Adjust the PFN voltage. (Refer to 7.5.4 and 7.5.5.)
Yes
No
Yes
End
4-6
XYC16*RDA001A/E
End
No
4-7
XYC16*RDA001A/E
No
End
4-8
XYC16*RDA001A/E
4-9
XYC16*RDA001A/E
Yes
Is there a break in the cable assy. (13706-EA16)? Replace the cable assy.
No
4 - 10
XYC16*RDA001A/E
4 - 11
XYC16*RDA001A/E
Yes
Is the illumination lamp (34530-E003) blown? Replace the illumination lamp.
No
Change the pupil distance while looking through the binocular tube for each eye.
When the illumination light shifts from the center of the visual field through the eyepiece, adjust
the optical axis of the binocular microscope. (See 7.1.1.)
4 - 12
XYC16*RDA001A/E
4.7 The Illumination Light is not in the Center of the Visual Field
when the Magnification is 12.5×.
4 - 13
XYC16*RDA001A/E
No
Yes
Is the slide plate broken? Replace the slide plate. (Refer to 6.22.)
No
Yes
Is the rack (34348-M009) broken? Replace the rack. (Refer to 6.23.)
No
Yes
Is the pinion (32903-M138) broken? Replace the pinion. (Refer to 6.24.)
No
No
4 - 14
XYC16*RDA001A/E
Yes
It the aiming beam emitted when the voltage No Replace the aiming LD assy. (13706-EA10).
between TP12 and GND on the master board? (Refer to 6.2.6.)
Yes
Yes
Is the trouble resolved? END
No
END
4 - 15
XYC16*RDA001A/E
4.15 The Burn Pattern is not Aligned with the Aiming Beam.
Adjust the optical axis of the YAG laser beam. (Refer to 7.4.)
Be sure to tape up a burn paper so that the optical system is not grimed with soot.
Is the burn pattern uneven at the front of the beam No Blow off the dirt on the beam expander with the
expander (expander assy.)? (D in the figure below) blower.
Yes
Is the burn pattern uneven at the front of the dichroic No Blow off the dirt on the dichroic mirror with the
mirror (D mirror assy.)? (C in the figure below) blower.
Yes
Yes
&
$
4 - 16
XYC16*RDA001A/E
Yes
Can the number of pulses set? End
No
End
Yes
Can the number on the COUNTER display become 0? End
No
End
4 - 17
XYC16*RDA001A/E
4.19 The Energy Output Measured with the YAG Energy Meter is
Abnormal.
Adjust the YAG laser energy output.
(Refer to 7.6.)
End
4.20 The YC-1800 does not Stop with the Emergency Stop Switch.
Disconnect P404 (J4) on the relay board (13706-
BA04).
4 - 18
XYC16*RDA001A/E
§5 COVER REMOVEMENT
%-
5-1
XYC16*RDA001A/E
4. Unscrew CK2 × 42 to remove the vertical %-
movement cover (13706-M332).
5-2
XYC16*RDA001A/E
2. Remove the vertical movement cover (13706-
M332). (Refer to 5.4.)
26
5-3
XYC16*RDA001A/E
5-4
XYC16*RDA001A/E
§6 PARTS REPLACEMENT
6.1 Boards
6.1.1 Master board
6-1
XYC16*RDA001A/E
(-
**
6-2
XYC16*RDA001A/E
7. After reassembling the YAG shutter, adjust Not shielded Shielded Shielded Not shielded
its sensor position so that the shutter plate
(13706-M116) is as shown in the right figures.
6-3
XYC16*RDA001A/E
2%
12'0 %.15'
6-4
XYC16*RDA001A/E
6-5
XYC16*RDA001A/E
6-6
XYC16*RDA001A/E
26
6-7
XYC16*RDA001A/E
5$
6-8
XYC16*RDA001A/E
6-9
XYC16*RDA001A/E
6.8 LD Assy.
6 - 10
XYC16*RDA001A/E
2%
6 - 11
XYC16*RDA001A/E
5$
6.12 Microscope
6 - 12
XYC16*RDA001A/E
6 - 13
XYC16*RDA001A/E
5$
6.17 Joystick
6 - 14
XYC16*RDA001A/E
2%
6.19 Head Rest
2%
6 - 15
XYC16*RDA001A/E
6 - 16
XYC16*RDA001A/E
6.23 Rack
6 - 17
XYC16*RDA001A/E
6.24 Pinion
6.25 Gear A
6.26 Gear B
5$
6 - 18
XYC16*RDA001A/E
§7 ADJUSTMENT
aligned in center.
1) Up and down: Tighten or loosen SB4 × 64
and HH4 × 64.
2) Right and left: Loosen SB4 × 64 to move the
microscope to the right or left.
7-1
XYC16*RDA001A/E
1) When the slit image moves opposite to the 5$
lamp house movement: Tighten HH3 × 82 on ** **
the patient’s side so that the focusing rod tilts
to the operator’s side.
5$
2) When the slit image moves along with the ** **
lamp house movement: Loosen HH3 × 82 on
the operator’s side so that the focusing rod
tilts to the patient’s side.
7-2
XYC16*RDA001A/E
7.3.1 LD assy.
()
7-3
XYC16*RDA001A/E
7-4
XYC16*RDA001A/E
7-5
XYC16*RDA001A/E
1. Mount the focal point jig on the chinrest with SB3 × 152.
2. Minimize the energy output with the control on the energy adjustment assy. (13706-1110).
3. Focus the aiming spot on the focal point jig
and emit the YAG laser beam. Check the
burn pattern position. 2%
7-6
XYC16*RDA001A/E
1. Mount the focal point jig on the chinrest with SB3 × 152.
2. Minimize the energy output with the control on the energy adjustment assy. (13706-1110).
3. Set the focal point to 0 µm.
1) Focus the aiming spot on the focal point jig.
2) Turn the knob so that the plunger (82001-
PR006) edge of the attachment plate (13706-
**
M114) is pointing down.
3) Loosen HH3 × 3 in the attachment plate.
4) Turn the control on the focus assy. (13706-
1130) so that the burn pattern on the focal
point jig is at the minimum.
5) Tighten HH3 × 3 so that the plunger edge of
the attachment plate is pointing down and
that the control on the focus assy. is fixed.
6) When the burn pattern on the focal point jig
is not at the minimum, perform the adjustment 24
again.
4. Check that the YAG laser beam is not obscured.
1) Maximize the energy output with the control on the energy adjustment assy.
2) Put a burn paper beyond the aiming focus point and emit several shots of the YAG laser
beam.
3) Check the burn pattern. If the laser is obscured, perform the following:
a. Mirror holder 3 and YAG laser head adjustment (Refer to 7.4.1.)
b. Focus assy. position adjustment (Refer to 7.4.2.)
c. D mirror assy. adjustment (Refer to 7.4.3.)
5. Adjust the focal point.
1) Minimize the energy output with the control on the energy adjustment assy.
2) Set the focal point jig to 0.
3) Move the slit so that the aiming spot is focused on the jig.
4) Set the focus shift to 0 µm with the focus shift control. Emit the YAG laser beam and check
the burn pattern.
5) Turn the jig scale one rotation toward you and set it to 0 again.
6) Set the focus shift to ATN500 µm with the focus shift control. Emit the YAG laser beam and
check the burn pattern.
7) Turn the jig scale two rotations away from ǴO ǴO
focus shift control. Emit the YAG laser beam ǴO ǴO
7-7
XYC16*RDA001A/E
** '
5. Set the focus shift control to ANT500 µm.
6. Press the aiming switches to display Address 40 on the COUNTER display.
7. Press the reset switch to set the parameter address.
8. As the same manner, set the parameters as shown in the table below (ANT375 to POST500).
Focus ANT POST
shift 500 375 250 125 0 125 250 375 500
Address 40 41 42 43 44 45 46 47 48
9. Set DIP SW1-6 on the display board in the display panel assy. to the ON (lower) position.
10. Turn the focus shift control and check that the YAG FOCUS SHIFT display is proper.
* When the display is abnormal, perform this adjustment again.
7-8
XYC16*RDA001A/E
7-9
XYC16*RDA001A/E
7 - 10
XYC16*RDA001A/E
$#
)
)
%-
7 - 12
XYC16*RDA001A/E
7 - 13
XYC16*RDA001A/E
1. Set DIP SW 1-1, 1-5, and 1-6 on the display board (13704-BA02) to the OFF (upper) position.
2. Turn the energy control to minimize the YAG laser energy.
3. Turn the master switch on while pressing the STATUS switch.
4. Emit five shots of the YAG laser beam and press the reset switch.
5. Perform the same adjustment for the energy settings of 1, 3, 5, 7.5 mJ and the maximum.
6. Turn the master switch off.
7. Set DIP SW 1-1, 1-5, and 1-6 to the ON (lower) position.
8. Turn the energy control to maximize the YAG laser energy. Turn the master switch on.
9. Press the burst switch to set the number of pulses to 1, emit five shots of the YAG laser
beam, and press the reset switch.
10. Press the burst switch to set the number of pulses to 2, emit five shots of the YAG laser
beam, and press the reset switch.
11. Press the burst switch to set the number of pulses to 3, emit five shots of the YAG laser
beam, and press the reset switch.
12. Turn the master switch off and on.
13. Turn the energy control to minimize the YAG laser energy.
14. Mount the detector of the powermeter on the chinrest.
15. Turn the illumination off.
16. Mount the dispersion lens on the objective lens at the edge of the microscope.
This step is unnecessary when using S200 (PHDX50).
17. Align the two aiming spots to the center of the detector.
18. Emit five shots of the YAG laser beam and check that the fifth actual energy output is
within the following standard.
[Standard]
Actual energy output less than 1.5 mJ: Displayed energy output ±15% ±0.1 mJ
Actual energy output of 1.5 mJ or more: Displayed energy output ±15%
Measure the actual energy output with each pulse number of 1, 2, and 3. Check that the difference
between the actual and displayed energy outputs is within the standard.
19. When the actual energy output is out of Address Display Contents Initial value
the standard, change the parameters in the 30 PPL Preset potentiel Min. 35
right table and repeat Steps 1 to 18. 31 PP1 Preset potentiel 1 mJ 50
32 PP3 Preset potentiel 3 mJ 70
33 PP5 Preset potentiel 5 mJ 85
34 PP7 Preset potentiel 7.5 mJ 100
35 PPH Preset potentiel Max. 125
20. Check that the actual energy output when the displayed one is 0.5 mJ satisfies the following
standard.
1) Set the YAG laser energy output to 0.5 mJ.
2) Press the burst switch to set the pulse number to 1.
3) Emit five shots of the YAG laser beam and check that the fifth actual energy output is within
the following standard.
[Standard]
Actual energy output of 0.5 mJ: Displayed energy output ±15% ±0.1 mJ
7 - 14
XYC16*RDA001A/E
Set the energy to 4 mJ and emit the YAG laser beam. Check that the air plasma occurs.
1. Start the YC-1800 while pressing the STATUS (READY mode) and burst switches.
2. Press the + aiming switch to change the following.
1) Total shot number (displayed number × 1000 shots)
2) Historical error number
3) Program version number
Start the YC-1800 while pressing the STATUS (READY mode) and two aiming switches
simultaneously. So, the energy display setting and aiming beam output adjustment are possible
with the control panel.
7 - 15
XYC16*RDA001A/E
1. Adjust parameter for Address 109 so that the energy control is rotated and the value on
the ENERGY display is changed slightly when the smart switch is moved to the right and
left.
2. Move the smart switch to either of the right and left and immediately move it to another
side. At that time, adjust parameter for Address 108 so that the energy control is rotated
and the value on the ENERGY display is changed slightly.
3. Perform Step 1 again.
7 - 16
XYC16*RDA001A/E
§8 INFORMATION
8.1 Wiring Diagram
8-1
XYC16*RDA001A/E
13706-EA01
13706-EA02
13706-EA03
8-2
XYC16*RDA001A/E
13706-EA04
13706-EA05
13706-EA06
8-3
XYC16*RDA001A/E
13706-EA07
13706-EA08
13706-EA09
8-4
XYC16*RDA001A/E
13706-EA10
13706-EA11
13706-EA12
8-5
XYC16*RDA001A/E
13706-EA13
13706-EA14
13706-EA15
8-6
XYC16*RDA001A/E
13706-EA16
8-7
XYC16*RDA001A/E
17306-EA17
8-8
XYC16*RDA001A/E
8.3 Appearance
Forehead rest
Control panel
Grips
Tonometer mount
8-9
XYC16*RDA001A/E
Chinrest
Hand switch
Microscope arm fixing knob
Smart switch
Grip
AC adapter
Power cable
Connecting cable
8 - 10
XYC16*RDA001A/E
2 Unassigned
3 Unassigned
4 Unassigned
5 Unassigned
6 For permitting access to the EEPROM
7 For adjustment
8 For running the system
8 - 11
XYC16*RDA001A/E
8 - 12
XYC16*RDA001A/E
Add Abbre
Default Contents Setting
ress viation
36 n-A 100 Unassigned —
37 n-A 100 Unassigned —
38 n-A 100 Unassigned —
39 n-A 100 Unassigned —
40 F-5 70 Focus shift value –500 µm
41 F-3 85 Focus shift value –375 µm
42 F-2 105 Focus shift value –250 µm
43 F-1 125 Focus shift value –125 µm
44 F0 145 Focus shift value 0 µm
45 F1 165 Focus shift value +125 µm
46 F2 185 Focus shift value +250 µm
47 F3 205 Focus shift value +375 µm
48 F5 225 Focus shift value +500 µm
49 n-A 0 Unassigned —
50 brt 30 Soft blinking time
Time that the shutter functions during
51 ruS 250
running
52 ruF 10 Number of shots during running ×100
53 n-A 0 Unassigned —
54 n-A 0 Unassigned —
55 ErG 100 Err15 gain adjustment (%)
56 ErO 3 Err15 offset adjustment (mJ)
57 n-A 100 Unassigned
58 StG 20 Gain adjustment of stability (%)
59 StO 1 Offset adjustment of stability (mJ)
60 PEL 3 Preset energy Min.
61 PE1 10 Preset energy 1 mJ
62 PE3 30 Preset energy 3 mJ
63 PE5 50 Preset energy 5 mJ
64 PE7 75 Preset energy 7.5 mJ
65 PEH 100 Preset energy Max.
66 n-A 0 Unassigned
67 n-A 0 Unassigned
68 n-A 0 Unassigned
69 n-A 0 Unassigned
Time required for resetting the protective Unassigned at
70 PF 0
filter present
Stand-by time required for opening each Value × (10 msec)
71 StO 50
shutter
Stand-by time required for closing each Value × (10 msec)
72 StC 50
shutter
73 n-A 10 Unassigned
74 n-A 10 Unassigned
75 n-A 0 Unassigned
8 - 13
XYC16*RDA001A/E
Add Abbre
Default Contents Setting
ress viation
76 ESb 1 Detection bit of the energy control
77 FSb 1 Detection bit of the focus shift control
78 n-A 0 Unassigned
79 n-A 0 Unassigned
80 Sti 33 Latency time of single (×10 msec) Value × (10 msec)
81 bti 66 Latency time of burst (×10 msec) Value × (10 msec)
82 Uti 200 Latency time of others (×10 msec) Value × (10 msec)
83 n-A 100 Unassigned
84 n-A 100 Unassigned
85 n-A 100 Unassigned
86 n-A 100 Unassigned
162: 4800
87 bAr 162 Baud rate
80: 9600
88 n-A 100 Unassigned
89 n-A 100 Unassigned
Pulse number for automatic PFN
90 PuL 0
adjustment
0: Fixed
91 End 1 Display change 1: Weight average
2: Actual indication
92 n-A 0 Unassigned —
93 Std con 0 Setting depending on destination Unassigned at present
94 bSt 0 Burst (reserved)
95 bno 3 Burst (reserved)
96 n-A 0 Unassigned
0: Display enabled
97 tOE 0 Total energy display
From 1: Enabled (×10
msec)
98 n-A 0 Unassigned —
0: Disabled
99 FPr 2 Motion control of the sub-switches 1: Foot SW enabled
2: Hand SW enabled
100 n-A 0 Unassigned —
0: Function disabled
1: ENERGY DOWN
2: ENERGY UP
3: AIMING DOWN
101 F01(R) 0 SW functions
4: AIMING UP
5: READY
6: BURST
7: RESET
0: Function disabled
1: ENERGY DOWN
2: ENERGY UP
3: AIMING DOWN
102 F02(L) 0 SW functions
4: AIMING UP
5: READY
6: BURST
7: RESET
8 - 14
XYC16*RDA001A/E
Add Abbre
Default Contents Setting
ress viation
0: Function disabled
1: ENERGY DOWN
2: ENERGY UP
3: AIMING DOWN
103 H01(R) 2 SW functions
4: AIMING UP
5: READY
6: BURST
7: RESET
0: Function disabled
1: ENERGY DOWN
2: ENERGY UP
3: AIMING DOWN
104 H02(L) 1 SW functions
4: AIMING UP
5: READY
6: BURST
7: RESET
H03
105 5 SW functions
(press)
106
to n-A 0 Unassigned
108
109 Ont 4 Motor ON time Value × (10 msec)
110
to n-A 0 Unassigned
197
198 PuP 0 Update of the ROM
199 EEP 0 Default setting of the EEPROM 99: Default setting
8 - 15