Resuscitation (2005) 67S1, S25—S37

European Resuscitation Council Guidelines for


Resuscitation 2005
Section 3. Electrical therapies: Automated
external defibrillators, defibrillation,
cardioversion and pacing
Charles D. Deakin, Jerry P. Nolan

Introduction the rhythm while CPR is in progress is required to


prevent unnecessary delays in CPR. Waveform anal-
This section presents guidelines for defibrillation ysis may also enable the defibrillator to calculate
using both automated external defibrillators (AEDs) the optimal time at which to give a shock.
and manual defibrillators. All healthcare providers
and lay responders can use AEDs as an integral com-
ponent of basic life support. Manual defibrillation is A vital link in the chain of survival
used as part of advanced life support (ALS) therapy.
In addition, synchronised cardioversion and pacing Defibrillation is a key link in the Chain of Survival
are ALS functions of many defibrillators and are also and is one of the few interventions that have been
discussed in this section. shown to improve outcome from VF/VT cardiac
Defibrillation is the passage across the myocard- arrest. The previous guidelines, published in 2000,
ium of an electrical current of sufficient magnitude rightly emphasised the importance of early defib-
to depolarise a critical mass of myocardium and rillation with minimum delay.1
enable restoration of coordinated electrical activ- The probability of successful defibrillation and
ity. Defibrillation is defined as the termination of subsequent survival to hospital discharge declines
fibrillation or, more precisely, the absence of ven- rapidly with time2,3 and the ability to deliver
tricular fibrillation/ventricular tachycardia (VF/VT) early defibrillation is one of the most important
at 5 s after shock delivery; however, the goal of factors in determining survival from cardiac
attempted defibrillation is to restore spontaneous arrest. For every minute that passes following
circulation. collapse and defibrillation, mortality increases
Defibrillator technology is advancing rapidly. AED 7%—10% in the absence of bystander CPR.2—4 EMS
interaction with the rescuer through voice prompts systems do not generally have the capability to
is now established, and future technology may deliver defibrillation through traditional paramedic
enable more specific instructions to be given by responders within the first few minutes of a call,
voice prompt. The ability of defibrillators to assess and the alternative use of trained lay responders

0300-9572/$ — see front matter © 2005 European Resuscitation Council. All Rights Reserved. Published by Elsevier Ireland Ltd.
doi:10.1016/j.resuscitation.2005.10.008
S26 C.D. Deakin, J.P. Nolan

to deliver prompt defibrillation using AEDs is now Automated external defibrillators have been
widespread. EMS systems that have reduced time to tested extensively against libraries of recorded
defibrillation following cardiac arrest using trained cardiac rhythms and in many trials in adults18,19
lay responders have reported greatly improved and children.20,21 They are extremely accurate in
survival-to-discharge rates,5—7 some as high as rhythm analysis. Although AEDs are not designed to
75% if defibrillation is performed within 3 min of deliver synchronised shocks, all AEDs will recom-
collapse.8 This concept has also been extended to mend shocks for VT if the rate and R-wave mor-
in-hospital cardiac arrests where staff, other than phology exceed preset values.
doctors, are also being trained to defibrillate using
an AED before arrival of the cardiac arrest team.
When bystander CPR is provided, the reduction in In-hospital use of AEDs
survival rate is more gradual and averages 3%—4%
At the time of the 2005 Consensus Confer-
per minute from collapse to defibrillation;2—4
ence, there were no published randomised trials
bystander CPR can double2,3,9 or treble10 sur-
comparing in-hospital use of AEDs with manual
vival from witnessed out-of-hospital cardiac
defibrillators. Two lower level studies of adults
arrest.
with in-hospital cardiac arrest from shockable
All healthcare providers with a duty to perform
rhythms showed higher survival-to-hospital dis-
CPR should be trained, equipped, and encouraged
charge rates when defibrillation was provided
to perform defibrillation and CPR. Early defibrilla-
through an AED programme than with manual defib-
tion should be available throughout all hospitals,
rillation alone.22,23 A manikin study showed that
outpatient medical facilities and public areas of
use of an AED significantly increased the likelihood
mass gathering (see Section 2). Those trained in
of delivering three shocks, but increased the time
AED use should also be trained to deliver at least
to deliver the shocks when compared with manual
external chest compressions before the arrival of
defibrillators.24 In contrast, a study of mock arrests
ALS providers, to optimise the effectiveness of early
in simulated patients showed that use of monitor-
defibrillation.
ing leads and fully automated defibrillators reduced
time to defibrillation when compared with manual
defibrillators.25
Automated external defibrillators
Delayed defibrillation may occur when patients
sustain cardiac arrest in unmonitored hospital beds
Automated external defibrillators are sophisti-
and in outpatient departments. In these areas sev-
cated, reliable computerised devices that use voice
eral minutes may elapse before resuscitation teams
and visual prompts to guide lay rescuers and health-
arrive with a defibrillator and deliver shocks.26
care professionals to safely attempt defibrillation
Despite limited evidence, AEDs should be consid-
in cardiac arrest victims. Automated defibrillators
ered for the hospital setting as a way to facilitate
have been described as ‘‘. . . the single greatest
early defibrillation (a goal of <3 min from collapse),
advance in the treatment of VF cardiac arrest since
especially in areas where staff have no rhythm
the development of CPR.’’11 Advances in technol-
recognition skills or where they use defibrillators
ogy, particularly with respect to battery capacity,
infrequently. An effective system for training and
and software arrhythmia analysis have enabled the
retraining should be in place. Adequate numbers of
mass production of relatively cheap, reliable and
staff should be trained to enable achievement of
easily operated portable defibrillators.12—15 Use of
the goal of providing the first shock within 3 min of
AEDs by lay or non-healthcare rescuers is covered
collapse anywhere in the hospital. Hospitals should
in Section 2.
monitor collapse-to-first-shock intervals and resus-
citation outcomes.
Automated rhythm analysis
Automated external defibrillators have micropro-
cessors that analyse several features of the ECG, Strategies before defibrillation
including frequency and amplitude. Some AEDs are
programmed to detect spontaneous movement by Safe use of oxygen during defibrillation
the patient or others. Developing technology should
soon enable AEDs to provide information about In an oxygen-enriched atmosphere, sparking from
frequency and depth of chest compressions dur- poorly applied defibrillator paddles can cause a
ing CPR that may improve BLS performance by all fire.27—32 There are several reports of fires being
rescuers.16,17 caused in this way, and most have resulted in
European Resuscitation Council Guidelines for Resuscitation 2005 S27

significant burns to the patient. The risk of fire dur- Shaving the chest
ing attempted defibrillation can be minimised by
taking the following precautions. Patients with a hairy chest have air trapping
beneath the electrode and poor electrode-to-skin
• Take off any oxygen mask or nasal cannulae and electrical contact. This causes high impedance,
place them at least 1 m away from the patient’s reduced defibrillation efficacy, risk of arcing
chest. (sparks) from electrode to skin and electrode
• Leave the ventilation bag connected to the tra- to electrode and is more likely to cause burns
cheal tube or other airway adjunct. Alterna- to the patient’s chest. Rapid shaving of the
tively, disconnect any bag-valve device from the area of intended electrode placement may be
tracheal tube (or other airway adjunct such as necessary, but do not delay defibrillation if a
the laryngeal mask airway, combitube or laryn- shaver is not immediately available. Shaving the
geal tube), and remove it at least 1 m from the chest per se may reduce transthoracic impedance
patient’s chest during defibrillation. slightly and has been recommended for elective DC
• If the patient is connected to a ventilator, for cardioversion.35
example in the operating room or critical care
unit, leave the ventilator tubing (breathing cir-
Paddle force
cuit) connected to the tracheal tube unless chest
compressions prevent the ventilator from deliv- If using paddles, apply them firmly to the chest
ering adequate tidal volumes. In this case, the wall. This reduces transthoracic impedance by
ventilator is usually substituted for a ventila- improving electrical contact at the electrode—skin
tion bag, which can itself be left connected or interface and reducing thoracic volume.36 The
detached and removed to a distance of at least defibrillator operator should always press firmly
1 m. If the ventilator tubing is disconnected, on handheld electrode paddles, the optimal force
ensure it is kept at least 1 m from the patient being 8 kg in adults37 and 5 kg in children aged
or, better still, switch the ventilator off; mod- 1—8 years when using adult paddles38 ; 8-kg force
ern ventilators generate massive oxygen flows may be attainable only by the strongest mem-
when disconnected. During normal use, when bers of the cardiac arrest team, and therefore it
connected to a tracheal tube, oxygen from a ven- is recommended that these individuals apply the
tilator in the critical care unit will be vented from paddles during defibrillation. Unlike self-adhesive
the main ventilator housing well away from the pads, manual paddles have a bare metal plate that
defibrillation zone. Patients in the critical care requires a conductive material placed between the
unit may be dependent on positive end expiratory metal and patient’s skin to improve electrical con-
pressure (PEEP) to maintain adequate oxygena- tact. Use of bare metal paddles alone creates high
tion; during cardioversion, when the spontaneous transthoracic impedance and is likely to increase
circulation potentially enables blood to remain the risk of arcing and to worsen cutaneous burns
well oxygenated, it is particularly appropriate to from defibrillation.
leave the critically ill patient connected to the
ventilator during shock delivery.
Electrode position
• Minimise the risk of sparks during defibrillation.
Theoretically, self-adhesive defibrillation pads No human studies have evaluated the electrode
are less likely to cause sparks than manual pad- position as a determinant of return of spontaneous
dles. circulation (ROSC) or survival from VF/VT cardiac
arrest. Transmyocardial current during defibrilla-
The technique for electrode contact with tion is likely to be maximal when the electrodes
the chest are placed so that the area of the heart that is fib-
rillating lies directly between them, i.e., ventricles
Optimal defibrillation technique aims to deliver in VF/VT, atria in atrial fibrillation (AF). Therefore,
current across the fibrillating myocardium in the the optimal electrode position may not be the same
presence of minimal transthoracic impedance. for ventricular and atrial arrhythmias.
Transthoracic impedance varies considerably with More patients are presenting with implantable
body mass, but is approximately 70—80  in medical devices (e.g., permanent pacemaker,
adults.33,34 The techniques described below aim to automatic implantable cardioverter defibrillator
place external electrodes (paddles or self-adhesive (AICD)). MedicAlert bracelets are recommended for
pads) in an optimal position using techniques that such patients. These devices may be damaged dur-
minimise transthoracic impedance. ing defibrillation if current is discharged through
S28 C.D. Deakin, J.P. Nolan

electrodes placed directly over the device. Place fibrillation.43 Most,44,45 but not all,46,47 studies
the electrode away from the device or use an alter- have shown that anteroposterior electrode place-
native electrode position as described below. On ment is more effective than the traditional antero-
detecting VF/VT, AICD devices will discharge no apical position in elective cardioversion of atrial
more than six times. Further discharges will occur fibrillation. Efficacy of cardioversion may be less
only if a new episode of VF/VT is detected. Rarely, dependent on electrode position when using bipha-
a faulty device or broken lead may cause repeated sic impedance-compensated waveforms.48 Either
firing; in these circumstances, the patient is likely position is safe and effective for cardioversion of
to be conscious, with the ECG showing a relatively atrial arrhythmias.
normal rate. A magnet placed over the AICD will
disable the defibrillation function in these circum- Respiratory phase
stances. AICD discharge may cause pectoral muscle
contraction, but an attendant touching the patient Transthoracic impedance varies during respiration,
will not receive an electric shock. AICD and pacing being minimal at end expiration. If possible, defib-
function should always be re-evaluated following rillation should be attempted at this phase of
external defibrillation, both to check the device the respiratory cycle. Positive end-expiratory pres-
itself and to check pacing/defibrillation thresholds sure (PEEP) increases transthoracic impedance and
of the device leads. should be minimised during defibrillation. Auto-
Transdermal drug patches may prevent good PEEP (gas trapping) may be particularly high in
electrode contact, causing arcing and burns if the asthmatics and may necessitate higher than usual
electrode is placed directly over the patch during energy levels for defibrillation.49
defibrillation.39,40 Remove medication patches and
wipe the area before applying the electrode. Electrode size
For ventricular arrhythmias, place electrodes
(either pads or paddles) in the conventional The Association for the Advancement of Medi-
sternal—apical position. The right (sternal) elec- cal Instrumentation recommends a minimum elec-
trode is placed to the right of the sternum, below trode size of for individual electrodes and the
the clavicle. The apical paddle is placed in the mid- sum of the electrode areas should be a mini-
axillary line, approximately level with the V6 ECG mum of 150 cm2 .50 Larger electrodes have lower
electrode or female breast. This position should impedance, but excessively large electrodes may
be clear of any breast tissue. It is important that result in less transmyocardial current flow.51 For
this electrode is placed sufficiently laterally. Other adult defibrillation, both handheld paddle elec-
acceptable pad positions include: trodes and self-adhesive pad electrodes 8—12 cm
in diameter are used and function well. Defib-
• each electrode on the lateral chest wall, one
rillation success may be higher with electrodes
on the right and the other on the left side (bi-
of 12-cm diameter compared with those of 8-cm
axillary);
diameter.34,52
• one electrode in the standard apical position and
Standard AEDs are suitable for use in children
the other on the right or left upper back;
over the age of 8 years. In children between 1 and
• one electrode anteriorly, over the left pre-
8 years, use paediatric pads with an attenuator
cordium, and the other electrode posterior to the
to reduce delivered energy, or a paediatric mode,
heart just inferior to the left scapula.
if they are available; if not, use the unmodified
It does not matter which electrode (apex/ machine, taking care to ensure that the adult pads
sternum) is placed in either position. do not overlap. Use of AEDs is not recommended in
Transthoracic impedance has been shown to be children less than 1 year.
minimised when the apical electrode is not placed
over the female breast.41 Asymmetrically shaped Coupling agents
apical electrodes have a lower impedance when
placed longitudinally rather than transversely.42 If using manual paddles, gel pads are preferable
The long axis of the apical paddle should therefore to electrode pastes and gels because the latter
be orientated in a craniocaudal direction. can spread between the two paddles, creating the
Atrial fibrillation is maintained by functional potential for a spark. Do not use bare electrodes
re-entry circuits anchored in the left atrium. As without a coupling material, because this causes
the left atrium is located posteriorly in the tho- high transthoracic impedance and may increase the
rax, an anteroposterior electrode position may be severity of any cutaneous burns. Do not use med-
more efficient for external cardioversion of atrial ical gels or pastes of poor electrical conductivity
European Resuscitation Council Guidelines for Resuscitation 2005 S29

(e.g., ultrasound gel). Electrode pads are preferred with immediate defibrillation. In contrast, a sin-
to electrode gel because they avoid the risk of gle randomised study in adults with out-of-hospital
smearing gel between the two paddles and the sub- VF or VT failed to show improvements in ROSC or
sequent risk of arcing and ineffective defibrillation. survival following 1.5 min of paramedic CPR.80 In
animal studies of VF lasting at least 5 min, CPR
Pads versus paddles before defibrillation improved haemodynamics and
survival.81—83 It may not be possible to extrapo-
Self-adhesive defibrillation pads are safe and effec- late the outcomes achieved by paramedic-provided
tive and are preferable to standard defibrillation CPR, which includes intubation and delivery of 100%
paddles.52 Consideration should be given to use oxygen,79 to those that may be achieved by laypeo-
of self-adhesive pads in peri-arrest situations and ple providing relative poor-quality CPR with mouth-
in clinical situations where patient access is diffi- to-mouth ventilation.
cult. They have a similar transthoracic impedance51 It is reasonable for EMS personnel to give a period
(and therefore efficacy)53,54 to manual paddles, of about 2 min of CPR (i.e., about five cycles at
and enable the operator to defibrillate the patient 30:2) before defibrillation in patients with pro-
from a safe distance rather than leaning over the longed collapse (>5 min). The duration of collapse
patient (as occurs with paddles). When used for ini- is frequently difficult to estimate accurately, and it
tial monitoring of a rhythm, both pads and paddles may be simplest if EMS personnel are instructed to
enable quicker delivery of the first shock compared provide this period of CPR before attempted defib-
with standard ECG electrodes, but pads are quicker rillation in any cardiac arrest they have not wit-
than paddles.55 nessed. Given the relatively weak evidence avail-
When gel pads are used with paddles, the elec- able, individual EMS directors should determine
trolyte gel becomes polarised and thus is a poor whether to implement a CPR-before-defibrillation
conductor after defibrillation. This can cause spu- strategy; inevitably, protocols will vary depending
rious asystole that may persist for 3—4 min when on the local circumstances.
used to monitor the rhythm; a phenomenon not Laypeople and first responders using AEDS should
reported with self-adhesive pads.56,57 When using deliver the shock as soon as possible.
a gel pad/paddle combination, confirm a diagnosis There is no evidence to support or refute CPR
of asystole with independent ECG electrodes rather before defibrillation for in-hospital cardiac arrest.
than the paddles. We recommend shock delivery as soon as possible
following in-hospital cardiac arrest (see Section 4b
Fibrillation waveform analysis and c).
The importance of early uninterrupted external
It is possible to predict, with varying reliability, chest compression is emphasised throughout these
the success of defibrillation from the fibrillation guidelines. In practice, it is often difficult to ascer-
waveform.58—77 If optimal defibrillation waveforms tain the exact time of collapse and, in any case,
and the optimal timing of shock delivery can be CPR should be started as soon as possible. The res-
determined in prospective studies, it should be pos- cuer providing chest compressions should interrupt
sible to prevent the delivery of unsuccessful high- chest compressions only for rhythm analysis and
energy shocks and minimise myocardial injury. This shock delivery, and should be prepared to resume
technology is under active development and inves- chest compressions as soon as a shock is delivered.
tigation. When two rescuers are present, the rescuer operat-
ing the AED should apply the electrodes while CPR
CPR versus defibrillation as the initial is in progress. Interrupt CPR only when it is nec-
treatment essary to assess the rhythm and deliver a shock.
The AED operator should be prepared to deliver a
Although the previous guidelines have recom- shock as soon as analysis is complete and the shock
mended immediate defibrillation for all shockable is advised, ensuring all rescuers are not in contact
rhythms, recent evidence has suggested that a with the victim. The single rescuer should practice
period of CPR before defibrillation may be bene- coordination of CPR with efficient AED operation.
ficial after prolonged collapse. In clinical studies
where response times exceeded 4—5 min, a period
of 1.5—3 min of CPR by paramedics or EMS physi- One-shock versus three-shock sequence
cians before shock delivery improved ROSC, sur-
vival to hospital discharge78,79 and 1-year survival79 There are no published human or animal studies
for adults with out-of-hospital VF or VT, compared comparing a single-shock protocol with a three-
S30 C.D. Deakin, J.P. Nolan

stacked-shock protocol for treatment of VF car-


diac arrest. Animal studies show that relatively
short interruptions in external chest compression
to deliver rescue breaths84,85 or perform rhythm
analysis86 are associated with post-resuscitation
myocardial dysfunction and reduced survival.
Interruptions in external chest compression also
reduce the chances of converting VF to another
rhythm.87 Analysis of CPR performance during out-
of-hospital16,88 and in-hospital17 cardiac arrest has
shown that significant interruptions are common,
with external chest compressions comprising no
more than 51%16 to 76%17 of total CPR time.
In the context of a three-shock protocol being
recommended in the 2000 guidelines, interruptions
in CPR to enable rhythm analysis by AEDs were
Figure 3.1 Monophasic damped sinusoidal waveform
significant. Delays of up to 37 s between delivery
(MDS).
of shocks and recommencing chest compressions
have been reported.89 With first shock efficacy of of repetitive shocks, which in turn limits myocardial
biphasic waveforms exceeding 90%,90—93 failure to damage.95
cardiovert VF successfully is more likely to suggest After a cautious introduction a decade ago,
the need for a period of CPR rather than a fur- defibrillators delivering a shock with a biphasic
ther shock. Thus, immediately after giving a single waveform are now preferred. Monophasic defibril-
shock, and without reassessing the rhythm or feel- lators are no longer manufactured, although many
ing for a pulse, resume CPR (30 compressions:2 ven- remain in use. Monophasic defibrillators deliver cur-
tilations) for 2 min before delivering another shock rent that is unipolar (i.e., one direction of cur-
(if indicated) (see Section 4c). Even if the defibril- rent flow). There are two main types of monopha-
lation attempt is successful in restoring a perfusing sic waveform. The commonest waveform is the
rhythm, it is very rare for a pulse to be palpable monophasic damped sinusoidal (MDS) waveform
immediately after defibrillation, and the delay in (Figure 3.1) which gradually returns to zero current
trying to palpate a pulse will further compromise flow. The monophasic truncated exponential (MTE)
the myocardium if a perfusing rhythm has not been waveform is electronically terminated before cur-
restored.89 In one study of AEDs in out-of-hospital rent flow reaches zero (Figure 3.2). Biphasic defib-
VF cardiac arrest, a pulse was detected in only 2.5% rillators, in contrast, deliver current that flows in
(12/481) of patients with the initial post shock pulse a positive direction for a specified duration before
check, though a pulse was detected sometime after reversing and flowing in a negative direction for the
the initial shock sequence (and before a second remaining milliseconds of the electrical discharge.
shock sequence) in 24.5% (118/481) of patients.93 If There are two main types of biphasic waveform: the
a perfusing rhythm has been restored, giving chest biphasic truncated exponential (BTE) (Figure 3.3)
compressions does not increase the chance of VF and rectilinear biphasic (RLB) (Figure 3.4). Bipha-
recurring.94 In the presence of post-shock asystole sic defibrillators compensate for the wide varia-
chest compressions may induce VF.94 tions in transthoracic impedance by electronically
This single shock strategy is applicable to both
monophasic and biphasic defibrillators.

Waveforms and energy levels

Defibrillation requires the delivery of sufficient


electrical energy to defibrillate a critical mass of
myocardium, abolish the wavefronts of VF and
enable restoration of spontaneous synchronised
electrical activity in the form of an organised
rhythm. The optimal energy for defibrillation is
that which achieves defibrillation while causing the
minimum of myocardial damage.33 Selection of an Figure 3.2 Monophasic truncated exponential wave-
appropriate energy level also reduces the number form (MTE).
European Resuscitation Council Guidelines for Resuscitation 2005 S31

The optimal current for defibrillation using a


monophasic waveform is in the range of 30—40 A.
Indirect evidence from measurements during car-
dioversion for atrial fibrillation suggest that the cur-
rent during defibrillation using biphasic waveforms
is in the range of 15—20 A.100 Future technology
may enable defibrillators to discharge according to
transthoracic current: a strategy that may lead to
greater consistency in shock success. Peak current
amplitude, average current and phase duration all
need to be studied to determine optimal values,
Figure 3.3 Biphasic truncated exponential waveform and manufacturers are encouraged to explore fur-
(BTE). ther this move from energy-based to current-based
defibrillation.
adjusting the waveform magnitude and duration.
The optimal ratio of first-phase to second-phase First shock
duration and leading-edge amplitude has not been
First-shock efficacy for long-duration cardiac arrest
established. Whether different waveforms have dif-
using monophasic defibrillation has been reported
fering efficacy for VF of differing durations is also
as 54%—63% for a 200-J monophasic truncated
unknown.
exponential (MTE) waveform97,101 and 77%—91%
All manual defibrillators and AEDs that allow
using a 200-J monophasic damped sinusoidal (MDS)
manual override of energy levels should be labelled
waveform.96—98,101 Because of the lower efficacy
to indicate their waveform (monophasic or bipha-
of this waveform, the recommended initial energy
sic) and recommended energy levels for attempted
level for the first shock using a monophasic defib-
defibrillation of VF/VT. First-shock efficacy for
rillator is 360 J. Although higher energy levels risk
long-duration VF/VT is greater with biphasic than
a greater degree of myocardial injury, the bene-
monophasic waveforms,96—98 and therefore use of
fits of earlier conversion to a perfusing rhythm are
the former is recommended whenever possible.
paramount. Atrioventricular block is more common
Optimal energy levels for both monophasic and
with higher monophasic energy levels, but is gen-
biphasic waveforms are unknown. The recommen-
erally transient and has been shown not to affect
dations for energy levels are based on a consensus
survival to hospital discharge.102 Only 1 of 27 animal
following careful review of the current literature.
studies demonstrated harm caused by attempted
Although energy levels are selected for defibril-
defibrillation using high-energy shocks.103
lation, it is the transmyocardial current flow that
There is no evidence that one biphasic wave-
achieves defibrillation. Current correlates well with
form or device is more effective than another. First-
the successful defibrillation and cardioversion.99
shock efficacy of the BTE waveform using 150—200 J
has been reported as 86%—98%.96,97,101,104,105 First-
shock efficacy of the RLB waveform using 120 J is up
to 85% (data not published in the paper but sup-
plied by personnel communication).98 The initial
biphasic shock should be no lower than 120 J for
RLB waveforms and 150 J for BTE waveforms. Ide-
ally, the initial biphasic shock energy should be at
least 150 J for all waveforms.
Manufacturers should display the effective wave-
form dose range on the face of the biphasic device.
If the provider is unaware of the effective dose
range of the device, use a dose of 200 J for the
first shock. This 200 J default energy has been cho-
sen because it falls within the reported range of
selected doses that are effective for first and subse-
quent biphasic shocks and can be provided by every
biphasic manual defibrillator available today. It is
a consensus default dose and not a recommended
Figure 3.4 Rectilinear biphasic waveform (RLB). ideal dose. If biphasic devices are clearly labelled
S32 C.D. Deakin, J.P. Nolan

and providers are familiar with the devices they use Blind defibrillation
in clinical care, there will be no need for the default
200 J dose. Ongoing research is necessary to firmly Delivery of shocks without a monitor or an ECG
establish the most appropriate initial settings for rhythm diagnosis is referred to as ‘‘blind’’ defibril-
both monophasic and biphasic defibrillators. lation. Blind defibrillation is unnecessary. Handheld
paddles with ‘‘quick-look’’ monitoring capabilities
Second and subsequent shocks on modern manually operated defibrillators are
widely available. AEDs use reliable and proven deci-
With monophasic defibrillators, if the initial shock sion algorithms to identify VF.
has been unsuccessful at 360 J, second and sub-
sequent shocks should all be delivered at 360 J. Spurious asystole and occult ventricular
With biphasic defibrillators there is no evidence to fibrillation
support either a fixed or escalating energy proto-
col. Both strategies are acceptable; however, if the Rarely, coarse VF can be present in some leads, with
first shock is not successful and the defibrillator is very small undulations seen in the orthogonal leads,
capable of delivering shocks of higher energy, it which is called occult VF. A flat line that may resem-
is rational to increase the energy for subsequent ble asystole is displayed; examine the rhythm in
shocks. If the provider is unaware of the effective two leads to obtain the correct diagnosis. Of more
dose range of the biphasic device and has used the importance, one study noted that spurious asystole,
default 200 J dose for the first shock, use either a flat line produced by technical errors (e.g., no
an equal or higher dose for second or subsequent power, leads unconnected, gain set to low, incor-
shocks, depending on the capabilities of the device. rect lead selection, or polarisation of electrolyte
If a shockable rhythm (recurrent ventricular fib- gel (see above)), was far more frequent than occult
rillation) recurs after successful defibrillation (with VF.120
or without ROSC), give the next shock with the There is no evidence that attempting to defib-
energy level that had previously been successful. rillate true asystole is beneficial. Studies in
children121 and adults122 have failed to show bene-
fit from defibrillation of asystole. On the contrary,
Other related defibrillation topics repeated shocks will cause myocardial injury.

Defibrillation of children Precordial thump


Cardiac arrest is less common in children. Aetiology There are no prospective studies that evaluate
is generally related to hypoxia and trauma.106—108 use of precordial (chest) thump. The rationale for
VF is relatively rare compared with adult cardiac giving a thump is that the mechanical energy of
arrest, occurring in 7%—15% of paediatric and ado- the thump is converted to electrical energy, which
lescent arrests.108—112 Common causes of VF in may be sufficient to achieve cardioversion.123
children include trauma, congenital heart disease, The electrical threshold of successful defibrillation
long QT interval, drug overdose and hypothermia. increases rapidly after the onset of the arrhythmia,
Rapid defibrillation of these patients may improve and the amount of electrical energy generated falls
outcome.112,113 below this threshold within seconds. A precordial
The optimal energy level, waveform and shock thump is most likely to be successful in convert-
sequence are unknown but, as with adults, bipha- ing VT to sinus rhythm. Successful treatment of
sic shocks appear to be at least as effective as, VF by precordial thump is much less likely: in all
and less harmful than, monophasic shocks.114—116 the reported successful cases, the precordial thump
The upper limit for safe defibrillation is unknown, was given within the first 10 s of VF.123 Although
but doses in excess of the previously recommended three case series124—126 reported that VF or pulse-
maximum of 4 J kg−1 (as high as 9 J kg−1 ) have less VT was converted to a perfusing rhythm by
defibrillated children effectively without signifi- a precordial thump, there are occasional reports
cant adverse effects.20,117,118 The recommended of thump causing deterioration in cardiac rhythm,
energy level for manual monophasic defibrillation such as rate acceleration of VT, conversion of VT
is 4 J kg−1 for the initial shock and for subsequent into VF, complete heart block or asystole.125,127—132
shocks. The same energy level is recommended for Consider giving a single precordial thump when
manual biphasic defibrillation.119 As with adults, if cardiac arrest is confirmed rapidly after a wit-
a shockable rhythm recurs, use the energy level for nessed, sudden collapse and a defibrillator is not
defibrillation that had previously been successful. immediately to hand. These circumstances are
European Resuscitation Council Guidelines for Resuscitation 2005 S33

most likely to occur when the patient is monitored. randomised study comparing escalating monophasic
Precordial thump should be undertaken immedi- energy levels to 360 J and biphasic energy levels to
ately after confirmation of cardiac arrest and only 200 J found no difference in efficacy between the
by healthcare professionals trained in the tech- two waveforms.137 An initial shock of 120—150 J,
nique. Using the ulnar edge of a tightly clenched escalating if necessary, is a reasonable strategy
fist, a sharp impact is delivered to the lower half of based on current data.
the sternum from a height of about 20 cm, followed
by immediate retraction of the fist, which creates Atrial flutter and paroxysmal
an impulse-like stimulus. supraventricular tachycardia

Atrial flutter and paroxysmal SVT generally


Cardioversion require less energy than atrial fibrillation for
cardioversion.138 Give an initial shock of 100-J
If electrical cardioversion is used to convert atrial monophasic or 70—120 J biphasic waveform. Give
or ventricular tachyarrhythmias, the shock must be subsequent shocks using stepwise increases in
synchronised to occur with the R wave of the elec- energy.99
trocardiogram rather than with the T wave: VF can
be induced if a shock is delivered during the rel- Ventricular tachycardia
ative refractory portion of the cardiac cycle.133
Synchronisation can be difficult in VT because of The energy required for cardioversion of VT
the wide-complex and variable forms of ventricular depends on the morphological characteristics and
arrhythmia. If synchronisation fails, give unsynchro- rate of the arrhythmia.139 Ventricular tachycardia
nised shocks to the unstable patient in VT to avoid with a pulse responds well to cardioversion using
prolonged delay in restoring sinus rhythm. Ventric- initial monophasic energies of 200 J. Use biphasic
ular fibrillation or pulseless VT requires unsynchro- energy levels of 120—150 J for the initial shock.
nised shocks. Conscious patients must be anaes- Give stepwise increases if the first shock fails to
thetised or sedated before attempting synchronised achieve sinus rhythm.139
cardioversion.
Pacing
Atrial fibrillation
Consider pacing in patients with symptomatic
Biphasic waveforms are more effective than bradycardia refractory to anticholinergic drugs or
monophasic waveforms for cardioversion of other second-line therapy (see Section 4f). Immedi-
AF100,134,135 ; when available, use a biphasic defib- ate pacing is indicated, especially when the block is
rillator in preference to a monophasic defibrillator. at or below the His—Purkinje level. If transthoracic
pacing is ineffective, consider transvenous pacing.
Monophasic waveforms Whenever a diagnosis of asystole is made, check the
ECG carefully for the presence of P waves, because
A study of electrical cardioversion for atrial fib-
this may respond to cardiac pacing. Do not attempt
rillation indicated that 360-J MDS shocks were
pacing for asystole; it does not increase short-term
more effective than 100-J or 200-J MDS shocks.136
or long-term survival in or out of hospital.140—148
Although a first shock of 360-J reduces overall
energy requirements for cardioversion, 360 J may
cause greater myocardial damage than occurs with References
lower monophasic energy levels, and this must be
taken into consideration. Commence cardioversion 1. American Heart Association in collaboration with Inter-
of atrial fibrillation using an initial energy level of national Liaison Committee on Resuscitation. Guide-
lines 2000 for Cardiopulmonary Resuscitation and Emer-
200 J, increasing in a stepwise manner as necessary. gency Cardiovascular Care, Part 6: Advanced Cardiovas-
cular Life Support: Section 2: Defibrillation. Circulation
Biphasic waveforms 2000;102(Suppl.):I90—4.
2. Larsen MP, Eisenberg MS, Cummins RO, Hallstrom AP.
More data are needed before specific recommen- Predicting survival from out-of-hospital cardiac arrest: a
dations can be made for optimal biphasic energy graphic model. Ann Emerg Med 1993;22:1652—8.
3. Valenzuela TD, Roe DJ, Cretin S, Spaite DW, Larsen
levels. First-shock efficacy of a 70-J biphasic wave- MP. Estimating effectiveness of cardiac arrest interven-
form has been shown to be significantly greater than tions: a logistic regression survival model. Circulation
that with a 100-monophasic waveform.100,134,135 A 1997;96:3308—13.
S34 C.D. Deakin, J.P. Nolan

4. Waalewijn RA, de Vos R, Tijssen JGP, Koster RW. Survival ysis in infants and children. Ann Emerg Med 2003;42:185—
models for out-of-hospital cardiopulmonary resuscitation 96.
from the perspectives of the bystander, the first responder, 21. Cecchin F, Jorgenson DB, Berul CI, et al. Is arrhythmia
and the paramedic. Resuscitation 2001;51:113—22. detection by automatic external defibrillator accurate for
5. Myerburg RJ, Fenster J, Velez M, et al. Impact of children? Sensitivity and specificity of an automatic exter-
community-wide police car deployment of automated nal defibrillator algorithm in 696 pediatric arrhythmias.
external defibrillators on survival from out-of-hospital car- Circulation 2001;103:2483—8.
diac arrest. Circulation 2002;106:1058—64. 22. Zafari AM, Zarter SK, Heggen V, et al. A program encour-
6. Capucci A, Aschieri D, Piepoli MF, Bardy GH, Iconomu aging early defibrillation results in improved in-hospital
E, Arvedi M. Tripling survival from sudden cardiac resuscitation efficacy. J Am Coll Cardiol 2004;44:846—52.
arrest via early defibrillation without traditional edu- 23. Destro A, Marzaloni M, Sermasi S, Rossi F. Automatic exter-
cation in cardiopulmonary resuscitation. Circulation nal defibrillators in the hospital as well? Resuscitation
2002;106:1065—70. 1996;31:39—43.
7. van Alem AP, Vrenken RH, de Vos R, Tijssen JG, Koster RW. 24. Domanovits H, Meron G, Sterz F, et al. Successful automatic
Use of automated external defibrillator by first responders external defibrillator operation by people trained only in
in out of hospital cardiac arrest: prospective controlled basic life support in a simulated cardiac arrest situation.
trial. BMJ 2003;327:1312. Resuscitation 1998;39:47—50.
8. Valenzuela TD, Bjerke HS, Clark LL, et al. Rapid defibrilla- 25. Cusnir H, Tongia R, Sheka KP, et al. In hospital cardiac
tion by nontraditional responders: the Casino Project. Acad arrest: a role for automatic defibrillation. Resuscitation
Emerg Med 1998;5:414—5. 2004;63:183—8.
9. Swor RA, Jackson RE, Cynar M, et al. Bystander CPR, 26. Kaye W, Mancini ME, Richards N. Organizing and imple-
ventricular fibrillation, and survival in witnessed, unmon- menting a hospital-wide first-responder automated exter-
itored out-of-hospital cardiac arrest. Ann Emerg Med nal defibrillation program: strengthening the in-hospital
1995;25:780—4. chain of survival. Resuscitation 1995;30:151—6.
10. Holmberg M, Holmberg S, Herlitz J. Effect of bystander 27. Miller PH. Potential fire hazard in defibrillation. JAMA
cardiopulmonary resuscitation in out-of-hospital cardiac 1972;221:192.
arrest patients in Sweden. Resuscitation 2000;47:59—70. 28. Hummel IIIrd RS, Ornato JP, Weinberg SM, Clarke AM. Spark-
11. Monsieurs KG, Handley AJ, Bossaert LL. European Resus- generating properties of electrode gels used during defib-
citation Council Guidelines 2000 for Automated External rillation. A potential fire hazard. JAMA 1988;260:3021—4.
Defibrillation. A statement from the Basic Life Support and 29. Fires from defibrillation during oxygen administration.
Automated External Defibrillation Working Group (1) and Health Devices 1994;23:307—9.
approved by the Executive Committee of the European 30. Lefever J, Smith A. Risk of fire when using defibrillation in
Resuscitation Council. Resuscitation 2001;48:207—9. an oxygen enriched atmosphere. Medical Devices Agency
12. Cummins RO, Eisenberg M, Bergner L, Murray JA. Sensitivity Safety Notices 1995;3:1—3.
accuracy, and safety of an automatic external defibrillator. 31. Ward ME. Risk of fires when using defibrillators in an oxygen
Lancet 1984;2:318—20. enriched atmosphere. Resuscitation 1996;31:173.
13. Davis EA, Mosesso Jr VN. Performance of police first 32. Theodorou AA, Gutierrez JA, Berg RA. Fire attributable
responders in utilizing automated external defibrillation to a defibrillation attempt in a neonate. Pediatrics
on victims of sudden cardiac arrest. Prehosp Emerg Care 2003;112:677—9.
1998;2:101—7. 33. Kerber RE, Kouba C, Martins J, et al. Advance prediction of
14. White RD, Vukov LF, Bugliosi TF. Early defibrillation by transthoracic impedance in human defibrillation and car-
police: initial experience with measurement of critical dioversion: importance of impedance in determining the
time intervals and patient outcome. Ann Emerg Med success of low-energy shocks. Circulation 1984;70:303—8.
1994;23:1009—13. 34. Kerber RE, Grayzel J, Hoyt R, Marcus M, Kennedy J.
15. White RD, Hankins DG, Bugliosi TF. Seven years’ experi- Transthoracic resistance in human defibrillation. Influence
ence with early defibrillation by police and paramedics of body weight, chest size, serial shocks, paddle size
in an emergency medical services system. Resuscitation and paddle contact pressure. Circulation 1981;63:676—
1998;39:145—51. 82.
16. Wik L, Kramer-Johansen J, Myklebust H, et al. Quality of 35. Sado DM, Deakin CD, Petley GW, Clewlow F. Comparison
cardiopulmonary resuscitation during out-of-hospital car- of the effects of removal of chest hair with not doing so
diac arrest. JAMA 2005;293:299—304. before external defibrillation on transthoracic impedance.
17. Abella BS, Alvarado JP, Myklebust H, et al. Quality of Am J Cardiol 2004;93:98—100.
cardiopulmonary resuscitation during in-hospital cardiac 36. Deakin CD, Sado DM, Petley GW, Clewlow F. Differen-
arrest. JAMA 2005;293:305—10. tial contribution of skin impedance and thoracic volume
18. Kerber RE, Becker LB, Bourland JD, et al. Automatic exter- to transthoracic impedance during external defibrillation.
nal defibrillators for public access defibrillation: recom- Resuscitation 2004;60:171—4.
mendations for specifying and reporting arrhythmia analy- 37. Deakin C, Sado D, Petley G, Clewlow F. Determining the
sis algorithm performance, incorporating new waveforms, optimal paddle force for external defibrillation. Am J Car-
and enhancing safety. A statement for health professionals diol 2002;90:812—3.
from the American Heart Association Task Force on Auto- 38. Deakin C, Bennetts S, Petley G, Clewlow F. What is the opti-
matic External Defibrillation, Subcommittee on AED Safety mal paddle force for paediatric defibrillation? Resuscitation
and Efficacy. Circulation 1997;95:1677—82. 2002;55:59.
19. Dickey W, Dalzell GW, Anderson JM, Adgey AA. The accu- 39. Panacek EA, Munger MA, Rutherford WF, Gardner SF. Report
racy of decision-making of a semi-automatic defibrillator of nitropatch explosions complicating defibrillation. Am J
during cardiac arrest. Eur Heart J 1992;13:608—15. Emerg Med 1992;10:128—9.
20. Atkinson E, Mikysa B, Conway JA, et al. Specificity and 40. Wrenn K. The hazards of defibrillation through nitroglycerin
sensitivity of automated external defibrillator rhythm anal- patches. Ann Emerg Med 1990;19:1327—8.
European Resuscitation Council Guidelines for Resuscitation 2005 S35

41. Pagan-Carlo LA, Spencer KT, Robertson CE, Dengler A, Bir- and nonparametric classification of ventricular fibrillation
kett C, Kerber RE. Transthoracic defibrillation: importance in patients with out-of-hospital cardiac arrest. Circulation
of avoiding electrode placement directly on the female 2000;102:1523—9.
breast. J Am Coll Cardiol 1996;27:449—52. 60. Eftestol T, Wik L, Sunde K, Steen PA. Effects of cardiopul-
42. Deakin CD, Sado DM, Petley GW, Clewlow F. Is the ori- monary resuscitation on predictors of ventricular fibrilla-
entation of the apical defibrillation paddle of impor- tion defibrillation success during out-of-hospital cardiac
tance during manual external defibrillation? Resuscitation arrest. Circulation 2004;110:10—5.
2003;56:15—8. 61. Weaver WD, Cobb LA, Dennis D, Ray R, Hallstrom AP, Copass
43. Kirchhof P, Borggrefe M, Breithardt G. Effect of electrode MK. Amplitude of ventricular fibrillation waveform and out-
position on the outcome of cardioversion. Card Electro- come after cardiac arrest. Ann Intern Med 1985;102:53—5.
physiol Rev 2003;7:292—6. 62. Brown CG, Dzwonczyk R. Signal analysis of the human elec-
44. Kirchhof P, Eckardt L, Loh P, et al. Anterior-posterior ver- trocardiogram during ventricular fibrillation: frequency
sus anterior-lateral electrode positions for external car- and amplitude parameters as predictors of successful coun-
dioversion of atrial fibrillation: a randomised trial. Lancet tershock. Ann Emerg Med 1996;27:184—8.
2002;360:1275—9. 63. Callaham M, Braun O, Valentine W, Clark DM, Zegans C. Pre-
45. Botto GL, Politi A, Bonini W, Broffoni T, Bonatti R. Exter- hospital cardiac arrest treated by urban first-responders:
nal cardioversion of atrial fibrillation: role of paddle posi- profile of patient response and prediction of outcome
tion on technical efficacy and energy requirements. Heart by ventricular fibrillation waveform. Ann Emerg Med
1999;82:726—30. 1993;22:1664—77.
46. Alp NJ, Rahman S, Bell JA, Shahi M. Randomised comparison 64. Strohmenger HU, Lindner KH, Brown CG. Analysis of
of antero-lateral versus antero-posterior paddle positions the ventricular fibrillation ECG signal amplitude and fre-
for DC cardioversion of persistent atrial fibrillation. Int J quency parameters as predictors of countershock success
Cardiol 2000;75:211—6. in humans. Chest 1997;111:584—9.
47. Mathew TP, Moore A, McIntyre M, et al. Randomised com- 65. Strohmenger HU, Eftestol T, Sunde K, et al. The predictive
parison of electrode positions for cardioversion of atrial value of ventricular fibrillation electrocardiogram signal
fibrillation. Heart 1999;81:576—9. frequency and amplitude variables in patients with out-
48. Walsh SJ, McCarty D, McClelland AJ, et al. Impedance com- of-hospital cardiac arrest. Anesth Analg 2001;93:1428—33.
pensated biphasic waveforms for transthoracic cardiover- 66. Podbregar M, Kovacic M, Podbregar-Mars A, Brezocnik M.
sion of atrial fibrillation: a multi-centre comparison of Predicting defibrillation success by ‘genetic’ programming
antero-apical and antero-posterior pad positions. Eur Heart in patients with out-of-hospital cardiac arrest. Resuscita-
J 2005. tion 2003;57:153—9.
49. Deakin CD, McLaren RM, Petley GW, Clewlow F, Dalrymple- 67. Menegazzi JJ, Callaway CW, Sherman LD, et al. Ventricular
Hay MJ. Effects of positive end-expiratory pressure on fibrillation scaling exponent can guide timing of defibrilla-
transthoracic impedance—–implications for defibrillation. tion and other therapies. Circulation 2004;109:926—31.
Resuscitation 1998;37:9—12. 68. Povoas HP, Weil MH, Tang W, Bisera J, Klouche K, Barbatsis
50. American National Standard: Automatic External Defibril- A. Predicting the success of defibrillation by electrocardio-
lators and Remote Controlled Defibrillators (DF39). Arling- graphic analysis. Resuscitation 2002;53:77—82.
ton, Virgina: Association for the Advancement of Medical 69. Noc M, Weil MH, Tang W, Sun S, Pernat A, Bisera J. Electro-
Instrumentation; 1993. cardiographic prediction of the success of cardiac resusci-
51. Deakin CD, McLaren RM, Petley GW, Clewlow F, Dalrymple- tation. Crit Care Med 1999;27:708—14.
Hay MJ. A comparison of transthoracic impedance using 70. Strohmenger HU, Lindner KH, Keller A, Lindner IM, Pfen-
standard defibrillation paddles and self-adhesive defibril- ninger EG. Spectral analysis of ventricular fibrillation and
lation pads. Resuscitation 1998;39:43—6. closed-chest cardiopulmonary resuscitation. Resuscitation
52. Stults KR, Brown DD, Cooley F, Kerber RE. Self-adhesive 1996;33:155—61.
monitor/defibrillation pads improve prehospital defibrilla- 71. Noc M, Weil MH, Gazmuri RJ, Sun S, Biscera J, Tang W.
tion success. Ann Emerg Med 1987;16:872—7. Ventricular fibrillation voltage as a monitor of the effec-
53. Kerber RE, Martins JB, Kelly KJ, et al. Self-adhesive preap- tiveness of cardiopulmonary resuscitation. J Lab Clin Med
plied electrode pads for defibrillation and cardioversion. J 1994;124:421—6.
Am Coll Cardiol 1984;3:815—20. 72. Lightfoot CB, Nremt P, Callaway CW, et al. Dynamic nature
54. Kerber RE, Martins JB, Ferguson DW, et al. Exper- of electrocardiographic waveform predicts rescue shock
imental evaluation and initial clinical application of outcome in porcine ventricular fibrillation. Ann Emerg Med
new self-adhesive defibrillation electrodes. Int J Cardiol 2003;42:230—41.
1985;8:57—66. 73. Marn-Pernat A, Weil MH, Tang W, Pernat A, Bisera J. Opti-
55. Perkins GD, Roberts C, Gao F. Delays in defibrillation: influ- mizing timing of ventricular defibrillation. Crit Care Med
ence of different monitoring techniques. Br J Anaesth 2001;29:2360—5.
2002;89:405—8. 74. Hamprecht FA, Achleitner U, Krismer AC, et al. Fibrilla-
56. Bradbury N, Hyde D, Nolan J. Reliability of ECG monitor- tion power, an alternative method of ECG spectral analysis
ing with a gel pad/paddle combination after defibrillation. for prediction of countershock success in a porcine model
Resuscitation 2000;44:203—6. of ventricular fibrillation. Resuscitation 2001;50:287—
57. Chamberlain D. Gel pads should not be used for monitoring 96.
ECG after defibrillation. Resuscitation 2000;43:159—60. 75. Amann A, Achleitner U, Antretter H, et al. Analysing ven-
58. Callaway CW, Sherman LD, Mosesso Jr VN, Dietrich TJ, Holt tricular fibrillation ECG-signals and predicting defibrillation
E, Clarkson MC. Scaling exponent predicts defibrillation success during cardiopulmonary resuscitation employing
success for out-of-hospital ventricular fibrillation cardiac N(alpha)-histograms. Resuscitation 2001;50:77—85.
arrest. Circulation 2001;103:1656—61. 76. Brown CG, Griffith RF, Van Ligten P, et al. Median
59. Eftestol T, Sunde K, Aase SO, Husoy JH, Steen PA. Predict- frequency—–a new parameter for predicting defibrillation
ing outcome of defibrillation by spectral characterization success rate. Ann Emerg Med 1991;20:787—9.
S36 C.D. Deakin, J.P. Nolan

77. Amann A, Rheinberger K, Achleitner U, et al. The predic- 93. Rea TD, Shah S, Kudenchuk PJ, Copass MK, Cobb LA. Auto-
tion of defibrillation outcome using a new combination of mated external defibrillators: to what extent does the algo-
mean frequency and amplitude in porcine models of car- rithm delay CPR? Ann Emerg Med 2005;46:132—41.
diac arrest. Anesth Analg 2002;95:716—22. 94. Hess EP, White RD. Ventricular fibrillation is not pro-
78. Cobb LA, Fahrenbruch CE, Walsh TR, et al. Influence voked by chest compression during post-shock organized
of cardiopulmonary resuscitation prior to defibrillation in rhythms in out-of-hospital cardiac arrest. Resuscitation
patients with out-of-hospital ventricular fibrillation. JAMA 2005;66:7—11.
1999;281:1182—8. 95. Joglar JA, Kessler DJ, Welch PJ, et al. Effects of repeated
79. Wik L, Hansen TB, Fylling F, et al. Delaying defibrillation to electrical defibrillations on cardiac troponin I levels. Am J
give basic cardiopulmonary resuscitation to patients with Cardiol 1999;83:270—2. A6.
out-of-hospital ventricular fibrillation: a randomized trial. 96. van Alem AP, Chapman FW, Lank P, Hart AA, Koster
JAMA 2003;289:1389—95. RW. A prospective, randomised and blinded compari-
80. Jacobs IG, Finn JC, Oxer HF, Jelinek GA. CPR before defibril- son of first shock success of monophasic and biphasic
lation in out-of-hospital cardiac arrest: a randomized trial. waveforms in out-of-hospital cardiac arrest. Resuscitation
Emerg Med Australas 2005;17:39—45. 2003;58:17—24.
81. Berg RA, Hilwig RW, Kern KB, Ewy GA. Precountershock 97. Carpenter J, Rea TD, Murray JA, Kudenchuk PJ, Eisenberg
cardiopulmonary resuscitation improves ventricular fibril- MS. Defibrillation waveform and post-shock rhythm in out-
lation median frequency and myocardial readiness for suc- of-hospital ventricular fibrillation cardiac arrest. Resusci-
cessful defibrillation from prolonged ventricular fibrilla- tation 2003;59:189—96.
tion: a randomized, controlled swine study. Ann Emerg Med 98. Morrison LJ, Dorian P, Long J, et al. Out-of-hospital car-
2002;40:563—70. diac arrest rectilinear biphasic to monophasic damped sine
82. Berg RA, Hilwig RW, Ewy GA, Kern KB. Precountershock defibrillation waveforms with advanced life support inter-
cardiopulmonary resuscitation improves initial response vention trial (ORBIT). Resuscitation 2005;66:149—57.
to defibrillation from prolonged ventricular fibrillation: 99. Kerber RE, Martins JB, Kienzle MG, et al. Energy, current,
a randomized, controlled swine study. Crit Care Med and success in defibrillation and cardioversion: clinical
2004;32:1352—7. studies using an automated impedance-based method of
83. Kolarova J, Ayoub IM, Yi Z, Gazmuri RJ. Optimal timing for energy adjustment. Circulation 1988;77:1038—46.
electrical defibrillation after prolonged untreated ventric- 100. Koster RW, Dorian P, Chapman FW, Schmitt PW, O’Grady
ular fibrillation. Crit Care Med 2003;31:2022—8. SG, Walker RG. A randomized trial comparing monophasic
84. Berg RA, Sanders AB, Kern KB, et al. Adverse hemo- and biphasic waveform shocks for external cardioversion of
dynamic effects of interrupting chest compressions for atrial fibrillation. Am Heart J 2004;147:e20.
rescue breathing during cardiopulmonary resuscitation 101. Martens PR, Russell JK, Wolcke B, et al. Optimal Response
for ventricular fibrillation cardiac arrest. Circulation to Cardiac Arrest study: defibrillation waveform effects.
2001;104:2465—70. Resuscitation 2001;49:233—43.
85. Kern KB, Hilwig RW, Berg RA, Sanders AB, Ewy GA. 102. Weaver WD, Cobb LA, Copass MK, Hallstrom AP. Ventricu-
Importance of continuous chest compressions during lar defibrillation: a comparative trial using 175-J and 320-J
cardiopulmonary resuscitation: improved outcome dur- shocks. N Engl J Med 1982;307:1101—6.
ing a simulated single lay-rescuer scenario. Circulation 103. Tang W, Weil MH, Sun S, et al. The effects of
2002;105:645—9. biphasic and conventional monophasic defibrillation on
86. Yu T, Weil MH, Tang W, et al. Adverse outcomes of inter- postresuscitation myocardial function. J Am Coll Cardiol
rupted precordial compression during automated defibril- 1999;34:815—22.
lation. Circulation 2002;106:368—72. 104. Gliner BE, Jorgenson DB, Poole JE, et al. Treatment of out-
87. Eftestol T, Sunde K, Steen PA. Effects of interrupting pre- of-hospital cardiac arrest with a low-energy impedance-
cordial compressions on the calculated probability of defib- compensating biphasic waveform automatic external defib-
rillation success during out-of-hospital cardiac arrest. Cir- rillator. The LIFE Investigators. Biomed Instrum Technol
culation 2002;105:2270—3. 1998;32:631—44.
88. Valenzuela TD, Kern KB, Clark LL, et al. Interruptions 105. White RD, Blackwell TH, Russell JK, Snyder DE, Jorgenson
of chest compressions during emergency medical systems DB. Transthoracic impedance does not affect defibrillation,
resuscitation. Circulation 2005;112:1259—65. resuscitation or survival in patients with out-of-hospital
89. van Alem AP, Sanou BT, Koster RW. Interruption of car- cardiac arrest treated with a non-escalating biphasic wave-
diopulmonary resuscitation with the use of the automated form defibrillator. Resuscitation 2005;64:63—9.
external defibrillator in out-of-hospital cardiac arrest. Ann 106. Kuisma M, Suominen P, Korpela R. Paediatric out-of-hospital
Emerg Med 2003;42:449—57. cardiac arrests: epidemiology and outcome. Resuscitation
90. Bain AC, Swerdlow CD, Love CJ, et al. Multicenter study 1995;30:141—50.
of principles-based waveforms for external defibrillation. 107. Sirbaugh PE, Pepe PE, Shook JE, et al. A prospective,
Ann Emerg Med 2001;37:5—12. population-based study of the demographics, epidemi-
91. Poole JE, White RD, Kanz KG, et al. Low-energy impedance- ology, management, and outcome of out-of-hospital
compensating biphasic waveforms terminate ventricular pediatric cardiopulmonary arrest. Ann Emerg Med
fibrillation at high rates in victims of out-of-hospital car- 1999;33:174—84.
diac arrest. LIFE Investigators. J Cardiovasc Electrophysiol 108. Hickey RW, Cohen DM, Strausbaugh S, Dietrich AM. Pedi-
1997;8:1373—85. atric patients requiring CPR in the prehospital setting. Ann
92. Schneider T, Martens PR, Paschen H, et al. Multicenter, Emerg Med 1995;25:495—501.
randomized, controlled trial of 150-J biphasic shocks com- 109. Appleton GO, Cummins RO, Larson MP, Graves JR. CPR and
pared with 200- to 360-J monophasic shocks in the resusci- the single rescuer: at what age should you ‘‘call first’’
tation of out-of-hospital cardiac arrest victims. Optimized rather than ‘‘call fast’’? Ann Emerg Med 1995;25:492—4.
Response to Cardiac Arrest (ORCA) Investigators. Circula- 110. Ronco R, King W, Donley DK, Tilden SJ. Outcome and cost
tion 2000;102:1780—7. at a children’s hospital following resuscitation for out-of-
European Resuscitation Council Guidelines for Resuscitation 2005 S37

hospital cardiopulmonary arrest. Arch Pediatr Adolesc Med 129. Gertsch M, Hottinger S, Hess T. Serial chest thumps for the
1995;149:210—4. treatment of ventricular tachycardia in patients with coro-
111. Losek JD, Hennes H, Glaeser P, Hendley G, Nelson DB. nary artery disease. Clin Cardiol 1992;15:181—8.
Prehospital care of the pulseless, nonbreathing pediatric 130. Krijne R. Rate acceleration of ventricular tachycardia after
patient. Am J Emerg Med 1987;5:370—4. a precordial chest thump. Am J Cardiol 1984;53:964—5.
112. Mogayzel C, Quan L, Graves JR, Tiedeman D, Fahrenbruch 131. Sclarovsky S, Kracoff OH, Agmon J. Acceleration of ven-
C, Herndon P. Out-of-hospital ventricular fibrillation in chil- tricular tachycardia induced by a chest thump. Chest
dren and adolescents: causes and outcomes. Ann Emerg 1981;80:596—9.
Med 1995;25:484—91. 132. Yakaitis RW, Redding JS. Precordial thumping during cardiac
113. Safranek DJ, Eisenberg MS, Larsen MP. The epidemiol- resuscitation. Crit Care Med 1973;1:22—6.
ogy of cardiac arrest in young adults. Ann Emerg Med 133. Lown B. Electrical reversion of cardiac arrhythmias. Br
1992;21:1102—6. Heart J 1967;29:469—89.
114. Berg RA, Chapman FW, Berg MD, et al. Attenuated adult 134. Mittal S, Ayati S, Stein KM, et al. Transthoracic cardiover-
biphasic shocks compared with weight-based monophasic sion of atrial fibrillation: comparison of rectilinear biphasic
shocks in a swine model of prolonged pediatric ventricular versus damped sine wave monophasic shocks. Circulation
fibrillation. Resuscitation 2004;61:189—97. 2000;101:1282—7.
115. Tang W, Weil MH, Jorgenson D, et al. Fixed-energy bipha- 135. Page RL, Kerber RE, Russell JK, et al. Biphasic versus
sic waveform defibrillation in a pediatric model of cardiac monophasic shock waveform for conversion of atrial fibril-
arrest and resuscitation. Crit Care Med 2002;30:2736—41. lation: the results of an international randomized, double-
116. Clark CB, Zhang Y, Davies LR, Karlsson G, Kerber RE. Pedi- blind multicenter trial. J Am Coll Cardiol 2002;39:1956—63.
atric transthoracic defibrillation: biphasic versus monopha- 136. Joglar JA, Hamdan MH, Ramaswamy K, et al. Initial energy
sic waveforms in an experimental model. Resuscitation for elective external cardioversion of persistent atrial fib-
2001;51:159—63. rillation. Am J Cardiol 2000;86:348—50.
117. Gurnett CA, Atkins DL. Successful use of a biphasic wave- 137. Alatawi F, Gurevitz O, White R. Prospective, randomized
form automated external defibrillator in a high-risk child. comparison of two biphasic waveforms for the efficacy and
Am J Cardiol 2000;86:1051—3. safety of transthoracic biphasic cardioversion of atrial fib-
118. Atkins DL, Jorgenson DB. Attenuated pediatric electrode rillation. Heart Rhythm 2005;2:382—7.
pads for automated external defibrillator use in children. 138. Pinski SL, Sgarbossa EB, Ching E, Trohman RG. A comparison
Resuscitation 2005;66:31—7. of 50-J versus 100-J shocks for direct-current cardioversion
119. Gutgesell HP, Tacker WA, Geddes LA, Davis S, Lie JT, McNa- of atrial flutter. Am Heart J 1999;137:439—42.
mara DG. Energy dose for ventricular defibrillation of chil- 139. Kerber RE, Kienzle MG, Olshansky B, et al. Ventricular
dren. Pediatrics 1976;58:898—901. tachycardia rate and morphology determine energy and
120. Cummins RO, Austin Jr D. The frequency of ‘occult’ ventric- current requirements for transthoracic cardioversion. Cir-
ular fibrillation masquerading as a flat line in prehospital culation 1992;85:158—63.
cardiac arrest. Ann Emerg Med 1988;17:813—7. 140. Hedges JR, Syverud SA, Dalsey WC, Feero S, Easter R, Shultz
121. Losek JD, Hennes H, Glaeser PW, Smith DS, Hendley G. Pre- B. Prehospital trial of emergency transcutaneous cardiac
hospital countershock treatment of pediatric asystole. Am pacing. Circulation 1987;76:1337—43.
J Emerg Med 1989;7:571—5. 141. Barthell E, Troiano P, Olson D, Stueven HA, Hendley G. Pre-
122. Martin DR, Gavin T, Bianco J, et al. Initial countershock in hospital external cardiac pacing: a prospective, controlled
the treatment of asystole. Resuscitation 1993;26:63—8. clinical trial. Ann Emerg Med 1988;17:1221—6.
123. Kohl P, King AM, Boulin C. Antiarrhythmic effects of acute 142. Cummins RO, Graves JR, Larsen MP, et al. Out-of-hospital
mechanical stiumulation. In: Kohl P, Sachs F, Franz MR, edi- transcutaneous pacing by emergency medical technicians
tors. Cardiac mechano-electric feedback and arrhythmias: in patients with asystolic cardiac arrest. N Engl J Med
form pipette to patient. Philadelphia: Elsevier Saunders; 1993;328:1377—82.
2005. p. 304—14. 143. Ornato JP, Peberdy MA. The mystery of bradyasystole
124. Befeler B. Mechanical stimulation of the heart; its thera- during cardiac arrest. Ann Emerg Med 1996;27:576—
peutic value in tachyarrhythmias. Chest 1978;73:832—8. 87.
125. Volkmann HKA, Kühnert H, Paliege R, Dannberg G, 144. Niemann JT, Adomian GE, Garner D, Rosborough JP. Endo-
Siegert K. Terminierung von Kammertachykardien durch cardial and transcutaneous cardiac pacing, calcium chlo-
mechanische Herzstimulation mit Präkordialschlägen. ride, and epinephrine in postcountershock asystole and
(‘‘Termination of Ventricular Tachycardias by Mechan- bradycardias. Crit Care Med 1985;13:699—704.
ical Cardiac Pacing by Means of Precordial Thumps’’). 145. Quan L, Graves JR, Kinder DR, Horan S, Cummins RO.
Zeitschrift für Kardiologie 1990;79:717—24. Transcutaneous cardiac pacing in the treatment of out-
126. Caldwell G, Millar G, Quinn E. Simple mechanical meth- of-hospital pediatric cardiac arrests. Ann Emerg Med
ods for cardioversion: Defence of the precordial thump and 1992;21:905—9.
cough version. Br Med J 1985;291:627—30. 146. Dalsey WC, Syverud SA, Hedges JR. Emergency department
127. Morgera T, Baldi N, Chersevani D, Medugno G, Camerini F. use of transcutaneous pacing for cardiac arrests. Crit Care
Chest thump and ventricular tachycardia. Pacing Clin Elec- Med 1985;13:399—401.
trophysiol 1979;2:69—75. 147. Knowlton AA, Falk RH. External cardiac pacing during in-
128. Rahner E, Zeh E. Die Regularisierung von Kammertachykar- hospital cardiac arrest. Am J Cardiol 1986;57:1295—8.
dien durch präkordialen Faustschlag. (‘‘The Regulariza- 148. Ornato JP, Carveth WL, Windle JR. Pacemaker insertion for
tion of Ventricular Tachycardias by Precordial Thumping’’). prehospital bradyasystolic cardiac arrest. Ann Emerg Med
Medizinsche Welt 1978;29:1659—63. 1984;13:101—3.

Você também pode gostar