UNIT-I: PMMC Instruments, Galvanometers, DC Ammeter &


Voltmeter, Series Ohmmeter
ELECTRONIC MEASUREMENTS AND INSTRUMENTATION
(REC-403)
ECE-IV SEM

Dr. ASHISH GUPTA


Associate Professor
TOPICS TO BE COVERED

❖PMMC Instruments

❖ Galvanometers

❖ DC Ammeters

❖ DC Voltmeters

❖ Series Ohmmeters

27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 2


Permanent Magnet Moving Coil (PMMC)
Instruments

Fig. 1: The deflecting force in a PMMC instrument is provided by a


current-carrying coil pivoted in a magnetic field.
27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 3
Permanent Magnet Moving Coil (PMMC)
Instruments

Fig. 2: The controlling force from the springs balances the deflecting
force.
27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 4
Permanent Magnet Moving Coil (PMMC)
Instruments

Fig. 3: Lack of damping causes


the pointer to oscillate.

Fig. 4: The damping force in a


PMMC instrument is provided by
eddy currents induced in the
aluminum coil former as it moves
through the magnetic field.

27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 5


Permanent Magnet Moving Coil (PMMC)
Instruments

Fig. 5: A typical PMMC


instrument is constructed
using a horse-shoe magnet,
soft-iron pole shoes, a soft-
iron core and a suspended coil
that moves in the air gap
between the core and the pole
shoes.

27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 6


GALVANOMETERS

Fig. 6: A galvanometer is simply an


extremely sensitive PMMC
instrument with a center-zero scale

Fig. 7: An adjustable shunt resistor is


employed to protect the coil of a
galvanometer from destructively
excessive current levels. The shunt
resistance is initially set to zero and then
gradually increased to divert current
through galvanometers.
27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 7
DC AMMETERS

Fig. 9: Circuit diagram of DC


Ammeter. The meter current is
directly proportional to the shunt
current, so that the meter scale
can be calibrated to indicate the
Fig. 8: A DC ammeter consists of a total current.
PMMC instrument and a low resistance
shunt.
27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 8
DC AMMETERS
Fig. 10: A swamping resistance made of a
material with a near-zero temperature
coefficient can be connected in series with
the coil of a PMMC instrument to minimize
temperature errors.

Fig. 11: Multirange ammeter using


switched shunts
A Multirange ammeter consists of a
PMMC instrument, several shunts
and a rotary switch that makes
contact with the next shunt before
loosing contact with the previous
one when switching between
different ranges.
27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 9
DC AMMETERS

(a) (R1 + R2 + R3) in parallel


with Rm

(b) (R1 + R2) in parallel


with (Rm + R3)

Fig. 12: An Ayrton shunt used with an ammeter consists of several series-
connected resistors all connected in parallel with the PMMC instrument.
Range change is effected by switching between the resistor junctions
27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 10
DC VOLTMETERS

(a) Construction of DC Voltmeter (b) DC Voltmeter Circuit

Fig. 13: A DC Voltmeter is made up of a PMMC instrument and a series


multiplier resistor. The meter current is directly proportional to the applied
voltage, so that the meter scale can be calibrated to indicate the voltage.

27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 11


DC VOLTMETERS

(a) Multirange voltmeter using (b) Multirange voltmeter using


switched multiplier resistors series-connected multiplier resistors

Fig. 14: A Multirange DC Voltmeter consists of a PMMC instrument, several


multiplier resistors and a switch for range selection. Individual or series-
connected resistors may be used.

27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 12


SERIES OHMMETERS

Fig. 15: Basic series ohmmeter


circuit consists of a PMMC
instrument and a series-
connected standard resistor
(R1). When the ohmmeter
terminals are shorted (RX = 0)
meter full-scale deflection
(a) Basic circuit of series ohmmeter occurs. At half-scale deflection
RX = R1 and at zero deflection
the terminals are open-
circuited.

(b) Ohmmeter Scale

27 March 2019 Dr. ASHISH GUPTA, Asso. Professor 13


SERIES OHMMETERS
Ohmmeter with Zero Adjust

Fig. 16: An adjustable resistor (R2) connected in parallel with the meter
provides an ohmmeter zero control. The ohmmeter terminals are
initially short-circuited and the zero control is adjusted to give zero-
ohms reading. This eliminates errors due to variations in the battery
voltage.