Você está na página 1de 10

14 Power Supply

14.1 View of Power Supply

Fig. 14-1 Monitor cart 5 power supply

Fig. 14-1 Monitor cart 6 power supply

Ziehm Vision 14-1


26451/28121 - 09/2007
14 Power Supply

14.2 General Description

The Monitor Cart Module 10 Power Supply is the main power module for the Ziehm Vision C-arm
and provides the following:
− power isolation from mains input
− internal power distribution of 230 VAC to the Monitor Cart and for the Mobile Stand via the XP0
cable
− additional 115 VAC insulated mains distribution
− inrush current limiter
− on/off circuitry
− emergency stop and key switch lock for power

14.3 Main Components of the Power Module 10

The TR1 isolation transformer provides:


− isolation from the mains line power supply
− input voltage taps for matching the input voltage to maintain 230 VAC output
− the main supply voltage for the complete C-arm
The U539 power distribution board provides:
− power on/off control
− distribution of power
− U539 allows easy method for matching the input line voltage to the facility power
− matching TR1 input to mains line voltage
− the ability to set the correct voltage is accomplished by following the power selection table in
the diagram below
− X40 to X49 terminals for matching input line to TR1
− output 230 VAC for system
Matching the line input to TR1:
Matching the line input to a facility is accomplished by performing a simple measurement of the
wall outlet connector using a digital volt meter. Monitor the voltage for a period of time long enough
to detect any unusual dips in the mains line (10 to 20 minutes).

NOTE
Those facilities demonstrating large fluctuations in the mains line voltage may
require a more detailed measurement test. Attach a voltage monitoring device to
the mains line and monitor the line voltage for 12 to 24 hours. This is an important
issue and should not be overlooked.

14-2 Ziehm Vision


26451/28121 - 09/2007
14 Power Supply

Once the line voltage is determined to be stable and within the range of the selection table, the
service technician must follow the tap setting and jumper placement as required in the Voltage
Selection table for the U539 board.
230 VAC power distribution on U539:
− the output of the TR1 transformer (X15 and X17) supplies 230 VAC to the U539 board
− distributes 230 VAC to auxiliary equipment by means of the terminal connectors X32 to X38
− supplies power to devices such as video monitors, printers, image system, etc.
− distributes 230 VAC power to the mobile stand by means of the power connector terminals X30
and X31
115 VAC power distribution:
− U539 board receives 115 VAC at connectors X15 and X16 from TR1
− distributes 115 VAC to the auxiliary equipment by means of the terminal connectors X20 to X22

14.4 Module 10 without HC08 Controller on U539

14.4.1 Module 10 power distribution diagram

Fig. 14-2 Module 10 wiring

Ziehm Vision 14-3


26451/28121 - 09/2007
14 Power Supply

Select Voltage
Transformer RTO 2500-000585
105 VAC 110 VAC 115 VAC 120 VAC 210 VAC 220 VAC 230 VAC 240 VAC

Connections with AWG 14 (Blue)

A–H A–F A–D A–B A–H A–F A–D A–B


H–I F–G D–E B–C K–I K–G K–E K–C
K–L K–L K–L K–L

F4 F3 F2

F1A = F1B = 20 A fast F1A = F1B = 15 A slow

→ Ch. 14.4.5, p. 14-6

Table of Colors:

bk = black = schwarz gr = grey = grau vi = violet = violett

bl = blue = blau or = orange = orange wh = white = weiß

br = brown = braun pi = pink = rosa yl = yellow = gelb

gn = green = grün rd = red = rot

14-4 Ziehm Vision


26451/28121 - 09/2007
14 Power Supply

14.4.2 Connectors

The module 10 has many connectors and voltage tap settings. The following will provide informa-
tion on the name and function of the listed connector.

X1: Mains Input I / Earth

1: Earth

Provides connection to the earth or chassis ground for safety

F1X1: Mains Input II / Line

1: Line in

Has to be wired to connectors and according to valid standards

F1X2: Mains to PCB distribution / Line

1: Line

Has to be wired to connectors and according to valid standards

F1X3: Mains Input III / Neutral

1: Neutral in

Has to be wired to connectors and according to valid standards

F1X4: Mains to PCB distribution / Neutral

1: Neutral

Has to be wired to connectors and according to valid standards

X2: Mains Input I / Line

1: Line

X3: Mains Input I / Neutral

1: Neutral

Ziehm Vision 14-5


26451/28121 - 09/2007
14 Power Supply

14.4.3 Insulated XP0 PCB outlet

X30: XP0 connection

screw connection 1: XP0 connector / Internal mains to Mobile Stand

X31: XP0 connection

screw connection 1: XP0 connector / Internal mains to Mobile Stand

14.4.4 Internal mains distribution

X32 – X38: Cart internal mains distribution 230 VAC

3-pin Phoenix MSTBV 2.5/3-G connector


Order no. Phoenix: 17 53 45 3
Part No.: 21368

Pin 1: Protective earth

Pin 2: 230 VAC insulated mains I / Neutral

Pin 3: 230 VAC insulated mains I / Line

! Secondary transformer coil – insulated / 16 mm distance to primary

X20 – X22: Cart internal mains distribution 115 VAC

3-pin Phoenix ICV 2,5/3-G-5,08 connector


Order no. Phoenix: 17 85 95 5
Part No.: 21547

Pin 1: Protective earth

Pin 2: 115 VAC insulated mains I / Neutral

Pin 3: 115 VAC insulated mains I / Line

! Secondary transformer coil – insulated / 16 mm distance to primary

14.4.5 Mains input voltage selection

Wiring

Selectable input voltages: 105 VAC; 110 VAC; 115 VAC; 120 VAC;
210 VAC; 220 VAC; 230 VAC; 240 VAC

Hardwired combinations

14-6 Ziehm Vision


26451/28121 - 09/2007
14 Power Supply

105 VAC wiring

X50 / A connected with X43 / H

X43 / H connected with X42 / I

X41 / K connected with X40 / L

Hardwired combinations

Wire type: AWM Style 1015

Main Fuse F1: 20 A fast (Part No.: 17044)

Main Fuse F2: 20 A fast (Part No.: 17044)

Supply voltage 100 mA slow (Part No.: 17010)


Fuse F4:
Fuse F2; F3 not connected

110 VAC wiring

X50 / A connected with X45 / F

X45 / F connected with X44 / G

X41 / K connected with X40 / L

Mains fuse F1: 20 A fast (Part No.: 17044)

Mains fuse F2: 20 A fast (Part No.: 17044)

Supply voltage 100 mA slow (Part No.: 17010)


Fuse F4:
Fuse F2; F3 not connected

Hardwired combinations

Wire type: AWM Style 1015

115 VAC wiring

X50 / A connected with X47 / D

X47 / D connected with X46 / E

X41 / K connected with X40 / L

Mains fuse F1: 20 A fast (Part No.: 17044)

Mains fuse F2: 20 A fast (Part No.: 17044)

Supply voltage 100 mA slow (Part No.: 17010)


Fuse F4:
Fuse F2; F3 not connected

Hardwired combinations

Wire type: AWM Style 1015

Ziehm Vision 14-7


26451/28121 - 09/2007
14 Power Supply

120 VAC wiring

X50 / A connected with X49 / B

X49 / B connected with X48 / C

X41 / K connected with X40 / L

Mains fuse F1: 20 A fast (Part No.: 17044)

Mains fuse F2: 20 A fast (Part No.: 17044)

Supply voltage 100 mA slow (Part No.: 17010)


Fuse F4:
Fuse F2; F3 not connected

Hardwired combinations

Wire type: AWM Style 1015

210 VAC wiring

X50 / A connected with X43 / H

X41 / K connected with X42 / I

Mains fuse F1: 15 A slow (Part No.: 17076)

Mains fuse F2: 15 A slow (Part No.: 17076)

Supply voltage 100 mA slow (Part No.: 17010)


Fuse F3:
Fuse F4; F2 not connected

Hardwired combinations

Wire type: AWM Style 1015

220 VAC wiring

X50 / A connected with X45 / F

X41 / K connected with X44 / G

Mains fuse F1: 15 A slow (Part No.: 17076)

Mains fuse F2: 15 A slow (Part No.: 17076)

Supply voltage 100 mA slow (Part No.: 17010)


Fuse F3:
Fuse F4; F2 not connected

Hardwired combinations

Wire type: AWM Style 1015

14-8 Ziehm Vision


26451/28121 - 09/2007
14 Power Supply

230 VAC wiring

X50 / A connected with X47 / DMains fuse

X41 / K connected with X46 / E

Mains fuse F1: 15 A slow (Part No.: 17076)

Mains fuse F2: 15 A slow (Part No.: 17076)

Supply voltage 100 mA slow (Part No.: 17010)


Fuse F3:
Fuse F4; F2 not connected

Hardwired combinations

Wire type: AWM Style 1015

240 VAC wiring

X50 / A connected with X49 / B

X41 / K connected with X48 / C

Mains fuse F1: 15 A slow (Part No.: 17076)

Mains fuse F2: 15 A slow (Part No.: 17076)

Supply voltage 100 mA slow (Part No.: 17010)


Fuse F2:
Fuse F4; F3 not connected

Hardwired combinations

Wire type: AWM Style 1015

Ziehm Vision 14-9


26451/28121 - 09/2007
14 Power Supply

14.4.6 Fuses

Fuse F1 A

210-220-230-240 VAC system


Mains fuse: 15 A slow (Part No.: 17076)

105-110-115-120 VAC system


Mains fuse: 20 A fast (Part No.: 17044)

Fuse F1 B

210-220-230-240 VAC system


Mains fuse: 15 A slow (Part No.: 17076)

105-110-115-120 VAC system


Mains fuse: 20 A fast (Part No.: 17044)

Fuse F2

240 VAC system


Supply voltage: 100 mA slow (Part No.: 17010)

Not installed for 230 VAC; 115 VAC; 110 VAC; 105 VAC

Fuse F3

210-220-230 VAC system


Supply voltage: 100 mA slow (Part No.: 17010)

Not installed for 240 VAC; 115 VAC; 110 VAC; 105 VAC

Fuse F4

105-110-115-120 VAC system


Supply voltage: 100 mA slow (Part No.: 17010)

Not installed for 240 VAC; 230 VAC

Fuse F5

Supply voltage: 315 mA slow

Installed for all mains input voltages

14-10 Ziehm Vision


26451/28121 - 09/2007

Você também pode gostar