Escolar Documentos
Profissional Documentos
Cultura Documentos
LABORATORY: SECTION of 1
PROCESS CHEMICAL REVISION NO:
TECHNOLOGY
29 SEPT
EFFECTIVE DATE:
EXPERIMENT: GAS FLOW 2014
PROCESS CONTROL AMENDMENT
DATE:
COURSE CODE
EXPERIMENT NO. EXPERIMENT 4
INTRODUCTION: /5%
PROCEDURE: /5%
RESULTS& CALCULATIONS /15%
ANALYSIS /15%
DISTRIBUTION OF MARKS FOR DISCUSSIONS: /20%
LABORATORY REPORT: ADDITIONAL QUESTIONS /15%
CONCLUSION /10%
REFERENCES: /5%
TOTAL: /100%
I hereby declare that I have prepared this report with my own efforts. I also admit to
not accept or provide any assistance in preparing this report and anything
that is in it is true.
1. Switch the "PANEL, SCADA/DDC" selector switch at the front of the cubicle to "PANEL,
SCADA" position. Switch ON the main power supply at the front of the cubicle. All the panel
instruments will lit up.
2. From the process air supply, trace the two pipelines, PLI and PLII as follows:
Pipeline I (PLI):
AR900-MV900A-T90-MV900D-PLI-FI911-TE911-FE91/FT91-FCV91-FCV1-VF
Pipeline II (PLII):
AR900-MV900A-T90-MV900D-PLII-FCV90-PG902-V
3. Open valves of the pipelines PLI and PLII at their various manual valves. Close the manual
by-pass valve of the control valve FCV91 and the ON/OFF control valve FCV90.
4. The pipeline PLI has two discharge manual valves, VF and MV901. Open the vent VF and
close the valve MV901.
5. Make sure the valves MV900A and MV900D are open and valves MV900B and MV900C are
shut.
6. Place the panel controller FIC91 in the Manual (M) mode. Open fully the control valve
FCV91 by setting the MV of FIC91 to -6.3%.
7. Adjust the manual valve MV900A till the air flow rate is about 50 kg/hr (Read the flow
rate from PV of FIC91).
8. Check that vent valve and bottom drain valve of vessel T90 are shut. Set MV = 106.3% at
FIC91 to fully shut the control valve FCV91.
9. Check control valve FCV91 (in the plant) is fully shut and FI911 reads the air volumetric
flow rate as ZERO. Also check the mass flow rate Fm at FIC91 is ZERO.
10. Now set the MV = 50% at FIC91 and check control valve FCV91 (in the plant) is 50%
opened.
FACULTY : UniKL MICET EDITION: 1
LABORATORY: SECTION of 1
PROCESS CHEMICAL REVISION NO:
TECHNOLOGY
29 SEPT
EFFECTIVE DATE:
EXPERIMENT: GAS FLOW 2014
PROCESS CONTROL AMENDMENT
DATE:
11. Note the variables that are recorded in each channel of the recorder FPTR91.
Red Pen - Mass flow rate (kg/hr)
Green Pen - Volumetric flow rate (m3/hr)
Blue Pen - Temperature (°C)
Locate the rotameter FI911 and pressure gauge PG900. Take both instruments’
readings and also note the units used. (Rotameter reading should be taken as ±0.5
Nm3/hr)
Now, take another sets of readings for MV = 50% and MV = 70% and record them in
the Table 1 in the results section.
FACULTY : UniKL MICET EDITION: 1
LABORATORY: SECTION of 1
PROCESS CHEMICAL REVISION NO:
TECHNOLOGY
29 SEPT
EFFECTIVE DATE:
EXPERIMENT: GAS FLOW 2014
PROCESS CONTROL AMENDMENT
DATE:
II III
PB = 150% PB = 100%
TI = 6s TI = 6s
TD = 0s TD = 0s
D. SHUTDOWN PROCEDURE
1. Stop the recorder chart by pressing ”RCD” press button at recorder FPTR91.
2. Switch FIC91 to Manual mode with MV = 0%.
3. Switch off the main power supply.
4. Shut off the process air supply at AR900.
5. Shut off the instrument air supply.
6.
6.0 RESULTS & CALCULATIONS
A. The recorder FPTR91 records the results of the experiments. Suitable portions of the
recorder chart paper should be submitted as RESULTS of the experiment.
B. The report should contain:
i) The P&I Diagram with all major components marked clearly.
ii) The instruments readings taken simultaneously and recorded for different valve
opening positions.
iii) Comparison between direct instrument data with calculated data for Fvb and Fm
iv) The PID flow control responses for disturbances and setpoint changes for different
controller settings.
v) Discussion on the response characteristics for different controller settings.
FACULTY : UniKL MICET EDITION: 1
LABORATORY: SECTION of 1
PROCESS CHEMICAL REVISION NO:
TECHNOLOGY
29 SEPT
EFFECTIVE DATE:
EXPERIMENT: GAS FLOW 2014
PROCESS CONTROL AMENDMENT
DATE:
vi) Tuning by Ziegler Nichols method and its response.
Analysed critically each results obtain in section 6.0 by referring to the theory/introduction
parts. Find reasons (suggested reasons) on each occurrence shown in the results.
8.0 DISCUSSIONS
Discuss the overall experiment on data obtain based on your understanding. Discussion
should relate reasons for the phenomena/results obtained, theoretical parts, suggested reasons
(if necessary but back with sound theoretical knowledge) and your overview understanding.
FACULTY : UniKL MICET EDITION: 1
LABORATORY: SECTION of 1
PROCESS CHEMICAL REVISION NO:
TECHNOLOGY
29 SEPT
EFFECTIVE DATE:
EXPERIMENT: GAS FLOW 2014
PROCESS CONTROL AMENDMENT
DATE:
9.0 ADDITIONAL QUESTIONS
i. Water has been analysed for pH, conductivity and dissolve oxygen. What does each of this
measurement tell you about the water?
ii. Includes short essay on your understanding on pH, conductivity and dissolve oxygen. Relate
your discussion on the requirement of industrial effluent released (if applicable).
9.0 CONCLUSION
Conclusion is merely a summary of the important findings already reported in the discussion
section. It also relates to the objectives stated earlier.
11.0 REFERENCES
Minimum of at least 5 references (with 3 of them are paper/book/chapter based)