H23 Hi Pressure High motor current 1. Check for free manual movement of
Level detected on startup plunger using manual knob
motor test.
2. Check motor calibration for a reason
why motor current is high
AVIDIA ERROR CODES
Code Error Text Explanation Actions
HW23 Motor Test System startup motor 1. Check current plunger position has not
test has failed. activated one of the override switches
and cut motor power from the base by
moving piston with manual knob
H30 Backup Battery Small monitor battery 1. Replace small monitor assembly
Lo low.
HW47 H/W watchdog Not used: should not 1. Replace head 1. Replace head
occur assembly
H48 BASE Hi Amps1 Higher than expected 1. Check battery capacity and charge
current required from battery
battery 1.
2. Replace battery pack
H49 BASE Hi Amps2 Higher than expected 1. Check battery capacity and charge
current required from battery
battery 2.
2. Check motor calibration
HW49 BASE armed & Base armed but FDE 1. Test head cable or replace
moving counts detected,
indicating the plunger 2. Replace and reprogram the base board
is moving.
AVIDIA ERROR CODES
Code Error Text Explanation Actions
H51 Volume left in End override switch 1. Manually retract the plunger away from
syringe has been activated the end override switch and press
but does not agree UNLOCK. The plunger should retract
with the plunger fully and reset the position count.
position.
2. Power the plunger to the end position
and check the end override switch is not
activated
3. Adjust if required
H53 Footswitch not Footswitch released 1. Retry endurance test with footswitch
pressed during endurance pressed
test.
2. Check footswitch cable for continuity
H54 Onboard Clock During startup, the 1. Go to Options, reset the clock and cycle
stopped real time clock on the power
controller board did
2. Check for damage to the control board
not change.
caused by the acrylic shield for the
plunger lock solenoid
H55 Syringe Sensor Syringe sensor value 1. Check for open circuit sensor cable by
Lo has dropped to zero. monitoring value in Self Test mode while
moving the plunger manually
H56 Syringe Sensor Syringe sensor value is 1. Check for short circuit sensor cable as in
AVIDIA ERROR CODES
Code Error Text Explanation Actions
H60 5VON Supply < 5VON is the low 1. Replace and reprogram the control
4.5V power supply for the board
head wakeup circuit.
H67 END switch Plunger is too far 1. Manually retract plunger and then press
forward. UNLOCK
H69 5V Supply Out 5V supply on the 1. Disconnect motor encoder from the
of Range controller board is low controller board. If this fixes the
(<4.5V). problem, replace the entire head
H70 12V Supply < Regulated 12V for the 1. Connect mains power to first eliminate
11V back board is low. low battery capacity
H71 STARTPOS- Plunger has retracted 1. Manually move the plunger forward and
retract too far. then press UNLOCK
H72 Please turn Error in motor 1. Run motor calibration test and correct
motor calibration detected. as required
AVIDIA ERROR CODES
Code Error Text Explanation Actions
H79 BASE FDE Base is not receiving 1. Test head cable by checking pin-to-pin
expected FDE counts. continuity
H81 Auto Power Off Failed to power off 1. Check setting of Auto Off option in
Fail automatically. service mode (should be enabled)
H86 Check BASE Battery voltage high 1. Relays may be jammed closed on base
Relays (over 24 V). board
H87 BASE time Injection ran over 1. Test head cable or replace
overrun time according to
base monitoring 2. Replace and reprogram the base board
H89 BASE volume Over volume injection 1. Test head cable or replace
overrun according to FDE
counts received by 2. Replace and reprogram the base board
base.