This chapter explains the setting method for performing program operation, the method of
starting and ending operation, and the displays of monitors.
Reference
By using a Web application, the setting of program operation, operation, and monitoring
can be performed remotely. For details, see the Network guide.
Program operation changes the settings of the temperature and humidity in the test area in
accordance with the program that you create by combining temperature and humidity steps.
For each program pattern, up to 99 steps of operation data can be edited or registered.
By setting the time, temperature, humidity and ramp for each step, you can operate the
equipment according to an arbitrary time control as shown below.
Temperature (°C)
1st 2nd 3rd 4th
step step step step
100 5th
step
50
20
Time
Figure 4.1 Example of program operation
ESPEC CORP. | 69
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
Enter the temperature, humidity, and setting time for each step to create or edit an operation
pattern.
<Procedure>
1) Press the Program Setup tab.
To create a new program, select an unregistered program number from the program
list.
If no programs are registered, no names are displayed in the program list.
To edit a registered program, press the number of the program that you want to edit.
* Depending on whether you create a new program or edit a registered program, the
keys displayed on the program selection pop-up screen are different.
70 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
ESPEC CORP. | 71
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
The items that you can edit on the edit step screen are shown in Table 4.2.
72 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
Reference
Setting Set Temp/Hum CTRL to OFF causes only the time to pass.
If Set Temp/Hum CTRL of step 1 is set to OFF, this creates a wait time until starting
temperature/humidity control. In other words, this can be used as an operation start timer.
The upper and lower limits of the input range are determined by the temperature range of
the equipment and High/Low Limit setting of [Others].
Reference
For the upper and lower limits, the values entered in "4.4.4 Entering high/low limit setup"
are reflected.
ESPEC CORP. | 73
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
Reference
When you perform ramp control in Step 1, select the start conditions (Process Value, Set
Point) in the start condition setup that you access by pressing [DETAIL] on the Edit
Program screen.
* If you set start conditions to OFF, ramp control is not performed even if the set point is
entered in the step input screen.
For details, see "4.4 Setting program details."
7) Set the soak time control to maintain operation at the temperature/humidity set point.
Select [ON] or [OFF] to configure this setting.
ON: Soak time control is performed.
OFF: Soak time control is not performed.
When [ON] is set: After the step operation starts, the equipment waits until the
process value reaches the control attainment range, and then
starts counting the time ( see "5.11.3 Setting the control
attainment range").Therefore, the soak time in the
temperature/humidity set point is the same as the setting time.
If the soak time control is set to [ON], it is not possible to set
Ramp Control to [ON]. When set to [ON], the message “Incorrect
ramp and soak control setup entered” appears. Change the
setting.
When [OFF] is set: The equipment starts counting the time at the same time that
step operation starts.
74 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
9) After you finish setting the step, press [ENT NEXT] to move to the next step. If the
next step has not been set, the previous values are copied to it.
To return to the previous step, press [ENT PREV].
To end step entry, press [ENT] or [QUIT].
To perform detailed settings, press [DETAIL].
ESPEC CORP. | 75
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
<Procedure>
1) To perform detailed settings, press [DETAIL] to move to the Details screen.
The items that you can set on this screen are shown below.
Details 1 (Pause, Refrigeration, DRN (optional))
Details 2 (Time signal 1)
Time signals 2 to 11 (optional)
Option (air flow adjusting, specimen temperature control)
For details on operating the optional functions, see the Option guide.
* For pause operation performed when the door is opened, see "5.11.2 Setting the
operation when the chamber is running."
Reference
To go to the next step after the program is paused, select [RESUME] or [NEXT STEP] on the
Operation Mode screen.
For details, see "4.9 Starting/ending or pausing/resuming operation."
76 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
Reference
You cannot change the refrigeration capacity during operation. Set the refrigeration capacity
when you set the program prior to operation.
Reference
Immediately after the program moves to the next step, the time signal changes the set
contact output.
You can assign names to time signals. (Use the network function.)
ESPEC CORP. | 77
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
5) If you press [ENT NEXT] after setting the step, the program moves to the next step. If
the next step has not been set, the previous values are copied to it.
To return to the previous step, press [ENT PREV].
To end step entry, press [ENT] or [QUIT].
ENT: Enters the input value on the step input screen, closes the step input
screen, and returns to the Edit Program screen. (The program does
not move to the next step.)
ENT NEXT: Enters the input value on the step input screen and moves to the next
step.
ENT PREV: Enters the input value on the step input screen and moves to the
previous step.
QUIT: Abandons the input value on the step input screen, closes the step
input screen, and returns to the Edit Program screen.
If setting is stopped without pressing [ENT], [ENT NEXT], or [ENT
PREV], the step setup being edited is abandoned.
78 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
ESPEC CORP. | 79
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
80 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
ESPEC CORP. | 81
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
After an operation pattern is created and registered, steps can be edited, inserted, copied, and
deleted.
<Procedure>
1) Press the Program Setup tab and select the program number to be edited from the
program list.
82 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
Reference
After a step is inserted, copied, or deleted, be sure to press [SAVE].
If you do not want to reflect the edited data, press [NO] on the confirmation screen. The
edited data is abandoned and program editing is ended.
For the operating program, the pattern editing and program details can be checked by
using the editing function but they cannot be saved and the program cannot be deleted.
<Procedure>
1) Pressing [EDIT] at the bottom of the screen pops up the following Edit screen.
Select [INSERT].
ESPEC CORP. | 83
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
2) Check that the edit mode display frame is changed to <Insert> and specify the place
to insert.
* You cannot insert steps beyond the maximum number of steps (99 steps).
3) When the step to be inserted is selected, the step edit screen appears.
Because the input value is displayed at the left of the step selected in the step 2,
modify it as required and press [ENT].
To cancel operation after the step edit screen is displayed, press [ENT] to close the
step edit screen.
Because the step has already been inserted at this point, remove unnecessary steps.
For details, see "4.3.3 Deleting steps."
84 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
To cancel insertion, press [QUIT]. The program returns to the program list.
* You cannot copy steps beyond the maximum number of steps (99 steps).
2) Select the top step number and last step number of the copying origin to specify the
copying range.
* To insert only one step, press the same number.
To cancel copying, press [QUIT]. The program returns to the input mode.
ESPEC CORP. | 85
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
[YES]: Copies the selected step and returns to the Step Input screen.
(The step is copied and inserted into the left side of the place in which the
step number is displayed in red.)
[NO]: Returns to the copying destination step selection screen.
Reference
Example:
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 2 3 4 8
If steps 2 to 4 are selected and step 8 is specified as the copying destination, steps 2 to 4 are
copied and inserted into the place before step 8.
86 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
<Procedure>
1) Pressing [EDIT] at the bottom of the screen pops up the Edit screen.
Select [DELETE].
2) Select the top step number and last step number of the steps to be deleted to specify
the range.
To cancel deletion, press [QUIT]. The program returns to the input mode.
[YES]: Deletes the selected step and returns to the step input screen.
[NO]: Returns to the screen on which the last step to be deleted is selected
(previous screen).
ESPEC CORP. | 87
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
In [DETAIL] in the Edit Program screen, you can set the start/end condition, counter, and
high/low limit setup.
Press [DETAIL].
<Procedure>
1) Press [DETAIL] on the Edit Program screen, and then press To Start.
Set it when you want to perform ramp operation in Step 1. Enter the process values
(temperature and humidity in the test area) or the set points (temperature and
humidity) and set the ramp control in Step 1 to [ON].
PV (Process Value): Select this when you want to start process control from the
current temperature and humidity in the test area.
SP (Set Point): Select this when you want to set arbitrary temperature and
humidity values when starting ramp control.
OFF: Disables the temperature and humidity settings.
88 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
2) When [SP] is selected, the start temperature and humidity can be changed.
* The ramp operation controls the temperatures from the set point of the previous
step to that of the present step at a certain ramp.
• Start condition: OFF • Start condition: Process Value or Set Point
With ramp
Without With ramp With ramp Without With ramp
ramp ramp
ESPEC CORP. | 89
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
<Procedure>
1) Press [DETAIL] on the Edit Program screen, and then press To End.
Select one condition for ending operation.
Stop: Keeps the operation stop status. (Power of the instrumentation panel
remains ON.)
Power OFF: Turns off the power of the instrumentation panel. (The primary circuit
breaker of the equipment remains turned on.)
Const 1, 2, or 3: After program operation is ended, constant operation No.1, No.2, or
No.3 is performed continuously.
Hold Last Step: Retains the setting at the end of the last step.
Start Next Program:
After program operation is ended, another program is executed.
For setup, see step 2.
Reference
The default value of the end condition is "Power OFF." If the end condition is not changed,
the instrumentation panel is turned off after the end of the program and the temperature in
the test area approaches the ambient temperature.
Depending on the ambient temperature, the temperature may approach the zero C in
winter. If you want to keep the temperature in the test area, select "Hold Last Step."
90 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
ESPEC CORP. | 91
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
Reference
The number of operations from the repeating start step to the repeating end step is
(Repeat Cycle +1).
92 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
<Procedure>
1) Press [DETAIL] on the Edit Program screen, and then press [Counter].
2) Set Counter A. Press [ON], and then enter Repeat Cycle , the repeating start step
, and the repeating end step .
When Counter B is also specified, set it in the same manner as Counter A.
Reference
Counter setting example
Example 1
Counter A Counter B
(repeat three times) (repeat six times)
Step 1 2 3 4 5 6 7 8 9 10
Repeating start step Repeating end step Repeating start step Repeating end step
Counter A: Repeating start step 1 Counter B: Repeating start step 6
Repeating end step 4 Repeating end step 9
Repeat cycle 3 Repeat cycle 6
Example 2
Step 1 2 3 4 5 6 7 8 9 10
Repeating start step Repeating start step Repeating end step Repeating end step
Counter A: Repeating start step 1 Counter B: Repeating start step 3
Repeating end step 9 Repeating end step 7
Repeat cycle 3 Repeat cycle 6
(Continued on the next page.)
ESPEC CORP. | 93
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
Step 1 2 3 4 5 6 7 8 9 10
Repeating start step Repeating end step
Counter A: Repeating start step 1 Counter B: Repeating start step 1
Repeating end step 9 Repeating end step 9
Repeat cycle 3 Repeat cycle 6
If the repeating end step and the repeating start step of Counters A and B are the same,
Counter B is executed first, and then Counter A is executed. In this case, the number of
operations from the repeating start step to the repeating end step is
[(Repeat Cycle of Counter B +1) x (Repeat Cycle of Counter A +1)].
Counter A (B)
Counter B (A)
Step 2 3 4 5 6 7 8 9 10
Repeating Repeating Repeating Repeating
start step start step end step end step
If both Counters A and B are used, it is not possible to set the repeating end step or the
repeating start step of a counter within the other counter.
94 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
+90°C 10°C
Absolute high limit temperature
Range of temperatures that
can be set in the test area
<Procedure>
1) Press [DETAIL] on the Edit Program screen, and then press Hi/Lo LMT.
ESPEC CORP. | 95
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
2) Enter the Abs High and Abs Low values for the temperature and humidity in the test
area.
Press [Set Point], enter a numeric value, and press [ENT].
3) Enter the upper limit deviation alarm values of the temperature in the test area.
For protective actions, see Tables 3.1 and 3.2 (in chapter 3).
Reference
The program name is automatically set to "PGM-program number".
Example: PGM-01
For details on changing the program name, see the Network guide.
96 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
[YES]: Copies the selected program and returns to the program list.
[NO]: The program returns to the program list.
ESPEC CORP. | 97
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
<Procedure>
1) From the program list, press the number of the program to be deleted, and select
[DELETE] from the pop-up screen.
[YES]: Deletes the selected program and returns to the program list.
[NO]: The program returns to the program list.
Reference
Programs in the following state cannot be deleted.
1. Program for which a timer is set
2. Program set as the end condition
3. Running program
4. Program for which Prevent Data Update is set to ON.
5. Program set to measuring instrument synchronization
98 | ESPEC CORP.
Controller guide Chapter 4 Program operation Bench-top type temperature (and humidity) chamber
Created and registered program patterns can be displayed in the list for each step to check
the settings.
The start and end conditions, counter setting status, and high/low limit setup set in "4.4
Setting program details" can also be checked.
The list can be displayed from both the program list and the Edit Program screen.
On the list screen, data cannot be edited.
<Procedure>
1) From the program list, select the number of the program to be listed.
Select [EDIT] from the pop-up screen.
* Selecting [VIEW] from the pop-up screen displays the View Program screen
directly.
ESPEC CORP. | 99
Bench-top type temperature (and humidity) chamber Controller guide Chapter 4 Program operation
3) Press [Details] to check the contents set in [DETAIL] on the Edit Program screen.
4) Press [MENU] to check the refrigerator setup, equipment setup, time signals, and
counter setup.
Be sure to perform this operation when the operation pattern entry is completed.
Reference
For the operating program, the pattern editing and program details can be checked by
using the editing function but they cannot be saved.
Before starting operation, check that the steps of "2.1 Refrigerator startup preparation," "2.2
Setting the protective device for protecting the specimen," and "2.3 Setting the absolute
high/low limits of temperature (and humidity) warning" in "Chapter 2 Preparing for operation"
have been performed properly. Also check that the information icon (on the Information tab) is
displayed.
Reference
Preventing condensation on the specimen
When conducting a test of the electronic components, make certain that the specimen is not
affected by condensation.
Condensation occurs when the surface temperature of the specimen is lower than the dew point
of the air in the test area.
To avoid condensation on the specimen, it is necessary to perform temperature-only operation in
advance and start temperature/humidity control operation after the surface temperature of the
specimen becomes the same as the temperature in the test area.
In addition, condensation can be avoided by using humidification delay control.
If the temperature and humidity in the test area is 85°C and 85%RH, respectively, the dew-point
temperature of the air in the test area is 80.9°C. Therefore, condensation occurs if the surface
temperature of the specimen is less than 80.9°C. The following shows the dew-point temperature.
Table 4.6 Dew-point temperature
Dry-bulb Relative humidity Dew-point temperature
temperature (°C) (%RH) (°C)
60 85 56.5
85 66.3
70
90 67.7
85 80.9
85
90 82.3
Temperature
set point
85°C
20°C
Humidity set
point
85%rh
Step No.
Figure 4.3 Example of a program to avoid condensation
(1) In Step 2, increase the temperature to 85°C. If the humidity is increased to 85%RH at the same
time, the surface temperature of the specimen becomes lower than the dew-point temperature
and condensation may occur. Therefore, control only the temperature to 85°C. The time during
which Step 2 should be performed differs depending on the heat capacity of the specimen. As
a general guide, the time during which the temperature in the test area is stabilized to 85°C
may be used.
(2) In Step 3, perform operation at 85°C and 85%RH. Though humidity control starts,
condensation can be avoided because the surface temperature of the specimen is nearly
85°C.
<Procedure>
1) Press [Program Selection] on the operation mode screen.
Step
3) Start operation.
Check the selected program name and the start step number, and then start operation.
Press [Start] to display the confirmation screen.
[YES]: Starts operation and returns to the original screen.
"PROGRAM: RUN" is displayed on the operation display area.
[NO]: Returns to the Operation Mode screen.
4.9.2 Pausing operation (only when the program operation is being executed)
<Procedure>
1) Press [Pause] in Program Run.
<Procedure>
1) Press [RESUME] in Program Run.
4.9.4 Proceeding operating step to next step forcibly (only when program operation is
being executed)
<Procedure>
1) Press [NEXT STEP] in Program Run.
<Procedure>
1) Press [Stop] in Stop Operation.
Note
If the operation is stopped during low-temperature operation, condensation may occur on
the equipment surface depending on the ambient conditions. In some cases, this can
result in water leakage in the chamber installation location.
Return the temperature in the test area to room temperature before stopping operation.
If the chamber will not be used for an extended period of time, turn off the circuit breaker
for power conservation and safety.
Reference
Press the instrumentation power switch to also turn off the display of the Instrumentation
screen.
4.11.1 Temperature/humidity
The temperature and humidity in the test area can be monitored.
Temp/Hum screen
Reference
The icons of unavailable functions (optional functions) are masked.
4.11.2 Details
The process values and set points of the temperature and humidity and the status of each
output can be monitored.
Details screen
4.11.3 Program
The detailed information on the pattern during operation can be monitored.
Program Operation Detailed Display screen
Step display frame
End step
Present step
Next step
Table 4.10 Program monitor
Display
Displays the program number and the name of the program in operation.
Displays the step number in operation and the last step number.
Displays the step information.
Top row: Displays the ended step. Nothing is displayed when the initial step is
operated.
Second row: Displays the step in operation. It is enclosed by a yellow frame.
Displays the set points and process values of the temperature/humidity of the step in
operation.
Displays the remaining time of the step in operation. (Subtraction)
Displays the scheduled end time of the program in operation.
The following cases are not included in the scheduled time.
When the soak time control is set to ON.
When the step in operation is forced to proceed to the next step.
* If the scheduled operation end time exceeds 23:59:59 on December 31, 2099,
[----] is displayed.
Displays the accumulated operating time of the program operation.
Note that even if an alarm (alert) is generated or the operation pauses, they are all
accumulated in the total operation time.
Displays the end condition after the program is ended.
(STOP, CONSTANT, Hold Last Step, Start Next Program)
Displays the status of the counter setup.
Pressing the [Counter] key displays the figure.
Close with .
Reference
For the program number and step number when an alarm occurs, check the Program
Operation Details screen before pressing the Power OFF switch. Step numbers during
operation will not be displayed if you press the Power OFF switch.
Accessory Function
Outputs a signal for 3 seconds when the time of the
Time up Optional
start, end, or quick timer elapses.
Outputs a signal when it is determined that the
humidity in the test area has reached a stable state
Hum (temperature and humidity type).
Optional
Attainment For details on humidity attainment condition
settings, see "5.11.3 Setting the control
attainment range."
The contact turns on when all the following
conditions are met.
Door Open Optional
In operation
Door is open
Reference
The names of time signals can be changed.
For details, see "9.14 Time signal name registration" under "Web application" in the
Network guide.