Escolar Documentos
Profissional Documentos
Cultura Documentos
:
27001 La Paz Road, Suite 312 Checklist
Mission Viejo, CA 92691 (ISO 9001/2, EN 46001/2, FDA QSR’s, and
USA 93/42/EEC)
SUMMARY STATEMENTS:
The scope and aim of the audit is for the qualification of this Company (at this one facility only). Any
other facility must be evaluated on its own merits.
Critical Observations: __
Major Observations: __ (None have an impact on Safety, Form, Fit, Function)
Minor Observations : __
Recommendations: __
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
SUMMARY OF ASSESSMENT
QM-SECTION QSR-SECTION SCORE N/C NO.
1. Management responsibility Subpart A - § 820.20 Management responsibility
2. Quality management (QM)-system Subpart B - § 820.5 Quality System
3. Contract review
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
820.25- Personnel
4.1.2.2 Resources (a)General
There are sufficient personnel with the
Is there identification of resource
requirements? necessary education, background,
training, and experience to assure that
Is there provision of adequate all activities required by this part are
resources? correctly performed.
Is there provision of trained
personnel (e.g. internal audits) ? § 820.20 Management Responsibility
(2)Resources
The manufacturer has provided
adequate resources, including the
assignment of trained personnel, for
management, performance of work,
and assessment activities, including
internal quality audits.
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Identification of suitable
verification at appropriate stages in
the realization process.
Development of new
instrumentation
Identification of any
measurement requirement involving
capability that exceeds the known
state of the art.
Clarification of standards of
acceptability for all features and
requirements , inclusive of those
which contain a subjective section.
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
4.3.1 General
Is the procedure for contract
review established?
Are there established and maintained
documented procedures for contract
review and for the coordination of
these activities?
4.3.2 Review
Is the statement of requirements
adequately defined and documented?
Does a review take place
before making an offer in respect to
acceptance of a contract/order?
Are requirements adequately
defined and documented?
Will differences between the
contract or the order requirements and
those in the tender be resolved?
Is it ensured that requirements
can be fulfilled/kept?
In the instance of an order
received by verbal means (word of
mouth/telephone), does there exist
an insurance of requirements agreed
before acceptance?
4.3.4 Records
Are the quality records in Section 4.16
taken in consideration?
Are channels for communication
and interfaces with the customer’s
organization in these contract matters
established? (see comment 9 of
standard)
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
4.4.4 CONTINUATION
DIRECTIVE 93/42/EEC
ANNEX I / essential requirements
Is there a record of how the essential
requirements are to be fulfilled?
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
4.4.4 CONTINUATION
ANNEX 1, 13 / Information supplied by the
manufacturer
Are the requirements for the
symbol considered (13.3 a to m, e.g.
Symbols w.r.t. EN 980)?
Is the intended purpose (13.4)
obvious or is this clearly stated?
Are all the necessary instructions
covered (13.6 a to p)?
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s – 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s – 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
For example
Standards,
Customer drawings,
Laws,
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s – 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
DIRECTIVE 93/42/EEC
ANNEX II 6., V 5., VI 5.
The following administrative provisions
must, at least, after the manufacture of the
last product be controlled:
declaration of conformity,
documentation concerning the QM
system
alterations/changes to the QM system,
documentation for the product
construction (only w.r.t. annex II),
technical documentation,
all decisions and reports from the
notified body
Note:
Documents are input, where they serve as a
revision to a standard.
Data is defined as digitized documents or
records.
The standard makes explicit reference
to the least quality record in a QM
system (see Section 4.16).
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s – 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Current evaluation?
Who will evaluate and select
subcontractors on their ability to meet
requirements including quality systems
and any specific quality assurance
requirements?
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
4.6.3 CONTINUED
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
DIRECTIVE 93/42/EEC
Annex ll 3.2.d
Does there exist adequate product
identification procedures?
Can it be stipulated that the product
identification procedures are kept up to
date (from drawings, specifications) at
every stage of construction?
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
(h)Manufacturing material
Are records maintained for qualified 1.Where a manufacturing material
processes, equipment and personnel? could reasonably be expected to have
an adverse effect on product quality,
Has there been consideration of
Section 4.16/control of quality records? the manufacturer has established and
maintained procedures for the use
and removal of such manufacturing
material to ensure that it is removed or
limited to amount that does not
adversely affect the device’s quality.
2.The removal or reduction of such
manufacturing material is
documented.
(i)Automated processes
1.When computers or automated data
processing systems are used as part
of production of the quality system,
the manufacturer has validated
computer software for its intended use
according to an established protocol.
2.All software changes are validated
before approval and issuance.
3.These validation activities are
documented.
§ 820.70 Production and process controls
Particular requirements for all medical (d)Personnel
devices (EN 46001/2)
1.The manufacturer has established
A) Personnel and maintained requirements for the
Are there established, documented and health, cleanliness, personal
maintained requirements for practices, and clothing of personnel if
health contact between such personnel and
cleanliness product or environment could have an
clothing adverse effect on product quality.
of personnel if contact between such 2.The manufacturer ensures that
personnel and product or environment maintenance and other personnel who
could adversely affect the quality of are required to work temporarily under
product? special environmental conditions are
appropriately trained or supervised by
a trained individual.
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
(2)Inspection
1.The manufacturer conducts periodic
inspections in accordance with
established procedures to ensure
adherence to applicable equipment
maintenance schedules.
2.The inspections, including the date
and individual(s) conducting the
inspections, are documented.
.
§ 820.170 Installation
E)Installation
If appropriate, are there 1.The manufacturer of a device
established and documented, requiring installation has established
both instructions and acceptance and maintained adequate installation
criteria for : and inspection instructions, and
installing and, where appropriate, test procedures.
checking the medical device? 2.Instructions and procedures include
Will records of such be retained see directions for ensuring proper
Section 4.16? installation so that the device will
perform as intended after installation.
3.The manufacturer distributes the
instructions and procedures with the
device or otherwise make them
available to the person(s) installing the
device.
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
F) Special processes
If, when necessary, the quality tests 1.Where the results of a process
can not be reviewed in complete cannot be fully verified by subsequent
scope/comprehensively, will these inspection and test, the process is
processes be performed through validated with a high degree of
quality personnel and/or the process assurance and approved according to
parameters be constantly reviewed? established procedures.
2.The validation activities and results,
Do the quality records identify: including the date and signature of the
a) the work instructions used?, individual(s) approving the validation
b) the date the special process was and where appropriate the major
performed?, equipment validated, are documented.
c) the identity of the operator of the
special process? (b) The manufacturer has established
and maintained procedures for
monitoring and control of process
parameters for validated processes to
ensure that the specified requirements
continue to be met.
(1)The manufacturer has ensured that
the validated processes are performed
by qualified individual(s).
(2)For validated processes, the
monitoring and control methods and
data, the date performed, and where
appropriate, the individual(s)
performing the process or the major
equipment used is documented.
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
4.10.1 General
Are the inspection and testing
procedures sufficiently detailed in
writing?
Are the required inspection and
testing, and the records to be
established, defined in a quality plan
or documented procedures?
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Are all inspection, measuring and 820.70 Inspection, measuring, and test
test equipment
identified, equipment
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
calibrated or (b)
adjusted?
Is this undertaken at prescribed intervals or Calibration
prior to use? 1.
Is the inspection, measuring and test Calibration procedures include specific
equipment calibrated and adjusted
against certified equipment having a directions and limits for accuracy and
known valid relationship? (With precision.
respect to internationally or nationally 2.
recognized standards!).
In the case where no such standards When accuracy and precision limits are not
exist, is the basis used for calibration met, there are provisions for remedial action
documented? to reestablish the limits and to evaluate
whether there was any adverse effect on
the device’s quality.
3.
These activities are documented.
(1)
Calibration standards
Calibration standards used for inspection,
measuring, and test equipment are
traceable to national or international
standards. If national or international
standards are not practical or available, the
manufacturer uses an independent,
reproducible standard. If no applicable
standard exists, the manufacturer has
established and maintained an in-house
standard.
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Where are the calibration records for 820.70 Inspection, measuring, and test
inspection, measuring and test
equipment maintained? equipment
2.These records are displayed on or
Please refer to Section 4.16? near each piece of equipment or are
readily available to the personnel
using such equipment and to the
individuals responsible for calibrating
the equipment.
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Is it ensured (in accordance with 2.The procedures set forth the review
documented procedures) how a
nonconforming product is reviewed? and disposition process.
This review maybe
reworked 3.Disposition of nonconforming
accepted with or without repair by
concession product is documented.
regarded for alternative
application
rejected or scrapped
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
4.13.2 CONTINUATION
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
QUESTIONS COMMENTS
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
4.15.1 General
Are there established and
maintained documented
procedures for handling, storage,
packaging, preservation and delivery of
product?
§ 820.140 Handling
4.15.2 Handling
The manufacturer has established
and maintained procedures to ensure
Are the methods such to prevent
product handling and deterioration? that mixups, damage, deterioration,
contamination, or other adverse
effects to product do not occur during
handling.
§ 820.150 Storage
4.15.3 Storage
1.The manufacturer has established
Are there designated storage areas or
stock rooms to prevent damage or and maintained procedures for the
deterioration of product, pending use or control of storage areas and stock
delivery? rooms for product to prevent mixups,
damage, deterioration, contamination,
Are the storage areas or stock rooms
sufficiently clearly identified? or other adverse effects pending use
or distribution and to ensure that no
Is there appropriate methods stipulated obsolete, rejected, or deteriorated
for authorizing: product is used or distributed.
receipt?
store? 2.When the quality of product
dispatch? deteriorates over time, it is stored in a
manner to facilitate proper stock
rotation, and its condition is assessed
Are the intervals of stock assessment as appropriate.
appropriate.
(b)The manufacturer has established
and maintained procedures that
describe the methods for authorizing
receipt from and dispatch to storage
areas and stock rooms.
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
(c)Labeling storage
The manufacturer stores labeling in a
manner that provides proper
identification and is designed to
prevent mixups.
(d)Labeling operations
1.The manufacturer controls labeling
and packaging operations to prevent
labeling mixups.
2.The label and labeling used for each
production unit, lot, or batch is
documented in the DHR.
(e)Control number
Where a control number is required by
§ 820.65, that control number is on or
accompanies the device through
distribution.
QM-Section 4.15 - Handling, storage, packaging, preservation and delivery / 3
QUESTIONS COMMENTS
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
DIRECTIVE 93/42/EEC
ANNEX l / Sterile products or products for
prior sterilization:
Does the packaging ensure
maintenance of the medical devices
until use (8.3)?
Is the packaging of non-sterile devices
suitable to ensure a suitable level of
cleanliness until prior sterilization
(8.6)?
Does the packaging/label distinguish
between identical and similar products
in both sterile and non-sterile
conditions (8.7)?
Does the label indicate the
recommended word STERILE
according to EN 980 (13.3)?
4.15.5 Preservation
Are methods appropriate?
Is there segregation (where necessary)
of the products?
§ 820.160 Distribution
4.15.6 Delivery
1.The manufacturer has established
Is there an arrangement for the
protection of the quality of product after and maintained procedures for control
final inspection and test? and distribution of finished devices to
ensure that only those devices
Is it contractually specified, that this approved for release are distributed
protection shall be extended to include
delivery to destination? and that purchase orders are reviewed
to ensure that ambiguities and errors
are resolved before devices are
released for distribution.
2.Where a device’s fitness for use or
quality deteriorates over time, the
procedures ensure that expired
devices or devices deteriorated
beyond acceptable fitness for use are
not distributed.
(b)The manufacturer maintains
distribution records which include or
refer to the location of:
The name and address of the
initial consignee;
The identification and quantity of
devices shipped;
The date shipped; and
Any control number(s) used.
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
Will the records be verified and signed 820.184 Device history record
(authorized)?
Important: The standard explicitly 1.The manufacturer maintains device
sets down, in each element the least history records (DHR’s).
quality record standard in a QM 2.The manufacturer has established
system. and maintained procedures to ensure
Are there respective standards w.r.t.: that DHR’s for each batch, lot, or unit
QM-valuation (4.1.3) are maintained to demonstrate that the
contract review (4.3.4) device is manufactured in accordance
design control (4.4.6) with the DMR and the requirements of
design verification (4.4.7) this part.
evaluation of subcontractors 3.The DHR includes, or refers to the
(4.6.2) location of, the following information:
control of customer-supplied
product (4.7) The dates of manufacture;
identification - where necessary - The quantity manufactured;
(4.8) The quantity released for
process control (4.9) distribution;
Preliminary release in incoming The acceptance records which
products (4.10.2.3) demonstrate the device is
test records (4.10.5) manufactured in accordance
control of inspection, measuring with the DMR;
and test equipment (4.11.1) The primary identification label
calibration (4.11.2 e) and labeling used for each
concession (4.13.2) production unit; and
corrective action (4.14.2) Any device identification(s) and
results of internal quality audits control number(s) used.
(4.17)
corrective action in following
audits (4.17)
training (4.18)
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Are the internal audits carried out by 2.The quality audits are conducted by
personnel independent of those
having direct responsibility for the individuals who do not have direct
activity being audited? responsibility for the matters being
audited.
Will the results of the internal audits be 5.The dates and results of quality
recorded?
brought to the attention of the audits and reaudits are documented.
personnel having responsibility in
the audited area?
brought to the attention of the
highest management personnel
( also for the QM
assessment?)
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
4.17 CONTINUATION
DIRECTIVE 93/42/EEC
ANNEX ll, V, VI 3.2 b)
Will the functioning of the QM system
be regularly investigated?
In the frame of the internal audit, is the
QM system investigated to see if it is
suited for the quality lay out and
product?
Will it be invetsigated in internal audits
if the QM system is suited for
identification of non-conforming
products?
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Are the personnel performing activities 3.As part of their training, personnel
affecting quality trained?
are made aware of device defects
Will the means of training be which may occur from the improper
assessed? performance of their specific jobs.
Are personnel
that work under special environment
conditions,
who performs special processes or
functions,
appropriately trained?
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
(ISO 9001/2 - EN 46001/2 ) (FDA QSR’s - 21 CFR 820 et.al.) Where Documentation Sc Practice Sc
found
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
AUDIT - SUMMARY
POSITIVE ASPECTS OF THE QM-SYSTEMS
DEFICIENCIES
GENERAL COMMENTS
Scoring (Sc.): co = Conforms / pc = Conforms with slight deficiencies / nc = Non-conformance / na = Not applicable
Deviation Nr. of
Stipulated deviations
Corrective procedures will be carried out until Responsible signature of the audited firm
Endorsement of consideration
Date Signature
___________________
______________________
_________________________
The audit performed on __________ at ______________ located in ___________ was designed to evaluate the quality
system implemented according to the FDA Quality System Requirements (QSR’s) as given in 21 CFR § 820, ISO 9001/2,
and EN 46001/2. There was a particular emphasis on the use of the Linvatec Supplier Selection Review form, SOP
06204.
The scope and aim of the audit is for the qualification of this supplier for ______________.
Critical Observations: 0
Major Observations: 0 (Major items marked as Form/ Fit Function/ Safety must be corrected)
Minor Observations: 0
Recommendations: 0
City, State
Date
____________________
Lead Auditor and Expert