Press the key. 9 Filter the contents of both test tubes into an empty round cell each.
(round filter, Ederol N°2).
If the liquid runs off turbid, filter the filtrate once again through the same
2 Pipette 2.5 ml of sample solution into a test tube with screw cap
filter. Filter the whole liquid.
(solution sample).
Wait until "fleaction time" or "measure blank appears in the display. 13 Place the blank cell in the cell compartiment.
6 Press the timer key. Reaction time of 5 minutes beggins. 14 Press the key.
Shake several times during this period. 15 Remove the blank cell from the cell compartment and insert the
sample cell.
16Press the key. 5 Pulse la tecla del temporizador. El tiempo de reacción de 5 minutos
comienza.
The result appears in the display.
Pulse la tecla.
7 Añadir 2,5 ml de S04-3A a cada uno y mezclar.
9 Si el líquido se sale turbio, filtre el filtrado una vez más a través del
2 Pipetee 2,5 ml de agua destilada en un segundo tubo de ensayo con mismo filtro. Filtrar todo el líquido.
tapón de rosca (solución en blanco).
4 Presione la tecla de valor en blanco. 11 Calentar los tubos de cata en un baño de agua a 40 ° C durante 7
minutos.
Espere hasta que aparezca "tiempo de reacción" o "medir en blanco" en
la pantalla.
12 Cuando suene el temporizador, transfiera las soluciones a celdas
separadas.
14 Pulse la tecla.
16 Pulse la tecla.