26 SERVICE NOTE Page 1 of 4

FILE TITLE VIVID I/Q SN 78668


Vivid i/q - CardioLab DICOM Workflow BT10 SW Issue

APPLIES TO:
Vivid i/q BT10.2.0 systems

PROBLEM:
ON BT10 systems, the CardioLab data, workflow context box does not appear (it has been inserted in
Figure 1-1 below).

"CardioLab
data,
CardioLab data, workflow workflow"
- On BT10
systems this
does not appear

Figure 1-1 "CardioLab data, workflow" context box shown above as an inset

SOLUTION:
See "work-around" described below.

Copyright© 2011 by General Electric Company Inc. All Rights Reserved


2011.04.26 SERVICE NOTE Page 2 of 4

FILE TITLE VIVID I/Q SN 78668


WORKAROUND:
On BT10 systems use the following work-around:

1.) On the system console press Patient button and in the screen that displays, select Patient List.
The Search/Create Patient screen display - see Figure 1-2 .

Figure 1-2 Search/Create Patient

2.) From the Dataflow drop-down list, select the required CardioLab data flow see Figure 1-2.
3.) Click the yellow folder on screen - see Figure 1-2 - and the Select Input context box opens - see
Figure 1-3.
4.) Select Worklist (shown in Figure 1-3) to open the Search/Create Patient list - see Figure 1-4.

Work list

Figure 1-3 Select Input context list.

Copyright© 2011 by General Electric Company Inc. All Rights Reserved


2011.04.26 SERVICE NOTE Page 3 of 4

FILE TITLE VIVID I/Q SN 78668


5.) Click Query button to display the required CardioLab patient’s name - see Figure 1-4..

Figure 1-4 CardioLab patient’s name selected

6.) Click the Select Patient soft-key at the bottom of the monitor screen - Figure 1-4.
The system starts to process the CardioLab examination.

Copyright© 2011 by General Electric Company Inc. All Rights Reserved


2011.04.26 SERVICE NOTE Page 4 of 4

FILE TITLE VIVID I/Q SN 78668


Submitted by: Geoff Bethlehem - Technical Communicator

For any further information, please contact:

Technical Support
GEHC - Americas GEHC - Europe GEHC - Asia

Online Services Ultrasound Asia


OLC - EMEA
Australia +(61) 1-800-647-855
Online Services Ultrasound Phone: +49 (0)212-2802-652 (-OLC)
China +(86) 800-810-8188
+33 1 3083 1300
India +(91) 1-800-425-8025
Phone: 1-800-321-7937 (USA/Canada) (English/German all segments including
Japan +(81) 42-648-2940
1-262-524-5300 (Latin America) training):
Korea +(82) 1544-6119
Fax: +49 (0)212-2802-431
Singapore +(65) 6277-3444

REVISION HISTORY:

Revision Date Reason for change

0 April 2010 Initial Release

Copyright© 2011 by General Electric Company Inc. All Rights Reserved