Instrument line for medical diagnostics
 
American LAbor
Home
Labor BioMedical Tech. 
How to trouble shoot.
Contact us
About American LAbor  Accustasis 2000 and CoaData 2000 analyzers
Videos
How to place an order GENERAL:
Sales terms & conditions Appliances such as refrigerators, centrifuges, and other motorized appliances
  should NOT be on the same circuit with CoaCalc, CoaData, Accustasis or
Disposables & prices
Accustasis 1000, AS 1000 other computerized equipment. We suggest that this analyzer be connected to a
Accustasis 2000, AS 2000 surge protected electrical circuit or a UPS device, to protect it’s sensitive
CoaData 1000, CD 1000 electrical components.
CoaData 2000, CD 2000
CoaData 3000, CD 3000 If your results are inconsistent the optics may have become contaminated by
CoaSystem fluids being pipetted directly into an empty measuring well, without a cuvette
CoaScreener in place. Turn off the analyzer. The measuring well optics must be clean. DO
CoaLab, Nycomatic, BFA
Disposable details
NOT spray any liquids onto or into the analyzer. Never use acetone, alcohol or
Price list & order form 2013 any petroleum based products to clean this piece of equipment.
Price list & order form 2014
Cleaning may be done by moistening a q­tip with water and a mild soap, such
Other items for sale as a dishwashing liquid, and wiping it around in a circular motion in the
New coagulation analyzers bottom of the test well. DO NOT saturate the q­tip. After using the soap
Demo or Refurbished solution, clean the soap solution by using only water on a q­tip, then use just a
Research analyzers dry q­tip to remove the moisture. Turn on your analyzer and test with QC
Pipettors
reagents.
Printers
 
Service & FAQ SENS LIGHT stays on:
Accustasis 1000, AS 1000 If you turn on your analyzer and the SENS light located on the keypad module
Accustasis 2000, AS 2000 will not go out, neither of the two channels will reset to zero, it is most likely
CoaData 1000, CD 1000 the power supply board in the analyzer has quit working. Sometimes this is all
CoaData 2000, CD 2000 that dies. However on rare occasions other components do not survive. This
CoaData 3000, CD 3000 unit should come in to be repaired contact us.
CoaSystem
CoaScreener
CoaLab, Nycomatic, BFA STIR MOTOR:
Service contract Check each measuring well with a clean cuvette & stir bar to see if the
     features & benefits  measuring well stir motor is working. This is done by having the instrument on
Service rates and warmed up, press reset to clear the display. Slide the spring loaded cover
  back to expose the measuring well. Place the cuvette into the measuring well,
the timer for that channel should begin to count up and the mixer in the bottom
  of the cuvette should be spinning. If this did not occur contact us.
If you have one of the WILL NOT ZERO:
above instruments and If the one channel resets and the other channel does not there may be a bulb
are no longer using it burned out. This may be checked by blocking the light path in the measuring
please contact us, we do channel with a dry q­tip. With the instrument powered on slide the measuring
sometimes buy units to channel cover back and while looking down into the well insert the q­tip. You
refurbish and resell. should see the end of it light up as the intensity of the lamp is increased. If you
do not see it illuminate contact us, this is not usually a repair that can be done
in the field.

TEMPERATURE:
If your unit will not maintain 37 degrees it can be adjusted in your lab. You
will need a mercury or traceable digital thermometer, a #1 and #2 flat blade
screwdriver. Simply turn off the analyzer remove all chemicals from the
  heating block. Gently turn the analyzer over with the bottom up. Remove the
outer corner screws indicated by blue circles, DO NOT REMOVE OTHER
SCREWS, see figure 1 below.
Figure 1

Restore the unit to the upright operating position. With caution connect the
power cord and turn unit on. Take a cuvette insert the thermometer, add ample
water to cover the end of the thermometer, place in incubation block as close to
center as possible, see figure 2.
Figure 2

The temperature is adjusted by turning a screw on an electronic device known
as a potentiometer also referred to as a pot. Using a small #1 flat blade
screwdriver the temperature may be adjusted up by turning the pot clockwise,
or if necessary turn the temperature down by adjusting the pot counter
clockwise. See figure 3 below for correct adjustment screw location marked by
a red arrow.
Figure 3             Click here for a closer view.

The temperature adjustment should be made in small increments, for example
1/4 turn of the pot at a time. Then allow the temperature to stabilize for 20 ­ 30
minutes before any other adjustments. After the temperature is stable, turn off
the analyzer,  replace the cover and secure with screws.

Please see service options for service charges.