Escolar Documentos
Profissional Documentos
Cultura Documentos
-
0
DS 55-82
CJ'I
6
CJ'I
8
I\)
?.....
~
NOTE
1
Bafle/le is not engaged in research for adi·ertising, sales
promotion, or publicity purposes, and this report may
not be reproduced in full or in part for such purposes.
l
Page
INTRODUCTION. • • 1
SUMMA.RY • • • • • • • • • 2
EXPERIMENTAL PROGRAM. • 3
Statistical Measures • 10
Reproducibility • 10
Repeatability • 11
Accuracy • • 11
Experimental Results • • 11
Analysis of Reproducibility. 15
Analysis of Repeatability. 19
Analysis of Accuracy • • 22
Analysis of Between-Rack and Within-Rack Variability • • 27
RECOMMENDATIONS • • • • • • • 28
ACKNOWLEDGEMENTS. • • 29
REFERENCES • • 31
APPENDIX
STANDARD METHODS OF TEST FOR SULFATE ION IN WATER AND WASTE WATER • • 39
i
LIST OF TABLES
Page
LIST OF FIGURES
ii
DS55S2-EB/Apr. 1974
by
INTRODUCTION
SUMMARY OF RESULTS
Sb = 0.0136 ~
.
wh ere, sb' and , m, are given . mg I cm2 - d ay.
in Th'is
relationship is based on measurements at three
sulfation rates over the range of 0.00178 to 0.01371
2
mg/cm -day.
• The mean of the combined measurements between labora-
2
tories at all sites is 0.0063"0 mg/cm -day and the overall
2
standard deviation is 0.00135 mg/cm -day.
• The standard deviation, s , for variations among repeated
w
sulfation rate measurements within laboratories (repeat-
ability) is related to the mean sulfation rate, m, as
follows:
s
w
= 0.00504 ~
2
w.ere,
h s , an,
d m, are given
· · mg I cm - d ay. Th ere 1 ation-
in ·
w
ship is based on duplicate determinations at three sulfation
2
rates over the range of 0.00178 to 0.01371 mg/cm -day.
• The overall standard deviation of combined duplicate deter-
2
minations within laboratories, at all sites is 0.00034 mg/cm -
day and is associated with a mean sulf ation rate of
2
0.00630 mg/cm -day.
3
EXPERIMENTAL PROGRAM
Test Procedure
were sealed in jars under Battelle supervision and returned to the respective
laboratories for sulfate analysis by Method D 516 (Referee Method).
Test Pattern
Table 1 shows the statistical design of the tests for all three
sites with random distribution of total-sulfation candles on the four racks.
At all sites, seven laboratories participated in the test. Figure 1 shows
diagrammatically the placement of containers according to the design of
Table 1.
Spiking Procedures
Test Sites
(a) Entries are laboratory code letters. Parentheses designate samples which were
spiked following exposure.
RACK E RACK F
RACK G RACK H
1 2 3
I
6
4
I I
7
6
8
Rack
Position
Designations
S = Spiked
~ U • Unspiked
Code: S-El
RLk ~ition
FIGURE 1. DIAGRAM OF PLACE1:1ENT OF THE SULFATION STATIONS (D 2010)
7
The racks were placed on the roof as shown in the foreground and
far background of Figure 2. The locations of the individual containers were
randomized on the racks. The exposure area also included sampling stations
for measurement of dustfall. These sampling stations, which also appear in
the photograph, were placed so that no direct interaction would be expected
between the two tests.
Participating Laboratories
Statistical Measures
Reproducibility
Repeatability
Accuracy
Experimental Results
Estimated
Unspiked Sam2les S2iked Sam2les Spiking SJ;!ike Added
Spike
Exposed Sulf ation, Exposed Sulfation, Rate, Sulfa2ion Recovery (E),
Exposure Exposure 2
Area, 2 Area, 2 Code Weight,
Candle Time, mg/cm -day Candle 2 Time, mg/cm -day mg/cm -day mg/cm -day percent
2 mg so2
Lab Location cm daIS ~U} Location cm daIS ~S) ~S-U) Number (S 'l of ~s 'l
Ll El
Gl
100 30 0.00165
0.00166
E7
G7
100
100
30 0.00522
0.00429
0.00357
0.00263
1 10.20 0.00340 105 "'
100 30 30 9 8.79 0.00293 90
Estimated
Uns2iked Sam2les S2ik!d Sam2les Spiking S2ike Added
Sulfation, Exposed Sulfation, Rate Sulfation Spike
Exposed
Exposure Exposure Recovery (E),
Area, 2 Area, 2 2 Weight, 2
Candle Time, mg/cm -day Candle Time, mg/cm -day mg/cm -day Code mg so mg/cm -day percent
2 2 2
1 1
Lab Location cm daIS (U} Location cm daxs !S2 (S•U} Number {S } of (S }
Estimated
Unseiked Sam2les Seiked sameles Spiking Seike Added Spike
Exposed Sulfation, Exposed Sul fat ion, Rate Sulfation
Area, Exposure 2 Area, Exposure 2 2 Weight, 2 Recovery (E),
Candle Time, mg/cm -day Candle Time, mg/cm -day mg/cm -day Code mg/cm -day percent
Lab Location cm2 daxs {U} Location cm2
daxs {Sl {S-U} Number mg so2 {S '} of {S '}
Analysis of Reproducibility
the number of degrees of freedom (df) associated with the statistical analysis,
the mean square (MS) of the sulfation rate measurements, the between-laboratory
standard deviation (sb)' and the coefficient of variation (CV).
Between-Laboratory Variation
m, MS (b) Sb'
2 2 2 2
Site n mg/cm -da~ df (mg/cm -day) mg/cm -day CV(%)
- - 2
l.: n. (x. -x)
MS = ~~1~-1~~ Equation (1)
df
The mean square calculated from Equation (1) includes both the
within- and between-laboratory variance. The relationship is described
by the equation
MS Equation (2)
2 2
where, s denotes the within-laboratory variance, sb denotes the between-
w
laboratory variance, and K is the number of replicate measurements by each
laboratory( ). The between-laboratory variance (s~) is determined by
6
substitution of values for MS, s , and Kin Equation (2). Duplicate deter-
w
minations were performed by each laboratory at each site, therefore K = 2.
The reproducibility of the Test Method, defined in this report
as the standard deviation (sb) of the sulfation rate measurements between
laboratories, is computed by extracting the square root of the between-
laboratory variance. Another measure of reproducibility is provided by the
coefficient of variation (CV) computed from the following equation
>.
Ill
reI
-
NEi
u
bO
Ei 0
0.0020
"'):i
Cll
'-' ,.
Cll ;
, ;
-1-1 ;
;
i:: ;
~Q)
;
,,
;
;
1-1
::l
Cll
Ill 0.0015
Q)
~
>.
1-1
0
-1-1
Ill
1-1
0
~
...:I
I
i:: 0.0010
Q)
Q)
~
-1-1
Q)
IX1
4-1
0
i::
0
•rl
-1-1
Ill 0.0005
•rl
>
Q)
~
re
1-1
Ill
'g
Ill
-1-1
tl.l
Analysis of Repeatability
MS = df
CV = 100 s w/m.
20
Within-Laboratory Variation
m, MS (b) SW'
2 2 ' 2 2
Site n mg/cm -day df {mg/ cm -day2 mg/cm -day CV{%2
Table 6 shows, for each site, the number (n) of sulfation rate measurements
performed, the mean (m) of the sulfation rate of all laboratories, the number
of degrees of freedom (df) associated with the statistical analysis, the mean
square (MS) of the sulfation rate measurements, the repeatability (s ), and
w
the coefficient of variation (CV).
The site values for repeatability are plotted versus the mean
sulfation rate in Figure 5. A curve of the form ~ = b-{;: was fitted to the
data points by the method of weighted least squares using the weighting formula,
W = f/m. The resultant curve is characterized by the equation~= 0.00504......;-;;.
2
The standard deviation of the residuals is 0.000056 mg/cm -day. The curve
sunnnarizes the relationship between repeatability and sulfation rate based on
the limited number of observations which were made.
The last line of Table 6 sunnnarizes the pooled data from the three
2
test sites. The mean sulfation rate is 0.00630 mg/cm -day and the over-all
2
repeatability is 0.00034 mg/cm -day. The repeatability expressed as a per-
centage of the mean value is five percent.
21
0.0025
:>-
QI
"O
I
-
N
13
tJ
~
0.0020
"""
Cll
::i:
.....,,
Cll
.u
i::
~
QJ
""::l
Cll
QI 0.0015
~
:>-
""0
.u
QI
""0
~
~
I
~
i:: 0.0010
,.c:::
.u
•r-1
:::.:
4-1
"O
""QI
'8QI
.u
CJ)
Analysis of Accuracy
(a) Column headings: n, number of measurements; m, mean spike recovery; df, degrees of
freedom; MS, mean square; crb' between-laboratory standard deviation of recovery; crw'
within-laboratory standard deviation of recovery; CV, coefficient of variation.
(b) Mean square contains both within- and between-laboratory variations and is related to
these variations by the equation MS = crw2 + Kcrb 2 where K=2.
(c) Data from Laboratory P at Los Angeles were excluded.
1
(d) These variations could not be calculated because the mean square of between-laboratory
measurements is less than the mean square of within-laboratory measurements. The model
for computing crb does not hold in this case.
24
--
:>..
Cll
"O
N
sCJ 80%
bO
s
("") 3
I
0
.-1
-::::>
Ul
I
...._,
Q)
.w 60%
Cll
P::
bO
s::
......
~
......
p..
Ul 2
"O
Q)
.µ
Cll
s
•rl
.µ
tll
J:;i::I 40%
pl (2.84.0)
•p
1
1
1 2 3 4
80%
(\")
3
I
0
r-1
Recovery (94%)
60%
40%
(2.26, 0.01)
1
1 2 3 4
--
:;....
Cll
'C / 95%
N 34 /
sCJ /
/
bD /
s /
('\"')
I /
0 /
...-!
/
....... ean Recover~ (96%) /
:=> / 90%
I
(/)
._, /
/
.µ
Q) 32 /
Cll
i:i::: /
-rl
~
bD
i:::: J 3,,,.
/
-rl /
p. /
(/)
/
't:I / 85%
Q)
.µ /
Cll
/
!I.µ 30 /
Cll /
IJ;l
/
/
/
/
/ 80%
3 /
/
/
/
/
28 "-~~~~~_...~ ......~~~~.........~~~_.._~~~--~~~~---~~~---~~~~
28 30 32 34 36
-3 2
Actual Spiking Rate (R), 10 mg/cm /day
Sb = 0.0136-v;;-
where, sb, and, m, the mean sulfation rate are given
in mg/cm2-day.
28
s = 0.00504 ~
w
RECOMMENDATIONS
ACKNOWLEDGEMENTS
REFERENCES
(1) Annual Book of ASTM Standards, Part 23, American Society for Testing
and Materials, 1915 Race Street, Philadelphia, Pennsylvania 19103.
(6) Kempthorne, o., "The Design and Analysis of Experiments", John Wiley and
Sons, Inc., pgs. 103-110 (1952).
APPENDIX
REPRINT OF ASTM
and
STANDARD METHODS OF TEST FOR SULFATE ION IN WATER AND WASTE WATER
35
D 2010
candle. The Sampling frequency should be not exceed 3 percent gum tragacanth. Warm
uniform and may be determined by the re- the mixture gently on a low-temperature hot
quirements of the survey. Monthly, bi- plate until a clear, uniform gel has been ob-
monthly, and seasonal sampling periods have tained. Take care not to overheat the gum.
been shown to provide consistent and reliable 5.3.1 Prepare batch quantities of paste (3)
data (5). by adding the desired weight of lead peroxide
4.2 Sampling Station-The sampling sta- to the glue, in small portions, with continuous
tion may be a louvered box as described in stirring to make the paste entirely free from
Fig. 1 (I). The dimensions of the louvered lumps. The paste then may be applied to the
area on each side should be not less than 203 support to provide the minimum reactive sur-
mm (8 in.) and the louvers should be placed at face.
an angle of 45 deg to provide maximum 5.3.1.1 When large quantities of paste are
shelter from rainfall. The box may be made made care must be taken to maintain an even
of metal or wood. If it is of metal it should be dispersion of the lead peroxide reagent and
of material that will have a low order of reac- excessive heating of the mixture avoided. The
tivity with sulfuric acid and sulfur dioxide. viscosity of the paste can be controlled by
When wood is used it may be shellacked or using an adequate water bath (6). When
varnished for protection. The use of a lead single lead peroxide candles are prepared (I)
base paint is not advised. The center of the 8 g of the lead peroxide shall be triturated in
louvered box should be fitted with a suitable sufficient glue to make paste of desirable
holder for the sampling device. working consistency. This may be done con-
· 4.2.I Sampling stations should be selected veniently with a spatula on a glass plate. The
at random on a uniform network grid over same spatula, or a stiff bristled brush I in.
the area to be studied and the density of the (25 mm) wide can be used to spread the paste
sampling stations should be not less than two on the support.
per square mile. Spacing of sampling stations 5.3.1.2 When quantity batches of candles
should be uniform. are prepared a small, hand-operated centri-
4.3 Location of Sampling Device-The box fuge may be adapted to facilitate the task.
should be located in a manner that will assure This is done by removing the tube holders and
protection from tampering, and it should be fastening to the rotating shaft an appropriate
secured from falling. The height of the box device for securing the candle (6).
from ground level should be the same at all
stations. 6. Procedure
6.1 Apply the lead peroxide paste to a
5. Preparation of Lead Peroxide Candle fabric binder on the support (I). Wash a
5.1 Support of Reactive Surface-A suit- fabric subsurface, such as tapestry cloth (I),
able support for the reactive surface may be a or stockinette (5) in boiling water and dry in
glass jar, test tube, plastic container, or any a sulfur dioxide-free atmosphere. Secure an
inert, impervious cylinder that conveniently adequate area of fabric to the support by
provides the desirable area. means of cotton thread (1) or rubber bands
5.2 Reactive Surface-There shall be a (3) and then apply the peroxide paste to the
minimum reactive surface of 100 cm 2 con- fabric.
taining not less than 8 g of lead peroxide. The 6.2 Drying of the coating must be accom-
weight of lead peroxide per candle in any plished in a sulfur dioxide-free atmosphere.
batch should not vary more than 10 percent. Retain a blank, unexposed candle from every
5.3 Bonding Paste-Gum tragacanth glue batch of candles prepared. Keep candles in
has been found satisfactory (l-5) and may be sealed containers, away from exposure to
prepared as follows: Disperse a sufficient sulfur dioxide or other gases that would con-
weight of good grade powdered gum traga- taminate the reactive surface, until used. At
canth in five times its weight of ethyl alcohol the end of the exposure period return the can-
(95 percent) and add hot water carefully, with dles to containers that can be sealed against
st1rnng, until a 2 to 3 percent' mixture is further contamination.
formed. The concentration of the glue shall 6.3 Treatment of Exposed Candles-
36
D 2010
Measure the surface area at this time. Sepa- ance with the referee ~ethod of ASTM
rate the impregnated cloth surface from the Methods D 516, Test for Sulfate Ion in In-
support, using a spatula or knife point, if nec- dustrial Water and Industrial Waste Water. 3
essary. The fabric may be cut into smaller Barium sulfate precipitates are slightly soluble
pieces. Transfer the lead peroxide-covered in dilute HCl (7). The precipitation with
fabric to a 250-ml beaker containing 5 g of BaCl 2 , therefore, must be performed in a di-
Na 2C0 3 di.ssolved in 60 ml of water. Allow lute acid solution of 0.05 N HCl for optimum
the immersed pieces to soak for 3 h, with results. It should be borne in mind during this
occasional stirring. Then simmer the mixture phase of the analysis that the rapid addition
gently on a thermostatically controlled hot of a hot (if possible, boiling) solution of
plate for 30 min. With reasonable care this BaCl 2 to a gently boiling solution of the sul-
operation can be conducted with minimum fate in dilute acid will yield a granular and
evaporation of water when properly covered easily filterable barium sulfate precipitate.
beakers are used. However, care should be
taken to maintain an approximately constant 7. Precision
volume. Filter the beaker contents through a 7.1 The standard deviation from the mean
fast filter paper, with appropriate washings, of data using this method has been found to
and neutralize the filtrate with 2N HCl to a be 7 percent (8 ).
pH range from 3.0 to 4.0 with methyl orange.
Care should be taken to prevent loss of 8. Report
sample by foaming, particularly when the 8.1 Report the results as milligrams of
point of neutralization is approached. sulfur dioxide per square centimeter per day,
6.4 Determination of Sulfate as Barium taking into consideration the possibility of the
Sulfate-Determine the sulfate ion in accord- interferences described in Section l.
REFERENCES
(I) Department of Scientific and Industrial Re- Annual Meeting, May 22-26, 1960.
search, Atmospheric Pollution Research, Tech- (5) Keagy, D. M., Stalker, W. W., Zimmer, C. E.,
nical Paper 1, pp. 20-23, H. M. Stationery Of- and Dickerson, R. C., "Sampling Station and
fice, London, 1945. Time Requirements for Urban Air Pollution
(Z) Measurement of Sulfur Dioxide with the Lead- Surveys; Part 1: Lead Peroxide Candles and
Peroxide Instrument, Investigation of Atmos- Dustfall Collectors," Paper No. 60-14, Air Pol-
pheric Pollution, Department of Scientific and lution Control Association, APMPA Cincin-
Industrial Research, Fuel Research Station, nati, Ohio, 53d Annual Meeting, May 23, 1960.
1948. (6) Levadie, B., Observations, Vermont Industrial
(3) Foran, M. R., Gibbons, E. V., and Wellington, Hygiene Laboratory, Barre, Vt.
J. R ., "The Measurement of Atmospheric 17) Kolthoff, I. M. and Sandell, E. B., Quantitative
Sulfur Dioxide and Chlorides," Chemistry in Inorganic Analysis, Macmillan Co., New York,
Canada, CH CAA, May, i958. N.Y.
(4) Thomas, F. W. and Davidson, C. M., "Moni- (8) Hickey, Jr., H. R., Studies on Two Types of
toring Sulfur Dioxide with Lead Peroxide Cyl- Lead Peroxide Cylinders for Monitoring At-
inders," Paper No. 56-1, Air Pollution Control mospheric Sulfur Dioxide, Thesis, University of
Association, APMPA, Cincinnati, Ohio, 53rd Florida, 1962.
37
~~l~ D 2010
1
....------- 7 "--------1...il
I !
le
!'
By publication of this standard no position is taken with respect to the validity of any patent rights in connection there-
with, and the American Society for Testing and Materials does not undertake to insure anyone utilizing the standard
against liability for infringement of any Letters Patent nor assume any such liability.
38
~~l~ Designation: D 516 - 68f
39
D 516
40
D 516
cable over the range of 10 to 100 mg/liter volume of glycerin with I volume of water.
(ppm) of sulfate ion (SO 4 - - ). NoTE 5-A stabilizing solution containing so-
dium carboxymethylcellulose ( 10 g/liter) may be
13. Summary of Method used instead of the glycerol solution.•
13.1 Sulfate ion is converted to a barium 16.3 Sodium Chloride Solution (240
sulfate suspension under controlled condi- g I liter)-Dissolve 240 g of sodium chloride
tions. Glycerin solution and a sodium chloride (NaCl) in water containing 20 ml of concen-
solution are added to stabilize the suspension trated hydrochloric acid (HCI, sp gr 1.19),
and minimize interferences. The resulting tur- and dilute to I liter with water. Filter the so-
bidity is determined by a photoelectric color- lution if turbid.
imeter or spectrophotometer and compared to 16.4 Sulfate, Standard Solution (I ml =
a curve prepared from standard sulfate solu- 0./00 mg so.- -)-Dissolve 0.1479 g of an-
tions. hydrous sodium sulfate (Na 2S0 4 ) in water,
and dilute with water to I liter in a volumetric
14. Interferences flask. Standardize by the procedure pre-
14.1 Insoluble suspended matter in the scribed in Section 9.
sample must be removed. Dark colors that
can not be compensated for in the procedure 17. Calibration
interfere with the measurement of suspended 17.1 Follow the procedure given in Section
barium sulfate (BaS0 4 ). 18, using appropriate amounts of the standard
14.2 Although other ions normally found in sulfate solution prepared in accordance with
water do not appear to interfere, the forma- 16.4, and prepare a calibration curve showing
tion of the barium sulfate suspension is very sulfate ion content in milligrams per liter
critical. This method is more suitable as a plotted against the corresponding photometer
control procedure where concentration and readings (Note 6 ). Prepare standa.-ds by di-
type of impurities present in the water are luting with water 0.0, 2.0, 5.0, IO.O, 15.0,
relatively constant. Determinations that are in 20.0, 30.0, 40.0, and 50.0 ml of standard sul-
doubt should be checked by the Method A in fate solution to 50-ml volumes in volumetric
some cases, or by the procedure suggested in flasks. These solutions will have sulfate ion
Note 7. concentrations of 0.0, 4.0, IO.O, 20.0, 30.0,
40.0, 60.0, 80.0 and 100.0 mg/liter (ppm),
15. Apparatus respectively.
15.1 Photometer-A filter photometer or NOTE 6-A separate calibration curve must be
spectrophotometer suitable for measurements prepared for each photometer and a new curve must
be prepared if it is necessary to change the cell,
between 350 and 425 nm, the preferable wave- lamp, or filter, or if any other alterations of instru-
length range being 380 to 400 nm. The cell for ment or reagents are made. Check the curve with
the instrument should have a light path each series of tests by running two or more solu-
through the sample of approximately 40 mm, tions of known sulfate concentrations.
and should hold about 50 ml of sample. Filter 18. Procedure
photometers and photometric practices pre-
scribed in this method shall conform to 18.1 Filter the sample if it is turbid, and
ASTM Recommended Practice E 60, Photo- adjust the temperature to between 15 and 30
metric Methods for the Chemical Analysis of c.
Metals•; spectrophotometers shall conform to 18.2 Pipet into a 200-ml beaker 50 ml or
ASTM Recommended Practice E 275, for less of the clear sample containing between
Describing and Measuring Performance of 0.5 and 5 mg of sulfate ion (Note 7). Dilute
Spectrophotometers. 5 to 50 ml with water if required, and add IO.O
•Annual Book of ASTM Standards, Part 32.
16. Reagents 'Annual Book of ASTM Standards, Part 30.
'The following commercial reagents also have been
16.1 Barium Chloride-Crystals of barium found suitable: Colloresine L V, obtainable from the Gen-
chloride (BaCl,·2H 2 0) screened to 20 to 30- eral Aniline and Film Corp., New York, N. Y.; or from
the Irwin Dyestuff Corp .. Ltd., Montreal, Canada: or Her-
mesh. cules CMC-70 Premium Low, obtainable from Hercules
16.2 Glycerin Solution (/+I )-Mix Incorporated, Wilmington, Del.
41
D 516
42
D 516
mg/liter when the sulfate present is low I volume of ammonium hydroxide (NH,OH,
(about 5 mg/liter). The noninterfering concen- sp gr 0.90) with 99 volumes of water.
tration of chloride increases with increasing 25.3 Barium Chloride, Standard Solution
sulfate content. (I ml = 0.500 mg SO 4 - - )-Dissolve 1.221 g
23.7 Chromium present as chromates and of barium chloride (BaCl2·2H 20) in I liter of
dichromates is converted by treatment with water that has been adjusted to pH 3.8 to 4.0
hydrogen peroxide to the cation, Cr+++, with dilute HCI. Standardize in accordance
which is then removed by ion exchange. with 26. I and 26.2, against standard sodium
sulfate solutions that have been passed
24. Apparatus through the ion-exchange resin.
24.1 Titration Assembly-For high accu- 25.4 Hydrochloric Acid (I +4)-Mix I
racy and determination of low SO, - - con- volume of concentrated hydrochloric acid
centration, a microburet reading to 0.0 I ml is HCI, (sp gr 1.19) with 4 volumes of water.
necessary. Efficient magnetic stirring im- 25.5 Hydrochloric Acid (I +99)-Mix I
proves the speed and convenience of titration. volume of HCI (sp gr 1.19) with 99 volumes
24.2 Ion Exchange Column-A suitable of water.
continuous flow column may be prepared by 25.6 Hydrogen Peroxide (30 percent)-
pouring 30 cm of a washed, wet resin, 20 to Concentrated hydrogen peroxide (H20 2).
25-mesh size into a glass column of 9 to 10- 25.7 Iodine, Standard Solution (I ml =
mm inside diameter and 500 ± 5 mm in 0.480 mg SO• - - )-Dissolve 10 g of potas-
length, the top of which widens to a reservoir sium iodide (Kl) in 100 ml of water, add 1.27
of 50 to 55-mm inside diameter and 100 ± 5 g of iodine crystals, and stir until solution is
mm in length. complete. Dilute to I liter with water and
24.2.1 The resin is held on a suitable screen store in a dark bottle. Standardize against
or filter plug and the flow is controlled by a 0.01 N Na 2S20 3 solution.
length of 2-mm bore capillary tubing joined to 25.8 Ion-Exchange Resin"-A cationic
the bottom of the column. This tubing is bent exchange resin, 20-- to 25-mesh or 100 to 200-
into a U shape and rises to about 13 mm mesh.
above the resin bed and then makes a U bend 25.9 Magnesium Carbonate (MgC0 3 ). -
downward for about 50 to 70 mm. 25.10 Phenolphthalein Indicator Solution
24.2.2 Other designs of exchange columns (5.0 g/liter)-Dissolve 0.5 g of phenol-
are suitable such as a small glass column, phthalein in 50 ml of 95 percent ethyl alcohol. 8
using only 25 mm of washed resin, 100 to Dilute to 100 ml with water.
200-mesh size, 65 to 70-mm total length and 25.11 Potassium Dichromate' 0 -Heat po-
15-mm inside diameter widening to a reser- tassium dichromate (K 2Cr 20,) in a platinum
voir cup 50 to 55 mm in length and 30 to 35- crucible to a temperature just above its fusion
mm inside diameter. This column is useful for point (396 C), taking care to exclude all dust
very small samples which must be washed and organic matter. After cooling, crush the
through the bed with water. Correction must fused salt to a powder in an agate mortar and
be made for the amount of wash water used. preserve in a glass-stoppered bottle.
24.3 The exchange columns shall be regen- 25.12- Potassium Iodide ' 0-This should not
erated when about two thirds exhausted. Re- yield a blue color when I g is dissolved in
generation shall be carried out by passing freshly boiled reagent-grade deaerated water
HCI (I +4) through the resin column and treated with 5 drops (0.25 ml) of I N sulfuric
thorough washing with water. If the resin acid (H 2SO,) and I ml of freshly prepared
column shows no visual change as exhaustion starch solution.
proceeds, it is advisable to regenerate after 25.13 Sodium Bicarbonate' 0 (NaHC0 3 ).
one or two samples have passed through it. 25.14 Sodium Carbonate, ' 0 anhydrous
43
D 516
44
D 516
NOTE 12-When the amount of sample is lim- 29. Procedure in the Presence of Negligible
ited, the sample may be passed through the small Interferences
ion exchange column described in 24.2 and rinsed
through with four or five times its volume of water, 29.1 If interfering cations are low and high
so that the final el utriate is 50.0 ml. Concentrate accuracy is not requi~ed, as in certain control
this elutriate to 10.0 ml, or take a 10.0-ml aliquot.
tests, directly titrate 10.0 ml of the filtered
27.4 Add 40 ml of alcohol and 2 drops of sample in accordance with the procedure de-
thorin indicator. Adjust the pH to 3.8 to 4.0 scribed in 27.4 and 27 .5.
by carefully adding dropwise NH ,OH (I +99)
until the solution just turns pink (Note 13). 30. Calculation
Then add HCI (I +99) dropwise until the pink 30.1 Calculate the sulfate ion (SO, - - )
color disappears. concentration in the original sample, in milli-
NOTE 13-lf the NH,OH is added too fast, it is grams per liter, as follows:
possible to overrun the color change from yellow to
pink and the sample continues to be yellow. It is Sulfate, mg/liter (ppm)
then impossible to develop the pink color by addi- [(V, - B,) x 500]/S,
tion of NH,OH. - [(V, - 8 2 ) x 480]/S,
By publication of this standard no position is taken with respect to the validity of any patent rights in connection there-
with, and the American So<iety for Testing and Materials does not undertake to insure anyone utilizing the standard
against liability for infringement of any Letters Patent nor assume any such /iabi/it.v.
45