Você está na página 1de 23

19

CHAPTER 2

CHARACTERIZATION TECHNIQUES

2.1 INTRODUCTION

The characterization of materials is important for understanding their


properties and applications. This chapter describes the instruments and experimental
methods utilized for various measurements towards the characterization of the
synthesized TiO2 nanostructures. The techniques adopted to characterize the
nanoparticles are: X - ray diffraction (XRD), UV - visible spectroscopy, Fourier
transforms infrared spectroscopy (FT-IR), scanning electron microscopy (SEM),
transmission electron microscopy (TEM) and surface area analysis (BET). The
fabrication of DSSC is also discussed.

Spectroscopy is a powerful tool to study the structure of


nanocrystalline, organic and inorganic materials. Spectroscopy is a technique that
uses the interaction of energy with a sample to perform an analysis. The data
obtained from spectroscopy is called a spectrum. A spectrum is a plot of the intensity
of energy detected versus the wavelength (or mass or momentum or frequency, etc.)
of the energy [59].

A spectrum can be used to obtain information about atomic and


molecular energy levels, molecular geometries, chemical bonds, interactions of
molecules and related processes. Often, spectra are used to identify the components
of a sample (qualitative analysis).
20

Table 2.1 Characterization techniques and their applications

TECHNIQUES APPLICATIONS
1. X-Ray Diffraction (XRD) To study the crystalline properties of solid
substances

2.Ultraviolet-Visible To analyze molecular (organic) and ionic species


Spectroscopy (UV/Vis) capable of absorbing at UV or Visible
wavelengths in dilute solutions
To analyze only molecular compounds (organic
3. Fourier Transform Infrared
compounds, natural products, polymers, etc.)
Spectroscopy (FT IR)

4.Scanning Electron Microscopy To study the topography, structure and


(SEM) compositions of metals, ceramics, polymers,
composites and biological materials

To study the local structures, morphology, and


5. Transmission Electron
dispersion of multi component polymers, cross
Microscopy (TEM)
sections and crystallizations of metallic alloys,
semiconductors, microstructure of composites,
etc.
6. Surface area analysis (BET) To study the surface area, pore volume, pore
diameter, and pore size distribution of
nanoparticles.

Spectra may also be used to measure the amount of material in a sample


(quantitative analysis). Table 2.1 shows the various characterization techniques and
their applications. This chapter also encompasses a detailed view of the theoretical
aspects, instrumentation techniques of the three major spectroscopic methods
namely powder X – ray Diffraction, UV-visible, and Fourier Transform Infrared
spectroscopy (FT - IR).

The next level of characterization is to examine the surface, arrangement


of surface atoms and the electronic structure of the nanoparticles. This can be done
only by viewing the particles in the nanometre range. There are two main kinds of
21

microscopy. The first class of microscopy involves a stationary sample in line with a
high - speed electron gun. Both the scanning electron microscope (SEM) and
transmission electron microscope (TEM) are based on this technique. Detailed
information of surface area measurement and pore size measurement of BET
analysis has described in this chapter.

2.2 POWDER X - RAY DIFFRACTION (XRD)

2.2.1 Principle

Figure 2.1 shows a schematic representation of powder X - ray


diffractometer. X-rays are electromagnetic radiation with wavelength in the range
0.01 - 10 nm. It is used in diffraction experiments in the typical wavelength range
0.5 - 1.5 A. For electromagnetic radiation to be diffracted, the spacing in the grating
should be of the same order as the wavelength. In crystals, the typical inter atomic
spacing 2 - 3Å the suitable radiation is X-rays. Hence X-rays can be used to the
study crystal structures [60].

Figure 2.1 Powder X - ray diffractometer

In this technique, the primary X - rays are made to fall on the sample
substance to be investigated. Because of its wave nature, like light waves, X - ray
gets diffracted to a certain angle. This angle of diffraction, which differs from that of
the incident beam, will give the information regarding the crystal nature of the
substance. The wavelength of the X - rays can be varied for the application by using
a grating plate.
22

2.2.2 Bragg’s Law

Bragg diffraction occurs when electromagnetic radiation or subatomic


particle waves with wavelength comparable to atomic spacings are incident upon a
crystalline sample, scattered by the atoms in the system and undergo constructive
interference in accordance to Bragg's law.

Figure 2.2 Bragg’s Law

Considering the conditions necessary to make the phases of the beams


coincide when the incident angle equals and reflecting angle (Figure 2.2). The rays
of the incident beam are always in phase and parallel up to the point at which the top
beam strikes the top layer at atom Z. The second beam continues to the next layer
where it is scattered by atom B. The second beam must travel the extra distance
AB + BC if the two beams are to continue traveling adjacent and parallel. This extra
distance must be an integral (n) multiple of the wavelength (λ) for the phases of the
two beams to be the same:

nλ = AB+BC (1)

Recognizing ‘d’ as the hypotenuse of the right triangle ABZ, we can use
trigonometry to relate ‘d’ and θ to the distance (AB + BC). The distance AB is
opposite θ so,

AB = d sinθ (2)

Because AB = BC eq. (1) becomes,


23

nλ = 2AB (3)

Substituting eq. (2) in eq. (3) we have,

nλ = 2d⋅sinθ (4)

and Bragg's Law has been derived.

where,

 n - order of diffraction

 λ - the wavelength of x-rays

 d - the spacing between the planes in the atomic lattice.

 θ - the angle between the incident ray and the scattering planes.

Note that if only two rows of atoms are involved, the transition from
constructive to destructive interference as θ changes is gradual. However, if
interference from many rows occurs then the constructive interference peaks become
very sharp with mostly destructive interference in between. This sharpening of the
peaks as the number of rows increases that is very similar to the sharpening of the
diffraction peaks from a diffraction grating as the number of slits increases.

2.2.3 Instrumentation

A powder X - ray diffractometer consists of a X - ray source (usually a


X - ray tube), a sample stage, a detector and provision to change angle θ
(Figure 2.3). The X - ray is focused on the sample at some angle θ, while the
detector opposite to the source reads the intensity of the X - ray that receives at
2θ away from the source path. The incident angle is the increased over time while
the detector angle always remains 2θ above the source path.
24

Figure 2.3 Powder X - ray diffraction

2.2.3.1 X - ray Tubes

The most common source of X - rays is an X - ray tube. The tube is


evacuated and contains a copper block with a metal target anode and a tungsten
filament cathode with a high voltage is applied between them. The filament is heated
by a separate circuit, and the large potential difference between the cathode and
anode electrons at the metal target. The accelerated electrons knock core electrons
out of the metal, and electrons in the outer orbitals drop down to fill the vacancies
and thereby, emitting X-rays. The X-rays exit the tube through a beryllium window.
Due to massive amounts of heat being produced in this process, the copper block
must usually be water cooled.

2.2.3.2 X-ray Detectors

Most modern equipment do use transducers that produce an electrical


signal when exposed to radiation. These detectors are often used as photon counters,
so intensities are determined by the number of counts in a certain period of time.
25

2.2.3.3 Gas-Filled Transducers

A gas-filled transducer consists of a metal chamber filled with an inert


gas, with the walls of the chamber as a cathode and a long anode in the center of the
chamber. When X - rays enters the chamber, its energy ionizes many molecules of
the gas. The free electrons then migrate towards the anode and the cations migrate
towards the cathode. The electrons that reach the anode cause current to flow, which
can be detected. The sensitivity and dead time (when the transducer will not respond
to radiation) both depend on the voltage of the transducer is operated at. At high
voltage, the transducer is very sensitive but it has a long dead time, whereas at low
voltage, the transducer has a short dead time but low sensitivity.

2.2.3.4 Scintillation Counters

In a scintillation counter, phosphor is placed in front of a photomultiplier


tube. When X - rays strike on the phosphor, it produces flashes of light, which are
detected by the photomultiplier tubes.

2.2.3.5 Semiconductor Transducers

A semiconductor transducer has a gold coated p-type semiconductor


layered on lithium containing semiconductor intrinsic zone, followed by an n-type
semiconductor on the other side of the intrinsic zone. The semiconductor is usually
composed of silicon and germanium but germanium is used if the radiation
wavelength is very short. The n - type semiconductor is coated by an aluminum
contact, which is connected to a pre-amplifier. The entire crystal has a voltage
applied across it. When X - ray strikes the crystal, it elevates many electrons in the
semiconductor into the conduction band, which causes a pulse of current [61, 62]

2.2.4 Applications

The X - ray diffraction is a good tool to study the nature of the crystalline
substances. In crystals, the ions or molecules are arranged in well-defined positions
in planes in three dimensions. The impinging X - rays are reflected by each crystal
26

plane. So the planes can not be same or identical for any two chemical substances,
this technique provides vital information regarding the arrangement of atoms and the
spacing in between them and also to find out the chemical compositions of
crystalline substances. The sample under study can be of either a thin layer of crystal
or in a powder form. Since, the power of a diffracted beam depends on the quantity
of the corresponding crystalline substance. It is also possible to carry out
quantitative determinations [63].

2.3 ULTRAVIOLET - VISIBLE SPECTROSCOPY

2.3.1 Principle

Molecules absorb energy and this energy can bring out translational,
rotational or vibrational motion or ionization of the molecules depending upon the
frequency of the electromagnetic radiation. Excited molecules are unstable and
quickly drop down to ground state again giving off the received energy in the form
electromagnetic radiation. The wavelength and intensity of the electromagnetic
radiation absorbed or emitted can be recorded to get a spectrum (Figure 2.4).
Spectral analysis yields qualitative and quantitative information about the materials
under study [64].

Figure 2.4 Electromagnetic spectrum


27

Ultraviolet-visible (UV-vis) spectroscopy is useful tool to characterize


the absorption, transmission, and reflection of a variety of compounds and materials,
such as pigments, coatings etc. The UV - vis spectra has broad features including
sample identification and very useful for quantitative measurements.

2.3.2 Instrumentation

The instruments has the following components:

 A light source that generates a broad band of electromagnetic


radiation

 A dispersion device that selects a particular wavelength (or, more


correctly a waveband) from the broadband radiation of the source

 A sample area (component)

 One or more detectors to measure the intensity of radiation

 Other optical components, such as lenses or mirrors, relay light


through the instrument.

 A schematic representation of a UV/vis spectrophotometer is shown


in Figure 2.5. Normal working range for a spectrometer is
190 – 900 nm, working beyond 180 nm requires special
arrangements [65].

Figure 2.5 Functional block diagram of UV - visible spectrophotometer


28

2.3.2.1 The Light Source

 A deuterium discharge lamp for UV region (160-375 nm)

 A tungsten filament lamp or tungsten-halogen lamp for Visible and


NIR regions (350 - 2500 nm)

 The instrument automatically swaps lamps when scanning between


the UV and VIS-NIR regions

2.3.2.2 The Monochromator

All monochromators contain the following component parts:

 An entrance slit

 A collimating lens

 A dispersing device

 A focusing lens

 An exit slit

Ideally, the output from a monochromator is monochromatic light.


However, in practice, the output is always a band, optimally symmetrical in shape.

2.3.2.3 Dispersion devices

Dispersion devices cause different wavelengths of light to be dispersed at


different angles. When combined with an appropriate exit slit, these devices can be
used to select a particular wavelength (or, more precisely, a narrow waveband) of
light from a continuous source. Two types of dispersion devices, prisms and
holographic gratings are commonly used in UV-vis spectrophotometers.

Light falling on the grating is reflected at different angles, depending on


the wavelength. Holographic gratings yield a linear angular dispersion with
29

wavelengths and are temperature insensitive. However, they reflect light in different
orders, which may overlap. As a result, filters must be used to ensure that only the
light from the desired reflection order reaches the detector.

2.3.2.4 Detectors

A detector converts a light signal into an electrical signal. Ideally, it


should give a linear response over a wide range with low noise and high sensitivity.
Spectrophotometers normally contain either a photomultiplier tube detector or a
photodiode detector.

The photomultiplier tube combines signal conversion with several stages


of amplification within the body of the tube. It consists of a photoemissive cathode,
a number of dynodes (which emit several electrons for each electron striking them)
and an anode.

Photodiodes are increasingly being used as detectors in modern


spectrophotometers. Photodiode detectors have a wider dynamic range and are more
robust than photomultiplier tube detectors. In a photodiode, light falling on the
semiconductor material allows electrons to flow through it, thereby depleting the
charge in a capacitor connected across the material. The amount of charge needed to
recharge the capacitor at regular intervals is proportional to the intensity of the light.

2.3.2.5 Cells

These are containers for the sample and reference solutions. They must
be transparent to the radiation passing through.

 For UV region: Quartz or fused silica cuvettes are usually used.

 VIS/NIR regions: Silicate glass or plastic cuvettes (350 - 2000 nm)


can also be used [66].
30

2.3.3 Applications

It is the most widely used technique for quantitative molecular analysis


and obeys Beer - Lambert law. Sometimes, it is used in conjunction with other
techniques such as NMR, IR, etc., in the identification and structural analysis, of
organic compounds. For qualitative analysis it provides valuable information
through the absorption spectrum which is unique for a given compound [67 - 72].

2.4 SCANNING ELECTRON MICROSCOPY (SEM)

2.4.1 Principle

In this technique, an electron beam is focused onto sample surface kept in


a vacuum by electro-magnetic lenses (since electron possesses dual nature with
properties of both particle and wave, hence an electron beam can be focused or
condensed like an ordinary light). The beam is then rastered or scanned over the
surface of the sample. The scattered electron from the sample is then fed to the
detector and then to a cathode ray tube through an amplifier, where the images are
formed, which gives the information of the sample [73].

2.4.2 Instrumentation

It comprises of a heated filament as a source of electron beam, condenser


lenses, aperture, evacuated chamber for placing the sample, electron detector,
amplifier, CRT with image forming electronics, etc.

The SEM is an instrument that produces a largely magnified image by


using electrons instead of light to form an image. A schematic diagram of the
FE-SEM is shown in Figure 2.6. A beam of electrons is produced at the top of the
microscope by an electron gun. The electron beam follows a vertical path through
the microscope, which is held within a vacuum chamber.
31

Figure 2.6 Functional Block diagram of field emission scanning electron


microscope (FE-SEM)

The beam travels through electromagnetic fields and lenses, which focus
the beam down towards the sample. Once the beam hits the sample, electrons and
X - rays are ejected from the sample. Detectors collect these X - rays, backscattered
electrons and secondary electrons and convert them into a signal that is sent to a
screen similar to a television screen. This produces the final image. In this research
work, the powder samples were placed on the carbon tape which is attached to the
sample holder. JEOL JSM 6320F (FESEM), F E I Quanta FEG 200 (HRSEM) are
used to study the surface morphology of the sample.
32

2.4.3 Applications

Scanning electron microscopy has been applied to the surface studies of


metals, ceramics, polymers, composites and biological materials for both topography
as well as compositional analysis. An extension of this technique is Electron Probe
Micro Analysis (EPMA), where the emission of X-rays, from the sample surface, is
studied upon exposure to a beam of high energy electrons. Depending on the type of
detectors used this method is classified in to two as: Energy Dispersive
Spectrometry (EDS) and Wavelength Dispersive Spectrometry (WDS).
This technique is used extensively in the analysis of metallic and ceramic inclusions,
inclusions in polymeric materials and diffusion profiles in electronic components.

2.5 ENERGY DISPERSIVE X - RAY ANALYSIS (EDAX)

Energy dispersive X - ray spectroscopy (EDS or EDX) is an analytical


technique used predominantly for the elemental analysis or chemical
characterization of a specimen. Being a type of spectroscopy, it relies on the
investigation of a sample through interactions between electromagnetic radiation and
matter, analyzing X - rays emitted by the matter in this particular case.
Its characterization capabilities are due in large part to the fundamental principle that
each element of the periodic table has a unique atomic structure allowing X - rays
that are characteristic of an element's atomic structure to be uniquely distinguished
from each other.

To stimulate the emission of characteristic X - rays from a specimen, an


high energy beam of charged particles such as electrons or protons or a beam of
X - rays is focused into the sample to be characterized. At rest, an atom within the
sample contains ground state (or unexcited) electrons situated in discrete energy
levels or electron shells bound to the nucleus. The incident beam may excite an
electron in an inner shell, prompting its ejection and resulting in the formation of an
electron - hole within the atom’s electronic structure. An electron from an outer,
higher - energy shell then fills the hole, and the difference in energy between the
higher - energy shell and the lower energy shell is released in the form of a X - ray. The
X - ray released by the electron is then detected and analyzed by the energy dispersive
33

spectrometer. These X - rays are characteristic of the difference in energy between the
two shells, and the atomic structure of the element form which they were emitted.

Figure 2.7 The principle of EDX

The excess energy of the electron that migrates to an inner shell (in order
to fill the newly - created hole) can do more than emitting a X - ray. Often, the
excess energy is transferred to a third electron from a further outer shell, prompting
its ejection. This ejected species is called an Auger electron and the method for its
analysis is known as Auger Electron Spectroscopy (AES).

Energy Dispersive X-Ray spectroscopy (EDS or EDX) is a technique


used in conjunction with chemical microanalysis by scanning electron microscopy
(SEM) (Figure 2.7). EDS technique detects X - rays emitted from the sample during
bombardment by an electron beam to characterize the elemental composition of the
volume analyzed.

When a sample is bombarded by the electron beam in SEM, the electrons


are ejected from atoms comprising the sample surface. EDS X - ray detector
measures the relative abundance of X - rays against their energy. The detector is
typically a lithium-drifted silicon solid-state device. When an incident X - ray hits
34

the detector, which creates a pulse of charge that is proportional to the energy of
X - ray. The pulse charge is converted to a pulse voltage (which is proportional to
the energy X - ray) by a charge sensitive preamplifier. The signal is then sent to a
multichannel analyzer where the pulses are sorted by the tension. The energy, as
determined by measuring the voltage per incident X - ray is sent to a computer for
display and further data evaluation. The spectrum of X - ray energy versus counts is
evaluated to determine the elemental composition of the sample volume.

2.6 TRANSMISSION ELECTRON MICROSCOPY (TEM)

2.6.1 Principle

In this technique, a beam of high-energy electrons (typically 100 – 400 keV)


is collimated by magnetic lenses and allowed to pass through a specimen under high
vacuum. The transmitted beam and a number of diffracted beams can form a
resultant diffraction pattern, which is imaged on a fluorescent screen kept below the
specimen. The diffraction pattern gives the information regarding lattice spacing and
symmetry of the structure under consideration. Alternatively, either the transmitted
beam or the diffracted beams can be made to form a magnified image of the sample
on the viewing screen as bright-and dark field imaging modes respectively. This
gives information about the size and shape of the micro-structural constituents of the
material. High - resolution image contains information about the atomic structure of
the material. This can be obtained by recombining the transmitted beam and
diffracted beams together [74, 75].

2.6.2 Instrumentation

It comprises of a tungsten filament or LaB6 or a field emission gun as


source of electron beam, objective lens, imaging lens, CCD camera, monitor, etc.
The ray of electrons is produced by a pin-shaped cathode heated up by current. The
electrons are vacuumed up by a high voltage at the anode. The acceleration voltage
is between 50 and 150 kV. The higher it is, the shorter are the electron waves and the
higher is the power of resolution, but this factor is hardly ever limiting. The power
of resolution of electron microscopy is usually restrained by the quality of the
35

lens-systems and especially by the technique with which the preparation has been
made. Modern gadgets have powers of resolution that range from 0.2 - 0.3 nm.

Schematic representation of TEM image is shown in Figure 2.8.


The accelerated ray of electrons passes a drill-hole at the bottom of the anode.
Its following way is analogous to that of a ray of light in a light microscope.
The lens-systems consist of electronic coils generating an electromagnetic field.
The ray is first focused by a condenser and then passes through the object, where it
is partially deflected. The degree of deflection depends on the electron density of the
object. The greater the mass of the atoms, the greater is the degree of deflection.
Biological objects have only weak contrasts since they consist mainly of atoms with
low atomic numbers (C, H, N, O). Consequently, it is necessary to treat the
preparations with special contrast enhancing chemicals (heavy metals) to get at least
some contrast. Additionally, they are not thicker than 100 nm, because the
temperature rises due to electron absorption. It is generally impossible to examine
living things.

Figure 2.8 Functional Block diagram of transmission electron microscope


36

After passing through the object, the scattered electrons are collected by
an objective. Thereby an image is formed, that is subsequently enlarged by an
additional lens - system (called projective with electron microscopes). The formed
image is made visible on a fluorescent screen or it is documented on photographic
material. Photos taken with electron microscopes are always black and white.
The degree of darkness corresponds to the electron density (differences in atom
masses) of the candled preparation.

2.6.3 Applications

Transmission electron microscopy is used to study the local structures,


morphology, dispersion of multi - component polymers, cross sections and
crystallization of metallic alloys semiconductors, microstructure of composite
materials, etc. The instrument can be extended to include other detectors like Energy
Dispersive Spectrometer (EDS) or Energy Loss Spectrometer (ELS) to study about
the local chemistry of the material similar to SEM technique [76]

2.7 FOURIER TRANSFORM INFRARED SPECTROSCOPY (FT-IR)

2.7.1 Principle

It involves the absorption of electromagnetic radiation in the infrared


region of the spectrum which results in changes in the vibrational energy of
molecule. Since, usually all molecules will be having vibrations in the form of
stretching, bending, etc. The absorbed energy will be utilized in changing the energy
levels associated with them. It is a valuable and formidable tool in identifying
organic compounds which have polar chemical bonds (such as OH, NH, CH, etc.)
with good charge separation (strong dipoles) [77 - 79].

2.7.2 Instrumentation

It was originally designed as a double beam spectrophotometer


comprising IR source (red hot ceramic material), grating monochromator,
thermocouple detector, cells made of either sodium chloride or potassium bromide
37

materials, etc. In this process, the light is dispersed by the monochromator but, this
type of basic design for IR measurements has been outdated. Instead, a newer
technique termed Fourier Transform-Infrared (FT-IR) has been in practice.
This technique utilises a single beam of un-dispersed light has the instrument
components similar to the previous one.

Figure 2.9 Functional Block diagram of Fourier transform infrared


spectrometer

In FT-IR, the un-dispersed light beam is passed through the sample and
the absorbance at all wavelengths is received at the detector simultaneously.
A computerized mathematical manipulation (known as “Fourier Transform”) is
performed on this data to obtain absorption data for each and every wavelength.
To perform this type of calculations interference of light pattern is required for
which the FT-IR instrumentation contains two mirrors; one fixed and one moveable
with a beam splitter in between them. Before scanning the sample, a reference or a
blank scanning is required. The following is the simplified design of the instrument
(Figure 2.9).
38

2.7.2 Applications

It finds extensive use in the identification and structural analysis of


organic compounds, natural products, polymers, etc. The presence of particular
functional group in a given organic compound can be identified. Since every
functional group has unique vibrational energy, the IR spectra can be seen as their
fingerprints.

2.8 BET ANALYSIS

2.8.1 Surface Area Determination

The surface of a material is the dividing line between a solid and its
surroundings, liquid, gas or another solid. Therefore, the amount of surface or
surface area is an important factor in the behavior of a solid. Surface area has strong
influence on many factors such as dissolution rates of pharmaceuticals, the activity
of industrial catalyst, adsorption capacity of air and water purifiers, and the
processing of most powders and porous materials, etc. Whenever solid matter is
divided into smaller particles, new surfaces are created thereby increasing the
surface area. Similarly, when pores are created within the particle interior
(by dissolution, decomposition or some other physical or chemical means) the
surface area is also increased. There may be more than 2000 m2 of surface area in a
single gram of activated carbon as an example for gas absorption. The true surface
area, including surface irregularities and pore interiors, cannot be calculated from
particle size information but is rather determined at the atomic level by the
adsorption of an unreactive gas or inert gas. The amount adsorbed, let’s call it X, is a
function not only of the total amount of exposed surface but also (i) temperature, (ii)
gas pressure and (iii) the strength of interaction between gas and solid. Because most
gases and solids interact weakly, the surface must be cooled substantially in order to
cause measurable amounts of adsorption - enough to cover the entire surface. Where
the gas pressure increases, more gas is adsorbed on the surface (in a non - linear
way). However, adsorption of a cold gas does not stop when it covers the surface in
a complete layer of one molecule thick (let’s call the theoretical monolayer amount
39

of gas Xm)! As the relative pressure is increased, excess gas is adsorbed to form
“multilayers”. [80, 81]

So, gas adsorption - as a function of pressure - does not follow a simple


relationship, and we must use an appropriate mathematical model to calculate the
surface area. We use the BET equation as follows:

1 1 C 1  P 
    (5)
x [( P0  P )  1] X mC X m C  P0 

where P/P0 is the gas’s relative pressure and constant ‘C’ is related to the strength of
interaction between gas and solid.

2.8.2 The Principle

The gas most commonly used is nitrogen for a number of reasons. In the
classical manometric technique, relative pressures less than unity are achieved by
creating conditions of partial vacuum (absolute pressures of pure nitrogen below
atmospheric pressure). This method is easily automated and the amount of gas
adsorbed is made at a number different relative pressures. Usually, the analyzer
obtains at least three data points in the relative pressure range between 0.025 and
0.30 Pa. Experimentally measured data are recorded as pairs of values: the amount
of gas adsorbed is expressed as STP volume (VSTP) and the corresponding relative
pressure (P/Po). A plot of these data is called an isotherm.

The Principle of calculation, the computer program takes over and a least
- squares linear regression is used to fit the best straight line through a transformed
data set consisting of the following pairs of values: 1/VSTP (Po/P)-1 and P/Po.
The monolayer capacity, Vm, is calculated from the slope,
40

C 1
S  (6)
VmC

and the intercept, i, of the straight line

1
i  (7)
VmC

Solving for Vm

1
Vm  (8)
si

The number of molecules in the monolayer is obtained through the


number of moles. Vm value is calculated by dividing Vm by the molar volume
(MV) for the number of moles. The number of molecules covering the surface in a
layer one molecule thick can be determined by multiplying moles by Avogadro’s
number. If we know how much area one molecule occupies, then the total area can
be calculated. Thus the area called as “cross - sectional area”. Therefore, the total
surface area, St, is then calculated from the below equation

V m L AV Am
St  (9)
M v

where LAV is Avogadro’s number and Am is the cross - sectional area. All surface
area results are finally reported normalized by sample weight, or mass, as square
meters per gram, written m2 /g.

2.9 SOLAR CELL FABRICATION

The DSSC consists of four components namely: photoanode, dye


molecules, electrolytes and photocathode.

For the DSSC fabrication, ruthenium (II) N - 719 dyes, iodide and
tri - iodide electrolytes and Pt - coated photocathodes were purchased from
41

commercially available sources and utilized. The surface of a photoanode was


modified with synthesized TiO2 nanostructures by using spray deposition method.
The fabrication of DSSC is described clearly here. 1.5 g of synthesized TiO2
nanostructures and 50 ml of ethanol were grounded in a mortar for few minutes to
form colloidal suspensions. Thereafter, five drops of triton - X were added to the
solution as an organic binder. Fluorine doped tin oxide (FTO) substrates were
cleaned ultrasonically using a mixture of acetone and ethanol.

Figure 2.10 I - V curve measurements system at Prof. Kenji Murakami


laboratory, Shizuoka University

The TiO2 nanostructures suspension in ethanolic solution was sprayed


over the FTO substrate at a substrate temperature of 150 ˚C by spray deposition.
TiO2 coated FTO substrates (photoanodes) were annealed at 450 ˚C for 2 h.
Photoanodes were immersed in ethanolic solution with 0.03 M di - tetrabutyl
ammonium is - bis (isothiocyanato) bis (2,2″- bipyridyl 4,4′dicarboxylato)
ruthenium (II) (N - 719). The dye sensitized photoanode and Pt - coated counter
electrode were clamped using clips. Finally, an iodide redox electrolyte was filled
between the electrodes via capillary action and DSSC’s device were subjected in
I - V instrument JASCO, CEP - 25BX, as shown in Figure 2.10.

Você também pode gostar