a r t i c l e in fo abstract
Article history: In a recent series of papers evidence has been presented for correlations between solar activity and
Received 29 December 2009 nuclear decay rates. This includes an apparent correlation between Earth–Sun distance and data taken
Received in revised form at Brookhaven National Laboratory (BNL), and at the Physikalisch-Technische Bundesanstalt (PTB).
12 March 2010
Although these correlations could arise from a direct interaction between the decaying nuclei and some
Accepted 18 March 2010
particles or fields emanating from the Sun, they could also represent an ‘‘environmental’’ effect arising
from a seasonal variation of the sensitivities of the BNL and PTB detectors due to changes in
Keywords: temperature, relative humidity, background radiation, etc. In this paper, we present a detailed analysis
Ionization chambers of the responses of the detectors actually used in the BNL and PTB experiments, and show that
Proportional counters
sensitivities to seasonal variations in the respective detectors are likely too small to produce the
Beta decay
observed fluctuations.
Gamma decay
Energy loss & 2010 Elsevier B.V. All rights reserved.
0168-9002/$ - see front matter & 2010 Elsevier B.V. All rights reserved.
doi:10.1016/j.nima.2010.03.129
Please cite this article as: J.H. Jenkins, et al., Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.03.129
ARTICLE IN PRESS
2 J.H. Jenkins et al. / Nuclear Instruments and Methods in Physics Research A ] (]]]]) ]]]–]]]
similar analyses of other past and future experiments on half-life bias voltage changes, since the ionization current is essentially
determinations to evaluate possible seasonal environmental independent of the applied voltage in the ion chamber region
influences. [12]. This is due to the fact that, as pointed out above, the bias
voltage applied in an ionization chamber is chosen to be
sufficiently high to prevent appreciable recombination of the
2. The PTB ionization chamber detector generated electron–ion pairs, while at the same time not
providing enough energy to cause multiplication of electron–ion
In the course of experiments at the PTB [5] which were pairs within the detector. It follows that the output current from
intended to evaluate the stability of detector systems employed in the detector will be based solely on the number of electron–ion
the measurement of half-lives, e.g. 152Eu (T1/2 ¼4936.6(20) d [11]), pairs generated by the incident radiation and collected at the
the authors reported an oscillation in the decay rate of the long- electrodes within.
lived standard, 226Ra (T1/2 ¼1600(7) y [11]) used in the measure- Since it is the ionization of the argon gas within the detector
ments. These fluctuations were attributed by the authors to a that provides the measure of the energy deposited within by the
‘‘discharge effect on the charge collecting capacitor, the cables and incident radiation, and thus the activity of the sources being
insulator to the ionization chamber electrode caused by back- analyzed, it follows that changes in the properties of the gas
ground radioactivity such as radon and daughters which are within the detector, such as the atom density, could affect the
known to show seasonal concentration changes [5]’’. It was also measurements. However, since the number of moles (n) of argon
proposed elsewhere [8] that the variations in the measured gas within the PTB detector is constant, the density n=V ¼ r of the
activity of the 226Ra source were caused by variations of the gas will also be constant, unless the volume (V) of the detector
density of the argon working gas within the detector chamber changes. Using the ideal gas law, PV ¼nRT, where R is the gas
itself, caused by seasonal changes in the ambient temperature. constant, T is the temperature, and P is the pressure, it follows
This and other potential systematic effects, such as changes in that if the volume is also constant (a non-constant volume will be
ambient pressure, humidity, background radiation or electronic addressed below) then the detector system is described by
fluctuations, are indeed possible explanations for the observed P1/T1 ¼ P2/T2, where 1 and 2 denote the initial and final conditions,
variations and should be examined. An alternative proposal to respectively. Hence, the only effect of a change in room
these explanations, as suggested in Ref. [2], was that the temperature is to change the pressure inside the detector. And,
oscillations of the 226Ra measurements were not caused by local since the ionization potential of argon is known to be a constant
perturbations to the detector system, but were due to the over all pressures from 1 to 200 bar [13], it follows that there
influence of fields or particles such as neutrinos emanating from should be no change in the response of the detector to changes in
the Sun. In what follows, we will examine possible environmental the room temperature if the volume of the detector is constant.
influences on the detector system, and determine which of these, We next consider the small effects that may arise from a
if any, could be responsible for the fluctuations in the 226Ra volume change due to expansion or contraction of the steel walls
measurements. of the PTB detector as the room temperature increases or
To evaluate which environmental influences may have caused decreases. The outer case and the bulkheads of the detector
the fluctuations in the detector, it is helpful to understand how chamber are manufactured using Mild Steel (Trevor Alloway,
the detector utilized in the PTB experiments works. The Centronic Ltd., private communication, 2009), and typical coeffi-
measurements were taken with a high-pressure 4pg ionization cients of linear expansion, a, for steel range from 9 106 to
chamber, Model IG12, manufactured by 20th Century (now 17 106 K1 . For our analysis here, we will use an average value
Centronic, Ltd.), which was filled with argon to 20 atm (2 MPa) of 13 10 6 K 1, and assume that the expansion of the metal is
by the manufacturer, then welded shut prior to delivery to the isotropic. Based on the stated dimensions of the detector, radius
user (Trevor Alloway, Centronic Ltd., private communication, 9.25 cm and length 42.7 cm, and neglecting the small volume of
2009). Therefore, the number of atoms of the working gas, argon, the re-entrant tube (well), the initial volume, V0, of the chamber
within the detector is constant throughout the lifetime of the at the initial temperature can be calculated to be 11,477:9 cm3 .
detector, irrespective of the external temperature, pressure or The change in volume, DV, arising from a change, DT, in
relative humidity. temperature is
The ionization chamber for the PTB experiments was operated
V0 þ DV ¼ p½r2 ð1 þ aDTÞ2 ½Lð1 þ aDTÞ ffi V0 ð1 þ3aDTÞ ð1Þ
in a ‘‘current’’ mode, as is typical with ionization chambers in this
3
application. Here, the output current of the detector is generated and hence the new volume of the cylinder will be 11,478.3 cm for
by the collection of electron–ion pairs created by the ionization of a 1 3 C increase in temperature in the room. The change in volume
the argon gas inside. These electron–ion pairs arise from photon is thus 4 parts in 100,000, and would then result in a fractional
interactions within the gas, and from the interaction of photo- density change of 3.5 10 5, which is two orders of magnitude
electrons and Compton electrons generated by photon interac- smaller than the density change proposed in Ref. [8] to explain the
tions in the wall of the re-entrant tube, which then enter the PTB data. Additionally, it should be noted that this is a decrease in
working gas also causing ionization. Ionization chambers operate density, not an increase as claimed in Ref. [8], since the same
at sufficiently high bias voltages to prevent recombination of the number of gas atoms now occupy a negligibly larger volume. We
electron–ion pairs generated by the interactions of photons, can therefore conclude that temperature increases would lead to
photoelectrons, and Compton electrons in the gas. In the case of lower density of the working medium in the detector, and thus to
the PTB measurements, that voltage was 500 V. The electrodes lower counts in the summer, as observed in the PTB data.
(anode and cathode) of the detector are contained within the However, in order to achieve a 3 10 3 fractional change in
chamber of the detector itself, and these collect the electrons and density, which would cause a corresponding fractional change in
ions created via the radiation interactions. The current output is the ionization and subsequent current, the room temperature
then the result of the of electrons collected at the anode, which is would need to change by 86 3 C, which is unrealistic.
1
1:44 108 Amp Roentgen1 h (http://www.centronic.co.uk/ In contrast to the operation of the Centronic IG12 ionization
ionisation.htm, 2009). One obvious systematic variable that can chamber employed in Ref. [5] and described above, whose density
be quickly examined is variation in the bias voltage. However, dependence is characterized rðTÞ constant, we next consider a
ionization chambers are known for their insensitivity to applied hypothetical detector in which the pressure, P, and the volume, V,
Please cite this article as: J.H. Jenkins, et al., Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.03.129
ARTICLE IN PRESS
J.H. Jenkins et al. / Nuclear Instruments and Methods in Physics Research A ] (]]]]) ]]]–]]] 3
are held constant, which is the setup studied in Ref. [8]. In this respectively. Plots of the time-averaged data and the monthly
case, we find from the ideal gas law that rT constant, so that averaged data from Ref. [19] are shown in Figs. 1 and 2. The
Dr ¼ DT=T. The change in intensity DI of grays passing diurnal effect is clearly evident, with the early morning maximum
through a thickness Dx of absorbing material with density r is activity concentration possibly happening while the heating,
given by DI=I ¼ mrDx where m is the absorption coefficient, and ventilation and air conditioning (HVAC) systems are off or in a
this leads to the familiar result [12] I ¼ I0 expðmrxÞ. Assuming lower functional state, as is typical in large institutional buildings.
100% detection efficiency, we can express the counting rate (C) as It should also be noted that 222Rn concentrations are known to be
C ¼C0(1 I/I0), where C0 is an appropriate normalization constant, dependent on air density, which is in turn dependent on
and hence, DC=Dr ¼ C0 mxexpðmrxÞ. To obtain the temperature temperature, relative humidity and pressure, as well as on
dependence of C, we then replace jDrj by jrjDT=T which gives precipitation. It follows that since the warmest part of the day
is generally late afternoon, and the coolest is early morning, a
DC mrxemrx DT
¼ : ð2Þ diurnal effect is an understandable result. The annual variation is
C 1e mrx T
less obvious, but it appears that a maximum occurs in
Although this reproduces the result in Ref. [8, Eq. (1)], we see from Summer/Autumn.
the preceding discussion that this result only applies to a detector While the preceding information and analyses do not com-
for which rT ¼ constant. This is not the case for the PTB detector, pletely rule out an effect caused by 222Rn, it appears that not only
which is characterized by rðTÞ constant as noted above. One are the resulting activities quite small, but additionally the phase
can illustrate the practical difference between these two cases and period of the 222Rn concentrations do not match those of the
by using the values for m, r, and x quoted in Ref. [8]:
m ¼ 0:07703 cm2 =g, r ¼ 0:03567 g=cm3 , and x ¼14.4 cm. We find
mrx ¼ 0:0396, and hence, jDC=Cj ¼ 0:98jDT=Tj. By way of contrast, 75
for the actual PTB detector, the only dependence of rðTÞ on T
is through the small thermal expansion of the detector chamber, 70
as discussed above. Using a coefficient of volume expansion
g ¼ 39 106 ð3 CÞ1 for steel, we find from the preceding analysis
jDr=rj ¼ 0:98jDV=Vj ¼ 4 105 . For T¼300 K, we then find Rn Activity (Bq/m ) 65
3
60
3
the order of 15–90 Bq/m3, which are far below the activities of the
sources being measured. Specifically, an analysis performed by 55
Abbady et al. in Hannover, Germany [19] (which is 60 km from
the PTB in Braunschweig), measured indoor concentrations of 50
222
Rn inside the Centre for Radiation Protection and Radioecology
222
Please cite this article as: J.H. Jenkins, et al., Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.03.129
ARTICLE IN PRESS
4 J.H. Jenkins et al. / Nuclear Instruments and Methods in Physics Research A ] (]]]]) ]]]–]]]
226
Ra measurement variations reported in Ref. [5] and later efficiency plot fall near the beginning of the calendar year for the
analyzed in Ref. [2]. Ge(Li) detector, which is opposite to the signal seen in the
Other contributions to background in the PTB measurements ionization chamber measurements of 226Ra. This appears to call
would be of terrestrial and cosmic origins, such as muons, or into question the conclusion of Jenkins et al. [2] that decay rates
photons from 40K decay. Efforts were made by Siegert et al. to are simply dependent on the distance of the Earth from the Sun.
reduce the background by use of significant lead shielding around However, the fact that the period of the 152Eu data is the same, but
the detector [5,22]. However, it should be noted that the 226Ra shifted by half a period, still suggests a possible link to the Earth–
data provided by Schrader (private communication, 2006) for the Sun distance.
analysis presented in Ref. [2] were already corrected for back- This possibility is supported by an examination of the decay
ground, which obviates the need for analyses of these background modes of 152Eu which has two branches, K-capture (72.10%) and
effects. Nevertheless, a discussion of those possible backgrounds b decay (27.90%) [23]. The 1408 keV photon measured with the
is presented below. An analysis of environmental radiation at Ge(Li) detector [5] only arises from the K-capture branch, and is
ground level was conducted by Wissmann at the PTB [20], seen in 21.07% of the decays (there is no photon of that energy in
utilizing data from the Cosmic Radiation Dosimetry Site, situated the b decay branch) [23]. In contrast to the Ge(Li) detector, the
on a lake near the PTB laboratory. The results of the analysis, ionization chamber is incapable of discriminating among
which corrected for fluctuations in weather and solar effects, the different photons measured, and hence could not track the
exhibited an annual variation of 76.9 nSv h 1, which is equiva- 1408 keV photon separately, as it would be lost in the sum of all
lent to a dose variation equal to 71.92 10 3 nSv s 1. Given the the photons emitted by the sample. The phase of the 152Eu data
initial amount of 226Ra in the PTB source, 300 mg, which would could thus be understood if K-capture modes responded to an
have had an initial activity of 11.1 MBq, the activity would be external influence oppositely from b decays.
88:8 MBq after 40 years due to the ingrowth of the radioactive In Refs. [1,2] the conjecture was put forward that a field or
daughters. Utilizing the more conservative activity value of particle emanating from the Sun, possibly neutrinos, might be
11.1 MBq, the gamma dose rate from the source can be estimated enhancing or interfering with the decay of radioactive nuclides.
to be 2:1 104 nSv s1 , which is 7 orders of magnitude larger
Since the quantum mechanical details of the K-capture and b
than the variation in the cosmic dose rate reported in Ref. [20]. decays are quite different, it is in fact possible that the response of
We can therefore conclude that seasonal variations in cosmic the K-capture branch would be different from that of the b decay
radiation background would likely have little if any effect on the branch. If these were in fact opposite effects, i.e. the decay rate for
PTB detector system. the b decay branch would be increasing when the K-capture rate
Returning briefly to the background radiation from 222Rn, it was decreasing, these competing effects would tend to damp out
should also be noted since 226Ra decays to 222Rn, the decay chain any fluctuations. That such a possibility could actually occur is
of 222Rn is identical to the decay chain of 226Ra. If the 222Rn supported by data from 54Mn (which also decays via K-capture)
daughters are in equilibrium with the relatively constant 222Rn acquired during the solar flare of 2006 December 13 [1]. It was
concentration of 55 Bq, the activity of 222Rn and its daughters at observed that the 54Mn count decreased from the expected rate
any time will be 300 Bq, which equates to a gamma dose rate of during the three-day period encompassing the solar flare. Thus, if
0:58 nSv s1 . This value is five orders of magnitude smaller than we assume that the solar neutrino flux increased during the solar
the dose rate of the 226Ra source itself. Hence, it can be safely flare, then the response of 54Mn and 152Eu to increased solar flux
assumed that even with a variation of the atmospheric 222Rn would have the same phase in the sense that both decay rates
concentration of 7100% (which is significantly larger than the decreased in response to an increase in solar neutrino flux.
actual variation noted in Ref. [19]), the even larger activity of
600 Bq would yield a new gamma dose rate of 1:16 nSv s1 ,
which is still four orders of magnitude smaller than the dose from 4. The BNL proportional counter
the 226Ra source itself. This effectively rules out a discharge effect
on cables, etc. arising from charged particles or photons produced We next turn to an examination of the environmental
by 222Rn daughters as the cause of the measured variations sensitivity of the gas proportional counter and sample changing
reported in Ref. [5]. system used by Alburger et al. [4] and Harbottle et al. [9] to
measure the half-life of 32Si. The detector in question was
cylindrical, 1:5 in. diameter, 2 in. depth, with a 1 in. end
3. The PTB solid state detector window, operating on P-10 gas. The detector pressure was held
constant by a device on the gas outlet vented to a barostated
The 154Eu and 155Eu samples studied by Siegert et al. [5] were enclosure to limit fluctuations in the density of the P-10 gas. The
measured with three different semi-conductor detectors as part of detector itself was mounted on an automatic precision sample
the evaluation of detector system stability previously discussed. changer, which was entirely contained within the aforementioned
In the course of these measurements, a separate 152Eu calibration barostated enclosure, as described in Refs. [4,9]. The window
source was used to measure the changes in efficiency of the Ge(Li) material was 0.006 mm Kaptons (Dupont H-film), with gold
detector which was one of the three semi-conductor detectors vacuum-deposited on both sides (40–50 Å on the outer surface,
examined. A plot of measurements of the efficiency of a Ge(Li) 120–150 Å on the inner surface). A bias voltage of 2150 V was
detector versus time, which is presented in Ref. [5, Fig. 2], exhibits applied to the detector, which was the voltage closest to the
a similar period of 1 year, shifted by half a period (or about 6 center of the beta counting plateau [4].
months), with an amplitude of about 0.5%. The efficiency of a Gas proportional detectors utilize gas multiplication, where
detector is determined by comparison of the measured counts free electrons generated by the ionization of the gas by the
taken with the detector to the expected output of a known source incoming particles are accelerated by the strong electric field
in a fixed geometry. Thus, for the Ge(Li) measurement of the 152Eu (created by the bias voltage) to sufficient kinetic energy to cause
point source, the minima of the efficiency curve represent a deficit further ionizations. Electrons freed in these ionizations can also be
in counts (with respect to the expected counts based on the accelerated to cause additional ionizations, as long as their energy
known activity and half-life of the standard), while the maxima is greater than the ionization energy of the neutral gas molecules.
represent an excess of counts. Specifically, the minima of the This gas multiplication process, or cascade, is known as a
Please cite this article as: J.H. Jenkins, et al., Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.03.129
ARTICLE IN PRESS
J.H. Jenkins et al. / Nuclear Instruments and Methods in Physics Research A ] (]]]]) ]]]–]]] 5
Please cite this article as: J.H. Jenkins, et al., Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.03.129
ARTICLE IN PRESS
6 J.H. Jenkins et al. / Nuclear Instruments and Methods in Physics Research A ] (]]]]) ]]]–]]]
1.2
diameter 3.96 cm, length 5.08 cm, and 3 mm wall thickness. The
detector window was modeled following Harbottle et al. [9] as
1 0.006 mm Kaptons plated with vacuum-deposited gold on both
the inner (125 Å) and outer (45 Å) surfaces. The detector volume
0.8
was filled with P-10 gas (90% argon, 10% methane by volume)
0.6 with a calculated density of 1.6773 10 3 g/cm3. This gas density
within the detector was held constant.
0.4 Due to the possibility that the lower energy betas from 36Cl
0.2 20°C, 35%RH
(708.6 keV) were more easily affected by changes in air density
25.56°C, 76%RH than the higher energy 32Si–32P betas (1710.5 keV), each source
0 was modeled separately, which also allowed us to account for the
0 0.005 0.01 0.015 0.02 0.025 0.03 0.035 0.04
difference in proximity of each source to the detector. The 32Si
β− Energy (MeV)
source, which was modeled in detail as described by Alburger
Fig. 4. Linear range of betas in air as a function of beta energy. The graphs are et al. [4], consisted of a brass base 3.18 cm in diameter and
calculated using the results of Katz and Penfold in Ref. [31]. The air densities approximately 0.3175 cm ð18 in:Þ thick with a 1.91-cm-diameter,
shown are based on the maximum and minimum values of temperature and 0.8-mm-deep recess. The recess contained 47.7 mg of SiO2
humidity quoted by Alburger et al. in Ref. [4].
covered by 9.263 10 5 cm of aluminum foil. The MCNPX
electron source volume was defined uniformly over the volume
due to transport through the detector window or sources, which occupied by the SiO2. According to Alburger et al., the 32SiO2
are assumed to be independent of temperature. It should be noted source was created approximately 13 years prior to the start
again, that the 36Cl source was 4 mm from the detector window, of the half-life experiment [4], and hence 32Si was certainly in
and the 32Si source 1 mm away. Fig. 4 confirms the conclusions of secular equilibrium with its radioactive daughter, 32P, during the
the BNL group that changes in T and/or RH would have been too experimental time frame. The source energy spectrum used in
small to explain the fluctuations in their data. These conclusions these MCNPX models consisted of equal contributions from the
are further supported by an MCNP analysis which we have carried 32
Si (Emax ¼225 keV) and 32P (Emax ¼ 1710.5 keV) beta energy
out, the results of which will be described below. Before turning spectra calculated according to Fermi theory, which allowed us to
to this analysis, we first consider the remaining question of incorporate the loss of the much lower energy betas from 32Si.
backgrounds which could impact the BNL detector system. As was the case for the 32Si source, the 36Cl source was placed
The BNL group measured an initial background counting rate on a brass base 3.18 cm in diameter and 0.3175 cm thick, but
of 6.6(3)/min. This background measurement was repeated after without a recess machined into the top. The 36Cl was spread over
the conclusion of all runs by counting with the 32Si source a diameter of 1.91 cm in the center of the top surface of the base.
removed from its holder, and the 36Cl still in place (to ensure that Thus, the MCNPX electron source was defined as a flat disk
there was no cross-talk between the sources). The rate was found 1.91cm in diameter on the surface of the brass base with an
to be 6.5(3)/min. Since the average count rates for the measure- energy spectrum equal to that of 36Cl (Emax ¼ 709.2 keV)
ments were initially substantially larger, 21,500/min and calculated according to Fermi theory.
14,800/min for the 36Cl source and 32Si source, respectively, the According to the BNL experimental notes, the distance of each
background count rates were considered insignificant, and hence source from the detector window was adjusted in order to obtain
were ignored by the BNL group. Additionally, since the sources similar count rates from each, and this was achieved by having
were alternated during the counting procedure for each run, the 32Si source 1.000 mm from the detector window, while the
30 min each for 20 cycles, the background would likely be the 36
Cl was placed at a distance of 4.000 mm. The source position in
same for both of the isotopes. the MCNPX models for each source reflected these distances.
As stated before, Alburger et al., reported [4] that the
temperature variation was well within the range 70276 3 F. We
5. Monte Carlo modeling of the BNL detector found that tally values varied so little over this range that they
were statistically constant to within the minimum achievable
The Monte Carlo radiation transport software package, MCNPX relative error in MCNPX. Thus, a large temperature range of 40 3 F
[32], was used to determine the effects of changes in ambient air to 90 3 F was chosen to improve the statistical determination of the
temperature on the proportional detector system used by function relating each tally to air temperature (and associated
Alburger et al. [4]. The air density between the source and density). The results presented here were obtained using 1 108
detector window was varied over the temperature range 40290 3 F (for 36Cl) or 2 108 (for 32Si-32P) source electrons for each density
by use of the ideal gas law, corresponding to an approximate variation, and this provided relative errors of 104 for the
density range between 1:1350 103 and 1:2486 103 g=cm3 particle/energy current tallies, and 2 104 for the energy
for dry air. While humidity also affects total air density, this effect deposition tallies.
is presumed to be small (see Fig. 3) relative to the effects of A number of MCNPX tallies were used to test the assertion that
temperature change, and hence was ignored in this study. variations in air density with temperature would result in an
Additionally, the 40290 3 F range that was analyzed encompasses associated increase or decrease in the probability that an electron
both the minimum and maximum air densities expected for a with a given energy will traverse the air gap between the source
wide range of both temperature and humidity in the laboratory. and detector window. A linear least-squares fit was applied to the
The physical parameters of the detector system (including the results of each tally as a function of air temperature, and these
proportional detector, 32Si source, and 36Cl source) were obtained results are summarized in Table 1.
Please cite this article as: J.H. Jenkins, et al., Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.03.129
ARTICLE IN PRESS
J.H. Jenkins et al. / Nuclear Instruments and Methods in Physics Research A ] (]]]]) ]]]–]]] 7
Table 1
Table of MCNPX results showing the sensitivities of the system per degree Fahrenheit.
32
Si–32P 36
Cl (32Si–32P)/36Cl
0.597 1.427
1.4265
0.5965
0.596 1.4255
1.425
0.5955
1.4245
0.595
1.424
0.5945 1.4235
1.423
0.594
1.4225
0.5935 1.422
30 40 50 60 70 80 90 100 30 40 50 60 70 80 90 100
Temperature (°F) Temperature (°F)
Fig. 5. The ratio of 32Si to 36Cl energy deposition in the proportional detector. Fig. 7. The ratio of 32Si to 36Cl energy current (MeV) across the proportional
Dashed lines indicate the 95% confidence interval for the calculated linear least- detector window (into the detector). Dashed lines indicate the 95% confidence
squares fit of the data. interval for the calculated linear least-squares fit of the data.
Please cite this article as: J.H. Jenkins, et al., Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.03.129
ARTICLE IN PRESS
8 J.H. Jenkins et al. / Nuclear Instruments and Methods in Physics Research A ] (]]]]) ]]]–]]]
0.961
1.14
_
In Fig. 10, we exhibit the count rate ratio Si=Cl NðSiÞ= _
NðClÞ as The preceding discussion leads naturally to the question of
a function of time, along with the outdoor temperature variation whether the observed effects in the BNL and PTB data could arise
for the area surrounding BNL [33], and 1/R2. We see that Si/Cl from a change in the gravitational potential of the Sun at the
approximately correlates with 1/R2 and anti-correlates with respective detectors as the Earth passes from perihelion to
temperature, in both cases with observable phase shifts. Since aphelion and back. In the framework of conventional General
at this stage we have no theoretical reason to suppose that Si/Cl Relativity (GR) all clocks at a given space-time point run at the
should correlate (rather than anti-correlate) with 1/R2, both of the same rate, irrespective of their internal mechanism. In terms of
correlations in Fig. 10 are significant. However, we have already Eq. (4) this would correspond to setting eðSiÞ ¼ eðClÞ, in which case
Please cite this article as: J.H. Jenkins, et al., Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.03.129
ARTICLE IN PRESS
J.H. Jenkins et al. / Nuclear Instruments and Methods in Physics Research A ] (]]]]) ]]]–]]] 9
there would be no fluctuation in the ratio Si/Cl. Since this is in removed from the ROI as a result of changes to the cosmic ray
obvious disagreement with the BNL data, conventional GR can be flux. An alternative explanation might be that an increase in the
ruled out as an explanation of the observed effects. cosmic ray flux would generate a large distribution of energy
However, this conclusion does not necessarily hold for various depositions across the energy spectrum, and thus change the
alternatives to GR. Will [35] has shown that in non-metric system dead time, leading to a loss of counts within the ROI. Dead
theories, clocks of different construction can in fact behave time was tracked as a variable, however, and did not show
differently in a gravitational field. Although such a mechanism changes over the course of the flare when compared to the counts
is possible in principle, it fails in practice on quantitative grounds. preceding or following the flare period. Additionally, inspection of
The Sun’s gravitational potential F ¼ GM } =Rc2 at the Earth is the spectra files also shows no significant changes in the counts of
F ¼ 9 109 , and the fractional change DF between perihelion channels above or below the 834.8 keV 54Mn peak, as might be
and aphelion is DF ¼ 3 1010 . An effect this small would be expected from a change in cosmic ray flux.
undetectable given the BNL and PTB statistics, even if the The above argument is bolstered by observations made by
coefficients multiplying DF in a non-metric theory were rela- other groups during the 2006 December 13 flare. Although an
tively large. Additionally, constraints on parameters which analysis of data from world-wide neutron monitors by Bütikofer
measure deviations from GR are sufficiently restrictive [36] to et al. and Plainaki et al. [37,38] exhibited a 90% increase in the
preclude the possibility of explaining the BNL and PTB data in 1-min data at Oulu and Apatity, the overall neutron flux was still
terms of any known alternative to GR. so small as to have a negligible effect on the detector system. This
A class of gravity-related theories which cannot be excluded at is particularly true in light of the fact that the NaI detector is
present are those in which scalar fields are introduced to induce a insensitive to neutrons, and the flux of neutrons was too low to
time variation in fundamental constants such as the fine structure affect the 54Mn sample directly.
constant a ¼ e2 =_c. We have discussed elsewhere [3] the possibi- Similar arguments allow us to exclude a change in the flux of
lity that although the simplest such theories cannot account for cosmic ray muons resulting from a Forbush decrease [39] as a
the BNL or PTB data, theories in which two or more scalar fields possible explanation of the flare data. A Forbush decrease is a rapid
are introduced which influence both a and the electron/proton change in the flux of cosmic rays resulting from plasma clouds
mass ratio (and possibly other quantities) might work. emitted by the Sun during a solar storm [39]. Evidence for a Forbush
decrease during the 2006 December 13 flare obtained from the
muon telescopes at the Basic Environmental Observatory (BEO) in
8. Environmental influences during the solar flare of 2006 Bulgaria and from Nagoya, Japan are shown in Figs. 5 and 6 of Ref.
December 13 [39], respectively. The BEO data exhibited a sharp Forbush decrease
beginning at 15 : 3016 : 30 UT on 2006 December 14, and
We turn in this section to a discussion of possible environ- reached a maximum at 05:00 UT on December 15. The beginning of
mental influences on the NaI detector system used in Ref. [1] the Forbush decrease thus occurred 37 h after the dip in the 54Mn
during the 2006 December 13 solar flare. This is motivated by the counting rate reported in Ref. [1], which coincided in time with the
potential connections between the 54Mn decay data collected in solar flare at 02: 40 UT on December 13. Additionally, the
the time period surrounding the flare, and the BNL and PTB decay minimum in the muon count rate associated with the Forbush
data which are the primary focus of the present paper. As we have decrease was observed more than 2 days later. The data from the
already noted in the Introduction, the short duration of the solar BEO are supported by data obtained at Nagoya, which exhibited a
flare allows us to rule out seasonal temperature, humidity and similar time-dependence [39]. Given the timing of the Forbush
pressure influences on the detector system as potential explana- decrease relative to the dip in the 54Mn counting rate we can thus
tions of the observed decrease in the 54Mn counting rate. Hence reasonably conclude that the dip was not likely the result of the
the focus of this section will be on other possible ‘‘environmental’’ response of the detector system to the Forbush decrease.
influences on the detector system during the solar flare, such as The preceding arguments can be further strengthened by noting
neutrons, muons, etc. from Ref. [1] that the 54Mn count rate started to decrease 1:7 days
Although cosmic ray particles are known to be capable of before the flare. It follows that the beginning of the decrease in the
influencing some detectors, there is a generic argument against 54
Mn count rate preceded the beginning of the Forbush decrease by
any mechanism which attributes the drop in the 54Mn count rate more than 70 h, and hence was not caused by it.
to cosmic rays. This is based on the observation that the signal for In summary, we can rule out environmental explanations of
a 54Mn decay in our system is the detection of the characteristic the correlation between the solar flare of 2006 December 13 and
834.8 keV photon resulting from the electron-capture process, the observed dip in our 54Mn data [1] on several grounds. These
e þ 54 P þ 54 P þ 54 include: (a) the observation that the solar flare lasted approxi-
25 MnðJ ¼ 3 Þ-ne þ 24 CrðJ ¼ 2 Þ-24 Crðg:s:Þ þ gð834:8 keVÞ:
mately 43 min, which is too short a time for seasonal environ-
ð5Þ
mental influences to have affected the detector system; (b) the
Hence the 834.8 keV photon which signals the electron-capture failure of these mechanisms to produce the characteristic
process is uniquely associated with the specific transition of the 834.8 keV photon arising from 54Mn electron capture, or to
daughter 54 P + P +
24Cr from the excited J ¼2 state to the J ¼0 ground account for the 1:7 day precursor signal observed in the 54Mn
state (g.s.). The data points utilized in Ref. [1] represent the decay data; and (c) the timing of the Forbush decrease which
integral count within the Region of Interest (ROI) set on the followed, rather than preceded, the dip in the 54Mn data.
834.8 keV photon for a 4 h live-time. Even if there had been a large Additionally, the observed changes in cosmic ray fluxes would
change in the cosmic ray flux of neutrons or muons during the have been too small quantitatively to account for the 54Mn data.
flare, the chance that the nuclear or electromagnetic interactions
induced by these particles would have accidentally produced a
photon at 834.8 keV, or deposited the energy equivalent of 9. Comments on analyses of decay data by other groups
834.8 keV within the NaI crystal, is extremely small. Moreover,
the fact that a deficit of counts appeared in the ROI, and not an Two recently published papers have questioned the results
increase, indicates that this already very remote possibility is published in Refs. [1–6]. The first of these, from Norman et al.
highly unlikely. It is also unlikely that counts were selectively [40], included the analyses of decay data from several isotopes of
Please cite this article as: J.H. Jenkins, et al., Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.03.129
ARTICLE IN PRESS
10 J.H. Jenkins et al. / Nuclear Instruments and Methods in Physics Research A ] (]]]]) ]]]–]]]
various decay modes, which showed no evidence of oscillations at observed in the BNL and PTB data: (a) they arise from the
the same level as seen in either the BNL or PTB data. It should be responses of the respective detector systems to seasonal varia-
noted, however, that none of the isotopes examined by the tions in temperature, pressure and humidity, and possibly other
Norman group were utilized in the experiments at BNL or PTB that factors such as radon buildup, and (b) the fluctuations arise from
did exhibit these oscillations. We have also received 36Cl data the decay process itself, due to some as yet unknown influence
from another institution, which did exhibit an oscillation similar possibly originating from the Sun. By modeling the respective
to the one observed at BNL. The analysis of these additional data is detector systems in detail, we have shown here that alternative
in process, and will be submitted for publication in the near future (a) is an unlikely explanation for the observed BNL and PTB data,
(Jenkins et al., in preparation). We note that the standard model and hence by implication alternative (b) is more likely to be
Lagrangian which describes beta decays, leads to half-lives which correct. The resulting inference, that nuclear decay rates are being
span a broad range, from fractions of a second to tens of billions of directly influenced by solar activity, is further supported by our
years. This range of half-lives reflects the extreme sensitivity of b analysis of 54Mn decay data acquired during the solar flare of 2006
decays to differences in nuclear structure, including the available December 13, which we have also analyzed. Our conclusion that
phase space and the overlap of parent and daughter wavefunc- the dip in the 54Mn data, which was coincident in time with the
tions. It is thus not surprising that different nuclei could exhibit a flare, was not likely attributable to changes in the cosmic ray flux
range of sensitivities to a common external perturbation, and this during the flare, further strengthens the case that nuclear decays
may account for the apparent absence of an indication of an are being directly affected by solar activity. This inference can
annual oscillation in the data reported by Norman et al. [40]. evidently be checked in a variety of experiments such as those
The second paper, by Cooper [41], was an examination of the data described on Ref. [3], some of which are already in progress.
from the radioisotope thermoelectric generators (RTGs) on board the
Cassini spacecraft, which was launched in 1997 and reached Saturn
in 2004. The Cassini mission provided an opportunity to examine the Acknowledgments
dependence of nuclear decay on the distance from the Sun, since
Cassini’s trajectory varied the range of R (the distance of the The authors wish to thank D. Alburger and G. Harbottle for
spacecraft from the Sun) from 0.7–1.6 AU. Each of the three RTGs that making their raw BNL data, apparatus, and samples available to
provide power for the mission initially contained 7.7 kg of 238Pu, us, and for many helpful conversations. We are also deeply
which decays by a emission (T1/2 ¼87.7(1) y [11]). The daughter of grateful to H. Schrader for the raw PTB decay data, as well as for
the 238Pu decay is 234U, which decays by a emission (T1/2)¼ 245,500 y invaluable discussions on his experiment during our visit to the
[11]) to 230Th, which also decays by a emission (T1/2 ¼75,400 y [11]). PTB. We are indebted to our colleagues who have helped us in
Since the half-lives of the daughters of 238Pu are so long, the only this work, including V. Barnes, J. Buncher, T. Downar, D. Elmore,
decays that are actually being measured in the RTGs are from the A. Fentiman, J. Heim, M. Jones, A. Karam, D. Krause, A. Longman,
238
Pu itself, via an indirect measurement of the power output. Cooper J. Mattes, E. Merritt, H. Miser, T. Mohsinally, S. Revankar, B. Revis,
claimed [41] that since the power measurement from the Cassini A. Treacher, J. Schweitzer, and F. Wissman. The work of EF was
RTGs showed no evidence of change with respect to the distance of supported in part by the U.S. Department of Energy under
the spacecraft from the Sun, this questioned the implication of Contract no. DE-AC02-76ER071428.
Jenkins et al. [1–3] that the Sun could be playing a role in these decay
rate changes. Cooper’s analysis was based upon the assumption that References
the 226Ra data measured in the PTB should be similar to the 238Pu
data, since both isotopes decay by a emission. However, as noted [1] J.H. Jenkins, E. Fischbach, Astropart. Phys. 31 (6) (2009) 407.
above, the PTB ionization chamber was not just measuring the [2] J.H. Jenkins, E. Fischbach, J.B. Buncher, J.J. Mattes, D.E. Krause, J.T. Gruenwald,
photons from the a decay of 226Ra and its a emitting daughters, but Astropart. Phys. 32 (1) (2009) 42.
[3] E. Fischbach, J. Buncher, J. Gruenwald, J. Jenkins, D. Krause, J. Mattes,
also from a series of b decays in the decay chain as well. When J. Newport, Space Sci. Rev. 145 (3) (2009) 285.
the PTB experiment began, the 226Ra source was already 40 years [4] D.E. Alburger, G. Harbottle, E.F. Norton, Earth Planet. Sci. Lett. 78 (2–3) (1986)
old [5], and most of the b emitting daughters (except 210Pb and its 168.
[5] H. Siegert, H. Schrader, U. Schoetzig, Appl. Radiat. Isot. 49 (9–11) (1998) 1397.
daughters, which will reach an approximate secular equilibrium after
[6] H. Schrader, Transmission of 226Ra data (Private communication).
222 years), were in secular equilibrium with the 226Ra parent, and [7] O.P. Verkhoglyadova, L. Gang, G.P. Zank, H. Qiang, AIP Conference Proceed-
comprised 42:2% of the total activity. The b decay share of the ings, vol. 1039. American Institute of Physics, 2008, pp. 214–219.
total activity of the 226Ra source will continue to increase slowly, [8] T.M. Semkow, D.K. Haines, S.E. Beach, B.J. Kilpatrick, A.J. Khan, K. O’Brien,
Phys. Lett. B 675 (5) (2009) 415.
until the 210Pb (T1/2 ¼22.20 y [11]) daughter and its progeny reach [9] G. Harbottle, C. Koehler, R. Withnell, Rev. Sci. Instr. 44 (1) (1973) 55.
secular equilibrium as well. By the end of the 20 year duration of [10] H. Schrader, Metrologia 44 (4) (2007) 53.
the PTB experiment, the b decay share of the total activity would [11] NNDC, National Nuclear Data Center: Chart of the Nuclides (March 2010)
/http://www.nndc.bnl.gov/chart/chartNuc.jspS.
have been 43:3%. When the b decay fraction of the 226Ra decay [12] G.F. Knoll, Radiation Detection and Measurement, third ed., John Wiley &
chain is compared to the b decay fraction of the 238Pu decay chain Sons, New York, 2000.
after 10 years, which is 4 1010 %, it is easy to see that there [13] C.G. Found, Phys. Rev. 16 (1) (1920) 41.
[14] S. Okabe, T. Nishikawa, M. Aoki, M. Yamada, Nucl. Instr. and Meth. A 255
is not likely any contribution to the RTG power from the b decaying (1987) 371–373.
daughters of 238Pu. Since the dynamics of b and a decays are [15] T. Nishikawa, S. Okabe, M. Aoki, Radiat. Protect. Dosim. 24 (1-4) (1988) 93–95.
completely different, it is therefore conceivable that only the b [16] J.C.H. Miles, R.A. Algar, J. Radiol. Protect. 8 (2) (1988) 103.
[17] M.I. Gaso, M.L. Cervantes, N. Segovia, V.H. Espindola, Radiat. Meas. 23 (1)
decays are being significantly affected by solar activity. Hence
(1994) 225.
not seeing the effect in the RTG measurements on Cassini does not [18] M.C. Thorne, J. Radiol. Protect. 23 (1) (2003) 29.
directly contradict the conclusions drawn in Refs. [1–3]. [19] A. Abbady, A.G.E. Abbady, R. Michel, Appl. Radiat. Isot. 61 (6) (2004) 1469.
[20] F. Wissmann, Radiat. Protect. Dosim. 118 (1) (2006) 3, doi:10.1093/rpd/
nci317.
[21] T. Szegvary, F. Conen, P. Ciais, Atmos. Environ. 43 (8) (2009) 1536.
10. Discussion and conclusions [22] H. Schrader, Appl. Radiat. Isot. 52 (3) (2000) 325.
[23] A. Artna-Cohen96, Nuclear data sheets for Eu-152, Nuclear Data Sheets 79(1).
[24] B. Singh, Nuclear data sheets for 32Si, Technical Report, 2006.
As we noted in the Introduction, there are at present two [25] P.M. Endt, R.B. Firestone, Nuclear data sheets for 32P, Technical Report, 1999.
competing general explanations for the apparent fluctuations [26] P.M. Endt, R.B. Firestone, Nuclear data sheets for 36Cl, Technical Report, 1999.
Please cite this article as: J.H. Jenkins, et al., Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.03.129
ARTICLE IN PRESS
J.H. Jenkins et al. / Nuclear Instruments and Methods in Physics Research A ] (]]]]) ]]]–]]] 11
[27] J.B. Cumming, Radiochem. Radioanal. Lett. 58 (5–6) (1983) 297. [36] C.M. Will, Theory and experiment in gravitational physics, revised ed.,
[28] P. Giacomo, Metrologia 18 (3) (1982) 171. Cambridge University Press, Cambridge, England; New York, NY, USA,
[29] R.S. Davis, Metrologia 29 (1) (1992) 67. 1993.
[30] A. Picard, R.S. Davis, M. Glaser, K. Fujii, Metrologia 45 (2) (2008) 149. [37] R. Bütikofer, E.O. Flückiger, L. Desorgher, M.R. Moser, B. Pirard, Adv. Space
[31] L. Katz, A.S. Penfold, Rev. Mod. Phys. 24 (1952) 28. Res. 43 (4) (2009) 499.
[32] Los Alamos National Laboratory, MCNPX User’s Manual, Version 2.6.0, la-cp- [38] C. Plainaki, H. Mavromichalaki, A. Belov, E. Eroshenko, V. Yanke, Adv. Space
07-1473 Edition, April 2008. Res. 43 (4) (2009) 474.
[33] NCDC, National Climatic Data Center: Land-based data (June 2009) /http:// [39] I. Angelov, E. Malamova, J. Stamenov, Adv. Space Res. 43 (4) (2009) 504.
www.nndc.noaa.gov/oa/land.htmlS. [40] E.B. Norman, E. Browne, H.A. Shugart, T.H. Joshi, R.B. Firestone, Astropart.
[34] K.J. Ellis, Phys. Med. Biol. 35 (8) (1990) 1079. Phys. 31 (2) (2009) 135.
[35] C.M. Will, Phys. Rev. D 10 (8) (1974) 2330. [41] P. Cooper, Astropart. Phys. 31 (4) (2009) 267.
Please cite this article as: J.H. Jenkins, et al., Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.03.129