Você está na página 1de 10

HIV Ag/Ab Combo

E
2G83-20
34-9962/R8
B2G830
Read Highlighted Changes
Revised March, 2008

HIV Ag/Ab Combo


Customer Service
For additional product information, please contact your local customer service organization.
This package insert must be read carefully prior to use. Package insert instructions must be carefully followed. Reliability
of assay results cannot be guaranteed if there are any deviations from the instructions in this package insert.

Key to symbols used


List Number
Reagent Pack
In Vitro Diagnostic Medical
Device
Lot Number Reaction Vessels

Expiration Date
Matrix Cells
Store at 2-8°C

Sample Cups
CAUTION: Consult
accompanying documents
Consult instructions
Manufacturer for use

See REAGENTS section for a full explanation of symbols used in reagent


component naming.

ABBOTT
Diagnostics Division

1
NAME Specimens that are initially reactive in the AxSYM HIV Ag/Ab Combo MEIA
AxSYM HIV Ag/Ab Combo should be retested in duplicate. Repeat reactivity is highly predictive of
the presence of HIV-1 and/or HIV-2 antibodies and/or HIV p24 antigen.
INTENDED USE However, as for all enzyme immunoassays, AxSYM HIV Ag/Ab Combo may
AxSYM HIV Ag/Ab Combo is a microparticle enzyme immunoassay yield nonspecific reactions due to other causes, particularly when testing
(MEIA) for the simultaneous qualitative detection of antibodies to human in low prevalence populations. A repeatedly reactive specimen should be
immunodeficiency virus type 1 and/or type 2 (HIV-1/HIV-2) and HIV p24 investigated further in sensitive, supplemental HIV specific tests, such
antigen in human serum or plasma. AxSYM HIV Ag/Ab Combo is used as as immunoblots, antigen tests, and HIV nucleic acid tests. Supplemental
an aid in the diagnosis of HIV infection. AxSYM HIV Ag/Ab Combo does testing of repeat reactive specimens obtained from individuals at risk
not discriminate between HIV-1 and HIV-2 antibody reactivity or HIV p24 for HIV infection usually confirms the presence of HIV antibodies or
antigen reactivity. HIV antigen, and HIV nucleic acid. A full differential diagnostic work-up for
SUMMARY AND EXPLANATION OF THE TEST the diagnosis of AIDS and AIDS-related conditions includes an examination
Acquired immunodeficiency syndrome (AIDS) is caused by two types of of the patient’s immune status and a clinical history.
human immunodeficiency viruses, collectively designated as HIV.1-7 BIOLOGICAL PRINCIPLES OF THE PROCEDURE
HIV is the etiologic agent of AIDS.1,3,6,7 HIV is transmitted by sexual AxSYM HIV Ag/Ab Combo is based on MEIA technology utilizing
contact, exposure to blood or blood products, and prenatal or perinatal recombinant HIV (E. coli) antigens and HIV p24 monoclonal (mouse)
infection of a fetus or newborn, respectively.8 Antibodies against HIV are antibodies coated on microparticles to capture antibodies against HIV-1/
nearly always detected in AIDS patients and HIV infected asymptomatic HIV-2 and HIV p24 antigen, respectively. Captured antibodies/antigen react
individuals,8,9 and HIV nucleic acid (RNA and/or proviral DNA) is always with biotin-labeled recombinant antigens, peptides, and p24 monoclonal
detected in AIDS patients and seropositive individuals.8,10 antibodies. The biotin-labeled complexes are detected using an anti-biotin:
Phylogenetic analysis classifies HIV-1 into groups M (major), N (non-M, alkaline phosphatase conjugate. Sample and all AxSYM HIV Ag/Ab Combo
non-O), and O (outlier).4,5 Group M viruses have spread throughout the reagents required for one test are pipetted by the sampling probe into
world to cause the global AIDS pandemic. In contrast, groups N and O are designated wells of a reaction vessel (RV) in the sampling center. The
relatively rare and endemic to west central Africa.11-17 However, group O RV is immediately transferred into the processing center. Further pipetting
infections have been identified in Europe and the USA.18-22 HIV-1 group is done in the processing center by the processing probe. The reactions
M is composed of genetic subtypes (A, B, C, D, F, G, H, J, and K) and occur in the following sequence:
circulating recombinant forms (CRFs).5,23 The geographic distribution and • A reaction mixture is formed by combining specimen diluent, sample,
regional predominance of HIV-1 subtypes and CRFs vary. All subtypes and and microparticles coated with recombinant antigens and HIV p24
many recombinant strains exist in Africa with CRF02_AG the predominant monoclonal antibodies in the sample well of the reaction vessel.
strain in west and west central Africa, subtypes A, C, and D predominant • When human antibodies to HIV-1/HIV-2 or HIV p24 antigen are present
in east central Africa, and subtype C predominant in southern Africa.23-28 in the sample, they bind to the coated microparticles, forming antigen-
HIV-1 subtype B is the predominant subtype in the USA, Europe, Japan, antibody complexes on the microparticles.
and Australia. However, a significant percentage of new HIV-1 infections • A portion of the reaction mixture (containing microparticles) is
in Europe are caused by non-B subtypes.29,30 In Asia, subtype C is found transferred to the matrix cell.
in India, and CRF01_AE (formerly called subtype E) and subtype B are
found in Thailand and southeast Asia.31 South America predominantly • The matrix cell is washed to remove materials not bound to the
has subtypes B and F.32,33 microparticles.
Human immunodeficiency virus type 2 (HIV-2) is similar to HIV-1 in • Biotinylated recombinant antigens, synthetic peptides, and HIV p24
its structural morphology, genomic organization, cell tropism, in vitro monoclonal antibodies are dispensed onto the matrix cell, forming
cytopathogenicity, transmission routes and ability to cause AIDS.6-8 immune complexes composed of recombinant antigen-human
However, HIV-2 is less pathogenic than HIV-1, and HIV-2 infections have antibody-biotinylated antigen or monoclonal antibody-p24 antigen-
a longer latency period with slower progression to disease, lower viral biotinylated monoclonal antibody.
titers, and lower rates of vertical and horizontal transmission.34-37 HIV-2 is • The matrix cell is washed to remove materials not bound to the
endemic to west Africa but HIV-2 infections, at a low frequency compared microparticles.
to HIV-1, have been identified in the USA, Europe, Asia, and other regions • The anti-biotin: alkaline phosphatase conjugate is dispensed onto the
of Africa.31,37 HIV-2 is classified into genetic subtypes A-G with most matrix cell and binds with the immune complexes.
infections caused by subtypes A and B.38,39 • The matrix cell is washed to remove materials not bound to the
The key immunogenic protein and antigenic target for serodetection of microparticles.
HIV infection is the viral (HIV) transmembrane protein (TMP). Antibodies • The substrate, 4-Methylumbelliferyl phosphate, is added. The
against the TMP (anti-TMP) consistently are among the first to appear alkaline phosphatase-labeled conjugate catalyzes the removal of a
at seroconversion of HIV infected individuals.9,39-43 The anti-TMP phosphate group from the substrate, yielding the fluorescent product,
response remains relatively strong throughout the course of the disease, 4-Methylumbelliferone. This fluorescent product is measured by the
as evidenced by the near universal presence of antibodies against the MEIA optical assembly.
TMP in asymptomatic and symptomatic stages of HIV infection.9,39-44 The presence or absence of antibodies to HIV-1/HIV-2 and/or HIV p24
TMPs from HIV-1 groups M and O and HIV-2 are represented in AxSYM antigen in the sample is determined by comparing the rate of formation of
HIV Ag/Ab Combo on the solid phase and in the probe by three pairs of fluorescent product to the cutoff rate which is previously calculated using
recombinant proteins and two synthetic peptides derived from native TMP the AxSYM HIV Ag/Ab Combo Index Calibrator. If the rate of formation of
sequences. The rationale for including three pairs of TMPs is derived from fluorescent product in the sample is greater than or equal to the cutoff
the genetic diversity within HIV-1 and between HIV-1 and HIV-2.5 Serologic rate, the sample is considered reactive for antibodies and/or p24 antigen.
studies indicate that although HIV-1 and HIV-2 share multiple common For further information regarding MEIA technology, refer to the AxSYM
epitopes in their core antigens, the envelope glycoproteins are much less System Operations Manual, Section 3.
cross-reactive.7,45-49 Antibodies elicited against the TMP (or portions of
the TMP) of a viral strain within one group or type may react well, poorly, REAGENTS
or not at all with the TMP (or portions of the TMP) from a viral strain of REAGENT KIT, 100 TESTS
a different group or type.15,50-55 An exception may be antibodies elicited AxSYM HIV Ag/Ab Combo Reagent Pack (2G83-20)
against HIV-1 group N.11,12 • 1 Bottle (6.8 mL) HIV-1/HIV-2 antigen/antibody coated microparticles
Early after infection with HIV, prior to seroconversion, HIV antigen(s) may in TRIS buffer. Minimum concentration: 0.025% solids. Preservative:
be detected in serum or plasma specimens.55-63 The HIV structural protein sodium azide. (Reagent Bottle 2)
most often used as the marker of antigenemia is the core protein, p24. The • 1 Bottle (14.3 mL) biotinylated probes: HIV-1/HIV-2 antigens and
AxSYM HIV Ag/Ab Combo employs pairs of mouse monoclonal (anti-) p24 anti-HIV p24 antibodies in TRIS buffer with surfactant, protein blockers,
antibodies on microparticles and in the probe to detect HIV p24 antigen stabilizers, and 10% goat serum. Minimum concentration: 100 ng/mL.
prior to seroconversion, thereby decreasing the seroconversion window Preservatives: sodium azide and ProClin 300. (Reagent Bottle 3)
and improving early detection of HIV infection.

2
• 1 Bottle (11.7 mL) anti-biotin (rabbit): alkaline phosphatase conjugate WARNINGS AND PRECAUTIONS
in TRIS buffer with protein stabilizers. Minimum concentration: 0.05 μg/mL. For In Vitro Diagnostic Use.
Preservative: sodium azide. (Reagent Bottle 1) SAFETY PRECAUTIONS
• 1 Bottle (38.2 mL) Specimen Diluent in TRIS buffer with surfactant, CAUTION: This product contains human sourced infectious and/or
protein blockers, and stabilizers. Preservatives: sodium azide and potentially infectious components. Refer to the REAGENTS section of
ProClin 300. (Reagent Bottle 4) this package insert. HIV-1 Positive Control has been tested and found
Index Calibrator to be reactive for anti-HIV-1. HIV-2 Positive Control has been tested and
• 1 Bottle (4.3 mL) AxSYM HIV Ag/Ab Combo Index found to be reactive for anti-HIV-2. Remaining components have been
Calibrator. Recalcified human plasma, nonreactive for HBsAg, tested and found to be nonreactive for antibodies to HCV, HIV-1/HIV-2 and
HIV-1 RNA or HIV-1 Ag, anti-HCV, and anti-HIV-1/HIV-2. Preservative: nonreactive for HBsAg and HIV-1 RNA or HIV-1 Ag. No known test method
sodium azide. Dye: green (Yellow No. 23 and Blue No. 9). can offer complete assurance that products derived from human sources
or inactivated microorganisms will not transmit infection. Therefore, it is
CONTROLS recommended that all human sourced materials be considered potentially
AxSYM HIV Ag/Ab Combo Controls: (NEG, PC1, PC2, VLPC) (2G83-11) infectious and handled with appropriate biosafety practices.
4 Bottles (8.0 mL each) of AxSYM HIV Ag/Ab Combo Controls. • Some components of this product contain methylisothiazolones, which
• The Negative Control, prepared in recalcified human are components of ProClin and are classified per applicable European
plasma, is nonreactive for HBsAg, HIV-1 RNA or HIV-1 Ag, anti-HCV, Community (EC) Directives as: Irritant (Xi). For a specific listing refer
and anti-HIV-1/HIV-2. Preservative: sodium azide. to the REAGENTS section of this package insert. The following are
• The HIV-1 Positive Control (PC1), prepared the appropriate Risk (R) and Safety (S) phrases.
in recalcified human plasma (inactivated), is reactive for anti-HIV-1 R43 May cause sensitization by skin contact.
and nonreactive for HBsAg, HIV-1 RNA or HIV-1 Ag, and anti-HCV. S24 Avoid contact with skin.
Preservative: sodium azide. S35 This material and its container must be
• The HIV-2 Positive Control (PC2), prepared disposed of in a safe way.
in recalcified human plasma (inactivated), is reactive for anti-HIV-2 S37 Wear suitable gloves.
and nonreactive for HBsAg, HIV-1 RNA or HIV-1 Ag, and anti-HCV. S46 If swallowed, seek medical advice immediately
Preservative: sodium azide. and show this container or label.
• All components of this product contain sodium azide. Contact with
• The Viral Lysate Positive Control (VLPC) acids liberates very toxic gas. This material and its container must be
is purified HIV viral lysate prepared in TRIS buffer with stabilizers. disposed of in a safe way.
Preservative: sodium azide. • For product not classified as dangerous per European Directive
AxSYM HIV Ag/Ab Combo Controls: (NEG, PC1, VLPC) (2G83-12) 1999/45/EC as amended - Safety data sheet available for professional
3 Bottles (8.0 mL each) of AxSYM HIV Ag/Ab Combo Controls. user on request.
• The Negative Control, prepared in recalcified human HANDLING PRECAUTIONS
plasma, is nonreactive for HBsAg, HIV-1 RNA or HIV-1 Ag, anti-HCV, • AxSYM HIV Ag/Ab Combo Reagents are susceptible to bubbles/
and anti-HIV-1/HIV-2. Preservative: sodium azide. foaming and require inspection and removal of bubbles
• The HIV-1 Positive Control (PC1), prepared before loading. Refer to the AxSYM System Operations Manual,
in recalcified human plasma (inactivated), is reactive for anti-HIV-1 Section 9.
and nonreactive for HBsAg, HIV-1 RNA or HIV-1 Ag, and anti-HCV. • Do not use Solution 1 (MUP) beyond the expiration date or a
Preservative: sodium azide. maximum of 14 days on-board the AxSYM System. When loading
new Solution 1 (MUP), it is important to immediately tighten the
• The Viral Lysate Positive Control (VLPC) instrument cap for MUP to minimize exposure to air. Prolonged
is purified HIV viral lysate prepared in TRIS buffer with stabilizers. exposure of MUP to air may compromise performance.
Preservative: sodium azide. • Do not use reagent pack beyond the expiration date, or a maximum
Control Range of 336 cumulative hours on-board the AxSYM System.
Control Color Dye S/CO • Do not mix reagents from different reagent packs. Do not mix reagents,
controls, and index calibrators from different lots. The reagent pack
Natural None 0.01 - 0.89
and its associated index calibrator must be discarded together.
Blue Blue No. 9 1.00 - 10.00 • Avoid microbial contamination of specimens and reagents. Use of
Red Red No. 33 1.00 - 10.00 disposable pipettes or pipette tips is recommended.
• Avoid chemical contamination of reagents and equipment.
Red No. 33,
Purple Blue No. 9 1.00 - 10.00 • Ensure that sufficient sample volume is present. If sample volume is
insufficient, the AxSYM System may indicate an error code and no
AxSYM HIV Ag/Ab Combo Index Calibrator and Controls should be mixed result will be reported. For a description of the system error codes,
by gentle inversion prior to use. refer to the AxSYM System Operations Manual, Section 10.
OTHER REAGENTS • Inadequate adherence to these package insert instructions may result
AxSYM Probe Cleaning Solution (9A35-05) in inconsistent results.
• 2 Bottles (220 mL each) AxSYM • Use accurately calibrated equipment.
Probe Cleaning Solution containing 2% tetraethylammonium hydroxide • Use caution in handling patient specimens to prevent cross
(TEAH). contamination.
Solution 1 (MUP) (8A47-04) Refer to the AxSYM System Operations Manual, Sections 7 and 8, for a
• 4 Bottles (230 mL each) Solution 1 (MUP) more detailed discussion of the safety and handling precautions during
containing 4-Methylumbelliferyl Phosphate, 1.2 mM, in AMP buffer. system operation.
Preservative: sodium azide. STORAGE INSTRUCTIONS
Solution 3 (Matrix Cell Wash) (8A81-04)
• 4 Bottles (1000 mL each)
Solution 3 (Matrix Cell Wash) containing 0.3 M sodium chloride in TRIS • The AxSYM HIV Ag/Ab Combo Reagent Pack, Index
buffer. Preservatives: sodium azide and antimicrobial agents. Calibrator, and Controls must be stored upright at 2-8°C. The AxSYM
Solution 4 (Line Diluent) (8A46) HIV Ag/Ab Combo Reagent Pack, Index Calibrator, and Controls may
• 1 Bottle (10 L) Solution 4 (Line Diluent) be used immediately after removal from the refrigerator.
containing 0.1 M phosphate buffer. Preservatives: sodium azide and NOTE: The AxSYM HIV Ag/Ab Combo Reagent Kit is shipped on wet
antimicrobial agent. ice and should be stored at 2-8°C after receipt.

3
When stored and handled as directed, reagents are stable until Assay Parameters
expiration date. The AxSYM HIV Ag/Ab Combo Reagent Pack may be
on-board the AxSYM System for a maximum of 336 cumulative hours 70 Min correlation coefficient for low rates: 0.9700
(for example, 42 eight hour shifts). Recalibration may be required 71 Min correlation coefficient for high rates: 0.9700
to obtain maximum onboard reagent stability. More frequent use of 72 MUP T Delay-Time delay following MUP: 3.0500
controls may be required to monitor reagent performance within the 75 Low Extreme Value > 0.00
same lot. After 336 hours, the reagent pack and its associated index 76 High Extreme Value > 0.00
calibrator must be discarded. Refer to the AxSYM System Operations
Manual, Sections 2, 5, and Appendix C, for further information on 80 Interpretation Option to use > 1
tracking on-board time. 84 Hold results with POS interpretation > ON
Solution 1 (MUP) must be stored at 2-8°C (do not freeze). It may 85 Hold results with NEG interpretation > OFF
be used immediately after removing it from the refrigerator. It may 86 Hold results with GRY interpretation > ON
be on-board the AxSYM System for a maximum of 14 days. After 91 Low Range Undiluted: 0.00
14 days, it must be discarded. 92 High Range Undiluted: 80.00
** AxSYM HIV Ag/Ab Combo Assay File Version 1.00.1
• The AxSYM Probe Cleaning Solution, Solution 3 *** AxSYM HIV Ag/Ab Combo Assay File Version 1.00.2 or higher
(Matrix Cell Wash), and Solution 4 (Line Diluent) must be stored at NOTE: Parameters 43, 44, and 45 cannot be edited. There are no other
15-30°C. options for this assay.
INDICATIONS OF INSTABILITY OR DETERIORATION OF The AxSYM HIV Ag/Ab Combo values available for Assay Parameter 80
(Interpretation Option to use) are:
REAGENTS
When AxSYM HIV Ag/Ab Combo Controls, Negative Control, Positive POS Interp NEG Interp GRY Interp
Control 1, Positive Control 2, or Viral Lysate Positive Control values are 1. REACTIVE NEGATIVE GRAYZONE
out of the expected range, it may indicate deterioration of the reagents or 2. REACTIVE NEGATIVE NEGATIVE
errors in technique. Associated test results are invalid and the specimens
must be retested. Assay recalibration may be necessary. Refer to the To utilize the grayzone option, option 1 for parameter 80 (Interpretation
AxSYM System Operations Manual, Section 10, for further troubleshooting Option to use) must be selected. Specimens with assay S/CO values
information. equal to or greater than 0.90, and less than the cutoff are indicated as
“GRAYZONE” in the interpretation field. (Refer to INTERPRETATION OF
INSTRUMENT PROCEDURE RESULTS section.)
NOTE: AxSYM HIV Ag/Ab Combo must only be used with AxSYM System Refer to the AxSYM System Operations Manual for a detailed description
Software Version 3.60 or higher. of Instrument Procedures.
Assay File Installation
The AxSYM HIV Ag/Ab Combo Assay File Version 1.00.1 or higher must
SPECIMEN COLLECTION AND PREPARATION FOR ANALYSIS
be installed on the AxSYM System from the assay disk, 2G84-01 or • Human serum (collected in Kaolin-coated or serum separator tubes)
higher, prior to performing the AxSYM HIV Ag/Ab Combo assay. Refer to or plasma (collected in EDTA, sodium heparin, lithium heparin, sodium
the AxSYM System Operations Manual, Section 2, for proper installation citrate, ACD, CPD, CPD/Adsol, or CPDA-1) may be used in the AxSYM
procedures. HIV Ag/Ab Combo assay. Follow the manufacturer’s processing
instructions for serum and plasma collection tubes.
AxSYM HIV Ag/Ab Combo Assay Parameters
• The AxSYM System does not provide the capability to verify the
The default values for the assay parameters for the AxSYM HIV Ag/Ab sample type. It is the responsibility of the operator to verify that
Combo assay are listed below. Values for the assay parameters that the correct sample type(s) is (are) used in the AxSYM HIV Ag/Ab
can be edited contain a (>) symbol. These parameters can be displayed Combo assay.
and edited according to the procedure in the AxSYM System Operations
Manual, Section 2. In order to obtain values for the parameters with an • This assay was designed and validated for use with human serum
asterisk (*), review the specific Assay Parameters screen. Press PRINT or plasma from individual patient and donor specimens. Pooled
to print the assay parameters. specimens must not be used since the accurracy of their test results
has not been validated.
Assay Parameters • Gravity separation is not sufficient for specimen preparation.
1 Long Assay Name (English): HIV_Ag/Ab_Combo • All patient specimens to be tested in primary tubes must be centrifuged
6 Abbrev Assay Name (English): HIV_Ag_Ab to remove red blood cells or particulate matter. Follow the tube
11 Assay Number: 817 manufacturer’s instructions for centrifugation.
12 Assay Version: * • After specimens have been initially separated per the collection tube
13 Calibration Version: * manufacturer’s instructions, they must be clarified by centrifugation
at 150,000-300,000 g-minutes* if:
14 Assay File Revision: *
• they contain clots, red blood cells, or particulate matter
15 Assay Enabled > ON
• they require repeat testing, or
17 Assay Type: MEIA
• they have been frozen and thawed.
18 Standard Cal Reps > 3 Transfer clarified specimen to a sample cup or secondary tube for
43 Default Dilution Protocol > UNDILUTED testing.
44 Default Calibration Method > Index Cal * g-minute = relative centrifugal force (RCF) x minutes spun
45 Selected Result Concentration Units > S/CO The following chart lists examples of acceptable time and force ranges
46 Selected Result Decimal Places > 2 that meet this criteria.
64 Max Intercept-Max MUP intercept: 15000.0000 Time RCF x Time
65 Min Intercept-Min MUP intercept: 1850.0000 (minutes) RCF (x g) (g-minutes)
66 Upper Limit for NRMSE for low rates: 9999.9900 15 10000 150000
67 Upper Limit for NRMSE for high rates: 0.3900** or 20 7500 - 12000 150000 - 240000
0.5100***
25 6000 - 12000 150000 - 300000
68 Max Rate-Max rate used to check Min
NOTE: The AxSYM System offers an Auto Retest/Auto Dilution
MUP Intercept: 150.0000
feature. Due to the requirements discussed above, this feature
69 Min Rate-Rate cutoff for NRMSE and must not be used with this assay.
Corr. Coef.: 8.0000 • Centrifuged specimens with a lipid layer on top of the liquid must be
transferred to a sample cup or secondary tube. Care must be taken to
transfer only the clarified specimen and not the lipemic material.

4
• Specimens from heparinized patients may be incompletely coagulated AxSYM HIV Ag/Ab Combo PROCEDURE
and inconsistent results could occur due to the presence of fibrin. To Materials Provided
prevent partially coagulated specimens, draw the specimen prior to • 2G83-20 AxSYM HIV Ag/Ab Combo Reagent Kit containing:
heparin therapy or into a plasma collection tube.
AxSYM HIV Ag/Ab Combo
• Specimens may be stored on or off the clot or red blood cells for up
AxSYM HIV Ag/Ab Combo
to 7 days at 2-8°C prior to being tested.
100
• Specimens which are not tested within 7 days must be stored frozen
(-20°C or colder). Specimens must be removed from the clot or red 100
blood cells prior to freezing. Materials Required but Not Provided
• Multiple freeze/thaw cycles should be avoided although no qualitative • 2G83-11 AxSYM HIV Ag/Ab Combo Controls
differences were observed between the experimental controls and (NEG, PC1, PC2, VLPC):
the 36 nonreactive or 36 spiked reactive specimens subjected to
6 freeze/thaw cycles.
Mix by inverting thawed specimens 180 degrees from upright
and return, for a total of 10 inversion cycles. Visually inspect
the specimens for the absence of stratification. If layering or
stratification is observed, repeat until specimens are visibly
homogeneous.64 or
Centrifuge at 150,000 - 300,000 g-minutes prior to use to remove • 2G83-12 AxSYM HIV Ag/Ab Combo Controls
particulate matter and to ensure consistency in the results. (NEG, PC1, VLPC):
• When shipped, specimens must be packaged and labeled in
compliance with applicable state, federal, and international regulations
covering the transport of clinical specimens and infectious substances.
It is recommended to ship specimens off the clot or red blood cells
and frozen or refrigerated.
• 8A47-04
• All samples (patient specimens, controls, and calibrators) must be
• 8A81-04
tested within 3 hours of being placed on board the AxSYM System.
Refer to the AxSYM System Operations Manual, Section 5, for a more • 8A46
detailed discussion of on-board sample storage constraints. • 9A35-05 AxSYM
• Inspect all samples for bubbles. Remove bubbles prior to analysis. • 8A76-01
• For optimal results, specimens must be free of fibrin, red blood cells, • Pipettes and pipette tips (optional) to deliver the volumes specified
or other particulate matter. on the Orderlist screen.
• Do not use heat-inactivated specimens. CAUTION:
• Specimens with obvious microbial contamination should not be • AxSYM HIV Ag/Ab Combo Index Calibrator and Controls should be
used. mixed by gentle inversion prior to use.
• Performance has not been established using cadaver specimens or • When manually dispensing samples, verify that dispensing equipment
body fluids other than human serum or plasma. does not introduce cross contamination and that it delivers the specified
• No qualitative performance differences were observed between the sample volume. Use a separate pipette tip for each sample.
experimental controls and the 36 nonreactive or 36 spiked reactive • For optimal performance, it is important to follow the routine
specimens tested with elevated levels of bilirubin (≤ 20 mg/dL), maintenance procedures defined in the AxSYM System Operations
hemoglobin (≤ 500 mg/dL), total protein (≤ 12 g/dL), or red blood Manual, Section 9. If your laboratory requires more frequent
cells (≤ 0.4% v/v). maintenance, follow those procedures.
• No qualitative performance differences were observed between the Assay Procedure
experimental controls and the 35 nonreactive or 35 spiked reactive CAUTION: The System status must be WARMING, PAUSED, READY, or
specimens tested with elevated levels of lipids (≤ 3000 mg/dL). STOPPED before adding or removing sample segments, reagent packs,
SAMPLE VOLUME or reaction vessels.
The sample volume required to perform a single AxSYM HIV Ag/Ab Combo NOTE: The AxSYM System Auto Retest/Auto Dilution feature must not
test on the AxSYM System varies according to the type of sample container be used for this assay. Refer to the SPECIMEN COLLECTION AND
used. For sample cups the minimum (STAT and ROUTINE) sample volume PREPARATION FOR ANALYSIS section of this package insert.
required is 194 μL. For every additional AxSYM HIV Ag/Ab Combo test 1. Check for sufficient onboard inventory of Matrix Cells, bulk solutions,
performed from the same sample container, an additional 144 μL of and sample segment availability.
sample is required. 2. Check for sufficient waste collection capacity.
The sample cup minimum volumes for both STAT and ROUTINE tests are
calculated by the AxSYM System. They are displayed on the Orderlist
CAUTION:
screen at the time the test(s) is(are) ordered and printed in the Orderlist • Do not open the Interior Waste Door or the AxSYM Processing Center
Report. When using Host Order Query, the Orderlist screen information and Cover while any test is in progress. If opened, all processing will stop.
Orderlist Report are not available. Refer to the AxSYM System Operations All tests will be terminated and must be repeated.
Manual, Section 5, for a description of the Host Order Query option. • The cap for reagent bottle 4 must be manually opened prior to running
To obtain the recommended volume requirements for the AxSYM HIV Ag/Ab an AxSYM HIV Ag/Ab Combo assay. Upon completion of the run, close
Combo Index Calibrator and Controls, hold the bottles vertically and the reagent bottle 4 cap securely.
dispense 10 drops of Index Calibrator or 8 drops of Negative or Positive 3. Order the AxSYM HIV Ag/Ab Combo Index Calibrator, AxSYM HIV
Controls into each respective sample cup. Ag/Ab Combo Controls, and/or patient specimens as required.
For sample volume requirements in primary or aliquot tubes, and control Assign or modify sample segment position (S/P) for each sample, as
volume requirements for multiple AxSYM HIV Ag/Ab Combo reagent lots, necessary. Refer to the QUALITY CONTROL PROCEDURES section
refer to the AxSYM System Operations Manual, Section 5. in this package insert for calibration and control requirements.
Index Calibration
Perform AxSYM HIV Ag/Ab Combo calibration by testing three
replicates of the Index Calibrator. Invert gently to mix and dispense
at least 10 drops of the Index Calibrator into a sample cup. Do not
simultaneously calibrate more than one AxSYM HIV Ag/Ab Combo
reagent lot.

5
Controls QUALITY CONTROL
Perform quality control by testing Positive and Negative Controls (one The minimum control requirement for an AxSYM HIV Ag/Ab Combo assay
test each). Invert gently to mix and dispense at least 8* drops each is a single sample of the Positive and Negative Controls tested every
of the Positive and Negative Controls into individual sample cups. 24 hours, each day of use for each reagent lot. Controls may be placed
* When more than one AxSYM HIV Ag/Ab Combo reagent lot is in any position in the sample carousel.
on board the AxSYM System, multiply the control volume by the NOTE: It is recommended that all three AxSYM HIV Ag/Ab Combo Positive
number of lots. Controls, Positive Control 1, Positive Control 2, and Viral Lysate Positive
Patient Specimens Control, be run to verify the calibration.
Ensure that sufficient volume is present in the sample cups or tubes. If the AxSYM HIV Ag/Ab Combo assay is used to evaluate HIV-2, the HIV-2
The sample cup minimum volume is 194 μL for the first AxSYM HIV Ag/Ab Positive Control must be tested every 24 hours.
Combo test plus 144 μL for each additional AxSYM HIV Ag/Ab Combo If the quality control procedures in your laboratory require more frequent
test. For volume requirements in primary or aliquot tubes, refer to the use of controls to verify test results, follow those procedures.
AxSYM System Operations Manual, Section 5. The AxSYM HIV Ag/Ab Combo Control values must be within the acceptable
NOTE: The operator may obtain an Orderlist Report by pressing PRINT. ranges specified in this package insert (see REAGENTS section). If a
The printout contains sample placement information and minimum STAT control value is out of its specified range, the associated test results are
sample cup volume requirements for all tests ordered. When using the invalid and must be retested. Recalibration may be indicated.
Host Order Query, the Orderlist Report is not available. Refer to the The AxSYM System has the capability to generate a Levey-Jennings plot
AxSYM System Operations Manual, Section 5, for a description of the of each assay’s quality control performance. Refer to the AxSYM System
Host Query option. Operations Manual, Section 5. At the discretion of the laboratory, selected
4. Place sample segments containing the ordered samples into the quality control rules may be applied to the quality control data.
sample carousel. FLUORESCENCE BACKGROUND ACCEPTANCE CRITERIA
5. Open reagent bottle 4. Place the AxSYM HIV Ag/Ab Combo Reagent Quality control with regard to the MUP substrate blank is automatically
Pack into the reagent pack carousel. determined by the instrument and checked under Assay Parameter 64
6. Ensure that reaction vessels (RVs) are present on the RV Carousel. (Max Intercept-Max MUP intercept) each time a test result is calculated.
Additional RVs may be added as needed. If the MUP intercept value is greater than the maximum allowable value,
7. Press RUN. All entries on the Order screen are transferred to the the result is invalid. The test request will be moved to the Exceptions List
Order Status screen for sample processing. where it will appear with the message “1064 Invalid test result, intercept
8. Review the results to determine whether retesting is required. too high” and the calculated intercept value. Refer to the AxSYM System
9. When testing is completed, close reagent bottle 4 and remove the Operations Manual, Section 10, when this error message is obtained.
samples and the AxSYM HIV Ag/Ab Combo Reagent Pack from the Refer to the AxSYM System Operations Manual, Section 2, for further
sampling center. Store at 2-8°C. information on parameter files.
NOTE: When using the onboard tracking feature, the operator must perform RESULTS
a reagent pack scan after removing any pack from the system in order to CALCULATION
maintain the validity of the reagent pack stability time. The AxSYM HIV Ag/Ab Combo assay protocol calculates the cutoff rate
Refer to Sections 5 and 6 of the AxSYM System Operations Manual for a (CO) from the mean rate of three Index Calibrator replicates and stores
detailed explanation of performing assay calibration and sample testing the result. The cutoff rate is determined by adding 27.5 to the mean Index
procedures. Calibrator rate.
QUALITY CONTROL PROCEDURES Cutoff rate (CO) = Index Calibrator mean rate + 27.5
CALIBRATION The AxSYM HIV Ag/Ab Combo assay protocol calculates a result based
A minimum of three replicates of the AxSYM HIV Ag/Ab Combo Index on the ratio of the sample rate (S) to the cutoff rate for each sample
Calibrator must be tested for an AxSYM HIV Ag/Ab Combo calibration. A and control.
single sample of both the Positive and Negative Controls must be tested S/CO = sample rate/cutoff rate
as a means of evaluating the calibrated equipment. FLAGS
NOTE: It is recommended that all three AxSYM HIV Ag/Ab Combo Positive Some results may contain information in the Flags field. For a description
Controls, Positive Control 1, Positive Control 2, and Viral Lysate Positive of the flags that may appear in this field, refer to the AxSYM System
Control, be run to verify the calibration. Operations Manual, Sections 1 and 2.
If the AxSYM HIV Ag/Ab Combo assay is used to evaluate HIV-2, the HIV-2
Positive Control must be tested every 24 hours. INTERPRETATION OF RESULTS
Once the AxSYM HIV Ag/Ab Combo calibration is accepted and stored, Initial AxSYM HIV Ag/Ab Combo Results
all subsequent samples may be tested without further calibration unless Initial Result Retest
one or more of the following occur: (S/CO) Instrument Flag Interpretation Procedure
• A reagent pack with a new lot number is used. ≥ 1.00 REACTIVE Reactive Retest in duplicate
• Any of the AxSYM HIV Ag/Ab Combo Control values is out of its 0.90 to < 1.00 GRAYZONE Grayzone Reactive Retest in duplicate
specified range. < 0.90* NEGATIVE Negative No retest required
• The MEIA Optics Verification Update has been performed.
* If parameter 80-option 2 is selected, NEGATIVE is designated as
Refer to the AxSYM System Operations Manual, Section 6, for additional
< 1.00 S/CO.
information.
NOTE: Specimens that are reactive or grayzone on initial testing must
The operator must verify that the AxSYM HIV Ag/Ab Combo Control values
be centrifuged according to directions in the SPECIMEN COLLECTION
are within the acceptable ranges specified in this package insert. Refer to
AND PREPARATION FOR ANALYSIS section of this package insert and
the REAGENTS section for AxSYM HIV Ag/Ab Combo control ranges.
retested in duplicate.
The AxSYM System verifies that the results of an assay calibration meet
the specifications assigned to selected validity parameters. An error
message occurs when the calibration fails to meet a specification. Refer
to the AxSYM System Operations Manual, Section 10, for an explanation
of the corrective actions for the error code. Refer to the AxSYM System
Operations Manual, Appendix E, for an explanation of the calibration validity
parameters that may be used by the AxSYM System.

6
Final AxSYM HIV Ag/Ab Combo Results EXPECTED VALUES
Initial Result Retest Results Final Result Interpretation In a random population of 7900 volunteer blood donor specimens,
10 (0.13%) were repeatedly reactive by AxSYM HIV Ag/Ab Combo. Among
Reactive/ Both of the Negative HIV-1/HIV-2 792 specimens from individuals with HIV-1 classified disease status, HIV-1
Grayzone duplicate retests antibodies and/or unknown disease status, HIV-1 subtyped, and unknown disease status
are negative. HIV p24 antigen for HIV-1 group O, HIV-1 p24 antigen, and HIV-2, the AxSYM HIV Ag/Ab
not detected. Combo assay detected 100.00% as reactive.
One or both of Reactive or Presumptive
the duplicate Grayzone evidence of SPECIFIC PERFORMANCE CHARACTERISTICS
retests are HIV-1/HIV-2 PRECISION
reactive antibodies and/or Assay reproducibility was determined during the clinical evaluation of
or grayzone. HIV p24 antigen; AxSYM HIV Ag/Ab Combo. A six-member panel and assay controls were run
a supplemental at 10 sites on 10 instruments in total, with the exception of the HIV-1 p24
assay should be panel which was tested at eight sites only. Nine sites tested two different
performed. clinical lots of the controls and one site tested three different clinical lots of
Nonreactive No retest required. Negative HIV-1/HIV-2 the controls. All samples were tested in triplicate in three runs. At one site,
antibodies and/or the panel and assay controls were tested on one clinical lot; at eight sites,
HIV p24 antigen they were tested on each of two clinical lots; and at one site, they were
not detected. tested across four different clinical lots. The intra-assay and inter-assay
standard deviation (SD) and percent coefficient of variation (%CV) were
• If repeat testing shows the assay S/CO value for both retests to be analyzed with a variance components analysis 65 using a mixed analysis
less than 0.90, the specimen is classified as negative. of variance model 66 (Table 1).
• If either duplicate retest S/CO value is greater than or equal to 0.90,
the specimen should be tested by supplemental tests. Analysis of Table 1
follow up samples is recommended. Reproducibility of AxSYM HIV Ag/Ab Combo
• Interpretation of results of specimens with a final result of reactive Total Mean Intra-assay Inter-assay
or grayzone by AxSYM HIV Ag/Ab Combo and indeterminate by Sample (N) (S/CO) SD %CV SD %CV
supplemental testing is unclear; further clarification may be obtained Undiluted 189 25.96 1.29 5.0 2.41 9.3
by testing another specimen taken three to six weeks later. anti-HIV-1
• AxSYM HIV Ag/Ab Combo and supplemental assay results should Undiluted 189 54.15 2.00 3.7 2.47 4.6
be interpreted in conjunction with the patient’s clinical presentation, anti-HIV-2
history, and other laboratory results. Diluted HIV-1 189 11.85 0.54 4.6 0.74 6.3
LIMITATIONS OF THE PROCEDURE medium
• It is recognized that this method for the detection of HIV p24 antigen Diluted HIV-2 189 11.79 0.52 4.4 0.91 7.7
and/or antibodies to HIV-1/HIV-2 may not detect all infected individuals. medium
A negative test result does not exclude the possibility of exposure to Anti-HIV-1 189 3.92 0.28 7.2 0.33 8.5
or infection with HIV-1/HIV-2. group O
• The AxSYM HIV Ag/Ab Combo assay does not discriminate between HIV-1 p24 162 1.59 0.07 4.2 0.10 6.4
HIV-1/HIV-2 antibodies and/or HIV p24 antigen reactivity.
Negative 414 0.38 0.02 4.5 0.02 5.5
• The presence of HIV-1/HIV-2 antibodies and/or HIV p24 antigen is
Control
not a diagnosis of AIDS. It is recommended that repeatedly reactive
specimens be investigated by supplemental testing. Individuals who Positive 414 3.84 0.21 5.3 0.25 6.6
are repeatedly reactive should be referred for medical evaluation Control 1
which may include additional testing. Positive 414 3.55 0.17 4.7 0.22 6.2
• This assay was designed and validated for use with human serum or Control 2
plasma from individual patient specimens. Data are not available to Viral Lysate 414 3.23 0.16 4.9 0.19 5.9
interpret tests performed on pooled blood. Pooled specimens must Positive
not be used. Control
• Centrifuged specimens with a lipid layer on the top of the liquid must Representative performance data are shown. Results obtained at individual
be transferred to a sample cup or secondary tube. Care must be taken laboratories may vary.
to transfer only the clarified specimen and not the lipemic material. SPECIFICITY
• Thawed specimens must be mixed by inverting 180 degrees from Specificity is based on testing of random blood donors and hospitalized
upright and return, for a total of 10 inversion cycles. Visually patient populations (serum and plasma specimens).
inspect the specimens for the absence of stratification. If layering
or stratification is observed, repeat until specimens are visibly Specificity is defined as the ability of the AxSYM HIV Ag/Ab Combo assay
homogeneous.64 to detect randomly selected specimens as negative in populations at low
risk for infection.
• Specimens containing red blood cells or particulate matter must be
clarified by centrifugation prior to testing. Refer to the SPECIMEN • Specificity based on zero prevalence of antibodies to HIV-1 or HIV-2,
COLLECTION AND PREPARATION FOR ANALYSIS section in this and/or HIV-1 p24 antigen in random blood donors (7900 tested) is
package insert. estimated to be 99.87% (7890/7900) for the AxSYM HIV Ag/Ab Combo
assay (Table 2).
• Performance has not been established using cadaver samples, or
body fluids such as urine, saliva, semen, or amniotic fluid. • Specificity based on low prevalence of antibodies to HIV-1 or HIV-2,
and/or HIV-1 p24 antigen in a hospitalized population (1938 tested)
• The AxSYM System “Automatic Sample Retest” feature must not is estimated to be 99.90% (1929/1931) for the AxSYM HIV Ag/Ab
be used due to the AxSYM HIV Ag/Ab Combo assay requirement to Combo assay (Table 2).
centrifuge all specimens prior to retesting.
The data from 7900 random blood donors from seven geographically
• Specimens from heparinized patients may be incompletely coagulated distinct regions and 1938 hospitalized patients from six geographically
and inconsistent results could occur due to the presence of fibrin. To distinct regions are summarized in Table 2. The specificity of the assay
prevent partially coagulated specimens, draw the specimen prior to was calculated as follows:
heparin treatment or into a plasma collection tube.
• Specimens containing red blood cells (0.4% v/v or greater, which are True negative specimens
x 100 = % Specificity
visibly red) may give inconsistent results with the AxSYM HIV Ag/Ab (True negative specimens +
Combo assay, and therefore must be centrifuged prior to testing. False positive specimens)

7
Table 2 SENSITIVITY
Specificity Results Using Random Blood Donors The sensitivity for anti-HIV-1 (including anti-HIV-1 group O), HIV-1 p24
and Hospitalized Patients antigen, and anti-HIV-2 is expressed in terms of detection rate using
confirmatory assay results (e.g., Western blot, HIV-1 p24 antigen,
Non-Confirmed HIV-1 PCR) as a basis for comparison.
Specimens Initially Repeatedly
The ability of AxSYM HIV Ag/Ab Combo to detect antibodies to HIV-1/HIV-2
Population Tested Reactive Reactive
and/or HIV-1 p24 antigen in individuals clinically diagnosed with HIV-1
Group (Site) (N) (N) (%) (N) (%) infection and classified disease status or from seropositive individuals
Blood Donors (Total) 7900 8 0.10 10a 0.13 (unknown disease status) is shown in Table 5.
Hospitalized • The HIV-1 antibody detection rate in a population of 615 HIV-1
1938 11 0.57 2b 0.10
Patients (5 sites) antibody confirmed seropositive individuals is 100% (615/615). This
a Two initial grayzone specimens were repeatedly reactive upon retest. rate includes 453 clinically diagnosed patients from different disease
Four of the 10 non-confirmed repeatedly reactive specimens were stages of HIV-1 infection and 55 HIV-1 subtyped samples.
indeterminate by HIV-1 or HIV-2 Western blot. None of the specimens • The HIV-2 antibody detection rate in a population of 108 HIV-2 antibody
could be confirmed by HIV-1/ HIV-2 Western blot, HIV-1 p24 antigen, confirmed seropositive individuals is 100% (108/108).
and/or HIV-1 PCR. • The HIV-1 p24 antigen detection rate in a defined (≥ 25 pg/mL)
b Six samples, which were initially and repeatedly reactive, could be population of 50 HIV-1 p24 antigen confirmed positive individuals is
confirmed by HIV-1 Western blot. One sample, which was initially and 100% (50/50).
repeatedly reactive, could be confirmed by HIV-1 PCR. • The HIV-1 group O antibody detection rate in a population of 19 HIV-1
Representative performance data are shown. Results obtained at individual group O antibody confirmed positive specimens tested is 100%
laboratories may vary. (19/19).
AxSYM HIV Ag/Ab Combo was used to test 319 specimens containing Table 5
potentially interfering substances. Results are shown in Table 3. This Detection of Anti-HIV-1 (Groups M and O), Anti-HIV-2 and/or
group contained specimens from the following categories: (viral infection) HIV-1 p24 Antigen in Serum or Plasma Specimens from Patients with
CMV, EBV, HSV, HAV, HBV, anti-HCV, anti-HTLV-I, anti-HTLV-II, rubella; Classified Disease Status and from Seropositive Individuals
(bacterial/other) fungal infection/anti-yeast (e.g., Candida albicans), Specimens
toxoplasmosis, syphilis, E. coli; (autoimmune) rheumatoid factor (RF), HIV Tested Reactive
autoimmune antibodies (ANA); (other conditions) pregnant females all
Group Infection (N) (N) (%)
trimesters, multiparous females, elevated IgG, elevated IgM, multiple
myeloma, monoclonal gammopathy, flu vaccine recipients, and human Classified Disease HIV-1 453a 453 100
antimouse antibodies (HAMA). Status
Unknown Disease HIV-1 107 107 100
Table 3 Status
Plasma Specimens Containing Potentially Interfering Substances
Subtyped HIV-1 55b 55 100
Non-Confirmed
Specimens Repeatedly Repeatedly Unknown Disease HIV-1 Group O 19c 19 100
Tested Reactive Reactive Status
Group (N) (N) (%) (N) (%) HIV-1 p24 Ag 50 50 100
Potentially Interfering HIV-2 108 108 100
319 5a,b 1.57 4 1.26
Substances Total HIV-Positive HIV-1 / HIV-2 / 792 792 100
a Four samples, which were initially and repeatedly reactive, could HIV-1 p24 Agd
not be confirmed by HIV-1/ HIV-2 Western blot, HIV-1 p24 antigen, a Specimens drawn from individuals in CDC classifications Stage A,
and/or HIV-1 PCR. These samples belong to the following categories: Stage B, or Stage C.
E. coli (2), RF (1), monoclonal gammopathy (1). E. coli infections were
b HIV-1 specimen subtypes: subtype A (11), subtype B (3), subtype
detected in urine samples.
B/D (1), subtype C (7), subtype D (5), subtype E (6), subtype F (3),
b One sample, which was initially and repeatedly reactive, could be subtype F1 (13), subtype G (5), subtype J (1).
confirmed by HIV-1 Western blot. This sample belongs to the following
c Nine of the HIV-1 group O specimens were diluted 1:100, while the
sample category: monoclonal gammopathy.
remaining 10 specimens were undiluted.
Representative performance data are shown. Results obtained at individual
d All specimens were sourced from different geographical regions:
laboratories may vary.
Argentina, China, Ghana, Ivory Coast, Mozambique, Thailand, Uganda,
AxSYM HIV Ag/Ab Combo was used to test specimens from individuals USA, Zimbabwe, Belgium, Cameroon, Germany, Iran, Italy, Kenya,
presumed to have an increased risk for blood transmissible infections. Korea, Netherlands, Nigeria, Poland, Spain, Syria, Togo, Turkey, Zaire,
Results are shown in Table 4. This group contains specimens from the Romania, Arabia, Brazil, Ethiopia, France, Great Britain, Lebanon, and
following categories: hemodialysis, hemophiliacs, multiple transfusion Russia.
recipient, homosexual males, intraveneous drug users (IVDU), and sexually
Representative performance data are shown. Results obtained at individual
transmitted diseases (STD).
laboratories may vary.
Table 4 Sensitivity of the assay to detect antibodies to HIV-1 and/or HIV-2 was
Increased Risk Population evaluated in sequential specimens from 22 seroconverting donors. These
Non-Confirmed well-characterized specimens are commercially available from Boston
Specimens Repeatedly Repeatedly Biomedica (BBI, Massachusetts, USA), North American Biologicals Inc.
Tested Reactive Reactive (Nabi, USA), Pyramid (USA) or Bioclinical Partners (BCP, Massachusetts,
Group (N) (N) (%) (N) (%) USA). The sensitivity of the assay in seroconverting donors is better than
High Risk 179 42a 23.5 0 0 the methods of comparison. The AxSYM HIV Ag/Ab Combo assay was
reactive 1-2 bleeds earlier than the method of comparison for 16 out of
a 42 specimens were confirmed by HIV-1 Western blot. These samples 22 seroconversion panels and equal in 4 out of 22 seroconversion
belong to the following categories: homosexual males (29), STD (5), panels. Two seroconversion panels were not detected by the method of
hemophiliacs (6), and multiple transfusion (2). comparison. Table 6 shows a subset of six representative seroconversion
Representative performance data are shown. Results obtained at individual panels.
laboratories may vary.

8
Table 6 Table 7
Results for Seroconverting Donors (S/CO) Analytical Sensitivity on AFSSAPS Panel
AxSYM AxSYM HIV-1 p24
HIV Ag/Ab HIV 1/2 Antigen
Day of Combo gO Sample ID (pg/mL) Lot 1 Lot 2 Lot 3
Dona- (LN 2G83) (LN 3D41) Western HIV-1 p24 S2023 500 14.85 16.70 14.54
Donor tion Result Result Blota Antigena PCRa S2024 250 8.95 9.43 7.00
BBI 0 0.42 0.49 Neg 0.4 Neg S2025 100 3.91 4.15 3.22
PRB923 7 0.41 0.48 Neg 0.4 Neg S2026 50 2.23 2.30 1.64
12 0.39 0.41 Neg 0.5 Neg S2027 25 1.35 1.45 1.16
14 0.36 0.43 Neg 0.4 Neg S2028 10 0.73 0.79 0.71
28 0.40 0.44 Neg 0.4 Neg S2029 5 0.55 0.65 0.57
30 0.38 0.49 Neg 0.4 Neg Analytical sensitivity
35 0.42 0.51 Neg 0.4 Pos 16.7 14.5 21.2
(pg/mL)
37 1.45 0.45 Neg 1.0 Pos Representative performance data are shown. Results obtained at individual
47 49.01 1.28 Neg >23.3 Pos laboratories may vary.
84 6.12 8.59 Ind 0.4 Pos
BIBLIOGRAPHY
86 5.97 8.57 Ind 0.4 Pos 1. Barré-Sinoussi F, Chermann JC, Rey F, et al. Isolation of a
145 11.56 15.25 Pos 0.4 Pos T-lymphotropic retrovirus from a patient at risk for acquired immune
161 12.49 16.39 Pos 0.5 Pos deficiency syndrome (AIDS). Science 1983;220:868-71.
BBI 0 0.30 0.42 Neg 0.3 Neg 2. Popovic M, Sarngadharan MG, Read E, et al. Detection, isolation
PRB926 2 0.37 0.54 Neg 0.3 Pos and continuous production of cytopathic retroviruses (HTLV-III) from
7 2.44 0.46 Neg 1.6 Pos patients with AIDS and pre-AIDS. Science 1984;224:497-500.
3. Gallo RC, Salahuddin SZ, Popovic M, et al. Frequent detection and
9 23.75 0.39 Neg 10.5 Pos
isolation of cytopathic retroviruses (HTLV-III) from patients with AIDS
27 8.30 10.73 Pos 0.5 Pos and at risk for AIDS. Science 1984;224:500-3.
32 8.16 10.12 Pos 0.5 Pos 4. Leitner T. Genetic subtypes of HIV-1. In: Myers G, Korber BT, Foley
BCP 0 0.39 0.40 p51, p66 0.09 <50 BT et al., editors. Human Retroviruses and AIDS 1996. Los Alamos,
9013 7 0.36 0.37 p51, p66 0.06 <50 NM: Los Alamos National Laboratory, 1996:III-28-III-40. (Available
9 0.42 0.43 p51, p66 0.09 <50 on-line at http://hiv-web.lanl.gov)
14 0.43 0.40 p51, p66 0.08 <50 5. Robertson DL, Anderson JP, Bradac JA, et al. HIV-1 nomenclature
proposal: a reference guide to HIV-1 classification. In: Kuiken CL,
18 0.36 0.41 p51, p66 0.09 58
Foley B, Hahn B, et al., editors. Human Retroviruses and AIDS 1999.
23 0.78 0.41 p51, p66 1.56 56,350 Los Alamos, NM: Los Alamos National Laboratory; 1999:492-505.
25 2.09 0.38 p51, p66 7.84 185,800 (Available on-line at http://hiv-web.lanl.gov)
BCP 0 0.48 0.47 no bands 0.361 <50 6. Clavel F, Guétard D, Brun-Vézinet F et al. Isolation of a new
9028 5 0.47 0.48 no bands 0.723 <50 human retrovirus from West African patients with AIDS. Science
16 0.50 0.46 no bands 0.482 <50 1986;233:343-6.
18 0.47 0.43 no bands 0.361 <50 7. Clavel F. HIV-2, the West African AIDS virus. AIDS 1987;1:135-40.
8. Schochetman G, George RF, editors. AIDS Testing: A Comprehensive
34 0.49 0.46 no bands 0.482 <50
Guide to Technical, Medical, Social, Legal, and Management Issues.
53 1.42 0.44 no bands 2.169 99,250 2nd ed. New York: Springer-Verlag;1994.
55 8.02 0.43 no bands 18.795 >500,000 9. Sarngadharan MG, Popovic M, Bruch L, et al. Antibodies reactive with
Nabi 0 0.45 0.31 Ind 0.31 <400 human T-lymphotropic retroviruses (HTLV-III) in the serum of patients
SV-0400 5 0.46 0.34 Ind 0.38 <400 with AIDS. Science 1984;224:506-8.
7 0.45 0.31 Ind 0.37 1,028 10. Jackson JB, Kwok SY, Sninsky JJ, et al. Human immunodeficiency
12 9.04 0.31 Ind 6.78 >743,884 virus type 1 detected in all seropositive symptomatic and asymptomatic
individuals. J Clin Microbiol 1990;28:16-9.
14 42.98 0.33 Ind >20.13 >750,000
11. Simon F, Mauclère P, Roques P, et al. Identification of a new human
23 56.84 8.38 Pos 20.13 >750,000 immunodeficiency virus type 1 distinct from group M and group O.
25 35.70 7.56 Pos 19.65 >750,000 Nature Medicine 1998;4:1032-7.
Nabi 0 0.47 0.34 Ind 0.27 484 12. Ayouba A, Souquières S, Njinku B, et al. HIV-1 group N among HIV-1-
SV-0401 4 0.55 0.33 Ind 0.48 19,427 seropositive individuals in Cameroon. AIDS 2000;14:2623-5.
7 1.30 0.36 Ind 2.57 88,176 13. Gürtler LG, Hauser PH, Eberle J, et al. A new subtype of human
11 6.80 2.25 Ind 9.36 >750,000 immunodeficiency virus type 1 (MVP-5180) from Cameroon. J Virology
14 12.16 6.79 Ind 12.06 >750,000 1994;68(3):1581-5.
14. Peeters M, Gueye A, Mboup S, et al. Geographical distribution of
18 6.73 9.96 Pos 2.42 208,979
HIV-1 group O viruses in Africa. AIDS 1997;11:493-8.
22 5.60 8.61 Pos 1.47 167,423
15. Hunt JC, Golden AM, Lund JK, et al. Envelope sequence variability and
a Data from the vendor serologic characterization of HIV type 1 group O isolates from Equatorial
Representative performance data are shown. Results obtained at individual Guinea. AIDS Res Hum Retroviruses 1997;13(12):995-1005.
laboratories may vary. 16. Hackett J Jr, Zekeng L, Brennan CA, et al. Genetic analysis of HIV
The analytical HIV-1 p24 antigen sensitivity evaluated with the AFSSAPS type 1 group O p24gag sequences from Cameroon and Equatorial
HIV antigen panel (VIH-SFTS96 Ag) across three different clinical reagent Guinea. AIDS Res Hum Retroviruses 1997;13(13):1155-8.
lots was less than 25 pg/mL (Table 7). The HIV-1 p24 antigen concentration 17. Vanden Haesevelde M, Decourt JL, De Leys RJ, et al. Genomic cloning
is calculated using a linear regression on panel members 2025-2029. and complete sequence analysis of a highly divergent African human
immunodeficiency virus isolate. J Virology 1994;68(3):1586-96.
18. Hampl H, Sawitzky D, Stöffler-Meilicke M, et al. First case of HIV-1
subtype O infection in Germany. Infection 1995;23:369-70.
19. Loussert-Ajaka I, Chaix M-L, Korber B, et al. Variability of human
immunodeficiency virus type 1 group O strains isolated from
Cameroonian patients living in France. J Virology 1995;69:5640-9.

9
20. Soriano V, Guitiérrez M, García-Lerma G et al. First case of HIV-1 46. Barin F, M’Boup S, Denis F, et al. Serological evidence for virus
group O infection in Spain. Vox Sang 1996;71:66. related to simian T-lymphotropic retrovirus III in residents of West
21. Rayfield MA, Sullivan P, Bondea CI, et al. HIV-1 group O virus identified Africa. Lancet 1985;2:1387-9.
for the first time in the United States. Emerging Infectious Diseases 47. Kanki PJ, Barin F, M’Boup S, et al. New human T-lymphotropic
1996;2:209-12. retrovirus related to simian T-lymphotropic virus type III (STLV-IIIAGM).
22. Sullivan PS, Do AN, Robbins K, et al. Surveillance for variant strains Science 1986;232:238-43.
of HIV: subtype G and group O HIV-1. JAMA 1997;278:292. 48. Kanki PJ, M’Boup S, Ricard D, et al. Human T-lymphotropic virus type
23. Peeters, M. Recombinant HIV sequences: their role in the global 4 and the human immunodeficiency virus in West Africa. Science
epidemic. In: Kuiken C, Foley B, Hahn B, et al., editors. HIV Sequence 1987;236:827-31.
Compendium 2000. Los Alamos, NM: Los Alamos National Laboratory, 49. Clavel F, Guyder M, Guétard D, et al. Molecular cloning and
2000:54-73. (Available on-line at http://hiv-web.lanl.gov) polymorphism of the human immune deficiency virus type 2. Nature
24. Montavon C, Toure-Kane C, Liegeois F, et al. Most env and gag 1986;324:691-5.
subtype A HIV-1 viruses circulating in West and West Central Africa 50. Cot MC, Poulain M, Delagneau JF et al. Dual HIV-1 and HIV-2 infection
are similar to the prototype AG recombinant virus IBNG. J Acquir in West Africa supported by synthetic peptide analysis. AIDS Res Hum
Immune Defic Syndr 2000;23:363-74. Retroviruses 1988;4:239-41.
25. Brennan CA, Lund JK, Golden A, et al. Serologic and 51. Simon F, Cot M-C, Lesager C, et al. Differentiation between HIV-1 and
phylogenetic characterization of HIV-1 subtypes in Uganda. AIDS HIV-2 infection by radioimmunoprecipitation and synthetic peptides in
1997;11:1823-32. double reactive sera. AIDS 1989;3:401-4.
26. Vidal N, Peeters M, Mulanga-Kabeya C, et al. Unprecedented degree of 52. Norrby E, Biberfeld G, Chiodi F, et al. Discrimination between antibodies
human immunodeficiency virus type 1 (HIV-1) group M genetic diversity to HIV and to related retroviruses using site-directed serology. Nature
in the Democratic Republic of Congo suggests that the HIV-1 pandemic 1987;329:248-50.
originated in Central Africa. J Virology 2000;74:10498-507. 53. Gürtler LG, Zekeng L, Simon F, et al. Reactivity of five anti-HIV-1
27. Hussein M, Abebe A, Pollakis G, et al. HIV-1 subtype C in commercial subtype O specimens with six different anti-HIV screening ELISAs and
sex workers in Addis Ababa, Ethiopia. J Acquir Immune Defic Syndr three immunoblots. J Virol Methods 1995;51:177-84.
2000;23:120-7. 54. Loussert-Ajaka I, Ly TD, Chaix ML, et al. HIV-1/HIV-2 seronegativity in
28. Van Harmelen JH, Van Der Ryst E, Loubser AS, et al. A predominately HIV-1 subtype O infected patients. Lancet 1994;343:1393-4.
HIV type 1 subtype C-restricted epidemic in South African urban 55. Devare SG, Desai Sm, Dawson GJ, et al. Diagnosis and monitoring
populations. AIDS Res Hum Retroviruses 1999;15:395-8. of HIV-1 and HIV-2 infection. In: Khan NC, Melnick JL, editors.
29. Couturier E, Damond F, Roques P, et al. HIV-1 diversity in France, Human Immunodeficiency Virus: Innovative Techniques for Isolation
1996-1998. AIDS 2000;14:289-96. and Identification (Monographs in Virology, vol.18). Basel: S Karger;
30. Fransen K, Buvé A, Nkengasong JN, et al. Longstanding presence in 1990; 18:105-21.
Belgians of multiple non-B HIV-1 subtypes. Lancet 1996;347:1403. 56. Kessler HA, Blaauw B, Spear J, et al. Diagnosis of human
31. Weniger BG, Takebe Y, Ou CY, et al. The molecular epidemiology of immunodeficiency virus infection in seronegative homosexuals
HIV in Asia. AIDS 1994;8(S2):S13-28. presenting with an acute viral syndrome. JAMA 1987;258:1196-9.
32. Russell KL, Carcamo C, Watts DM, et al. Emerging genetic diversity 57. Phair JP. Human immunodeficiency virus antigenemia. JAMA
of HIV-1 in South America. AIDS 2000;14:1785-91. 1987;258:1218.
33. Tanuri A, Swanson P, Devare S, et al. HIV-1 subtypes among blood 58. Allain J-P, Laurian Y, Paul DA, et al. Serological markers in early
donors from Rio de Janeiro, Brazil. J Acquir Immune Defic Syndr stages of human immunodeficiency virus infection in haemophiliacs.
1999;20:60-6. Lancet 1986;ii:1233-6.
34. Marlink RG, Ricard D, M’Boup S, et al. Clinical, hematologic, and 59. Kenny C, Parkin J, Underhill G, et al. HIV antigen testing. Lancet
immunologic cross-sectional evaluation of individuals exposed 1987;1(8532): 565-6.
to human immunodeficiency virus type-2 (HIV-2). AIDS Res Hum 60. Wall RA, Denning DW, Amos A. HIV antigenaemia in acute HIV
Retroviruses 1988;4:137-48. infection. Lancet 1987;1(8532):566.
35. Simon F, Matheron S, Tomalet C, et al. Cellular and plasma viral load 61. Stute R. HIV antigen detection in routine blood donor screening. Lancet
in patients infected with HIV-2. AIDS 1993;7:1411-7. 1987;1(8532):566.
36. Kanki PJ, Travers KU, MBoup S, et al. Slower heterosexual spread of 62. Goudsmit J, De Wolf F, Paul DA, et al. Expression of human
HIV-2 than HIV-1. Lancet 1994;343:943-6. immunodeficiency virus antigen (HIV-Ag) in serum and cerebrospinal
37. DeCock KM, Adjorlolo G, Ekpini E, et al. Epidemiology and transmission fluid during acute and chronic infection. Lancet 1986;ii:177-80.
of HIV-2: why there is no HIV-2 pandemic. JAMA 1993;270:2083-6. 63. von Sydow M, Gaines H, Sonnerberg A, et al. Antigen detection in
38. Gao F, Yue L, Robertson DL, et al. Genetic diversity of human primary HIV infection. Brit Med J 1988;296:238-40.
immunodeficiency virus type 2: evidence for distinct sequence subtypes 64. Constantine NT, Abouseif N, Fox E, et al. To mix or not to mix: the
with differences in virus biology. J Virology 1994;68:7433-47. effects of not mixing sera on HIV serologic results. Laboratory Medicine
39. Yamaguchi J, Devare SG, Brennan CA. Identification of a new HIV-2 1990;21:749-51.
subtype based on phylogenetic analysis of full-length genomic 65. Box GEP, Hunter WG, Hunter JS. Statistics for Experimenters; An
sequence. AIDS Res Hum Retroviruses 2000;16:925-30. Introduction to Design, Data Analysis, and Model Building. New York:
40. Dawson GJ, Heller JS, Wood CA, et al. Reliable detection of John Wiley & Sons, Inc, 1978: 510-539, 571-583.
individuals seropositive for the human immunodeficiency virus (HIV) 66. SAS Institute Inc. SAS Technical Report P-229, SAS/STAT Software:
using Escherichia coli-expressed HIV structural proteins. J Infect Dis Changes and Enhancements, Release 6.07. Cary, NC: SAS Institute,
1988;157:149-55. 1992:289-366.
41. Montagnier L, Clavel F, Krust B, et al. Identification and antigenicity
of the major envelope glycoprotein of lymphadenopathy-associated All trademarks are property of their respective owners.
virus. Virology 1985;144:283-9.
42. Barin F, McLane MF, Allan JS, et al. Virus envelope protein of HTLV-III Abbott
represents major target antigen for antibodies in AIDS patients. Science Max-Planck-Ring 2
1985; 228:1094-6. 65205 Wiesbaden
43. Schulz TF, Aschauer JM, Hengster P, et al. Envelope gene-derived Germany 0088
recombinant peptide in the serodiagnosis of human immunodeficiency + 49-6122-580
virus infection. Lancet 1986;2:111-2.
44. Allan JS, Coligan JE, Barin F, et al. Major glycoprotein antigens that March 2008
induce antibodies in AIDS patients are encoded by HTLV-III. Science © 2002, 2008 Abbott Laboratories
1985;228:1091-4.
45. Hunt JC, Johnson-Paepke J, Boardway K, et al. Discrimination between
HIV-1 and HIV-2 seropositive individuals using mouse monoclonal
antibodies directed to HIV transmembrane proteins. AIDS Res Hum
Retroviruses 1990;6:883-98.

10

Você também pode gostar