1. GENERALITIES
The supply module DL 10017 has been planned to satisfy the requires of the MICROLAB electric
machine laboratory: all the deliveries are increased as regards the power effectively required by the
machines under test.
The regulation of the ac/dc variable voltages can be carried out manually by the operator or
automatically by means of the DE LORENZO data acquisition system and relative software.
The safety of the operator is guaranteed by a differential magnetothermic general switch and by a
particular connector for the connection of the maximum speed protection of the machines under
test.
All the variable voltage outputs with single control are protected by magnetothermic automatic
while the direct current output that can be manually varied is protected by magnetothermic circuit
breaker.
Since all the output circuit are galvanically isolated from the mains and they are carried out in low
voltage the safety standards are implicitly satisfied.
ver a 20110415 1
DL 10017
2. ELECTRICAL CHARACTERISTICS
The module DL 10017 is arranged in the factory for a three-phase supply voltage corresponding to
the local mains.
Before connecting the supply cable to the mains it is a good idea to check that the voltage is the one
shown in the data plate placed on the back of the module.
2.3 Rectified direct current outputs with double single phase bridge.
Variable voltage from 0 to 42 V/5 A.
ver a 20110415 2
DL 10017
Front panel
S6 S9
S5 S7 H3 H4 S11
H1 H2
S1 AC H5 H6 DC
off 40 60
man auto AC off on off on
S2 a b
20 80
T1A
ver a 20110415 3
DL 10017
Back panel
ver a 20110415 4
DL 10017
4. OPERAZIONI PRELIMINARI
The module DL 10017 has to be connected to the supply three-phase mains by means of the back
cable.
After having checked the value of the mains voltage and the one shown on the data plate, the
conductors (3P+N+PE) have to be connected respecting the shown cyclic direction of the phases.
4.1 Supply
After having connected the module DL 10017 to the three-phase mains, close the Q1 differential
magnetothermic switch Q1 set on the back..
The mains voltage is present: the back XS socket is active.
In these conditions it is possible to use only the direct current variable output 0 ÷ 42 V/5 A, closing
the S10 switch (on position): the H3 lamp switches on and the output voltage at the corresponding
terminals (X11 – X12) can be regulated by means of the T1A knob.
We remember that the T1A voltage changer is protected by means of the F magnetothermic circuit
breaker.
ver a 20110415 5
DL 10017
The deliveries of the ac/dc variable voltages controlled in "man" or "auto" mode are possible only if
the protection remote controlled switch internal to the module DL 10017 is enabled.
All the MICROLAB equipments foreseeing an instrument for the measurement of the machine
rotation speed are equipped with a connector identified with the abbreviation K1: this connector has
to be connected to the analogue K1 connector of the module DL 10017 by means of the suitable
cable supplied with the equipment themselves each time that we want to foresee the protection
against the maximum speed and this to enable the internal remote controlled switch in lower speed
conditions at the limit foreseen for the intervention of the protection itself that, disabling the remote
controlled switch, nullifies the delivery.
Alternatively, of no maximum speed protection is foreseen, the it is possible to enable the internal
remote controlled switch by using the flying connector equipped with the module DL 10017: in this
case no protection is foreseen and the delivery can be nullified only with the intervention of the
operator on the output switches or on the "off" palm push-button.
N.B.: If the K1 connector of the module DL 10017is not connected the internal remote controlled
switch is not enabled and no ac/dc variable delivery is possible.
ver a 20110415 6
DL 10017
5. START-UP
The regulation of the direct voltage 0 ÷ 42 V/5 A is possible only manually by means of the T1A
knob (see point 4.1).
The regulation of the ac/dc variable voltages can be carried out manually by the operator or
automatically by means of the software.
S6 S9
S5 S7 H3 H4 S11
H1 H2
S1 AC H5 H6 DC
off 40 60
man auto AC off on off on
S2 a b
20 80
T1A
5. Verify that the S7 and S11 switches are open (off position).
7. Activate the module DL 10017 by closing the Q1 differential magnetothermic switch set on the
back.
ver a 20110415 7
DL 10017
S6 S9
S5 S7 H3 H4 S11
H1 H2
S1 AC H5 H6 DC
off 40 60
man auto AC off on off on
S2 a b
20 80
T1A
3. By means of the S8 selector set the voltage range that you want.
Position "a": 0 ÷ 26 V/14 A.
Position "b": 0 ÷ 46 V/10 A.
4. Push the S3 "start" push-button and activate the output by setting the S7 switch to position
"on".
Attention
When the S8 selector is in "off" position, or when the voltage range is changed with consequent
passage for this position, the ac/dc variable outputs are deactivated and it is necessary to act on the
S3 "start" push-button again after the S8 selector has been set to the wanted range
5 Turn the RP knob to carry out the voltage vale that you want.
6. At the end of the tests set in the order the S7 switch and the S8 selector to "off" position.
Set the RP knob to "min" position again.
The module DL 10017 remains ready for further applications.
ver a 20110415 8
DL 10017
S6 S9
S5 S7 H3 H4 S11
H1 H2
S1 AC H5 H6 DC
off 40 60
man auto AC off on off on
S2 a b
20 80
T1A
3. By means of the S8 selector set the voltage range that you want.
Position "a": 0 ÷ 32 V/14 A.
Position "b": 0 ÷ 46 V/14 A.
4. Push the S3 "start" push-button and activate the output by setting the S11 switch to position
"on".
Attention
When the S8 selector is in "off" position, or when the voltage range is changed with consequent
passage for this position, the ac/dc variable outputs are deactivated and it is necessary to act on the
S3 "start" push-button again after the S8 selector has been set to the wanted range.
5. Turn the RP knob to carry out the voltage vale that you want.
6. At the end of the tests set in the order the S11 switch and the S8 selector to "off" position.
Set the RP knob to "min" position again.
The module DL 10017 remains ready for further applications.
ver a 20110415 9
DL 10017
S6 S9
S5 S7 H3 H4 S11
H1 H2
S1 AC H5 H6 DC
off 40 60
man auto AC off on off on
S2 a b
20 80
T1A
2. Connect the X1-X2 terminals for the "auto analog" checking to the ANALOG OUT terminals
corresponding to the DE LORENZO acquisition system.
Checking analogue signal: 0 ÷ 10 Vcc.
3. Connect the X3-X4 terminals for the "auto start-stop" control to the DIGITAL OUT terminals
corresponding to the DE LORENZO acquisition system.
Control on-off signal: NC contact, activated outputs;
NO contact, deactivated outputs.
4. Verify that the control and checking software is set-up in the right way and that it operates.
8. Verify that the S7 and S11 switches are open (off position).
9. Activate the module DL 10017 by closing the Q1 differential magnetothermic switch set on the
back: the H1 lamp switches on.
ver a 20110415 10
DL 10017
S6 S9
S5 S7 H3 H4 S11
H1 H2
S1 AC H5 H6 DC
off 40 60
man auto AC off on off on
S2 a b
20 80
T1A
ver a 20110415 11
DL 10017
S6 S9
S5 S7 H3 H4 S11
H1 H2
S1 AC H5 H6 DC
off 40 60
man auto AC off on off on
S2 a b
20 80
T1A
3. By means of the S8 selector set the voltage range that you want.
Position "a": 0 ÷ 32 V/14 A.
Position "b": 0 ÷ 46 V/14 A.
ver a 20110415 12
DL 10017
6. MAINTENANCE
The supply module DL 10017 has no need of any special maintenance intervention.
In case of bad operation, or of fault, AFTER HAVING DISCONNECTED THE MODULE FROM
THE THREE-PHASE MAINS, remove the upper cover and look inside: check that the components
(switches, selectors, lamps...) and the electrical connections are efficient.
Check the fuse placed on the control electronic card of the motor.
When you think that the module DL 10017 needs some particular interventions, we suggest to call a
specialized staff of to call De Lorenzo.
ver a 20110415 13