Escolar Documentos
Profissional Documentos
Cultura Documentos
The main details of this method are taken specifically from Standard
Methods for the Examination of Water and Wastewater (Method 507:
1985, and Method 218B: 1971) and the United States Environmental
Protection Agency of 1979 (Method 405.1). Slight variations and
additional insight are added from my experience as an analyst, and
modifications will be noted. Other methods may exist amongst
laboratories performing this test, so it must be stressed that this
method, although approved, is not all inclusive. In addition, this
procedure is only suitable for samples void of serious matrix
interferences. To gain a broader appreciation of oxygen demand,
additional avenues of interest may be explored including CBOD
(carbonaceous oxygen demand), COD (chemical oxygen demand),
and TOC (total organic carbon).
Equipment
Reagents
Sample Preparation
Procedure
Analysis
Calculations
Equipment
The following is a list of necessary materials and equipment for starting the
procedure:
Series of 250-300 ml BOD bottles with ground glass stoppers, and caps
pH meter
Millipore water
Seed
50 ml class A buret
Series of beakers
Deionized water
Reagents
Starch indicator
Bleach (5.25%)
Acetic Acid
Potassium Iodide
Manganese sulfate
Sodium sulfite
The remaining reagents are prepared within the laboratory. Caution must be
taken since the shelf life of these reagents should not exceed 6 months
unless otherwise noted.
1. 0.1 N Sulfuric Acid: Add 2.8 mls concentrated acid to 1 liter distilled
water.
7. 1+1 Acetic Acid: 500 ml pure grade acetic acid is added to 500 ml
deionized water with stirring. Caution must be taken since the
temperature of the solution will increase rapidly.
Sample Preparation
Chlorine Content
Procedure
Once reagents have been properly prepared, the sample is ready for
analysis. The BOD procedure including the DO analysis is actually
quite lengthy, so time maintenance is important. This portion can be
carried out in four separate steps including carboy setup, adjustment
of DO, preparation of seed inoculum, and BOD sample preparation.
It is best to set up the carboy as soon as possible to ensure proper
results. In addition, the BOD sample preparation is the last step of
the procedure.
Carboy Set Up
Seed Preparation
DO Preparation
Carboy Set Up
The BOD procedure calls for two carboys to be used each time the
procedure is carried out. The carboys will contain the water
necessary for the procedure, and two are used instead of one to
serve as a source of comparison amongst carboys. Also, it is a good
idea to have two sources of water in case something goes terribly
wrong with one of the carboys, there will be some sort of back up.
To ensure that the two carboys involved in the procedure are free of
contaminants, the carboys must first be rinsed with acid and water.
Sufficient water should be rinsed to eliminate all traces of acid. Once
the acid is flushed from the carboys, a small amount of bleach is
added to eliminate excess organisms that may serve as sources of
interference. The carboys are again flushed with adequate deionized
water until the presence of bleach is eliminated.
The carboys are then filled with pure water that has been finely filtered.
Millipore water seems to work best in this procedure. The carboys
should be filled up to supply atleast six liters for the first sample, and
three liters for each successive sample. The six liters for the first
sample are necessary because quality controls will also be included
as part of the run. Once filled, one phosphate buffer pillow is added
to the carboy per six liters water. These pillows are commercially
available, and save the analyst time from preparing the reagents.
After addition, the water solutions must be aerated until the point of
saturation. Commonly, many laboratories will prepare this water the
day before the procedure, but five hours prior to analysis appears to
be sufficient. After aeration, the water is ready to be used to fill the
BOD bottles, and must be capped until then.
Seed Preparation
The seed solution must be set up some time before the BOD bottle are
filled with water. This is done by adding 0.045 grams of a polyseed
inoculum (or sometimes a BOD sample can be used as the
inoculum) to 250 mls distilled water. The seed will not dissolve but it
is important that the seed is stirred continously at moderate speed
for about two hours. At approxiamtley 30 minutes prior to use, the
seed solution is allowed to settle undisturbed. Caution must be taken
not to disturb the seed particles because the liquid portion of the
solution is used in the procedure. Remnants of actual seed within
the prepared BOD bottles could greatly effect the results.
DO Preparation
Once the DO readings are calculated, the two bottles that were withdrawn
at the beginning are now used. The carboy bottle from which the
samples will be made up from is used to adjust the DO meter (in our
case, the pH meter). Once adjusted, the DO of the other bottle is
measured to determine whether the meter is reading correctly. If the
meter reading comes within 0.1 of the Winkler reading, the
calibration is complete. If not, the entire Winkler procedure must be
completed for both carboys.
Finally, the last step in the BOD procedure involves inoculating the sample
with various dilutions along with standards and blanks for quality
control. The following quality control must always accompany each
BOD run:
- 4 standards
- 2 seeded blanks
Usually, many laboratories will include one set of QC for every ten
samples. Additional QC is necessary for more than 10 samples.
Preparation:
4. Samples- 4-5 bottles are usually necessary for each sample. Some
samples may have to be diluted in order for the DO range to be
detected by the meter. Observation of the sample will usually give
an indication to its dilution. Clean samples usually require small
dilutions whereas wastewater samples will need high dilutions due to
their high BOD values. Once reasonable dilutions have been
determined, the specific volume of sample is added to each bottle
along with 2 mls polyseed. The bottles are filled to the rim with
carboy water.
Once all the bottles have been filled, the initial DO's of each solution is
determined on the meter and recorded. Once recorded, the bottles
are capped with ground glass stoppers to avoid excess bubbles and
capped. The bottles are placed in an incubator at approxiamtely 20
degrees celsius where they will remian for five days.
Analysis
After five days of incubation, the samples are ready to be analysed. The
samples are removed from the incubator and allowed to equilibrate
to room temperature. In the meantime, the analyst should calibrate
the meter again with the carboy water as discussed in the DO
Sample Preparation section. It may be desired to use fresh carboy
water for the calibration as discussed in the Carboy Set Up page.
Once the meter is calibrated, the samples are read starting with the blanks
and ending with the actual samples. The final DO of each solution is
recorded and the initial and final readings will be used to calculate
the BOD. The best results come about when the initial and final DO
values for the blanks are similar indicating the absence of organisms
and reliable equipment. The blank DO should normally be less than
0.2 mg/L
BOD Calculations
The calculations for BOD take into account the unseeded blanks and the
seeded solutions. These values must be subtracted out in order to
obtain reasonable BOD results. The following calculations are taken
from Standard Methods (Method 507: 1985, p. 531).
A. The BOD of the blanks are calculated by subtracting the final DO from
the initial DO:
BOD blank=DO1-DO2;
B. The BOD of the seeds are calculated by subtracting the final DO from
the initial DO and multiplying this number by the dilution factor:
If the testing procedure was carried out correctly and the dilutions of the
sample were made appropriately, the analyst should have obtained
BOD values that are within a reasonable percent error and relative
percent difference. Generally, values are discarded for a specific
sample dilution if the final DO of the sample is < 1.0 mg/L of if delta
DO is < 1.0 mg/L. This also stresses the importance of using
different dilutions for each sample to key in on the appropraite BOD
when little is known about how the sample will react and how high its
BOD levels are.
Why 5 days ?
Note the vertical line for 5 days. If the samples are quite different in their
composition, the error in comparing them at 5 days will be great, and
a longer time for the test would be better. This must be balanced by
a long wait before having results, and delay in making adjustments
based on these results may be costly.
DO Preparation
Once the DO readings are calculated, the two bottles that were withdrawn
at the beginning are now used. The carboy bottle from which the
samples will be made up from is used to adjust the DO meter (in our
case, the pH meter). Once adjusted, the DO of the other bottle is
measured to determine whether the meter is reading correctly. If the
meter reading comes within 0.1 of the Winkler reading, the
calibration is complete. If not, the entire Winkler procedure must be
completed for both carboys.