Você está na página 1de 10

Validation of Sterile

Filtration
Leesa McBurnie and Barry Bardo*

This article discusses the

T
he Parenteral Drug Association
basics of sterile filter (PDA) published the authorita-
qualification and validation, tive summary of best practices
with emphasis on bacterial in sterile filtration and validation of
challenge protocol sterile filtration in its 1998 technical
development and testing. report, “Sterilizing Filtration of Liq-
Reference is made to uids” (1). It highlights the history of
Technical Report no. 26 of sterile filtration, explains how filters
the Parenteral Drug work, details selection criteria, and
explains validation considerations
Association.
and integrity testing methods.
Much of this article is based on that
monograph.
The purpose of sterile filtration
validation is to prove that a particu-
lar filtration process generates a ster-
ile filtrate. This is achieved by choos-
ing a sterilizing grade filter that is
compatible with the process, non-
toxic, integrity testable, sterilizable,
that does not adsorb formula com-
ponents or add extractables to the
Leesa McBurnie is senior process and can remove the biobur-
microbiologist and Barry den associated with the product.
Bardo is director of business
The filter then is challenged with
development at Meissner
Filtration Products, Inc., .107 colony forming units (cfu) of
Camarillo, CA 93012. Brevundimonas diminuta (ATCC
barry.bardo@meissner.com. 19146) per cm2 under process con-
*To whom all correspondence ditions and demonstrated by testing
should be addressed. to produce a sterile filtrate.

Pharmaceutical Technology FILTRATION 2004 s13


Filtration

Sterilizing grade filters Qualifying pharmaceutical filters


The removal rating of a given mem- Safety and purity. Before they can be
brane filter type refers to the size, or used in a sterilizing filtration
narrower dimension, of microor- process, filters must meet or exceed
ganisms and particles removed by minimum safety standards.
the filter, rather than the actual size The first step is to qualify the filter
or shape (i.e., morphology) of the in several important areas of safety.
filter pore structure. The industry- To qualify as a pharmaceutical filter,
accepted rating for a sterilizing the filter must be nontoxic, according
grade filter is 0.2 or 0.22 mm, de- to USP-specified tests (e.g., USP ^87&
pending on the manufacturer, which “Biological Reactivity Tests,” in vitro,
is validated as capable of removing USP ^88& “Biological Reactivity Tests,”
.107 cfu/cm2 B. diminuta under in vitro, USP ^88& “Biological Reactiv-
certain extreme processing condi- ity Tests,” in vivo, including the Class
tions. Tighter filters such as 0.1 mm VI Plastics Tests) and be tested free of
filters that demonstrate the same pyrogen or endotoxin to acceptably
bacterial retention also may be rated low levels.
as sterilizing grade. Pharmaceutical-grade filters also
In validating and performing must have very low extractables levels,
sterile filtration, it is essential to although neither USP nor the Food
identify the bioburden or endemic and Drug Administration have speci-
microorganism(s) in a given fied minimum or maximum levels. It
process, to use the grade of filter is the responsibility of pharmaceutical
that quantitatively removes the mi- manufacturers to set allowable specifi-
croorganism(s), and to demonstrate cations for extractables.
quantitative removal by test before As a further indicator of safety
using the filter in production. This and purity, filters also should rinse
is the essence of filter validation. quickly when exposed to high-
For filter manufacturers, a critical purity water, must be compatible
requirement is to provide users with with pharmaceutical process fluids
a reasonably convenient, safe, easy- and pharmaceutical products, and
to-perform integrity test which con- be both sterilizable and integrity
firms the integrity of the filter, the testable.
seals, and contiguous process equip- All this pharmaceutical filter safety
ment. Liquid-sterilizing filters can qualification information typically is
be integrity tested by the bubble provided in the filter manufacturers’
point, forward flow or diffusion/ validation guide and product litera-
diffusive flow test, or the pressure ture for a specific filter.
hold test. Hydrophobic filters can be Performance qualification. Filters
tested with any of those methods or must be qualified by the user to
by the water intrusion test. Filter in- demonstrate that their performance
tegrity tests are explained in another in processing will meet or exceed
article in this issue. minimum process requirements.

s14 Pharmaceutical Technology FILTRATION 2004 www.phar mtech.com


Filtration

atures of normal process conditions,


Performance qualification occasional runaway process condi-
requirements tions, and the temperature ranges of
• flow rates processing (minimum to maxi-
• throughput mum) and steam or autoclave steril-
• pressure and temperature resistance ization, including both temperature
• hydrophilic or hydrophobic ranges and duration of the steriliza-
• membrane composition tion cycle(s).
• compatibility Sterilizing-grade filters for aque-
• membrane support layers,core,or cage ous pharmaceutical liquids are nor-
• o-rings mally hydrophilic, or water-
• housings wettable, membrane filters. In the
case of solvent or chemical liquids
to be filter-sterilized, hydrophobic,
The filter must be tested to verify or nonwater-wetting, filters may be
that it provides the flow rates re- used. They can be wetted by a low-
quired by the pharmaceutical surface-tension liquid.
process. The filter system must be Compatibility. The filter system
sized to provide flow rates and vol- must be qualified to ensure that all
umes adequate to keep pace with product-contact surfaces of the filter
filling machines or other production and its constituent parts (mem-
equipment requirements, with some brane, support layers, core, cage,
reserve capacity for use in case of and end caps), o-rings, piping,
batch contaminant variability and hoses, seals, pumps, gaskets, and any
premature plugging. The total liquid other components of the sterilizing
volume passed through the filters, filtration system can withstand the
its throughput, should be adequate hydraulic, thermal, and chemical
to process a complete batch without challenges of the sterilization and
interruption. production processes. None of these
Small-scale sizing or filterability should extract into the filtered phar-
tests are used as the basis for extrap- maceutical product in any signifi-
olating or scaling-up filtration sys- cant amount.
tems. Sizing of final, sterilizing fil- Chemical compatibility questions
ters, and any upstream prefilters generally are resolved by reference
used to remove coarse contaminants to compatibility tables generated by
and thereby extend the life of the manufacturers of elastomers or
final filters, is based upon antici- polymers used in o-rings, gaskets,
pated flow rates and throughput in a and seals. Membrane compatibilities
given pharmaceutical liquid. generally are well established for the
Sterilizing filters and filter hous- commonly used membrane materi-
ings, stainless steel, or disposable als. Any specific questions can read-
plastic, must be rugged enough to ily be resolved by testing during the
withstand the pressures and temper- qualification stage.

s16 Pharmaceutical Technology FILTRATION 2004 www.phar mtech.com


Sterilization of sterilizing filters. facturer’s validation guide. Product
Sterilizing grade filters can be steril- integrity test values are correlated to
ized in a number of ways. Capsule the water or model solvent values.
filters can be gamma irradiated or Bacterial retention. The bacterial
autoclaved. Disk filter holders are challenge test validates the ability of
autoclaved with the wetted filter in a filter to provide sterile effluent in a
place. Cartridge filter installations specific pharmaceutical liquid. It is
frequently are sterilized by steam- also the ultimate compatibility test,
in-place (SIP) operations. because the bacterial challenge si-
Common steaming temperatures multaneously tests the physical-
used in the United States are 121– chemical interaction of the liquid
135 8C, sustained for 30–60 min in product and the filter, under process
the filter installation. Whatever conditions. Any filter inadequacy
time–temperature parameters are caused by this interaction will be
specified, it is critical that these pa- detected by the bacterial challenge.
rameters be validated by the phar- Validation of bacterial retention
maceutical manufacturer under op- normally is performed by the filter
erating conditions. manufacturer or an independent lab-
oratory, using 47-mm diameter disks
Validation to minimize the volume of pharma-
There are four major elements of ceutical product required. Larger sur-
the filtration validation process: face area filters also can be used.
• physical/chemical compatibility, Bacterial challenge tests are usu-
usually established during the ally performed with an industry
qualification phase before valida- standard concentration of 107 cfu of
tion, is confirmed during the val- B. diminuta per cm2, using pharma-
idation process ceutical product, whenever possible,
• binding and adsorption filter for the most realistic validation. The
characteristics are measured in high bacterial concentration used in
the qualification phase the challenge test constitutes a
• bacteria retention capability of the worst-case scenario. The manufac-
filter, which is established by chal- turer qualifies the filter using a simi-
lenging the filter with B. diminuta lar challenge.
• integrity of the process filtration B. diminuta is grown to produce
installation, as verified by the fil- monodispersed cells capable of pen-
ter integrity test. etrating a 0.45-mm filter, typically in
Concerning integrity testing, the accordance with ASTM Standard
user must demonstrate that they F838. Following that standard, the
know how to install, sterilize, and organism is cultured in saline lac-
integrity test the filters. Filter in- tose broth (SLB) and either used
tegrity test values provided by the freshly cultured or concentrated
filter manufacturer are correlated to into a frozen cell paste that is
the bacterial challenge in the manu- thawed immediately prior to use. A

Pharmaceutical Technology FILTRATION 2004 s17


Filtration

0.4-µm rated membrane filter is Validation parameters


used as a positive control; if the or-
ganisms pass through this control • product contact time
filter, it proves that they are alive, • differential pressure
small, and nonaggregated, thus veri- • flow rates per unit area
fying that the challenge is a strin- • temperature
gent test. • bioburden
Validation parameters for bacterial • integrity test correlation
challenge using pharmaceutical prod-
uct. Key considerations for using sound rationale to support the deci-
pharmaceutical product liquid to sion and use it in challenge testing.
validate sterile filters are listed in the Temperature. If the liquid tempera-
following chart. ture is outside the viable range of
Product contact time. The bacterial the challenge bacteria, it may be
challenge using pharmaceutical necessary to recirculate the product
product must be run for at least the at process temperature, condition-
same duration as a product batch ing the filter first, and then chal-
will be run in processing. If the lenge the filter at a temperature at
batch requires eight hours to filter, which the bacteria survive.
the challenge must be run for at Bioburden. Bioburden levels can in-
least eight hours. It is good practice fluence process filtration efficacy.
to run the challenge for a little The probability of passage increases
longer to anticipate unusual pro- when the bioburden is high. The
cessing circumstances. area-specific bioload (Ba) is the
Differential pressure and flow rates per unit bioburden per unit area of filter
area. Maximum process differential (cfu/cm2) or Ba = BV/A, where B is
pressures and flow rates should be the bioburden in cfu/mL, V is the
incorporated into the protocol. total volume to be filtered (mL) and
Often, it is virtually impossible to A is the surface area of the filter in
match both simultaneously. At the cm2. It is therefore best to control
start of the challenge, when the filter the bioburden of the raw materials
is clean, extremely high flow rates to avoid approaching or exceeding
per unit area are necessary to gener- the validated limit.
ate process-level differential pres- Integrity test correlation. Filters used in
sures (pressure drop across the fil- the bacterial challenge must be in-
ter). But, as the test filter disks tegrity tested to form the correlation
accumulate bacteria and pressure to retention. Because the user can-
builds, the flow drops. One solution not use a destructive challenge test
is to match flow rates at the start of in processing, the filter manufac-
the challenge and pressures near the turer must supply a correlated, non-
end. Another solution is to decide destructive integrity test that reli-
whether pressure or flow is more rel- ably assesses the integrity of a given
evant, and then develop a technically filter installation. In performing the

s18 Pharmaceutical Technology FILTRATION 2004 www.phar mtech.com


Select filter

Filter extractable Microbial retentivity Physical parameters


validation validation validation

Determine and document Develop nonbactericidal


viability of B. diminuta surrogate with same product
in product under characteristics for toxic,
process conditions high abuse potential
or limited supply drugs

B. diminuta viable in B. diminuta not viable


product under process in product under process
conditions conditions

Direct inoculation of Precondition filter with product–


B. diminuta into product followed by microbial challenge.
under process conditions One or a combination of the following
methods may possibly be used

Modify process– Modify formulation– Use product for Change from B. diminuta–
adjust temperature, etc. adjust pH, remove time period challenge use bacteria isolated from
bactericidal component organism is viable formulation or environment
product surrogate

Figure 1: Establishing a microbial challenge protocol (1).

challenge test, three lots of filters pharmaceutical product is compared


typically are chosen, including a to a control over time. If there is less
low-bubble-point lot, one with a than one log difference in the counts,
bubble point specification that ap- the organism is considered viable
proximates the manufacturer’s pro- under those conditions. Results are
duction limit. This is important be- documented. If a pharmaceutical
cause the end user is limited to product is toxic, has high potential
using filters in their process that for abuse, or is in limited supply, a
match or exceed the lowest bubble nonbactericidal surrogate fluid with
point validated in product. very similar properties may be devel-
Establishing the challenge protocol. oped. It is recommended that the
The above chart highlights the steps user consult with local FDA inspec-
needed to develop a suitable chal- tors before developing this approach.
lenge protocol. If the microbe is viable. If B. diminuta
After the filter is qualified and se- is found to be viable in the pharma-
lected, filter extractables and micro- ceutical product, it is directly inocu-
bial retentivity are validated. Physi- lated into the product as the chal-
cal parameters such as sterilization lenge microorganism.
and integrity test methods also are If the microbe is nonviable. If B. dimin-
validated. Viability of the test mi- uta is found to be nonviable under
croorganism, B. diminuta, in the process conditions, the test filter is

Pharmaceutical Technology FILTRATION 2004 s19


Filtration

the apparatus, integrity testing post-


Factors affecting microbial sterilization and recirculating prod-
retention uct through the filters at process
• filter type:polymer,structure flow rates or pressures. Bacteria are
• fluid properties:pH,viscosity,osmolarity, added according to the results of the
and ionic strength viability study. In nonbactericidal
• process conditions:temperature,differential products, the microorganisms are
pressure,and flow rate added at intervals throughout the
• bioburden:quantity,composition challenge. Before introducing the
bacteria, analysis membranes are in-
stalled downstream of all test and
preconditioned with pharmaceutical control filters. Analysis filters must
product liquid, and one of several have a pore size of at least 0.45 mm
options can be followed, prior to or tighter. Membranes that are 0.45
conducting the bacterial challenge. mm will retain 104–106 logs of Bre-
Options are to: vundimonas, which generally is ac-
• modify the process by adjusting cepted as adequate for a valid test.
temperature or otherwise
• modify the formulation, by ad- Bracketing
justing pH, removing bactericidal
components, or using a product • challenge extreme products
surrogate liquid • carefully choose family groupings
• use the bacteria for a brief enough • API concentration levels
time such that the challenge mi- • pH,other
croorganism retains its viability dur- • validates intermediate products
ing the test period
• switch the test organism from B. Factors affecting bacterial retention.
diminuta to a process isolate, or na- The bacterial challenge procedure
tive bioburden endemic to the ac- addresses the possible influences
tual pharmaceutical process. It is upon retention of the membrane
recommended to consult first with polymer, liquid, process conditions,
FDA inspectors on this option. and bioburden level. Adverse effects
If the product is bactericidal, the from any of these parameters would
volume and time required to thor- result in bacterial passage.
oughly rinse the filter free of the The only factor not routinely ad-
product must be established, after dressed in the filter validation or
the product has been recirculated, microbial challenge process is com-
prior to adding the bacteria and position of the bioburden. Compo-
challenge solution. This is docu- sition of bioburden should be ad-
mented in an inhibition test. dressed prior to choosing the
Bacterial challenge. The challenge removal rating of the filter, during
procedure typically consists of in- filter qualification.
tegrity testing the filters, autoclaving Bracketing. It is not always necessary

s20 Pharmaceutical Technology FILTRATION 2004 www.phar mtech.com


to perform a microbial chal-
lenge on every pharmaceuti-
cal product. The practice of
bracketing makes this ac-
ceptable. With a family of
pharmaceutical products
having high degrees of simi-
larity but only slightly differ-
ing characteristics such as
differing concentrations of
active ingredient, we can
perform the microbial chal- Figure 2: Bacterial challenge test apparatus
lenge on the extreme prod- showing differential pressure between two inline
ucts (i.e., those having the 47-mm diameter filter disk holders in series, with
highest and lowest concen- the challenge filter upstream and the analysis filter
trations) while products downstream, connected by sanitary piping to a
with intermediate-level con- peristaltic pump which recirculates the challenge
centrations need not be in- microorganism.
dividually challenge-tested.
However, we must generate data to vent, rather than pharmaceutical
show that these product extremes product. It involves soaking the fil-
have been tested. ter, boiling off the solvent, drying
Extractables. Another important as- the residue, and weighing it. The
pect of sterile filter validation is ex- weight of residue per filter is calcu-
tractables testing. It is important lated. Pharmaceutical products sel-
that the filters are not themselves a domly are used for NVR assay, be-
source of physical (e.g., particles) or cause their constituents mask filter
chemical contaminants. extractables. Various analytical tech-
niques may be applied to identify
filter extractables.
Adsorption analysis Adsorption. The flip side of the phe-
• binding of formulation components to filter
nomenon of filter extractables is the
• potential for OOS ingredients
phenomenon of filter adsorption.
• identify any problem
The filter should not remove active
• address problem if necessary
ingredients, excipients, carriers, dilu-
ents, proteins, preservatives, or any
other formulation component. It is
The nonvolatile residue (NVR) important to test the product to en-
test from USP ^661& Containers is sure that the filter does not cause any
used to quantify the amount of ex- ingredient to fall below specification.
tractables released by a filter in a With most sterilizing grade filters,
particular process stream. NVR typ- adsorptive levels are so low as to be
ically is tested by using a model sol- insignificant and well within limits

Pharmaceutical Technology FILTRATION 2004 s21


Filtration

Table I: Tests commonly performed by filter users and the filter


manufactures–general industry practices.
Filter user Filter manufacturer
Criteria Filter device Membrane disk Device
Bacteria retention/integrity — Q Q
test relationship data
Integrity test
water/solvent V Q, R, L Q, R, L
product V — —
Integrity test methodology V R —
and selection
Bacterial retention
water, SLB, etc. — Q, L Q, L
product V, membrane disk — —
Bacterial retention/integrity V Q Q
test methodology
Effects of chemical compat- V Q Q
ibility on filter integrity
Toxicity testing — Q Q
Extractables V Q, R, L Q, R, L
Effects of sterilization V Q Q
methods on filter integrity
Q denotes qualification testing process specific; V denotes validation testing; R
denotes recommendation for validation; and L denotes filter lot specific release
criteria.

set by pharmaceutical manufactur- both must pass the post-use in-


ers. However, if an adsorption prob- tegrity test in process operations. If
lem is experienced with a certain only one filter in the series is neces-
formulation, it should be addressed. sary to validate a sterile filtrate, then
Redundant filtration. One final cir- only one is required to pass the in-
cumstance to be addressed in vali- tegrity test.
dation is the use of redundant fil- Where redundant filters are in-
ters, two final filters of identical stalled, the validation process also
rating deployed in series as “insur- must address the impact of the sec-
ance” against nonsterile filtration. ond filter on extractables levels and
This practice is common in Europe adsorption (i.e., twice as much).
and with some American manufac-
turers of serum, tissue culture PDA Technical Report No. 26
media, and biologicals. Filter validation recommendations.
If both filters are needed to PDA’s Technical Report No. 26 sum-
achieve sterility in a given process, marizes the principles and best

s22 Pharmaceutical Technology FILTRATION 2004 www.phar mtech.com


practices of sterile filtration and its that the pharmaceutical or biophar-
validation. In this article, we have maceutical company bears complete
described details of some of the responsibility for:
more significant aspects of filter val- • filter validation
idation that follow the recommen- • use of the validated sterile filtra-
dations of “TR26.” tion system in compliance with cur-
TR26 recommends that the phar- rent good manufacturing processes
maceutical manufacturer perform a and the recommendations of the
process-specific validation which filter manufacturer as found in the
includes: validation report and supporting
• establishing an integrity test documentation.
methodology and demonstrating Pharmaceutical users of sterile fil-
integrity of the sterilizing filter tration are urged to read PDA’s
• performing bacterial retention Technical Report No. 26, to follow
studies its guidelines, and to work closely
• having a correlation between bac- with their filtration manufacturer
terial retention and the integrity test in resolving any questions or issues
method relating to sterile filtration and its
• verifying chemical compatibility validation.
• performing extractables testing
• evaluating the effects of steriliza- References
tion on filter integrity. 1. PDA Technical Report #26, “Sterilizing
Filtration of Liquids,” J.Pharm. Sci. and
Technol. 52 (1 Supp) (1998).
Summary and conclusion 2. T.H. Meltzer, ed., Filtration in the Phar-
The validation report summarizes maceutical Industry (Marcel Dekker, Inc.,
the protocols, results, and interpre- New York, NY, 1987).
tations of all testing performed. It 3. American Society for Testing and Ma-
also provides parameters and proto- terials (ASTM), Standard Test Method
for Determining Bacterial Retention of
cols for performing steam or auto- Membrane Filters Utilized for Liquid Fil-
clave sterilization and integrity test- tration, ASTM Standard F838-83,
ing. The report is an important (ASTM, Philadelphia, PA, 1983).
documentation of the validation 4. P. Stinavage, “Perspective on Filtration”
process. It should be read and un- presented at PDA, Washington DC, 3
December 1999. PT
derstood by all end users involved in
validating the specific sterile filtra-
tion process. The report also serves
as an important reference document
in an FDA inspection.
While filter manufacturers per-
form much of the qualification and
validation work in sterile filtration
validation, it is crucial for the phar-
maceutical processor to remember

Pharmaceutical Technology FILTRATION 2004 s23

Você também pode gostar