PROCESS INSTRUMENTS

®
CHANDLER ENGINEERING
Dew Point Testers
Bureau of Mines-Type Dew Point Tester from AMETEK

Features Bureau of Mines Model Operation


zx Meets ASTM D1142 The Chandler Bureau of Mines Dew The Chandler “Bureau of Mines-
Point Tester is a primary instrument type” Dew Point Tester consists
zx Portable and rugged for
used to measure the dew point of a small high pressure chamber
repeatable results
temperature of any gas. It is rugged, through which the gas sample
zx Nothing to calibrate portable, and can provide very flows. A polished mirror is at one
zx Traceable to NIST repeatable results. Because of its end of the chamber and a viewing
primary nature, the standard dew window at the other. The operator
zx Easy to maintain and use point tester requires no calibration. controls an expandable gas through
Consequently, it is frequently a valve, cooling the polished mirror
used to verify results of electronic until the dew point is observed. The
Applications instruments. temperature and pressure are then
zx Natural Gas The Chandler Dew Point Tester is a recorded and plotted on a graph
Designed for accuracy in the chilled mirror device that operates to obtain the water content of the
rugged environment typically by achieving the conditions required sample. Hydrocarbon dew points in
found in the gas industry. A by the definition of dew point. natural gas can be obtained using
special model is available for When a gas is cooled out of contact the same method.
offshore use. with liquid water, the humidity or
zx Sour Gas water content remains constant,
Corrosion-resistant materials are but the percentage of saturation
used for H2S atmospheres. increases until it reaches 100% and
moisture begins to condense. The
zx Industrial Gases
temperature at this point is known
The tester is ideally suited for the
as the dew point or frost point.
measurement of the dew point
of pressurized N2, O2, H2, Ar, air,
and other bottled gases.
zx Heat Treating Furnaces
Chandler instruments are easy
to use and produce reliable,
accurate answers. They have
been used for years in this
application.
The Chandler Dew Point Tester
is a chilled mirror device that
operates by achieving the
conditions required by the
definition of dew point.

The Chanscope II version allows lighted,


magnified viewing of the dew point with a digital
temperature display superimposed on the viewing
window in tenths of degree.

www.chandlerengineering.com
Dew Point Testers

CHANDLER ENGINEERING
Specifications Options zx Quick Chiller
Chanscope II Dew Point Tester zx Digital Readout Uses the same expandable
Accuracy: The Chanscope II version gases but directs them onto
±0.1°C (0.2°F) down to 0°C (32°F) allows lighted, magnified the back of the polished mirror.
±0.5°C (1.0°F) down to -100°C (- viewing of the dew point with Fast reaction time is achieved,
148°F) a digital temperature display but there is an increased risk of
±1.0°C (2.0°F) down to -200°C (- superimposed on the viewing overshooting the dew point until
328°F) window in tenths of degree. all components reach thermal
Dew Point Temperature Range: The unit also has a button-hold equilibrium.
-200°C (-148°F) to ambient feature to retain the measured zx Liquid Nitrogen Chiller
Operating Pressure: Up to 5000 value. The model uses accurate Uses an immersion rod and
psig (34.6 MPa) platinum RTD and LED Dewar flask containing N2 for
Connections: Sample inlet and readout to give the operator temperatures down to -129°C
outlet: 1/8” NPT; Refrigerant, 1/8” instantaneous feedback during a (-200°F).
tube test.
zx Thermometers
Power: Battery-powered with
115VAC or 230VAC charger. zx Pressure Range A wide range of thermometers
Models are available in ranges of are available in Centigrade or
Ship Weight: 25 lb (11 kg)
1,000, 3,000 or 5,000 psig (7.0, Fahrenheit , including certified
Carrying Case: 14” x 18” x 7”
20.8, and 34.6 MPa). thermometers traceable to NIST.
(35.5cm x 20cm x 17.8cm)
zx Regular Chiller Thermometers are filled with a
Uses liquid propane: -29°C biodegradable liquid, eliminating
Standard Dew Point Tester the safety concerns of mercury.
(-20°F) or liquid carbon dioxide
Accuracy: ± 0.5°F (± 0.25°C)
-62°C (-80°F) as a refrigerant.
Dew Point Temperature Range:
Its geometry optimizes thermal
Dependant on which thermometer is
uniformity for superior control
chosen.
and accurate determination.
Operating Pressure: Up to 5000
psig (34.6 MPa)
Connections: Sample inlet and
outlet: 1/8” NPT; Refrigerant, 1/8”
tube
Ship Weight: 25 lb (11 kg)
Carrying Case: 14” x 18” x 7”
(35.5cm x 20cm x 17.8cm)

SALES AND MANUFACTURING:


USA - Delaware
455 Corporate Blvd., Newark DE 19702 • Tel: +1-302-456-4400, Fax: +1-302-456-4444
USA - Oklahoma
2001 N. Indianwood Ave., Broken Arrow OK 74012 • Tel: +1-918-250-7200, Fax: +1-918-459-0165
150 Freeport Road, Pittsburgh, PA 15238 CANADA - Alberta
Ph. +1-412-828-9040, Fax +1-412-826-0399 2876 Sundridge Way N.E., Calgary, AB T1Y 7H9 • Tel: +1-403-235-8400, Fax: +1-403-248-3550
www.chandlerengineering.com
WORLDWIDE SALES AND SERVICE LOCATIONS:
USA - Texas FRANCE
Tel: +1-281-463-2820, Fax: +1-281-463-2701 Tel: 33 1 30 68 89 20, Fax: 33 1 30 68 89 29
®

CHINA GERMANY
© 2007, by AMETEK, Inc. Beijing / Tel: 86 10 8526 2111, Fax: 86 10 8526 2141 Tel: 49 21 59 91 36 0, Fax: 49 21 59 91 3639
All rights reserved. Printed in the U.S.A. Chengdu / Tel: 86 28 8675 8111, Fax: 86 28-8675 8141 MIDDLE EAST - Dubai
Model 2920 (1/07) Shanghai / Tel: 86 21 6426 8111, Fax: 86 21 6426 7818 Tel: 971 4 881 2052, Fax: 971 4 881 2053
One of a family of innovative process analyzer solutions from AMETEK Process Instruments. SINGAPORE
Specifications subject to change without notice. Tel: 65 6484 2388, Fax: 65 6481 6588