Você está na página 1de 3

Harries.

qxp 17/5/05 12:25 pm Page 23

Regulations & Legislation

ISO 11137 – Requirements for the Validation and Routine Control of


Radiation Sterilization of Healthcare Products – A Discussion of
a report by Current Revisions
John Harries

UK Expert to WG2 and WG8, International Organization for Standardization (ISO)

There are currently two standards that regulate the Part 1 – Requirements for the John Harries is the UK Expert to
WG2 and WG8 of the International
sterilisation of healthcare products using ionising Development, Validation and Routine Organization for Standardization
radiation – the European Standard (EN) 5521 published Control of a Sterilization Process for (ISO), which is developing the new
in 1994 and International Organization for Medical Devices combined European Standard (EN)
552/ISO 11137 Standard for
Standardization (ISO) 111372 published in 1995. These Control and Validation of
two standards are very similar in their requirements; There are nine sections to Part 1, which is Sterilization of Healthcare Products
using Ionising Radiation. He is also
however, ISO 11137 contains more detail with regard normative. In addition, there is an Annexe A, which Corporate Quality Compliance
to dose setting methods. contains guidance to each requirement where Manager for Isotron plc. He spent
available. The key changes are as follows. seven years with a medical device/
microbiological diagnostic
Both standards are currently being revised and will be manufacturer where, as the
combined into a new standard ISO EN 11137. This • Increased emphasis on the establishment of laboratory manager, he was
responsible for quality assurance
article will summarise the major changes that are maximum acceptable dose. The standard suggests
(QA) and research and development
likely to be incorporated in the new standard. It must that a test programme shall be undertaken to (R&D). He is an active member of
be emphasised that this article references requirements establish the effect on product of treatment with a the Association of British Healthcare
Industries (ABHI) Sterilization
that are currently in a draft form and have not been wide range of doses. Working Group and Chairman of
published. There is therefore a possibility that some of the Microbiology Group of the
the content discussed in this article may change in the • Although the requirement to reference a Panel on Gamma and Electron
irradiation. Mr Harries is also a
final form of the published standard. maximum dose exists in the current standards the member of British Standards
addition of the need to undergo a test programme Institution (BSI) Sterilization
Standards Committee CH/67.
The new combined draft standard is a three- will undoubtedly lead to increased costs in the
part document: development of new products.

• Part 1 – Requirements for the Development, • There is more detailed reference to the
Validation and Routine Control of a Sterilization application of X-rays to sterilise healthcare
Process for Medical Devices; products in this draft standard. This underlines the
increasing importance of this technology in the
• Part 2 – Establishing the Sterilization Dose for field of healthcare sterilisation.
Radiation Sterilization; and
• The transfer of dose between different radiation
• Part 3 – Guidance on Dosimetric Aspects. technologies has more emphasis and greater clarity
in the new standard. There are specific
The format and style of the standard has changed requirements laid down for the transference of
significantly. The new standard is based on ISO maximum acceptable, verification and sterilisation
14937: 2000 – Sterilization of Healthcare Products – dose between radiation sources. A distinction is
General Requirements for Characterization of a made between products containing water and
Sterilizing Agent and the Development, Validation products that do not contain water.
and Routine Control of a Sterilization Process for
Medical Devices. This ‘umbrella’ standard is • The new standard contains much more emphasis
intended to be used for the regulation of any new or on product definition. There are sections within
novel sterilisation techniques; however, it is also the requirements regarding the use of product
being used as the format for the revision of all three families for establishing the sterilisation dose and
major sterilisation technology standards, i.e. ethylene dividing into product categories for the purpose
oxide, radiation and moist heat. of routine processing. This will enable

1. BS EN 552, “Sterilization of Medical Devices – Validation and Routine Control of Sterilization by Irradiation” (1994).
2. ISO 11137, “Sterilization of Health Care Products – Requirements for Validation and Routine Control – Radiation
Sterilization” (1995). 23

BUSINESS BRIEFING: MEDICAL DEVICE MANUFACTURING & TECHNOLOGY 2005


Harries.qxp 17/5/05 12:26 pm Page 24

Regulations & Legislation

validation cost savings; however, there is set a Method 2. The verification dose is set at a sterility
criteria that must be followed to ensure assurance level (SAL) of 10-1 not 10-2 as in Method
compliance with the standard. 1 and the pass criteria is one failure in 10 sterility
tests. Both VDmax methods will only substantiate
• The section on validation in the new standard is a minimum dose to achieve a sterility assurance
now divided into three distinct stages: level of 10-6.

– installation qualification (IQ), The table of radiation dose required to achieve a


– operational qualification (OQ) and given SAL for an average bioburden utilised in
– performance qualification (PQ). Method 1 has changed significantly. The table now
starts at 0.1 average bioburden, which is a higher
• The need to perform quarterly dose audits bioburden than the current table, and the table in
required by the current standards has caused the draft now references bioburden value to one
problems for many medical device manufacturers decimal place to a bioburden of 10 and thereafter
primarily due to the availability and cost of whole numbers of bioburden. This should make
product and the volume of microbiological the table easier to use and is a more logical method
testing involved. This has been recognised by the of representing the data. This will change the
standards committee and greater flexibility has average bioburden that equates to the minimum
been introduced. sterilisation dose, e.g. in the new table the average
bioburden equivalent to a minimum dose of 25
• A completely new section, The Maintenance of kGy for an SAL 10-6 is now 1,100.
Process Effectiveness, covers bioburden determ-
ination and sterilisation dose audits. This The draft also contains a different method of dealing
provides an opportunity to reduce the frequency with dose augmentation. Dose augmentation is
of dose audits when combined with regular required when a dose audit fails. It is based on a
bioburden determination and successful periodic method that utilises extrapolation, which is linked to
dose audits. the number of audit sterility test failures.

• It is possible to reduce the microbiological This method of augmentation of the sterilisation


monitoring required to a bioburden determ- dose, established using Methods 1, Method 2A or
ination every quarter and a sterilisation dose audit Method 2B is based on a method propounded by
annually within three years provided dose audits Herring in 1999.4 It uses the information from the
are successful and bioburden determinations are failed dose audit and the principles underlying
within specification. However, for products with Method 2, together with a conservative estimate of
low bioburden there is a need to perform monthly the resistance of the most radiation-resistant
bioburden determinations. component of the microbial population contam-
inating the product. The draft standard also
Part 2 – Establishing the recognises that the cause of the dose audit failure
Sterilization Dose for Radiation often cannot be identified. It contains a statement
Sterilization that it is not possible to assess the extent to which the
SAL of previously sterilised batches has been affected.
This provides guidance on the establishment of Augmentation of the dose should begin with the
sterilisation dose and is considered informative. next batch to be sterilised and no action should be
However, if a decision is made to follow one of the taken with previously released batches.
methods it becomes normative with respect to
the methodology. Part 3 – Guidance on
Dosimetric Aspects
The main change is the addition of VDmax3 for the
substantiation of 25kGy and 15kGy as a sterilising This gives useful guidance on dosimetry used in the
dose. The specific method for the substantiation is in measurement of absorbed dose during radiation
addition to Method 1 and Method 2, which are sterilisation. Part 3 is informative.
carried forward from the current ISO 11137.
There are sections on the measurement of dose
The method for VDmax utilises only 40 product establishing the maximum acceptable and sterilising
units as opposed to 130 for a Method 1 or 840 for dose, selection and calibration of dosimetry systems,

3. Kowalski J, Tallentire, “A Substantiation of 25 kGy as a Sterilization Dose: A Rational Approach to Establishing


Verification Dose”, Radiat. Phys. Chem. (1999);54; pp. 55–64.
24 4. Herring, C, “Dose Audit Failures and Dose Augmentation”, Radiat. Phys. Chem. (1999);54; pp 77–81.

BUSINESS BRIEFING: MEDICAL DEVICE MANUFACTURING & TECHNOLOGY 2005


Harries.qxp 17/5/05 1:25 pm Page 25

ISO 11137 – A Discussion of Current Revisions

installation qualification, operational qualification distribution provided by mathematical modelling can


and performance qualification. be used to ensure that a sufficient number of
dosimeters are distributed in the expected zones for
There is also an annex A to Part 3 that details minimum and maximum doses during irradiator dose
mathematic modelling. Mathematic modelling may mapping studies.
be used to estimate doses in certain applications.
There are a number of methods for mathematical The use of mathematical modelling in industrial
modelling of radiation transport; however, most irradiators is relatively new but there is increased
modelling is performed using either the Point Kernel interest in these techniques and it is predicted that
Method or the Monte Carlo Method. it will have more prominence in the future.

The Point Kernel Method is used for calculating Conclusion


the dose distribution in gamma and X-ray
irradiators, while the Monte Carlo Method can The revision process for this standard is now ending.
be used for gamma, X-ray and electron It is anticipated the final draft ISO (F-Dis) will be
beam irradiators. issued in June or July 2005 with publication of the
final standard early in 2006. There will be a three-year
Mathematical modelling is used extensively in the transitional period for implementation of the
design of irradiators. Calculations are performed to requirements of the standard. This will allow
optimise the irradiation geometry to achieve the manufacturers, contract sterilisers and regulatory
desired throughputs and dose homogeneity. Data authorities time to ensure effective implementation.
from mathematical modelling is used to determine
the radiation performance of the irradiator when The new standard will provide clarification of many
filled with homogeneous product. issues related to sterilisation of healthcare products
using ionising radiation. However, medical device
Mathematical modelling can also be used in manufacturers should be aware that it will present
conjunction with dose mapping. For gamma some challenges to ensure compliance with the
irradiators, information on the expected dose requirements of the new combined standard. ■

GAMMA
ETO
E BEAM
LABORATORY SUPPORT

VWHULOLW\  &OR STERILE MEDICAL DEVICE AND PHARMACEUTICAL

D FULWLFDO
MANUFACTURERS ENSURING PATIENT SAFETY IS AS
IMPORTANT TODAY AS IT WAS WHEN *OSEPH ,ISTER

LVVXH
PIONEERED ANTISEPTICS IN THE S
E
ICIN
MED
S IN !CROSS THE 5+ %UROPE AND 3OUTH %AST !SIA
PTIC
IS E
ANT MANUFACTURERS RELY ON )SOTRONS TECHNICAL EXPERTISE
E E RE D
PION
* OS E
P H , IS IC AL S C IE NTIS T
WHO
*DPPD (WK\OHQH 2[LGH AND KNOWLEDGE ON A RANGE OF STERILISATION ISSUES
D ME D
TE R     
    A " RITIS H S URGEON AN DQG (OHFWURQ %HDP FROM INITIAL DESIGN TO VOLUME PRODUCTION

 SODQWV LQ  FRXQWULHV )SOTRON SUPPORT EXTENDS ACROSS THE THREE PRINCIPAL

&KRLFH DQG IOH[LELOLW\ STERILISATION TECHNOLOGIES AND OUR IN HOUSE


,ABORATORIES ARE FULLY EQUIPPED TO PROVIDE
$OO VLWHV DSSURYHG WR ,62  DQG (1 
COMPREHENSIVE TESTING CAPABILITIES REQUIRED
6XSSRUWLQJ LQKRXVH PLFURELRORJLFDO ODERUDWRULHV TO COMPLETE STERILISATION VALIDATIONS

& 86720(5 )2&86('‡48$/,7 < $6685('

)NFORMATION (EAD /FFICE 5NITED +INGDOM 4ELEPHONE     WWWISOTRONCOM

Você também pode gostar