Escolar Documentos
Profissional Documentos
Cultura Documentos
601
S
SSt
tta
aan
nnd
dda
aar
rrd
dd P
PPr
rro
ooc
cce
eed
ddu
uur
rre
ee
f
ffo
oor
rr
C
CCa
aal
lli
iib
bbr
rra
aat
tti
iio
oon
nn
o
oof
ff
P
PPr
rre
ees
sss
ssu
uur
rre
ee M
MMe
eea
aas
ssu
uur
rri
iin
nng
gg
I
IIn
nns
sst
ttr
rru
uum
mme
een
nnt
tts
ss
T
TTy
yyp
ppe
ee]
]]M
MMo
ood
dde
eel
ll :
:: H
HHy
yyd
ddr
rra
aau
uul
lli
iic
cc T
TTy
yyp
ppe
ee
!ssued on January, 200+
First revision on October, 200+
Approved by N!NT
N
I
M
T
-
C
A
L
I
B
R
A
T
I
O
N
N
E
T
W
O
R
K
L
A
B
O
R
A
T
O
R
I
E
S
NIMT-CNL-SP No. 601
Page 2 of 18
HYDRAULIC PRESSURE MEASURING INSTRUMENT
1. Calibration Description:
Parameter Performance Specifications Test Method
Cauge Pressure
Neasurement
Range : 0,1 NPa to 120 NPa
(Note 1)
Uncertainty : Lower than 0,18 kPa
in the range up to 6 NPa
(Note 2)
Lower than 1,18 kPa
in the range up to 120 NPa
Comparison with
Standard Hydraulic
Pressure Balance
2. Equipment Requirements: Note 2, Note 3, Note 5
Equipment used Specifications Requirement Calibration Equipments
1. Pressure Balance,
Nodel AAA with Piston
8 Cylinder
Nodel : xxx
S/N : 111
Range : 0,1 NPa to 6 NPa
Uncertainty (k=2) : 8,0*10
-5
*Pe
Not lower than 0,18 kPa
1 Piston 8 Cylinder
Nodel : xxx
S/N : 111
Nodel : yyy
S/N : 222
Range : 2 NPa to 120 NPa
Uncertainty (k=2): 8,0*10
-5
*Pe
Not lower than 1,18 kPa
2 Piston 8 Cylinder
Nodel : yyy
S/N : 222
Nass set
Nodel : zzz
S/N : 333
Range : 50 kg
Uncertainty (k=2) : 10*10
-6
*m
3 Nass set
Nodel : zzz
S/N : 333
2. Piston 8 Cylinder
!ndicator
(Note +)
Nodel : uuu
S/N : +++
Uncertainty (k=2, for temperature
measurement) : 0,5 C
+ Piston 8 Cylinder !ndicator
Node : uuu
S/N : +++
3. Steel Ruler
S/N : 555 Uncertainty (k=2) : 0,0+ mm 5 Steel Ruler
S/N : 555
+. Pressure Transmitting
(Note 6)
Fluid = Hydraulic oil
Density : (91+ 10) kg/m
3
Surface Tension : (0,031 0,003) N/m
5. Digital Nultimeter
Nodel : mmm
S/N : 666
Range : 0 y 100 vdc
Uncertainty : 10 ppm
Range : 0 y 100 mAdc
Uncertainty : 50 ppm
NIMT-CNL-SP No. 601
Page 3 of 18
3. Preliminary Operation:
3.1 Read the entire procedure before beginning a calibration.
3.2 visual check the Unit Under Calibration (UUC) to ensure that:
3.2.1 The line voltage required of the UUC has bee n set to 220 vAC, 50 Hz. !f the UUC line voltage
cannot be selected to be 220 vAC, the proper adapter is required.
TmPORTAAT AOTE:
- CIEAA GIOVE5 WTTHOOT TAICOm POWOER mO5T BE WORA WHTIE
HAAOITAG PT5TOA & CYITAOER OF THE 5TAAOARO PRE55ORE BAIAACE
AAO OORTAG CAITBRATTOA.
- REFEREACE IEVEI OF THE OOC 5HAII BE TOEATTFTEO.
3.3 Connect equipment as shown in figure 1 or figure 2 by using short, clean and appropriate
pressure tubes. The reference level of UUC should be the same level as the reference of the
standard 2 millimeters. !f not the different level shall be measured and taken into
consideration.
3UHVVXUH,QGLFDWRU
8QLW8QGHU&DOLEUDWLRQ
+\GUDXOLF
3UHVVXUH
7UDQVGXFHU
UHIHUHQFHOHYHO
Figure 1
NIMT-CNL-SP No. 601
Page + of 18
AK
8QLW8QGHU&DOLEUDWLRQ
UHIHUHQFHOHYHO
%RXUGRQWXEH
3UHVVXUHJDXJH
Figure 2
3.+ !f the UUC is needed to measure an electrical output, voltage or current ; and the measuring
instrument is not provided. !t can be calibrated using the laboratory digital multimeter.
However it shall be stated in the calibration certificate provided to the customer.
3.5 Care must be taken to ensure that no leakage of pressure between the standard pressure
balance and the UUC which will cause some confusion or incorrect reading.
4. Pressure Cycle Exercise:
Unless otherwise specified, the pressure cycles must be applied to exercise (preloading) the UUC
according to the accuracy class stated in the DKD R 6-1: 2002, pressure calibration guideline. The
pressure cycle must be consisted of starting at zero, going to full scale then back to zero pressure as
described in table 1 below.
NIMT-CNL-SP No. 601
Page 5 of 18
Table 1: Calibration sequences
Calibration
sequence
Neasurement
uncertainty
aimed at, in
of the meas-
urement span
Number of
meas-
urement
points
Number
of
preload-
ings
Load
change +
waiting
time
Waiting time at
upper limit of
measurement
range
Number of
measurement
series
(*) with zero
up/down
(**)
seconds
(***)
minutes
up down
A < 0,1 9 3 >30 2 2 2
B 0,1 h 0,6 9 2 >30 2 2 1
C > 0,6 5 1 >30 2 1 1
(*) Reference to the span was used to allow the sequence (necessary calibration effort) to be
selected from the table, as the accuracy specifications of the manufacturers are usually
related to the measurement span.
(**) One has in any case to wait until steady-state conditions (sufficiently stable indication) of
standard and calibration item are reached.
(***) For Bourdon tube pressure gauges, a waiting time of five minutes is to be observed. For
quasi-static calibrations (piezoelectric sensor principle), the waiting times can be reduced.
TmPORTAAT AOTE:
PRE55ORE 5ORGE5 CAA OAmAGE THE PRE55ORE 5EA5OR, 5mOOTH PRE55ORE CHAAGE5
T5 REOOTREO THROOGHOOT THE CAITBRATTOA.
5. Calibration Procedure:
TmPORTAAT AOTE:
- BEFORE CAITBRATTOA, A ZERO REAOTAG AT THE OOC WHTIE THE HYORAOITC T5
OPEAEO TO AA ATmO5PHERE AFTER EXERCT5E 5HAII BE OB5ERVEO. TF AOT, ZERO
5ET T5 AEEOEO.
- THE AOmBER OF THE PRE55ORE CAITBRATTOA POTAT5 OF THE OOC CAA BE
OETERmTAEO ACCOROTAG TO THE REOOTREmEAT TA THE OKO R J : 2DD2
CAITBRATTOA OF PRE55ORE GAOGE5 GOTOEITAE WHTCH OERCRTBEO TA TABIE J.
NIMT-CNL-SP No. 601
Page 6 of 18
- AFTER APPIYTAG THE TAPOT PRE55ORE AT EVERY mEA5OREmEAT POTAT, 3D
5ECOAO5 mTATmOm ARE AEEOEO TO WATT BEFORE TAKTAG A REAOTAG.
5.1 Cenerate a known pressure, step by step; from the minimum to maximum range which
explained in table 1 to the UUC. An accurate input pressure which being developed by
standard pressure balance can be calculated from the following equation.
@ >
U
W W
F S S
$
F J
D
I
9
L P D
J
L
P
H
3
o + o +
o + p p + p p
=
1
1
where :
P
A
m
a
J g
m
P
]
]
]
\
|
=
U
U
6
D
p
is the linear thermal expansion coefficient of the piston, K
-1
.
D
6GP
= GP
sI
- GP
rsIuIn
- GP
zrvaIn
- GP
rpaIaDIIv
- GP
nvsIrss
- GP
'n
U =
2
n
u
2
nvsIrss
u
2
Iv rpaIaDI
u
2
n zrvaI
u
2
rsIuIn
u
2
ar Ian s
u K
'
+ + + + +
Where:
P
n
= UUC reading, Pa
P
sI
= Pressure standard at the the point to be measured, Pa
GP
sI
= The correction value of pres sure standard, Pa
GP
rsIuIn
= The correction value due to resolution of the UUC reading, Pa
GP
zr vaIn
= The correction value due to the UUC zero reading, Pa
GP
rpaIaDIIv
= The correction value due to the UUC reading repeatability, Pa
GP
nvsIrss
= The correction value due to hysteresis of the UUC reading, Pa
GP
('n!
= The correction value due to the head correction
measurement system, Pa
Since pressure balance is used as a standard equipment.
Therefore:
GP
sI
= GP
rI/aI
- GP
ImpraIur
- GP
gravIv
- GP
(Ua!
2
sI
u =
2
a
2
gravIv
2
ImpraIur
2
rI/aI
u u u u + + +
Where:
GP
rI/aI
= The correction value due to the calibration certificate of the standard Pa
hydraulic pressure balance,
GP
ImpraIur
= The correction value of the piston 8 cylinder temperature measurement, Pa
GP
gravIv
= The correction value due to local gravity acceleration, Pa
GP
(Ua!
= The correction value due to an ambient air density measurement, Pa
NIMT-CNL-SP No. 601
Page 12 of 18
Note:
u
ImpraIur
= - (D-E!x p x
2
I
a
3
J
u
'n
= ('p x g! x
2
h
a
3
1
u
gravIv
=
2
g
a
3
J
g
p
]
]
\
|
u
Ua
=
2
,
,
]
]
\
|
Where:
a
I
, a
n
, a
g
, a
Ua
= Uncertainty attributed to the input quantity
Example
Uncertainty budget for the calibration of a hydraulic pressure transducer
Calibration effort for calibration sequence A
Statement of mean value ( Mw) with measurement deviation ('p), repeatability (DK) and hysteresis (n)
Calibration item
Accuracy stated by manufacturer : 0,0+ Rdg + 0,005 FS
MasurmnI rang : 2UU Dar
Resolution : 0,01 bar
Standard device
Expanded uncertainty (standard) : 0,008 rdg but not smaller than 0,000 + bar
Lnar InrmaI xpansn //nI (D
-D
p
! : U,UUU UU9 K
-J
Temperature of piston (t
piston
) : (20 0,5)C
Density of mass (p
m
) : / 800 kg/m
3
Calibration conditions
Pressure-transmitting medium : Oil
Fluid density (p
f ,20C, 1 bar
) : (91+ 10) kg/m
3
Air density (p
a ,20C, 1 bar
) : (1,199 0,015) kg/m
3
Height different (Ah) : (0 0,002) m
Cravity (g) : (9,/82 9/0 0,000 + 9) m/s
2
Ambient temperature : (20 1)C
Ambient pressure : (1 013 10) mbar
AmDnI numIv . (55 r JU! PH
NIMT-CNL-SP No. 601
Page 13 of 18
*Table 2 : Measurement data
Reading from calibration item
Pn (bar)
P standard
(bar)
M1 M2 M3 M4
0,000 0,00 0,00 0,00 0,00
20,022 19,99 19,99 19,9/ 19,98
+0,030 39,9/ 39,98 39,95 39,9/
59,9/0 59,8/ 59,90 59,86 59,88
/9,9/9 /9,88 /9,90 /9,86 /9,88
99,988 99,88 99,89 99,85 99,8/
119,99/ 119,8/ 119,89 119,85 119,8/
1+0,005 139,8/ 139,89 139,85 139,8/
160,01+ 159,88 159,91 159,8/ 159,89
180,022 1/9,8/ 1/9,91 1/9,8/ 1/9,88
200,031 199,90 199,91 199,88 199,90
*Note:
- All value are presented after normalization.
Table 3: Evaluation
Mean value
Mw
N
i
/+
Deviation
'p
Mw-p
Repeatability
DK
Max(_M3-MJ_,
_M-M2_!
Hysteresis
n
(_M2-MJ_-
_M-M3_!72
Expanded
uncertainty
U
bar bar bar bar bar
0,00 0,00 0,00 0,00 0,013
19,98 -0,0+ 0,02 0,01 0,018
39,9/ -0.06 0,02 0,02 0,020
59,88 -0,09 0,02 0,03 0,023
/9,88 -0,10 0,02 0,02 0,022
99,8/ -0,12 0,03 0,02 0,025
119,8/ -0,13 0,02 0,02 0,023
139,8/ -0,1+ 0,02 0,02 0,02+
159,89 -0,12 0,02 0,02 0,026
1/9,88 -0,1+ 0,03 0,03 0,030
199,90 -0,13 0,02 0,02 0,026
NIMT-CNL-SP No. 601
Page 1+ of 18
Uncertainties distributed
GP
sIanar, r
Width of distribution (k = 2) = ]
\
|
- 031 , 200
100
008 , 0
bar
= 1,60*10
-2
bar
Probability distribution = Normal
Uncertainty, u(x
! =
2
10 * 60 , 1
2
bar
= 8,00*10
-3
bar
Sensitivity coefficient, (
! = 1
?? Uncertainty contribution, u(v! = (
!' u(x
!
= 8,00*10
-3
bar
GP
ImpraIur
Width of distribution (2a) = 0,5 K
Probability distribution = Rectangular
Uncertainty, u(x
! =
3
5 , 0
K
= 2,89*10
-1
K
Sensitivity coefficient, (
! = -(o+|)
*
p
= -9*10
-6
K
-1
*200,031 bar
= -1,80*10
-3
bar/K
Uncertainty contribution, u(v! = (
!' u(x
!
= -1,80*10
-3
bar/K*2,89*10
-1
K
= -5,20*10
-+
bar
NIMT-CNL-SP No. 601
Page 15 of 18
GP
gravIv
Width of distribution (2a) = 0,000 +9 ms
-2
Probability distribution = Rectangular
Uncertainty, u(x
! =
3
10 * 9 , +
+
ms
-2
= 2,83*10
-+
ms
-2
Sensitivity coefficient, (
! =
g
p
=
9/ /82 , 9
031 , 200
bar/ m s
-2
= 20,+5 bar/ m s
-2
Uncertainty contribution, u(v! = (
!' u(x
!
= 20,+5 bar/ m
.
s
-2
* 2,83*10
-+
ms
-2
= 5,/9*10
-3
bar
GP
(Ua!
Width of distribution (2a) = 0,015 kg m
-3
Probability distribution = Rectangular
Uncertainty, u(x
! =
3
015 , 0
kg m
-3
= 8,66*10
-3
kg m
-3
Sensitivity coefficient, (
! =
m
p
p
=
800 /
031 , 200
bar/ kg m
-3
= -2,56*10
-2
bar/ kg m
-3
Uncertainty contribution, u(v! = (
!' u(x
!
= -2,56*10
-2
bar/ kg
.
m
-3
*8,66*10
-3
kg
.
m
-3
= -2,22*10
-+
bar
NIMT-CNL-SP No. 601
Page 16 of 18
GP
('n!
Width of distribution (2a) = 0,002 m
Probability distribution = Rectangular
Uncertainty, u(x
! =
3
002 0
m
= 1,15*10
-3
m
p
f(p,t)
= p
f(20C, 1bar)
]
]
]
\
|
]
]
\
|
+
20
2
C) ( t C) ( t
/8 0,000 1
piston room
(1+0,000 08P(bar)) (1)
= 91+
]
]
\
|
\
| +
20
2
20 20
/8 000 , 0 1 (1+0,000 08 * 200,031)
= 928,63 kg m
-3
Note : Equation (1) reference from PTB
Sensitivity coefficient, (
! = Ap * g
= (928,63-1,199) kg m
-3 *
9,/82 9/ m s
-2
= 9,0/*10
-2
bar/m
Uncertainty contribution, u(v! = (
!' u(x
!
= 9,0/*10
-2
bar/m
*
1,15*10
-3
m
= 1,0+*10
-+
bar
GP
rsIuIn
Width of distribution (2a) = 0,01 bar
Probability distribution = Rectangular
Uncertainty, u(x
! =
3 2
01 , 0
bar
= 2,89*10
-3
bar
Sensitivity coefficient, (
! = 1
Uncertainty contribution, u(v! = (
!' u(x
!
= 2,89*10
-3
bar
NIMT-CNL-SP No. 601
Page 1/ of 18
GP
zr vaIn
Uncertainty contribution, u(v! = 0,00 bar
GP
rpaIaDIIv
Width of distribution (2a) = 0,02 bar
Probability distribution = Rectangular
Uncertainty, u(x
! =
3 2
02 , 0
bar
= 5,//*10
-3
bar
Sensitivity coefficient, (
! = 1
Uncertainty contribution, u(v! = (
!' u(x
!
= 5,//*10
-3
bar
GP
nvsIrss
Width of distribution (2a) = 0,02 bar
Probability distribution = Rectangular
Uncertainty, u(x
! =
3 2
02 , 0
bar
= 5,//*10
-3
bar
Sensitivity coefficient, (
! = 1
Uncertainty contribution, u(v! = (
!' u(x
!
= 5,//*10
-3
bar
Expanded uncertainty =
2
hys
2
repeat
2
zero
2
res
2
h
2
a
2
g
2
temp
2
cer
u u u u u u u u u k + + + + + + + +
A
p
= 2
*
0,013
= 0,026 bar
NIMT-CNL-SP No. 601
Page 18 of 18
Table 4: Uncertainty budget for load step p = 200,031 bar
Ouantity
X
Estimate
x
Width of
distribution
2a
Diviso
r
Uncertainty
u(x
!
Sensitivity
coefficient
Uncertainty
contribution
u(v!
variance
u
2
bar bar
2
P sIanar
200,031
bar
1,60*10
-2
bar
2
8,00*10
-3
bar
1 8,00*10
-3
6,+0*10
-5
GP
sIanar,
ImpraIur
0,5 K \3 2,89*10
-1
K
-1,80*10
-3
bar/K
-5,20*10
-+
2,/0*10
-/
GP gravIv +,90*10
-+
m/s
2
\3
2,83*10
-+
m/s
2
20,+5
bar/ms
-2
5,/9*10
-3
3,35*10
-5
GP (U
a
! 0,015
kg/m
3
\3
8,66*10
-3
kg/m
3
-2,56*10
-2
bar/kg m
3
-2,22*10
-+
+,93*10
-8
''GP ngnI 0,002 m \3
1,15*10
-3
m
9,0/*10
-2
bar/m
1,0+*10
-+
1,08*10
-8
P n,
rsIuIn
199,90
bar
0,01 bar 2\3
2,89*10
-3
bar
1 2,89*10
-3
8,35*10
-6
GP zr
vaIn
0,00 bar 2\3 0,00 bar 1 0,00 0,00
GP
rpaIaDIIv
0,02 bar 2\3
5,//*10
-3
bar
1 5,//*10
-3
3,33*10
-5
GP
nvsIrss
0,02 bar 2\3
5,//*10
-3
bar
1 5,//*10
-3
3,33*10
-5
'P -0,13 bar u = 0,013
u
i
2
=
1,/3*10
-+
'P -0,13 bar U = k.u (k = 2) 0,026 bar
**allowing for the pressure-dependent fluid density