Detector Maintenance...................................53
Detector Maintenance Tips ...............................................................54 Maintenance Procedures ..................................................................55 Ordering Information .........................................................................59
General Supplies............................................79
Ordering Information .........................................................................80
Essential facts about Agilent supplies and accessories Vital maintenance procedures and troubleshooting tips All the Agilent part numbers you need in one easy-to-use guide
Infinite power range of the Agilent 1290 Infinity LC combines ultra high pressure up to
1200 bar and high flow rates up to 5 mL/min for maximum chromatographic performance, compatibility, flexibility and investment protection New 1290 Infinity Binary Pump combines active damping combined with lowest delay volumes facilitated by proprietary multi-layer technology in the new Agilent Jet Weaver Mixer to provide ultra-fast gradients and superior LC/UV and LC/MS performance New optical design in 1290 Infinity Diode Array Detector including Agilent Max-Light Cartridge Flow Cell offers a new level of UV-sensitivity and baseline robustness, including fast spectral acquisition at data rates up to 160 Hz Next generation flow-through design of the 1290 Infinity Autosampler achieves highest precision for both large and small injection volumes without changing sample loops New 1290 Infinity Flexible Cube combines with Autosampler to facilitate fixed-loop injection mode for ultra-fast cycle times, or automatic backflushing of needle seat for a new benchmark in lowest carryover New Agilent Quick-Change Valves in the 1290 Infinity Thermostatted Column Compartment boosts usability and paves the way for ultra high-throughput, multi-method and automated method development solutions
The power range of the Agilent 1290 Infinity LC is the widest available, combining ultra high pressures up to 1200 bar and high flow rates up to 5 mL/min for maximum chromatographic performance, compatibility, flexibility and investment protection.
High resolution chromatography 90,000 plates in 4 minutes Full compatibility with existing HPLC methods More detection capabilities from UV-Visible and ELSD through LC/MS Near-zero sample carryover for uncompromised data quality High system flexibility for automated method development
Entry-level system for world-class LC technology Easy and fast maintenance Flow rates up to 10 mL/min for a wide range of column dimensions and applications Upgrade options from isocratic to automated quaternary LC system
Greatest flexibility for multi-solvent gradients by accessing up to 4 solvents Flow rates up to 10 mL/min support narrow bore, standard and
semi-preparative applications Wide selection of autosamplers for full flexibility in injection volumes and sample containers
High pressure gradient mixing for superior gradient performance even at lower flow rates Optimized delay volumes for fast chromatography Flow rates from 0.05 to 5 mL/min provide ideal support for narrow and standard
bore applications
Related Literature
Order the Agilent HPLC Column Selection Guide from your local Agilent Representative or Agilent Authorized Distributor using publication number 5990-4435EN.
Easy assembly that saves you time and trouble (see diagram below) Adjustable shelves that let you tailor the rack to suit your needs Open design that ensures proper airflow management and power distribution
of equipment and cabling
For more information, contact your local Agilent Representative or Agilent Authorized Distributor.
Related Literature
Order the complete 1200 Series brochure from your local Agilent Representative or Agilent Authorized Distributor using publication number 5989-5200EN.
Pump Maintenance
Regular pump maintenance will help you lower your operating costs and generate precise results that make you feel confident.
You can count on Agilent isocratic, binary, quaternary, capillary, and preparative pumps for superior flow and composition stability. And by following a regular maintenance routine, you can also count on maximum uptime and a steady, accurate solvent flow for the life of the pump. On the following pages, you will find step-by-step instructions for these procedures:
Exchanging the purge valve frit Replacing the active inlet valve cartridge Cleaning the solvent filter Exchanging the outlet ball valve Changing the pump seals Changing and inspecting the sapphire pistons Seal wear-in procedure Replacing the wash seals
You may perform these procedures all at once, or as needed. Keep in mind that some parts may need replacing more often than others, depending on your application and solvent preparation procedures.
Pump Maintenance
Dirty active inlet valve cartridge Run leak test for verification and exchange the active inlet valve cartridge Leak on pump head Solvent filter is blocked Dirty frit High seal wear Run leak test for verification and exchange the outlet ball valve sieve or the complete valve Change the solvent filter Exchange the purge valve frit or the purge valve Run leak test for verification and exchange the pump seals Check plungers while changing the seals Exchange the wash seals
10
Pump Maintenance
Tools required
Maintenance Procedures
Exchanging the Purge Valve Frit
Maintenance Procedure
1. Using a 1/4 inch wrench, disconnect the pump outlet capillary at the purge valve. 2. Disconnect the waste tube. Beware of leaking solvents due to hydrostatic pressure. 3. Using a 14 mm wrench, unscrew the purge valve and remove it from the purge valve holder. 4. Remove the plastic cap with the gold seal from the purge valve. 5. Use a pair of tweezers or a toothpick to remove the frit. 6. Place a new frit into the purge valve with the slit facing the gold seal. 7. Replace the cap with the gold seal. Note: Before reinstallation, always check the gold seal. A deformed seal or a cracked cap should be replaced. 8. Insert the purge valve into the purge valve holder and orient the waste outlet nozzle downward. 9. Tighten the purge valve and reconnect outlet capillary and waste tubing.
11
Pump Maintenance
Tools required
Maintenance Procedure
1. Using a 14 mm wrench, loosen the active inlet valve and remove the valve from pump head. 2. Use tweezers to remove the valve cartridge from the actuator assembly.
Valve cartridge
3. Clean the area in the actuator assembly. Flush the cartridge area thoroughly. 4. Insert a new valve cartridge into the actuator assembly (make sure the valve cartridge is fully inserted into the actuator assembly). 5. Insert the valve into the pump head. Using the 14 mm wrench, turn the nut until it is hand tight. 6. Position the valve so that the solvent inlet tube connection points towards the front. 7. Using the 14 mm wrench, tighten the nut by turning the valve in its final position (do not overtighten the valve). 8. Reconnect the active inlet valve cable to the connector and the inlet tube to the valve. Note: After an exchange of the valve cartridge, it may take several mL of pumping with the solvent used in the current application before the flow stabilizes at a low percentage-ripple as low as when the system was working properly.
Plug
12
Pump Maintenance
Always filter solvents Never use the module without solvent inlet filter
Maintenance Procedure
1. Remove the blocked solvent filter from the bottle head assembly and place it in a beaker with concentrated nitric acid (35%) for one hour. 2. Thoroughly flush the filter with bidistilled water (remove all nitric acid, some capillary columns can be damaged by nitric acid). 3. Replace the filter.
13
Pump Maintenance
Tools required
Maintenance Procedure
Valve body
1. Using a 1/4 inch wrench, disconnect the valve capillary from the outlet ball valve. 2. With a 14 mm wrench, loosen the valve and remove it from the pump body. 3. Remove the plastic cap with the gold seal. 4. Remove the sieve with a pair of tweezers. Note: Check the gold seal. It should be exchanged when badly deformed. Place the valve in an upright position, insert the sieve into the recess and replace the gold seal with the cap. Make sure that the sieve cannot move and is away from the gold seal area.
5. Place a new sieve into the recess of the outlet ball valve and replace the cap with the gold seal.
Plastic cap
6. Reinstall the outlet ball valve and tighten the valve. 7. Reconnect the valve capillary.
Valve capillary
14
Pump Maintenance
Meticulously cut at just the right angle, making them the most durable and longest
Sapphire piston and seals
Precisely aligned in their stainless steel holder to minimize wear on the piston and seal
Agilent seals are designed to fit snugly around our pistons, and are capable of adapting to a wide range of flow rates and pressures. Agilent piston seals are:
Spring-loaded and engineered to deliver optimal performance over highly dynamic flow
and pressure ranges
Manufactured from a proprietary polymer blend, and feature a spring made from the same
high-quality stainless steel that is used in our pump's flow path The combination of our piston and seal has undergone extensive testing under temperature stress, with all common HPLC solvents, and in many instruments. More importantly, they yield consistent, reproducible results.
15
Pump Maintenance
Maintenance Procedure
All pumps except the G1361A preparative pump and the G4220A Infinity pump
1. Disassemble the pump head and use a piston to remove the two seals. 2. Clean the pump chambers and seal areas by flushing with solvent and wiping with a lint-free cloth. 3. Insert new seals. 4. Reassemble the pump head. 5. Perform the seal wear-in procedure (standard seals 5063-6589 only). Note: The wear-in procedure involves flushing the seals with isopropanol at high pressure while they are in place.
Maintenance Procedure
All pumps except the G1361A preparative pump
1. Disassemble the pump head assembly. 2. Remove any deposits particularly buffers from the piston surface using alcohol or toothpaste. Replace the pistons if scratched. 3. Rinse thoroughly in alcohol and HPLC-grade water. 4. Reassemble the pump head. 5. Check to make sure there are no fractures in the springs. Replace piston housing assembly if springs are damaged. 6. Reinsert the pistons.
16
Pump Maintenance
Tools required
Maintenance Procedure
Recommended after replacing standard seals only (5063-6589)
1. Fill the bottle with 100 mL of isopropanol, and place it in the solvent cabinet. Then, place the new pump head tubing (including bottle head assembly) into the bottle. 2. Screw the adapter to the active inlet valve. Connect the inlet tube from the bottle head directly to the adapter (quaternary pumps and binary pumps with selection valves only). 3. Connect the restriction capillary to the purge valve. Insert the other end of the capillary into a waste container. 4. Open the purge valve and purge the system for 5 minutes with isopropanol at a flow rate of 2 mL/min. 5. Close the purge valve and set the flow to a rate that is adequate to achieve a pressure of 350 bar. Pump for 15 minutes at this pressure to wear in the seals. You can monitor the pressure at your analog output signal using the handheld controller, ChemStation or any other controlling device connected to your pump. 6. Turn OFF the pump flow and slowly open the purge valve to release pressure from the system. 7. Disconnect the restriction capillary and reconnect the outlet capillary to the purge valve. Reconnect the appropriate tubing to the active inlet valve. 8. Rinse your system with the solvent that you plan to use for your next application.
17
Pump Maintenance
Tools required
18
Pump Maintenance
Ordering Information
Purge Valve
Description Purge valve assembly Purge valve assembly, 400 bar Purge valve assembly, 600 bar
Purge valve assembly, G1311-60009
Comments For 1120 RRLC pump For 1100/1200 pumps Includes PTFE frit For 1200 RRLC pump Includes PTFE frit For 1290 purge valve For 1290 purge valve For 1120 RRLC pump For 1100/1120/1200 RRLC, 1120 and 1290 pumps For 1120 RRLC pump For 1100/1200/1200 RRLC, 1120 and 1290 pumps For 1100/1200/1200 RRLC and 1120 pumps
Part No. G4280-60031 G1311-60009 G1312-60023 5068-0005 5068-0004 G4280-60033 01018-22707 0905-1192 5001-3707 5062-2485
5 position/7 port rotor seal 5 position/7 port stator Purge valve actuator PTFE frits, 5/pk Seal for purge valve Gold seal, outlet Outlet caps, 4/pk
19
Pump Maintenance
Inlet Valves
Description
Active Inlet Valve (AIV)
Active inlet valve, G1312-60025
Cartridge for active inlet valve, 400 bar For 1100/1200 pumps Cartridge for active inlet valve, 600 bar For 1200 RRLC pump Gold seal for inlet valve
Passive Inlet Valve (PIV)
For 1100/1200/1200 RRLC and 1120 pumps For 1120 Compact LC System For 1290 Infinity LC For 1200 Prep Pump
For binary SL pump, 600 bar For 1290 Infinity LC Gold seal, outlet Outlet caps, 4/pk Sieves for outlet ball valve, 10pk For 1100/1200/1200 RRLC and 1120 pumps For 1100/1200/1200 RRLC and 1120 pumps For G1312-60012 valve
20
Pump Maintenance
Part No. 9301-1420 9301-1421 9301-0656 9301-1450 5065-4421 9301-6341 9301-6342 G1311-60003 G1312-68716 G1376-60003 G1361-60022 G4220-60007 5063-6531
Solvent reservoir, amber, 1 L Solvent bottle, clear, 2 L, 2 inlets Solvent bottle, amber, 2 L Solvent bottle, clear, 2 L
Bottle Head Assembly
Bottle head assembly for screw bottle Bottle head assembly for F29/32 tapered solvent bottle Bottle head assembly with tubing and filter For capillary and nano systems (with stainless steel solvent filter) Bottle head assembly for prep system Bottle head assembly for 1290 Infinity LC System Bottle cap with 3-hole insert
21
Pump Maintenance
Solvent Filters
Description Glass filter, solvent inlet, 20 m pore size Glass filter, solvent inlet, 40 m pore size Glass filter, solvent inlet, 40 m pore size Stainless steel filter Recommended Use Analytical scale, micro scale Preparative LC For G2258A Dual Loop Autosampler For use in capillary and nano systems Part No. 5041-2168 3150-0944 3150-0944 01018-60025 Frit Adapter Frit adapter, PTFE, 3 mm, 4/pk Frit adapter, PTFE, 4 mm Frit adapter, PTFE, for 1/8 in. OD tubing Part No. 5062-8517 G1361-23204 G2258-23201
22
Pump Maintenance
Seal Wash
Description Peristaltic pump cassette with silicone tubing
Peristaltic pump, 5042-8507
Silicone tubing, 1 mm ID, 3 mm OD, 5 m Peristalitic pump with ChemSure tubing ChemSure tubing for peristaltic pump
Wash Seal and Wash Keeper
5065-9952 5042-8954
Wash seal Seal keeper Wash seal gasket, 6/pk Wash seal PE Support ring Backup ring for seal holder
Seal Wash Kit
For 1100/1200/1200 RRLC and 1120 pumps For 1100/1200/1200 RRLC and 1120 pumps For 1100/1200/1200 RRLC and 1120 pumps For 1290 Infinity Binary Pump For 1290 Infinity Binary Pump with seal wash For 1290 Infinity LC Pump
Active seal wash kit for isocratic or quaternary pumps Active seal wash kit for binary pumps
Includes 2 wash seal gaskets, 2 pump seals, peristaltic pump (includes pump cassette and motor), 2 seal keepers, 2 support ring assemblies, seal insert tool and silicone tubing Includes 4 wash seal gaskets, 4 pump seals, 2 peristaltic pumps (includes pump cassette and motor), 4 seal keepers, 4 support ring assemblies, seal insert tool and silicone tubing Includes 2 wash seal gaskets, 4 m flex tubing, 2 pump seals, 1 flow regulator, 2 seal keepers, 2 support ring assemblies, 20 mL luer lock syringe, seal insert tool and abrasive paper
G1311-68711
G1312-68711
01018-68722
23
Pump Maintenance
Vacuum Degassers
Description Tubing kit, degasser to pump, 4/pk, 30 cm pieces of tubing with screws and bushings Mounting tool for flangeless nut
Plastic tubing cutter, 8710-1930
Comments For G1322A, G1379A/B For G1322A, G1379A/B For G1322A, G1379A/B For G1322A, G1379A/B For G1322A, G1379A/B For G1322A, G1379A/B For G1322A, G1379A/B For G1322A For G1322A For G1379A For G1379A For G1379B For G1322A, G1379A/B
Part No. G1322-67300 0100-1710 8710-1930 5063-6598 5063-6599 5022-2155 5062-2483 5062-8534 9301-1337 G1379-60010 G1379-67310 5042-8922 G1322-68705
Plastic tubing cutter Tefzel ferrules and SS lock rings, 1/8 in., 10/pk PPS nuts, 1/8 in., 1/4-28 thread, 10/pk Union, 1/4-28 thread, polypropylene PTFE solvent tubing, 5 m, 1.5 mm ID, 3 mm OD Disposable syringes, 20 mL, 10/pk Syringe adapter, 1/16 in. OD, 2 in. long Dual channel micro degasser vacuum chamber Micro vacuum degasser tubing kit for G1379A
Micro vacuum degasser tubing kit for G1379B Online degasser accessory kit Includes 8 screws, 8 bushings, 4 markers, tubing, syringe and syringe adapter
24
Pump Maintenance
Maintenance Kits
Description
Start-up Kits
Comments Includes 1 outlet cap, 5 PTFE frits, 4 piston seals, 1 outlet gold seal, 2 glass solvent inlet filters, 20 m, and 1 cartridge for active inlet valve Includes PEEK coated fused silica capillaries, column and fittings to start up a Nanoflow LC System Includes 2 wash seal gaskets, 2 pump seals, peristaltic pump (includes pump cassette and motor), 2 seal keepers, 2 support ring assemblies, seal insert tool and silicone tubing Includes 4 wash seal gaskets, 4 pump seals, 2 peristaltic pumps (includes pump cassette and motor), 4 seal keepers, 4 support ring assemblies, seal insert tool and silicone tubing Includes 2 wash seal gaskets, 4 m flex tubing, 2 pump seals, 1 flow regulator, 2 seal keepers, 2 support ring assemblies, 20 mL luer lock syringe, seal insert tool and abrasive paper Includes piston seal, PTFE frits, 2 gold seals and 2 outlet caps Includes 4 piston seals, PTFE frits, 3 gold seals, 2 sieves and 3 outlet caps Includes 5063-6589 piston seals (2), 01018-22707 PTFE frits (5/pk), 5062-8562 cartridge active inlet valve, G1311-60012 outlet ball valve and 5063-6586 pistons (2) Includes 4 pump seals, 4 stainless steel sieves, 4 gold seals, 1 stainless steel frit and 4 outlet caps Includes 3150-0942 filter, 5022-2192 filter assembly, 5065-4445 peristaltic pump, 0890-1764 tubing and 4 prep pump seals Includes 2 wash seals and 1 pack of 6 wash seal gaskets For 1120 manual injector systems Includes piston seals, PTFE frits, gold seals and rotor seal For 1120 automated injector systems Includes piston seals, PTFE frits, gold seals, rotor seal, needle and needle seat
Pump start-up kit For 1100/1200/1290 RRLC pumps Nanoflow LC start-up kit
Seal Wash Kits
Active seal wash kit for isocratic or quaternary pumps Active seal wash kit for binary pumps Continuous seal wash kit For 1100/1200/1290 RRLC pumps
Preventive Maintenance Kits
PM kit for isocratic/quaternary pump PM kit for binary pump Extended PM kit For 1100/1200/1290 RRLC pumps Capillary pump PM kit PM kit for prep pump PM kit for seal wash option PM kit for manual injector PM kit for automated injector
25
Pump Maintenance
Maintenance Kits
Description
Accessory Kits
(Continued)
Comments Includes 3 wrenches, 5 PTFE frits, tubing, capillary and wrist strap Includes side cover with fixed rail, top and right cover for pump housing, plus 6 connecting capillaries. Allows automatic switching between different delay volumes to optimize the system for 2.1 mm ID or 4.6 mm ID columns. Includes purge valve and holder, hex keys 2.5, 3 mm, 2 wrenches 1/2 in. x 1/16 in., wrenches 1/4 in. x 5/16 in. and 14 mm, wrist strap, torque adaptor and stainless steel frit, 0.5 m Includes stainless steel connecting capillaries, solvent mixer, 2 L solvent bottle, bottle head assembly, filter, glass stop valve, stainless steel union, tubing and other parts Includes 8 screws, 8 bushings, 4 markers, tubing, syringe and syringe adapter Includes all parts to go from 20 L/min to 100 L/min flow rate in a capillary LC system Includes position sensing, 10 mL loop, 25 mL syringe, ring mounting bracket, start cable and SS connecting capillaries, 0.5 mm ID, 40 cm and 60 cm Part No. G1311-68705 G1312-68726
1100/1200 pump accessory kit Pump configuration kit for G1312B Needs G1158B 2 position/6 port valve Capillary pump accessory kit
G1376-68705
Accessory kit prep pump/gradient G1361A Online degasser accessory kit Extended flow range kit, 100 L/min Manual prep injection valve kit, stainless steel
26
Autosampler Maintenance
Ensuring peak performance, productivity, and peace-of-mind.
Your Agilent autosampler is designed to deliver accurate measurements, precise injection volumes, and high-quality data. And by following a regular schedule of preventive maintenance, you can ensure a lifetime of defensible results. This section will give you step-by-step instructions on how to perform crucial procedures such as:
Exchanging the needle Replacing the seat assembly Changing the rotor seal
You will also find the high-quality Agilent supplies you need to keep your autosampler running at top efficiency.
27
Autosampler Maintenance
Tools required
Maintenance Procedures
Exchanging the Needle
Warning: To avoid personal injury, keep fingers away from the needle area during autosampler operation.
Preparations
1. Select "Start" in the maintenance function "Change Needle." 2. When the needle is positioned approximately 15 mm above the needle seat, remove the front cover.
28
Autosampler Maintenance
Maintenance Procedure
1. Select "Needle Down" until the needle screw is aligned with the hole in the safety cover.
4. Insert the new needle. Align the needle in the seat, then tighten the screw firmly.
6. Use "Needle Up" to lift the needle to a position approximately 2 mm above the seat.
7. Ensure the needle is aligned with the seat. 8. On completion of this procedure, install the front cover. 9. Select "End" in the maintenance function "Change Needle."
29
Autosampler Maintenance
Tools required
Maintenance Procedure
1. Disconnect the seat capillary fitting from the injection valve (port 5).
2. Use a small flat head screwdriver to ease out the needle seat.
3. Insert the new needle seat assembly. Press the seat firmly into position.
6. Ensure the needle is aligned with the seat. If required, bend the needle slightly until the needle is aligned correctly. 7. On completion of this procedure, install the front cover. 8. "Select "End" in the maintenance function "Change Needle."
30
Autosampler Maintenance
Tools required
Maintenance Procedure
2. Loosen each fixing bolt two turns at a time. Remove the bolts from the head.
5. Install the new rotor seal and isolation seal. Ensure the metal spring inside the isolation seal faces towards the valve body.
6. Install the stator ring with the shorter of the two pins facing towards you at the 12 o'clock position. Ensure the ring sits flat on the valve body.
(Continued)
31
Autosampler Maintenance
8. Install stator head and stator face. Tighten the bolts alternating two turns at a time until the stator head is secure.
10. Slide the waste tube into the waste holder in the leak tray. 11. On completion of this procedure, install the front cover.
32
Autosampler Maintenance
Ordering Information
Injection Valves
Valve 2 position/6 port injection valve Use With G1313A, G1329A, G1367A, 1120 G1377A, G1389A Part No. Rotor Seal Material Tefzel PEEK 0101-1050 Vespel Rotor Seal Part No. 0100-1853 0100-1849 0100-2231 0100-2088 0101-1416 0100-2089 0101-1417 Stator Part No. Stator Face Part No. RheBuild Kit Part No. 0101-1257 0101-0921 Vespel 0100-1850 0100-1851
2 position/6 port micro injection valve 2 position/6 port injection valve, 600 bar 10 port, dual loop valve 2 position/6 port MBB injection valve 2 position/6 port ultra high pressure valve, 1200 bar
*Includes seal and stator face
G1367C SL, 0101-1422 PEEK G1367D SL Plus, G1329B G2258A G2260A G4226A 0101-1385 Vespel 0101-1267 PEEK 5067-4117 Vespel
33
Autosampler Maintenance
Needle seat, 600 bar, with seat capillary, G1367-87017 0.17 mm ID x 100 mm, 0.8 mm OD Needle seat, 600 bar, with seat capillary, G1367-87012 0.12 mm ID x 100 mm, 0.8 mm OD
34
Autosampler Maintenance
(Continued)
Part No. G1377-87201 Compatible with Needle Seat Micro needle seat with seat capillary, 100 m Micro needle seat with seat capillary, 75 m Micro needle seat with seat capillary, 50 m Part No. G1377-87000 G1377-87001 G1377-87002 G2258-87102 G2260-87101
G2258A G2260A G1313A, G1120A (automated sampler only) G1313A, G1329A, G1389A, G1120A (automated sampler only) G1377A G4226A
Needle assembly, dual loop autosampler Needle assembly, prep autosampler Seat adapter Finger caps, 15/pk
Twin needle seat, dual loop autosampler Needle seat, prep autosampler 0.5 mm ID, 20 L
Tool for micro seal capillary mounting Needle assembly, 1290 Infinity Autosampler
35
Autosampler Maintenance
36
Autosampler Maintenance
Autosampler Trays
Description 100 position tray for 2 mL vials 100 position tray for 2 mL vials, thermostattable 40 position tray for 2 mL vials 15 position tray for 6 mL vials External vial tray for 17 vials (disposal position) Disposal tube for external vial tray Well plate tray, 2 well plates, 10 vials (supports 50 mm plates) Vial plate for 54 x 2 mL vials, 6/pk Vial plate for 15 x 6 mL vials 100 position tray for micro vials Tray for 27 Eppendorf safe lock tubes, 0.5/1.5/2 mL 8.5 in. well plate rack, 2/pk For 16 shallow well plates, 4 deep well plates (max 48 mm height) or 6 vial racks 10 in. well plate rack, 2/pk For 20 shallow well plates (max height 16 mm), not compatible with deep well plates 8.5 in. well plate rack extension Includes 3 racks for 3 x 16 shallow well plates, 2 x 4 deep well plates (max 48 mm height) or 3 x 6 vial racks 10 in. well plate rack extension Includes 3 racks for 3 x 20 shallow well plates (max height 16 mm), not compatible with deep well plates 205H rack, two 96 deep well plates 200 rack, 13 x 100 mm tubes (9 mL), 96 207 rack, 16 x 100 mm tubes (12 mL), 75 209 rack, 12 x 32 mm tubes (12 mL), 96 94A special holding 1100 tray For G2250A For G1367A/B/C/D, G4226A For G1367A/B/C/D For G2257A For G1367A/B/C/D, G2258A, G4226A Comments For G1313A, G1329A, G1120A Part No. G1313-44510 G1329-60011 G1313-44512 G1313-44513 G1313-60004 G1313-27302 G2258-60011 G2255-68700 5022-6539 G4226-60021 5022-6538 G2255-68709 G2255-68710 G2255-68720 G2255-68730 G2250-04504 G2250-04503 G2250-04502 G2250-04501 G2250-04500
Tray, 5022-6538
37
Autosampler Maintenance
Autosampler Kits
Description PM kit for 1200 autosampler SL G1313/27A autosampler maintenance kit G1313/27A autosampler accessory kit Comments Includes PEEK rotor seal, needle seat, needle, 2 seals and 15 finger caps Includes 3 hex keys, 2 wrenches, tubing, 1 label halftray, wrist strap, 15 finger caps, 0.17 ID x 180 mm capillary, 100 screw top 2 mL vials and caps, tray for 40-2 mL vials and tray for 15-6 mL vials Includes 1 each: rotor seal, isolation seal, stator face, needle and needle seat Includes Vespel rotor seal, standard needle seat and needle Includes opaque front and side doors and front cover Includes transparent front and side doors Includes 100/pk 2 mL screw top vials and caps, 10/pk 0.5 mL 96 well plates, tools, 380 mm 0.17 mm capillary and CAN cable Includes 100/pk 2 mL screw top vials and caps, 10/pk 0.5 mL 96 well plates, tools, 40 L loop, 50 cm-50 m fused silica/PEEK capillary, 75 m seat capillary and CAN cable Part No. G1329-68719 G1313-68705 Includes 1 rotor seal (Vespel), 1 needle, 1 needle seat, 2 metering seals and 15 finger caps G1313-68709
PM kit for standard autosampler PM kit for standard autosampler Light protection kit for G1329A Door replacement kit for G1329A Well plate sampler accessory kit PM kit for well plate autosamplers Micro well plate sampler accessory kit
Includes Vespel rotor seal, needle, needle seat, peristaltic pump cartridge and seal tight nut G1367-68730 G1377-68705 G2260-68705 G1313-68711 G4280-68710 G4280-68730
Accessory kit for G2260A prep autosampler Includes stainless steel connecting capillaries, hex keys, wrenches, 100/pk 2 mL screw cap vials and caps, tray for 6 mL vials, stainless steel union and other parts Multidraw upgrade kit for G1313A/G1327A/G1329A autosamplers PM kit, manual injector PM kit, auto injector Includes 500 L capillary, 1500 L capillary and ZDV union For 1120 manual injector systems Includes piston seals, PTFE frits, gold seals and rotor seal For 1120 automated injector systems Includes piston seals, PTFE frits, gold seals, rotor seal, needle and needle seat
38
Fraction collector maintenance, including simple repairs you can perform without
a service call
Guidelines for maintaining the micro fraction collector and spotter Agilent-engineered parts, such as trays, collecting tubes, plates, and needle assemblies MALDI spotting accessories
39
Analytical and Preparative Fraction Collector Maintenance Replace the inlet/waste tubing Replace the valve-to-needle tubing Exchange the preparative needle assembly Exchange the analytical needle assembly Once per year or when you notice signs of damage or wear Once per year or when you notice signs of damage or wear When the needle shows signs of damage or blockage When the needle shows signs of damage or blockage or when using the short needle assembly with high test tubes (>45 mm) When the valve is leaking or not switching properly When the flow delay sensor no longer works When defective, leaky, blocked or contaminated
Exchange the diverter valve Exchange the internal tray Repair or exchange a funnel within the internal tray or funnel tray
Micro Fraction Collector/Spotter Maintenance Replace fraction collector capillary Exchange the capillary guiding assembly Exchange the internal tray Exchange the flap septum and waste tubing At least every six months or when worn, blocked or damaged When bent or damaged When the flow delay sensor no longer works properly At least every six months or when defective or contaminated
40
Maintenance Procedures
Tubing Replacement
There are two types of tubing available to connect the valve with the needle: preparative scale PTFE tubing (0.8 mm ID) and analytical scale tubing (0.25 mm ID). Both types should be replaced when they become contaminated, worn or damaged.
Maintenance Procedure
1. Position the fraction collector's transport unit in the "Home" position. 2. Remove all installed trays from the tray base. 3. Position the transport unit in the "Change Parts" position. 4. Turn off the instrument.
Capillary Replacement
Agilent offers three types of connecting capillaries each with different internal diameters. We recommend replacing the capillaries and flap septum every six months or when contaminated, blocked, worn, or visibly damaged.
Maintenance Procedure
1. Place the transport unit of the micro fraction collector/spotter in the "Home" position. 2. Remove all installed trays from the tray base. 3. Place the transport unit in the "Change Parts" position. 4. Turn off the instrument. 5. Unscrew the micro fraction collector/spotter capillary from the flow cell of the detector. Consult your user's manual for detailed replacement instructions.
41
Ordering Information
Fraction Collector Capillary Kits and Needles
Module G1364B G1364C Max Flow Rate 100 mL/min 1 mL/min 10 mL/min 10 mL/min 100 mL/min G1364D 4 L/min 4-30 L/min 30-100 L/min Tubing ID (mm) 0.8 mm ID 0.15 mm ID 0.25 mm ID 0.25 mm ID 0.8 mm ID 25 m ID 50 m ID 100 m ID Tubing Kit G1364-68711 G1364-68723 G1364-68712 G1364-68712 G1364-68711 G1364-87304 G1364-87305 G1364-87306 50 mm 50 mm 20 mm 20 mm Needle Length Needle Typical Use G1364-87201 Tubes (max 100 mm) G1367-87200 Tubes (max 48 mm), well plates, vials G1367-87200 G1364-87202 Funnel tray (tubes max 75 mm) G1364-87202 MALDI targets, well plates
42
Well Plates
Description 96 well plates, 0.5 mL, polypropylene 96 well plates, 0.5 mL, polypropylene 96 deep well plates, 1 mL, polypropylene Closing mats for 96 well plates, silicone, pre-slit, fits 96 well plates 5042-1385 and 5042-1386 only 96 well plates, 150 L, conical, polypropylene
Collection plate, showing 96-position closing mat, 5042-1389
Part No. 5042-1385 5042-1386 5042-6454 5042-1389 5042-8502 5042-1388 5188-5375 5065-4402
384 well plates, 90 L, polypropylene 384 well plates, max 45 L, skirted 96 deep well collection plates with glass inserts, caps, and septa, pre-assembled, 0.35 mL Glass inserts, 350 L Caps/septa for glass inserts Vial plate for 54 x 2 mL vials Tray for 27 Eppendorf safe lock tubes, 0.5/1.5/2 mL Vial plate for 15 x 6 mL vials Plate for 24 tubes with 18 mm diameter
Tray, 5022-6538
43
44
Valve Maintenance
Engineered for consistent results and optimal system performance.
Agilent's industry-leading Manual Injection Valves are designed to ensure trouble-free operation with your HPLC System. Our valves also feature patented "Make-Before-Break" architecture that allows you to switch between LOAD and INJECT positions without interrupting the flow. So you can analyze more samples in less time. In this section, you will find the latest in LC injection valve technology, along with important maintenance information that will maximize reproducibility.
45
Valve Maintenance
PEEK offers a high chemical resistance and versatility and will tolerate the entire pH range
from 0 to 14.
Tefzel is recommended for use in applications where PEEK cannot be used, such as
methylene chloride or DMSO in higher concentrations.
pH Range Vespel PEEK Tefzel 0-7 7-10 10-14
Ordering Information
Injection Valves
Valve 2 position/6 port injection valve Use With G1313A, G1329A, G1367A, 1120 G1377A, G1389A G1367C SL, G1367D SL Plus, G1329B G2258A G2260A G4226A Part No. 0101-0921 RheBuild Kit Part No. 0101-1257 Rotor Seal Material Vespel Tefzel PEEK 0101-1050 0101-1422 Vespel PEEK Rotor Seal Part No. 0100-1853 0100-1849 0100-2231 0100-2088 0101-1416 0100-2089 0101-1417 Stator Part No. Stator Face Part No. 0100-1850 0100-1851
2 position/6 port micro injection valve 2 position/6 port injection valve, 600 bar 10 port, dual loop valve 2 position/6 port MBB injection valve 2 position/6 port ultra high pressure valve, 1200 bar
*Includes seal and stator face
46
Valve Maintenance
2 position/6 port switching valve, 600 bar 2 position/6 port micro switching valve 2 position/10 port micro switching valve 6 position/7 port selection valve
47
Valve Maintenance
Continuous flow path with "Make-Before-Break" design Sample capacity Choice of stainless or PEEK flow path Easy access to fittings due to wide 30 port angles
Manual Injection Valves with Position Sensing Switches
Description 7125 Comments Part No. N/A Rotor Seal Material Vespel Tefzel PEEK 7725i, stainless steel 9725i, PEEK 3725i, stainless steel prep valve 3725i, PEEK prep valve 7010/7000/7040 Manual prep injection valve kit, stainless steel Includes position sensing, 10 mL loop, 25 mL syringe, ring mounting bracket, start cable and SS connecting capillaries, 0.5 mm ID, 40 cm and 60 cm Analytical scale Analytical scale Preparative scale Preparative scale 5063-6502 0101-1253 0101-1232 0101-1231 N/A 5065-9922 PEEK PEEK PEEK PEEK PEEK Rotor Seal Part No. 0101-0623 0101-0620 0101-1255 0101-1255 0101-1255 0101-1233 0101-1233 0101-1256 0100-1860 0100-1859 0101-1254 Stator Part No. Stator Face RheBuild Kit Part No. Part No. 0101-0624
2 position/6 port Does not include 20 L loop manual injection valve and needle port for 1120
5067-4104
Vespel
5067-4105
0100-1850
48
Valve Maintenance
Stainless steel loops are square cut, free of burrs, for a flush connection Flexible PEEK loops have a clean, straight cut for low dead volume connections
Volume 5 L ID (mm) 0.18 0.18 0.18 10 L 0.30 0.30 0.25 20 L 0.51 0.30 0.25 50 L 0.51 0.51 0.51 100 L 0.51 0.51 0.51 Material SS SS PEEK SS SS PEEK SS SS PEEK SS SS PEEK SS SS PEEK Use With 7125 and 7010 7725 9725 7125 and 7010 7725 9725 7125 and 7010 7725 9725 7125 and 7010 7725 9725 7125 and 7010 7725 9725 Part No. 1535-4860 0101-1248 0101-1241 0101-0376 0100-1923 0101-1240 0101-0377 0100-1922 0101-1239 0101-0378 0100-1924 0101-1238 0101-0379 0100-1921 0101-1242
49
Valve Maintenance
Volume 200 L
Material SS SS PEEK SS SS PEEK SS SS PEEK SS SS PEEK PEEK SS SS PEEK PEEK PEEK PEEK
Use With 7125 and 7010 7725 9725 7125 and 7010 7725 9725 7125 and 7010 7725 9725 7125 and 7010 7725 9725 3725 7125 and 7010 7725 9725 3725 3725 3725
Part No. 0101-1252 0101-1247 0101-1237 0101-1251 0101-1246 0101-1236 0101-1219 0101-1245 0101-1235 0101-1250 0101-1244 0101-1234 0101-1229 0101-1249 0101-1243 0101-1230 0101-1228 0101-1227 0101-1226
500 L
1 mL
2 mL
5 mL
10 mL 20 mL
2.0 2.0
50
Valve Maintenance
Agilent color-coded manual syringes allow you to determine syringe volume with one quick glance, so you can more efficiently perform manual dilution, extraction, and sample prep. They also give you the advantages of:
Improved scale readability with a new vertical syringe scale orientation Unique package design that allows you to store syringes right on your lab bench Accuracy within 1% of nominal volume, and with precision within 1%, measured at 80%
of total scale volume
Earth-friendly cardboard and recyclable plastic packaging Certificate of Conformance to ensure the highest quality construction and performance Lot traceable for accurate identification
LC Manual Syringes with Fitted Plungers
Volume (L) Description 5 10 Fixed Fixed Removable Replacement needle for 10 L syringe 25 50 100 250 500 Fixed Fixed Fixed Fixed Fixed 3/pk 22 gauge/2 in./LC tip 22 gauge/2 in./LC tip 22 gauge/2 in./LC tip 22 gauge/2 in./LC tip 22 gauge/2 in./LC tip Unit Needle Gauge/Length/Tip Part No. 22 gauge/2 in./LC tip 22 gauge/2 in./LC tip 22 gauge/2 in./LC tip 5190-1480 5190-1484 5190-1485 5190-1486 5190-1494 5190-1501 5190-1508 5190-1515 5190-1522
Related Literature
To select the best syringe for your application, order the Agilent Syringe Specifications and Selection Guide from your local Agilent Representative or Agilent Authorized Distributor using publication number 5990-3563EN.
51
Valve Maintenance
52
Detector Maintenance
Ensure uncompromised chromatographic performance with these tips and techniques.
Agilent wavelength detectors combine exceptional flexibility with superior instrument control, data communication, and analytical capabilities. This section will show you how to maintain your detectors high level of selectivity and sensitivity by performing the following procedures:
Choosing the most compatible flow cell for your application Disassembling, cleaning, and installing the flow cell Changing the Deuterium lamp
Of course, you will also find the precision-engineered Agilent parts you need to get the job done quickly and keep your downtime to a minimum.
53
Detector Maintenance
54
Detector Maintenance
Maintenance Procedures
Tools required
Screwdriver (POZI 1 PT3)
Maintenance Procedure
1. Unscrew the three screws of the lamp housing cover and remove the cover. 2. Locate the two screws that secure the lamp and unscrew. 3. Disconnect the lamp connector and remove the lamp. 4. Place the lamp on a clean surface. Note: Do not touch the glass bulb of the new lamp with fingers. Doing so may reduce light output.
Long Life Deuterium Lamp with RFID tag, 2140-0820
5. Insert the new lamp and reconnect the lamp connector. 6. Secure the lamp with the two screws. 7. Fit the lamp wires inside the lamp housing cover so that they are not scratched by the cover. 8. Replace the lamp housing cover and secure the three screws. Note: Perform a wavelength re-calibration after lamp warm up.
55
Detector Maintenance
Tools required
Hexagonal wrench
Maintenance Procedure
1. Unlock the cell cartridge holder by pulling the lever to the front. The lever should be in the final down position. 2. Pull the cell cartridge holder completely out towards the front. 3. Remove the cell from the cartridge holder. 4. Replace the black hoods (A) to the cell interfaces (IN/OUT) and insert plugs (B) for safe storage. Note: The hoods and plugs should always be in place in order to protect the flow cell. For long-term storage, the flow cell should be flushed and filled with isopropanol to prevent the growth of algae. 5. Slide the cell cartridge holder completely into the module. 6. Lift the two levers into the upper final position to secure the cell. 7. Remove the plugs from the CELL IN and CELL OUT and set aside. 8. Connect the inlet capillary to CELL IN (left) and the waste tubing to CELL OUT (right).
56
Detector Maintenance
1. Remove both capillaries from the flow cell. 2. Remove the flow cell from the holder. 3. Remove the cell screw with a 4 mm hex key. 4. Remove the gasket. 5. Use ethanol to clean the cell body. 6. Insert the new gasket. 7. Insert the new cell screw. 8. Carefully tighten the cell screw.
57
Detector Maintenance
58
Detector Maintenance
Ordering Information
Detector Lamps
Description Comments Part No. G1314-60101 G1314-60100
For G4212A
Long life Deuterium lamp with RFID tag For G1315D/E and G1365D/E For G1315A/B/C and G1365A/B/C For G1315A/B/C and G1365A/B/C For G1315A/B/C and G1365A/B/C For G1315A/B/C/D/E and G1365A/B/C/D/E Long life Deuterium lamp Deuterium lamp Tungsten lamp
59
Detector Maintenance
Use With G1314DE G1314A/B/C G1314D/E G1314A/B/C G1314D/E G1314A/B/C G1314A/B/C G1314D/E G1314A/B/C
Comments 10 mm, 14 L, 40 bar 10 mm, 14 L, 40 bar 6 mm, 5 L, 40 bar 6 mm, 5 L, 40 bar 2 L, 120 bar 2 L, 120 bar 1 L, 40 bar 10 mm, 14 L, 400 bar 10 mm, 14 L, 400 bar
Repair Procedure* B A A
G1314-60186 G1314-65061 G1314-60183 G1315-68713 G1314-60083 G1314-60187 G1315-68713 G1314-60087 G1314-60081 G1314-60182 G1314-65054 G1314-60082
60
Detector Maintenance
Use With G1315C/D, G1365C/D G1315A/B, G1365A/B G1315C/D, G1365C/D G1315A/B, G1365A/B G1315C/D, G1365C/D G1315A/B, G1365A/B G1315A/B, G1365A/B G4212A Infinity DAD Must be used to perform detector build-IN tests
Comments 10 mm, 13 L, 120 bar 10 mm, 13 L, 120 bar 6 mm, 5 L, 120 bar 6 mm, 5 L, 120 bar 3 mm, 2 L, 120 bar 6 mm, 1.7 L, 400 bar 3 mm, 120 bar, stainless steel 10 mm, 1.0 L Must be used to perform detector build-IN tests
Repair Procedure* A A A A
G1315-60022 G1315-68712 G1315-60025 G1315-68713 G1315-60011 G1315-60024 G1315-68713 G1315-60015 N/A G1315-60016 G1315-68712 G4212-60008 N/A G4212-60011 N/A
G1315-87319 Outlet capillary, 0.12 mm ID, 200 mm long Outlet capillary, 0.17 mm ID, 200 mm long Outlet capillary, 0.12 mm ID, 200 mm long Outlet capillary, 0.17 mm ID, 200 mm long G1315-87339 Outlet capillary, 0.12 mm ID, 200 mm long Outlet capillary, 0.17 mm ID, 200 mm long G1315-87325 Outlet capillary, 0.12 mm ID, 200 mm long
DAD heat exchanger capillary, 0.12 mm ID, 310 mm long Inlet capillary with heat exchanger, 0.12 mm ID, 290 mm long
G1315-60015
61
Detector Maintenance
Sealing kit
G1315-68715
Quartz cell body, 10 mm Cell seal assembly, 500 nL Fused silica/PEEK capillary, 100 m ID, 30 cm long Fused silica/PEEK capillary, 50 m ID, 40 cm long Fused silica/PEEK capillary, 100 m ID, 12 cm long Fused silica/PEEK capillary, 50 m ID, 12 cm long Inlet Inlet Outlet Outlet
62
Detector Maintenance
G1315-68725
Quartz cell body, 80 nL, 6 mm path length Fused silica/PEEK capillary, 50 m ID, 40 cm long Fused silica/PEEK capillary, 50 m ID, 12 cm long Fused silica/PEEK capillary, 25 m ID, 20 cm long Fused silica/PEEK capillary, 25 m ID, 60 cm long Inlet Outlet Inlet Outlet
63
Detector Maintenance
Standard "D" type flow cell kit Semi-micro flow cell kit
Micro flow cell kit Cell repair kit, semi-micro High-pressure flow cell kit
1100/1200 Series Diode Array Detector (DAD)/ Multiple Wavelength Detector (MWD)
Cell repair kit Cell repair kit, semi-micro Sealing kit for 500 nL flow cell Includes window screw kit, 4 mm hexagonal wrench and seal kit Includes window screw kit, 4 mm hexagonal wrench and seal kits Includes torque adapter, 2 cell seal assemblies, 5 litetouch front and back ferrules Includes torque adapter, 2 cell seal assemblies, 5 litetouch front and back ferrules and 5 sleeves for 360 m OD capillaries G1315-68712 G1315-68713 G1315-68715
G1315-68725
64
Detector Maintenance
65
Detector Maintenance
66
Detector Maintenance
Flow cell Cuvette kit, 8 L, 20 bar Includes tubing, SS fitting, front and back ferrule, PEEK fitting, syringe needle and syringe Cut-off filter kit: 389, 408, 450, 500, 550 nm 380, 399, 418, 470, 520 nm 280, 295, 305, 335, 345 nm Corrugated tubing, polypropylene, 6.5 mm ID, 5 m PTFE tubing, FEP, 0.7 mm ID, 5 m 1/16 in. finger-tight PEEK fittings, beige, 2/pk Column connecting capillary with fittings, 380 x 0.17 mm 1/16 in. front ferrule, stainless steel, 10/pk 1/16 in. back ferrule, stainless steel, 10/pk 1/16 in. male fittings, stainless steel, 10/pk Hex key, 4.0 mm, 10 cm long straight handle Hex key, 2.5 mm, 10 cm long straight handle Fluorescence detector calibration sample, 1 g glycogen Hex key set, 1-5 mm Open end wrench, 1/4 and 5/16 in. Glass syringe Syringe needle
5061-3327 5061-3328 5061-3329 5062-2463 5062-2462 0100-1516 G1315-87311 5180-4108 5180-4114 5061-3303 5965-0027 5965-0028 5063-6597 8710-0641 8710-0510 9301-1446 9301-0407
67
Detector Maintenance
68
LC/MS maintenance, including routine procedures Supplies, chemicals, calibrants, and tools for maintaining your system Quiet Cover II for easy movement and maintenance of your rough pumps
69
70
Ordering Information
LC/MS Supplies
Description ES nebulizer assembly Used for analytical sprayer API-ES nebulizer needle replacement kit APCI nebulizer assembly APCI nebulizer needle replacement kit Needle assembly Corona needle APCI Capillary cap (G1946A/B/C) Capillary cap (G1946D and G1956A/B) Capillary (G1946A/B/C and G1956A) Capillary (G1946D, G1956B) Capillary (G6140A, G6460A, G6530A) Spring canted coil, 0.25 in. ID, 0.53 mm 1/6 in. tee, low dead volume, stainless steel ] ] ] ] 6100 Series Single Quadropole LC/MS ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] 6200 Series TOF LC/MS ] ] ] ] ] ] 6300 Series Ion Trap LC/MS ] ] ] ] ] ] ] 6400 Series Triple Quadropole LC/MS ] ] ] ] 6500 Series Accurate-Mass Q-TOF LC/MS Part No. ] G1946-60098 ] ] ] ] ] G2427A G1946-60037 G2428A G1947-60103 G1947-20029 G1946-20056 G1946-20301 G1946-80009 59987-20040 G1960-80060 1460-2571 0100-0969
71
LC/MS Supplies
Description Syringe adapter Syringe pump
(Continued)
6100 Series Single Quadropole LC/MS 6200 Series TOF LC/MS 6300 Series Ion Trap LC/MS ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] 6400 Series Triple Quadropole LC/MS 6500 Series Accurate-Mass Q-TOF LC/MS Part No. 9301-1291 3162-0178 0100-1516 0100-2304 0890-1915 5182-9710 G1969-20302 G1946-20089 G1946-20302 G1956-20302 G1946-20087 G1946-80019 G2571-80103 G1956-80000 05971-80103 G2441-80010
1/16 in. finger-tight PEEK fittings, beige Female luer to female 10/32 adapter PEEK tubing Gas-tight syringe, PTFE luer lock High-throughput skimmer, 2 mm Skimmer 1 (G1946A/B/C) Skimmer 1 (G1946D) Skimmer 1 (G1956A/B) Skimmer 2 (G1946A/B/C and G1956A only) HED electron multiplier (G1946A/B/C/D) HED electron multiplier (G6140A, G6460A, G6530A) Detector assembly (G1956A/B) Electron multiplier replacement horn Replacement horn and dynode
72
LC/MS Chemicals
Description ES/APCI positive ion performance standard, 5 x 1 mL ampoules Electrospray LC demo sample (Sulfamix) ESI+APCI LC demo sample ES Negative Ion Performance Standard, 5 x 1 mL ampoules APCI Negative Ion Performance Standard, 5 x 1 mL ampoules ES/APCI Positive Ion Performance Standard Multiple-charge compound performance evaluation sample (horse heart myoglobin) Caffeine Standards Kit for LC/MS OQ/PV ES/APCI positive ion performance standard, 5 x 1 mL ampoules ES-TOF reference mix, 6 x 2 mL ampoules ES-TOF Biopolymer reference standard kit Flushing solvent High purity water, 4 L Methyl alcohol, 1 L Ammonium formate Formic acid, 5 mL Acetonitrile, 1 L
*Recommended item for familiarization
6500 Series Accurate-Mass Q-TOF LC/MS Part No. ] G2423A 59987-20033 G1978-85000
8500-6917 G2423A ] ] ] G1969-85001 G1969-85003 ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] ] G1969-85026 8500-2236 8500-1867 G1946-85021 G2453-85060 G2453-85050
] ] ] ] ] ]
] ] ] ] ] ]
73
74
75
Part No. 0100-2051 0100-1854 0100-1855 G1946-60180 G1969-60086 1460-2571 8660-0827 G1946-80054 5190-1401 05980-60051 5080-5400 BHT-4 RMSN-2 5183-2003 5183-2004 5183-2014
Filter element, 5 m, 5/pk Rotor seal, Tefzel, pH 0 to 14 Rotor seal, Vespel, pH 0 to 10 Inlet filter assembly SSV long drain tubing assembly Spring canted coil
Cleaning Supplies
Abrasive mesh, 4000 grit Capillary cleaning wire for dip tube Cleaning powder, dielectric capillary, Alconox Cloths, lint free Cotton swabs, 100/pk
Gas Purifiers and Generators
Nitrogen gas generator, 5183-2003
Big hydrocarbon trap, 1/4 in. fittings Big universal trap, 1/8 in. fittings, nitrogen Nitrogen gas generator Nitrogen gas generator with oxygen monitor, 120/220V Maintenance kit for LC MSD nitrogen generator
*These parts are common to all LC MSD systems
76
(Continued)
Part No. G1946-60157 G1946-20215 G1946-80049 8710-1930 8710-1622 8710-1615 8710-0510 8710-0806 8710-0004 8710-2699 5190-1443
Nebulizer 25X magnifier Plastic tubing cutter Screwdriver, Torx T15 Screwdriver, Torx T20 Open end wrench, 1/4 and 5/16 in. Wrench, 1/2 and 7/16 in. Needle nose pliers, pointed serrated jaws 3 mm wrench for nebulizer needle adjustment
Maintenance Kit
LC/MS Preventive Maintenance kit Includes vacuum pump oil, oil element, big hydrocarbon trap, 5 m filter, rotor seal and canted coil spring
*These parts are common to all LC MSD systems
77
Quiet Cover
Agilent has a solution to the frequent maintenance and inherent noise of LC/MS rough pumps. The Quiet Cover II was designed for easy movement, maintenance, and better living with rough pumps used with Agilent and other LC/MS systems.
Locking castors to move heavy pump for maintenance No tools necessary to remove sectioned cover for easy access to pump Built in Lift and Tilt lever raises end of pump to drain oil Removable drip pan with well and hand holds to collect and transport oil Sound absorbing cabinet with resistant foam insulation to reduce pump noise Pump mounted on cushioned grommets to minimize vibration 2 Integrated fans maintain temperature inside cover LEDs and audible alarm if temperature exceeds 35 C limit Maximum ambient temperature of 35 C when airflow is neither restricted nor recycled Standard one-year warranty. Installation and familiarization included with new LC/MS orders
The Quiet Cover II is compatible with these Agilent LC/MS Systems:
6100 Single Quads: G6110AA, G6120AA, G6130AA, G6140AA 6300 Traps: G2440DA, G2451AA, G4533AA, G2474SS 6410 QQQ: G6410AA 6210 TOF: G3250AA, G3252A 6510AA Q-TOF: G6510AA
Or any analytical system using BOC Edwards pumps (lbs/kg): E2M28, E2M18, E1M18.
Quiet Cover
78
General Supplies
They may be small, but they have a BIG effect on your productivity and results.
While supplies and accessories are the least expensive components of your LC system, they can contribute to problems such as injector damage, ghost peaks, and analyte degradation. That is why Agilent supplies are engineered with the same reliability and reproducibility you expect from Agilent instruments. And they are precisely designed to work with your 1100, 1120, 1200 or 1290 Series HPLC System. The following pages feature a wide selection of Agilent-engineered supplies, including:
Connection capillaries Fittings and unions LC tubing In-line and solvent filters Frits and adapters Sample filtration supplies
These supplies are also designed for fast, easy replacement, so you can get your system running again quickly.
79
General Supplies
Ordering Information
Made of flexible stainless steel (0.6 mm OD) with 1/16 in. OD tubing at both ends to accept standard fittings Pre-swaged fittings are assembled according to Swagelok specifications
80
General Supplies
(Continued)
Color Code Fittings/Connection Red Red Red Green Green Green Green Green Red Green Red Green 1 end pre-swaged 1 end pre-swaged 1 end pre-swaged 1 end pre-swaged 1 end pre-swaged 1 end pre-swaged 1 end pre-swaged Non-swaged 1 end pre-swaged Non-swaged Pre-swaged Pre-swaged Part No. G1313-87304 01090-87610 01090-87611 G1313-87305 01090-87304 01048-87302 01090-87305 G2260-87300 G1328-87600 G1316-87303 G1316-87300 G1315-87312 G1315-87311 5062-8522 SS SS SS SS SS SS SS SS SS SS SS SS SS PEEK
Inlet capillary, With fittings 600 mm, 0.17 mm ID Red Red Red Red Red Green Green Green Green Green Without fittings Without fittings Without fittings Without fittings 1 end pre-swaged Without fittings Without fittings Without fittings Without fittings With fittings
0.12 0.12 0.12 0.12 0.12 0.17 0.17 0.17 0.17 0.17 VWD Waste 0.25
SS SS SS SS SS SS SS SS SS SS PEEK
5021-1820 5021-1821 5021-1822 5021-1823 G1316-87303 5021-1816 5021-1817 5021-1818 5021-1819 G1316-87300 5062-8535
Waste Without fittings capillary, PEEK, 0.25 mm ID Blue Color Code Fittings/Connection Red Green Red Pre-swaged Pre-swaged 1 end pre-swaged
Injector valve Injector valve Description Stainless steel capillary Stainless steel capillary Stainless steel capillary
*SS = stainless steel
SS SS SS SS SS
81
General Supplies
Length (mm) Connection 400 700 150 105 150 500 310 310 100 130 340 75 500 50 70 90 130 150 170 210 Male to male Male to male Male to male Male to male Male to female Male to male Male to male Male to male Male to male Male to female Male to male Male to male Male to male Male to female Male to female Male to female Male to female Male to female Male to male Male to male
Fittings Pre-swaged Pre-swaged Pre-swaged Pre-swaged Non-swaged 1 end pre-swaged 1 end pre-swaged 1 end pre-swaged Non-swaged Non-swaged 1 end pre-swaged Non-swaged 1 end pre-swaged Non-swaged Non-swaged Non-swaged Non-swaged Non-swaged Non-swaged Non-swaged
Part No. G1312-87303 G1312-87304 G1312-87305 G1312-87306 G1315-87303 G1315-87307 G1315-87319 G1315-87339 G1316-27301* G1316-87304* G1316-87305* G1316-87306* G1316-87309 G1316-87312 G1316-87313 G1316-87314 G1316-87315 G1315-87303 G1316-87316 G1316-87317
82
General Supplies
(Continued)
Part No. G1316-87318 G1316-87319 G1316-87321 G1316-87323 G1316-87326* G1316-87327 G1316-87328* G1367-87303* G1367-87304 Part No. G1315-87319 G1315-87339 G1312-87300 G1312-67500 G1312-87301
To TCC Cooler (50-150 mm column) Cooler (20-30 mm column) Detector DAD cell 10 port valve Column TCC
ID (mm) 0.12 0.12 0.17 0.17 0.12 0.12 0.12 0.12 0.17 ID (mm) 0.17 0.12 0.17 0.17 0.17 0.17 0.17 0.17 0.17
Length (mm) Connection 300 340 105 170 75 170 210 100 250 310 310 1770 Male to male Male to male Male to male Male to male Male to female Male to female Male to male Male to male Male to male Male to male Male to male Male to male
Fittings Non-swaged Non-swaged Non-swaged Non-swaged Non-swaged Non-swaged Pre-swaged Non-swaged Pre-swaged Fittings 1 end pre-swaged 1 end pre-swaged Pre-swaged
83
General Supplies
CTC Autosampler Column Autosampler Second pump Valve Column Column Valve Valve Filter Pressure sensor Mixer Valve
84
General Supplies
PEEK Coated Fused Silica Capillaries for use with 20 L/min Flow Range
From EMPV Flow sensor Injection valve Metering device Injection valve Column Detector -switching valve To Flow sensor Injection valve Metering device Needle Column Detector Waste Column ID (m) 50 50 50 100 50 50 75 50 Length (mm) 220 550 200 110 500 400 700 280 Fitting Part No. B/B B/C B/C B/B C/D D/E E/C/D G1375-87301 G1375-87310 G1375-87302 G1375-87303 G1375-87304 G1315-68703 G1315-68708 G1375-87309
PEEK Coated Fused Silica Capillaries for use with 100 L/min Flow Range
From EMPV Flow sensor Injection valve Metering device Injection valve Column Detector -switching valve To Flow sensor Injection valve Metering device Needle Column Detector Waste Column ID (m) 100 100 100 100 75 75 75 50 Length (mm) 220 550 200 110 500 400 700 280 Fitting Part No. B/B B/C B/C B/B C/D D/E E/C/D G1375-87305 G1375-87306 G1375-87312 G1375-87303 G1375-87311 G1375-87308 G1315-68708 G1375-87309
85
General Supplies
Replacement Fittings and Ferrules for Capillary and Nano Flow System
Graphic Description 1/16 in. stainless steel fittings, front and back ferrules, 10/pk 1/16 in. stainless steel fittings, male, 4 mm, 10/pk Type A Part No. 5062-2418
5063-6593
1/32 in. ferrule and stainless steel lock ring, 10/pk 6 fittings, 2 plugs, PEEK for -valves
5065-4423
5065-4410
5065-4422
5022-6536
86
General Supplies
Part No. 5062-2418 5065-4454 5065-9967 5061-3303 5180-4108 5180-4114 5063-6591 0100-1516 5062-8541 5065-4426 5042-6500 0100-1631 0100-2175 0350-1402 01080-83202 5067-1540 5067-1547 0100-1259 5042-8957 5067-1557 5067-1558
1/16 in. stainless steel fittings, front and back ferrules, 10/pk 1/16 in. stainless steel long fittings, front and back ferrules, 10/pk 1/16 in. extra long stainless steel fittings and ferrules, 10/pk 1/16 in. male fittings, stainless steel, 10/pk 1/16 in. front ferrule, stainless steel, 10/pk 1/16 in. back ferrule, stainless steel, 10/pk 1/16 in. finger-tight PEEK fittings, beige, 10/pk 1/16 in. finger-tight PEEK fittings, beige, 2/pk 1/16 in. finger-tight PEEK long fittings, beige, 10/pk 1/16 in. finger-tight PEEK fittings, 10/pk 2 each: yellow, blue, black, green, red 1/16 in. double winged 2 piece fitting, 10/pk PEEK RheFlex 2-piece fittings, 5/pk PEEK RheFlex 2-piece fittings, colored, 10/pk ChromTrac identifiers, 20/pk 2 each: black, green, white; 4 each: yellow, blue, red 1/16 in. blanking nut, stainless steel Stainless steel hex nut fitting, PEEK ferrule, max 600 bar, 6/pk 1/16 in. PEEK ferrule, max 600 bar, 6/pk, for use with 5067-1540 fitting 1/16 in. plastic fitting (plug)
1/16 in. finger-tight polyketone fittings, max pressure 600 bar, 10/pk 0.8 mm ID stainless steel ferrules, 6/pk 0.8 mm ID stainless steel fittings, M4/4 mm, 6/pk
87
General Supplies
Unions
Description True ZDV union, no fittings
Micro T-connector, PEEK, 5042-8519
Part No. 5022-2145 5022-2184 0100-0900 0100-2441 5022-2133 0100-1847 0100-2298 5065-9971 5042-8517 5042-8518
Universal ZDV union, stainless steel, no fittings ZDV union, with fittings ZDV union, PEEK with fittings High flow union, no fittings PEEK adapter 1/4-28 to 10-32
Adapter, PEEK int. 1/4-28 to ext. 10-32 Barbed Y-Connector PP for 3/16 in. ID tube, 10/pk Adapter, female to female 1/4-28 Adapter, male luer to female 1/4-28 T-connector, PEEK, swept volume 0.57 L For 1/16 in. OD tubing
5022-2144 5042-8519
Micro T-connector, PEEK, swept vol 29 nL, with 1/32 in. ID fittings
88
General Supplies
PEEK Tubing
ID (mm) 0.50 0.25 0.25 0.18 0.18 0.13 0.13 Length (m) 1.5 1.5 5 1.5 5 1.5 5 Color Code Orange Blue Blue Yellow Yellow Red Red Part No. 0890-1761 0890-1762 5042-6463 0890-1763 5042-6462 0890-1915 5042-6461
89
General Supplies
Other Tubing
Description PTFE tubing, FEP, primary use for valve solutions PTFE solvent tubing, primary use for flow path from solvent bottle to degasser, to pump Corrugated tubing, polypropylene Silicone tubing Clamps and micro clamps, 10/pk Barbed Y-Connector PP for 3/16 in. ID tube, 10/pk Front seat tube, SS Back seat tube, SS Front seat tube, PTFE Back seat tube, PTFE Waste tube Waste tube Drawing tube assembly for flush solvent with filter and bottle cap Tubing assembly, solvent flush Waste tube Waste tubing, 5 m, 6.5 mm ID, corrugated polypropylene Waste tube, FEP For G1313/27/29A 1100/1200 Series Autosampler For G1387A 1100/1200 Series Micro Autosampler 5 6.5 0.8 1.6 For G2258A 1100/1200 Series Dual Loop Autosampler 0.1 0.12 0.1 0.12 0.15 0.1 0.5 0.5 0.2 0.25 0.8 0.8 Comments Length (m) 5 5 5 5 ID (mm) 0.7 1.5 6.5 1 3 OD (mm) 1.6 3.1 Part No. 5062-2462 5062-2483 5062-2463 5065-9978 5065-9976 5065-9971 G2258-87316 G2258-87315 G2258-87312 G2258-87313 G2258-87310 G2258-87311 G2258-87307 G2258-87314 G1313-87300 5062-2463 G1375-87326
90
General Supplies
Accessories
Description Plastic tubing cutter Blades for plastic tubing cutter, 5/pk
Plastic tubing cutter, 8710-1930
Fitting screws, SS, 10-32, 4 mm, 5/pk PEEK ferrule and stainless steel ring for 2 mm tube, 5/pk Union, PEEK for 1/8 in. OD tubing Waste adapter, 1200 Series autosamplers, gray
5061-3315
5065-9939
91
General Supplies
(Continued)
Part No. 5065-9937
Kit includes: PEEK capillary, 0.13 mm ID, 1.5 m Qty 4 Stainless steel capillary, 105 x 0.12 mm Qty 4 Stainless steel capillary, 150 x 0.12 mm Qty 2 Stainless steel capillary, 200 x 0.12 mm Qty 2 Stainless steel capillary, 280 x 0.12 mm Stainless steel capillary, 400 x 0.12 mm Tubing cutter for PEEK capillaries 1/16 in. stainless steel fittings, 10/pk 1/16 in. PEEK fittings, color, 10/pk 1/16 in. PEEK fittings, 10/pk Qty 3 ZDV fitting Rheotool Cybertool
5065-9938 Kit includes: Qty 2 Fused silica/PEEK capillary, 50 m, 55 cm Qty 1 Fused silica/PEEK capillary, 50 m, 20 cm Qty 1 Fused silica/PEEK capillary, 100 m, 110 cm Qty 2 Fused silica/PEEK capillary, 50 m, 50 cm Qty 2 Fused silica/PEEK capillary, 50 m, 40 cm Qty 4 4 mm stainless steel fitting, male 10-32 Qty 4 1/32 in. PEEK ferrule and stainless steel lock ring Qty 4 PEEK fittings for -valves Qty 4 Double winged PEEK nuts and 1/32 in. ferrules Qty 1 Cybertool Kit includes: Qty 5 Capillaries to optimize 1100 Series LC for use with RRHT columns Includes capillaries to optimize 1200 LC for low dispersion operation 5065-9947
Rapid Resolution High Throughput capillary kit 1200 capillary kit for 0.12 mm ID
G1316-68716
92
General Supplies
0.2
2.1
5067-1551
5067-1555, 10/pk
2 0.5
2.1
< 1 mL/min
4.8
1-5 mL/min
0.5 10 10
12.7 19
5064-8273 5022-2165
Recommended when high salt concentrations are used 0.3 2.0 1200 bar
5023-0271, 5/pk
93
General Supplies
Solvent Filters
Description Glass filter, solvent inlet, 20 m pore size Glass filter, solvent inlet, 40 m pore size Glass filter, solvent inlet, 40 m pore size Recommended Use Analytical scale, micro scale Preparative LC For G2258A Dual Loop Autosampler Part No. 5041-2168 3150-0944 3150-0944 Frit Adapter Part No. Frit Inlet Frit Inlet ID (mm) ID (mm) 5 7 7 3.2 4 3.2 Frit adapter, PTFE, 3 mm, 4/pk 5062-8517 Frit adapter, 4 mm, PTFE Frit adapter, PTFE for 1/8 in. OD tubing G1361-23204 G2258-23201
94
General Supplies
Sample Filtration
Syringe Filters, Membrane Econofilters, 200/pk
Description Regenerated cellulose Regenerated cellulose
Membrane Econofilters, 5185-5830
Diameter (mm) 25 25 25 25 25 25
Part No. 5190-1415 5190-1416 5190-1417 5190-1418 5190-1419 5190-1420 5190-1421 5190-1422
0.45 m PP For solvent based samples, low water breakthrough values 0.45 m regenerated cellulose For aqueous or organic solvents, very low non-specific protein binding membrane 0.20 m PTFE For aggressive samples 0.20 m Nylon For aqueous and organic samples, pH 3-10 0.20 m PP For solvent-based samples, low water breakthrough values 0.20 m regenerated cellulose For aqueous or organic solvents, very low non-specific protein binding membrane
95
General Supplies
Solvent Filters/Degassers
Description HPLC solvent filter/degasser assembly
Replacement Parts for 3150-0577
Part No. 3150-0577 5188-2743 5188-2744 5188-2745 5188-2746 3150-0576 9301-0895 3150-0509
Glass funnel, 250 mL PTFE coated sieve PTFE seal Funnel base, glass
Filter Membranes
Regenerated Cellulose filter membranes Diameter 47 mm, pore size 0.45 m, 100/pk Nylon filter membranes Diameter 47 mm, pore size 0.45 m, 100/pk PTFE filter membranes Diameter 47 mm, pore size 0.45 m, 10/pk
96
Agilent e-Seminars
The latest technology and newest developments are just a click away with free e-Seminars from Agilent
With different series and topics for troubleshooting, maintenance, informatics, compliance, and industry applications, Agilent e-Seminars are the easy way to get the latest information without the hassle or expense of travel. These free 60 to 90 minute online seminars are offered continuously:
Choose from live or recorded events to best fit your schedule Participate wherever you are from the office, home, or hotel Ask Agilent experts and industry leaders questions during live Q&A sessions
For more information and to see a complete listing of available sessions, visit www.agilent.com/chem/eseminars
Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2009 Printed in the USA November 9, 2009 5990-4957EN