Você está na página 1de 55

Cat. No.

57050-18

2200 PCX Particle Counter

INSTRUMENT MANUAL
, continued

2
57050-18

2200 PCX Particle Counter


INSTRUMENT MANUAL

© Hach Company, 2000. All rights reserved. Printed in the U.S.A. dd/dp 1/00 1 ed
Trademarks of Hach Company
AccuGrow® H2O University™ Pond In Pillow™
AccuVac® H2OU™ PourRite™
AccuVer™ Hach Logo® PrepTab™
AccuVial™ Hach One® ProNetic™
Add-A-Test™ Hach Oval® Pump Colorimeter™
AgriTrak™ Hach.com™ QuanTab®
AluVer® HachLink™ Rapid Liquid™
AmVer™ Hawkeye The Hach Guy™ RapidSilver™
APA 6000™ HexaVer® Ratio™
AquaChek™ HgEx™ RoVer®
AquaTrend® HydraVer® sension™
BariVer® ICE-PIC™
Simply AccurateSM
BODTrak™ IncuTrol®
SINGLET™
BoroTrace™ Just Add Water™
SofChek™
BoroVer® LeadTrak®
SoilSYS™
C. Moore Green™ m-ColiBlue24®
SP 510™
CA 610™ ManVer®
Specê
CalVer® MolyVer®
StablCal®
ChromaVer® Mug-O-Meter®
StannaVer®
ColorQuik® NetSketcher™
SteriChek™
CoolTrak® NitraVer®
StillVer®
CuVer® NitriVer®
SulfaVer®
CyaniVer® NTrak®
Surface Scatter®
Digesdahl® OASIS™
TanniVer®
DithiVer® On Site Analysis.
Results You Can TrustSM TenSette®
Dr. F. Fluent™
OptiQuant™ Test ‘N Tube™
Dr. H. Tueau™
OriFlow™ TestYES!SM
DR/Check™
OxyVer™ TitraStir®
EC 310™
PathoScreen™ TitraVer®
FerroMo®
PbEx® ToxTrak™
FerroVer®
PermaChem® UniVer®
FerroZine®
PhosVer® VIScreen™
FilterTrak™ 660
Pocket Colorimeter™ Voluette®
Formula 2533™
Pocket Pal™ WasteAway™
Formula 2589™
Pocket Turbidimeter™ ZincoVer®
Gelex®

Page 2
Trademarks of Hach Company 57050ttl.fm
Table of Contents
Trademarks of Hach Company ................................................................................................................ 2
Safety Precautions................................................................................................................................... 4
General Specifications ............................................................................................................................. 5

OPERATION............................................................................................................................................. 7

Section 1 General Description.............................................................................................................. 9


1.1 Interpreting the Front Panel ............................................................................................................. 10

INSTALLATION ...................................................................................................................................... 13

Section 2 Installation........................................................................................................................... 15
2.1 Plumbing the Instrument.................................................................................................................. 17
2.1.1 Plumbing Connections ............................................................................................................ 19
2.1.2 Adjusting the Flow Rate .......................................................................................................... 20
2.1.3 Electrical Connections ............................................................................................................ 22
2.1.4 Connecting Multiple Sensors .................................................................................................. 22
2.2 Using the Optional Analog I/O Kit .................................................................................................... 24
2.3 Wiring the Optional Junction Box..................................................................................................... 28

Section 3 Configuration Programming .............................................................................................. 29


3.1 Changing Operating Configuration .................................................................................................. 29
3.1.1 Changing Operating Configuration For RS485 Operation ...................................................... 29
3.1.2 Changing Operating Configuration For RS232 Operation ...................................................... 31
3.2 Setting Up Analog Inputs to the Sensor and Analog Outputs from the Sensor ............................... 31
3.2.1 Setting Up Analog Inputs ........................................................................................................ 31
3.2.2 Setting Up the Analog Outputs ............................................................................................... 32
3.2.3 Determining the Full Scale Value............................................................................................ 33
3.2.4 Understanding the 4-20 mA Output Signal ............................................................................. 34
3.2.5 Calculation at the SCADA ....................................................................................................... 35
3.3 4-20 mA Output Test........................................................................................................................ 36
SCADA Worksheet ................................................................................................................................ 37

Section 4 Serial Port Protocol and Commands ................................................................................ 39


4.1 Communication Protocol.................................................................................................................. 39
4.2 Command and Data Syntax............................................................................................................. 39

Section 5 Maintenance ........................................................................................................................ 43


5.1 Cleaning the Sensor ........................................................................................................................ 43
5.1.1 Cleaning Frequency ................................................................................................................ 43
5.1.2 Brush Cleaning ....................................................................................................................... 43
5.1.3 Stain Cleaning......................................................................................................................... 44
5.2 Long-term Storage or Shipping........................................................................................................ 45
5.3 Replacing or Removing the Sensor Flow Cell ................................................................................. 46
5.4 Tubing Replacement ........................................................................................................................ 46

GENERAL INFORMATION .................................................................................................................... 47


Certification............................................................................................................................................ 48
How to Order ......................................................................................................................................... 50
Repair Service ....................................................................................................................................... 51
Warranty ................................................................................................................................................ 52

Page 3
5705018TOC.fm Table of Contents
Safety Precautions
Please read this entire manual before unpacking, setting up,
or operating this instrument. Pay particular attention to all danger and
caution statements. Failure to do so could result in serious injury to
the operator or damage the equipment.
To ensure the protection provided by this equipment is not impaired,
do not use or install this equipment in any manner other than that
which is specified in this manual.

Use of Hazard Information


If multiple hazards exist, this manual will use the signal word
(Danger, Caution,Note) corresponding to the greatest hazard.
DANGER
Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or
moderate injury.
NOTE
Information that requires special emphasis.

Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed.

! This symbol, if noted on the instrument, references


the instruction manual for operation and/or safety information.

Laser Safety Information


This product contains a laser-based sensor that is a Class 1 product (as defined by 21 CFR, Subchapter J, of the
Health and Safety Act of 1968) when used under normal operation and maintenance. Service procedures on the
sensor can result in exposure to invisible radiation. Service should be performed only by factory–authorized
personnel. Accession number 8921784.

The particle counter has been evaluated and tested in accordance with EN 61010-1:1993, “Safety Requirements
For Electrical Equipment For Measurement, Control, and Laboratory Use” and IEC 825-1:1993, “Safety of Laser
Products”.

MODEL PCX CE W/DIS 115V NEMA –WARNING– CLASS 1


DO NOT RETURN THIS UNIT FOR REPAIR
P/N 57050-00 OR CALIBRATION IF IT CONTAINS ANIMAL
BY–PRODUCTS, TOXIC OR RADIOACTIVE
LASER PRODUCT
MATERIALS. REF: OSHA HAZARD COMMU-
MFG DT OCTOBER 1999 NICATION STD 29 DFR SECTION 1910.1200 IEC 825-1 1993

S/N 991000014
Warning Label
OPTION 2082310-2 Warning Label
VOLTS +/-15V 1A
SIZES 2u
FLOW 100 ML/MIN
PLE DANGER
SAM INVISIBLE LASER RADIATION WHEN OPEN
—AVOID DIRECT EXPOSURE TO BEAM—
THIS PRODUCT COMPLIES WITH
21 CFR, CHAPTER 1, SUBCHAPTER J Warning Label
Identification Label

Page 4
Safety Precautions 57050safety.fm
General Specifications
Sensor:

Smallest Particles Counted: 2 microns

Largest Particles Counted: 750 microns

Distance from Computer to Sensor: 4,000 feet maximum (for entire


RS 485 signal path)

Power: 115 V ac (±10%); Optional 220 V ac (±10%); 50/60 Hz

Enclosure: NEMA 4X

Indicators: Power, Counting Display, Clean Sensor, and Alarm

Flow Rate: 100 mL/Minute, Nominal

Max. Pressure:
65 psig, not more than 1 minute duration; 55 psig continuous

Operating Temperature Range: 0 – 50 °C (32 – 122 °F)

Fluid Connections:

Inlet: Quick Disconnect. Connects to ¼-inch O.D Tubing

Outlet: Quick Disconnect. Connects to ¼-inch O.D. Tubing

Accessories/Options:

Flow Control: Both active and passive/manual control devices available

Computer: IBM compatible with the following minimum specifications:


Pentium 166 MHz with 32 MB of RAM and 2 GB hard drive. SVGA color
graphics, 800 x 600 video resolution, Windows® 95, 98 or NT 4.0
or higher.

Vista Software: For monitoring filter performance, generating reports

Wiring/Cabling:

Wall Plug: 120 V ac

NEMA-Enclosed Power Supply: optional

Analog Input/Output Card:

Input: accepts input signals from external devices;

Output: provides an analog output level proportional to total number of


particles counted (raw count); includes installation manual

Junction Box: For permanent installation at test site locations so that a


sensor can be easily connected and removed without interrupting
RS485 network.

Page 5
57050specs.fm General Specifications
Visit http: //www.hach.com
OPERATION

DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the necessary
Material Safety Data Sheets and become familiar with all safety procedures before handling any
chemicals.

DANGER
La manipulation des échantillons chimiques, étalons et réactifs peut être dangereuse. Lire les
Fiches de Données de Sécurité des Produits (FDSP) et se familiariser avec toutes les
procédures de sécurité avant de manipuler tous les produits chimiques.

PELIGRO
La manipulación de muestras químicas, estándares y reactivos puede ser peligrosa. Revise las
fichas de seguridad de materiales y familiarícese con los procedimientos de seguridad antes
de manipular productos químicos.

GEFAHR
Das Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren verbunden.
Es wird dem Benutzer dieser Produkte empfohlen, sich vor der Arbeit mit sicheren
Verfahrensweisen und dem richtigen Gebrauch der Chemikalien vertraut zu machen und alle
entsprechenden Materialsicherheitsdatenblätter aufmerksam zu lesen.

PERIGO
A manipulação de amostras, padrões e reagentes químicos pode ser perigosa. Reveja a folha
dos dados de segurança do material e familiarize-se com todos os procedimentos de
segurança antes de manipular quaisquer produtos químicos.

Page 7
57050op_stop.fm OPERATION
Visit http: //www.hach.com
Section 1 General Description
The 2200 PCX Instrument is supplied with a local display of particle counts. If
you are operating the 2200 PCX Instrument in “AUTO” (stand-alone) mode or
with the Vista data collection software program, the 2200 PCX Instrument with
Local Display indicates the "normalized" count (counts/mL) based on the
selected size range, count period, and flow rate. These parameters are either
set in the 2200 PCX Instrument or set by the data collection software.

When operating the 2200 PCX Instrument with Vista data collection software,
or in manual mode, the display gives the total counts over the size range and
count period selected in the 2200 PCX Instrument. This displayed information
is useful in working areas where a quick check of “localized” counts at
selected stations is needed in detecting filter problems.

Particle removal information is collected from 1 to 32 size ranges from


2 to 250 µm. These size ranges can be easily changed (re-programmed)
based on state or federal regulations, or to mimic sizes corresponding to
Giardia or Cryptosporidium. The system also records signals from other
devices measuring turbidity, sample flow rate, pH, and more.

Expansion is available by connecting additional sensors into the computer-


based data acquisition system. See Figure 3.

Figure 1 shows a typical system using the on-line particle counting sensors
(2200 PCX Instrument) with power supply (NEMA-enclosure is optional),
a controlling computer and Vista Windows-based software.

Figure 1 On-Line Sensor (Model 2200 PCX w/Local Display)

Power Alarm
Counting Clean Sensor

2200 PCX
Particle Counter

Hach’s laser-diode-based particle counting sensors are specifically designed


for drinking water applications. Water is directed into the sensor and passes
through an optical flow cell measuring 750 x 750 microns. Each particle that
passes through the sensor generates a signal corresponding to its size.

Page 9
57050desc.fm General Description
General Description, continued

Each sensor comes with a calibration curve showing the signal response
versus size of each particle. Hach uses NIST-traceable spheres of
defined size to calibrate each sensor. Calibration information is stored in
sensor memory, and is used to separate the particle counts into the proper
size category.

1.1 Interpreting the Front Panel


The clear front panel of the NEMA 4X-rated enclosure (shown in Figure 2) is
designed for easy viewing of diagnostic LEDs as well as the local particle
count display. Refer to Table 1.

Figure 2 Instrument Front Panel

Power Alarm
Counting Clean Sensor
1 3

2 4

2200 PCX
Particle Counter
5

Table 1 Status Indicators


Item
Indicator/Feature Function
Number
1 Power LED Illuminates when power is supplied to the unit.
2 Counting LED Illuminates during the count cycle (when the actual counting occurs).
3 Alarm LED Illuminates when programmed count alarm limit has been exceeded.
4 Clean Sensor LED Illuminates if a fault in the sensor is detected that may be caused by a dirty sensor. Refer to
Section 5.1 on page 43.
5 Count Display Totalized (raw) particle count in the size range limits and for the count period set in the 2200
Window PCX Instrument Load Command Menu (see Changing Operating Configuration) and while
operating with a PCC or with Vista data collection software program.
—or—
Normalized particle count (counts/mL) in the size range limits for the count period, and flow
rate set in the 2200 PCX Instrument Load Command Menu while in the AUTO (stand-alone)
mode, or as controlled by Vista data collection software program.

Page 10
Interpreting the Front Panel 57050desc.fm
57050desc.fm

Figure 3

General Description, continued


PCX Remote Sensor
(Location “?”)
PCX On-Line Sensor PCX Remote Sensor
(Location “0” = influent) (Location “1”) Power Alarm
Counting Clean Sensor

Networked System
2200 PCX
Particle Counter
Power Alarm Power Alarm
Counting Clean Sensor Counting Clean Sensor

2200 PCX 2200 PCX


Particle Counter Particle Counter

Total RS485 distance


Up to 4000 feet
without repeater.

System Computer
Sensor Communications Cable, RS485 with
(Beldon 9841 or equal) Data Collection Software
Interpreting the Front Panel

RS232
to
RS485
Standard 25-pin
Converter
or 9-pin Serial Computer Cable
(or plug converter directly into computer).
Page 11

Water Wier Flow Controllers are shown (other automatic flow controllers available)
Visit http: //www.hach.com
INSTALLATION

Some of the following manual sections contain information in the form of warnings, cautions and
notes that require special attention. Read and follow these instructions carefully to avoid personal
injury and damage to the instrument. Only personnel qualified to do so, should conduct the
installation/maintenance tasks described in this portion of the manual.

Certains des chapitres suivants de ce mode d’emploi contiennent des informations sous la forme
d’avertissements, messages de prudence et notes qui demandent une attention particulière. Lire et
suivre ces instructions attentivement pour éviter les risques de blessures des personnes et de
détérioration de l’appareil. Les tâches d’installation et d’entretien décrites dans cette partie du
mode d’emploi doivent être seulement effectuées par le personnel qualifié pour le faire.

Algunos de los capítulos del manual que presentamos contienen información muy importante en
forma de alertas, notas y precauciones a tomar. Lea y siga cuidadosamente estas instrucciones a
fin de evitar accidentes personales y daños al instrumento. Las tareas de instalación y
mantenimiento descritas en la presente sección deberán ser efectuadas únicamente por personas
debidamente cualificadas.

Einige der folgenden Abschnitte dieses Handbuchs enthalten Informationen in Form von
Warnungen, Vorsichtsmaßnahmen oder Anmerkungen, die besonders beachtet werden müssen.
Lesen und befolgen Sie diese Instruktionen aufmerksam, um Verletzungen von Personen oder
Schäden am Gerät zu vermeiden. In diesem Abschnitt beschriebene Installations- und
Wartungsaufgaben dürfen nur von qualifiziertem Personal durchgeführt werden.

Algumas das seguintes secções do manual contêm informações em forma de advertências,


precauções e notas que requerem especial atenção. Leia e siga atentamente as presentes
instruções para evitar ferimentos pessoais e não danificar o instrumento. As tarefas de instalação/
manutenção descritas nesta parte do manual só poderão ser executadas por pessoal qualificado
para o fazer.

Page 13
57050install_stop.fm INSTALLATION
Visit http: //www.hach.com
Section 2 Installation
The Model 2200 PCX Particle Counting Sensor consists of a sensor and a
power supply. The sensor has a NEMA-4X enclosure and is designed for
indoor environments. The standard power supply must be plugged into a wall
outlet above the flood level. An optional NEMA-4X-enclosed power supply
may be hard-wired to the sensor and is necessary for applications requiring
extra power for analog outputs to external data acquisition systems. See
Figure 4.

DANGER DANGER
This instrument should be Cet appareil doit être installé par du personnel technique qualifié, afin d'assurer
installed by qualified technical le repect de toutes les normes applicables d'électricité dt de plomberie.
personnel to ensure adherence to
all applicable electrical and GEFAHR
plumbing codes. Um zu gewährleisten, daß alle elektrischen und sanitärinstallationstechnischen
VDE-Vorschriften und gegebenefalls die Zusatzvorschriften der zuständigen
Elektrizitäts- und Wasserwerke erfüllt werden, darf dieses Gerät nur von
geschultem Fachpersonal installiert werden.

PELIGRO
Este instrumento debe ser instalado por personal técnico capacitado para
asegurar elo cumplimiento con todos los códigos eléctricos y de plomería
aplicables.

PERICOLO
Questa macchina deve essere installata da un personale tecnico qualificato
affincchè ogni codice elettrico e di tubature adeguato sia rispettuto.

Figure 4 Sensor and Water Weir Orientation

AC Power Adapter Water Weir Flow Controller

Sensor Outlet Adjust height to


(Quick Disconnect Fitting)
control the flow rate.

Power Alarm
Counting Clean Sensor

The distance
2200 PCX
Particle Counter between the
sensor and
the weir
can vary.

To Drain or Waste

From Sample Source


Sensor Outlet
(Quick Disconnect Fitting)
Mounting clips can be used to mount the
weir to the sensor.

Page 15
57050install.fm Installation
Installation, continued

To ensure the sensor will prime by gravity feed, the standard Water Weir flow
controller must be mounted with the top of the 2200 PCX Instrument plumbing
connector lower than the Water Weir overflow, see Figure 4.

Sensor dimensions and additional mounting information is shown in Figure 5.

Several other flow controllers are available to address different installation


needs. Mechanical and electronic controllers for the Water Weir are optional.
Detailed plumbing procedures appear on the following pages.

Figure 5 Sensor Mounting Information

Plug-In AC Power Adapter Note: Water Weir


5.5 inches x 3.9 inches x 2.6 inches deep Flow Controller requires
additional space.

210.8 mm
8.30 inches

177.8 mm
152.4 mm
7.00 inches
6.00 inches

Power Alarm
Counting Clean Sensor
350.5 mm
13.80 inches

330.2 mm
13.00 inches 2200 PCX
Particle Counter

9.52 mm
.375 inches
4 places

Page 16
Installation 57050install.fm
Installation, continued

2.1 Plumbing the Instrument


DANGER DANGER
This instrument is not designed Cet appareil n'est pas prévu pour utilisation avec des échantillons de nature
for use with samples that are inflammable ou explosive. Pour toute solution d'échantillon autre que de l'eau
flammable or explosive in nature. utilisée avec cet appareil, tester la compatibilité échantillon/appareil pour
If any sample solution other than assurer la sécurité de l'utilisateur et le fonctionnement correct de l'appareil.
water is used in this product, test
the sample/product compatibility GEFAHR
to assure user safety and proper Dieses Instrument darf nicht in Verbindung mit Proben benutzt werden, die
product performance. entflammbar oder explosiv sind. Wenn irgendeine andere Lösung als Wasser in
diesem Gerät analysiert werden soll, muß die Proben/Gerät-Kompatibilität
getestet werden, um die Sicherheit des Benutzers und korrektes Arbeiten des
Gerätes zu gewährleisten.

PELIGRO
Este instrumento no está diseñado para usarse con muestras de naturaleza
inflamable o explosiva. Si se empleara en este producto alguna solución de
muestra que no fuera a base de agua, ponga a prueba la compatibilidad de la
muestra/producto, para cerciorarse de la seguridad y del correcto
funcionamiento del producto.

PERICOLO
Questa macchina non è concepita per essere utilizzata con campioni di natura
infiammabile o esplosiva. Se una soluzione campione oltre l’acqua è utilizzata in
questo prodotto, si deve testare la compatibilità campione / prodotto per
garantire la sicurezza dell’utente e le proprie performanze del prodotto.

Plumbing connections involve connecting the particle sensor and a flow


controller to the water system. See Figure 6.

The most important part of connecting the plumbing is tapping the best
sampling location in the water system. A good sampling location ensures a
meaningful measurement. Listed below are guidelines for tapping the best
location. Refer also to Figure 7.

• Minimize the distance from the tap to the sensor.

• Tap in a location that can supply 300 – 500 mL/min. flow with four feet of
head above the elevation of the sensor.

• Tap upstream of a pressure reduction valve or a point where the plumbing


becomes larger. These locations tend to create bubbles that can be
erroneously counted as particles.

• Tap 18 inches downstream of a bend in the plumbing. Bends create


turbulence, which mixes the particles—resulting in uniform particle-size
distribution. The 18" distance from the bend restores an even flow,
ensuring complete mixing while reducing particle settling.

• Tap a location with a high flow rate to keep heavy particles suspended.

• Tap the pipe as shown in Figure 7 to avoid air or sediment contamination


of the sample.

Page 17
57050install.fm Plumbing the Instrument
Installation, continued

Figure 6 Plumbing Connections

Maximum Head Loss


plus 4 feet

AC Power
Adapter

Power Alarm
Counting Clean Sensor

Filter
2200 PCX
Particle Counter

Sample Inlet tubing Plumbing Tap


should not extend with Shut Off
beyond 10 feet.

Drain

Page 18
Plumbing the Instrument 57050install.fm
Installation, continued

Figure 7 Sampling Options

Air (Typical)

Sediment (Typical)

Poor Good Best

Poor

2.1.1 Plumbing Connections

Use the following procedure and Figure 8 to plumb the water source to the
Water Weir Controller and 2200 PCX Instrument.

1. Install a separate plumbing tap, including a shut-off valve, for the


2200 PCX Instrument.

2. Install a quick-disconnect fitting on the plumbing tap. A white compression


fitting for ¼" tubing is provided. Use the appropriate adaptors to install
this fitting.

3. Install one end of the ¼" black semi-rigid tubing in the quick-disconnect
fitting as shown below. A 10-foot length of this tubing is provided with the
Water Weir.

Note: Do not lengthen the tubing. Tubing lengths longer than 10 feet will cause the
larger particles to "drop out" of the sample, distorting the accuracy of the particle
size reading.

Figure 8 Installing a Quick Disconnect Fitting on the Tubing

4. Route the black semi-rigid tubing (already installed in Water Weir) to the
water source; connect tubing to the source.

5. Install an 18" long piece of ¼" flexible tubing (provided with Water Weir) in
the Water Weir outlet. Attach to fitting as shown in Figure 8.

6. Attach quick-disconnect fitting to other end of tubing in step 5 then fasten


tubing to sensor inlet port. See Figure 11.

Page 19
57050install.fm Plumbing the Instrument
Installation, continued

7. Attach quick-disconnect fitting to another 12 inch long piece of ¼-inch


flexible tubing then insert into particle sensor outlet port.

8. Place other end of tubing in step 7 into Water Weir return, see Figure 9.

9. Install drain line on Water Weir. Slip clear ½-inch I.D. hose over barbed
fitting on Water Weir drain and route to appropriate drain.

10. Open shut-off valve on plumbing tap and check for leaks.

Figure 9 Water Weir Plumbing Connections

Flow Control Device


(Water Weir)

609.6 mm
2 feet

Inlet from sample source

¼″ Black Tubing
Outlet to sensor Check for kinks in tubing as it enters the fitting.

2.1.2 Adjusting the Flow Rate

Adjust the flow rate of the Water Weir for 100 mL/min. Flow rate is adjusted
by changing the head. Refer to the instructions below and Figure 10 and
Figure 11.

1. Move the white cap up to lower the head/flow; move it down to increase
the head/flow. For the Model 2200 PCX Instrument, 33" from the overflow
to the white cap is about 100-mL/min flow. Moving the cap one vertical
inch changes the flow rate about one or two milliliters per minute.

2. Measure the flow for one full minute into a 200 mL graduated cylinder.
Record the result as mL/min.

3. Adjust flow as in step 1 if necessary and remeasure.

Note: The more accurately you set the flow, the more accurate the resulting count
concentration data.

Page 20
Plumbing the Instrument 57050install.fm
Installation, continued

Figure 10 Adjusting the Water Weir Flow Rate

Sensor Outlet Flow Control Device


(Water Weir)
Sensor

To adjust the flow rate, slide this


cap up or down.
25.4 mm (1 inch) = 1-2 mL/min

To measure the flow rate, rotate


the cap as shown and use a
graduated cylinder and a stop
2200 PCX
watch. Measure for one minute.
Particle Counter

832.2 mm (33 inches) from the


overflow to the white cap
is about 100-mL/minute.

To Drain or Waste

4. Allow the water to flow through the system for 24 hours before recording
measurements. New plastic tubing can “shed” particles for the first
several hours of use. These particles could affect the accuracy of the
particle count readings.

If the optional NEMA-rated power supply was ordered with the on-line sensor,
all primary AC power can be physically isolated from areas that contain
flowing water. The power supply is housed in a NEMA enclosure, which
should be wired by certified/licensed electricians in accordance with all
applicable laws and codes.

Figure 11 Positioning the Sensor Outlet Tube in the Flow Adjustment Cap

Cut the end of this


tube straight across.

Point the tubing straight down.


Extend the tube no more than ¼″ inside the cap.

Page 21
57050install.fm Plumbing the Instrument
Installation, continued

2.1.3 Electrical Connections

Proper fusing and power interruption means must be provided on site. If AC


power is wired in conduit, provide a local disconnect next to the instrument.
Some programming functions require the power to be cycled.

Mount the AC power supply enclosure above flood stage areas. The sensor is
shipped with the cable wired between the AC power supply and the sensor.
An optional NEMA-rated power supply is available. See Figure 12 for
connection information.

The 2200 PCX sensor features serial RS-485 data communication.


An RS485 serial network provides asynchronous communications between
multiple sensors and a controlling computer. The total distance from the
RS232 to RS485 converter to the farthest sensor may be up to 4000 feet
without an amplifier/repeater.

The use of a junction box allows a 2200 PCX Instrument to be brought to the
site and plugged into the J-box for sampling, then disconnected and moved to
another site. See Figure 17.

2.1.4 Connecting Multiple Sensors

A typical connection of multiple sensors is described below. The example


presented describes the connection of two sensors. Refer to Figure 13 and
the instructions below.

1. Route RS485 cable between the RS232 to RS485 converter and PCX #1
(do not connect wiring at this time). Use RS485-type shielded,
low-capacitance twisted-pair cable (Belden PN 9841 or equivalent).

2. Assuming power supplies have been wired to the 2200 PCX Instruments
(described earlier), verify that line power to power supply and line power
to RS-232 to RS-485 converter have been disconnected.

3. Dress cable wires at both ends of the cable; at the RS232 to


RS485 converter end, connect the white wire to the terminal location
marked "A", attach the blue wire to the terminal marked "B", and attach
the shield to "S".

4. At PCX #1, run cable through a liquid-tight bulkhead fitting then connect
blue wire to terminal lug marked "485B". Connect white wire to terminal
lug marked "485A". Connect shield to "SGND". If using optional J-box
(see description above), wire into the line as described in Section 2.3 on
page 28.

5. To connect another 2200 PCX Instrument to the system, dress cable


wires at both ends of the cable then route cable through a second PCX #1
bulkhead fitting and to a bulkhead fitting of PCX #2.

6. Connect blue wire to terminal lug marked "485B" at both ends of the
cable; connect white wire to terminal lug marked "485A", and shield to
"SGND" at both ends of the cable.

7. Repeat step 4 through step 6 above for each remaining 2200


PCX Instrument.

Page 22
Plumbing the Instrument 57050install.fm
Installation, continued

8. For the last 2200 PCX Instrument in the string, set jumper JP1 (located in
lower left corner of the box) so that two pins are shorted together, see
Figure 13. Close and secure the cover and reconnect power disconnected
in step 2 above.

Note: Leave jumper JP1 open (not terminated) in all instruments except the last one.

Figure 12 Wiring Optional Power Supply to the On-line Sensor

Pre-wired NEMA Power Supply


I/O MCA I/O G
ANALOG
4-20mA OUT SERIAL 4-20mA OUT IN
MCA BANK 1
POWER BANK 2
I/O 4-20Ma IN
ALARM OUT

J6 J8 J2 J7 J3 J4 J5

J1
RS232

10 (GRY)

11 (BLK)

12 (VIO)
2 (WHT)

3 (GRN)

9 (RED)
1 (BLK)

8 (YEL)
INVISIBLE LASER RADIATION WHEN OPEN
—AVOID DIRECT EXPOSURE TO BEAM—

DANGER

VOID

7
ANTY
WARRWARRAN NTY VOID
ROKE
IF BIF BROKVO
ENID
WARRANTY
IF BROKEN

AC (L)

AC (N)

GND

+15V

+5V

A GND

D GND

-15V
+15V I_PWR OUT4 AIN2

AGND OUT0 OUT5 AIN3

-15V OUT1 OUT6 AIN4

+5V OUT2 OUT7 AIN5

DGND OUT3 IN0 AIN6

SGND I_RET RET0 AIN7

JP1
485A ALM1 IN1 D
AINGND

485B ALM2 RET1 SPARE

Wire Clamp/
Strain Relief

Cable:
To facility power
Power Supply to
Particle Counter
(6 feet long, supplied
by Hach Company)

Page 23
57050install.fm Plumbing the Instrument
Installation, continued

Figure 13 Connecting RS485 Wires For Multiple Sensor Applications

PCX #1 PCX #2 (last)

I/O MCA I/O G


ANALOG I/O MCA I/O G
ANALOG
4-20mA OUT SERIAL 4-20mA OUT IN 4-20mA OUT SERIAL 4-20mA OUT IN
MCA BANK 1 MCA BANK 1
POWER BANK 2 POWER BANK 2
I/O 4-20Ma IN I/O 4-20Ma IN
ALARM OUT ALARM OUT

J6 J8 J2 J7 J3 J4 J5 J6 J8 J2 J7 J3 J4 J5

J1 J1
RS232 RS232

INVISIBLE LASER RADIATION WHEN OPEN

INVISIBLE LASER RADIATION WHEN OPEN


—AVOID DIRECT EXPOSURE TO BEAM—

—AVOID DIRECT EXPOSURE TO BEAM—


DANGER

DANGER
VOID VOID
ANTY ANTY
WARRWARRANNTY VOID WARRWARRANNTY VOID
IF BR OKE IF BR OKE
IF BROKVO
ENID IF BROKVO
ENID
WARR TYAN WARR TYAN
IF BROKEN IF BROKEN

+15V I_PWR OUT4 AIN2 +15V I_PWR OUT4 AIN2

AGND OUT0 OUT5 AIN3 AGND OUT0 OUT5 AIN3

-15V OUT1 OUT6 AIN4 -15V OUT1 OUT6 AIN4

+5V OUT2 OUT7 AIN5 +5V OUT2 OUT7 AIN5

DGND OUT3 IN0 AIN6 DGND OUT3 IN0 AIN6

SGND I_RET RET0 AIN7 SGND I_RET RET0 AIN7


JP1 485A ALM1 IN1 AINGND
D JP1 485A ALM1 IN1 AINGND
D

485B ALM2 RET1 SPARE 485B ALM2 RET1 SPARE

JP1 is shown
JP1 is shown
TERMINATED
NOT-TERMINATED
in this unit.
in this unit.

RS485-type shielded cable


(Beldon pn 9841 or exact equivalent)

RS485 to RS232 Standard 25-pin Computer Cable


Converter (or plug directly into computer)

JP1 will be terminated in only the last unit on the network.

2.2 Using the Optional Analog I/O Kit


When the optional analog I/O kit is installed along with Vista software,
0 to +5V [or 0 to +10V]) from other measurement devices such as
turbidimeters and flow meters. Connect and analyze analog signals of any
combination of 4 to 20 mA and 0 to +5 V (or 0 to +10V) inputs simultaneously.
See Figure 15 and Figure 16.

Page 24
Using the Optional Analog I/O Kit 57050install.fm
Installation, continued

Use the following operating configuration to enable the two desired analog
inputs:

If inputs are voltage:

• Select +5V full scale by removing jumper JP2 through JP7 corresponding
to the input being used (AIN2 through AIN7).

• Select +10V full scale by installing jumper JP2 through JP7,


corresponding to the input being used (AIN2 through AIN7).

• Configure voltage inputs to accept 4-20mA inputs by connecting a


250 ohm, 1% (or better) shunt resistor in parallel with the analog signal
cable and setting the appropriate jumper for 5 V operation. See Figure 14.

Figure 14 Voltage Inputs

The voltage inputs can be configured to accept 4-20 mA inputs by


connecting a 250Ω, 1% (or better) shunt resistor in parallel with the analog
signal cable and setting the appropriate jumper (JP2 to JP7) for 5V operation. AIN2

AIN3

AIN4

AIN5

AIN6

AIN7

AINGND
D

SPARE

If inputs are current (4-20 mA):

• Use IN0 (RET0 is ground) and IN1 (RET1 is ground) on 2200 PCX
Instrument Interconnect Card. The incoming data is sent along with
particle count data, via serial communications, to the computer where it is
displayed and recorded (when on-line software is being used). Connect
analog inputs from external devices to the 2200 PCX Instrument
interconnect card as shown in step 16.

Each sensor can also provide up to eight 4 to 20 mA analog outputs of


particle count data if the optional I/O kit is installed. The 4 to 20 mA output
levels are proportional to the total number of particles counted during the
sample period. This output data is also communicated over the network to the
computer for display and recording (when on-line software is being used).

Each size category will have a unique analog output signal and will interface
to an individual analog input on corresponding data acquisition system input
terminals. Analog outputs are connected to the 2200 PCX Instrument
interconnect card similar to the example shown below.

Page 25
57050install.fm Using the Optional Analog I/O Kit
Installation, continued

Figure 15 Connecting Analog Input Signals to 2200 PCX Instrument Sensor

I/O MCA I/O G


ANALOG
4-20mA OUT SERIAL 4-20mA OUT IN
MCA BANK 1
POWER BANK 2
I/O 4-20Ma IN
ALARM OUT

J6 J8 J2 J7 J3 J4 J5

J1
RS232
JP2
JP3

INVISIBLE LASER RADIATION WHEN OPEN


—AVOID DIRECT EXPOSURE TO BEAM—

DANGER
VOID
ANTY
WARRWARRAN NTY VOID
ROKE
IF BIF
JP4
JP7
JP6
JP5

BRTY ENID
OKVO
WARRAN
IF BROKEN

+15V I_PWR OUT4 AIN2

AGND OUT0 OUT5 AIN3

-15V OUT1 OUT6 AIN4

+5V OUT2 OUT7 AIN5

DGND OUT3 IN0 AIN6

SGND I_RET RET0 AIN7

JP1
485A ALM1 IN1 D
AINGND

485B ALM2 RET1 SPARE

4-20 mA from turbidimeter, flow meter, etc. Sample Input

0 to +5V or 0 to +10V,
attach braided shield to AINGND
(4-20 mA with appropriate shunt resistor)

Page 26
Using the Optional Analog I/O Kit 57050install.fm
Installation, continued

Figure 16 Connecting Analog Output Signals from PCS Sensor

I/O MCA I/O ANALOG


G
4-20mA OUT SERIAL 4-20mA OUT IN
MCA BANK 1
POWER BANK 2
I/O 4-20Ma IN
ALARM OUT

J6 J8 J2 J7 J3 J4 J5

J1
RS232

JP2
JP3

INVISIBLE LASER RADIATION WHEN OPEN


—AVOID DIRECT EXPOSURE TO BEAM—

DANGER
VOID
ANTY
WARRWARRAN NTY VOID
ROKE
IF BIF

JP4
JP7
JP6
JP5
BRTY ENID
OKVO
WARRAN
IF BROKEN

+15V I_PWR OUT4 AIN2

AGND OUT0 OUT5 AIN3

-15V OUT1 OUT6 AIN4

+5V OUT2 OUT7 AIN5

DGND OUT3 IN0 AIN6

SGND I_RET RET0 AIN7

JP1
485A ALM1 IN1 AINGND

485B ALM2 RET1 SPARE

OUT0
OUT1
OUT2
OUT3

Analog In Analog In Analog In Analog In

Device Device Device Device


#1 #2 #3 #4

Typical Connections to Data Acquisition Systems (ie. SCADA, PLC)

From external +15 to +24V Power Supply


(150mA rating, if required)

Page 27
57050install.fm Using the Optional Analog I/O Kit
Installation, continued

2.3 Wiring the Optional Junction Box


Junction boxes are available for prewiring RS485 network sensor locations.
Junction boxes allow sensors to be moved between multiple locations, without
disturbing the network integrity. Connect wires from the RS485 cables
(Belden 9841 or exact equivalent) as shown in Figure 17.

Figure 17 Junction Box Wiring Information

From the last particle counter To the next particle counter


or from the last junction box. or to the next junction box.
(Beldon Cable 9841) (Beldon Cable 9841)

Wire Color Connection


Use the supplied pigtail to connect the Junction Box
White 485A to a 2200 PCX Particle Counter.
Blue 485B

Green SGND *
* The green drain wire must go through
a 100 ohm, ½ W resistor BEFORE To the particle counter
the SGND connection.

Page 28
Wiring the Optional Junction Box 57050install.fm
Section 3 Configuration Programming

3.1 Changing Operating Configuration


Most applications for the Model 2200 PCX Instrument are configured at the
factory prior to shipment. If custom configuration is required, configure the
particle sensor for either RS485 or RS232 operation using the information
below. A worksheet is provided for your convenience. See page 37.

3.1.1 Changing Operating Configuration For RS485 Operation

1. Using a computer terminal emulation software (Windows® Terminal,


Procomm, etc.) configure the computer for: COM 1 port or COM 2 port,
9600, N, 8, 1. Start the terminal emulation program before applying power
to the sensor.

2. Verify sensor power is off, then connect plumbing and cabling, including
serial communications to the 2200 PCX Instrument.

Note: The word "load" in the next 3. Turn on power only to the sensor to be configured (only one PCX on 485
step is case sensitive; all letters in bus at a time; turn all other PCX instruments off or disconnect them from
the word must be either lower or the RS-85 network).
upper case characters.
4. To change the operating configuration, send the ASCII command "load
<Enter>" to the PCX using Terminal. You will have 30 seconds after
initialization to access the change menu for configuration of the PCX
(initialization takes about 5 seconds after the power is turned on). The
opening command menu (with analog I/O card installed) has the
following format:

---MAIN MENU---RMCA [2082375-1E]


1 - UNIT ID32 [0...32*]
2 - COUNT PERIOD00:10 [MIN,SEC]
3 - CAL LIMITS0800–1200 [LOWER, UPPER mV]
4 - COUNT MODEMANUAL [AUTO, MANUAL]
5 - PANEL DISPLAY2.0 [CUMULATIVE 9U0]
6 - FLOW RATE 100 mL/min]
7 - CALIBRATE
8 - DEFAULT MEMORY
9 - SETUP ANALOG I/OQ = QUIT

5. To select a specific operating parameter from the main menu, type the
number (1–9) then follow the prompts at the Command line. See Table 2.

* The default ID number is 32. This number is set at the factory to a lower number prior to shipment; 32 typically means the
EPROM was changed (the unit was re-optioned).

Page 29
57050config.fm Configuration Programming
Configuration Programming, continued

Table 2 Selecting Operating Parameters


Menu
Command Line Message Notes
Number
1 ENTER ID "00" corresponds to LOC 01, etc., in the Vista Software
2 ENTER MIN:SEC Sets count period when in "AUTO" mode
3 ENTER LOWER THRESHOLD (mV) Calibration check threshold
ENTER UPPER THRESHOLD (mV) Calibration check threshold
4 ENTER COUNT MODE (A OR M) Auto = locally self-timed; Manual = remote control (Vista software)
5 ENTER LOWER SIZE Defines particle size range (for front panel numerical display units
only). Do not set lower size to less than 2 microns.
ENTER UPPER SIZE (0 FOR Defines particle size range (for front panel numerical display
CUMULATIVE) units only).
6 ENTER FLOW RATE If 0 is entered, display will show particles/count period (raw count)
If >0 is entered, display will show particles/mL (normalized count)
7 CALIBRATING DISPLAY Continuous display of calibration data. Used for 4–20mA analog output
calibration. Toggles from 4–20mA when space bar is pressed.
8 DEFAULT MEMORY Do not enter a message on this line unless instructed by the
manufacturer.
9 SETUP ANALOG I/O Use only when Analog I/O card is installed.

The following is an example when "9" (SETUP ANALOG I/O) is pressed:

---I/O CARD SETUP MENU ---

1 -ANALOG INPUTS [AN0,AN1,AN2,AN3,AN4,AN5,AN6,AN7]


[OFF,OFF,OFF,OFF,OFF,OFF,OFF,OFF]

2 -ANALOG OUTPUTS
CH LOWER UPPER FULL SCALE
0 --- --- --------
1 --- --- --------
2 --- --- --------
3 --- --- --------
4 --- --- --------
5 --- --- --------
6 --- --- --------
7 --- --- --------

R - RETURN TO MAIN MENU


Command [1,2,R]:

Note: For lower size, enter 0 for 6. After selecting "1" for analog inputs or "2" for analog outputs, enter
cumulative count (no upper limit). the command number but do not press <Enter> key. If "2" is selected, add
Upper size defines the count value the following prompt lines, one line at a time, to the above menu:
represented by 20 mA.
ENTER ANALOG OUTPUT CHANNEL [0..7]:
ENTER LOWER SIZE:
ENTER UPPER SIZE (0=cumulative):
ENTER FULL SCALE COUNT (0=disable channel):

Page 30
Changing Operating Configuration 57050config.fm
Configuration Programming, continued

When finished, end the sequence by typing “Q <Enter>”. All of the


configuration information will be saved in the 2200 PCX Instrument
memory.

7. Repeat step 6 above for additional channels.

8. Repeat steps 1 through 6 above for additional sensors.

3.1.2 Changing Operating Configuration For RS232 Operation

1. Turn power off to 2200 PCX Instrument to be configured.

2. Open the instrument cover then disconnect RS485 "COM" line connector
(485A, 485B, SGND) from 2200 PCX Instrument Interconnect Card.

3. Connect programming cable between connector J1-RS232 on the


2200 PCX Instrument Interconnect Card and serial port 1 or 2 of your
computer.

4. Using your computer's terminal emulation software (Windows


"Terminal", "Procomm", etc.) configure the computer for: COM 1 or 2
port, 9600, N, 8, 1.

5. Apply power to the 2200 PCX Instrument, then repeat steps 4 through 7
of Section 3.1.1, above.

3.2 Setting Up Analog Inputs to the Sensor and Analog Outputs


from the Sensor

3.2.1 Setting Up Analog Inputs

The 2200 PCX sensor equipped with an analog I/O card will permit
connection of up to eight (8) analog inputs from other devices. Two of the
inputs, IN0/RET0 and IN1/RET1, are configured for 4-20 mA inputs. Analog
inputs AIN2 through AIN7 are configured as either 0–5 V or 0–10 V inputs.
AIN 2–7 use a common ground connection, AINGND. Select 0–5 V or 0–10 V
operation by properly setting jumpers JP2 through JP7 on the analog board.
See Figure 14 on page 25 for proper jumper settings. Configure jumpers for
4-20 mA inputs by setting the inputs for 0–5 V operation then installing a 250
ohm, 1% (or better) shunt resistor between AIN 2, for example, and AINGND.

Note: If this option does not appear, When the analog input connections are made, the analog inputs can be set up
either the analog card is not for use. See Section 3.1 on page 29 for specific setup instructions. Also see
installed or is not functioning Figure 15 on page 26. Enter the programming sequence using the “Load”
correctly. If the analog card is command as described. When the configuration program is accessed, the
installed and the Setup Analog Main Menu will appear. If an Analog I/O card is installed, a menu option titled
I/O menu does not appear,
“Setup Analog I/O” will appear.
contact Hach.

Page 31
57050config.fm Setting Up Analog Inputs to the Sensor and Analog Outputs from the Sensor
Configuration Programming, continued

Select the SETUP ANALOG I/O option on the menu. The analog set up menu
similar to the following will appear:

---I/O Card Setup Menu---


1 –Analog Inputs [AN0,AN1,AN2,AN3,AN4,AN5,AN6,AN7]
[ON, ON, ON, ON, ON, ON, ON, ON]2 –
Analog Outputs

CH Lower Upper Full Scale


0 – – – – – – – – – – – –
1 – – – – – – – – – – – –
2 – – – – – – – – – – – –
3 – – – – – – – – – – – –
4 – – – – – – – – – – – –
5 – – – – – – – – – – – –
6 – – – – – – – – – – – –
7 – – – – – – – – – – – –

Note: Some operation problems In most cases, the analog inputs will be set to ON when shipped from the
have been reported if one or more factory. If the instrument reverts to default condition, the default is for the
unused analog inputs are set to analog inputs is ON. If the analog connection to be used is indicated as ON,
OFF when others are in use. Set all no further action is needed. If it is not set to ON, set it to ON before continuing.
analog inputs to ON, even if they are
not in use.

3.2.2 Setting Up the Analog Outputs

Up to eight analog output signals of particle count data can be configured for
2200 PCX sensors equipped with the analog I/O card. Connections for analog
outputs are labeled OUT0 through OUT7. Grounds for all analog outputs are
made to I-RET. All of these outputs are configured as 4-20 mA outputs. See
Figure 16 on page 27.

The analog particle count outputs are independent of the digital outputs via
the RS485 connection to the Vista software package.

The first step in setting up the analog outputs is to determine the proper count
period. For raw water or filter influent samples, it is recommended that a count
period between 6 and 15 seconds be set. For filter effluent samples, a count
period between 30 seconds and one minute be set. Set the count period in
the Main Menu, see Section 3.1 on page 29.

Note: If the digital RS485 signal is Once the count period has been determined and set, the output channels
connected to the Vista software must be set up. From 1 – 8 of the channels can be configured and used.
package, the count period set in Normally only 1 – 4 of these channels is used. Since each active channel
Vista will also be the count period requires a shielded, twisted-pair cable, wiring for eight channels and the
used for analog outputs. Connection connections to a PLC or other device can become quite expensive. However,
to Vista automatically overrides the
careful decisions can result in a large amount of data being available with only
count period set in the main menu
three or four connections.
during configuration programming.

Page 32
Setting Up Analog Inputs to the Sensor and Analog Outputs from the Sensor 57050config.fm
Configuration Programming, continued

The analog outputs can be configured for cumulative or differential count


formats in any combination. One output can be configured for cumulative and
another for differential if desired. Generally, setting all outputs as cumulative
counts will provide the greatest flexibility. This will be discussed in greater
detail below.

Use the blank table on page 37 as a work sheet when determining how the
analog output channels are to be set. Follow the instructions in Section 3.1.1,
step 6, to set the lower and upper size for a channel. Some calculations must
be performed to determine the proper setting for the Full Scale value, see
Section 3.2.3, below.

3.2.3 Determining the Full Scale Value


Full Scale (FS) = Counts/mL x mL sample
mL Sample = 100 mL/min* x count period (in minutes)
or = 100 mL/60 seconds x count period (in seconds)

Table 3 Sample Size Reference Table (assuming flow of 100 mL/min.)


Count Period (seconds) Sample Size (mL) Count Period (seconds) Sample Size (mL)
6 10 24 40
12 20 30 50
15 25 48 80
18 30 60 100

Counts/mL must be less than or equal to 17,000 counts/mL, this is the


concentration limit of the 2200 PCX sensor. The value used for counts/mL
should be as small as is practical for the sample to be measured.
Select the maximum counts/mL by considering the resolution of the
analog output signal.

Determine the full scale value by estimating the maximum cumulative particle
counts at the sensitivity of the instrument, >2 µm. Once that is determined,
calculate the estimated full scale value for other channels by using the
appropriate divisor from the tables below:

Table 4 Divisors for determining full scale values


Divide FS value for >2 µm by: Divide FS value for >2 µm by:
>3 µm 3.4 >9 µm 90
>4 µm 8 >10 µm 125
>5 µm 15.6 >11 µm 166
>6 µm 27 >12 µm 216
>7 µm 43 >14 µm 343
>8 µm 64 >15 µm 422

To estimate divisors for values not shown in the table use the
following formula:

Divisor = (Size/2)3.

For example, the divisor for >18 µm is (18/2) 3 = 729

* The recommended flow rate for the 2200 PCX is 100 mL/min ±5%. This flow rate should be used whenever possible.

Page 33
57050config.fm Setting Up Analog Inputs to the Sensor and Analog Outputs from the Sensor
Configuration Programming, continued

3.2.4 Understanding the 4-20 mA Output Signal

The 4-20 mA output signal is divided into 4,096 steps where a signal of 4 mA
will equal zero particles/mL and 20 mA will equal the full scale (FS) value. If
the full scale value is set too high, the resolution of the analog output signal
will be poor. For example, if the maximum concentration of the sensor, 17000,
is selected, then smallest increment of counts that will be recorded is 4.15
particles/mL (17,000/4,096). This may be sufficient signal resolution for raw
water, but may be insufficient for filter effluent.

If filter effluent never exceeds, for example, 1000 counts/mL, then resolution
of the analog output signal will be to the nearest 0.24 particles/mL
(1,000/4,096).

Setting Cumulative or Differential Modes

Setting the analog particle count outputs as cumulative channels will minimize
the number of channels that must be set and also provide the maximum
amount of flexibility.

For example, one set four differential channels:

CH0 = 2-5 µm, CH1 = 5-7 µm, CH2 = 7-10 µm, and CH3 = 10-15 µm.

Much more can be achieved if the channels are set to be cumulative instead.

For example:

CH0 = >2µm, CH1 = >5µm, CH2 = >7µm and CH3 = >10µm.

Then, once values are received at the SCADA, values can be subtracted to
create other combinations.

To achieve differential data as 2-5µm, program the SCADA to subtract


the value in CH1 from CH0. A total of ten distinctly different values can
be achieved:

Table 5 Calculating Differential Values From Cumulative Values

CH 0 >2 µm

CH 0 - CH 1 = 2-5 µm

CH 1 >5 µm CH 0 - CH 2 = 2-7 µm

CH 1 - CH 2 = 5-7 µm CH 0 - CH 3 = 2-10 µm

CH 2 >7 µm CH 1 - CH 3 = 5-10 µm

CH 2 - CH 3 = 7-10 µm

CH 3 >10 µm

Page 34
Setting Up Analog Inputs to the Sensor and Analog Outputs from the Sensor 57050config.fm
Configuration Programming, continued

3.2.5 Calculation at the SCADA

Once each analog output is defined for lower limit, upper limit, and full scale,
the SCADA programmer must be informed as to how the information is to be
handled. For each channel, the channel range and label must be defined,
along with the lower and upper limit signal values and the divisor to apply to
the upper limit value. In all cases, the lower limit signal (4 mA) will be 0 (zero)
particles, the upper limit signal (20 mA) will be equal to the full scale value.
The full scale value must then be divided by the sample volume (mL sample).

For example: Assume a sample flow rate of 100 mL/min, a count period
of 30 seconds, resulting in a sample volume of 50 mL. Assume also a
maximum expected particle count of 1000 particles/mL.

Then, FS value = 1000 particles/mL X 50 mL = 50,000 CH0 set to cumulative


particle counts > 2 µm, then:

CH Lower Upper Full Scale Label at SCADA 4 mA = 20 mA = Divide by*


0 2 0 50000 >2 µm 0 50000 50
1 5 0 3200** >5 µm 0 3200 50
2 7 0 1166** >7 µm 0 1166 50
3 10 0 400** >10 µm 0 400 50
4
5
6
7
* Divide by value equals the mL sample used to calculate the full scale value.
** Divisor applied from Table 4.

Page 35
57050config.fm Setting Up Analog Inputs to the Sensor and Analog Outputs from the Sensor
Configuration Programming, continued

3.3 4-20 mA Output Test


For 2200 PCX Instruments equipped with the optional analog I/O card, use
the following test to verify the analog current scaling of 4-20 mA data
acquisition instruments (SCADA, PLC, etc.) connected to the 2200 PCX
Instrument. The procedure will alternately force 4 mA or 20 mA from the
2200 PCX Instrument. An ammeter in the line will verify current flow.

Note: To avoid damage to internal 1. Turn power off to the 2200 PCX Instrument under test.
components, always power down
the 2200 PCX Instrument before 2. Disconnect RS485 "COM" line connector (485A, 485B, SGND) from the
making electrical connections. 2200 PCX Instrument Interconnect Card.

3. Connect the cable between your computer serial port 1 or 2 and


connector J1-RS232 on the 2200 PCX Instrument Interconnect Card.
Supply power to the 2200 PCX Instrument.

4. Connect a load (e.g., PLC) as described in Section 2.2 on page 24.


Attach a 100 mA DC ammeter in series with the load.

Note: The word "load" in the next 5. Access the load command menu by sending the ASCII command “load
step is case sensitive, that is, all <Enter>” to 2200 PCX Instrument using Windows Terminal. You will have
letters in the word must be either all 60 seconds after initialization (initialization takes about 5 seconds after
lower or all upper case characters. power turn-on) to access the change menu for configuration of the
2200 PCX Instrument.

6. Type the number “7” to activate the Calibrate mode. Calibration data will
begin to scroll on the monitor. Two lines of instruction while in the
Calibrate mode will also scroll. They are:

SPACE — toggles analog outputs between 4 mA and 20 mA

RETURN — return to main menu

7. Press the space bar once to initialize the system (begins current flow).
Press the space bar again to toggle either from 4 mA to 20 mA or from
20 mA to 4 mA current source as read on the ammeter. Press the space
bar repeatedly to toggle back and forth.

Note: Failure to enter the "quit" 8. When finished, press <Enter> to return to the main menu then end the
command may result in erroneous sequence by typing “Q <Enter>”. Turn power off to 2200 PCX Instrument
calibration and address values. and remove programming cable and ammeter.

9. Return all wiring to its original configuration.

Page 36
4-20 mA Output Test 57050config.fm
SCADA Worksheet
Setting the Analog Output Channels
CH Lower Upper Full Scale

Calculating Differential Values From Cumulative Values

Calculation at the SCADA


CH Lower Upper Full Scale Label at SCADA 4 mA = 20 mA = Divide by

Page 37
worksheet.fm SCADA Worksheet
Visit http: //www.hach.com
Section 4 Serial Port Protocol and Commands
This section describes 2200 PCX Instrument operation using a computer and
Vista Software. The Vista Software was designed to control 2200 PCX
Instrument operation using a computer and to download count data into
computer memory. If you are not using Vista, the information on the
following pages will still help you understand operation of the particle counter
with a computer.

Your particle counter has been set up for one of two serial data
communications capabilities: 1) RS232 serial interface circuitry provides for
using Vista Software and a computer to control a counter, or 2) RS485 serial
network circuitry provides asynchronous communications between multiple
counters and a controlling computer.

4.1 Communication Protocol


The 2200 PCX Instrument has the following fixed settings for serial
communications:

Eight(8) data bits No parity


One(1) stop bit9600 baud (RS232)

The Device Select [= 128 (ID# = 00)—equivalent to Vista sensor 01] mode is
selectable by means of a computer. This value can be changed as described
in Changing Operating Configuration on page 29.

4.2 Command and Data Syntax


The 2200 PCX Instrument responds to ASCII commands and sends a data
record that varies in length based on content. The command and data syntax
is defined below.

Commands

The following ASCII commands described below are supported by the particle
counter and are case sensitive (lower-case letter is a different command from
an upper-case letter):

“c” Start Counting (computer controlled): The counter will begin counting
without waiting for an even second boundary (quick start). Counting will
continue until stopped by the computer. The count cycle should be controlled
by the computer.

“d” Start Counting (counter controlled): The counter will begin counting
and control the count cycle based on the front-panel setting for period
(sample time).

“e” Stop Counting: The counter will immediately stop counting without
waiting for an even second boundary.

“o” Alarm Output: The analog multiplexer will set (or reset) a
current-sinking output for activating (or deactivating) an external alarm.

“C” Clear Buffer: The rotating buffer will be erased.

Page 39
57050serial.fm Serial Port Protocol and Commands
Serial Port Protocol and Commands, continued

“E” Send EPROM Revision: The counter will send the EPROM number
and revision.

“K” Sensor Cal Curve: The calibration curve will be placed (or verified)
in memory.

“M” Mode Request: The counter will send its present mode. If counting, a
“C” will be sent. If holding, an “H” will be sent. If stopped, an “S” will be sent.

“P” Program Sizes: The program size labels will be placed (or verified)
in memory.

“T” Identify Model: The counter will send a four character model number
(e.g., PCX[space]).

“A” Send Record: The next record in the rotating buffer will be sent. When
the rotating buffer is empty, a “#” will be sent. Each record is erased from the
buffer as it is sent.

If no count cycles have been completed since the counter was turned on,
then a “#” will be sent. The record can not be sent until the current count cycle
is complete.

“R” Resend Record: The last record sent will be resent. Records sent prior
to the last record have been permanently erased.

“l” Local Mode: (for factory test purposes) The counter will be set to off-line.

“U” Universal Select: The counter will respond to all commands after
receiving this command, regardless of which select code is programmed into
the counter.

“128 to 191” Counter Select: The counter will respond to all subsequent
commands when a number is sent that matches its select code, i.e. sending a
number between 128 (corresponding to ID# 00) and 191 (corresponding to
ID# 63). The counter is deselected, or made unresponsive to computer
commands, by selecting another counter.

Command Responses: The 2200 PCX Instrument will respond to all


commands and select codes by sending the command character back to the
computer. If the 2200 PCX Instrument does not recognize a command, it will
send a “?” character. If the computer is asking for a record from an empty
buffer, the 2200 PCX Instrument will send a “#” character. If the computer is
asking for a record that has already been sent, the 2200 PCX Instrument will
send a “#” character unless the computer uses the Resend Record command.

The 2200 PCX Instrument will not send any command characters if a parity or
framing error occurs.

Data: Each counter can send a stream of its data records. Data records are
strings of ASCII characters separated by end-of-record characters. Each data
record begins with 20 characters of Counter Data. The remainder of the data
record consists of Count Data for each of the six particle size ranges, Analog
Input Fields data (up to eight: two 4-20 mA, six 0-5V [in mV]—Vista supports
two inputs total), Calibration number (sensor calibration voltage in mV),
Location number that was assigned to this counter, and a Checksum
(hexadecimal number) for testing accuracy of the data transmission.

Page 40
Command and Data Syntax 57050serial.fm
Serial Port Protocol and Commands, continued

The length of the string may change with the number of data points—count
channels and analog inputs—available from the counter. Each data point
consists of a three-character tag that identifies the type of data and six data
characters, separated by spaces.

In Table 6 bold characters represent a data record showing the serial


communications format of a 2200 PCX Instrument with six size ranges and
two analog inputs. The analog inputs are tagged with AN0 (dc current level
from flow controller) and AN1 (dc current level from turbidimeter).

Table 6 Serial Communications Format


Counter Data Status $ —

Date 080193
Time 081350
Period 0130
Count Data Range 1 2.0 002682 The first two numbers designate
particle size and the final six
Range 2 5.0 000334 numbers report particle counts.
Range 3 8.0 000136
Range 4 10. 000102
Range 5 12. 000032
Range 6 15. 000009
Analog Input Fields Input A AN0 001730 The first three characters are the tag
and the remaining characters are the
Input B AN1 002481 value for that tag.
Calibration CAL 001000
Location LOC 000007
Checksum C/S 001676
End. msg CRLF —

Status: When translated to a binary byte, this character will indicate the
status of the 2200 PCX Instrument. For example, the ASCII character “$” has
a decimal value of 36, which when converted to a binary byte sets the 3rd and
6th bits:

Decoding the Status Character

ASCII Character Meaning Decimal Equivalent Binary Equivalent (bit 76543210)


Blank Space no alarms 32 00100000
! SENSOR fail alarm 33 00100001
$ alarm/count alarm 36 00100100

Page 41
57050serial.fm Command and Data Syntax
Serial Port Protocol and Commands, continued

In this example the ASCII character “$” translates to a status byte that is
indicating a “Count Alarm”.

Date: Date information is carried in the 3rd through 8th characters of the
record. The 2nd character is always a space, to separate the status character
from the date characters. The date is arranged as MMDDYY (Month Day
Year). The example in Table 6 shows the date as August 1, 1993, the day the
2200 PCX Instrument collected the data in the record.

Time: Time information is carried in the 10th through the 15th characters of
the record. The 9th character is always a space, to separate the date from the
time. The time is arranged as HHMMSS (Hours Minutes Seconds) military
time. In the example on the preceding page, the time is 8:13 A.M. and
50 seconds.

Period: The period is the sample time, or the length of counting time. The
period information is carried in the 17th through 20th characters. The 16th
character is always a space, to separate the time in the 17th through 20th
characters. The period is presented in minutes and seconds. In the example
above the period was 6 seconds. When the period is controlled by the
computer (e command), the period characters will be zeros. When the period
(sample time) is controlled by the 2200 PCX Instrument (d command), the
period characters will represent the sample time.

Data 1, 2, 3, 4, 5, and 6: These characters contain data from the different


programmed 2200 PCX Instrument size ranges. Each data value is preceded
by a three-character tag that identifies the size range of data that follows.

The tags and data are each preceded by a space character for separation.
The record will contain as many tag/data elements as required (a minimum of
seven and a maximum of ten).

Tags: The tags contain three characters that identify the type of data that will
follow. If the data is a particle count, the tag will indicate the particle size.
In the example above the first tag indicates that particle count data follows for
the 2.0 micron particle size range. The record data will be in the following
units of measure:

Particle Counts....counts

Analog Inputs......millivolts

Data: The data are six numeric characters, preceded by a space.

Checksum: The sum of the ASCII value of each character in the data string.
Used for testing accuracy of data transmission.

End of Message: The end of message characters will immediately follow the
last tag/data element. There will be no separating space. The end of message
characters are a carriage return and line feed.

Page 42
Command and Data Syntax 57050serial.fm
Section 5 Maintenance

5.1 Cleaning the Sensor


Each on-line sensor contains a cell assembly that channels water through the
laser beam. Occasionally, this cell may become dirty (or develop a film
coating), affecting the normal calibration of the sensor. If this occurs, the
Calibration Fail indicator (LED on the front of the unit) will illuminate. Cleaning
the sensor may be accomplished without opening the NEMA enclosure.

Note: Handle all chemical solutions Hach sensors are designed to make maintenance quick and easy. Hach
with care. Hach recommends use of particle counting sensors are designed to be cleaned with a specially
rubber gloves and goggles when designed brush. The cell of the sensor is constructed of material that is harder
handling all cleaning solutions. Treat than the brush bristles; hence, the brush will not scratch or otherwise damage
all chemical solutions with respect. the cell.

A variety of dilute non-abrasive cleaning solutions may be used; never


use concentrated acids or bases. Concentrated solutions may damage
sensor components.

5.1.1 Cleaning Frequency

Cleaning requirements vary widely. Typically, sensors used to monitor


clean samples such as filter effluent should be cleaned about once per
month. Sensors used to monitor untreated water or clarifier effluent may need
to be cleaned weekly. The best policy is to clean sensors as often as
experience dictates. Excursions of high turbidity, minerals (iron, manganese,
calcium, etc.) and algae or other microbiological growths may increase the
need for cleaning.

5.1.2 Brush Cleaning

Often, a cell becomes dirty when the sample dries out in the flow cell and
leaves small amounts of residue on the flow cell surface. For this reason,
Hach ships liquid particle sensors cleaned and dry. If the Calibration Fail LED
lights, clean the cell using Figure 18 and the procedure below. This procedure
will not damage the cell.

1. Remove the sample line quick-disconnect fitting at the inlet port (bottom)
of the NEMA enclosure. Leave the quick-disconnect connected at the top.

2. Insert the cleaning brush up through the bottom flow path as shown
in Figure 18. Use a gentle laboratory cleaning solution to enhance the
cleaning and lubricate the brush/cell. Gently insert the brush and twirl
it as it is being inserted into the bottom fitting. It will bottom out to a stop.
Repeat several times, then reconnect the inlet flow connector to flush
the cell.

3. Observe the Calibration Fail indicator. The indicator should go off within a
few seconds.

4. If the Calibration Fail indicator remains on, perform the stain cleaning
procedure in Section 5.1.3 or contact the Hach Customer Service
Department.

Page 43
57050maintenance.fm Maintenance
Maintenance, continued

5.1.3 Stain Cleaning

If brush cleaning is unsuccessful in restoring the sample cell and resolving the
Calibration Fail condition, then the cell may have been chemically stained.
Disconnect the particle sensor from the normal on-line flow path and
chemically dissolve the stain as follows:

• For microbiological (green) growths, soak the cell with 30 to 50 mL of 70%


or 90% isopropyl alcohol.

Dilute solutions of household chlorine bleach (5.25% available chlorine)


also may be used. Dilute the bleach approximately 1:1000 (1mL bleach to
1 liter of water) to prepare a 50 mg/L cleaning solution. Stronger bleach
solutions may be used to remove severe growths. Soak the cell with 30 to
50 mL of the bleach solution. Rinse with clean water.

• For red mineral deposits (iron, etc.), soak the cell with an iron reducing
agent (Hach RoVer®) then flush the cell with water.

• For calcium (white) deposits, soak the cell with white vinegar or
phosphoric acid then flush the cell with water.

• For mild manganese stains (purple or black), soak the cell with a solution
(by volume) of 1/3 water, 1/3 white vinegar, and 1/3 hydrogen peroxide;
flush the cell with water.

• For severe manganese stains, soak the cell with a solution by volume of
70% white vinegar and 30% hydrogen peroxide (3% strength).

Reconnect the sensor to the flow path. Check the cell condition indicator with
a fully wetted cell. If the cell indicator remains on, try the chemical soak using
one of the other solutions discussed above, or try soaking for a longer period
of time. Contact Hach Customer Service for additional recommendations.

Page 44
Cleaning the Sensor 57050maintenance.fm
Maintenance, continued

Figure 18 Cleaning the Cell

I/O MCA I/O ANALOG


G
4-20mA OUT SERIAL 4-20mA OUT IN
MCA BANK 1
POWER BANK 2
I/O 4-20Ma IN
ALARM OUT

J6 J8 J2 J7 J3 J4 J5

J1
RS232

INVISIBLE LASER RADIATION WHEN OPEN


—AVOID DIRECT EXPOSURE TO BEAM—

DANGER
ANTY
VOID The brush should stop here.
WARRWARRAN NTY VOID
ROKE
IF BIF BRTY ENID
OKVO
WARRAN
IF BROKEN

+15V I_PWR OUT4 AIN2

AGND OUT0 OUT5 AIN3

-15V OUT1 OUT6 AIN4

+5V OUT2 OUT7 AIN5

DGND OUT3 IN0 AIN6

SGND I_RET RET0 AIN7

JP1
485A ALM1 IN1 AINGND

485B ALM2 RET1 SPARE

Disconnect the
Sample Inlet Line

Gently insert the brush into the


Sample Inlet Port.

Twirl the brush to clean.

5.2 Long-term Storage or Shipping


Clean the sensor thoroughly prior to long-term storage or shipment.
“Long-term” storage is defined as a period of time longer than two weeks.
Do not leave water in the sensor; water spots could develop on the cell
windows and may be difficult to remove. Water left in the cell during storage or
shipment in cold weather may freeze and cause severe damage to the sensor.

To prepare the sensor for long-term storage or shipping, clean the sensor with
appropriate cleaning solutions (see Section 5.1). Then, flush approximately
30 mL of isopropyl alcohol through the sensor to remove any remaining water.

Page 45
57050maintenance.fm Long-term Storage or Shipping
Maintenance, continued

5.3 Replacing or Removing the Sensor Flow Cell


The particle counter sensor has a field-replaceable sample cell. If the cell
becomes damaged or coated with a substance that cleaning solutions will not
remove, the sample cell can be replaced and calibration is not affected.

The replacement must be done carefully and correctly to avoid


damage to the sensor. Please contact Hach Service Department for
replacement information.

5.4 Tubing Replacement


The 2200 PCX sensor is supplied with tubing that has been specially selected
to minimize accumulation of dirt and mineral deposits. Replace it only with the
same tubing size and type.

Change tubing as often as is necessary, depending on your sample


conditions. Typically, for treated water samples (filter effluent) change the
tubing once per year. For sensors used to monitor clarifier effluent, change
the tubing every six months. For sensors used to monitor untreated water,
replace the tubing on an approximately three-month interval.

Page 46
Replacing or Removing the Sensor Flow Cell 57050maintenance.fm
GENERAL INFORMATION

At Hach Company, customer service is an


important part of every product we make.

With that in mind, we have compiled the following


information for your convenience.

Page 47
57050gi_stop.fm GENERAL INFORMATION
Certification
Hach Company certifies this instrument was tested thoroughly, inspected
and found to meet its published specifications when it was shipped from
the factory.

The 2200 PCX Particle Counter has been tested and is certified as indicated
to the following instrumentation standards:

Product Safety
UL/CSA approved 100-115 V, 50/60 Hz external wall-style power supply

or

100-240 V, 50/60 Hz external NEMA-enclosed power supply certified with the


2200 PCX to EN 61010-1 (IEC1010-1) per 73/23/EEC, supporting test
records with Elliot Laboratories, Inc.

Immunity
The 2200 PCX Particle Counter tested the Sensor and 100-240 V, 50/60 Hz
external NEMA-enclosed power supply.

EN 50081-2: 1992 (Generic Immunity Standard) per 89/336/EEC EMC:


Supporting test records with Elliot Laboratories, Inc.

Required Standard/s include:


IEC 801-2:1991 Electro-static Discharge
IEC 801-3 Radiated RF Electro-magnetic Fields
IEC 801-4 Electrical Fast Transients/Burst

Emissions
The 2200 PCX Particle Counter tested the Sensor and 100-240 V, 50/60 Hz
external NEMA enclosed power supply.

EN 50081-2 (Generic Emissions Standard) per 89/336/EEC EMC:.


Supporting test records by Elliot Laboratories, Inc. (NVLAP # 200069-0).

Required Standard/s include:


EN 55011 (CISPR 11) Emissions, Class A Limits

Canadian Interference-causing Equipment Regulation, Chapter 1374, Class A


Supporting test records with Elliot Laboratories, Inc. (NVLAP # 200069-0).

This Class A digital apparatus meets all requirements of the Canadian


Interference- Causing Equipment Regulations.

Cet appareil numérique de la classe A respecte toutes les exigences du


Règlement sur le matériel brouilleur du Canada.

Page 48
Certification 5705018Certification.fm
Certification, continued

FCC Part 15, Class "A" Limits


Supporting test records with Elliot Laboratories, Inc. (NVLAP #200069-0).

This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions:

(1) this device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may
cause undesired operation.

Changes or modifications to this unit not expressly approved by the party


responsible for compliance could void the user’s authority to operate the
equipment.

This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when
the equipment is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference, in which case the user
will be required to correct the interference at his own expense. The following
techniques of reducing the interference problems are applied easily.

1. Disconnect the Model 2200 PCX Particle Counter from it’s power source
to verify that it is or is not the source of the interference.

2. If the Model 2200 PCX Particle Counter is connected into the same outlet
as the device with which it is interfering, try another outlet.

3. Move the Model 2200 PCX Particle Counter away from the device
receiving the interference.

4. Reposition the receiving antenna for the device receiving the interference.

5. Try combinations of the above.

Page 49
5705018Certification.fm Certification
How to Order
By Telephone: By Mail:
6:30 a.m. to 5:00 p.m. MST Hach Company
Monday through Friday P.O. Box 389
(800) 227-HACH (800-227-4224) Loveland, Colorado 80539-0389
U.S.A.
By Fax:
(970) 669-2932 Ordering information by E-mail:
orders@hach.com

Information Required
• Hach account number (if available) • Billing address
• Your name and phone number • Shipping address
• Purchase order number • Catalog number
• Brief description or model number • Quantity

Technical and Customer Service (U.S.A. only)


Hach Technical and Customer Service Department personnel are
eager to answer questions about our products and their use.
Specialists in analytical methods, they are happy to put their talents to
work for you. Call 1-800-227-4224 or E-mail techhelp@hach.com.

International Customers
Hach maintains a worldwide network of dealers and distributors.
To locate the representative nearest you, send E-mail to
intl@hach.com or contact:

Hach Company World Headquarters; Loveland, Colorado, U.S.A.


Telephone: (970) 669-3050; FAX: (970) 669-2932

In Canada:
Hach Sales & Service Canada Ltd.; Manitoba, Canada
Telephone: (204) 632-5598; FAX: (204) 694-5134

Page 50
How to Order 57050end.fm
Repair Service
Authorization must be obtained from Hach Company before sending
any items for repair. Please contact the HACH Service Center serving
your location.

In the United States:


Hach Company
100 Dayton Avenue
Ames, Iowa 50010
(800) 227-4224 (U.S.A. only)
Telephone: (515) 232-2533
FAX: (515) 232-1276

In Canada:
Hach Sales & Service Canada Ltd.
1313 Border Street, Unit 34
Winnipeg, Manitoba
R3H 0X4
(800) 665-7635 (Canada only)
Telephone: (204) 632-5598
FAX: (204) 694-5134
E-mail: canada@hach.com

In Latin America, the Caribbean, the Far East, the


Indian Subcontinent, Africa, Europe, or the Middle East:
Hach Company World Headquarters,
P.O. Box 389
Loveland, Colorado, 80539-0389 U.S.A.
Telephone: (970) 669-3050
FAX: (970) 669-2932
E-mail: intl@hach.com

Page 51
57050end.fm Repair Service
Warranty
Hach warrants most products against defective materials or workmanship
for at least one year from the date of shipment; longer warranties may apply
to some items.

HACH WARRANTS TO THE ORIGINAL BUYER THAT HACH


PRODUCTS WILL CONFORM TO ANY EXPRESS WRITTEN
WARRANTY GIVEN BY HACH TO THE BUYER. EXCEPT AS
EXPRESSLY SET FORTH IN THE PRECEDING SENTENCE,
HACH MAKES NO WARRANTY OF ANY KIND WHATSOEVER
WITH RESPECT TO ANY PRODUCTS. HACH EXPRESSLY
DISCLAIMS ANY WARRANTIES IMPLIED BY LAW, INCLUDING BUT
NOT BINDING TO ANY WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE.

LIMITATION OF REMEDIES: Hach shall, at its option, replace or


repair nonconforming products or refund all amounts paid by the buyer.
THIS IS THE EXCLUSIVE REMEDY FOR ANY BREACH OF WARRANTY.

LIMITATION OF DAMAGES: IN NO EVENT SHALL HACH BE LIABLE FOR


ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND FOR
BREACH OF ANY WARRANTY, NEGLIGENCE, ON THE BASIS OF STRICT
LIABILITY, OR OTHERWISE.

This warranty applies to Hach products purchased and delivered


in the United States.

Catalog descriptions, pictures and specification, although accurate to the


best of our knowledge, are not guarantee or warranty.

For a complete description of Hach Company’s warranty policy, request


a copy of our Terms and Conditions of Sale for U.S. Sales from our
Customer Service Department.

Page 52
Warranty 57050end.fm
LIST OF REFERENCES

../../ART/hach.eps ........................................................................................................................................................... 1
../../MASTERPG/85X11/BOILERS/Safety.eps ................................................................................................................ 5
../../ART/hach.eps ........................................................................................................................................................... 9
../../Art06000/06505a.eps.............................................................................................................................................. 11
../../Art06000/06507a.eps.............................................................................................................................................. 12
../../Art06000/06506a.eps.............................................................................................................................................. 13
../../art/hach.eps ............................................................................................................................................................ 15
../../Art06000/06508a.eps.............................................................................................................................................. 17
../../Art03000/03540.eps................................................................................................................................................ 18
../../Art06000/06510b.eps.............................................................................................................................................. 18
../../Art06000/06509a.eps.............................................................................................................................................. 19
../../Art06000/06511a.eps.............................................................................................................................................. 20
../../Art06000/06512a.eps.............................................................................................................................................. 21
../../Art06000/06512b.eps.............................................................................................................................................. 22
../../Art06000/06516a.eps.............................................................................................................................................. 24
../../Art06000/06513a.eps.............................................................................................................................................. 25
../../Art06000/06514a.eps.............................................................................................................................................. 27
../../Art06000/06515a.eps.............................................................................................................................................. 28
../../Art06000/06517a.eps.............................................................................................................................................. 29
../../Art06000/06518a.eps.............................................................................................................................................. 40
../../MASTERPG/85X11/BOILERS/order.eps................................................................................................................ 45
../../MASTERPG/85X11/BOILERS/repair.eps ............................................................................................................... 46
../../MASTERPG/85X11/BOILERS/warranty.eps .......................................................................................................... 47
../../Art00100/00568.eps................................................................................................................................................ 48

53

Você também pode gostar