Odyssey DR/2500 Illuminator Module Replacement

Replacing the Illuminator Module


1. Turn off the power switch and remove the cell holder. Disconnect the power
supply, the pH probe, and the computer or printer cables.
2. Place the instrument upside down on a padded surface.
Use the T10 L-Key supplied in the Illuminator Replacement Module Kit
(Cat. No. 59059-00) to remove the three Torx screws that secure the
illuminator module (Figure 1).
Warning: Some internal components may be damaged by static electricity. Use
static discharge protection.
3. Remove the illuminator module by pushing on the cell holder cavity from the
top of the instrument and pushing the module out the bottom of the
instrument (Figure 1).

Figure 1 Removing the Illuminator Module

T10 L-Key

Push Down
Torx Screws (3 places)

1
Odyssey DR/2500 Illuminator Module Replacement, continued

Note: Do not touch the 4. Insert the new illuminator module from the bottom of the instrument. Make
three optical surfaces of sure the connector pins on the module line up with the connector inside the
the illuminator module! Odyssey. Do not force the module into place. Secure the module with the
(See Figure 2.)
three Torx screws (Figure 1).

Figure 2 Inserting New Illuminator Module

2
3 Optical Surfaces (3 places)

Connector Pins

5. Return the instrument to the upright position and restore power and other
cable connections. No calibration is necessary for the illuminator module.

Components
Description Cat. No.
Illuminator Module Replacement................................................................................................................ 59059-00
Instruction Sheet ............................................................................................................................................. 59059-88
T10 L-Key......................................................................................................................................................... 59117-00
Torx Screws...................................................................................................................................................... 49440-00

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: HACH COMPANY


In the U.S.A. – Call toll-free 800-227-4224 WORLD HEADQUARTERS
Outside the U.S.A. – Contact the HACH office or distributor serving you. Telephone: (970) 669-3050
On the Worldwide Web – www.hach.com; E-mail – techhelp@hach.com FAX: (970) 669-2932

© Hach Company, 2001. All rights reserved. Printed in the U.S.A. te/dk 06/01 1 ed

Você também pode gostar