DIFFUSION IN AIR

AIM: To determine the diffusion coefficient of Liquid Gas System (acetone air system). APPARATUS: LiquidGas diffusion system setup, acetone, water (for water bath). THEORY: When a concentration gradient exists within a system consisting of one or more components, the components flow in such a direction, so as to reduce the concentration gradient. This process is called Mass Transfer. The mass transfer of components takes place by random motion of molecules and is known as diffusion. In a liquid gas system, when the liquid evaporates into still gas, the vapour formed on the surface of the liquid is transferred into the bulk of gas by diffusion. The rate of diffusion depends on the ease with which the molecules can travel through the gas. This factor is called diffusivity. Refer to the figure; here the liquid evaporates in a vertical glass tube in still air. By Stephens law for diffusion of vapour through a stationary gas, the mass transfer rate is given by the equation: NA = Where D = diffusivity of vapour through air (m2/s) y2 y1 = the distance through which the vapour travels in the direction of diffusion (m). CT = the total concentration of vapour and gas. CA1, CA2 = concentrations of vapour at the surface of the liquid and at the end of the tube respectively (kmol/m3). Replacing (y2-y1) by x and writing the concentration in terms of vapour pressure we get NA = DP C B 2 ln xRT C B1 DP PB 2 ln xRT PB1 DCT C C A2 ln T y 2 y1 CT C A1

Where CB1, CB2 = concentration of air the surface of the liquid and at the end of the tube respectively. (kmol/ m3). PB1, PB2 = partial pressure of air at the above mentioned points, total pressure (kN/m2). P = total pressure (kN/m2). This rate of mass transfer will be equal to the rate of evaporation of liquid, which in turn can be followed by the fall in level of liquid surface.

NA = Where,

L dx M dt

= density of liquid (kg /m3) M = molecular weight of liquid (kg/kmol) dx/dt = rate of fall of liquid level P dx D P ln B 2 = L x RT PB1 M dt

Integrating between the limits x = x1 to x2 when t changes from 0 to t x t 2 DP P ln dt = L xdx RT P 0 M x1


2 x 2 x12 1 P 2 t ln B 2 PB1 All the other values in the above equation being known, D can be calculated.

RT D = P M

EQUIPMENT DESCRIPTION: The set up consists of: A diffusion column immersed in a water bath. A blower for circulation of air. Orificemeter and manometer for measuring the air flow rate. An electrical heater for heating the water in the water bath. A temperature sensor for measuring the temperature of the liquid. Digital temperature indicator. PROCEDURE: 1. Fill acetone in the process column upto the desired level. 2. Check the water level in the water bath for heating acetone. 3. Switch on the main switch of the control panel. 4. Close the flow regulating valve partially and start the blower. 5. Adjust the opening to get the desired flow rate of air. 6. Note the level of liquid in the tube at time t = 0 as x1 (m). 7. Allow the liquid level to fall to another level (x2) and note the time required for it. 8. Repeat the experiment for different temperatures. 9. Switch off the heater and blower 10. Switch off the main control OBSERVATION TABLE: -

Sr. No.

Temperature C

Time (sec)

Liquid level Initial x1(m) Final x2(m)

PB1
2 kN/m

D m2/s

Using the formula below D is calculated D = RT P M


2 x 2 x12 1 P 2 t ln B 2 PB1

Where, P = atmospheric pressure (KN/m2) PB2 = atmospheric pressure (KN/m2) PB1=difference in atmospheric pressure and vapour pressure of acetone at the operating temperature (KN/m2) t = time for which evaporation occurs (s) RESULT: The calculated value of Diffusivity is

CONCLUSION:

LIQUID-LIQUID DIFFUSION
AIM: To determine diffusion co-efficient for a liquid-liquid system (Acetic acid- water) APPRATUS: Porous cylinder, beaker, conical flask, acetic acid, distilled water, burette, pipette. THEORY: If in a liquid there exist a concentration gradient, the molecules of the liquid will migrate in such a way so as to reduce concentration gradient. This phenomenon is called diffusion depends on magnitude of concentration gradient. The basic law defining mass transfer rate is known as Ficks law given by, NA = DAB dCA/dy Where, NA DAB = molar flux (Kmol/m2sec). = Diffusivity of compound A in B (m2/sec).

dCA/dy= Concentration gradient (Kmol/m3)/m. In addition to diffusion movement of A occurs by bulk flow also. N=N A+NB gives the net flux & total rate of transfer of A is given by, NA = ( N A + N B )C A dC A D AB C dy

This equation can be integrated if DAB & C are constant. In case of liquid, these vary considerably with concentration. Integrated form of above equation is given by; NA = NA D AB C N A /( N A + N B ) x A2 ln (N A + N B ) z N A /( N A + N B ) x A1

In case of liquid DAB & C both vary with concentration & therefore an average value of above equation is written as, NA = NA N /( N A + N B ) x A2 ln A ( N A + N B ) M avg N A /( N A + N B ) x A1

For steady state diffusion of A through non-diffusing B, NB=0. In this particular case we consider only diffusion of A in B NA = (1 x A2 ) M ln (1 x ) avg A1 D x = AB ln B 2 z M avg x B1 D AB z

PROCEDURE: 1. Fill the porous pot with acetic-acid of known concentration (4N) 2. Place it in beaker. 3. Pour known quantity of distilled water in beaker. 4. Start the stop clock. 5. Note thickness of porous pot. 6. Acetic acid diffuses in to water. 7. After every 15 minutes take a sample from beaker & titrate it against 0.05 N NaOH solution to find quantity of acetic acid in it, using phenolphthaliene indicator. 8. Using the formula given calculate DAB for each reading. OBSERVATIONS: Inner radius of porous pot (r1) Outer radius of porous pot (r2) Thickness of porous pot (z) = (r2- r1) Initial conc. of acetic acid Length of porous pot involved in transfer = = = = =

OBSERVATION TABLE: Sr. No. Time (sec) Burette Reading (ml) Normality Strength of of solution solution in in beaker beaker (g/lit) (N) Total amount Mass transfer Diffusivity of acetic acid rate (NA) DAB (m2/sec) 2 in beaker (kmol/m sec) (gm)

CALCULATIONS: 1. To calculate normality of solution in beaker using equation N1V1 = N2V2 N1V1 = Solution in beaker N2V2 = NaOH used for titration N1 = N 2V2 = V1

2. To calculate strength of solution in beaker (which gives g/lit of acetic acid diffused) = Normality (N1) x Equivalent weight of acetic acid = 3. Total gm of acetic acid present in the beaker = (Strength of solution in beaker *total volume of water in beaker). = 4. Initial conc. of acetic acid (A) = x N = _______N. = x * eq. wt. of A (gm/lit of solution) =

kg/lit of solution.

Initial mass fraction of acetic acid = (wt of acetic acid)/(wt.of acetic acid + wt. of water) =

Initial mass fraction of water = (wt.of water)/(wt.of acetic acid + wt. of water) = Density of solution inside pot (1) = (Mass fraction of acetic acid * density of acetic acid) + (mass fraction of water * density of water) kg/m3 = 5. Mole fraction of acetic acid (xA1) inside pot = (wt. of acetic acid inside the pot) / (molecular wt of AA)___ (wt of AA / mol wt of AA) + (wt of water / mol wt of water) = Mole fraction of water inside the pot (xB1) = 1 xA1 = 6. Mole fraction of acetic acid outside pot (xA2 ) = (wt of acetic acid diffused/mol.wt of acetic acid)__________ (wt of AA / mol wt of AA) + (wt of water / mol wt of water) = Mole fraction of water xB2 = 1xA2 = 7. Molecular wt of solution inside pot (M1) = xiMi. 8. 1/M1 = 2/M2 = (/M)Avg = x B 2 x B1 x = ln B 2 x B1

9.

x Bm =

10. Area of mass transfer A = . DO L 11. NA = N /At kmol / m2 s =

12. Diffusivity =

RESULT: Value of diffusivity is

CONCLUSION: .

WETTED WALL COLUMN


AIM: To find the Mass Transfer coefficient for air-water system using a wetted wall column. THEORY: Wetted wall column is mainly used for experimental purpose, to estimate the value of mass transfer coefficient for a gas-liquid system. In this equipment, liquid is allowed to flow down the inside wall of the column, so that, the entire inner surface of the column is wetted. The gas is allowed to enter at the bottom and mass transfer coefficient is evaluated. Ripples are avoided so that the inner surface area of the column can be taken as mass transfer area. The advantage of using a wetted wall column is that the mass transfer area is known. Though, industrially it has limited application, it provides a very useful means for conducting experiments in the lab. This provides a basis from which correlation can be developed for packed towers. It also provides a means for determining the importance of various factors affecting the mass transfer coefficient. The general equation for mass transfer used for a wetted wall column is: hD d PBm = 0.023 Re 0.83 Sc 0.44 DV P Where, hD = mass transfer coefficient = kG RT and kG = DV P RTz G PBm

= gas film thickness (m) Dv= diffusivity (m2/s) P = total pressure (kN/m2) d = diameter of the column Re = Reynolds no. based on gas phase = du/ Sc= Schmidt no. = / DV = gas phase viscosity = gas phase density.
ZG

EQUIPMENT DESCRIPTION: The wetted wall column set-up consists of the following: A glass column with a small liquid reservoir at the top. An inlet for air at the bottom A rotameter for measuring the inlet airflow rate. A second liquid reservoir kept at such a height so as to allow flow of liquid to the first reservoir by gravity. A hygrometer to measure the humidity of inlet and outlet air.

PROCEDURE: 1. Allow the water from the reservoir at the top to flow down to the small reservoir in such a way that the water overflows into the column, a thin layer covers the entire inner surface of the column. Keep the flow rate minimum. Measure this flow rate of water. Start the air supply and set it to the desired flow rate. Measure the humidity of inlet air. Measure the humidity of outlet air. Note the inlet air pressure. Note the temperature of inlet and outlet air, and inlet and outlet water.

2. 3. 4. 5. 6. 7.

OBSERVATIONS: 1. Diameter of the column 2. Height of the column = = = m m kN/m2

3. Inlet air pressure


OBSERVATION TABLE: Sr. No. Flow rate Flow Temp. of water rate of of (lpm) air inlet (lpm) water (oC)

Temp. of inlet air (oC)

Temp. of outlet air (oC)

Humidity of inlet air (RH1) %RH

Humidity of outlet air (RH2) %RH

Outlet water temp (oC)

1 2 3

CALCULATIONS: 1. Inlet partial pressure of water vapour PA1 = RH 1 PAS 1 100 (PAS1 = Vapour pressure of water at the inlet temperature of air) =

Inlet humidity of air H1 = = 2. The outlet air humidity in kmol vapour/kmol dry air PA 2 = RH 2 PAS 2 100 (PAS2 = Vapour pressure of water at the outlet temperature of air) = Outlet humidity of air H2 = = 3. The dry air flow rate in kmol/s Air flow rate in m3/s (moist) = air flow rate in lph x 10-3 3600 = Therefore, molar flow rate of air (moist) = Volumetric flow rate of air x inlet air pressure R x absolute temperature of inlet air = 4. The rate of mass transfer: NA = dry air flow rate (outlet molal humidity- inlet molal humidity) Area for mass transfer PA 2 P PA 2 (P = total pressure) PA1 P PA1 (P = total pressure)

= Area for mass transfer = dL = d = diameter of the column L = height of the column

5. The experimental value of mass transfer coefficient k


' k G = N A ( PA2 PA1 )

by

(where, kG is in kmol/(m2s)(kN/m2)

6. The theoretical value of k

is

hD d PBm = 0.023 Re 0.83 Sc 0.44 DV P PBm = ( PB 2 PB1 ) P = ln B 2 PB1

PB1 = P - PA1, PB2 = P - PA2 Re = du / = Sc = / DV = Dv= taken from literature at temperature (T) for air water system. hD = Where, hD/RT = kG, theoretical mass transfer coefficient kG = RESULT: where =PM/RT = obtained from literature

CONCLUSION:

PACKED BED ABSORPTION


Aim: To determine the overall mass transfer coefficient for gas, liquid system using packed bed absorption. Apparatus and Chemicals: - Packed bed absorption column assembly, NaOH pellets or flakes, hydrochloric acid, barium chloride and phenolphthalein indicator. Theory: - Gas absorption is a process in which a gas mixture is contacted with a liquid for the purpose of preferentially dissolving one or more components of the gas. The absorption process may be a physical one or may be accompanied by a chemical reaction. For absorption to take place it is necessary that the gas & liquid be brought in to intimate contact with each other. For this purpose packed or plate towers are generally used, packed towers being more common. In this system under consideration CO2 is being absorbed in NaOH solution in a packed tower .The chemical reaction taking place is CO2 + 2NaOH Na2CO3 + H2O As the solution is concentrated it may be assumed that there will be negligible vapour pressure of carbon dioxide over the solution, i.e. all the resistance to mass transfer lies in the gas phase. The basic equation, which can be used to calculate the overall mass transfer coefficient, is Gm (y1-y2) = KG a P (y-ye)ln Z Where: Gm = molar gas flow rate (kmol / m2s) y = mole fraction of CO2 KG a = overall mass transfer coefficient, kmol /m3s (KN/m2) P = operating pressure (kN/m2) Z = ht. of packing (m) Subscripts 1, 2 & e represents inlet, out late & equilibrium conditions. Equipment descriptions: The set up consists of A packed column (with Raschig rings) Feed tank (10 lit) to store NaOH solution fed to column. Receiving tank (7 lit) for collecting the solution leaving the column. A compressor for air circulation. CO2 cylinder with regulator to supply CO2 to feed gas stream. Rotameters for measuring feed solution, air and CO2 flow rates Procedure: 1.

Fill the feed tank with 10 lit of 1N NaOH solution and close the tank cover.

2. Prepare 100 ml of BaCl2 solution 25% w/w (25%BaCl2 + 75% water, by wt.) 3. Adjust the air pressure regulator to 1kg/cm2 pressure, to maintain the flowrate of NaOH. 4. The flow rate of the gas (air + CO2) and liquid (NaOH) are adjusted to and respectively by valves provided on the barometer 5. The CO2 flow rate should be such that the percentage in air-CO2 mixture is 8 to 10% 6. The system is allowed to attain to steady state after fixing the liquid level. This level is adjusted by changing the valve opening at the bottom. 7. The time for steady state should be 3-4 times the mean residence time of the liquid phase in the column. During this period the liquid level in the bottom section must be maintained at marked height. 8. After steady state is achieved, liquid samples are collected at the outlet for analysis 9. A known volume V of the sample is titrated against standard HCl using phenolphthalein. End point is noted as T1, colour change:- pink to colourless 10. To the volume V of the sample 25% BaCl2 solution is added, to precipitate Na2CO3. Go on adding BaCl2 till precipitation is complete, add some excess amount. This solution is then titrated against standard 1N HCl using phenolphthalein as indicator. End point as T2, colour change: colourless to pink 11. Repeat the experiment for different flow rates. Observations:Height of column (Z) = Blank reading (T1) = Normality of HCL = Operating pressure = Cross sectional area of column = Observation table:Sr. No. Feed flow (lph) Air flow (lph) CO2 flow Titration (lph) T1 (ml of HCl) Titration T2 after adding BaCl2 (ml of HCl) KGa

(m) (ml) (N) (KN/m2) (m2)

Calculations:1. Concentration of inlet NaOH, (CNaOH)in Normality of NaOH X1 (N) T1 x concentration of HCL = Volume of sample = X1 x Eq. Wt. of NaOH (CNaOH)in = mol. Wt. Of NaOH = 2. Concentration of out let NaOH,(CNaOH)out Normality of NaOH X2 (N) T2 x concentration of HCL = Volume of sample =

X2 x Eq. Wt. of NaOH (CNaOH)Out = mol. Wt. Of NaOH = 3. Concentration of Na2CO3 in outlet stream (CNa2CO3)Out =(1/2){(CNaOH)in (CNaOH)Out}

4. Rate of absorption of CO2 RCO2 = = Flow rate at inlet m3/s 5. CO2 flow rate at inlet = Molar volume m3/Kmol = (1/2){(CNaOH)in (CNaOH)Out} x Liquid flow rate m3/s

Flow rate at inlet m3/s 6. Air flow rate at inlet = Molar volume m3/Kmol

Flow rate of CO2 (Kmol/s)in 7. Mole fraction of CO2 entering y1 = Flow rate of CO2 + Flow rate of air (Kmol/s)

Flow rate of CO2 (Kmol/s)out Mole fraction of CO2 at outlet y2 = Flow rate of CO2 + Flow rate of air (Kmol/s) =

7.(y-ye)ln = The log mean average of the concentration differences at the end of tower. y1 y 2 y = ln 1 y2 8. Overall gas transfer coefficient ( y y e ) lm = Gm (y1-y2) = KGa P Z (y-ye)ln Gm = molar flow rate at inlet = Kmol/s (unit cross sectional area of Column) [Flow rate of CO2 + Flow rate of air (Kmol/s)]in Gm = Cross sectional area of column =

P = Operating pressure = Atmospheric pressure (101.325 kN/m2) KGa = = Result: The value of KGa calculated = Gm ( y1 y 2) Pz ( y y e ) ln

Conclusion:

ENHANCEMENT FACTOR
AIM: To determine the enhancement factor for absorption using CO2, H2O & NaOH System APPARATUS: Packed bed absorption set up CHEMICALS: CO2, NaOH solution (1N), distilled water, BaCl2 (25%W/W), HCl (1N), phenolphthalein indicator THEORY: Absorption is a process used for removal of unwanted elements from a gas mixture by contacting it with a suitable solvent. If the gaseous component is easily soluble in water (e.g. NH3) water can be used as solvent for removal of this component. But if solubility in water is low (CO2) a solution with gaseous component react can be used as solvent. The reaction of solution with the solvent leads to the consumption in the solution thus helping in maintaining a low concentration gradient between the bulk of the gas & interface. Consequently increasing the rate of absorption as compared to that with non-reactive solvent. The ratio of the two rates obtained above is called as Enhancement factor Rate of absorption with reaction --------------------------------------------Rate of absorption without reaction

Enhancement Factor =

EQUIPMENT DESCRIPTION: The set up consists of 1. A packed column (with Raschig rings) 2. Feed tank 10 lit. to store NaOH solution. 3. The receiving tank (1lit) for collection of the outlet solution. 4. The compressor for air feed. 5. CO2 cylinder with regulator to supply CO2 to the feed gas stream 6. Rotameter to measure feed solution, air & CO2 flow rates PROCEDURE: 1. 2. 3. 4. 5. Fill feed tank with 1N NaOH solution. Allow CO2 and air mixture to flow up the column at a fixed flow rate. Allow NaOH solution to flow down the column at a fixed flow rate. After steady state is reached collect the sample. Repeat the procedure for different flow rates of NaOH solution keeping CO2 and airflow rates constant. 6. Analyze the collected sample for Na2CO3 content to calculate rate of absorption of CO2. 7. Drain remaining NaOH from the tank and fill the tank with distilled water. 8. Wash entire system by flooding with water. 9. Next fill the tank with distilled water. 10. Allow it to flow down the column at the same flow rate as those used for NaOH solution. 11. Keeping the CO2 and air flow rates again at the same value, collect the samples for same flow rates of H2O as flowrates of NaOH solution. 12. Analyze these samples for their CO2 content and calculate the rate of absorption of CO2. 13. The enhancement factor can now calculated by taking the ratio of the rate calculated in steps from (1) to (6) to the rate calculated in remaining steps for respective NaOH and water flow rates. OBSERVATIONS: Blank reading = OBSERVATIONS TABLE:

Sr. No. 1 2 3

Feed Flow Rate (LPH)

Air Flow Rate (LPH)

CO2 Flow Rate (LPH)

Titration Readings NaOH T1 (ml) Water T2 (ml)

CALCULATIONS: A) For absorption of CO2 in NaOH 1) Concentration of NaOH at inlet Blank reading x 1 Normality = --------------------------25 = Normality * Equivalent weight C(NaOH)inlet = ------------------------------------------------Molecular weight = 2. Concentration of NaOH at outlet Normality = Normality * Equivalent weight = -----------------------------------------------Molecular weight = 3. Concentration of Na2CO3 at outlet C (Na2CO3) = (1/2) (C (NaOH)inlet C (NaOH)out) =

C(NaOH)out

4. Rate of absorption of CO2 1N NaOH (R co2 ) 1 = C( Na2CO3) (Kmole/m3 ) * liquid flow rate (m3/sec)= = Kmol/sec

B) For absorption of CO2 in water 1. Concentration of H2CO3: Normality = Normality * Eq. wt. C(H2CO3) = ------------------------------Molecular weight = 2. Rate of absorption of CO2 in water C(H2CO3) * liquid flow rate (m3/sec) (RCO2)2 = ------------------------------------------------Molecular weight of H2CO3 = 3. Enhancement Factor (RCO2)1 E = -------------------(RCO2)2 RESULT: Sr.No. Rate of absorption of CO2 In NaOH (RCO2) In water (kmol / sec) (RCO2) 2 (kmol/sec) Enhancement Factor E

1 2 3

CONCLUSION:

INDIVIDUAL AND OVERALL MASS TRANSFER COEFFICIENT

AIM: To determine individual and overall mass transfer coefficient for gas-liquid system. APPARATUS AND CHEMICALS: Packed bed absorption column, sodium hydroxide solution (1 N), Hydrochloric acid (1 N), BaCl2, CO2 gas cylinder, Air compressor, and Phenolphthalein indicator. THEORY: When gas is absorbed in solution, the solute component has to overcome interfacial resistance before it can be absorbed in the bulk of liquid using two-film theory to represent the process. The transfer of solute component from bulk of gas to bulk of liquid is as shown in the figure. The solute particles from gas diffuse through gas film near interface and partial pressure drops from PAG to PAi at interface. The component passes across the liquid film before it reaches bulk of liquid phase. The rate of mass transfer through liquid and gas film becomes equal when system is at steady state and is, NA = kG (PAG - PAi) = kL (CAi CAL) CAi = Concentration of solute at the interface PAi = Partial pressure in gas phase at the interface Relation between PAi and CAi is given by PAi = H CAi Where H = Henrys constant Measurement of kG and kL requires concentrations at interface, which are not easily measurable hence; overall mass transfer coefficients are measured. The rate of mass transfer is given by NA = KG (PAG - PA*) = KL (CA* CAL) CA* = Concentration of solute in liquid phase which is in equilibrium with PAG PA* = Partial pressure in gas phase in equilibrium with CAL Relation between Overall and Individual mass transfer coefficients is given as 1/KG = 1/kG + H/kL (1) 1/KL = 1/kL + 1/H kG (2) When NaOH is contacted with mixture of air-CO2 overall KG involves kG and kL but when NaOH solution is contacted with CO2 alone resistance to transfer of solute component in the gas phase becomes negligible i.e. kG becomes very large so the above equation reduces to 1/KG = H/kL (since 1/kG = 0) and hence overall mass transfer coefficient gives values of kL. This value along with KG is used to find out kG. Thus individual and overall mass transfer coefficients can be calculated.

PROCEDURE: 1. Fill the tank with 1 N NaOH solution. 2. Allow it to flow down through column at desired rate. 3. Allow mixture of air and CO2 to flow up column. 4. Let the system achieve steady state. 5. Collect the product sample at the outlet. 6. Repeat steps from 1 to 5 for minimum two readings. 7. Stop airflow and pass only CO2. 8. Allow NaOH to flow down the column. 9. Let the system achieve steady state. 10. Collect the product sample at the outlet. 11. Analyze samples collected in the step 5 and 10 for CO2 content by titrating it with 1N HCl. OBSERVATIONS: 1) Height of the column = z 2) Diameter of the column 3) C/s of the column 4) Volume of the packed bed 5) Blank reading 6) Normality of HCl 7) Operating pressure OBSERVATION TABLES: Part 1) For Air + CO2 mixture Sr. No. 1 2 3 Part 2) For only CO2 Sr. Flow rate of NaOH No. (LPH) 1 2 -3 -Flow rate of NaOH (LPH) Air flow rate (LPH) CO2 flow rate (LPH) Titration Reading (T1) (ml) = 0.5 m =d = 0.06 m = m2 = m3 = ml =1N = 101.325 KN/m2

Air flow rate (LPH) --

CO2 flow rate (LPH)

Titration reading (T2) (ml)

CALCULATIONS: Part 1) 1. Normality (N1)

= [ 1 x Blank reading] / 25 =

2. C(NaOH)in

= [N1 x 40] / 40 =

3. C(NaOH)out

= [T1 x 1 x 40] / [25 x 40] =

4. C(Na2CO3)

= 0.5 [C(NaOH)in - C(NaOH)out] =

5. Rate of absorption of CO2 (RCO2) RCO2 per unit volume of packed bed = C(Na2CO3) x liquid flow rate (m3/sec) = 6. Rate of CO2 at outlet = CO2 flow rate at inlet - RCO2 = = 7. y1 = = 8. y2 = = 9. (y - ye)ln = log mean concentration difference = [y1 - y2] / ln(y1 /y2) mole fraction of CO2 at outlet = (rate of CO2 out)/( flowrate of CO2 + air out) kmol/sec kmole/sec

mole fraction of CO2 at inlet = (flowrate of CO2 in)/(flowrate of CO2 + air in)

10. Gm = molar flow rate of air and CO2 mixture / c/s area of packed bed = 11. KGa = Gm(y1 - y2) / P z (y - ye) = Part ll) For Absorption of CO2 only, in NaOH Rate of absorption of CO2 1. C(NaOH)out = [T1 x 1 x 40] / [25 x 40] (N = Normality of NaOH solution) = 2. R(CO2) kmole/m3

= 0.5 [C(NaOH)out - C(NaOH)in] x liquid flow rate (m3/sec) = kmole/sec

3. CAi = PAi / H [Where H = Henrys constant = 30 and PAi is total pressure]

4. CAL = 0 5. R(CO2) = kL (CAi - CAL) = 6. 1/KGa = 1/ kGa + H/kLa

RESULT: Sr. Flow rate of Overall mass Individual mass Individual mass transfer

no.

NaOH (LPH)

transfer coefficient transfer coefficient coefficient on gas side (KGa) on liquid side (kLa) (kGa)

CONCLUSION:

TRAY DRYER
AIM: To determine the drying rate characteristics using tray dryer APPARATUS: Tray dryer assembly, sand and water. THEORY: Drying refers to the final removal of moisture from a substance. It is generally carried out after evaporation, filtration or crystallization. In most of the drying process water is removed by vaporization. The moisture content of a material is usually expressed as a percentage of weight of dry material. It may be present in the material in 2 forms: Bound moisture: - Moisture present in any material in such a way that it exerts an equilibrium vapor pressure less that of pure water at the same temperature. Eg: moisture in loose chemical combination with cellulosic material, moisture absorbed, etc. Unbound moisture: - Its the moisture contained by the substance, which exerts a vapor pressure at equilibrium, equal to that of the pure liquid. Rate-of-drying-curve: - When rate of drying is plotted against percentage moisture above equilibrium value, curve, as shown in the fig. is obtained. The curve can be divided into 2 parts: 1. Constant rate period: - In this region the rate of moisture removal / unit area remains constant and it is assumed that drying takes place from a saturated surface of the material. 2. Falling rate period: - In this region the rate of drying/unit area continuously decreases, till the equilibrium moisture content is reached. In some cases this curve may have two different slopes and this period is then divided into the first and second falling rate periods Critical moisture content: - the moisture content of the material at the end of the constant rate period is called critical moisture content. Equilibrium moisture content: - the moisture content of the material at the end of the falling rate period is called as the equilibrium moisture content. This does not change on further heating of the material. Free moisture: - Its the water in excess of the equilibrium moisture content. EQUIPMENT DESCRIPTION:

Tray dryer is a one pan dryer with a blower. The complete assembly consists of:1. A drying chamber with tray. 2. A blower for passage of air 3. An electrical heater placed after the blower 4. A pan placed in the path of air 5. A weighing balance for periodic weighing of the pan with the drying material. PROCEDURE: 1. Take a measured quantity of material and add a measured quantity of water to it and place it on the pan. 2. Note the total weight of the pan and wet material. 3. Start the blower and air heater. 4. As time elapses, drying proceeds and there is loss of water which leads to reduction in the weight of the wet material. 5. Record this weight reduction against time by weighing the pan with material. 6. Continue heating till no further change in weight is observed even on continued heating. 7. Note the total time required for drying & the temperature of dryer. OBSERVATIONS: Dry weight of the material Weight of water added Area of the tray (m2) Temperature of the dryer = = = = Dimensions of the tray (L x B) =

OBSERVATION TABLE:

S No. Weight of solid (w) kg on tray

Time (t) secs

w kg.

t secs.

NA=w/A (t) kg/s m2

X= kg moisture kg dry solid

CALCULATIONS: 1. Calculate rate of drying: NA = moisture evaporated Area x time kg m2s

2. Calculate the mass ratio of solvent as: X= Weight of solvent remaining = Weight of dry solid 3. Plot a graph of NA Vs. X. Identify the const. Rate & falling rate periods & find out the values for critical & euilibrium moisture content.

4. Plot a graph of X Vs. t & calculate the value of slope = dx/dt. Using this value calculate the rate of drying using the eqn:

NA = Kg of dry solids dx Area = dt

kg m2 s

RESULTS: Equilibrium moisture content = Critical moisture content = Rate of drying :In constant rate period = In falling rate period = CONCLUSION: kg/m2 s kg/m2 s kg/kg dry solid kg/kg dry solid

COOLING TOWER
AIM: To study the characteristics of cooling tower. APPARATUS: Cooling towers are equipment used for lowering the temperature of process water so that it can be reused, by bringing it in contact with ambient air. Cooling of water takes place both by transfer of sensible heat and by evaporative cooling. As a result of this the necessary latent heat for evaporation is provided by the sensible heat of water, resulting in the lowering of water temperature. In a cooling tower, air and water are brought into intimate contact in a counter-current manner. The warm process water flows down over a series of slots, which increase the interfacial area, and promotes turbulence. As the air rises it gets heated and humidified, which the water is cooled. The air circulation in cooling tower can be effected in two ways-by natural draught and by mechanical draught. Thus there are two types of cooling towers: 1. Natural draught type, which depend on chimney effect for air circulation. The air and vapour present in the cooling tower have a higher temperature than atmospheric air. Hence, they are less dense and rise upward through the tower, allowing fresh air to enter the tower at the bottom and thus setting up an air circulation pattern. 2. Mechanical draught type in which circulation of air is effected by either a forced draught in which a fan is placed at the bottom of the tower, or an induced draught in which a fan is placed at the top of the tower removing moist air. In this type of tower the air velocity is higher than in natural draught type. The performance factor for cooling tower is given by the relation:
1 ( T T2 ) 1 + 1 + 1 + 1 KaV 1 = '* dT = 1 CP ' L 4 H H 1 H 2 H 3 H 4 T2 H

Where, H = H* - H H*= enthalpy of saturated air (kJ/kg) H = enthalpy of air (kJ/kg) T1 = inlet temperature of water (K) T2 = outlet temperature of water (K) K = mass transfer coefficient a = interfacial area per unit volume (m2/m3) Cp= specific heat of water = 4.18 kJ/kg K V = active volume (m2) L = liquid flow rate (kg/s) H1 = value of H at T2+0.1 (T1-T2) H2= value of H at T2+0.4 (T1-T2) H3 = value of H at T1-0.4(T1-T2) H4 = value of H at T1+0.1 (T1-T2)

EQUIPMENT DESCRIPTION:
The setup consists of: Cooling tower A tank at the bottom for water collection A second tank with heater to produce hot water Pump for hot water circulation Rotameter for measuring hot water flow rate Thermometer for measuring inlet and outlet temperature of water Valve for controlling hot water flow Two sets of dry and wet bulb thermometer to measure dry and wet bulb temperature of air at inlet and outlet of tower Blower for supplying air to the tower at a fixed flow rate PROCEDURE: 1. Fill water in hot water tank and ensure that the heater is completely immersed. 2. Start the heater and heat the water upto the temperature of 40-430C 3. Start the hot water pump keeping the by-pass open and set the flow of hot water at the desired rate. 4. Start the blower for air circulation 5. Allow the system to stabilize 6. Note the inlet and outlet temperatures of water. 7. Note the inlet dry and outlet dry and wet bulb temperatures of air. 8. Repeat the above for different flow rate of water. OBSERVATIONS: 1. Height of tower = 2. Velocity of air inlet = 3. Velocity of air outlet = 4. Cross- section area of air inlet = 5. Relative Humidity of inlet Air =

m m/s m/s m2 %

OBSERVATION TABLE: -

Sr. No.

Hot water flow rate (lph)

Inlet water Temp. 0 C (T1)

Outlet water Temp. 0 C (T2)

Inlet air Temp. 0 C

Outlet air Temp. 0 C

Relative Humidity of Outlet Air %RH

Water flow Rate (kg/s) L

Gas flow rate (kg/s) (G)

1 2 3

CALCULATIONS: 1. Calculate the different temperatures using the formulae mentioned below, for each rate of water, Sr. No. 1. 2. 3. 4. 3. Calculate H* & H, the enthalpy of saturated air at these temperatures, using the humidity chart. 4. Calculate H, enthalpy of air, at all these temperatures using the operating line equation G (HG2-HG1) = L CL (L2- L1) Where: H = HG2 (kJ/kg) HG1=enthalpy of inlet air (kJ/kg) = G = gas flow rate (kg/s.m2) = L = liquid flow rate (kg/s.m2) = CL= specific heat of water (kJ/kg K) = L1= outlet temperature of water L2 = various temperatures calculated above T2+(0.1)(T1-T2) T2+0.4 (T1-T2) T1-0.1 (T1-T2) T1-0.4 (T1-T2) flow

Sr. No

T2+(0.1) (T1-T2) H*

T2+(0.1) (T1-T2) H

T2+0.4 (T1-T2) H*

T2+0.4 (T1-T2) H

T1-0.1 (T1-T2) H*

T1-0.1 (T1-T2) H

T1-0.4 (T1-T2) H*

T1-0.4 (T1-T2) H

1. 2. 3. 4.

5. Calculate H1, H2, H3, &H4 & their reciprocals. Sr. No. 1. 2. 3. 4. 6. Now calculate the performance factor using the formula given earlier. H1 1/ H1 H2 1/ H2 H3 1/ H3 H4 1/ H4

RESULT: The performance factors are calculated and their values are reported in the table below: Sr. No. Performance Factor

CONCLUSION: