Você está na página 1de 7

English

Hydraderm Lift
Double Ionisation

OPERATING MANUAL

THE HYDRADERM LIFT CAN ONLY BE USED BY BEAUTY THERAPIST WHO HAVEFOLLOWED THE TRAINING COURSE

THE TECHNICAL DESCRIPTION OF THE HYDRADERM LIFT EQUIPMENT


General characteristics

Manufacturer: Model: Classification:

ROSS ESTETICA S.L. HYDRADERM LIFT - GUINOT CLASS 1 EQUIPMENT (COSMETIC APPARATUS)

Mechanical characteristics
Size: Weight: L 420 mm / W 460 mm / H 215 mm 7.5 kg

Electrical characteristics
Voltage: Frequency: T Type fuses: Power: 230 V 50 Hz 500 mA / 250 V 25 W

Protection
The apparatus has 2 x 500 mA fuses / 250 V which protect against overloading

Application characteristics
Ionisation: Stimulation: Maximum current produces 0,5 mA per electrode Low frequency current (maximum frequency 165 Hz / maximum voltage 20 V)

Cleansing
After each use of the roller electrodes, the ball electrodes and the return handle, clean with a damp towel using 70 ethyl alcohol and dry with a sterile cloth. The neon light electrode must also be cleaned after each use with 70 ethyl alcohol.

CAUTION
Before using the HYDRADERM LIFT apparatus, you need to obtain authorisation from clients using a cardiac stimulator as they run a certain risk caused by the interference or deterioration of the cardiac stimulator. In case of a breakdown, contact the nearest technical service or: ROSS C/Del Molinot 46-48 Pol.Ind Cami Ral - 08860 Castelldefels (Spain) Tl : 00 34 93 42 66 017 - export@ross.es

-1-

All Guinot-trained beauty therapists must attend a compulsory training course on the HYDRADERM LIFT apparatus prior using it. During this training they will be given a detailed training manual explaining how to use the HYDRADERM LIFT apparatus and the associated skincare products.

-2-

THE HYDRADERM LIFT APPARATUS


A GALVANIC CURRENT
A constant strength of current, used to:

Increase effectiveness of skin care cosmetics. Moisturise the skin.

A STIMULATING CURRENT
Low frequency stimulating micro-current used to:

Stimulate the contraction of the surface muscular fibres.

A HIGH FREQUENCY CURRENT


Alternative current used to:

Strengthen the diffusion and the assimilation of skin care cosmetics.

THE ELECTRODES
To transmit the effect of the currents:

Galvanic 1 positive Roller electrode (+) 1 negative Roller electrod () Stimulation 2 stimulation Ball electrodes High Frequency 1 High Frequency handle 1 neon glass electrode 3

-3-

USING THE HYDRADERM LIFT APPARATUS


IONISATION ROLLER 1. Switch on the apparatus by pushing the button in the back.

Push the ON button; automatically the Roller indicator lights up (front panel), as well as the length of the treatment: 10 minutes.

Verify the connections with the help of the diagram below.

2. Put the potentiometer at level 5 using the + and buttons located under Galvanic,
then press the MARCH button to start the treatment (green light). Adjust the intensity to skin sensitivity.

3. Move the Roller electrodes as indicated on the diagrams until the beep sounds.

-4-

FUSE

0 I

STIMULATION 1. Push the Mono Ball button


. Check the connections according to the diagram below:

2. Put the potentiometer at level 2 for the Draining step. 3. Put the potentiometer at level 5 for the Stimulation step, then increase depending
on the sensitivity of the clients skin using the + and buttons located under Stimulation.

4. Adjust the timer to 10 minutes. Then push the MARCH button to begin (green
light).

5. Move the Ball electrodes (

) around the areas indicated, in accordance with the diagrams, until the melody is obtained.

-5-

HIGH FREQUENCY OUTPUT 1. Push the High Frequency button. Adjust the strength of the potentiometer with
the + and buttons situated beneath the High Frequency sign: begin at 5 and gradually increase up to 10 (for sensitive skin do not go over 7). Check the connections according to the diagram below:

2. Adjust the timer to five minutes, then press the MARCH button. 3. Move the High Frequency electrode over the gauze while exerting a little pressure
(to avoid sparks), until the beep sounds.

Contraindications for using the device:

Pacemaker Open cuts Eczema Metal plates and metal implants under the skin Use of Vitamin A acid Bruises Do not use on thyroid area Do not use the stimulation current in case of pregnancy

Please advise the client to check with a doctor prior to having the skincare treatment if any of the above apply.

-6-

Você também pode gostar