Moving from TRACE GC Ultra to..

Moving from TRACE GC Ultra to ......

How Thermo TRACE GC Ultra was structurally concepted: Mainframe - Oven Overview Injectors - Detectors Overview Detector base body concept Gas ModulesOverview Detectors Overview

Mainframe & Oven Overview


DGFC module Detectors DPFC module

Injectors detectors cassette Electronics Detectors base bodies

Keypad
3

Injections Overview

True cold On-Column

Split/Splitless

PTV

Packed and wide-bore columns

Detector base body concept

It allows

simple detector interchangeability

GC top deck

It ensures flexibility without


adding complexity
Detector Base body

Oven

Detector base body concept

A common base which


provides the detector with main heating and gas feeding

All detectors are


available with both Digital Gas Flow Control (DGFC) and conventional pressure regulators (nonDGFC)

Moving from TRACE GC Ultra to............

.....................................

.........TRACE 1300 Series Modular Concept

.....TRACE 1300 - 1310 Series Modular Concept


A main electronic module that includes: main CPU, oven control, power supply.
- 2 versions: 220V and 120V

Injector modules include injector body, electronics for injector temperature and carrier gas control. Detector modules include detector cell and electronics for signal amplifier digitalization, for temperature and detector gas control. TRACE 1300 TRACE 1310 can accommodate 2 inlet and 2 detector modules (2 channels) self-recognized Additional external devices can be located into dedicated housing.

10

TRACE 1300 - 1310 Series Modularity Advantages


Modules can be tested (also analytically) separately prior testing the assembled unit
Modules can be shipped separately and assembled during the installation in field Modules can be kept in stock for fast replacement, upgrading, or change of configuration Possibly modules would be replaced without requiring a field service engineer

11

TRACE 1300 - 1310 Modular Concept


Trace GC 1300: Local built-in ultrasimplified user interface - two buttons and four LEDs. Trace GC 1310: Touch screen interface provides instant access for ease of use and local control

12

TRACE 1300 - 1310 Series Modular Concept

Where the parts were on TRACE GC Ultra


and

Where they are now on TRACE 1300 Series

13

TRACE GC Ultra
Injector

TRACE 1300 - 1310


Injector Module

DPFC/DCC module

EPC Board Thermoregulator


(I/O Feedback board)

Power control
14

TRACE GC Ultra
Detector Base body DGFC module Detector board Thermoregulator
(I/O Feedback board)

TRACE 1300 1310


Detector Module

EPC board Power control

15

TRACE GC Ultra
CPU Board I/O Feedback board Mother board Power supply

TRACE 1300 1310


Electronic Compartment

16

TRACE GC Ultra
GC Keypad

TRACE 1300 1310


User Interfaces 1300

User Interfaces 1310

17

TRACE 1310
Instrument Control

18

TRACE 1310
Configuration

19

TRACE 1310
Diagnostics

20

TRACE 1310
Status

21

TRACE 1310
Maintenance

*When Graph Pack adapters will be used,


the column insertion into the Injector and Detector inlets must be added of 12mm.

22

TRACE 1300 - 1310 HW connectivity

GC Control through Ethernet IP address

Direct connection to PC (reversed RED cable) Connection to LAN (standard GREY cable)

Two RS-232 ports for Autosampler control Thermo Fisher bus I2C with 5V power for additional auxiliary devices

23

TRACE 1300 - 1310 Ethernet configuration

MAC address fixed and assigned in the factory

IP Address -> need to be assigned


TCP Port -> 2551 (different from the one used on Trace GC)

Chrom-Card and Xcalibur accepts TCP Port=2551


ChromQuest has a different TCP Port -> need to be changed in TRACE 1300-1310

Ethernet Configuration details listed within Diagnostic menu

24

TRACE 1300 1310 Set up

From the PC, launch the Setup.exe file named TR1300ST (Service Tool) entering ThermoFisher as password, then, press OK. This SW load the whole package zip file as it is. Do not change any file inside.

25

TRACE 1300 1310 Set up


Automatically the SCAN will start to check how many systems are present.

26

TRACE 1300 1310 Set up


Selecting the instrument on which IP set-up and Firmware updating must be performed, automatically the two dedicated options will be activated

27

TRACE 1300 1310 IP Set up


Selecting IP Set-up a new page will be shown where GC name and IP address can be modified.

28

TRACE 1300 1310 F/W update


Selecting F/W update a new page will be shown, click on Browse package name and select where the package with the new firmware version is present (C: or USB, or..... ) Do not change any file inside.

29

TRACE 1300 1310 F/W update


On the new page will be shown the TRACE 1300-1310 configuration with the current F/W version of the modules installed and the package of the new firmware version. Select Update to start the F/W updating.

30

TRACE 1300 1310 F/W update


F/W updating in progress..............

31

TRACE 1300 1310 F/W update


At the end of the F/W updating, all the status cell must be OK. Click Done to close the program.

32

TRACE 1300 - 1310 HMI SW update from USB stick

HMI SW can be updated through the USB port positioned below the touch screen

33

TRACE 1300 - 1310 HMI SW update from USB stick


Configuration

34

TRACE 1300 1310 Series

..............more in depth

35

TRACE 1300 1310 SSL Injector Module


Manifold (Plumbing) Removable SSL Injector Body 2 Electronic Boards 3 Solenoid Valves
(Carrier, Split & Purge)

2 Removable Filters (Carrier & Split)

36

TRACE 1300 1310 SSL Injector Module


0 10 20 30

Designed to have a cool head and septum.

Depth (mm)

40 50 60 70 80 90 100 100

Septa is kept below 150C even with inlet set to 350C Lower septum bleeding (siloxanes) No septum sticking Lower air (oxygen) diffusion Longer septa lifetime The cool head SSL is optimized for cold needle injection
150 200 250 T (C) 300 350 400

37

TRACE 1300 1310 SSL Technical Specification


Temperature range:
Maximum temperature 400 C with temperature settings resolution of 1 C units Temperature stability 1 C Heater cartridge 80W 48Va.c.

Pneumatics:
Electronic control only Constant pressure Constant flow Flow regulation through calculated pressure Pressure or flow programming in three ramps and four plateaus

Pressure range:
0 to 1000 KPa, with a resolution of .015 KPa, and stability of 0.1 KPa along the entire scale

Flow ranges:
Column flow from 0.1 to 100 ml/min, with resolution of 0.01 ml/min Split flow from 10 to 1250 ml/min He with resolution of 1mL/min Septum purge flow from 0 to 50 ml/min with resolution of 1mL/min Ambient temperature and pressure compensation Gas saver function Split /Split less Surged Split/Surged Split less Large Volume Split less Back-flush capability (*) NOTE: Standard SSL CANNOT BE converted to SSL BKF a new module MUST BE requested.

Operating modes:

38

TRACE 1300 1310 PTV Injector Module


Manifold (Plumbing) Removable PTV top head Removal of bottom column connection for easier cleaning of internal body 2 Electronic Boards 3 Solenoid Valves
(Carrier, Split & Purge)

2 Removable Filters (Carrier & Split) Cooling Fan

39

TRACE 1300 1310 PTV Injector Module


Removal top body plus tubing for easier cleaning

Removal of bottom column connection for easier cleaning of internal body

40

TRACE 1300 1310 PTV Technical Specification


Temperature operating range:
Maximum temperature 450 C with temperature settings resolution in 1 C units Temperature stability 1 C maximum at 250 C for 30 minutes Temperature program rate up to 14.5C/sec Heater assy 120W 48Va.c. Capable of sub ambient operation to -50 C with LN2, 0 C with CO2 Sub-ambient control will be optional and should not add cost to base unit. Three different temperature programmable ramps, four plateaus Electronic control only Constant pressure Constant flow (Flow regulation through calculated pressure) Pressure or flow programming in three ramps and four plateaus Pressure programming: 0 to 1000 KPa, with a resolution of .015 KPa, and stability of 0.1 KPa along the entire scale Flow programming: 0.1 to 100 ml/min, with resolution of 0.01 ml/min Independent pressure/flow set points during injection Ambient pressure compensation. Control of split flow in 1mL/min from 10 to 1250 mL/min He Gas consumption saver Purge flow from 0 to 50 ml/min

Pneumatics:

Operating modes:
Split, Splitless, Solvent Split (Large Volume) Back-flush capability (*) NOTE: Standard PTV CANNOT BE converted to PTV BKF a new module MUST BE requested.

41

TRACE 1300 1310 Injector Modules


SSL Body PTV Head

42

TRACE 1300 1310 FID Detector Module


Manifold (Plumbing) Detector body 2 Electronic Boards 3 Solenoide Valves
(Air, Hydrogen & Make up)

43

TRACE 1300 1310 Detector Specifications


FID Technical Specifications
Maximum operating temperature, 450 C Flameout detection and warning Automatic flame re-ignition with shutdown of gases upon failure (3 times) MDA: <2 pg carbon/sec as propane Linear dynamic range 10^7 Acquisition rate up to 300 Hz

44

TRACE 1300 1310 Detector Specifications


TCD Technical Specifications
Maximum operating temperature, 400 C, filament ambient to 450 C Operating modes: constant mean temperature Filament protection in case of carrier gas failure Compatible with Fe/Ni or W/Re filaments MDA: 600 pg ethane/ml He carrier Linear dynamic range >= 10^6

45

TRACE 1300 1310 Detector Specifications


NPD Technical Specifications Compatibility with DETs TID-2, Agilents BLOS and other DET TID types. MDL: Nitrogen: 0.05 pg N/s or better; Phosphorus: 0.02 pg P/s or better. Dynamic and Linear Range: Dynamic range of response: 10E6; Linear range: 10E4 or better. Selectivity: N/C = 50,000 to 1 gN/gC or better; P/C = 200,000 to 1 gP/gC or better. Data acquisition rate: Up to 300Hz to be compatible with fast and ultrafast GC. Maximum Operating Temperature: Up to 450C. Heating Current setting: 1.00 to 3.50 A (or different range according to the source requirements) in 0.01 steps. Source protection in case of gas failure
46

TRACE 1300 1310 Detector Specifications


ECD Technical Specifications
Maximum operating temperature: 400 C Cell volume <= 150 l Acquisition rate up to 100 Hz Operating mode: constant current, pulse modulated MDA <= 5 fg Lindane Linear dynamic range: >10^4 Ar/CH4 make-up Equipped with removable collector electrode for local maintenance

47

TRACE 1300 1310 Pneumatic Net

48

TRACE 1300 1310 Gas Net Assembly

49

TRACE 1300 1310 Gas Net Location

50

TRACE 1300 1310 Components location


Hermes back view

electronic box detector module injector module gases inlet

main switch

toroidal transform

fan motor

oven

Hot air outlet duct

door insulation material

H2 sensor (optional)

51

TRACE 1300 1310 Oven heating


The tension rate changes from 230V to 120V but not the baffle. The serial or parallel heaters wiring is driven by the electronic box, according to the tension rate.

heating element

RTD

front view

back view

52

TRACE 1300 1310 Module Electronics Backplane


CONTROL AREA Connections for: 2 x injector 2 x detectors 1 x PT100 1 x door switch 1 x stepper (flaps) 1 x H2 sensor 3 x 24V auxiliary POWER AREA Connections for: 1 x main line 1 x 500VA pri. 1 x 500VA sec. 1 x heater 1 x fan (oven)

53

TRACE 1300 1310 Module Electronics Electronic compartment


Fan
Handles

Breaker

Oven & Power

Power switching External Interface

Main CPU

54

TRACE 1300 1310 Series Autosamplers

Injection position
55

Inlet maintenance position

TRACE 1300 1310 Series Autosamplers

Injection position

Inlet Maintenance position

56

TRACE 1300 1310 Series Autosamplers

57

TRACE 1300 1310 Series Autosamplers

58

TRACE 1300 - 1310 Series & ISQ


TRACE 1310 & ISQ must look like a single instrument when coupled

59

TRACE 1300 1310 Series