Você está na página 1de 26

Sistema de Inmunoensayo

Access
Introduccin a Redes
Comunicar componentes del hardware a
las funciones de redes
Conocer los principios bsicos de
comunicaciones de redes del sistema
Access 2
Localizar y utilizar los documentos de
servicio del Access 2 especficos de
redes.
Objetivos
Documentos de Servicio Aplicables
Chapter 14B: Software - Access 2
Chapter 16: Troubleshooting, hardware
Troubleshooting, network, Access 2
Identificar PC IP address
Identificar AU IP address
Test PC Network Interface-ping

Glosario
XP (Microsoft Windows Operating System)
RTOS (OS9000 Real Time Operating System)
UI (User Interface)
AU (Analytical Unit)
Instrumento Access sin PC y monitor
HD con OS9000 RTOS
No tiene UI o base de datos
PC (consola)
Access 2 PC
Tiene UI (y quiz base de datos)
Tarjeta Ethernet
Network PCB en Access 2 (NIC)
Hub
Multi-port connection box for Ethernet (10Base-T)
Cable Crossover (cable para un sistema)
Used in older systems to connect PC to AU for single system
configuration (yellow cable), and troubleshooting
Cable de Red
Conecta PC al AU al Hub (azul = 14 feet, gris = 50 feet)
Required for workgroup
Provided, and required for new single system installs, eliminating
need to have PC powered on before AU
Glosario
Tarjeta de Red Access 2
I/O PCB
CPU PCB
Stepper Motor Controller PCB
Ultrasonics PCB
Stepper Motor Driver PCB
Front of Instrument
Power Driver PCB
Ethernet PCB (A2 Only)
Sistema Slo (sistema un solo equipo)
1 Access2 AU/PC par (no en red)


Grupo de Trabajo (Workgroup)
2 - 4 Acess2 AU/PC pares en red

Glosario
Apagado Agradable (Graceful
shutdown)
Cierre de archivos usando tanto la funcin de apagado
de la UI o la funcin de apagado del men de inicio de
Windows antes de apagar la PC.

Si la PC se apag agradablemente
La comunicacin con el AU esta formalmente
terminada.
El AU ya no intentar enviar datos despus de un
apagado suave.
Glosario
Apagado Agradable (Ungraceful
shutdown)
Apagado de la PC sin primero haber cerrado los archivos.
Esto resulta comunmente bajo muchas condiciones en una
base de datos corrompida.

Si la PC es apagada desagradablemente
La comunicacin AU/base de datos es repentinamente
interrumpida en medio del proceso.
Cualquier objeto que estuviera comunicando cuando se
presenta un apagn tiene alta posibilidad de ser
corrompido.
Glosario
Componentes de Comunicacin
AU
RTOS OS9000
PCs
UI uses Windows XP
Uno se configura como servidor al instalarlo
Hay una sola conexin LIS, en el servidor
Tarjetas Ethernet, cables y hub
Permiten al AU y PC interactuar (comunicar)
Permiten capacidades para enlazarse en red
Sistemas mltiples (hasta 4) corriendo independientemente
Todos usando la misma base de datos.
Comunicacin AU/PC Access2
Iniciando una corrida
Requiere comunicacin UI/RTOS
Manteniendo una corrida
Caso normal UI y RTOS comunican y actualizan
En ausencia de UI, RTOS puede completar una corrida
empezada
Transferencia de resultados
Requiere comunicacin UI/RTOS
Cualquiera que se apague el otro no se cae
Cualquiera que se reinicialice el otro no se cae y
no se reinicializa automticamente.
AU
RTOS
PC
UI
Servidor
Viene a ser el servidor durante
la instalacin por tener la virtud
de ser el primero donde se
carga el software
La PC servidor tiene lo
siguiente:
User interface
Ability to set date/time
Backup capability
Global LIS connection
Auto delete features
Database (for workgroup)
Registry
UPS connection
disc
Colorado GB hp
Server
8
Glosario...
Cliente
Any system in the
workgroup that is not the
server
The client PC has the
following
User interface (different)
Registry
Does not contain a database
Surge protector connection
disc
Colorado GB hp
Client
8
Glosario
Comunicacin Crtica
Servidor/Cliente Access2
Server PC supports
its AU and all clients
by providing a
location for the DB

Communication flows
between AU and DB
through UI
Server AU
Server PC
HD

Client AU
Client PC
HD
UI
DB
UI
*Note this is not how the cabling is set up
Access 2 Client UI Crash
If client UI
crashes (without
affecting DB)

Client AU can
continue a run

Server is
unaffected

Reboot client PC
Server AU
Server PC
HD

Client AU
Client PC
HD
UI
UI
UI
DB
*Note this is not how the cabling is set up
X
X
Access 2 Server UI Crash
If server UI crashes
(without affecting the
database.)

Server AU can
continue a run

Client is
unaffected

Reboot all PCs (take
clients down first, bring
up server first)
Server AU
Server PC
HD

Client AU
Client PC
HD
UI
UI
DB
X
UI
DB
*Note this is not how the cabling is set up
X
Access 2 Database Shutdown
If server UI is shut down or
crashes (affects the DB)
then DB is also shutdown

Effects
Client UIs will
crash

All AUs can
continue a run

Reboot all PCs (take clients
down first, bring up server
first)

Server AU
Server PC
HD

Client AU
Client PC
HD
UI
UI
X
DB
X
X
X
*Note this is not how the cabling is set up
IP address (internet protocol address)
Required specific identifier for communication between
the PC and AU
All are in the format 192.168.2.x, where x is the
identifying variable
PC/AU typical pair numbering convention
PC: odd last digit (i.e. 192.168.2.3)
AU: next higher even last digit (i.e. 192.168.2.4)

Registry
System for directing communication between and
among the PCs and AUs
Stand-alone Access 2
Access 2 workgroup
Glosario
IP Address
Cuando el par PC/AU es embarcado de fbrica
(sin software)

PC IP address number set to 192.168.2.1


AU IP address number set to 192.168.2.2

Individual units in a workgroup are given new IP
addresses automatically during software
installation
Exact address depends on the order software is
installed

Server
PC: xxx.xxx.x.3
AU: xxx.xxx.x.4
Client 1
PC: xxx.xxx.x.5
AU: xxx.xxx.x.6
Client 2
PC: xxx.xxx.x.7
AU: xxx.xxx.x.8
Client 3
PC: xxx.xxx.x.9
AU: xxx.xxx.x.10
Numeracn convencional para Grupo de trabajo






Direccines son fijas
Si una PC Servidor falla, un cliente puede ayudar
Base de datos cargada en una PC cliente (respaldo en USB)
The system registry must be modified to direct
communications to the database on the new server


IP Address
Determining IP Address

For PC- Start/run-type cmd, type ipconfig

For AU-Start/run-type cmd, type ping
NOTE: All IP addresses on the hub will be available for response. If
necessary isolate the AU by disconnecting other AU/PCs from the
network



Refer to service manual Ch. 16 Troubleshooting,hardware,
Troubleshooting, network errors, Access 2

Determining PC IP Address
Ping
A command used to check for network response of IP
address
Any PC can ping any other PC and all AUs in the
workgroup
Executed from Command Prompt Window in NT
START/Run, type Cmd, click OK
At C prompt, type Ping 192.168.2.x

Glosario
Ping Satisfactorio

2
4
Ping No Satisfactorio
Otros Comandos/Programas
tiles
Desde UI Access 2
Salir de UI
Presionar [Alt] [F4], introducir password (Access2Support)
Para acceder a Men Inicio sin salir de UI
Utilizar la tecla de Windows en el teclado
O presionar [Ctrl] [Esc]
Para prevenir que la UI del Access 2 se autoinicie
durante el boot up
Mantener presionada la tecla [Ctrl] cuando aparezca la
pantalla aqua y hasta que aparezca el Men Inicio