Você está na página 1de 40

Validation of analytical

methods
Rutendo Kuwana

Training workshop: Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

BACKGROUND-LAB METHOD FLOW

Method
Development

Method
Validation

Method
Transfer

Approved
2|

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

VALIDATION

THE PROCESS OF PROVIDING DOCUMENTED

EVIDENCE THAT SOMETHING DOES WHAT IT IS


INTENDED TO DO

3|

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

WHY VALIDATE?

To demonstrate that the method is suitable for its


intended use

Provides assurance of reliability

4|

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

WHO GMP - 4.11 Analytical methods,


computers and cleaning procedures
It is of critical importance that particular attention is paid to the
validation of analytical test methods, automated systems and
cleaning procedures.
Validation of analytical procedures used in the examination of
pharmaceutical materials (WHO Expert Committee on
Specifications for Pharmaceutical Preparations. 32nd Report.
Geneva, WHO, 1992 (WHO Technical Report Series, No. 823)
Text on Validation of Analytical Procedures Q2 (R1) Validation
of Analytical Procedures: Text and Methodology. ICH
Harmonized Tripartite Guidelines

5|

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

ANALYTICAL METHOD VALIDATION


PERFORMANCE CHARACTERISTICS
P
r
e
c
i
s
i
o
n
A
c
c
u
r
a
c
y
L
i
m
i
t
o
f
D
e
t
e
c
t
i
o
n

M
e
t
h
o
d
V
a
l
i
d
a
t
i
o
n

L
i
m
i
t
o
f
Q
u
a
n
t
i
t
a
t
i
o
n
S
p
e
c
i
f
i
c
i
t
y
L
i
n
e
a
r
i
t
y
R
a
n
g
e
R
o
b
u
s
t
n
e
s
s
S
y
s
t
e
m
S
u
i
t
a
b
i
l
i
t
y

6|

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

TYPES OF ANALYTICAL METHODS TO BE


VALIDATED
Identification tests
Quantitative tests for impurities content
Limit tests for the control of impurities
Quantitative tests of the active in samples of the drug
substance (raw material), finished product or other
selected components in the drug

7|

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

ACCURACY
Expresses the CLOSENESS of
agreement BETWEEN the value,
which is accepted either as a
conventional TRUE VALUE or an
accepted REFERENCE VALUE and
the VALUE FOUND i.e. individual
observation or mean of
measurements
The closeness of test results to the

true value obtained by the method


(trueness).
Established across the range
8|

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

DETERMINATION OF ACCURACY
Drug Substance
Analysis of reference material
Compare results to a second, well-characterized method
Drug Product
Analysis of synthetic mixtures spiked with known quantities of
components
Compare results to a second, well-characterized method
Determined concurrently with precision, linearity and specificity
9|

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

DETERMINATION OF ACCURACY cont


Impurities (Quantitation)
Analysis of samples (Drug substances/Drug product) spiked with known
amounts of impurities
If impurities are not available, see specificity
Recommended Data
Minimum of 9 determinations over a minimum of 3 concentration levels
covering the specified range (e.g. 3 concentrations/3 replicates each)
Reported as % recovery of known added amount or difference between
the mean and true value, with confidence intervals

10 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

10

PRECISION
Precision

The measure of the degree of


agreement (degree of scatter)
among test results when the

method is applied repeatedly


to multiple samplings of a
homogeneous sample
Expressed as %RSD for a
statistically significant number
of samples

11 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

11

Precision (of any process)


Measured mean
Real mean

The precision (VARIABILITY) of an


analytical procedure is usually
expressed as the standard deviation
(S), variance (S2), or coefficient of
variation (= relative standard deviation,
R.S.D.) of a series of measurements.

The confidence interval should be


reported for each type of precision
investigated.

12 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

12

PRECISION cont
May be considered at 3 levels:
Repeatability, a measure of variability under the same
operating conditions over a short interval (intra-assay
precision). Minimum of 9 determinations covering specified
range
Intermediate precision, a measure of within-laboratory
variations (different days, different analysts, different
equipment)
Reproducibility, expresses precision between laboratories (e.g.
in collaborative studies), also applies to method transfer

13 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

13

Repeatability (of any process)


Measured
mean

14 |

Repeatability expresses the


precision (spread of the data,
variability) under the same
operating conditions over a short
interval of time. Repeatability is also
termed intra-assay precision.

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

14

Intermediate Precision and


Reproducibility

Measured
means

Intermediate precision expresses withinlaboratories variations. #1, #2 and #3:


different days, different analysts,
different (manufacturing) equipment,
etc.
Reproducibility expresses the precision
between laboratories #1, #2 and #3
(collaborative studies, usually applied to
standardization of methodology).
(Transfer of technology)

15 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

15

ACCURACY & PRECISION

Inaccurate &
imprecise

Inaccurate but
precise

16 |

Accurate but
imprecise

Accurate and precise

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

16

SPECIFICITY/SELECTIVITY
The ability to measure accurately and

specifically the analyte in the presence of


components that may be expected to be
present in the matrix

The degree of interference


Active Ingredients
Excipients
Impurities (synthetic precursors, enantiomers)
Degradation Products
Placebo Ingredients
17 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

17

SPECIFICITY/SELECTIVITY
Combination of 2 or more analytical procedures may be
required to achieve necessary level of discrimination
Stability indicating analytical methods should always be
specific
Analysts should ascertain whether the peaks within a sample

chromatogram are pure or consist of more than one compound.


Therefore should know how many compounds are in the
sample or use procedures to detect peak purity

18 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

18

LOD, LOQ and SNR

Limit of Quantitation (LOQ)


Limit of Detection (LOD)
Signal to Noise Ratio (SNR)

Peak B
LOQ

Peak A
LOD

Baseline

19 |

noise

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

19

LIMIT OF DETECTION
The lowest concentration of an analyte in a sample that
can be detected, not quantified
Expressed as a concentration at a specified signal:noise
ratio

20 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

20

LIMIT OF QUANTIFICATION
The lowest concentration of analyte in a sample that can
be determined with acceptable precision and accuracy
under stated operational conditions
Expressed as concentration of analyte

21 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

21

LINEARITY
The Ability of the method to obtain test results that are directly
proportional to concentration within a given range
Method: dilution of stock solution/separate weighings

Expressed as the variance of the slope of the regression line


Correlation coefficient, y-intercept, slope of regression line and
residual sum of squares should be presented together with plot of
the data

22 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

22

Linearity
Linearity expresses differences in precision at different points of a
given range.
The linearity of an analytical procedure is its ability (within a given
range) to obtain test results, which are directly proportional to the
concentration (amount) of analyte in the sample

23 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

23

RANGE
Interval between upper and lower levels of analyte demonstrated
by the method
Confirms that the analytical procedure provides acceptable degree

of linearity, accuracy and precision when applied to samples


containing amounts of analyte within or at the extremes of the
specified range
Minimum 5 concentrations

24 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

24

Range (minimum requirements)


Assay of an API or a FPP: 20% of the test concentration.
Content uniformity: 30% of the test concentration (unless a wider more
appropriate range, based on the nature of the dosage form (e.g., metered
dose inhalers), is justified).

Dissolution testing: 20 % over the specified range.


Impurity: from the reporting level of an impurity to 120% of the
specification. (Unusually potent or toxic impurities, LOD and LOQ should be
commensurate with ICH requirement.)

If assay and purity are performed together as one test and only a
100% standard is used, linearity should cover the range from the
reporting level of the impurities to 120% of the assay specification

25 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

25

Typical expected results for Analyte concentration vs. Precision


(The AOAC manual for the Peer-Verified Methods program)

Analyte Conc (%)

Analyte Ratio

Unit

RSD%

100

100%

1.3

10

10-1

10%

2.8

10-2

1%

2.7

0.1

10-3

0.1%

3.7

0.01

10-4

100 ppm

5.3

0.001

10-5

10 ppm

7.3

0.0001

10-6

1 ppm

11

0.00001

10-7

100 ppb

15

0.000001

10-8

10 ppb

21

0.0000001

10-9

1 ppb

30

26 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

RUGGEDNESS
Ruggedness
Degree of reproducibility of test
results under a variety of
conditions

Different Analysts

Different Laboratories

Different Instruments

Different Reagents

Different Days

Etc.

Expressed as %RSD

27 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

27

ROBUSTNESS

Measure of the capacity to


remain unaffected by small
(deliberate) variations in

method parameters
Indication of reliability
during normal use

28 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

28

Sensitivity and robustness

Input-output
relationship

29 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

29

Stability of analytical solution


Solutes may readily decompose prior to chromatographic investigations e.g. during
sample preparation , extraction, cleanup, phase transfer or storage of prepared vials
(refrigerators or automatic sampler). Method development should investigate the
stability of the analytes AND standards.
System stability
stability of the samples being analyzed in a sample solution.
Measure of the bias in assay results generated during a preselected time
interval e.g. 1 48 hours using a single solution
should be determined by replicate analysis of the sample solution.
considered appropriate when the RSD, calculated on the assay results obtained
at different time intervals, Less than 20 percent of the corresponding value of
the system precision
30 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

30

SYSTEM SUITABILITY
The checking of a system, before or during analysis of unknowns, to
ensure system performance.
No sample analysis is acceptable unless the requirements for system
suitability have been met. (USP Chapter 621)
Plate Count, Tailing, Resolution
Determination of reproducibility (%RSD)
For %RSD < 2.0%, Five replicates
For %RSD > 2.0%, Six replicates
System Suitability "Sample - A mixture of main components and
expected by-products utilized to determine system suitability
Whenever There is a Significant change in Equipment or Reagents
System Suitability Testing Should be Performed (USP Chapter 621)

31 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

31

Classes of analytical tests


The objective of validation of an analytical procedure is to
demonstrate that it is suitable for its intented purpose.
Class A: To establish identity

Class B: To detect (Bd) and quantitate (Bq) impurities


Class C: To determine quantitatively the concentration, or assay
Class D: To assess characteristics
Other classes not covered in the guides

32 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

32

Criteria for analytical classes


Criteria

Bq

Accuracy

Precision

Robustness

Linearity and range


Specificity

X
X?

Limit of detection
Limit of quantitation
33 |

Bd

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

X
X

33

General requirements
Qualified and calibrated instruments
Documented methods
Reliable reference standards
Qualified analysts
Sample integrity
Change control (e.g., synthesis, FPP composition)
34 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

34

General requirements (2)

Analytical methods should be used within GMP and GLP


environments, and must be developed using the protocols
and acceptance criteria set out in the ICH guidelines Q2
(R1)

35 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

35

General requirements (3)


Validation Protocol important
Revalidation should accompany

36 |

formulation changes (new samples with new compounds or new matrices)

manufacturing batch changes

new analysts with different skills,

new instruments with different characteristics,

new location with different environmental conditions,

new chemicals and/or reference standards and

modification of analytical parameters.

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

36

Validation Report

Objective and scope of the method (applicability, type).

Summary of methodology.

Type of compounds and matrix.

All chemicals, reagents, reference standards, QC samples with purity, grade, their source
or detailed instructions on their preparation.

Procedures for quality checks of standards and chemicals used.

Method parameters.

Critical parameters taken from robustness testing.

Listing of equipment and its functional and performance requirements, e.g., cell
dimensions, baseline noise and column temperature range.

37 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

Validation Report (2)

Detailed conditions on conduct of experiments, including sample preparation

Statistical procedures and representative calculations.

Procedures for QC in routine analyses, e.g., system suitability tests.

Representative plots, e.g., chromatograms, spectra and calibration curves.

Method acceptance limit performance data and expected uncertainty of


measurement results.

Criteria for revalidation.

The person(s) who developed and validated the method.


References (if any).
Summary and conclusions.
38 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

Use and Validation of Pharmacopoeial methods


The degree of validation of a pharmacopoeial method should be adequate for
required purpose, and the laboratory be able to match any stated performance data.
Following product specific attributes should be considered
the type of compounds to be analyzed,
matrices,
the type of information required (qualitative or quantitative),
detection and quantitation limits,
Concentration range for analysis based on own product

precision and accuracy (sample preparation critical) as specified by the client of the
analytical data and
the type of equipmentits location and environmental conditions.

Therefore PQP three parameters required for PQP Specificity, Accuracy and
Precision
39 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

Main Points Again


Validation of analytical procedures is a critical
requirement in risk assessment and management:
establishment of product-specific acceptance criteria, and
stability of APIs and FPPs.

Validation should demonstrate that the analytical


procedure is suitable for its intented purpose.
HPLC systems and method validation deserves special
attention during the inspection of QC laboratories.
40 |

Assessment of Interchangeable Multisource Medicines, Kenya, August 2009

40

Você também pode gostar