Escolar Documentos
Profissional Documentos
Cultura Documentos
DATE: SIGNATURE:
ACKNOWLEDGMENT
LABELLING METHODS
Indirect labelling of probes involves the use of
compounds such as biotin as the primary labels and later streptavidins, which are used as
a conjugate for signal generation system. Otherwise, antibody of a hapten is incorporated
into the probe, the recognition of which leads to detection. Similarly, photobiotin is
detected either through luminescence or fluorescence or alkaline phosphatase conjugate
– colorimetrically with enzyme conjugates. In general, the indirect methods use probes
tagged with biotin, digoxigenin and dinitrophenol (DNP) as reporter molecules, which are
detected by fluorochrome-conjugated avidin or antibodies. The common fluorochromes
are FITC (fluorescin-isothiocyanate), rhodamine and AMCA (amino-r-methyl coumarin-
3acetic acid). Isomer of biotin (mol. wt. 244.31) is a natural vitamin of the animal cell,
pecially of liver, kidney and pancreas. It has a few chromophoric groups .
Advin is whereas a tetrametic group animal glyco-protein and all four tetramers have four
biotin binding sites. The avidin-biotin interaction is the strongest noncovalent biological
recognition so far known between ligand and protein. The reaction is rather rapid; not
affected by pH, solvents or denaturing agents and can withstand short exposure to high
temperatures more than 100 °C. Avidin denaturation may, however, affect the binding
capacity [15]. The long immunological detection method can be avoided by using
fluorescin-12-dUTP instead of dinitrophenol as reporter. The direct method on the other
hand, is characterized by probe labelling with fluorochrome-labelled antibodies. In direct
labelling, the signal generation system is directly attached to the probe which is detected
post-hybridization with enzyme and enzyme conjugate. The examples include horse
radish peroxidase where the detection is through high chemo luminescence. Such direct
coupling of fluorochromes to probes does not need detection through
immunocytochemical methods. The direct labelling of fluorochrome into the probe is
rapid and ensures good resolution. Probes for single, large and repetitive resolution
To find out the location of a specific gene located on a specific chromosome: FISH
hybridization is done with the appropriate gene-specific probe labelled with the
fluorescent dye. The test will give the binding of the gene-specific probe labelled with
the fluorescent dye to the respective chromosome at the specific position, where the
gene is located.
Detection of translocation: Chromosomes that have undergone translocation will have
two segments. When it is subjected to the technique of chromosome painting it will
take different probes and appear in two colours or multicoloured depending on the
number of translocations.
Detecting chromosome abnormalities: FISH has improved the efficiency of
screening cells for chromosome abnormalities in mutagenic studies and for testing
the mutagenic ability of chemicals and other potent mutagens in the environment.
It has also improved the detection of chromosome aberrations and rearrangements
associated with tumour and cancer.
PROCEDURE SAMPLE PREPARATION
AND HYBRIDISATION
DISADVANTAGES
oRestricted to those abnormalities hat can be detected with currently avliabke probe
oOnly one or a few abnormalities can be assessed simultaneously
oDue to failure to detect signal FISH is higher sensetive for trisomy but less sensitive for
detecting chromosome loss or deletion
oRequires flourescence microscopy and an image analysis system
Scanning electron microscope
A scanning electron microscope (SEM) is a type of electron microscope that produces
images of a sample by scanning the surface with a focused beam of electrons. The
electrons interact with atoms in the sample, producing various signals that contain
information about the sample's surface topography and composition. The electron beam is
scanned in a raster scan pattern, and the beam's position is combined with the detected
signal to produce an image. SEM can achieve resolution better than 1 nanometer.
Specimens can be observed in high vacuum in conventional SEM, or in low vacuum or wet
conditions in variable pressure or environmental SEM, and at a wide range of cryogenic or
elevated temperatures with specialized instruments.
The most common SEM mode is detection of secondary electrons emitted by atoms
excited by the electron beam. The number of secondary electrons that can be detected
depends, among other things, on specimen topography. By scanning the sample and
collecting the secondary electrons that are emitted using a special detector, an image
displaying the topography of the surface is created
History
An account of the early history of SEM has been presented by McMullan. Although Max
Knoll produced a photo with a 50 mm object-field-width showing channeling contrast by
the use of an electron beam scanner, it was Manfred von Ardenne who in 1937 invented a
true microscope with high magnification by scanning a very small raster with a demagnified
and finely focused electron beam.
Ardenne applied the scanning principle not only to achieve magnification but also to
purposefully eliminate the chromatic aberration otherwise inherent in the electron
microscope.
He further discussed the various detection modes, possibilities and theory of SEM, together
with the construction of the first high magnification SEM.
Further work was reported by Zworykin's group, followed by the Cambridge groups in the
1950s and early 1960s headed by Charles Oatley, all of which finally led to the marketing of
the first commercial instrument by Cambridge Scientific Instrument Company as the
"Stereoscan" in 1965, which was delivered to DuPont.
Principles and capacities
The signals used by a scanning electron microscope to produce an image result from
interactions of the electron beam with atoms at various depths within the sample.
Various types of signals are produced including secondary electrons(SE), reflected
or back-scattered electrons (BSE), characteristic X-rays and light
(cathodoluminescence) (CL), absorbed current (specimen current) and transmitted
electrons. Secondary electron detectors are standard equipment in all SEMs, but it is
rare that a single machine would have detectors for all other possible signals.
In secondary electron imaging, or SEI, the secondary electrons are emitted from very
close to the specimen surface. Consequently, SEM can produce very high-resolution
images of a sample surface, revealing details less than 1 nm in size. Back-scattered
electrons (BSE) are beam electrons that are reflected from the sample by elastic
scattering. They emerge from deeper locations within the specimen and consequently
the resolution of BSE images is less than SE images. However, BSE are often used in
analytical SEM along with the spectra made from the characteristic X-rays, because the
intensity of the BSE signal is strongly related to the atomic number (Z) of the specimen.
BSE images can provide information about the distribution of different elements in the
sample. For the same reason, BSE imaging can image colloidal gold immuno-labels of 5
or 10 nm diameter, which would otherwise be difficult or impossible to detect in
secondary electron images in biological specimens.
Sample preparation
A spider sputter-coated in gold, having been prepared for viewing with an SEM.
Low-voltage micrograph (300 V) of distribution of adhesive droplets on a Post-it note. No
conductive coating was applied: such a coating would alter this fragile specimen.
Samples for SEM have to be prepared to withstand the vacuum conditions and high energy
beam of electrons, and have to be of a size that will fit on the specimen stage. Samples are
generally mounted rigidly to a specimen holder or stub using a conductive adhesive. SEM is
used extensively for defect analysis' of semiconductor wafers, and manufacturers make
instruments that can examine any part of a 300 mm semiconductor wafer. Many
instruments have chambers that can tilt an object of that size to 45° and provide
continuous 360° rotation.
Nonconductive specimens collect charge when scanned by the electron beam, and especially
in secondary electron imaging mode, this causes scanning faults and other image artifacts
For conventional imaging in the SEM, specimens must be electrically conductive, at least at the
surface, and electrically grounded to prevent the accumulation of electrostatic charge. Metal
objects require little special preparation for SEM except for cleaning and conductively
mounting to a specimen
Nonconducting specimens may be imaged without coating using an environmental SEM
(ESEM) or low-voltage mode of SEM operation. In ESEM instruments the specimen is
placed in a relatively high-pressure chamber and the electron optical column is
differentially pumped to keep vacuum adequately low at the electron gun. The high-
pressure region around the sample in the ESEM neutralizes charge and provides an
amplification of the secondary electron signal.] Low-voltage SEM is typically conducted in
an FEG-SEM because field emission guns(FEG) are capable of producing high primary
electron brightness and small spot size even at low accelerating potentials. To prevent
charging of non-conductive specimens, operating conditions must be adjusted such that
the incoming beam current is equal to sum of out coming secondary and backscattered
electrons currents a condition that is more often met at accelerating voltages of 0.3–4 kV.
Synthetic replicas can be made to avoid the use of original samples when they are not
suitable or available for SEM examination due to methodological obstacles or legal issues.
This technique is achieved in two steps: (1) a mold of the original surface is made using a
silicone-based dental elastomer, and (2) a replica of the original surface is obtained by
pouring a synthetic resin into the mold.
Embedding in a resin with further polishing to a mirror-like finish can be used for both
biological and materials specimens when imaging in backscattered electrons or when
doing quantitative X-ray microanalysis.
The main preparation techniques are not required in the environmental SEM outlined
below, but some biological specimens can benefit from fixation.
Biological samples
• For SEM, a specimen is normally required to be completely dry, since the
specimen chamber is at high vacuum. Hard, dry materials such as wood, bone,
feathers, dried insects, or shells (including egg shells[) can be examined with little
further treatment but living cells and tissues and whole, soft-bodied organisms
require chemical fixation to preserve and stabilize their structure.
• Fixation is usually performed by incubation in a solution of a buffered chemical
fixative, such as glutaraldehyde, sometimes in combination
with formaldehyde and other fixatives,[23and optionally followed by postfixation
with osmium tetroxide. The fixed tissue is then dehydrated. Because air-drying
causes collapse and shrinkage, this is commonly achieved by replacement
of water in the cells with organic solvents such as ethanol or acetone, and
replacement of these solvents in turn with a transitional fluid such as
liquid carbon dioxide by critical point drying. The carbon dioxide is finally removed
while in a supercritical state, so that no gas–liquid interface is present within the
sample during drying.
• The dry specimen is usually mounted on a specimen stub using an adhesive such
as epoxy resin or electrically conductive double-sided adhesive tape, and sputter-
coated with gold or gold/palladium alloy before examination in the microscope.
Samples may be sectioned (with a microtome) if information about the organism's
internal ultra structure is to be exposed for imaging.
• If the SEM is equipped with a cold stage for cryo microscopy, cryofixation may be
used and low-temperature scanning electron microscopy performed on the
cryogenically fixed specimens.[ Cryo-fixed specimens may be cryo-fractured under
vacuum in a special apparatus to reveal internal structure, sputter-coated and
transferred onto the SEM cryo-stage while still frozen.
SEM Properties
The Scanning Electron Microscope developed by professor Dr. Charles Oatlev with
the assistance of graduate students in the 1950s, are one of the three types of
electron microscopes (EM).
Electron microscopes utilize the same basic principles as light microscopes, but
focus beams of energetic electrons rather than photons, to magnify an object.
SEMs consist of the following components:
Electron Source
Thermionic Gun
Field Emission Gun
Electromagnetic and/or Electrostatic Lenses
Vacuum chamber
Sample chamber and stage
Computer
Detectors (one or more)
Secondary Electron Detector (SED)
Backscatter Detector
Diffracted Backscatter Detector (EBSD)
X-ray Detector (EDS)
In addition, SEMs require a stable power supply, vacuum and cooling system,
vibration-free space and need to be housed in an area that isolates the instrument
from ambient magnetic and electric fields.
SEM Imaging
A Scanning Electron Microscope provides details surface information by tracing a sample in
a raster pattern with an electron beam.
The process begins with an electron gun generating a beam of energetic electrons down
the column and onto a series of electromagnetic lenses.
These lenses are tubes, wrapped in coil and referred to as solenoids.
The coils are adjusted to focus the incident electron beam onto the sample; these
adjustments cause fluctuations in the voltage, increasing/decreasing the speed in which
the electrons come in contact with the specimen surface.
Controlled via computer, the SEM operator can adjust the beam to control magnification as
well as determine the surface area to be scanned.
The beam is focused onto the stage, where a solid sample is placed. Most samples require
some preparation before being placed in the vacuum chamber.
Of the variety of different preparation processes, the two most commonly used prior to
SEM analysis are sputter coating for non-conductive samples and dehydration of most
biological specimens.
In addition, all samples need to be able to handle the low pressure inside the vacuum
chamber.
SEM Applications
SEMs have a variety of applications in a number of scientific and industry-related
fields, especially where characterizations of solid materials is beneficial.
SEM Advantages
Advantages of a Scanning Electron Microscope include its wide-array of applications,
the detailed three-dimensional and topographical imaging and the versatile
information garnered from different detectors.
SEMs are also easy to operate with the proper training and advances in computer
technology and associated software make operation user-friendly.
This instrument works fast, often completing SEI, BSE and EDS analyses in less than five
minutes. In addition, the technological advances in modern SEMs allow for the generation
of data in digital form.
Although all samples must be prepared before placed in the vacuum chamber, most SEM
samples require minimal preparation actions.
SEM Disadvantages
The disadvantages of a Scanning Electron Microscope start with the size and cost.
SEMs are expensive, large and must be housed in an area free of any possible electric,
magnetic or vibration interference.
Maintenance involves keeping a steady voltage, currents to electromagnetic coils and
circulation of cool water.
Special training is required to operate an SEM as well as prepare samples.
The preparation of samples can result in artifacts. The negative impact can be
minimized with knowledgeable experience researchers being able to identify artifacts
from actual data as well as preparation skill .
Transmission electron microscope
A TEM image of a cluster of poliovirus. The polio virus is 30 nm in diameter
DISADVANTAGES
Some cons of electron microscopes include:
TEMs are large and very expensive.
Laborious sample preparation.
Potential artifacts from sample preparation.
Operation and analysis requires special training.
Samples are limited to those that are electron transparent, able to tolerate the vacuum
chamber and small enough to fit in the chamber.
TEMs require special housing and maintenance.
Images are black and white.
G-banding, G banding, or Giemsa banding is a technique used in cytogenetics to produce a
visible karyotype by staining condensed chromosomes. It is useful for identifying genetic
diseases through the photographic representation of the entire chromosome
complement. The metaphasechromosomes are treated with trypsin(to partially digest the
chromosome) and stained with Giemsa stain. Heterochromatic regions, which tend to be
rich with adenine and thymine (AT-rich) DNA and relatively gene-poor, stain more darkly in
G-banding. In contrast, less condensed chromatin—which tends to be rich
with guanine and cytosine (GC-rich) and more transcriptionally active—incorporates less
Giemsa stain, and these regions appear as light bands in G-banding. The pattern of bands
are numbered on each arm of the chromosome from the centromere to the telomere. This
numbering system allows any band on the chromosome to be identified and described
precisely. The reverse of G-bands is obtained in R-banding. Banding can be used to identify
chromosomal abnormalities, such as translocations, because there is a unique pattern of
light and dark bands for each chromosome.
It is difficult to identify and group chromosomes based on simple staining because the
uniform colour of the structures makes it difficult to differentiate between the different
chromosomes. Therefore, techniques like G-banding were developed that made "bands"
appear on the chromosomes. These bands were the same in appearance on
the homologous chromosomes, thus, identification became easier and more accurate. The
less condensed the chromosomes are, the more bands that appear when G-banding. This
means that the different chromosomes are more distinct in prophase than they are in
metaphase acid-saline-Giemsa protocol reveals G-banding.
Chromosome Banding: Technique # 2. G-Bands:
The technique of G-banding involves Giemsa staining following pre-treatment with weak
trypsin solution, urea or protease. It provides greater detail than C-banding. It was first used
for human chromosomes by Summer et al. in 1971. G-bands may reflect a stronger chromatin
condensation. However, this technique is not suitable for plant chromosomes.
Chromosome Banding: Technique # 3. Q-Bands:
The method of Q-banding was developed by Caspersson et al. in 1968. The chromosomes
stained with Quinacrine mustard show bright and dark zones under UV light. This technique is
used to identify human and mice chromosomes.
Chromosome Banding: Technique # 4. N-Bands:
The technique of N-banding was originally described by Matsui and Sasaki in 1973. Briefly, air-
dried chromosomes slides are stained for 90 minutes with Giemsa (diluted 1 : 10 in 1/15 M
phosphate buffer at pH 7.0) following extraction with 5% trichloroacetic acid at 95°C for 30
minutes and then 0.1 NHCl at 60″C for 30 minutes.
The N-bands are generally located at the secondary constriction, satellites, centromeres,
telomeres and heterochromatic segments. It is suggested that the N-bands represent certain
structural non-histone proteins specifically linked to the nucleolar organizer region of the
eukaryotic chromosomes.
The N- banding patterns have been used for the location of nucleolar regions in the different
organisms, such as, mammals, birds, amphibians, fishes, insects and plants. N-banding
patterns differ in the chromosomes of different species.
In 1980, Islam used this method to identify the barley chromosomes from those of wheat in
the reciprocal wheat-barely F, hybrids, and to detect translocations between the wheat and
barley chromosomes. He also used this technique to isolate lines possessing a pair of barely
chromosomes substituted for particular pair of wheat chromosomes.
A modified Giemsa-N-banding technique was developed by Singh and Tsuchiya in 1982 for
the identification of barley chromosomes. This method is a combination of acetocarmine
staining and Giemsa-N-banding. After processing according to this method, the centromeric
region looks like a “diamond-shaped” structure; this is not seen in other techniques.
Early metaphase or prometaphase chromosomes are more suitable for this staining as they
show better banding pattern than the chromosomes at mid-metaphase in somatic cells.
Chromosome Banding: Technique # 5. Other Techniques of
Chromosome Banding:
Besides the above, there are other techniques for chromosome banding, e.g., R-banding
(Reverse Giemsa banding). H-banding, and T-banding (Terminal banding). Chromosome
banding patterns can be used not only for the identification of individual chromosomes of an
organism but also to establish evolutionary relationships between different species.
Banding patterns in human, chimpanzee, gorilla and orangutan have indicated that the
evolutionary relationship between human and chimpanzee is closer than that between
human and gorilla. It has further indicated that humans have a more distant evolutionary
relationship with orangutans.
Banding type Staining method
Q-banding quinacrine
T-banding telomeric
patterns lend each chromosome a distinctive appearance so the 22 pairs of human nonsex
chromosomes and the X and Y chromosomes can be identified and distinguished without
ambigchromosome banding
Chromosome banding: The treatment of chromosomes to reveal characteristic patterns of
horizontal bands like bar codes.
The banding uity. Banding also permits the recognition of chromosome deletions (lost
segments), chromosome duplications (surplus segments) and other types of structural
rearrangements of chromosomes.
The distinctive pattern, visible in a light microscope, that results from the selective binding of
certain dyes to individual chromosomes.
Banding patterns are patterns of light and dark transverse bands on chromosomes. The light
and dark bands become apparent by staining the chromosome with a chemical solution and
then viewed under a microscope. These bands describe the location of genes on a
chromosome.
Chromosomes in metaphase can be identified using certain staining techniques, so called
banding. Cells are cultured and then stopped in metaphase to maximize the number of suitable
cells. They are then spread on a slide, stained with a suitable dye and visualized in the
microscope. Most conventional cytogenetic analyses depend on the karyotyping of banded
metaphase chromosomes.
The banding techniques fall into two principal groups: 1) those resulting in bands distributed
along the length of the whole chromosome, such as G-, Q- and R-bands and 2) those that stain
a restricted number of specific bands or structures. These latter include methods which
reveal centromeric bands, C-bands, and nucleolus organizer regions, NOR's (at terminal
regions of acrocentric chromosomes). C-banding methods do not permit identification of every
chromosome in the somatic cell complement, but can be used to identify specific
chromosomes.
G- and R- bands can be bright field or fluorescent.
Bright field G-bands
These G-bands are most commonly used. They take their name from the Giemsa
dye, but can be produced with other dyes. In G-bands, the dark regions tend to
be heterochromatic, late-replicating and AT rich. The bright regions tend to
be euchromatic, early-replicating and GC rich.
Bright field R-bands
These R-bands are approximately the reverse of G-bands (the R stands for "reverse").
The dark regions are euchromatic and the bright regions are heterochromatic.
Q-bands are like fluorescent G-bands, but certain heterochromatic regions are
more brightly stained with Q-banding.
CONCLUSION
After completing this project I gained a lot of knowledge on
the topics sem tem chromosome painting and banding.
This project gave me a lot of knowledge on terms I have never
heard. I came to know about its advantages disadvantages procedure and my more.
I hope this project will be liked and all mistakes will be acknowledged.
REFRENCES