How MV Works – Volume Displacement

2
The Four Phases of Breath

1 2 3 4 Phases of Breath

Flow

Time

Triggering Inspiration Cycling Expiration Phase Variables


• Manual • Volume • Volume • PEEP
• Machine • Pressure • Pressure
• Patient • Time
• Pressure • Flow
• Flow

3
Inspiratory Trigger

• Manual Button
User Inisiatif (Begging) / Inspiratory Hold

• Machine Inisiatif
RR Setting, Periodically every 60/RR second
(Ex : RR=10, Triggered Periodically ever y 6 second)

• Patient Inisiatif
Trigger Sensitivity Setting : Pressure or Flow Trigger
(Ventilator Sensitivity in detecting Patient’s Triggers)

4
Inspiratory Trigger
Patient Trigger  Pressure vs Flow Trigger
• P-trigger (minus, cmH2O)
• Sensitive, but may be affected by noise (self-cycling)
• Any associated base-flow, intrinsic PEEP or Deadspace
worsens the performance (Increased WOB in triggering)

• F-trigger (liter/min)
• More Sensitive, and only rarely affected by noise
• Any associated base-flow improves the performance
but need Leakage Identification and Compensation

5
Inspiration

• Volume Breath
• Tidal Volume (Square Flow x Ti)

• Pressure Breath
• Pressure Inspirasi (Nilai diatas 0) atau
• Pressure Control (Pinsp = PEEP + PC) atau
• Pressure Support (PASB = PEEP + PS)

6
Cycling

Perpindahan dari fase inspirasi ke ekspirasi.


Katup ekspirasi terbuka sehingga terjadi proses ekspirasi.

• Volume (Flow x Ti)


• Pressure
• Time (Ti)
• Flow

7
Expiration

Pengaturan pada saat akhir ekspirasi:


a. Tanpa tekanan, berarti pada akhir ekspirasi,
alveoli hampir colaps
b. Dengan tekanan (PEEP), berarti pada akhir ekspirasi,
alveoli tetap mengembang, karena ada tekanan udara

8
9
Patient’s Breath
(from Apnea to Spontaneous Breath)

10
Ventilation Modes/Parameter following
the Patient’s Spontaneous Breath

11
Controlled Ventilation Modes

Full Controlled Modes :


Trigger Sensitivity = OFF  All patient’s Trigger unaccepted

Assist Controlled Modes :


Trigger Sensitivity = SET  All patient’s Trigger will be assisted with Controlled Breath (VT or Pinsp)

12
Intermittent Ventilation Modes

Intermittent Modes :
Trigger Sensitivity = SET  All patient’s Trigger will be Supported with Pressure Support (PS) Breath

Intermittent Modes is combination of Controlled Breath (VT or Pinsp) and Supported Breath (PS)

13
Spontaneous Ventilation Modes

Spontaneous Modes :
No Controlled Breath (VT or Pinsp) will be given
Trigger Sensitivity = SET  All patient’s Trigger will be Supported with Pressure Support

Continuous Positive Airway Pressure (CPAP) Modes :


Maintain PEEP as CPAP Level while No Support on any patient’s Triggers

14
Ventilation Graph
Perbandingan Mode Full Control, Assist Control, Intermittent & CPAP/ASB

Full Control Mode (IPPV )


Example :
PEEP=5cmH2O, VT=500 ml,
Ti = 2 sec, RR=10 bpm (TE=4 sec)
 I:E Ratio = TI:TE = 1:2
Trigger = Off  Tidak ada Trigger Detection

Assist Control Mode (IPPV Assist)


Example :
PEEP=5cmH2O, VT = 500 ml,
Ti = 2 sec, RR=10 bpm (TE=4 sec)
 I:E Ratio = TI:TE = 1:2
Trigger = 5 lpm (Set)  Sensitivitas Trigger

Intermittent Mode (SIMV or BIPAP/ASB)


Example :
PEEP=5cmH2O, PC = 20 cmH2O,
Ti = 2 sec, RR = 6 bpm (TE=8 sec)
Trigger = 5 lpm (Set)  Sensitivitas Trigger
PS = 12 cmH2O

Spontaneous Mode (CPAP/ASB)


Example :
PEEP=5cmH2O
PS = 12cmH2O

15
Ventilation Setting & Graph

F (Rate) = 60 / (Ti + Te) TI determine Inspiratory Time on Mandatory


and Assisted Breathing.
TE determine Expiratory Time on Mandatory
Breathing or Patient Triggers Window on
Assisted & Supported Breathing.

OR

16
Ventilation Graph
Mandatory & Assisted Breathing Comparison

Mandatory Breath Assisted Breath


Inspiratory Trigger : Ventilator Rate=60/(Ti+Te) Patient Trigger
Inspiratory Breath : Pinsp Pinsp
Inspiratory Time : TI TI
Expiratory Time: TE TE=Patient Trigger Window
(If there is no Trigger then
Mandatory Breathing will be given)

17
Ventilation Graph
Assisted & Supported Breathing Comparison

Assisted Breath Supported Breath


Inspiratory Trigger : Patient Trigger Patient Trigger
Inspiratory Breath : Pinsp Pressure Support (PASB-PEEP)
Inspiratory Time : TI Expiratory Trigger (15%PeakFlow)
Expiratory Time: TE=Patient Trigger Window Patient Trigger Window
(If patient is Apnea then Mandatory Breathing will be given)

18
Ventilation Graph
Perbandingan Mode Full Control, Assist Control, Intermittent & CPAP/ASB

Full Control Mode (IPPV )


Example :
PEEP=5cmH2O, VT=500 ml,
Ti = 2 sec, RR=10 bpm (TE=4 sec)
 I:E Ratio = TI:TE = 1:2
Trigger = Off  Tidak ada Trigger Detection

Assist Control Mode (IPPV Assist)


Example :
PEEP=5cmH2O, VT = 500 ml,
Ti = 2 sec, RR=10 bpm (TE=4 sec)
 I:E Ratio = TI:TE = 1:2
Trigger = 5 lpm (Set)  Sensitivitas Trigger

Intermittent Mode (SIMV or BIPAP/ASB)


Example :
PEEP=5cmH2O, PC = 20 cmH2O,
Ti = 2 sec, RR = 6 bpm (TE=8 sec)
Trigger = 5 lpm (Set)  Sensitivitas Trigger
PS = 12 cmH2O

Spontaneous Mode (CPAP/ASB)


Example :
PEEP=5cmH2O
PS = 12cmH2O

19
Draeger Ventilation Modes

Evita 4 Edition Pressure Volume


Controlled Modes Full PCV+ CMV
(Trigger=HIGH, Pinsp) (Trigger=OFF)
Assist PCV+ CMV
(Trigger=SET, PEEP+PS=Pinsp) (Trigger=SET, VT)
Intermittent Modes PCV+ SIMV
(Trigger=SET, PEEP+PS < Pinsp) (Trigger=SET, VT & PS)
Spontaneous Modes Spont Psupp
(Trigger=SET, PS=SET)
CPAP Psupp
(Trigger=SET, PS=0)

20
Waveform

Rectangular Sinusoidal Ascending Ramp


(Square)

Descending Ramp Exponential (rise) Exponential (decay)

21
22
23
24
Resistance & Compliance
AIRWAY

LUNG COMPLIANCE (COMPL)

AIRWAY RESISTANCE (RAW) 1. Rasio Perubahan Volume akibat terjadinya Perubahan


Pressure  ∆V/∆P
2. Kategori areanya
1. Obstruksi Jalan Nafas (Membatasi Aliran Gas) - Compl Paru (Edema Paru, Fibrosis)
2. Flow = Pressure / Resistance - Compl Dinding Dada (Obesitas, Distensi Abdomen)
(Jika Resistensi Meningkat maka Flow turun)
3. Kategori nilainya
3. Ditentukan oleh besarnya diameter jalan nafas - Low Compliance (Volume kecil pada Pressure besar
4. Pada nafas spontan, jika Resistance meningkat  ARDS, Edema Paru, Pneumonia berat,
maka response tubuh adalah meningkatkan usaha Hematopneumotoraks, Efusi Pleura, Abdominal
nafas (WOB dan RR meningkat) Pressure
- High Compliance (Volume besar pada Pressure
kecil)  COPD, Open Chest, Muscle relaxant
ETT
Terlalu kecil V
Tumor
Pressure
Bronkospasme /Sekret Peningkatan Penurunan
Peak

Resistensi Compliance
Pressure

Penurunan Peningkatan
Peak

Resistensi Compliance
Kolaps/
Atelektasis P

25
Compliance – Pressure & Volume Correlation

Pressure Control Volume Control


Target Pemberian Nafas adalah Pressure Target Pemberian Nafas adalah Volume
(Pressure Konstan) tapi Volume yang dihasilkan (Volume Konstan) tapi Pressure yang dihasilkan
dapat berubah-ubah sesuai perubahan dapat berubah-ubah sesuai perubahan
Compliance Paru pasien. Compliance Paru pasien.
Antisipasi Volutrauma dengan Alarm Tidal Volume Antisipasi Barotrauma dgn Alarm Peak Pressure
V
V

T T
P P

T T
Peningkatan Compliance
Peningkatan Volume pd Pressure yg sama
Penurunan Pressure pd Volume yg sama

Penurunan Compliance
Penurunan Volume pd Pressure yg sama
Peningkatan Pressure pd Volume yg sama
26
Free Breathing with PCV+ (BIPAP) and AutoFlow®

27
CPAP Concept :
Free Breathing on PEEP/CPAP Level

28
Inspiratory Phase on
Conventional PCV VS PCV+ (BIPAP)

29
Conventional PCV VS PCV+ (BIPAP)

30
Pressure & Volume Modes

Pressure Modes Volume Modes


Control/Assist PCV+ (BIPAP) CMV with AutoFlow
(=BIPAP with Volume Target)
Intermittent PCV+ (BIPAP) with PS SIMV with AutoFlow
(=BIPAP/ASB with Volume Target)
Spontaneous Psupp

31
Dual Control Modes
(Conventional)

32
Volume Modes with AutoFlow

33