Technical seminar
Disassembly
& Assembly
Celltac α
Disassembly
(1) Before disassembling
(2) Covers
(3) Units
(4) Exchange of the Electro
Magnetic Valve
(5) Practice time
Celltac α
(1) Before disassembling
CAUTION
・Be careful not to directly touch any place where there is blood or it might
spread to.
・Wear rubber gloves to protect yourself from infection before doing
maintenance.
WARNING
・For your safety, before disassembling the hematology analyzer, wait for
approximately 1 minutes after turning off the main power switch always.
Celltac α
(1) Before disassembling
Before disassembly procedure,
perform strong cleaning with CLEANAC3
detergent!!
Purpose is protection from infection
and removing some proteins.
Celltac α
(1) Before disassembling
~ STRONG CLEAN~ ~ DRAIN ALL ~
Contents
・ Covers
(2) Covers
NO!
NO!
or
PV-640VK MS-640V
FOR MEK-6400
MP-640V
MS-641V MC-640V
or MR-640V
Disassembly
PV-641VK
FOR MEK-6410/20
or
PV-640VK
FOR MEK-6400
(3) Units
• Removing the Display Unit
MC-640V
(3) Units
MS-640V
(3) Units
or
MR-640V
FOR MEK-6410/20
Adjusting the position of
Sampler Unit and Rinse Unit
Select “UNIT MAINTE”.
Select “MS-640V”.
If this case is
OPEN Rinse,
select “INITIALIZE”.
Confirm position
of Sample tube
and Rinse unit
by hand.
Adjusting the position of Sampler
Unit and Rinse Unit
Confirm position by
using OPEN Rinse or
OPEN Rinse CAP rinse.
CAP Rinse
good
The tube
is fixed
It would be the
cause of false
bad operation!!
Be careful!
(3) Units
MS-641V
or MR-640V
(3) Units
Hook the
three parts
(4)Exchange of the Electro Magnetic Valve
How to replace
New electro magnetic valve
are 2 type
•XP-602V for MC-640V only
No5 and No11
•XP-612V for others
Difference point is inside
rubber material
If you use XP-612V for MC-
640V (No5 or No11) then
PLT NOISE alarm will
appear
(4)Exchange of the Electro Magnetic Valve
Removing
Remove 2 screws
(4)Exchange of the Electro Magnetic Valve
Removing
Be careful!
(4)Exchange of the Electro Magnetic Valve
Replace new one
XP-602V / 612V must be
connected as these pictures
1,2,3,・・・
Replace
・Attention: Don’t miss inserting
the Electro Magnetic Valve!!
NG!
OK
(4)Exchange of the Electro Magnetic Valve
In case of exchange valves of MC-640V
XP-602V No.5
XP-602V No.11
MEK-6400 Hydraulic System Diagram (Real version)
(5)Practice time
① Covers
② Units (except UT-7192,7193)