1.Brief Introduction
2.Working principle
3.Composition of instrument
1. Brief Introduction
Some main technical date :
Instrument structure: Discrete system
Throughput: Constant speed, 400 tests/ hour (800 tests/ hour with ISE)
Sample volume:2 to 35μl(Stepping 0.1μl)
Reagent volume: 20 to 350μl,(Stepping 1μl)
Reaction solution volume:150~450μl
115 sample positions, double disks with total 90 reagent position
Integration of sample probe and reagent probe with touch sensor and
sample probe block test function.
20W /12V Long life quartz halogen lamp (water cooling)
120 pcs of reusable rigid optical plastic reaction cuvette.
Reaction cuvette optical diameter: 6mm
LIS/HIS interface, supports remote maintenance
2.Working principle
Beer-Lambert law
I0
A lg (1/T ) lg ( ) b c
It
A -Absorbency of the light when passing through
liquid
T -Transmitted intensity and incident intensity
ratio: transmittance It/I0;
I0 - Incident intensity
It - Transmitted intensity
ε - Molar absorption coefficient of solution
(ml×mmol-1×cm-1);
c - Mol concentration of the solution(mmol/ml);
b - Solution layer thickness(cm);
Abs
(1) 1-point Assay Blank
(B1+B2+B3)/3 R2~R4
R1
S A L A L 1
2
SB B1 B2 B3
AL AL 1
Time
Calculation of absorbance AX
2
Calculation of concentration C X K AX B C1 IFA IFB
Analytical items
TP, ALB, etc.
SB: Stopped cup blank
B1~B3: Passed cup blank
R1~R4: Reagent adding position
Cx: Concentration of standby sample
C1: Concentration of standard 1 solution(reagent blank)
K: Factor
B: Absorbance of blank
IFA and IFB: Instrument constants, representing slope and
intercept
(2) 2-point Assay
( AM AM 1 ) k ( AL AL 1 )
Calculation of absorbance AX
2
Calculation of concentration CX K AX B C1 IFA IFB
Analytical items
CRE, etc.
SB: Stopped cup blank
R1~R3: Passed cup blank
R1~R4: Reagent adding position
Ax: the deference between the photometric point M and L
Cx concentration of standby sample
C1: Concentration of standard 1 solution(reagent blank)
K: Factor
B: Absorbance of blank
IFA and IFB: Instrument constants, representing slope and
intercept
(3) 2-point Rate Assay
Calculation of absorbance:
C 2 C1
K
A2 B
Calculation of concentration
C X K AX B C1 IFA IFB
C 2 C1
K
A2 B
Cx: Concentration of standby sample
AX: Absorbance or change in absorbance per minute
IFA and IFB: Instrument constants, representing slope and
intercept.
Applicable assay
1-point assay, 2-point rate assay, 2-point assay, 3-point assay, 1-
point&rate assay, rate A assay and rate B assay
3.Composition of instrument
1.Analytical unit
switches
(not including
refrigeration power
supply)
2. refrigeration power
supply indicator
(green)
3. main power supply
(breaker)
4.power indicator(red)
12 3 4
1 Power entrance
2 RS232 interface
3 cooling fan
4 light liquid outlet
5 purified water entrance
6 concentrated liquid waste
outlet
7 concentrated waste liquid
level sensor interface
Reaction Unit
Reaction cup:120,
optical path: 6mm
20×6 sets hard optical
plastic
cup
Incubation bath
Digital liquid sensor
7-stop 11-step rinsing of
colorimetric cup
B F G F
C D A B A B E
Reagent Unit
Reagent position:45×2
R1disk:only for R1and R4
R2disk:only for R2 and R3
Position 45 of two reagent
disk:
Only for phosphor-free
cleaning liquid
Reagent bottle volume:
70mland 20ml
Probe mechanism:
1sample probe, 60nm polishing
treatment probe,Digital liquid
level detection,intelligentized
clot detection and collision
protection
2 reagent probes
High-precision digital liquid
detector
Stirring mechanism:2
High-speed hollow cup motor
Surface high-intensity Teflon
coating
115 sample positions,
including:
Routine sample position: 50
STAT sample position : 20
Standard solution position: 34
QC position: 8
Cleaning liquid position: 3
Barcode window:1
Supper high precision syringe
with long life ceram piston.
Guaranteed one million times
usage
Electrical board
1. Reaction disk
Circuit board
2. Sample disk
Circuit board
3. Reagent 1 disk
Circuit board
4. Reagent 2disk
Circuit board
5. ISE
Circuit board
6. Main controller
Board
Switching Power Supply Box:
Order: from left to right
+12V,-12V(simulation)
+12V(lamp)
+5V(digital circuit)
+24V(motor, valve)
1.Interface board