Escolar Documentos
Profissional Documentos
Cultura Documentos
Machine :-
Instrument calibration is not done regularly as per requirement and it is expected
that there shall be no drift and variation in the measured parameters which is
incorrect.
Regular maintenance of the instrument is not under practice.
Age of HVS/RDS is considered indefinite hence no instrument retired even after
completion of their age. Normal life of HVS/RDS is about 5 years but machine
having age of 10 years and above are still used for monitoring.
Over all it can be said that conditions of man and machine are not good for reliable data
generation.
Unintentional Errors In Air Pollution Monitoring
Major Cause of Unreliable Data
• Standard Methods are not followed . If followed several modifications
are incorporated for conveniences with out knowing their ultimate
impact on the results.
• Standard certified instruments are not used . Instruments calibration
is either not done or done for showing that we do calibration. Factors
are not account for.
• Monitoring are done with limited infrastructure and untrained staff
where many times compromises are done without understanding
there impact.
• Financial constraints are most of the time control quality of data.
• Senior responsible group hardly bothered about quality of data.
• Crosschecking of the data is hardly done.
• Management most of the time do not trust data generated by their
own staff.
• Monitoring Data are therefore never used for process optimization,
energy saving etc.
• Monitoring are done to show that we are also monitoring air pollution.
• There is no recognition and rewards for good work as Envt. Mangmt
is not linked with productivity.
Air Quality Monitoring
Monitoring of air pollutants need to be done for
specified period for which standard are notified. All
parameters must be monitored simultaneously.
Monitoring must be done twice a week Total 104 days
monitoring must be done in a year to work out annual
averages.
• Representative Site
• The site should be away from major pollution
sources
The values should not be reported below the detection limit as per
the method.
The interesting aspect of calibration is that user does not account for
errors/variation in calculation. Calibration certificate and data are used to
show that instrument is calibrated thats all.
REQUIRED METHODOLOGY
(A) For Field Calibration
Calibration is required only for those parameters which are taking part in the calculation.
Roots meter, a primary standard traceable to NIST has been set up at Envirotech for
calibration of Top Loading calibrator.
All units of HVS/RDS are possible to calibrate in the field using calibrated portable Top
Loading Calibrator (For Manometer Flow), Dry cal (For Rotameter) and Stop Watch (For
Time Totalizer).
• Validity of Certification
• One year for Calibrators
• EPA recommends a two tier check for individual instruments:
– Monthly Audit at ~1m3/min
– Annual three point calibration check
Correction Required in Calibration of
HVS/RDS
a) Calibration of Top Loading Calibrator has to be done at
Known Temperature and Pressure. Correction for pressure
drop need to be done as below.
Pb- Pm
VTrue = Vmeasured x
Pb
b) If climatic conditions are varying significantly then that of
calibration conditions. Temperature, Pressure corrections
are need to be done for finding out correct flow using.
T2P1
Qcorrect = Qincorrect x
T1P2
T1= absolute temp at orifice calibration conditions (ok)
T2= absolute temp when calibrating the sampler (ok)
P1= Barometric pressure at orifice calibration condition mm Hg
P2= Barometric pressure when calibrating the sampler mm Hg
Calibration Methodology
• Perform a Leak Check
– Block the air entry port on the inlet pipe and check that the flow
rate drops to zero.
• Position the calibrator with one of the perforated plats (Total 4
plates have been provided. (Range 1.3-0.7m3/min).
• Check Zero Level on the Manometer
• Start the Blower and allow it to warm up for at least 5 minutes.
• Record the flow rate indicated by the manometer on the
instrument and the pressure across the calibration orifice on
digital pressure indicator.
• Repeat readings at 3 or 4 points after replacing the perforated
plates.
• Plot the readings and determine the slope of the best-fit
straight line to get a calibration factor for the instrument.
CALIBRATION OF PM10/2.5 SAMPLER
• Leak Check:-
– Remove the PM10 impactor with inlet hood from the sampler.
– Install a filter cassette containing a new 47mm PTFE filter in to the unit.
– Install the leak check adapter on the upper end of the sampling tube of the
sampler
– Ensure that the valves on the leak check adopter and on dry gas meter is in
the open position.
– Switch ON the sampler and ensure that reading of dry gas meter moves
smoothly.
– Shut off the valve of leak check adapter fitted at inlet pipe.
– Shut off the valve of the dry gas meter when vacuum shown in dial gauge
reach at 15 inches of Hg.
– Record the reading on the vacuum gauge. This reading should not drop by
more than 8.5 inches Hg in 30 seconds period use your wrist watch for time
reading, if the reading does drop by more than 8.5 inches Hg in 30 seconds
trace the internal flow paths to identify place of leaks and problem is
connections.
– After leak check open the valve of the dry gas meter.
– Open the valve of the leak check adapter slowly and remove the leak check
adapter and reinstall the impactor with hood.
LEAK CHECK DATA SHEET FOR PM2.5 SAMPLER
Leak Check
Comments: -
CALIBRATION PROCEDURE
Measurement No. Dry Gas Meter Total Volume of Time Taken by Flow Rate
(m3) Air Passed (m3) Stop Watch (Total Volume of Air
Initial Final Passed (m3) x 1000/Time
Taken (Min)
Active silica gel bottles with holes must be placed inside the weighing
chamber.
Thank
you…
Activity Matrix for Calibration of Equipment : SPM
Monitoring
Equipment Acceptance limits Frequency and method of Action of requirements
measurement are
no met
Analytical balance Use 3 to 5 standard weights Gravimetric test; calibration Have balance maintained
covering normal range of at installation and periodic and calibrated by
filter weights; weight checks manufacturer
difference within 1 mg
Relative humidity indicator Indicator reading = Comparison with reading of Adjust or replace to attain
Psychrometer reading + 6% wet bulb/dry bulb acceptance limits
psychrometer receipt and
every 6 monthly after words
On-off timer * + 15 min/24 h Check at purchase and Adjust, and repeat test
quarterly with elapse-time
meter
Elapsed-time meter* + 2 min/ 24 h Standard timepiece of Adjust or replace time
known accuracy at receipt indicator to attain
and every 6 monthly acceptance limits
afterwards
Orifice calibration unit Difference in flowrate <+4 % Rootsmeter at installation Adopt new calibration curve
and after words yearly if orifice damaged or
replace orifice unit
Sampler % deviation = 100 (Qo- Calibration orifice unit on Rerun points for which %
(in operation) Qc)/Qc within 5% receipt and after major deviation exceeds + 5%
Qo = observed flow rate maintenance on sampler until acceptance limits
Qc = flow rate from attained
calibration curve