iPAL / PAL
with specific reference to:
1.Biodiesel in Diesel 2.Oil in Water 3.Anti-oxidants levels of an oil 4.Water in oil
Sensitive
Small changes in sample chemistry can be measured
Universal
Most covalent chemical bonds can be measured All organic chemicals Inorganics such as nitrates, phosphates, chlorates, sulfates, etc.
8 lbs 8 x 8 x 4.5 External Computer USB connection External Power Full spectral analysis Designed for mobile/temp. labs
15 lbs 8 x 11.5 x 7.5 Integrated PDA computer Optional PC Internal battery Designed for field use
Common Specifications
Frequency range 4000 650 cm-1 Maximum Resolution 4 cm-1 Non-hygroscopic optics ZnSe beam splitter Power 100 250 VAC 47 63 Hz, Output: 15VDC Operating temperature 0 to 50 C Humidity 95% non-condensing
2. Rotate TumblIR3. Once TumlIR is in place analyze into place the sample
4. Cleaning is easy!
Transformer Oil
ASTM D2668
Transformer oil ASTM D2668 Version 3
Biodiesel in diesel
EN 14078, LOQ of 1% A2 technologies Method, LOQ of 0.025%
Fuel Biodiesel % in Diesel Version4
1.5% Absorbance at 1745 cm-1 of biodiesel in diesel fuel at 0.0, 0.025, 0.05, 0.1, 0.5, 0.8 and 1.5 % (v/v). 0.8% 0.5% 0.1%
Biodiesel (FAME) has a strong infrared absorbance at 1745 cm-1 Completely free of any diesel interference
Validation results for the Biodiesel in diesel method showing an average relative error of only 1%.
Spectra (A) and calibration curve (B) for oil in water using the A2 filter based method utilising Oil Stabiliser. Mineral oil in water from 0.5 to 45 ppm.
10 mins
SAMPLE
RESULT
Advantages Completely Solvent free & novel oil extraction method. Only three steps. Well established FTIR technique. <ppm limit of detection. Sample to spectra then actual result in ten minutes. Very little sample preparation. On site testing possible. Much reduced glass ware usage compared to liquidliquid extractions.
Mro ic la bP C
N H
Absorbance
Absorbance
3630 3625 3620 3615 3610
3680
3675
3670
3665
3660
3655
3635
3460 3455 3450 3445 3440 3435 3430 3425 3420 3415 3410 3405 3400 3395 Wavenumber
16
19
Concentration
Absorbance
3700
0.0 0.2 0.4 0.6 0.8 Peak Area 1.0 1.2
3690
1.4
3680
3670
3640
3630
3620
3610
PASS 120.00
MONITOR FREQUENTLY
100.00
60.00
15.00
10.00
40.00
Phenolic
5.00 20.00
0.00
0.00
Day 1
Day 2
Day 5
Day 6
Day 8
Day 9
Day 12
Day 13
Day 16
Day 19
Day 22
Day 23
Day 24
Day 26
Oxidation
80.00
1. 2. 3.
Phenolic Diminishes 40% right away- Evaporation, Low molecular weight flash off Aminic stays above 70% until near the end of useful life Aminic Stages of depletion Stage 1: Mid-way point in oil lifespan, 25% depletion Stage 2: Decent from 80% to 40% after phenolic reaches 30%
Phenolic Aminic and Antioxidants (% of Conc. in New Oil)
120.00 PASS M N RFR Q E TL O ITO E UN Y C AN EIM D H G ME IATE Y L 25.00
100.00
Ox idation
80.00
3.Stage 1
60.00
3.Stage 2
15.00
10.00 40.00
5.00 20.00
0.00
New ISO 32
D 1 ay
D 2 ay
D 5 ay
D 6 ay
D 8 ay
D 9 ay
D 12 ay
D 13 ay
D 16 ay
D 19 ay
D 22 ay
D 23 ay
D 24 ay
D 26 ay
20.00
0.022 0.020 0.018 0.016 0.014 0.012 0.010 0.008 0.006 0.004 0.002 0.000 -0.002 4600 4400 4200 4000
3410.8
Reduced Absorbance
Poor Reproducibility
WITHOUT surfactant
Absorbance
Baseline shift
3454.3
3800 3600 3400 3200 3000 2800 2600 2400 2200 Wavenumber 2000 1800 1600 1400 1200 1000 800 600
0.045 0.040 0.035 0.030 0.025 0.020 0.015 0.010 0.005 0.000
Absorbance
3x Abs
WITH A2 surfactant
1127.5 0.3136
2000 1800 1600 1400 1200 1000
4200
4000
3800
3600
3400
3200
3000
2800
800
600
2200
1. 2. 3. 4. 5. 6.
Collect Kit and PAL Load Method Decant 20ml of Oil sample into a suitable container. Add 545 l of water-inoil stabiliser Gently Swirl both clockwise and anti (~30s) Run background for method then place a small drop of the stabilised sample into the well and run.
Addition of surfactant to the oil Creates a stable emulsion Maximum IR water absorbance achieved at 3% surfactant Two different non-ionic PEO based surfactants used Three fold increase in water absorbance Less IR scattered light, more absorbed Smaller water particle (micelle) sizes, with a narrower distribution Water measurements are very reproducible Variability in the non-surfactant method has been eliminated Compares favourably with Gold standard Karl Fischer method
t u r b in e w a t e r s u r f a c t a n t . t d f , 2 5 (R = 0 . 9 9 8 0 2 9 9 6 9 )
WITH
5 50 0
4 00 0
60 58 59 7
28 26 25 7
2 68 16 1 2 65
Predicted Concentration ( F4 C1 )
1 400
WITHOUT surfactant
1 88 10 6
54 56 55
13 5 254 1 253 72 13 6 3 256 13 4 255 308 307 1 74 1 73 1 75 306 1 99 2 00 2 01 2 02 1 85 1 86 1 87
14 5 14 6 14 7 2 79
2 40 42
15 8 16 0 15 9
2 67
2 66
2 38 2 39 2 77 2 14 80 8 2 78 2 81
2 50 0
80 0
22 21 24 3
20 0
500ppm Span
2 14 2 11 13 2 12 92 93 94
2 97 1 2 96 18 1 21 2 94 2 95 1 20 1 19
22 5 22 4 22 3 7 10 10 7 22 4 6 10 10 5
300ppm span
50 52 51 49
1 22
91 -4 00 -10 0 20 0 5 00 8 00 11 00 14 00 17 00
20 8 17 9 1 00 0 46 48
Range
ppm of water results take the same spectral acquisition time and 75 5000 ppm interpretation time regardless of the level of water.
0.075-5.0%
Accuracy
+/- 10%
Virtually independent from operator removing the requirement of subjective end point assignment. Does not interfere with any of the major additives or components in a oil therefore can measure water-in-oil concomitant with additives, breakdown products and contaminants
Application
1.Biodiesel in Diesel
Correlation of the carbonyl of the FattyAcidMethylEster (biodiesel) in Diesel.
1.
Requirements
2.
2.Oil-in-Water (OIW)
filtration with based technique requiring A2 technologies OIW surfactant and lipophilic filters.
3. 4.
optional GC-MS accredited biodiesel in diesel Stds of 0%, 0.05% and 0.5% are available. (12 packs each with a set of three abovementioned stds). iPAL + OIW kit / refill Kit For those without glassware and pipettes OIW kit is most appropriate alternatively the refill kit for those who either have the requisite glassware or have used up all their surfactant and filters). iPAL, no additional requirements.
4.Water-in-oil (WIO)
A 20ml aliquot of the oil is mixed with a 545ul volume of WIO stabiliser, gently swirled, rested then analysed as normal. Nb: 3.&4. can be combined into a single method