BIO-2000
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
KIT
PGINA
KIT
PGINA
ALBUMINA ....................................................... 01
FOSFATO ......................................................... 23
ALDOLASE ...................................................... 02
FOSFATO UV .................................................... 24
GAMA GT .......................................................... 25
AMILASE .......................................................... 04
BILIRRUBINA ................................................... 07
HEMOGLOBINA ............................................... 29
CLCIO ............................................................ 09
CK NAC ............................................................ 10
CLORETOS COLORIMTRICO ...................... 11
COLESTEROL 250 .......................................... 12
COLESTEROL ENZ. LQUIDO ........................ 13
COLESTEROL HDL ........................................ 14
COLINESTERASE ........................................... 15
CPK .................................................................. 16
CREATININA CINTICA ................................... 17
DHL .................................................................. 18
MAGNSIO ...................................................... 31
MICROPROTE ................................................. 32
MUCOPROTENAS .......................................... 33
POTSSIO ....................................................... 34
PROTENAS TOTAIS ........................................ 35
TRANSAMINASES ........................................... 36
TRIGLICRIDES 120 ...................................... 37
TRIGLICRIDES ENZ. LQUIDO ..................... 39
DHL UV ............................................................ 19
FERRO ............................................................ 20
FOSFATASES ................................................... 22
URIA UV ......................................................... 43
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
NDICE
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1 =
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
620
SIM
NO/NO
SIM
37C
900
003
SIM
TRIPLICATA
0004
g/dL
2
0.000
8.000
0.090
0.200
0.353
0.391
***
***
Calibrao
Para obteno do fator, recomenda-se
dosar em triplicata o padro, ler as
absorvncias e tirar a mdia aritmtica das
mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
ALBUMINA
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
01
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1 =
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
546
*****
NO
SIM/SIM
NO
37C
****
003
SIM
TRIPLICATA
0040
U/L
2
0.000
0080
****
****
0.273
0.345
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade e outro soro
controle de valor elevado.
ALDOLASE
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
02
SORO/PLASMA
MODO:
WL1:
WL2:
BLANK:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
FATOR:
UNID:
DEC:
IMPRESSO:
INT. CINTICO:
N INTERVALOS:
AMIN:
%LIM. LIN:
DIR:
ABS. REAT. MIN:
ABS. REAT. MAX:
VR/VN MIN.:
VR/VN MAX.:
CIN
340
*****
NO
SIM
37C
0500
060
1746
UI/L
0
EXTENSA
20
4
0.068
10
DECR
0.800
2.000
***
***
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
ALT/TGP CINTICA
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
03
MODO:
WL1:
WL2:
RETARDO:
CUB. FLUXO:
TEMP:
VOL.ASP:
Absorvncia
670
*****
003
NO
37C
****
AMILASE
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
04
SORO/PLASMA
MODO:
WL1:
WL2:
BLANK:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
FATOR:
UNID:
DEC:
IMPRESSO:
INT. CINTICO:
N INTERVALOS:
AMIN:
%LIM. LIN:
DIR:
ABS. REAT. MIN:
ABS. REAT. MAX:
VR/VN MIN.:
VR/VN MAX.:
CIN
405
*****
NO
SIM
37C
0900
030
3953
UI/L
0
EXTENSA
60
1
0.500
10
INCR
0.000
0.500
***
***
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
AMILASE CNPG
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
05
SORO/PLASMA
MODO:
WL1:
WL2:
BLANK:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
FATOR:
UNID:
DEC:
IMPRESSO:
INT. CINTICO:
N INTERVALOS:
AMIN:
%LIM. LIN:
DIR: DECR:
ABS. REAT. MIN:
ABS. REAT. MAX:
VR/VN MIN.:
VR/VN MAX.:
CIN
340
*****
NO
SIM
37C
0500
060
1746
UI/L
0
EXTENSA
20
4
0.068
10
DECR
0.800
2.000
***
***
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
AST/TGO CINTICA
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
06
1.TCNICA MACRO
1.1. Bilirrubina Total
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
546
*****
NO
SIM/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0010
mg/dL
2
0.000
0020
****
****
0.368
0.412
***
***
P.F.
546
*****
NO
SIM/NO
SIM
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
37C
0900
003
SIM
TRIPLICATA
0010
mg/dL
2
0.000
0020
****
****
0.368
0.412
***
***
2.TCNICA MICRO
2.1. Bilirrubina Total
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
P.F.
546
*****
NO
SIM/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0010
mg/dL
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
BILIRRUBINA
07
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
2.2. Bilirrubina Direta
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
2
0.000
0020
****
****
0.200
0.222
***
***
P.F.
546
*****
NO
SIM/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0010
mg/dL
2
0.000
0020
****
****
0.200
0.222
***
***
REVISO:
BILIRRUBINA
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
08
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
578
*****
SIM
NO/NO
NO
37C
****
003
SIM
TRIPLICATA
0010
mg/dL
2
0000
0015
0.325
0.532
0.372
0.444
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
CLCIO
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
09
SORO/PLASMA
MODO:
WL1:
WL2:
BLANK:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
FATOR:
UNID:
DEC:
IMPRESSO:
INT. CINTICO:
N INTERVALOS:
AMIN:
%LIM. LIN:
DIR:
ABS. REAT. MIN:
ABS. REAT. MAX:
VR/VN MIN.:
VR/VN MAX.:
CIN
340
*****
NO
SIM
37C
0500
120
8095
UI/L
0
EXTENSA
60
3
0.240
10
INCR
0.000
0.700
***
***
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
CK - NAC
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
10
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
100
mEq/L
0
0070
0125
0.000
0.040
0.194
0.250
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
CLORETOS COLORIMTRICO
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
11
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
200
mg/dL
0
0000
0500
0.000
0.200
0.289
0.353
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
COLESTEROL 250
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
12
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
SIM
NO/NO
SIM
37C
0500
003
SIM
TRIPLICATA
200
mg/dL
0
0000
0800
0.000
0.200
0.304
0.372
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
13
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
100
mg/dL
2
0000
0400
0.000
0.200
0.192
0.216
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
COLESTEROL HDL
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
14
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
405
*****
NO
SIM/SIM
SIM
37C
0900
003
SIM
TRIPLICATA
0007
U/mL
2
0000
0014
****
****
0.570
0.630
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade e outro soro
controle de valor elevado.
COLINESTERASE
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
15
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
670
*****
SIM
SIM/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0130
U/L
2
0000
0360
0.016
0.030
0.217
0.255
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade e outro soro
controle de valor elevado.
CPK
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
16
SORO/PLASMA
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
INT. CIN. 060
DIR:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
T.F.
505
*****
NO
NO/NO
SIM
37C
0500
030
SIM
TRIPLICATA
0005
mg/dL
2
INCR
0000
0010
0.000
0.200
0.131
0.149
***
***
SIM
37C
0500
030
SIM
TRIPLICATA
50
mg/dL
2
INCR
0000
0200
0.000
0.200
0.131
0.149
***
***
Procedimento Tcnico
Vide instrues de uso do kit.
URINA
MODO: T.F.
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
INT. CIN. 060
DIR:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
505
*****
NO
NO/NO
CREATININA CINTICA
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
17
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
NO
SIM/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0350
U/L
0
0000
0800
****
****
0.512
0.560
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
DHL
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
18
SORO/PLASMA
MODO:
WL1:
WL2:
BLANK:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
FATOR:
UNID:
DEC:
IMPRESSO:
INT. CINTICO:
N INTERVALOS:
AMIN:
%LIM. LIN:
DIR:
ABS. REAT. MIN:
ABS. REAT. MAX:
VR/VN MIN.:
VR/VN MAX.:
CIN
340
*****
NO
SIM
37C
0500
030
8109
UI/L
0
EXTENSA
60
3
0.155
10
DECR
0.800
2.000
***
***
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
DHL - UV
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
19
Absorvncia
546
*****
003
NO
37C
****
Procedimento Tcnico
Vide instrues de uso do kit
Clculos
Vide instrues de uso do kit
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle N
- Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
FERRO
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
20
SORO/PLASMA
MODO:
WL1:
WL2:
BLANK:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
FATOR:
UNID:
DEC:
IMPRESSO:
INT. CINTICO:
N INTERVALOS:
AMIN:
%LIM. LIN:
DIR:
ABS. REAT. MIN:
ABS. REAT. MAX:
VR/VN MIN.:
VR/VN MAX.:
CIN
405
*****
NO
SIM
37C
0500
30
2764
UI/L
0
EXTENSA
20
3
0.250
10
INCR
0.400
1.100
***
***
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
21
FOSFATASE ALCALINA
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
405
*****
NO
SIM/NO
SIM
37C
0900
003
SIM
TRIPLICATA
150
U/L
2
0000
0267
****
****
0.460
0.526
***
***
P.F.
405
*****
NO
SIM/NO
SIM
37C
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
0900
003
SIM
TRIPLICATA
28
U/L
2
0000
0050
****
****
0.460
0.526
***
***
FOSFATASES
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
22
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
670
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0005
mg/dL
2
0000
0012
0.000
0.050
0.224
1.100
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
FOSFATO
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
23
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
340
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0005
mg/dL
2
0000
0020
0.500
1.500
0.228
0.276
***
***
Calibrao
Para obteno do fator, recomenda-se
dosar em triplicata o padro, ler as
absorvncias e tirar a mdia aritmtica das
mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
FOSFATO - UV
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
24
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
546
*****
NO
SIM/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0100
U/L
0
0000
0175
****
****
0.573
0.669
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
GAMA GT
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
25
SORO/PLASMA
MODO:
WL1:
WL2:
BLANK:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
FATOR:
UNID:
DEC:
IMPRESSO:
INT. CINTICO:
N INTERVALOS:
AMIN:
%LIM. LIN:
DIR:
ABS. REAT. MIN:
ABS. REAT. MAX:
VR/VN MIN.:
VR/VN MAX.:
CIN
405
*****
NO
SIM
37C
0500
003
1111
UI/L
2
EXTENSA
60
3
0.270
10
INCR
0.500
2.000
***
***
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
GAMA GT CINTICA
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
26
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
100
mg/dL
0
0000
0600
0.000
0.080
0.267
0.299
***
***
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
Procedimento Tcnico
Vide instrues de uso do kit
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
27
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
100
mg/dL
0
0000
0600
0.000
0.080
0.299
0.339
***
***
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
Procedimento Tcnico
Vide instrues de uso do kit
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
REVISO:
GLUCOX 500
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
28
HEMOGLOBINA
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
VR/VN MIN.:
VR/VN MAX.:
Procedimento Tcnico
Vide instrues de uso do kit
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de amostra controle de referncia deve
ser uma prtica rotineira do laboratrio.
Recomenda-se utilizar uma amostra controle
com valor na faixa de normalidade e outra de
valor elevado.
REVISO:
HEMOGLOBINA
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
29
MODO:
WL1:
WL2:
RETARDO:
CUB. FLUXO:
TEMP:
VOL.ASP:
Absorvncia
405
*****
003
SIM
25C
0900
Procedimento Tcnico
Vide instrues de uso do kit
Clculos
Vide instrues de uso do kit
Controle de Qualidade
O uso de amostra controle de referncia deve
ser uma prtica rotineira do laboratrio.
Recomenda-se utilizar uma amostra controle
com valor na faixa de normalidade e outra de
valor elevado.
REVISO:
HEMOGLOBINA GLICOSILADA
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
30
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0002
mg/dL
2
0000
3.500
0.380
0.600
0.116
0.168
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
MAGNSIO
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
31
Urina
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
620
*****
SIM
NO/NO
NO
37C
****
003
SIM
TRIPLICATA
0010
mg/dL
2
0000
0030
0.073
0.100
0.119
0.137
**** (L)
****(H)
P.F.
620
*****
SIM
NO/NO
NO
37C
****
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
003
SIM
TRIPLICATA
0050
mg/dL
2
0000
0150
0.073
0.100
0.119
0.137
***
***
MICROPROTE
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
32
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
670
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0006
mg/dL
2
0000
0015
0.000
0.050
0.291
0.341
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
MUCOPROTENAS
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
33
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
578
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0005
mmol/L
2
0000
0010
0.000
0.020
0.191
0.237
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
POTSSIO
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
34
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
546
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0004
g/dL
2
0000
0012
0.080
0.110
0.233
0.285
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
PROTENAS TOTAIS
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
35
Programao em unF.R/mL
TGO
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
PAD2:
PAD3:
PAD4:
PAD5:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
NO
NO/NO
SIM
37C
0900
003
SIM
NICA
0000
24
61
114
190
unF.R/mL
0
0000
0191
****
****
****
****
***
***
P.F.
505
*****
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
PAD2:
PAD3:
PAD4:
PAD5:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
NO
NO/NO
SIM
37C
0900
003
SIM
NICA
0000
28
57
97
150
unF.R/mL
0
0000
0151
****
****
****
****
***
***
P.F.
505
*****
NO
NO/NO
SIM
37C
0900
003
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
TRANSAMINASES
36
PADRO:
PAD:
PAD1:
PAD2:
PAD3:
PAD4:
PAD5:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
SIM
NICA
0000
11.50
29.40
54.94
91.58
U.I./L
2
0000
91.60
****
****
****
****
***
***
P.F.
505
*****
NO
NO/NO
SIM
37C
0900
003
SIM
NICA
0000
13.50
PAD3:
PAD4:
PAD5:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
27.50
46.80
72.30
U.I./L
2
0000
72.35
****
****
****
****
***
***
TRANSAMINASES
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
37
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
200
mg/dL
0
0000
1000
0.000
0.200
0.254
0.306
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
TRIGLICRIDES 120
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
38
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
SIM
NO/NO
SIM
37C
0500
003
SIM
TRIPLICATA
200
mg/dL
0
0000
0800
0.000
0.200
0.239
0.290
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
39
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0007
mg/dL
2
0000
0014
0.000
0.200
0.219
0.245
***
***
Calibrao
Para obteno do fator, recomenda-se
dosar em triplicata o padro, ler as
absorvncias e tirar a mdia aritmtica das
mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
URATO 160
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
40
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
505
*****
SIM
NO/NO
SIM
37C
0500
003
SIM
TRIPLICATA
0007
mg/dL
2
0000
0014
0.000
0.200
0.203
0.227
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve
ser uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
41
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
LIM. LIN.MIN:
LIM. LIN. MAX:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
P.F.
620
*****
SIM
NO/NO
SIM
37C
0900
003
SIM
TRIPLICATA
0080
mg/dL
0
0000
0280
0.000
0.050
0.820
0.902
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle
N - Doles) e outro soro controle de valor
elevado (soro controle P - Doles).
URIA 500
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
42
SORO/PLASMA
MODO:
WL1:
WL2:
BLANK:
BLK.AMOS/PAD:
CUB. FLUXO:
TEMP:
VOL.ASP:
RETARDO:
PADRO:
PAD:
PAD1:
UNID:
DEC:
INT. CIN:
LIM. LIN.MIN:
LIM. LIN. MAX:
DIR:
ABS. REAT. MIN:
ABS. REAT. MAX:
ABS. PAD. MIN:
ABS. PAD. MAX:
VR/VN MIN.:
VR/VN MAX.:
T.F.
340
*****
NO
NO/NO
SIM
37C
0500
030
SIM
TRIPLICATA
80
mg/dL
0
060
0000
200
DECR
1.000
2.000
0.140
0.184
***
***
Calibrao
Para obteno do fator, recomenda-se dosar
em triplicata o padro, ler as absorvncias e
tirar a mdia aritmtica das mesmas.
Controle de Qualidade
O uso de soro controle de referncia deve ser
uma prtica rotineira do laboratrio.
Recomenda-se utilizar um soro controle com
valor na faixa de normalidade (soro controle N
- Doles) e outro soro controle de valor elevado
(soro controle P - Doles).
URIA - UV
REVISO:
13 (11/2004)
Servio de Atendimento ao Consumidor - DDG Fone: 0800 - 703 - 6433 - FAX: 0800 - 703 - 5961 - email: sac@doles.com.br
43