Você está na página 1de 1

PARÂMETROS PARA PROGRAMAÇÃO AUTOQUANT 200i®

Catálogo:K040 ALBUMINA MONOREAGENTE Nº de testes: 1250 Versão: 09/07/2020

Cód.: 00

APRESENTAÇÃO:
K040-1
Reagente Nº 1 – Reagente de Cor – 1 x 250 mL
Reagente Nº 2 – Padrão – 1 x 2 mL

Os reagentes são prontos para uso.

Test Code Test Name


ALB Albumina
Basic Parameter Reference Range Calibration
Method End Point
Primary Filter Secondary Filter Unit Decimal
620 - g/dL 1

R1 Volume R1 Position Incub Time (sec) Sample Volume Reading time


200 # 300 2 -

R2 Volume R2 Position Incub Time (sec) RB High RB Low


- - - 0.0000 0.0000

Co-relation Y= 1.000 X= 0.000


Dilution ratio
2.0

Reagent Lot Reagent Lot Number Expiry Date Linearity


Number
- - 6

Basic Parameter Reference Range Calibration


Normal Range Set Column
No. Gender Sample Type Age Ref Low Ref High
1 Male Serum - 3.5 5.5
2 Female Serum - 3.5 5.5

Basic Parameter Reference Range Calibration


Curve Type Number of Standards Cup or Tube Valid Days
Two Point 2 Tube 0

Standard Position #
Standard Value @
Absorbance -

# = valor definido pelo usuário


@ = valor disponível no rotulo do produto

A Bioclin recomenda o uso do Reagente Nº 2 (Padrão 3,8 mg/dL) ou do calibrador multiparâmetro de bioquímica Biocal – K072
para Calibração.
Para avaliar a precisão e a exatidão das dosagens, recomendamos o uso dos soros controle Biocontrol N – K073 e Biocontrol P –
K074.
Cada Laboratório Clínico deve possuir um programa interno de Controle de Qualidade.

BIOCLIN..QUIBASA . SAC Serviço de Assessoria ao Cliente: 0800 031 5454 | sac@bioclin.com.br . Pag. 1/1