IEB Reset Procedure

February 18, 2021

Purpose

Reset the IEB board procedure for field service. This procedure is to be used to re-establish
communication with IEB if for some reason the IEB network address can not be set via the
LUI. This procedure will not recover a board that has corrupted boot code as indicated by
no heart beat (the LED’s are not flashing when switch 4 is on).

CR ELITE COM PROBLEMA – HOSPITAL DAS CLÍNICAS – I.O.T. –


K46117349

SWITCH 1 => ON

SWITCHES 2 / 3 / 4 => OFF, DEPOIS SWITCH 4 => ON !

OBSERVAMOS APENAS AOS LED’S 1 E 10 PISCANDO !

OBSERVAMOS AO S LED’S 3 / 4 / 5 / 6 / 16 / 17 / 18 / 19 => ACESSOS !

CR ELITE SAUDÁVEL – HOSPITAL DAS CLÍNICAS – I.O.T –


K46117410

LED’S 2 E 3 PISCANDO JUNTOS A UMA VELOCIDADE DE 1


SEGUNDO ON E 1 SEGUNDO OFF ! ESTES DEVEM SER OS HEART
BEATS CITADOS ! CONFORME TABELA ABAIXO:

Page 1 of 8
PELA TABELA ACIMA, ESTAMOS NO CRELITE SAUDÁVEL COM A
OPÇÃO =>KERNEL & APPLICATION LOADED => 1HZ => DS2 &
DS3 SYNCHRONIZED !
Isso pode acontecer quando uma placa IEB carregada com o software 5.2 é colocada em um
scanner 5.1. Algumas placas foram enviadas sem o código do aplicativo LUI (não é
necessário estar na placa do IEB para 5.2), portanto, nenhuma LUI é exibida, talvez levando
o engenheiro de serviço a pensar que a LUI é ruim . Além disso, não há um endereço IP
conhecido na placa, tornando impossível a comunicação com a placa para definir o endereço
IP. Reinicializar a placa define-a para o endereço IP legado padrão (192.168.0.99),
permitindo que o engenheiro de serviço conecte temporariamente o scanner ao console e
baixe o aplicativo de exibição na placa IEB para que o endereço IP correto possa ser
inserido no scanner.

O endereço MAC definido de fábrica não será modificado e permanecerá na configuração


MAC de fábrica.

Observação

Este procedimento não deve ser usado se a conectividade de rede for perdida. Se o IEB já
estava funcionando corretamente e as luzes LED piscam quando o Dip Switch 4 está na
posição ON, o IEB não precisa ser reiniciado. Isso normalmente seria usado quando uma
placa 5.2 IEB é instalada em uma máquina 5.1 e o LUI não é exibido (portanto, não é
possível definir o endereço IP do scanner).

Aviso

A placa do IEB será reformatada e o software do aplicativo e todos os arquivos de


configuração serão perdidos. Se o backup dos arquivos foi feito anteriormente no PC, eles
podem ser restaurados. Se a versão do software for V5.1, o software do aplicativo de
exibição pode ser perdido. Os arquivos diferentes dos arquivos do aplicativo que precisarão
ser restaurados são a calibração do laser, calibrações de potenciômetro digital, calibrações
de placa, etc.

Page 2 of 8
IEB Board Switch Settings
1 Configurações normais do interruptor IEB.
Chave 1 = LIGADA Inicialização I2C. (0 = Inicialização I2C, 1 = Inicialização normal)
Chave 2 = OFF Permaneça na inicialização. (0 = inicialização normal, 1 =
permanecer na inicialização)
Switch 3 = OFF Formatar sistema de arquivos. (0 = inicialização normal, 1 =
sistema de arquivos de formato)
Chave 4 = OFF LED’s (0 = LEDs apagados, 1 = LEDs acesos) - definir para a
posição LIGADO - se
as luzes ainda estão piscando, o quadro está bom

1 A reinicialização do quadro levará aproximadamente 3 minutos.

Page 3 of 8
2 Durante a reinicialização, as taxas de intermitência das luzes no quadro variam. Você saberá
que o processo está completo quando as luzes na placa mudarem de piscar lentamente para piscar
rapidamente, e pelo menos 3 minutos se passaram.

Reset Procedure

1 Exit the Capture Console to the desktop.

2 Reformat the board.


a) Flip all 4 dip switches to the "ON" position. FIZEMOS MANTENDO O CRELITE
ENERGIZADO !
b) H.I.A.E. – PERDIZES ALTERAMOS AS MICRO CHAVES COM O CRELITE
ENERGIZADO. JÁ ESTAVA COM SW1 – ON, MUDAMOS A SW4 DE OFF
PARA ON.
c) Hit the red reset button on the IEB board.
d) H.I.A.E. – PERDIZES, NOTAMOS QUE MUDOU A FREQUÊNCIA DE
ATUAÇÃO DOS LED’s DS2 E DS3, PARA ALGO EM TORNO DE 1 SEGUNDO
ON / 1 SEGUNDO OFF.
e) H.I.A.E. – PERDIZES, REFIZEMOS OS ITENS A/C NOVAMENTE, PORÉM
COM CABO CROSSOVER ENTRE O SCANNER E A CPU UTILIZANDO A
SEGUNDA PORTA DE REDE ETHERNET, E DESTA VEZ MUDOU A
ATUAÇÃO DOS LED’s PARA ALGO EM TORNO DE 2 SEGUNDOS.

3 Bootloader while formatting file system


Blink out of phase slow (2 Second rate)

f) OBSERVAMOS QUE EM SEGUIDA AO PRESSIONARMOS O RESET


BUTTON, ACENDERAM AOS LED’S 3 E 12 !
g) The lights will be blinking slowly for 2 or 3 minutes and then they'll start blinking
rapidly when the reset is complete. NO NOSSO CASO NÃO VIMOS
ALTERAÇÃO NENHUMA APÓS CINCO (05) MINUTOS !
h) H.I.A.E. – PERDIZES, REALMENTE SE TORNARAM BEM MAIS RÁPIDOS.
i) H.I.A.E. – PERDIZES, NESTA SEGUNDA VEZ, REALMENTE DEMOROU O
TEMPO CITADO ACIMA NO ITEM G E DEPOIS SE TORNARMA MAIS
RÁPIDAS.

1 Bootloader only (or bootloader after format file system is complete)


Blink out of phase fast (5Hz)

Page 4 of 8
j) Put the switches back to the normal positions ON-OFF-OFF-OFF (see “IEB Board
Switch Settings”).
k) H.I.A.E. – PERDIZES, VOLTAMOS AO SETUP DAS CHAVES COMO PEDIDO,
MANTENDO APENAS SW1 – ON.

2 Temporarily disconnect the scanner network cable from the customer network and insert it
into the scanner NIC (on the Console PC’s MCPU board). Note: The scanner NIC may be
covered over with white tape. This configuration is similar to the legacy configuration where
the PC talks to the scanner directly through the dedicated NIC on the PC.
3 H.I.A.E. – PERDIZES, OBSERVAMOS NO CONECTOR RJ45, LED VERDE – ON /
LED VERDE – PISCANDO.

4 Extract the file factory service.zip to the console PC c:\temp folder.


5 Run c:\temp\factory service\ftp_IEB.bat. A command window will appear to download the
LUI display application to the IEB. This will take several minutes. When the prompt, “Press
any key to continue” appears, press a key and the command window will exit.

OBSERVAMOS AS SEGUINTES MENSAGENS:


 FTP: CONNECT: UNKNOW ERROR NUMBER
 FTP> KODAK
 INVALID COMMAND
 FTP > HISERV
 INVALID COMMAND
 FTP > BI
 NOT CONNECTED
 FTP > HASH ON
 HASK MARK PRINTING ON FTP: <2048 BYTES / HARSH MARK>
 FTP > PUT IEB_FPGA.RBF
 NOT CONNECTED
 FTP > PUT IEB_APPLICATION
 NOT CONNECTED
 FTP > PUT IEB_KERNEL
 NOT CONNECTED
 FTP > PUT DISPLAY_RAM
 NOT CONNECTED
 FTP > BYE
 POWER CYCLE THE SCANNER AND THEN ENTER THE IP ADDRESS
 PRESS ANY KEY TO CONTINUE…..

H.I.A.E. – PERDIZES, TIVEMOS AS SEGUINTES MENSAGENS:

CONNECTED TO 192.168.0.99
220 SERVICE READY
USER <192.168.0.99:<NONE>>:
331 USER NAME OK, NEED PASSWORD.
230 USER LOGGED IN.

Page 5 of 8
FTP> BI
200 COMMAND OK.
FTP> HASH ON
HASH MARK PRINTING ON FTP: <2048 BYTES / HASH MARK>
FTP> PUT IEB_FPGA.RBF
230 COMMAND OK.
451 ABORTED, CAN’T OPEN FILE.
FTP> PUT IEB_APPLICATION
200 COMMAND OK.
451 ABORTED, CAN’T OPEN FILE
FTP> PUT IEB_KERNEL
200 COMMAND OK.
451 ABORTED, CAN’T OPEN FILE
FTP> PUT DISPLAY_RAM
200 COMMAND OK.
451 ABORTED, CAN’T OPEN FILE
FTP> BYE
POWER CYCLE THE SCANNER AND THEN ENTER THE IP ADDRESS
PRESS ANY KEY TO CONTINUE…..

6 Turn off the scanner and wait at least 30 seconds before turning it back on (prevents possible
corruption of IEB).
7 MANTIDO CABO CROSS OVER CONECTADO ENTRE SCANNER E PORTA DE
REDE NÃO USADA NORMALMENTE EM UM SISTEMA CRELITE SAUDÁVEL !
8 Enter the customer-assigned scanner IP address using the LUI service screen. Exit out of the
service screen on the LUI to the main screen. The LUI will come up with a message to wait
while it reboots and an hour glass will appear on the LUI. DO NOT power cycle the scanner
at this time.

9 MENSAGEM => UNABLE TO SAVE CONFIGURATION DATA AT THIS TIME.


PLEASE RESTART SCANNER AND TRY AGAIN.

10 CONECTADO O CABO CROSS OVER NA PORTA DE REDE DO CLIENTE, E DESTA


VEZ OBTIVE AS SEGUINTES MENSAGENS APÓS A EXECUÇÃO DO PROGRAMA
“FTP IEB.BAT”:

11 CONNECTED TO 192.168.0.99
12 CONNECTION CLOSED BY REMOTE HOST
13 POWER CYCLE THE SCANNER AND THEN ENTER THE IP ADDRESS
14 PRESS ANY KEY TO CONTINUE…..
15 AGUARDADO PELOS TRINTA SEGUNDOS PEDIDOS !!!!!

Page 6 of 8
16 CONTINUAMOS COM A MENSAGEM => MENSAGEM => UNABLE TO SAVE
CONFIGURATION DATA AT THIS TIME. PLEASE RESTART SCANNER AND TRY
AGAIN.

17 PORTANTO NÃO HÁ MUITO O QUE SER FEITO COM ESTA I.E.B. BOARD.

18 While the scanner is rebooting itself, re-connect the scanner’s network cable to the customer
network. (Reverse of Step 3, above.)
19 Reboot the console PC.
20 At the console, navigate to Service->Diagnostics->Software Download and download the
IEB firmware for any versions that are different (for version 5.1, also download the display
application):

Page 7 of 8
 Display Application (v5.1, only)

 IEB FPGA

 IEB Kernel

 IEB Application

21 Restore the following configuration files, if they had previously been backed up.
 Plate Calibration
 Laser configuration
 Galvo digital pot calibrations

More Information:
LED Blink rates on the IEB during reset.
22 Bootloader only (or bootloader after format file system is complete)
Blink out of phase fast (5Hz)

23 Bootloader while formatting file system


Blink out of phase slow (2 Second rate)

24 Kernel with no app loaded


Blink DS2 only @ 1 second, DS3 ON

25 Kernel with app loaded


Blink DS2 and DS3 together @ 1 second

OBSERVAÇÃO !!!!!

EM UMA I.E.B. BOARD “SAUDÁVEL” IREMOS OBSERVAR PELO


SEGUINTE COMPORTAMENTO DOS LED’s => DS2 & DS3:

ELES IRÃO ATUAR SIMULTÂNEAMENTE, OU SEJA, ON E OFF…..ON &


OFF…..EM INTERVALOS DE 1 SEGUNDO !

Page 8 of 8