Esterilizador a Vapor HI VAC MX II


SELEÇÃO

Espaço útil da câmara (médio)


 B0112-140 - 460 x 460 x 660 mm
 B0112-200 - 460 x 660 x 660 mm
 B0112-263 - 460 x 460 x 1245 mm
 B0112-370 - 460 x 660 x 1245 mm
 B0112-418 - 660 x 660 x 960 mm
 B0112-542 - 660 x 660 x 1245 mm

(FOTO ILUSTRATIVA - DETALHES PODEM VARIAR)


 B0112-716 - 660 x 660 x 1645 mm
 B0112-890 - 660 x 660 x 2045 mm
Portas
 Porta Única
 Dupla Porta
Aplicação
 Uso geral em CME
 Ciclos Especiais
 Ciclos Formaldeído
Sistema de Vácuo
 Bomba de Vácuo simples estágio
 Bomba de Vácuo duplo estágio
 Bomba de Vácuo seca
Alimentação de Vapor
 Fonte Externa (Ver requisitos normas)
 Gerador Elétrico
 Gerador por Serpentina de vapor
 Com Comutador
Acabamentos
 Acabamento lateral removível
 Acabamento lateral articulável
 Acabamento traseiro (01 porta)
 Acabamento superior
Componentes opcionais
 Trocador de calor APLICAÇÃO esterilização confiáveis, precisos e validáveis.
 Válvula PID
 Registrador Gráfico Os esterilizadores da linha HI VAC MX II utilizam Custo Operacional Reduzido: Consumo zero de
 Hermeticidade vapor de água sobre pressão em altas temperaturas água na opção de bomba de vácuo seca, que opera
 Dispositivo DR como agente esterilizante, em ciclos com remoção sem a necessidade de água. Baixo consumo elétrico.
Alimentação elétrica
de ar por sistema de vácuo pulsante e controle Redução de até 85% do consumo de eletricidade
 220 V  380 V
 60 Hz  50 Hz automatizado através de controlador microprocessado com a opção de gerador por serpentina de vapor
Acessórios dedicado tipo CLP industrial. Construídos com o industrial.
 Rack de carga  Suporte de solo emprego da melhor tecnologia disponível nas áreas Rapidez e Segurança: Asseguradas por precisas
 Cestos DIN  Rodízios
 Suporte parede da mecânica, termodinâmica, hidráulica, eletrônica rotinas de controle implementadas por software
 Bancada preparo e software, são equipamentos robustos, seguros e proprietário e instrumentação de alta precisão.
 Carro para rack de alta performance. Em modelos com uma ou duas Operação Simplificada: Comando microprocessado
 Cestos ISO
portas e com capacidades que variam de 140 a 890 e Interface intuitiva com tela “Touch Screen”
Componentes para instalação
 Cavalete de água  Cavalete Vapor litros, são destinados ao setor de esterilização de colorida. Facilita a aderência dos usuários aos
 Abrandador  Quadro Elétrico materiais em hospitais e outros estabelecimentos requisitos do processo.
 Osmose de saúde, laboratórios médicos, laboratórios de Versatilidade: Flexibilidade na programação de
 Filtro sanitário
 Compressor de ar controle de qualidade e para processos produtivos processos. Acesso à programação e funções do
 Cavalete ar de indústrias farmacêuticas e outras. Seu uso é equipamento através de senhas de usuários com
Serviços adequado para esterilização de material poroso níveis de permissão configuráveis.
 Instalação
empacotado, instrumentais e utensílios empacotados Processos Sustentáveis: E q u i p a m e n t o
 QI (Qual.Instal.)
 QD (Qual.Des) ou não, vidrarias, luvas, seringas e outros materiais ambientalmente limpo e seguro, sem presença de
 Manutenção em borrachas ou ligas de plásticos resistentes a resíduos ao término do ciclo. Câmaras, portas e
 Start Up altas temperaturas, líquidos em frascos de vidro com tubulações de transporte de vapor com isolamento
 QO (Qual.Oper.)
 Treinamento fechamento ventilado e outros. térmico. Economia de energia, menor dissipação
 Kit peças Podem ainda ser fornecidos com ciclo de formaldeído de calor e menor carga térmica para o sistema de
para esterilização a baixa temperatura de materiais ar condicionado.
termossensíveis. Instalação e Manutenção: Facilitadas por painel
com componentes elétricos sobre trilho e acesso
CARACTERÍSTICAS PRINCIPAIS aos componentes pela parte frontal, traseira e uma
Alto desempenho: assegura processos de das laterais.

1/24
Esterilizador a Vapor HI VAC MX II
MODELOS E CAPACIDADES a face do lado de carga.
A família HI VAC MX II é composta por esterilizadores ALTERNATIVAS PARA ACABAMENTOS
de pequeno, médio e grande porte, com as capacidades Alternativas para a Câmara Material Gabinete frontal
abaixo:
B0112-140 - Capacidade de 140 l · AISI 316 L – 3/16” (4,76 mm) · Aço inox AISI 304 escovado
B0112-200 - Capacidade de 200 l · AISI 316 L – 5/16” (7,93 mm) · Aço inox sem níquel AISI 444 escovado
B0112-263 - Capacidade de 263 l Acabamento lateral
B0112-370 - Capacidade de 370 l · AISI 316 TI – 5/16” (7,93 mm)
B0112-418 - Capacidade de 418 l · Removível em aço inox AISI 304 escovado
Acabamentos
B0112-542 - Capacidade de 542 l · Removível em aço inox sem níquel AISI 444
B0112-716 - Capacidade de 716 l · Polido Sanitário escovado
B0112-890 - Capacidade de 890 l · Eletropolido · Articulável em aço inox AISI 304 escovado.
Exceto para equipamentos de 140 e 200 l
NORMAS APLICÁVEIS
· Articulável em aço inox sem níquel AISI 444
· EN ISO 17665-1:2006 - Sterilization of health care Estrutura: O conjunto da câmara é montado em
escovado. Exceto para equipamentos de 140 e
products - Moist heat - Part 1: Requirements for estrutura em aço inoxidável AISI 304 e altura dos pés
200 l
the development, validation and routine control of a reguláveis para permitir o perfeito nivelamento do
sterilization process for medical devices. equipamento. Acabamento superior
· EN ISO 15223-1:2016 - Medical devices — Symbols Tubulações: as tubulações soldadas na câmara são · Aço inox AISI 304 escovado
to be used with medical device labels, labelling construídas em aço inoxidável AISI 316 e conexões
· Aço inox sem níquel AISI 444 escovado
and information to be supplied — Part 1: General triclamp (rugosidade Ra<0,375µm).
requirements. As tubulações sobre o revestimento da câmara são Acabamento traseiro (porta única)
· NBR 11816:2003 - Esterilizadores a vapor com em aço inoxidável AISI 316 e isoladas por tubo de · Aço inox AISI 304 escovado
vácuo, para produtos de saúde. silicone com fibra de vidro no interior. As conexões são
· Aço inox sem níquel AISI 444 escovado
· NBR 15659:2009 – Esterilização de produtos para rosqueadas ou opcionalmente triclamp (rugosidade
saúde – Esterilizadores de vapor a baixa temperatura e Ra<0,375µm). Hermeticidade hospitalar (porta dupla)
formaldeído – Requisitos e métodos de ensaio. · Aço inox AISI 304, parafusada
· ASME, Section VIII, Division I - Boiler and pressure Alternativas para tubulações fora do Hermeticidade biológica
vessel code. revestimento
· NR 13 - Caldeiras e Vasos de Pressão. · Aço inox AISI 304, soldada
· Rosqueada
CONSTRUÇÃO · Triclamp Postas
Câmara Esterilização: Construída em dupla parede O equipamento pode ser fornecido com uma ou duas
(Câmara interna e Câmara externa), com vinco central O conjunto da câmara é protegido por válvula de portas quando instalado em barreira sanitária, com
e desnível, para facilitar a retirada do condensado, segurança construída em aço inoxidável AISI 316, abertura e fechamento automático, acionados através
otimizando o processo. Câmara interna de perfil selada e com mecanismo de verificação, tendo a da interface do comando. A porta é de movimentação
retangular construída em aço inoxidável AISI 316L, conexão roscada e certificado de calibração. vertical através de um par de pistões pneumáticos,
com espessura de 3/16” (4,76 mm) e acabamento elevando-se no fechamento e rebaixando-se por trás
polido sanitário. Câmara externa construída em gomos Acabamento e Hermeticidade do gabinete na abertura.
no formato de “U” invertidos com o mesmo material da Gabinete frontal e acabamentos laterais A pressão de ar dos pistões pode ser ajustada para
câmara interna. O conjunto é dimensionado conforme removíveis, fabricados em aço inoxidável AISI permitir uma movimentação suave da porta e um
norma ASME para suportar as seguintes pressões: 304 e acabamento escovado. Como padrão o dispositivo de segurança interrompe esse movimento
· Pressão Máxima de Trabalho de 3,0 kgf/cm². equipamento é comercializado sem os acabamentos sempre que haja uma obstrução involuntária ou não,
· Pressão de teste hidrostático de 4,5 kgf/cm² - 1,5 superior e traseiro. para proteção do operador.
vezes a PMTA. Um painel (opcional) fixado a câmara do Travas mecânicas de segurança impedem a abertura
Duas entradas de validação independentes, com 1” de equipamento completa a vedação entre os das portas durante o processo de esterilização e a
diâmetro, estão disponíveis na lateral do equipamento ambientes onde o esterilizador ficará ancorado. abertura simultânea das portas de carga e descarga
possibilitando a introdução de sensores de temperatura Essa vedação pode atender requisitos de no caso de equipamento com duas portas nas fases
e pressão em processos de qualificação operacional hermeticidade como são as exigências da indústria de carga e descarga do equipamento.
e de desempenho. farmacêuticas entre áreas controladas ou apenas Vedação da porta por guarnição de silicone alojada na
Externamente, uma camada em fibra mineral isenta de garantir a manutenção do diferencial de pressão canaleta na câmara, que no fechamento é pressurizada
cloretos, revestida com chapa de aço inoxidável AISI entre áreas contíguas, como são as exigências contra a face interna da porta e na abertura retraída
430 com acabamento polido, diminui a condensação para hospitais. através de pulso de vácuo. Tanto a face da porta como
de vapor e irradiação de calor. Dreno da câmara O padrão hospitalar é desmontável facilitando o a da canaleta são usinadas para um ajuste adequado.
interna protegido com filtro em chapa perfurada de aço deslocamento do equipamento em áreas internas A(s) porta(s) tem a parte interna construída em chapa
inoxidável com diâmetro de 1.½”, localizado próximo dos edifícios. de aço inoxidável AISI 316L. Isolamento interno feito

2/24
Esterilizador a Vapor HI VAC MX II
com manta de fibra mineral isenta de cloretos, estrutura Bombas de Vácuo disponíveis e supervisão de ciclos por “wat chdog timer” (tempo
interna através de perfis de aço e acabamento externo cão de guarda) conforme norma vigente.
· Bomba de Vácuo Simples Estágio
por chapa de aço inoxidável AISI 304 com acabamento Interface Homem Máquina (IHM) com tela Touch
escovado. · B. de Vácuo de 1,5 cv. Para esterilizadores de Screen colorida de 7”, grau de proteção IP65, permite
140, 200, 263, 370, 418 e 542 l. monitorar e controlar todas as funções do equipamento
Alternativas para portas · B. de Vácuo de 3,0 cv. Para esterilizadores de e parâmetros do ciclo. Com porta USB para atualização
140, 200, 263, 370, 418 e 542 l. de software.
· Porta única
· B. de Vácuo de 4,0 cv. Para esterilizadores de Porta Ethernet permite o acesso remoto com
· Dupla porta espelhamento da interface para monitoramento do
263, 370, 418, 542, 716 e 890 l.
Material (face interna) processo.
· B. de Vácuo de 5,0 cv. Para esterilizadores de A Interface permite selecionar e iniciar um ciclo de
· Aço inoxidável AISI 316L 716 e 890 l.
esterilização, acompanhar as etapas do processo com
· Aço inoxidável AISI 316TI Bomba de Vácuo Duplo Estágio a visualização das medições de temperatura, pressão,
Acabamento · B. Vácuo Duplo Estágio de 3,0 cv. Para esteri- estimativa de tempo para término da fase; gerenciar
lizadores de 140, 200, 263, 370, 418, 542, 716 l. usuários e senhas com possibilidade de cadastro
· Polido Sanitário
de 40 usuários com níveis de acesso configuráveis;
· Eletropolido · B. Vácuo Duplo Estágio de 5,0 cv. Para esteri-
lizadores de 890 l. configurar alerta de final de ciclo; visualizar alarmes de
falhas e de segurança; alterar data e hora do sistema;
Bomba de Vácuo Seca
Entrada de ar limpo: para quebra do vácuo através ajuste de tela; acesso ao registro dos últimos alarmes
de filtro hidrófobo esterilizável com eficiência de · B. de Vácuo Seca de 3,0 cv. Para esteriliza- com possibilidade de exportação desses dados; e
99,9997% para partículas maiores ou iguais a 0,22µm. dores de 263, 370, 542, 716 e 890 l. escolha do idioma de operação.
Trocador de calor: Item opcional, por placas em Também pela interface é feito o acesso às rotinas de
aço inoxidável AISI 316, para resfriamento dos fluídos GERAÇÃO DE VAPOR operação para programação dos parâmetros de ciclos
encaminhados ao sistema de efluentes. O vapor utilizado na esterilização de materiais é em faixas de segurança definidas em projeto, acesso
preferencialmente fornecido através de gerador a operação manual, calibração e manutenção com
Alternativas trocador de calor de vapor elétrico incorporado ao equipamento e utilização de login de usuário e senha, e a programação
alimentado por água com qualidade adequada para de atividades rotineiras de manutenção em função do
· Sem trocador de calor.
o processo. número de ciclos ou de horas de funcionamento.
· Com trocador de calor. Exceto para equipa- Sistema Supervisor (opcional) composto por
mentos de 140, 200 l ou equipados com bomba Alimentação de vapor - Alternativas um segundo CLP que recebe dados de sensores de
de vácuo seca.
temperatura independentes comparando os com as
· Com preparação para receber gerador de vapor
elétrico. leituras dos sensores de temperatura do controlador
Sistema de vácuo principal. Em caso de desvio entre as leituras um alarme
Para realização dos pulsos de vácuo no · Sem gerador de vapor. Vapor fornecido de é disparado do sistema supervisor para o controlador
fonte externa.
acondicionamento da carga no início do ciclo e a principal iniciando uma rotina de segurança.
secagem na fase final do processo, através de bomba · Com preparação para receber gerador de vapor
de vácuo de simples estágio, tipo monobloco com anel por serpentina. Alternativas para Comando
de água, dimensionada para atender os requisitos de OBS: O gerador deve ser configurado como item a
· Sem Sistema Supervisor
normas vigentes. parte. Veja detalhes construtivos dos geradores no
item “componentes opcionais”. · Com Sistema Supervisor
Opções para Sistemas de Vácuo
PAINEL ELÉTRICO Painel Secundário: no lado estéril (descarga), em
· Bomba de Vácuo Simples Estágio.
O equipamento possui dois painéis elétricos, sendo um equipamentos com porta dupla, dotado de manômetro
· Bomba de Vácuo de Duplo Estágio. para comando e instrumentação com fonte chaveada com glicerina para leitura de pressão da câmara interna,
Tipo monobloco com anel de água e dois está- de saída em extra baixa tensão 24 Vcc e um painel de botão de operação da porta de descarga, botão de
gios de sucção interligados.
potência com proteção para os elementos de carga emergência e lâmpadas para indicação de ciclo em
· Bomba de Vácuo Seca. O componente por disjuntores. Opcionalmente o painel de potência andamento, ciclo completo, porta de carga fechada
funciona sem a utilização de água ou outro fluído pode ser fornecido com proteção diferencial residual e alarme ativo.
líquido, permitindo um uso ambientalmente cor- (dispositivo DR). Controle do processo realizado pelo CLP principal
reto. com dados de temperatura enviado pelos 2 sensores
· Potência da bomba de vácuo: depende do COMANDO E INSTRUMENTAÇÃO: PT-100 de temperatura alocados junto ao dreno - um
volume da câmara e as opções de configuração Painel de comando: disposto em altura que facilita deles usado para controle de processo e outro para
disponíveis para cada modelo garantem a reali- visualização e operação do lado de carregamento do verificação dos dados de temperatura - e com os dados
zação com sucesso dos testes de Bowie & Dick esterilizador. de pressão enviados pelos transdutores da câmara
e de carga oca conforme definido em normas na- Controlador microprocessado dedicado tipo CLP interna e da câmara externa. O CLP tem o controle
cionais e internacionais
industrial com rotinas de autotestes, autodiagnóstico interno de tempos do processo.

3/24
Esterilizador a Vapor HI VAC MX II
Na fase de esterilização o CLP verifica a condição de · Com sensor certificado tempo de patamar com a manutenção da temperatura
saturação do vapor, através de algoritmo proprietário do processo nessa mesma faixa até o final da fase. Os
com os dados de pressão e temperatura medidos na Sensor para Câmara Externa - Opcional Sensor parâmetros para essa fase são:
câmara de esterilização. para acompanhamento da temperatura na câmara · Temperatura de esterilização com faixa de
Diferenças entre as temperaturas medidas nos sensores externa. programação entre 101 e 135OC.
de controle e verificação, fora da faixa definida em · Tempo de exposição com faixa de programação entre
· Sem sensor
normas e estado de saturação do vapor inadequados 1 e 9999 minutos.
ao processo, emitem alarme e interrompem o ciclo · Com sensor Fase 4- Exaustão
de esterilização · Com sensor certificado Composta pela liberação da pressão da câmara e
Todo o processo de esterilização pode ser acompanhado resfriamento dos fluídos que serão encaminhados
em tempo real pela interface que informa ao operador: CICLOS DE ESTERILIZAÇÃO - PERFIL para o sistema de drenagem. Sem parâmetros para
· Temperatura da câmara interna; A Linha de Esterilizadores HI VAC MX II pode ser programação.
· Temperatura da câmara externa (opcional fornecida com dois programas alternativos e distintos Fase 5 - Secagem
necessitando do sensor para a CE); com perfis com ciclos configuráveis. Em ambos estão Com nova fase de vácuo para eliminação da umidade
· Pressão da câmara interna; disponíveis ciclos com perfil para densidade e para da câmara e secagem da carga. Os parâmetros para
· Pressão da câmara externa; líquidos essa fase são:
· Tempo decorrido da fase de exposição / esterilização; O “Programa C” é carregado como padrão nos · Pulsos de secagem com faixa de programação
· Tempo decorrido para finalização da fase de esterilizadores e disponibiliza ao usuário 9 ciclos de entre 0 e 9999.
secagem; esterilização. · Nível de vácuo que assume o valor definido para os
· Informações sobre status do ciclo; O “Programa D” é opcional e inclui, além de todos os pulsos de vácuo iniciais como padrão.
· Visualização gráfica em tempo real dos dados ciclos do “Programa C”, uma série de ciclos opcionais · Tempo de secagem com faixa de programação entre
temperatura e pressão. com perfil diferenciado e faixas de configuração 1 e 9999 minutos.
O painel contém ainda os instrumentos de segurança distintas, disponíveis na quantidade de ciclos Fase 6 - Aeração
e controle indicados abaixo: desejadas pelo cliente. Consulte para detalhes do Com a entrada de ar filtrado através do filtro hidrófobo,
· Chave Liga / Desliga; programa “D”. a quebra do vácuo e equalização da pressão da câmara
· Botão de Emergência; com o ambiente possibilita a abertura da porta de
· Manovacuômetro da câmara interna; CICLOS MATERIAL DE DENSIDADE descarga. Sem parâmetros programáveis.
· Manômetro da câmara externa; Ciclos com perfil para esterilização de materiais têxteis,
· Manômetro do gerador de vapor ou vapor externo; vidrarias, instrumentos metálicos embalados em caixas CICLOS PARA LÍQUIDOS
· Impressora para registro dos parâmetros alcançados perfuradas ou contêineres, dispositivos com lúmen, Ciclos de esterilização com perfil para líquidos em
no processo. materiais termo sensíveis em látex e/ou plásticos com frascos abertos. Com programação flexível a definição
Instrumentação e controle resistência a temperatura de esterilização. O ciclo é dos parâmetros depende fundamentalmente das
Controle de pressão das câmaras interna e externa composto por 06 fases distintas: características do líquido e do volume do frasco onde
realizado por transdutores de pressão eletrônicos Fase 1 - Remoção de ar e acondicionamento ele é contido. Essas variáveis definem a temperatura
fabricados em aço inoxidável, com compensação Através de pulsos alternados de vácuo e vapor. Essa do processo, o tempo e as características da fase inicial
eletrônica da temperatura e leitura absoluta da pressão. fase só é configurável no ciclo de Bowie & Dick e de retirada de ar.
As indicações de pressão são realizadas digitalmente essa configuração é automaticamente carregada nos O ciclo permite que o usuário escolha o sensor de
através da interface do painel em bar. demais ciclos com esse perfil. Os parâmetros para testemunho, se esse estiver disponível, para o controle
Controle da temperatura por par de sensores PT- essa fase são: do processo.
100 classe A, alocados junto ao dreno da câmara, fazem · Pulso de vácuo inicial com faixa de programação Fase 1 - Remoção de ar e acondicionamento
a leitura da temperatura do processo. As indicações entre 0,05 e 0,90 bar absoluto. Para líquidos em frascos abertos a remoção de ar
de temperatura são realizadas digitalmente através da · Pulso de vapor com faixa de programação entre 1,0 se inicia com um pulso de vácuo pouco profundo
interface do painel em OC. e 2,0 bar absoluto. para evitar o extravasamento do líquido dos frascos
· Número de pulsos com faixa de programação entre e se complementa com uma varredura de vapor,
Sensores de Temperatura 1 e 99. onde a válvula de entrada de vapor e a de drenagem
Fase 2 - Aquecimento da carga permanecem abertas, para que o vapor arraste o
· Sensor de temperatura
Através de injeção de vapor até a que seja atingida residual de ar na câmara para o dreno. São parâmetros
· Sensor de temperatura certificado a temperatura de esterilização na câmara e na carga programáveis dessa fase:
a ser esterilizada. Sem parâmetros programáveis. O · Pulso único de vácuo com profundidade programada
Sensor de carga equipamento controla pressão e temperatura na câmara na faixa de 0,50 e 0,90 bar.
Sensor para medição de temperatura na carga, em função da temperatura de esterilização definida · Pulso de vapor com faixa de programação entre
principalmente em ciclos de esterilização de líquidos. para o processo. 1,0 e 2,0 bar.
Com cabo flexível para o melhor posicionamento no Fase 3 - Exposição · Varredura, com tempo programável entre 01 e 9999
interior da carga. Inclui um intervalo de tempo de até 60 segundos para minutos.
que a temperatura se acomode na faixa definida para Fase 2 - Aquecimento da carga
· Com sensor
o processo (temperatura de esterilização +2OC) e o Com injeção de vapor até alcançar a temperatura de

4/24
Esterilizador a Vapor HI VAC MX II
esterilização, na câmara. Controle da fase através de 134OC, tempo de exposição de 04 minutos e secagem de 1,15 bar; tempo de varredura de 10 minutos;
medições de temperatura e pressão da câmara. de 15 minutos, configuráveis nas faixas disponíveis temperatura de esterilização 110 OC, tempo de 10
Fase 3 - Exposição de cada parâmetro. minutos e temperatura de final de ciclo de 95OC. Todos
O tempo para alcance de temperatura na câmara Ciclo 2 - Termossensíveis parâmetros configuráveis nas faixas disponíveis para
depende basicamente dos volumes de líquido. Quando Fases de remoção de ar e acondicionamento, com os ciclos de líquido.
houver o sensor de testemunho, é possível determinar parâmetros definidos para o teste B&D. Temperatura Programa “F”
que o início da contagem do tempo de exposição 121OC, tempo de exposição de 20 minutos e secagem Composto pelo programa “C” e um ciclo para baixa
será definido pelas medições desse sensor. Sem o de 20 minutos, configuráveis nas faixas disponíveis temperatura a vapor de formaldeído.
sensor, o tempo para que seja atingida a temperatura de cada parâmetro. Ciclo 10 - Formaldeído
é definido na validação do processo e somado ao Ciclo 3 - Rápido(flash) Indicado apenas para os modelos B0112-263(P),
tempo de exposição. Para material desembalado ou embalado B0112-370(P) e B0112-542(P). Processo totalmente
São parâmetros programáveis dessa fase: individualmente. Fases de remoção de ar e validável, compatível com 29 tipos de polímeros além
· Definição do Sensor de Controle: entre o sensor acondicionamento, com os parâmetros definidos para de ser mais econômico que o processo por óxido de
localizado no dreno e o sensor de carga na câmara, o teste B&D. Temperatura 134OC, tempo de exposição etileno e possibilitar maior segurança para pacientes
quando disponível. de 04 minutos e secagem de 25 minutos, configuráveis e para a equipe de enfermagem. Como padrão, o ciclo
· Temperatura de esterilização: com faixa de nas faixas disponíveis de cada parâmetro. apresenta fase de condicionamento da carga através
programação entre 101 e 135OC. Ciclo 4 - Instrumental de pulsos de vácuo e vapor e pulsos de vácuo e vapor
· Tempo de exposição: após o alcance da temperatura Fases de remoção de ar e acondicionamento, com os de formaldeído, fase de exposição na temperatura de
quando utilizado o sensor de controle, ou tempo de parâmetros definidos para o teste B&D. Temperatura 55 ou 65 OC, lavagem com pulsos de vácuo e vapor de
exposição conforme validado na ausência do sensor 134 OC, tempo de exposição de 04 minutos e secagem água e fase de secagem sob vácuo mecânico, este ciclo
de controle. Com faixa de programação ente 01 e de 40 minutos, configuráveis nas faixas disponíveis de não permite configurações pelo usuário. A solução de
9999 minutos. cada parâmetro. formaldeído possui registro na ANVISA, em conjunto
· Controle do ciclo por F0: com a definição da Ciclo 8 - Usuário A com o equipamento.
letalidade pretendida equivalente a uma exposição a Permite a configuração da fase de remoção de ar Programa “D”
121OC. Com faixa de programação entre 01 e 9999. independente dos valores definidos para Bowie & O “Programa D” é opcional e inclui, além de todos os
Fase 4 - Resfriamento Dick. Temperatura de 134 OC, tempo de exposição de ciclos do “Programa C”, uma série de ciclos opcionais
Com a liberação lenta da pressão da câmara para evitar 4 minutos e tempo de secagem de 20 minutos. Todos com perfil diferenciado e faixas de configuração
que o líquido entre em ebulição, até a temperatura os parâmetros, inclusive os da fase de remoção de distintas, disponíveis na quantidade de ciclos
definida para o final de ciclo. É parâmetro programável ar, configuráveis nas faixas permitidas para cada desejadas pelo cliente. Consulte-nos para detalhes
dessa fase: parâmetro. do programa “D”.
Temperatura de final de ciclo: com faixa de programação Ciclos com perfil para líquidos:
entre 80 e 99OC Ciclo 5 - Líquidos OUTROS INSTRUMENTOS
Pulso único e inicial de vácuo até 0,50 bar; pressão Manovacuômetro para indicação do nível de vácuo
CICLOS ESTERILIZAÇÃO DISPONÍVEIS de varredura de 1,15 bar; tempo de varredura de 10 ou de pressão da câmara interna. Com tubo bourdon
Ciclos de teste: min; temperatura de esterilização de 121 OC; tempo de DN 63 mm, mecanismo do sistema em aço inox com
Ciclo 6 - Bowie & Dick 30 minutos e temperatura de final de ciclo de 95 OC. escala de 0 a 6 bar absoluto, calibrado e preenchido
Para verificação da fase de remoção de ar. Com 4 Todos parâmetros configuráveis nas faixas disponíveis com glicerina para estabilização.
pulsos de vácuo até 0,30 bar, pressão de vapor de para ciclos de líquido. Manômetros para indicação de pressão, um
1,15 bar, configuráveis nas faixas de programação Ciclo 9- Usuários B para câmara externa e um segundo para o gerador
disponíveis e tempo de espera de 30 segundos não Idem ao ciclo 5. de vapor. Com tubo bourdon DN 63 mm, mecanismo
configurável. Ciclo para Lactário do sistema em aço inox com escala de 0 a 6 bar
Temperatura de 134OC, tempo de exposição de 210 Tanto o ciclo 9 como o ciclo 5, podem ser programados absoluto, calibrado e preenchido com glicerina para
segundos e tempo de secagem de 01 minuto, conforme como ciclos de lactário. O nome do ciclo pode ser estabilização.
norma específica, sem possibilidade de alteração. alterado pela interface do comando.
Ciclo 7 - “Leak test” Os parâmetros indicados para o ciclo de lactários são: Alternativas para Manômetros
Para verificação da estanqueidade da câmara. Ciclo não pulso único e inicial de 0,50 bar; pulso de varredura
· Manômetro calibrado
programável, com parâmetros fixos de fábrica. Pulso de 1,15 bar; tempo de varredura de 10 minutos;
de vácuo até 0,30 bar, intervalo para acomodação da temperatura de esterilização 110 OC, tempo de 10 · Manômetro calibrado com certificação RBC
pressão de 05 minutos e acompanhamento da pressão minutos e temperatura de final de ciclo de 95OC. Todos
a partir do nível estabilizado por 10 minutos. O teste é parâmetros configuráveis nas faixas disponíveis para Impressora térmica de 40 colunas, constituída de
considerado “aceito”, quando o aumento de pressão ciclos de líquido. corpo termoplástico de alta resistência e tecnologia
for menor que 0,013 bar. Tanto o ciclo 9 como o ciclo 5, podem ser programados de impressão térmica, bobina de papel protegida por
Ciclos com perfil de densidade. como ciclos de lactário. O nome do ciclo pode ser tampa, interface serial, e instalada no painel frontal
Ciclo 1 - “Têxteis” alterado pela interface do comando. do esterilizador.
Fases de remoção de ar e acondicionamento, com os Os parâmetros indicados para o ciclo de lactários são: Os dados impressos garantem a completa
parâmetros definidos para o teste B&D. Temperatura pulso único e inicial de 0,50 bar; pulso de varredura documentação do processo de esterilização com

5/24
Esterilizador a Vapor HI VAC MX II
informações como: nas linhas de alimentação de vapor das duas câmaras, Comutador de vapor
· Nome do estabelecimento; identificação do das linhas de vácuo e exaustão, e da entrada de ar no Sistema para comutação da alimentação de vapor,
esterilizador; data; hora de início do ciclo; nº do ciclo; final do processo. acionado eletronicamente através da interface de
código da carga. Demais controles como pressurização e vácuo das comando do esterilizador. Para substituir a alimentação
· Número e nome do programa com parâmetros a canaletas das portas e entrada de água para bombas de vapor de um gerador incorporado ao equipamento
serem alcançados no processo. Informação de controle através de válvulas solenoides. para uma fonte central de fornecimento. Conexão com
do ciclo por F0, quando for o caso. Válvula PID com corpo em aço inoxidável AISI rosca em inox ou latão conforme a escolha de material
· Informação das fases executadas, com registro 316, com atuador pneumático e controlador PID para as tubulações do equipamento.
de hora em minutos e segundos e os dados integrado. Permite, com a associação das tecnologias
correspondentes de temperatura do sensor de controle, pneumáticas e eletrônica, uma precisão no controle de Alternativas Comutador de vapor
temperatura do sensor de verificação, pressão medida fluxo e de pressões na alimentação de vapor, facilitando Válvulas Pneumáticas
pelo transdutor da CI e se disponível temperatura do o ajuste da velocidade de subida da temperatura na
sensor de testemunho na carga. câmara e eliminando flutuações de temperatura durante · Sem comutador de vapor
· No final de ciclo a informação de ciclo completo e o o processo de esterilização. · Com comutador de vapor
tempo de processo. Quando controlado por “F0” será Bomba de alimentação do gerador
impresso o valor “F” correspondente. Alternativas Válvula de Controle
· A ocorrência de falha na execução do ciclo. · Com tratamento de níquel cromo
Válvulas Pneumáticas
· Em inox AISI 316 ou similar
Alternativas para impressoras · Em aço inox AISI 316
· Em liga de bronze A capacidade do gerador de vapor influi no tempo
· Impressora térmica
Válvula PID de ciclo de esterilização. Para atender a demandas
· Impressora matricial diferenciadas e a disponibilidade de potência elétrica
· Sem válvula PID nos locais de instalação existem variadas potências
Impressora matricial de 40 colunas, constituída · Com válvula PID em inox AISI 316 disponíveis para o mesmo modelo de equipamento.
de corpo termoplástico de alta resistência e tampa
protetora de bobina de papel e cabeça de impressão COMPONENTES OPCIONAIS Potência gerador de vapor
com mecanismo de impressão de impacto rápido de Geradores Elétricos
oito agulhas. Com interface serial e instalada no painel SISTEMA DE GERAÇÃO DE VAPOR
frontal do esterilizador. Gerador de vapor Elétrico com abastecimento · Com potência de 18 kW.
Para esterilizadores de 140 e 200 l.
Registrador Gráfico (opcional) com alta qualidade automático de água, através de bomba centrífuga com
para a aquisição de dados como tempo, temperatura tratamento à base de níquel cromo. O corpo do gerador · Com potência de 24 kW.
e pressão. Apresenta seis canais configuráveis que é construído em aço inoxidável AISI 316L e resistências Para esterilizadores de 263 e, 370 l.
permitem visualização de medidas, escalas ou barras blindadas em aço inoxidável AISI 316. · Com potência de 36 kW.
gráficas e sua interface é Touch Screen. A operação do Controle de nível de água por sistema de boia de aço Para esterilizadores de 263, 370, 418 e 542 l.
registrador gráfico pode ser realizada no equipamento inoxidável, para detecção de nível mínimo e máximo
· Com potência de 44 kW.
ou opcionalmente a distância via comunicação de água no gerador. Para esterilizadores de 263, 370, 418 e 542 l.
ETHERNET. Os dados podem ser salvos via USB. Drenagem parcial e automática feita no final de cada
· A alimentação elétrica deve ser fornecida a partir de ciclo, controlada por temporizador para diminuir a · Com potência de 52 kW.
Para esterilizadores de 263, 370, 418 e 542 l.
quadro elétrico, com as proteções indicadas na tabela concentração de minerais na água. O gerador possui
de dados técnicos. O quadro elétrico deve ser instalado chave seletora para drenagem total do gerador para · com potência de 72 kW.
próximo ao esterilizador; limpeza do fundo. Para esterilizadores de 716 e 890 l.
· Embora seja possível a instalação do equipamento Válvula de segurança com certificado de calibração, em Gerador por Serpentina a Vapor
com a lateral direita encostada na alvenaria (olhando aço inoxidável AISI 316 e conexões rosqueadas, selada
· Capacidade de 100 kg/h de vapor.
pelo frontal de carga do esterilizador), o ideal é prever e com mecanismo de verificação de funcionamento, Para esterilizadores de 263, 370, 418 e 542 l.
acesso por toda a periferia do equipamento; protege o conjunto do gerador. Construído conforme
· São necessários espaços de trabalho na frente das normas vigentes. · Capacidade de 200 kg/h de vapor.
Para esterilizadores de 716 e 890 l
portas, no lado de carga e descarga do equipamento. Gerador de Vapor por Serpentina Construído em
aço inoxidável AISI 316L e sistema de aquecimento por
Alternativas Registrador Gráfico serpentina de aço inoxidável AISI 316, alimentada por COMPONENTES OPCIONAIS
vapor industrial. Comando, controles e componentes O equipamento necessita de ar comprimido para
· Sem Registrador Gráfico
iguais ao modelo elétrico vedação de portas e movimentação de válvulas.
· Com Registrador Gráfico Para fornecimento de ar comprimido está disponível
Alternativas Registrador Gráfico compressor de ar, montado fora da estrutura do
Válvulas de controle de processo com acionamento equipamento. Com reservatório de 50 ou 100 litros
· Gerador de Vapor Elétrico
pneumático independente em aço inoxidável AISI 316. e capaz de fornecer ar comprimido com 8,0 kgf/cm²
Controladas por válvulas solenoides de ar comprimido, · Gerador de Vapor por Serpentina de pressão.
6/24
Esterilizador a Vapor HI VAC MX II
Alternativas para ar comprimido desligamento do esterilizador e despressurização da · pH entre 5 e 8.
câmara de esterilização; · Cloretos < 3 mg/l.
· Ar comprimido de rede
· Trava mecânica que impede a abertura da porta · Pressão Dinâmica: Mínimo 1 bar absoluto.
· Com Compressor de ar quando o esterilizador está em funcionamento. OBS: As vazões necessárias variam com a configuração
e modelo do equipamento. Consulte o Guia de
OSMOSE REVERSA NOTAS Instalação para maiores informações.
Sistema opcional para tratamento de água para a A largura total dos equipamentos desembalados está
alimentação do gerador de vapor, dotado de 3 estágios indicada na tabela de dados de engenharia. Para ALIMENTAÇÃO DE VAPOR
de filtração – com grau de retenção de 5 e 1 mícron, e facilitar a circulação do equipamento no interior do O suprimento de vapor direto para a câmara da
carvão ativado – bomba pressurizadora, membranas edifício, o frontal pode ser desmontado, reduzindo-se autoclave não é solução adequada para atendimento
de osmose, comando e reservatório de água tratada. a largura. É importante verificar o percurso que será dos requisitos das Boas Práticas para processos de
percorrido pelo esterilizador no interior do edifício para esterilização de materiais.
Alternativas para Osmose Reversa a verificação de pontos de estrangulamento. A liberação Alimentação de Vapor direta
desses pontos é de responsabilidade do comprador; · Qualidade conforme EN 285 e NBR ISO 17665
· Sem sistema de osmose
As bitolas das conexões e linhas de suprimentos parte 2.
· Com sistema de osmose para 20 l/h para es- indicadas em Dados Técnicos referem-se aos pontos de · Pressão 3,5 a 3,7 bar absoluto.
terilizadores de 140 a 370 l conexão com o esterilizador. As tubulações que levam · Título vapor acima de 0,95.
· Com sistema de osmose para 60 l/h para es- esses suprimentos até os locais de instalação devem OBS: A Baumer não se responsabiliza por manchas em
terilizadores de 418 a 890 l ser calculadas por terceiros e o dimensionamento instrumentos e tecidos e por corrosão em instrumentos
depende da distância, das pressões de fornecimento e partes do equipamento no caso de alimentação de
SEGURANÇA e das vazões necessárias; vapor direto fora dos padrões sugeridos pelas normas
O equipamento apresenta os seguintes itens de acima.
segurança: DADOS DE ENGENHARIA A melhor alternativa é o fornecimento de vapor
· Na falta de energia elétrica a entrada de vapor é DRENO: industrial para um gerador de vapor por serpentinas.
fechada; A linha de drenagem deve ser independente do esgoto Alimentação de Vapor para gerador de vapor
· Quando há excesso de pressão a passagem de predial e conectada a caixa de passagem externa. A por serpentina
vapor é fechada; ligação do equipamento com a linha de esgoto predial · Pressão 7 a 9 bar absoluto.
· Válvula de segurança normalizada e calibrada em só está indicada quando o equipamento possuir · Título vapor acima de 0,95.
3,0 kgf/cm² com dispositivo de limpeza e verificação trocador de calor para resfriar os efluentes liberados
de funcionamento; no processo. ALIMENTAÇÃO AR COMPRIMIDO
· Início do ciclo apenas na presença de pressão de Tubulação em material resistente ao calor. Ar comprimido livre de impurezas e de umidade e
vapor no gerador ou na linha de suprimento; ABASTECIMENTO DE ÁGUA com pressão de fornecimento de 6 a 7 bar. Conexão
· Sistema de elevação automática da porta com Os esterilizadores possuem entrada de água para o de engate rápido de 6mm.
pressão de trabalho calibrada, para proteção do gerador de vapor e para o sistema de vácuo. Na entrada
operador; de água para gerador, pode ser conectado um sistema ALIMENTAÇÃO ELÉTRICA
· Porta com sistema antiesmagamento que interrompe de osmose reversa para tratamento da água utilizada A linha HI VAC MX II possui quadro de distribuição
a movimentação da porta caso ela esteja obstruída; para geração de vapor. interno com ligação trifásica e terra com impedância
· Impossibilidade de abertura das portas após o Alimentação de água para Bomba de Vácuo máxima de 10 Ohms e proteções para cada elemento
início do ciclo e na presença de pressão de vapor na · Dureza máxima de 250 ppm de CaCO3. de carga por disjuntor.
câmara interna; · Temperatura ideal menor que 15 OC. Máxima OBS: Consulte o Guia de Instalação para identificação
· O ciclo só se inicia com portas fechadas e com as admissível 25OC. das proteções necessárias a cada configuração do
guarnições pressurizadas; · pH 5 a 8. equipamento.
· Impossibilidade de abertura simultânea das duas · Pressão entre 1,2 e 1,5 bar absoluto.
portas pelo operador (no caso de esterilizadores de Alimentação de água para Osmose Reversa
barreira); · Dureza menor que 60 ppm de CaCO3.
· Aborto do ciclo quando a temperatura da câmara na · Condutividade menor que 500 µS/cm.
fase de esterilização estiver abaixo da definida para · Temperatura da água menor que 25OC
o processo; · pH entre 5 e 8.
· Termostato de segurança para proteção das · Cloretos < 60 mg/l.
resistências elétricas; · Pressão Dinâmica: Entre 2,0 e 4,0 bar absoluto.
· Rotina de emergência que na falta de energia elétrica · Trava mecânica que impede a abertura da porta
mantém o travamento das portas e a liberação da quando o esterilizador está em funcionamento.
pressão de vapor da câmara interna; Alimentação de água para gerador de vapor
· Sistema de alarme para falha nos suprimentos de · Dureza menor que 10 ppm de CaCO3.
água, vapor e ar comprimido; · Condutividade menor que 50 µS/cm.
· Botão de emergência no painel do equipamento para · Temperatura da água entre 5 e 35OC
7/24
Esterilizador a Vapor HI VAC MX II
CAPACIDADE E DIMENSÕES DOS EQUIPAMENTOS

50

ABERTURA ALVENARIA - LARGURA


WALL OPENING - WIDTH

LADO DE CARGA
LOADING SIDE
B
ABERTURA ALVENARIA - ALTURA
WALL OPENING -HEIGHT

25
C + 50

ABERTURA ALVENARIA - LARGURA


C

WALL OPENING - WIDTH

LADO DE DESCARGA
25 UNLOADING SIDE
A

A + 50

VISTA V1 - LADO DE DESCARGA DIMENSÕES EM PLANTA


V1 VIEW - UNLOADING SIDE DIMENSIONS IN PLAN

ESTERILIZADOR DUPLA PORTA ESTERILIZADOR UMA PORTA


DOUBLE DOOR STERILIZER SINGLE DOOR STERILIZER

DISTÂNCIA MINIMA -
BF
BF

LADO DIREITO /FACE DE CARGA ÁREA OPERACIONAL


ÁREA OPERACIONAL - LADO DE CARGA .
OPERATIONAL AREA
MINIMIUM DISTANCE
OPERATIONAL AREA - LOAD SIDE
RIGHT SIDE / LOADING SIDE

50

AFASTAMENTO MÍNIMO
-
MINIMUM CLEARANCE
ÁREA MANUTENÇÃO
25 25
MAINTENANCE AREA
DISTÂNCIA MIN. - LADO ESQ.
FACE DE CARGA
-
MINIMUM DISTANCE - LEFT SIDE 600
LOADING SIDE
50 DISTÂNCIA MIN. - LADO ESQUERDO
FACE DE CARGA
25 25 -
MINIMIUM DISTANCE - LEFT SIDE
LOADING SIDE

V1 V1
25 ÁREA MANUTENÇÃO
600 ESTERILIZADOR 01 PORTA
600

MAINTENANCE AREA
ÁREA OPERACIONAL - LADO DESCARGA 01 DOOR STERILIZER
OPERATIONAL AREA - UNLOADING SIDE
BF

ÁREA OPERACIONAL E DE MANUTENÇÃO ÁREA OPERACIONAL E DE MANUTENÇÃO


OPERATIONAL AND MAINTENANCE AREA OPERATIONAL AND MAINTENANCE AREA

8/24
Steam Sterilizer HI VAC MX II
SELECTION

Chamber dimensions
 B0112-140 - 460 x 460 x 660 mm
 B0112-200 - 460 x 660 x 660 mm
 B0112-263 - 460 x 460 x 1245 mm
 B0112-370 - 460 x 660 x 1245 mm
 B0112-418 - 660 x 660 x 960 mm
 B0112-542 - 660 x 660 x 1245 mm

(FOTO ILUSTRATIVA - DETALHES PODEM


 B0112-716 - 660 x 660 x 1645 mm
 B0112-890 - 660 x 660 x 2045 mm
Door
 Single door
 Double door
Application
 Commonly used in CSSD
 Special cycles
 Formaldehyde cycles
Vacuum System
 Vacuum pump – one stage
 Vacuum pump - two stage
 Dry Vacuum Pump
Steam source
 External Source
 Electric steam generator
 Steam Generator with steam coil
 Switch (direct steam + steam generator)
Panels
 Removable side panel
 Articulated side panel
 Back panel (single door)
 Top panel
Optional components
 Heat exchanger
APPLICATION without the need for water. Low power consumption.
 PID valve
 Recorder Sterilizers HI VAC MX II line using steam under Reduction of up to 85% of electricity consumption
 Barrier panel pressure at high temperatures as a sterilizing agent with the generator option for industrial steam coil.
 Residual current circuit breaker in cycles by pulsing air removal and automated Speed and Security: Ensured by precise control
Electrical power
 220 V  380 V vacuum control system through dedicated type routines implemented by proprietary software and
 60 Hz  50 Hz microprocessor controller PLC industrial. Built with high precision instrumentation.
Accessories the use of the best technology available in the areas of Simplified operation: Microprocessor control
 Load rack  Floor support
mechanics, thermodynamics, hydraulics, electronics and intuitive interface screen with “Touch Screen”
 DIN baskets  Caster
 Wall support and software, equipment is robust, secure and high colorful. Facilitates adherence of users to process
 Preparation bench performance. On models with one or two doors and requirements.
 Rack trolleys with capacities ranging from 140 to 890 liters are Versatility: Flexibility in scheduling processes.
 ISO baskets
Installation components intended for material sterilization sector in hospitals Access to programming and equipment functions
 Water treatment  Softener and other healthcare facilities, medical laboratories, through user passwords with configurable
 Air treatment  Electrical panel quality control laboratories and production processes permission levels.
 Steam treatment
of pharmaceutical and others. Its use is suitable for Sustainable Processes: Environmentally clean
 Osmosis
 Air compressor sterilization of packaged porous material, instruments and safe equipment without residues to cycle
 Sanitary filter and utensils packaged or not, glassware, gloves, completion. Chambers, doors and vapor transport
Services syringes and other materials in rubber or alloys pipes with thermal insulation. Energy saving, lower
 Installation
 IQ (Instal. Qual.) resistant to high temperatures, liquids in glass bottles heat dissipation and lower thermal load on the air
 PQ (Perf. Qual.) with ventilated closure and others. conditioning system.
 Maintenance It can also be offered with formaldehyde cycle to Installation and Maintenance: Facilitated by
 Start Up
sterilization at low temperature for those thermosensitive panel for electrical components and rail access to
 OQ (Oper Qual.).
 Training materials. components from the front, rear and one side.
 kit parts MODELS AND CAPABILITIES
MAIN FEATURES HI VAC MX II family consists of sterilizers small,
High performance: ensures sterilization reliable, medium and large, with capacities below:
accurate and validable. B0112-140 - capacity of 140 l
Reduced Operating Cost: Zero water consumption B0112-200 - capacity of 200 l
in the dry vacuum pump option, which operates B0112-263 - capacity of 263 l

9/24
Steam Sterilizer HI VAC MX II
B0112-370 - capacity of 370 l · Polished Sanitary · Carbon steel with anticorrosion protection
B0112-418 - capacity of 418 l and painting
· Electropolished
B0112-542 - capacity of 542 l Side panel
B0112-716 - capacity of 716 l
B0112-890 - capacity of 890 l Structure: The chamber assembly is mounted on a · Removable stainless steel AISI 304 brushed
structure in AISI 304 stainless steel and the height of · Removable stainless steel without nickel AISI
STANDARDS the feet is adjustable to allow the perfect leveling of 444 brushed
· EN ISO 17665-1:2006 - Sterilization of health care the equipment. · Removable carbon steel with anticorrosion
products - Moist heat - Part 1: Requirements for Pipes: the pipes welded in the chamber are made protection and painting
the development, validation and routine control of a of AISI 316 stainless steel and triclamp connections
· Articulated in stainless steel AISI 304
sterilization process for medical devices. (roughness Ra< <0,375µm).
brushed. Except for devices 140 and 200 l
· EN ISO 15223-1:2016 - Medical devices — Symbols The pipes over the chamber lining are made of AISI
to be used with medical device labels, labelling 316 stainless steel and insulated by a silicone tube · Articulated in stainless steel without nickel
and information to be supplied — Part 1: General with fiberglass inside. The connections are threaded AISI 444 brushed. Except for devices 140 and
requirements. or optionally triclamp (roughness Ra< <0,375µm). 200 l
· NBR 11816:2003 - Esterilizadores a vapor com Top panel
vácuo, para produtos de saúde. Alternatives to Pipes · Stainless steel AISI 304 brushed.
· NBR 15659:2009 – Esterilização de produtos para
saúde – Esterilizadores de vapor a baixa temperatura e Outside coating · Stainless steel without nickel AISI 444
brushed
formaldeído – Requisitos e métodos de ensaio.
· stainless steel AISI 316
· ASME, Section VIII, Division I - Boiler and pressure · Carbon steel with anticorrosion protection
vessel code. · Brass / Bronze and painting
· NR 13 - Caldeiras e Vasos de Pressão. Back panel (single door)
The whole of the camera is protected by safety valve
· Stainless steel AISI 304 brushed
CONSTRUCTION built in AISI 316 stainless steel, sealed and verification
Sterilization chamber: Built in double wall (inner mechanism, with the threaded connection and · Stainless steel without nickel AISI 444
chamber and outer chamber) with central crease calibration certificate. brushed
gap and to facilitate the removal of the condensate, · Carbon steel with anticorrosion protection
optimizing the process. Rectangular Chamber internal Alternatives p / Safety Valve and painting
profile built in AISI 316L stainless steel with a thickness Barrier - hospital (double door)
of 3/16 “(4.76 mm) and sanitary polished finish. Jacket · stainless steel AISI 316
in “U” shaped sections inverted in the same material as · Cast iron / bronze · Stainless steel AISI 304, screwed
the chamber. The set is sized according to the ASME Barrier – Biological (double door)
standard to withstand the following pressures: Panels and Barrier panel
· Stainless steel AISI 304, welded
· Maximum working pressure of 3.0 kgf / cm². Front panel and removable side panel manufactured
· Hydrostatic test pressure of 4.5 kgf / cm² - 1.5 times in AISI 304 stainless steel and brushed finish. As
the PMTA. standard the equipment is marketed without the Doors
Two independent validation inputs, with 1” in diameter, top and back panels. The equipment can be supplied with one or two doors
are available on the side of the equipment, enabling A panel (option) attached to the equipment chamber when installed in a sanitary barrier, with automatic
the introduction of temperature and pressure sensors completes the seal between the environments opening and closing, activated through the command
in operational and performance qualification processes. where the sterilizer will be anchored. This seal interface. The door moves vertically through a pair of
Externally, a chloride-free mineral fiber layer, coated (barrier panel) can meet airtight requirements as pneumatic pistons, rising at the closing and lowering
with AISI 430 stainless steel plate with a polished are the requirements of the pharmaceutical industry behind the cabinet at the opening.
finish, reduces steam condensation and heat irradiation. between controlled areas or just guarantee the The air pressure of the pistons can be adjusted to allow
Chamber drain protected with 1.½” diameter perforated maintenance of the pressure differential between a smooth movement of the door and a safety device
stainless steel filter, located close to the face of the contiguous areas, as are the requirements for stops this movement whenever there is an unintended
loading side. hospitals. obstruction or not, to protect the operator.
The hospital standard is dismountable, facilitating Mechanical safety locks prevent the doors from
Alternatives to the chamber material the movement of equipment inside buildings. opening during the sterilization process and the
simultaneous opening of the loading and unloading
· AISI 316 L – 3/16” (4,76 mm) ALTERNATIVE FINISHES doors in the case of equipment with two doors in the
· AISI 316 L – 5/16” (7,93 mm) Front panel loading and unloading phases of the equipment.
Sealing the door by a silicone gasket housed in a
· AISI 316 TI – 5/16” (7,93 mm) · Stainless steel AISI 304 brushed channel in the chamber, which at pressurized closing
Finishes · Stainless steel without nickel AISI 444 is closed against the internal face of the door and in the
brushed retracted opening by means of a vacuum pulse. Both
10/24
Steam Sterilizer HI VAC MX II
the face of the door and that of the channel are rectify · Vacuum pump power: depends on the viewing and operation of the sterilizer load side.
for proper adjustment. volume of the chamber and the configuration op- Microprocessor controller dedicated PLC
The door (s) has an internal part made of stainless steel tions available for each model guarantee the suc- industrial type with routine self-tests, self-diagnosis
AISI 316L. Internal insulation made with chloride-free cessful completion of Bowie & Dick and hollow and supervision cycles by “watchdog timer” (watchdog
mineral fiber blanket, internal structure through steel load tests as defined in national and international time) according to the current standard.
profiles and external finish by AISI 304 stainless steel standards. Human Machine Interface (HMI) Touch Screen
plate with brushed finish. Available Vacuum pumps with color screen 7”, IP65, allows you to monitor and
control all machine functions and parameters of the
· Vacuum Pump - one Stage.
Alternative for door cycle. With USB port for software update.
· Vacuum Pump 1.5 HP Optional Ethernet port allows remote access with
· Single door For sterilizers 140, 200, 263, 370, 418 and 542 l. mirrored interface for monitoring the process.
· Double door · Vacuum Pump 3.0 HP The Interface allows you to select and start a
Material (inner face) For sterilizers 140, 200, 263, 370, 418 and 542 l. sterilization cycle, follow the steps of the process
· Vacuum Pump 4.0 HP with the visualization of the temperature, pressure
· AISI 316L stainless steel
For sterilizers 263, 370, 418, 542, 716 and 890 l. measurements, estimated time to end the phase;
· AISI 316TI stainless steel manage users and passwords with the possibility
· Vacuum Pump 5.0 HP of registering 40 users with configurable access
Finishing
For sterilizers 716 and 890 l.
levels; configure end of cycle alert; view fault and
· Polished Sanitary Vacuum Pump – Two stage security alarms; change system date and time; screen
· Electropolished · Vacuum pump – two stage 3.0 HP For steriliz- adjustment; access to the record of the last alarms
ers 140, 200, 263, 370, 418, 542,716 l. with the possibility of exporting this data; and choice
Entry of clean air: to break the vacuum through of operating language.
· Vacuum pump – two stage 5.0 HP
hydrophobic sterilizable filter with 99.9997% efficiency For sterilizer 890 l. The interface also provides access to the operation
for larger particles or equal to 0,22μm. routines for programming the cycle parameters in
Dry Vacuum Pump
Heat exchanger: Optional item for plates in AISI security bands defined in the project, access to manual
316 stainless steel, for cooling of the fluid sent to the · Dry vacuum pump 3.0 HP operation, calibration and maintenance using a user
effluent system. For sterilizers 263, 370, 542, 716 and 890 l. login and password, and the programming of routine
maintenance activities. depending on the number of
Alternative Heat Exchanger STEAM GENERATION cycles or hours of operation.
The steam used in the sterilization of materials is Supervisor System (optional): composed of a
· Without heat exchanger.
preferably supplied through an electric steam generator second PLC that receives data from independent
· With heat exchanger. Except for equipment incorporated into the equipment and fed by water with temperature sensors comparing them with the
140, or 200 L equipped with a dry vacuum adequate quality for the process. temperature sensor readings from the main controller.
pump.
In case of deviation between readings, an alarm is
Steam supply - Alternatives triggered from the supervisor system to the main
Vacuum System controller, initiating a safety routine.
· With preparation for receiving electric steam
For carrying out the vacuum pulses in the load generator.
conditioning at the beginning of the cycle and drying Command Alternatives
in the final stage of the process, through a single stage · With preparation for receiving Generator with
steam coil. · Without Ethernet port
vacuum pump, monobloc type with water ring, sized to
meet the requirements of current standards. · Without steam generator. Steam supplied from · With Ethernet port
an external source. · Without Supervisor system
Options for Vacuum Systems NOTE: The generator must be set up as a separate item.
· With Supervisor system
See construction details of the generators in “optional
· Vacuum Pump - one Stage.
components”.
· Vacuum Pump – two stage. Secondary panel on the sterile side (discharge),
monobloc with water ring and two interconnected ELECTRIC PANEL in equipment with double door, equipped with a
suction stages.
The equipment has two electrical panels, one for manometer with glycerin to read the pressure of the
· Dry Vacuum Pump. command and instrumentation with a switched source Chamber, operation button of the discharge door,
The component works without the use of water of output in extra low voltage 24 Vdc and a power emergency button and lamps for indicating the cycle
or other liquid fluid, allowing an environmentally panel with protection for the load elements by circuit in progress, complete cycle, door closed load and
friendly use. breakers. Optionally, the power panel can be provided active alarm.
with a residual current circuit breaker (RCCB). Process control carried out by the main PLC with
temperature data sent by the 2 PT-100 temperature
COMMAND AND INSTRUMENTS: sensors located next to the drain - one used for process
Control panel: arranged in height that facilitates control and the other for checking the temperature

11/24
Steam Sterilizer HI VAC MX II
data - and with the pressure data sent by the transducers · Without sensor the process (sterilization temperature + 2 OC) and
the inner chamber and the outer chamber. The PLC has the threshold time with maintaining the process
· With sensor
the internal control of process times. temperature in that same range until the end of the
In the sterilization phase, the PLC checks the condition · With sensor and certification phase. The parameters for this phase are:
of saturation of the steam, through a proprietary Jacket sensor - Optional Sensor for monitoring the · Sterilization temperature with a programming range
algorithm with the pressure and temperature data temperature in the outer chamber. between 101 and 135 OC.
measured in the sterilization chamber. · Exposure time with a programming range between
· Without sensor
Differences between the temperatures measured 1 and 9999 minutes.
in the control and verification sensors, outside the · With sensor Phase 4 - Exhaustion
range defined in standards and vapor saturation · With sensor and certification With a new vacuum phase to eliminate moisture from
status inappropriate to the process, emit an alarm and the chamber and dry the load. The parameters for this
interrupt the sterilization cycle. STERILIZATION CYCLES - PROFILE phase are:
The entire sterilization process can be followed in real The HI VAC MX II Line of Sterilizers can be supplied Phase 5 - Drying
time by the interface that informs the operator: with two alternative and distinct programs with profiles Com nova fase de vácuo para eliminação da umidade
· Chamber temperature; with configurable cycles. Both cycles with density and da câmara e secagem da carga. Os parâmetros para
· Jacket temperature (optional requiring the sensor liquid profiles are available essa fase são:
for the jacket); The “Program C” is loaded as standard in the sterilizers · Drying pulses with a programming range between
· Chamber pressure; and provides the user with 9 sterilization cycles. 0 and 9999.
· Jacket pressure; “Program D” is optional and includes, in addition to · Vacuum level that assumes the value set for the initial
· Time elapsed from the exposure / sterilization phase; all “Program C” cycles, a series of optional cycles with vacuum pulses as standard.
· Time elapsed to complete the drying phase; a different profile and different configuration ranges, · Drying time with programming range between 1 and
· Cycle status information; available in the number of cycles desired by the 9999 minutes.
· Real-time graphical display of temperature and customer. Refer to the “D” program details. Phase 6 - Aeration
pressure data. With the entry of filtered air through the hydrophobic
· The panel also contains the security and control DENSITY MATERIAL CYCLES filter, breaking the vacuum and equalizing the chamber
instruments indicated below: Profile cycles for the sterilization of textile materials, pressure with the environment makes it possible to
· On-off key; glassware, metallic instruments packed in perforated open the discharge door. No programmable parameters.
· Emergency button; boxes or containers, devices with lumen, thermo-
· Manovacuometer of the chamber; sensitive materials in latex and / or plastics with LIQUID CYCLES
· Jacket pressure gauge; resistance to sterilization temperature. The cycle Sterilization cycles with profiles for liquids in open
· Manometer of the steam generator or external steam; consists of 06 distinct phases: bottles. With flexible programming the definition
· Printer for recording the parameters reached in the Phase 1 - Air removal and conditioning of parameters depends fundamentally on the
process. Through alternating pulses of vacuum and steam. characteristics of the liquid and the volume of the
This phase is only configurable in the Bowie & Dick bottle where it is contained. These variables define the
Instrumentation and control cycle and this configuration is automatically loaded process temperature, the time and the characteristics
Pressure control of the internal and external in other cycles with this profile. The parameters for of the initial phase of air removal.
chambers performed by electronic pressure transducers this phase are: The cycle allows the user to choose the load sensor, if
made of stainless steel, with electronic temperature · Initial vacuum pulse with a programming range one is available, for process control.
compensation and absolute pressure reading. Pressure between 0.05 and 0.90 bar absolute. Phase 1 - Air removal and conditioning
indications are made digitally via the panel interface · Steam pulse with programming range between 1.0 For liquids in open flasks, air removal starts with
in bar. and 2.0 bar absolute. a shallow vacuum pulse to prevent the liquid from
Temperature control by pair of PT-100 class A · Waiting time after steam pulse with a programming leaking out of the flasks and is complemented by a
sensors, located next to the chamber drain, read the range between 01 and 9999 seconds. steam sweep, where the steam inlet and drain valve
process temperature. Temperature indications are made · Number of pulses with a programming range between remain open, to let the steam drag the residual air in
digitally via the panel interface in OC. 1 and 9999. the chamber into the drain. Programmable parameters
of this phase are:
Temperature sensors Phase 2 - Load heating · Single vacuum pulse with programmed depth in the
Through steam injection until the sterilization range of 0.50 and 0.90 bar.
· With sensor
temperature in the chamber and the load to be · Steam pulse with programming range between 1.0
· With sensor and certification sterilized is reached. No programmable parameters. and 2.0 bar.
The equipment controls pressure and temperature in · Scan, with programmable time between 01 and
Load sensor - Optional the chamber according to the sterilization temperature 9999 minutes.
Sensor for measuring temperature in the load, available defined for the process.
as a control option, mainly in liquid sterilization cycles. Phase 3 - Exhibition Phase 2 - Load heating
With flexible cable for the best positioning inside the Includes a time interval of up to 60 seconds for the With steam injection until reaching the sterilization
cargo. temperature to settle within the range defined for temperature, in the chamber. Phase control through
12/24
Steam Sterilizer HI VAC MX II
chamber temperature and pressure measurements. Cycle 2 - Thermosensitive at a temperature of 50 or 60 O C, washing with vacuum
Phase 3 - Exhibition Phases of air removal and conditioning, with the pulses and water vapor and phase mechanical vacuum
The time to reach temperature in the chamber basically parameters defined for the B&D test. Temperature 121 drying, this cycle does not allow user configurations.
depends on the volumes of liquid. When there is a O
C, exposure time of 20 minutes and drying time of The formaldehyde solution is registered with ANVISA,
core sensor, it is possible to determine that the start 20 minutes, configurable in the available ranges of together with the equipment.
of the exposure time count will be defined by the each parameter. Program “D”
measurements of that sensor. Without the sensor, the Cycle 3 - Fast (flash) The “Program D” is optional and includes, in addition
time to reach the temperature is defined in the process For unpacked or soaked material. Phases of air removal to all “Program C” cycles, a series of optional cycles
validation and added to the exposure time. and conditioning, with the parameters defined for the with a different profile and different configuration
Programmable parameters of this phase are: B&D test. Temperature 134 OC, exposure time of 04 ranges, available in the number of cycles desired by
· Control Sensor Definition: between the sensor minutes and drying time of 05 minutes, configurable the customer. Consult us for details of “D” program.
located in the drain and the load sensor in the chamber, in the available ranges of each parameter.
when available. Cycle 4 - Instrumental OTHER INSTRUMENTS
· Sterilization temperature: with a programming range Phases of air removal and conditioning, with the Manovacuometer for indicating the vacuum or
between 101 and 135 OC. parameters defined for the B&D test. Temperature 134 pressure level of the inner chamber. With DN 63 mm
· Exposure time: after reaching the temperature when O
C, exposure time of 04 minutes and drying time of bourdon tube, stainless steel system mechanism with
using the control sensor, or exposure time as validated 30 minutes, configurable in the available ranges of a scale of 0 to 6 bar absolute, calibrated and filled with
in the absence of the control sensor. With programming each parameter. glycerin for stabilization.
range between 01 and 9999 minutes. Cycle 8 - User A Pressure gauges for pressure indication, one
· Cycle control by F0: with the definition of the desired It allows the configuration of the removal phase for an external chamber and a second for the steam
lethality equivalent to an exposure to 121 OC. With of independent values defined for Bowie & Dick. generator. With DN 63 mm bourdon tube, stainless steel
programming range between 01 and 9999. Temperature of 134OC, exposure time of 4 minutes and system mechanism with a scale of 0 to 6 bar absolute,
drying time of 30 minutes. All parameters, including calibrated and filled with glycerin for stabilization.
Phase 4 - Cooling those of the air removal phase, configurable in the
With the slow release of pressure from the chamber to permitted ranges for each parameter. Alternatives to Gauges
prevent the liquid from boiling, until the temperature set Cycles profile for liquids:
· Gauge calibrated
for the end of the cycle. It is a programmable parameter Cycle 5 - Liquids
of this phase: Single and initial vacuum pulse up to 0.50 bar; · Gauge calibrated and certified RBC
End of cycle temperature: with programming range scanning pressure of 1.15 bar; scan time of 10 min;
between 80 and 99 OC sterilization temperature of 121 O C; time of 30 minutes Thermal printer 40 column thermal printer, high
and end of cycle temperature of 95 O C. All configurable resistance thermoplastic body and thermal printing
STERILIZATION CYCLES AVAILABLE parameters in the ranges available for liquid cycles. technology, paper roll protected by cover, serial
Test cycles: Cycle 9 - User B interface and installed on the front panel of the sterilizer.
Cycle 6 - Bowie & Dick Same as cycle 5. The printed data guarantees a complete sterilization
To check the air removal phase. With 4 vacuum pulses Lactate cycle process with information such as:
up to 0.20 bar, vapor pressure of 1.15 bar and a waiting Both cycle 9 and cycle 5 can be programmed as · Name of the establishment; sterilizer identification;
time of 30 seconds, configurable in the available lactation cycles. The cycle name can be changed via Dice; cycle start time; cycle number; charge code.
programming ranges. the command interface. · Program number and name with parameters reached
Temperature of 134 OC, exposure time of 210 seconds The parameters indicated for the lactation cycle in the process. Cycle control information by F0, when
and drying time of 01 minutes, according to specific are: single and initial pulse of 0.50 bar; 1.15 bar applicable.
norm, without possibility of alteration. scanning pulse; scan time of 10 minutes; sterilization · Information about the phases performed, with
Cycle 7 - “Leak test” temperature 110 OC, time of 10 minutes and end of time recording in minutes and seconds and data
To check the stability of the chamber. Non-programmable cycle temperature of 95 OC. All configurable parameters corresponding to the temperature of the control sensor,
cycle, with factory fixed parameters. Vacuum pulse up in the ranges available for liquid cycles. temperature of the verification sensor, pressure of the
to 0.20 bar, interval for accommodating the pressure “F” Program chamber transducer and available temperature of the
of 05 minutes and monitoring the pressure from the Composed by program “C” and a cycle for low test load sensor.
stabilized level for 10 minutes. The test is considered temperature with formaldehyde steam. · No final information on the cycle and time of the
“accepted”, when the pressure increase is less than Cycle 10 - Formaldehyde process. When controlled by “F0”, the corresponding
0.013 bar. Indicated only for models B0112-263 (P), B0112- “F” value will be printed.
Density profile cycles. 370 (P) and B0112-542 (P). Fully validable process, · Failure to execute the cycle.
Cycle 1 - Textiles compatible with 29 types of polymers in addition to
Phases of air removal and conditioning, with the being more economical than the ethylene oxide process Alternatives to printers
parameters defined for the B&D test. Temperature 134 and providing greater safety for patients and the nursing
· Thermal printer
O
C, exposure time of 04 minutes and drying time of team. As standard, the cycle has a load conditioning
15 minutes, configurable in the available ranges of phase through pulses of vacuum and steam and pulses · Matrix printer
each parameter. of vacuum and steam of formaldehyde, exposure phase

13/24
Steam Sterilizer HI VAC MX II
Matrix printer 40 columns composed of thermoplastic Partial and automatic drainage done at the end of each · With power of 44 kW.
body of high strength and printing paper and the cycle, controlled by timer to decrease the concentration For sterilizers 263, 370, 418 and 542 l.
head coil protective cover with an impact printing of minerals in the water. The generator has a selector
· With power of 52 kW.
mechanism quick eight needles. With serial interface switch for total drainage of the generator to clean the
For sterilizers 263, 370, 418 and 542 l.
and installed in the sterilizer front panel. bottom.
Recorder (optional) with high quality to acquire data Safety valve with calibration certificate, in AISI 316 · With power of 72 kW.
such as time, temperature and pressure. It features six stainless steel and threaded connections, sealed and For sterilizers 716 and 890 l.
configurable channels that allow viewing measures, with function check mechanism, protects the generator Serpentine generator for Steam
graphic scales or bar and its interface is Touch Screen. set. Built according to current regulations. · Capacity 100 kg/h of steam.
The chart recorder operation can be performed on the Serpentine Steam Generator For sterilizers 263, 370, 418 and 542 l.
equipment or optionally via Ethernet communication Built in AISI 316L stainless steel and AISI 316 stainless
distance. Data can be saved via USB. steel coil heating system, powered by industrial steam. · Capacity 200 kg/h of steam.
For sterilizers 716 and 890 l
Command, controls and components equal to the
Alternative Recorder electrical model
COMPRESSED AIR SUPPLY
· Without Recorder
Alternatives to Steam generator The machine requires compressed air to drive the
· With Recorder sealing ports and valves. For the supply of compressed
· Electric Steam Generator
air is available air compressor mounted outside of the
Process control valves with independent pneumatic · Steam generator with steam coil machine structure. With the reservoir 50 or 100 liters
actuation in AISI 316 stainless steel. Controlled by Generator Safety Valve and capable of supplying compressed air to 8.0 kgf /
compressed air solenoid valves, in the steam supply cm² pressure.
· Stainless steel AISI 316
lines of the two chambers, the vacuum and exhaust Alternatives for compressed air
lines, and the air inlet at the end of the process. · Cast iron or bronze
· Compressed air network
Other controls such as pressurization and vacuum of Steam switch
the door channels and water inlet for pumps through System for steam power switching, electronically · With air Compressor
solenoid valves. actuated through the sterilizer control interface. To
PID valve with AISI 316L stainless steel body, with replace the steam generator power to an embedded REVERSE OSMOSIS
pneumatic actuator and integrated PID controller. device to a central source of supply. Threaded Optional system for treating feed water for steam
It allows, with the association of pneumatic and connection stainless steel or brass as the material of generator, provided with three filter stages - with a
electronic technologies, precision in the control of choice for pipes of the equipment. degree of retention of 5 and 1 micron, and activated
flow and pressure in the steam supply, facilitating the carbon - pressurizing pump, reverse osmosis
adjustment of the temperature rise rate in the chamber Alternatives steam switch membranes, control and treated water tank.
and eliminating temperature fluctuations during the Válvulas Pneumáticas
sterilization process. Alternatives to Reverse Osmosis
· With steam switch
· Without osmosis system
Alternative Control Valve · Without steam switch
· With osmosis system to 20 l/h for sterilizing
Pneumatic valves Generator feed pump
140-370 l
· Stainless steel AISI 316 · Nickel chrome treatment
· With osmosis system of 60 l / h for sterilizing
· Bronze alloys · In stainless steel AISI 316 or similar 418-890 l
PID valve
The steam generator capacity influences the SAFETY
· Without PID valve sterilization cycle time. To meet different demands The equipment has the following safety items:
· With PID valve stainless steel AISI 316L and the availability of electrical power in the installation · In the absence of electricity, the steam inlet is closed;
locations are different powers available to the same · When there is excess pressure, the vapor passage
OPTIONAL COMPONENTS model of equipment. is closed;
· Safety valve standardized and calibrated at 3.0 kgf/
SISTEMA DE GERAÇÃO DE VAPOR Power steam generator cm² with cleaning device and function check;
Electric steam generator with automatic water Electrical generators · Start of the cycle only in the presence of steam
supply, through a centrifugal pump with nickel pressure in the generator or in the supply line;
chromium treatment. The generator body is constructed · With power of 18 kW. · Automatic door lift system with calibrated working
For sterilizers 140 and 200 l.
of AISI 316L stainless steel and armored heating pressure, to protect the operator;
element in stainless steel AISI 316. · With power of 24 kW. · Door with anti-crushing system that interrupts the
Control of water level by stainless steel float system, For sterilizers 263 and 370 l. movement of the door if it is obstructed;
for detection of minimum and maximum level of water · With power of 36 kW. · Impossibility of opening the doors after the beginning
in the generator. For sterilizers 263, 370, 418 and 542 l. of the cycle and in the presence of vapor pressure in
14/24
Steam Sterilizer HI VAC MX II
the internal chamber; Water supply for reverse osmosis
· The cycle starts only with closed doors and · Hardness less than 60 ppm CaCO3.
pressurized linings; · Conductivity less than 500 µS/cm.
· Impossibility of opening the two doors simultaneously · Water temperature less than 25 OC
by the operator (in the case of barrier sterilizers); · pH 5 to 8.
· Abortion of the cycle when the chamber temperature · Chloride <60 mg/l.
in the sterilization phase is below that defined for the · Dynamic pressure: min 1 bar and a maximum of
process; 4 bar.
· Safety thermostat to protect electrical resistances;
· Emergency routine that, in the absence of electricity, Water supply for steam generator
keeps the doors from locking and the pressure release · Hardness less than 10 ppm CaCO3.
from the internal chamber; · Conductivity less than 50 µS/cm
· Alarm system for water, steam and compressed air · Water temperature between 5 and 35OC
supply failure; · pH 5 to 8.
· Emergency button on the equipment panel to switch · Chloride <3 mg/l.
off the sterilizer and depressurize the sterilization · Dynamic Pressure: Minimum 1 bar absolute.
chamber;
· Mechanical lock that prevents the door from opening NOTE: The required flow rates vary with the configuration
when the sterilizer is in operation. and model of equipment. See the Installation Guide for
more information.
Note
The total width of the unpacked equipment is indicated STEAM SUPPLY
in the engineering data table. To facilitate the movement The direct steam supply to the autoclave chamber is
of the equipment inside the building, the front can be not suitable solution for meeting the requirements of
disassembled, reducing the width. It is important to Good Practices for material sterilization processes.
check the route that will be taken by the sterilizer inside
the building to check for bottlenecks. The release of Direct Steam supply
these points is the buyer’s responsibility; · Quality suggestion as EN 285 and ISO 17665-2.
The gauges of the connections and supply lines · Pressure 3.5 to 3.7 bar absolute.
indicated in technical data refer to the connection points · Steam title above 0.95.
with the sterilizer. The pipes that take these supplies to NOTE: The Baumer is not responsible for stains on
the installation sites must be calculated by third parties instruments and tissue and corrosion on instruments
and the dimensioning depends on the distance, supply and pieces of equipment in case of direct steam supply
pressures and necessary flows; outside the standards suggested by the standard above.
The best alternative is the provision of process steam
ENGINEERING DATA for a steam Generator with steam coil.
DRAIN: Steam supply for Steam Generator with steam
The drain line must be independent of the sewer coil
building and connected to an external junction box. · Pressure 7-9 bar absolute.
The connection of the equipment to the building sewer · Steam title above 0.95.
line is only indicated when the equipment has heat
exchanger to cool the effluents released in the process. COMPRESSED AIR SUPPLY
Piping made of heat resistant material. Compressed air free of impurities and moisture and
WATER SUPPLY supply pressure of 6 to 7 bar. Push-in connector 6 mm.
The sterilizers have water inlet to the steam generator
and the vacuum system. In generator for water inlet ELECTRICAL SUPPLY
can be connected to a reverse osmosis system to treat HI VAC MX II line has an internal electrical panel with
water used for steam generation. three-phase and earth connection with a maximum
impedance of 10 Ohms and protections for each load
Water supply for Vacuum Pump element per circuit breaker.
· Maximum hardness 250 ppm CaCO3. NOTE: Refer to the Installation Guide to identify
· Ideal temperature lower than 15 OC. Maximum the necessary protections for each equipment
allowable 25 OC. configuration.
· pH 5 to 8.
· Pressure between 1.2 and 1.5 bar absolute.

15/24
Steam Sterilizer HI VAC MX II
EQUIPMENT CAPACITY AND DIMENSIONING

50

ABERTURA ALVENARIA - LARGURA


WALL OPENING - WIDTH

LADO DE CARGA
LOADING SIDE
B
ABERTURA ALVENARIA - ALTURA
WALL OPENING -HEIGHT

25
C + 50

ABERTURA ALVENARIA - LARGURA


C

WALL OPENING - WIDTH

LADO DE DESCARGA
25 UNLOADING SIDE
A

A + 50

VISTA V1 - LADO DE DESCARGA DIMENSÕES EM PLANTA


V1 VIEW - UNLOADING SIDE DIMENSIONS IN PLAN

ESTERILIZADOR DUPLA PORTA ESTERILIZADOR UMA PORTA


DOUBLE DOOR STERILIZER SINGLE DOOR STERILIZER

DISTÂNCIA MINIMA -
BF
BF

LADO DIREITO /FACE DE CARGA ÁREA OPERACIONAL


ÁREA OPERACIONAL - LADO DE CARGA .
OPERATIONAL AREA
MINIMIUM DISTANCE
OPERATIONAL AREA - LOAD SIDE
RIGHT SIDE / LOADING SIDE

50

AFASTAMENTO MÍNIMO
-
MINIMUM CLEARANCE
ÁREA MANUTENÇÃO
25 25
MAINTENANCE AREA
DISTÂNCIA MIN. - LADO ESQ.
FACE DE CARGA
-
MINIMUM DISTANCE - LEFT SIDE 600
LOADING SIDE
50 DISTÂNCIA MIN. - LADO ESQUERDO
FACE DE CARGA
25 25 -
MINIMIUM DISTANCE - LEFT SIDE
LOADING SIDE

V1 V1
25 ÁREA MANUTENÇÃO
600 ESTERILIZADOR 01 PORTA
600

MAINTENANCE AREA
ÁREA OPERACIONAL - LADO DESCARGA 01 DOOR STERILIZER
OPERATIONAL AREA - UNLOADING SIDE
BF

ÁREA OPERACIONAL E DE MANUTENÇÃO ÁREA OPERACIONAL E DE MANUTENÇÃO


OPERATIONAL AND MAINTENANCE AREA OPERATIONAL AND MAINTENANCE AREA

16/24
Esterilizador a Vapor HI VAC MX II
SELECCIÓN:

Dimensiones de la Cámara
 B0112-140 - 460 x 460 x 660 mm
 B0112-200 - 460 x 660 x 660 mm
 B0112-263 - 460 x 460 x 1245 mm
 B0112-370 - 460 x 660 x 1245 mm
 B0112-418 - 660 x 660 x 960 mm
 B0112-542 - 660 x 660 x 1245 mm

(FOTO ILUSTRATIVA - DETALHES PODEM


 B0112-716 - 660 x 660 x 1645 mm
 B0112-890 - 660 x 660 x 2045 mm
Puertas
 Una Puerta
 Dos Puertas
Aplicación
 Uso general en Central de Esterilización
 Ciclos Especiales
 Ciclos Formaldehido
Sistema de Vacío
 Bomba de Vacío una etapa
 Bomba de Vacío dos etapas
 Bomba de Vacío seca
Alimentación de Vapor
 Fuente Externa (Ver requisitos normas)
 Generador Eléctrico
 Generador por Serpentina de vapor
 Con Conmutador
Acabados
 Acabados lateral removible
 Acabados lateral articulable
 Acabados trasero (01 puerta)
 Acabados superior
Componentes opcionales
 Termocambiador
APLICACIÓN Reducción del costo de operación: Consumo
 Válvula PID
 Registrador Gráfico Los esterilizadores de línea HI VAC MX II utilizan vapor cero de agua en la opción de bomba de vacío seco,
 Hermeticidad de agua bajo presión a altas temperaturas como agente que funciona sin necesidad de agua. Bajo consumo
 Dispositivo DR esterilizante, en ciclos con eliminación de aire mediante eléctrico. Reducción de hasta el 85% del consumo
Alimentación Eléctrica
 220 V  380 V sistema de vacío pulsante y control automatizado a de electricidad con la opción de generador de
 60 Hz  50 Hz través de controlador de microprocesado dedicado bobina de vapor industrial.
Accesorios tipo PLC industrial. Construidos con el uso de la Velocidad y seguridad: Garantizada por rutinas
 Rack de carga  Suporte de solo
mejor tecnología disponible en las áreas de mecánica, de control precisas implementadas por software
 Cestos DIN  Ruedas
 Suporte pared termodinámica, hidráulica, electrónica y software, son propietario e instrumentación de alta precisión.
 Bancada preparo equipos robustos, seguros y de alto rendimiento. En Funcionamiento simplificado: c o m a n d o
 Carro para rack modelos con una o dos puertas y con capacidades que microprocesado e interfaz intuitiva con pantalla
 Cestos ISO
Componentes para instalação van desde 140 a 890 litros, están destinados al sector táctil colorida. Facilita el cumplimiento por parte
 Cavalete de água  Cavalete Vapor de la esterilización de materiales en hospitales y otros del usuario de los requisitos del proceso.
 Abrandador  Quadro Elétrico centros de salud, laboratorios médicos, laboratorios Versatilidad: Flexibilidad en la programación de
 Osmose
de control de calidad y para procesos de producción procesos. Acceso a la programación de equipos y
 Filtro sanitário
 Compressor de ar de industrias farmacéuticas y otras industrias. Su funciones a través de contraseñas de usuario con
 Cavalete ar uso es adecuado para la esterilización de material niveles de permisos configurables.
Servicios poroso empaquetado, instrumental y envasado o sin Procesos Sostenibles: Equipos ambientalmente
 Instalación
 QI (Qual.Instal.) envasado utensilios, cristalería, guantes, jeringas y limpios y seguros, sin residuos al final del ciclo.
 QD (Qual.Des) otros materiales en cauchos o aleaciones de plásticos Cámaras, puertas y tuberías de transporte de vapor
 Manutención resistentes a altas temperaturas, líquidos en botellas con aislamiento térmico. Ahorro de energía, menor
 Start Up
de vidrio con cierre ventilado y otros. disipación de calor y menor carga térmica para el
 QO (Qual.Oper.)
 Capacitación Poden aun ser fornecidos con ciclo de formaldehido sistema de aire acondicionado.
 Kit piezas repuesto para esterilización a baja temperatura de aquellos Instalación y Mantenimiento: Facilitado por
materiales termosensibles. panel con componentes eléctricos en riel y acceso
a componentes desde la parte delantera, trasera
CARACTERÍSTICAS PRINCIPALES y lateral.
Alto rendimiento: garantiza procesos de
esterilización fiables, precisos y válidos.

17/24
Esterilizador a Vapor HI VAC MX II
MODELOS Y CAPACIDADES Alternativas para la Cámara Material ALTERNATIVAS PARA ACABADOS
La familia HI VAC MX II está formada por esterilizadores Gabinete frontal
· AISI 316 L – 3/16” (4,76 mm)
pequeños, medianos y grandes con las siguientes
capacidades: · Acero inox AISI 304 cepillado
· AISI 316 L – 5/16” (7,93 mm)
B0112-140 - Capacidad de 140 l · Acero inox sin níquel AISI 444 cepillado
· AISI 316 TI – 5/16” (7,93 mm)
B0112-200 - Capacidad de 200 l
· Acero carbono con protección anticorrosiva
B0112-263 - Capacidad de 263 l Acabados
y pintura
B0112-370 - Capacidad de 370 l
· Pulido Sanitario Acabado lateral
B0112-418 - Capacidad de 418 l
B0112-542 - Capacidad de 542 l · Electropulido · Desmontable en acero inox AISI 304 cepillado
B0112-716 - Capacidad de 716 l · Desmontable en acero inox sin níquel AISI
B0112-890 - Capacidad de 890 l Estructura: El conjunto de la cámara está montado 444 cepillado
en el marco de acero inoxidable AISI 304 y la altura
NORMAS Y LEGISLACIÓN · Desmontable en acero carbono con protec-
ajustable del pie para permitir una nivelación perfecta
ción anticorrosiva y pintura
· EN ISO 17665-1:2006 - Sterilization of health care del equipo.
products - Moist heat - Part 1: Requirements for Tuberías: los tubos soldados en la cámara están · Articulado en acero inoxidable AISI 304 cepil-
the development, validation and routine control of a construidos de acero inoxidable AISI 316 e conexiones lado. Excepto equipos de 140 y 200 l.
sterilization process for medical devices. triclamp (rugosidad Ra <0,375µm). · Articulado en acero inox sin níquel AISI 444
· EN ISO 15223-1:2016 - Medical devices — Symbols Los tubos en el revestimiento de la cámara son cepillado. Excepto para equipos de 140 y 200 l
to be used with medical device labels, labelling de acero inoxidable AISI 316 y aislados por tubo Acabado superior
and information to be supplied — Part 1: General de silicona con fibra de vidrio en el interior. Las
requirements. conexiones son roscadas o opcionalmente triclamp · Acero inox AISI 304 cepillado
· NBR 11816:2003 - Esterilizadores a vapor com (rugosidad Ra <0,375µm). · Acero inox sin níquel AISI 444 cepillado
vácuo, para produtos de saúde.
· Acero carbono con protección anticorrosiva
· NBR 15659:2009 – Esterilização de produtos para Alternativas para tuberías y pintura
saúde – Esterilizadores de vapor a baixa temperatura e Fuera do recubrimiento
formaldeído – Requisitos e métodos de ensaio. Acabado trasero (una porta)
· Acero inoxidable AISI 316
· ASME, Section VIII, Division I - Boiler and pressure · Desmontable en acero inox AISI 304 cepillado
vessel code. · Latón/Bronce
· Desmontable en acero inox sin níquel AISI
· NR 13 - Caldeiras e Vasos de Pressão. 444 cepillado
El conjunto de la cámara está protegido por una
CONSTRUCCIÓN válvula de seguridad construida en acero inoxidable · Desmontable en acero carbono con protec-
Cámara Esterilización: Construido en doble pared ción anticorrosiva y pintura
AISI 316, sellada y con un mecanismo de verificación,
(Cámara Interior y Cámara Externa), con pliegue con una conexión roscada y un certificado de Hermeticidad hospitalar (doble puerta)
central y desniveles, para facilitar la eliminación de calibración. · Acero inox AISI 304, atornillado
condensado, optimizando el proceso. Cámara interior
Hermeticidad biológica
de perfil rectangular construida en acero inoxidable Alternativas p/ Válvula Seguridad
AISI 316L, con espesor de 3/16” (4,76 mm) y · Acero inox AISI 304, soldada
acabado sanitario pulido. Cámara externa construida · Acero inoxidable AISI 316
en cogollos en forma de “U” invertida con el mismo · Hierro hundido/Bronce Puertas
material de la cámara interna. El conjunto se ajusta al El equipo se puede suministrar con una o dos puertas
estándar ASME para admitir lo siguiente presiones: Acabado y Hermeticidad cuando se instala en una barrera sanitaria, con apertura
· Presión Máxima de Trabajo de 3,0 kgf/cm². Gabinete frontal y acabados laterales removibles, y cierre automáticos, activados a través de la interfaz
· Presión de prueba hidrostático de 4,5 kgf/cm² - 1,5 fabricados en acero inoxidable AISI 304 y de comando. La puerta se puede mover verticalmente
veces la PMTA. acabado cepillado. El estándar, el equipo se a través de un par de pistones neumáticos, subiendo
Dos entradas de validación independientes, con un comercializa sin los acabados superior y trasero. al cerrarse y bajando detrás del gabinete en la abertura.
diámetro de 1”, están disponibles en el lateral del Un panel (opcional) unido a la cámara del equipo La presión de aire de los pistones se puede ajustar
equipo, lo que permite la introducción de sensores completa el sello entre los entornos donde se para permitir un movimiento suave de la puerta y un
de temperatura y presión en los procesos operativos anclará el esterilizador. Este sello puede cumplir dispositivo de seguridad detiene este movimiento
y de calificación de rendimiento. Externamente, una con los requisitos herméticos como lo son los siempre que haya una obstrucción involuntaria o no,
capa de fibra mineral libre de cloruro recubierta con requisitos de la industria farmacéutica entre para proteger al operador.
placa de acero inoxidable AISI 430 con acabado pulido áreas controladas o simplemente garantizar el Las cerraduras de seguridad mecánicas evitan que las
disminuye la condensación de vapor y la irradiación de mantenimiento de la presión diferencial entre puertas se abran durante el proceso de esterilización
calor. Drenaje de cámara interior protegido con filtro áreas contiguas, como lo son los requisitos y la apertura simultánea de las puertas de carga y
de placa perforada de acero inoxidable con diámetro para los hospitales. descarga en el caso de equipos con dos puertas en
de 1.1/2”, situado cerca de la cara del lado de carga. las fases de carga y descarga del equipo.
18/24
Esterilizador a Vapor HI VAC MX II
Sellando la puerta con una junta de silicona alojada en · Bomba de Vacío Seca. El componente fun- comando e instrumentación con una fuente conmutada
el canal de la cámara, que en el cierre presurizado se ciona sin el uso de agua u otro fluido líquido, de salida en voltaje extra bajo de 24 Vcc y un panel
cierra contra la cara interna de la puerta y en la abertura lo que permite un uso ambientalmente correcto. de alimentación con protección para los elementos de
retraída mediante un pulso de vacío. Tanto la cara de carga mediante disyuntores. Opcionalmente, el panel
· Potencia da bomba de vacío: depende del
la puerta como la del canal están mecanizadas para volumen de la cámara y las opciones de configu- de alimentación se puede proporcionar con protección
un ajuste adecuado. ración disponibles para cada modelo garantizan interruptor diferencial (dispositivo ID).
La (s) puerta (s) tiene una parte interna de acero la finalización exitosa de Bowie & Dick y las
inoxidable AISI 316L. Aislamiento interno realizado con pruebas de carga hueca tal como se define en los CONTROL Y INSTRUMENTACIÓN:
una manta de fibra mineral libre de cloruro, estructura estándares nacionales e internacionales Panel de control: dispuesto en altura que facilita
interna a través de perfiles de acero y acabado externo Bombas de Vacío disponibles la visualización y el funcionamiento del lado de carga
con placa de acero inoxidable AISI 304 con acabado del esterilizador.
· Bomba de Vacío Una Etapa
cepillado. PLC industrial controlado por microprocesador
· Bomba de Vacío 1,5 cv. dedicado con rutinas de autocomprobación,
ALTERNATIVAS PARA PUERTAS Para esterilizadores de 140, 200, 263, 370, 418 autodiagnóstico y supervisión de ciclos por “watchdog
y 542 l.
timer” (tiempo de vigilancia) de acuerdo con la
· Una porta
· Bomba de Vacío 3,0 cv. normativa vigente.
· Doble puerta Para esterilizadores de 140, 200, 263, 370, 418 La interfaz hombre-máquina (HMI) con pantalla
Material (parte interna) y 542 l. táctil a color de 7 “, grado de protección IP65, permite
· Bomba de Vacío 4,0 cv. monitorear y controlar todas las funciones del equipo y
· Acero inoxidable AISI 316L
Para esterilizadores de 263, 370, 418, 542, 716 parámetros de ciclo. Con puerto USB para actualización
· Acero inoxidable AISI 316TI y 890 l. de software.
Acabado · Bomba de Vacío 5,0 cv. Puerta Ethernet opcional permite el acceso remoto
Para esterilizadores de 716 y 890 l. con una interfaz reflejada para monitorear el proceso.
· Pulido Sanitario
La interfaz le permite seleccionar e iniciar un ciclo
· Electropulido Bomba de Vacío Dos Etapas
de esterilización, seguir los pasos del proceso con la
· Bomba de Vacío Dos Etapas 3,0 cv. visualización de la temperatura, mediciones de presión,
Entrada de aire limpio: para romper el vacío a Para esterilizadores de 140, 200, 263, 370, 418, tiempo estimado para finalizar la fase; gestionar
través de un filtro hidrofóbico esterilizable con una 542, 716 l. usuarios y contraseñas con la posibilidad de registrar
eficiencia del 99.9997% para partículas mayores o · Bomba de Vacío Dos Etapas 5,0 cv. 40 usuarios con niveles de acceso configurables;
iguales a 0.22 µm. Para esterilizadores de 890 l. configurar alerta de fin de ciclo; ver alarmas de falla
Termocambiador: Elemento opcional, por placas Bomba de Vacío Seca y seguridad; cambiar fecha y hora del sistema; ajuste
en acero inoxidable AISI 316, para enfriar los fluidos de pantalla; acceso al registro de las últimas alarmas
· Bomba de Vacío Seca 3,0 cv.
enviados al sistema de efluentes. con la posibilidad de exportar estos datos; y elección
Para esterilizadores de 263, 370, 542, 716 y 890
l. del lenguaje operativo.
Alternativas Termocambiador La interfaz también proporciona acceso a las rutinas
de operación para programar los parámetros del
· Sin termocambiador
GENERACIÓN DE VAPOR ciclo en las bandas de seguridad definidas en el
· Con termocambiador. Excepto para equipos El vapor utilizado en la esterilización de materiales se proyecto, acceso a la operación manual, calibración
de 140, 200 l o equipados con bomba de vacío suministra preferiblemente a través de un generador de y mantenimiento utilizando el inicio de sesión y
seca.
vapor eléctrico incorporado en el equipo y alimentado la contraseña del usuario, y la programación de
por agua con la calidad adecuada para el proceso. actividades de mantenimiento de rutina. dependiendo
Sistema de vacío del número de ciclos u horas de operación.
Para llevar a cabo los pulsos de vacío en el Alimentación de vapor - Alternativas Sistema de supervisión (opcional) compuesto
acondicionamiento de la carga al comienzo del ciclo por un segundo PLC que recibe datos de sensores
· Preparado para recibir un generador de vapor
y secar en la etapa final del proceso, a través de una eléctrico. de temperatura independientes comparándolos con
bomba de vacío de una sola etapa, tipo monobloque las lecturas del sensor de temperatura del controlador
con anillo de agua, dimensionada para cumplir con los · Sin generador de vapor. Vapor suministrado principal. En caso de desviación entre lecturas, se
desde una fuente externa.
requisitos de los estándares actuales. activa una alarma desde el sistema de supervisión
· Preparado para recibir generador de vapor por al controlador principal, iniciando una rutina de
Opciones para Sistemas de Vacío bobina. seguridad.
NOTA: El generador debe configurarse como un
· Bomba de Vacío Una Etapa.
elemento separado. Ver detalles constructivos de los Alternativas para Control
· Bomba de Vacío Dos Etapas. generadores en el ítem “componentes opcionales”.
Tipo monobloque con anillo de agua y dos etapas · Sin Puerta Ethernet
de succión interconectadas. · Con Puerta Ethernet
PAINEL ELÉTRICO
El equipo tiene dos paneles eléctricos, uno para
19/24
Esterilizador a Vapor HI VAC MX II
· Sin Sistema Supervisor El control de temperatura mediante un par de · Pulso de vacío único con profundidad programada
sensores PT-100 clase A, ubicados junto al drenaje en el rango de 0,50 y 0,90 bar.
· Con Sistema Supervisor
de la cámara, lee la temperatura del proceso. Las · Pulso de vapor con rango de programación entre
Panel Secundario: en el lado estéril (descarga), en indicaciones de temperatura se realizan digitalmente 1.0 y 2.0 bar.
equipo de doble puerta, equipado con un manómetro a través de la interfaz del panel en OC. · Escaneo, con un tiempo programable entre 01 y
con glicerina para leer la presión de la cámara interna, 9999 minutos.
botón de operación de la puerta de descarga, botón Sensores de Temperatura Fase 2 - Calentamiento da carga
de emergencia y lámparas para indicar el ciclo en Mediante inyección de vapor hasta alcanzar la
· Sensor de temperatura
progreso, ciclo completo, puerta de carga cerrada y temperatura de esterilización en la cámara y la carga
alarma activa. · Sensor de temperatura certificado a esterilizar. Sin parámetros programables. El equipo
Control de proceso realizado por el PLC principal controla la presión y la temperatura en la cámara de
con datos de temperatura enviados por los 2 sensores Sensor de carga - Opcional acuerdo con la temperatura de esterilización definida
de temperatura PT-100 ubicados al lado del drenaje, Sensor para medir la temperatura en la carga, para el proceso.
uno utilizado para el control del proceso y el otro para disponible como opción de control, principalmente en Fase 3 - Exposición
verificar los datos de temperatura, y con los datos de ciclos de esterilización líquida. Con cable flexible para Incluye un intervalo de tiempo de hasta 60 segundos
presión enviados por los transductores La cámara el mejor posicionamiento dentro de la carga. para que la temperatura se asiente dentro del rango
interior y la cámara exterior. El PLC tiene control interno definido para el proceso (temperatura de esterilización
· Sin sensor
de los tiempos de proceso. + 2 OC) y el tiempo umbral para mantener la temperatura
En la fase de esterilización, el PLC verifica la condición · Con sensor del proceso en ese mismo rango hasta el final de la
de saturación del vapor, a través de un algoritmo · Con sensor certificado fase . Los parámetros para esta fase son:
patentado con los datos de presión y temperatura · Temperatura de esterilización con un rango de
medidos en la cámara de esterilización. Sensor para Cámara Externa - Opcional Sensor programación entre 101 y 135 OC.
Las diferencias entre las temperaturas medidas en para acompañamiento da temperatura en la Cámara · Tiempo de exposición con un rango de programación
los sensores de control y verificación, fuera del rango externa. entre 1 y 9999 minutos.
definido en los estándares y el estado de saturación de Fase 4 - Escape
· Sin sensor
vapor inadecuado para el proceso, emiten una alarma Compuesto por liberar presión de la cámara y enfriar
e interrumpen el ciclo de esterilización. · Con sensor los fluidos que se enviarán al sistema de drenaje. Sin
La interfaz que informa al operador puede seguir todo · Con sensor certificado parámetros de programación.
el proceso de esterilización en tiempo real: Fase 5 - Secado
· Temperatura de la cámara interna; CICLOS DE ESTERILIZACIÓN - PERFIL Con una nueva fase de vacío para eliminar la humedad
· Temperatura de la cámara externa (opcional que La línea de esterilizadores HI VAC MX II se puede de la cámara y secar la carga. Los parámetros para
requiere el sensor para la CE); suministrar con dos programas alternativos y distintos esta fase son:
· Presión de la cámara interna; con perfiles con ciclos configurables. En ambos ciclos · Pulsos de secado con un rango de programación
· Presión de la cámara externa; con densidad y perfiles líquidos están disponibles. El entre 0 y 9999.
· Tiempo transcurrido desde la fase de exposición / “Programa C” se carga de serie en los esterilizadores · Nivel de vacío que asume el valor establecido para
esterilización; y proporciona al usuario 9 ciclos de esterilización. los pulsos de vacío iniciales como estándar.
· Tiempo transcurrido para terminar la fase de secado; El “Programa D” es opcional e incluye, además de · Tiempo de secado con rango de programación entre
· Información del estado del ciclo; todos los ciclos del “Programa C”, una serie de 1 y 9999 minutos.
· Visualización gráfica en tiempo real de datos de ciclos opcionales con un perfil diferente y rangos de Fase 6 – Aeración
temperatura y presión. configuración diferentes, disponibles en la cantidad de Con la entrada de aire filtrado a través del filtro
El panel también contiene los instrumentos de ciclos deseados por el cliente. Consulte los detalles hidrofóbico, la ruptura del vacío y la igualación de la
seguridad y control que se indican a continuación: del programa “D”. presión de la cámara con el entorno hacen posible abrir
· Tecla de encendido y apagado; la puerta de descarga. Sin parámetros programables.
· Botón de emergencia; CICLOS MATERIAL DE DENSIDADE
· Manovacuómetro de la cámara interna; Ciclos de perfil para la esterilización de materiales CICLOS PARA LÍQUIDOS
· Manómetro externo de la cámara; textiles, cristalería, instrumentos metálicos embalados Ciclos de esterilización con perfiles para líquidos
· Manómetro del generador de vapor o vapor externo; en cajas o contenedores perforados, dispositivos en botellas abiertas. Con una programación
· Impresora para registrar los parámetros alcanzados de lumen, materiales termosensibles en látex y flexible, la definición de los parámetros depende
en el proceso. / o plásticos con resistencia a la temperatura de fundamentalmente de las características del
Instrumentación y controle esterilización. El ciclo consta de 06 fases distintas: líquido y del volumen de la botella donde está
Control de presión de las cámaras internas y Fase 1 - Extracción de aire y acondicionamiento contenido. Estas variables definen la temperatura
externas mediante transductores electrónicos de A través de pulsos alternos de vacío y vapor. Esta del proceso, el tiempo y las características de la
presión de acero inoxidable, con compensación fase solo se puede configurar en el ciclo Bowie & fase inicial de eliminación de aire.
electrónica de temperatura y lectura de presión Dick y esta configuración se carga automáticamente El ciclo permite al usuario elegir el sensor
absoluta. Las indicaciones de presión se realizan en otros ciclos con este perfil. Los parámetros para central, si hay uno disponible, para el control
digitalmente a través de la interfaz del panel en bar. esta fase son: del proceso.
20/24
Esterilizador a Vapor HI VAC MX II
Fase 1 – Extracción de aire y acondicionamiento segundos, configurable en los rangos de programación minutos y temperatura de final de ciclo de 95 OC. Todos
Para líquidos en matraces abiertos, la extracción de disponibles. los parámetros configurables en los rangos disponibles
aire comienza con un pulso de vacío superficial para Temperatura de 134 OC, tiempo de exposición de 210 para ciclos de líquido.
evitar que el líquido se escape de los matraces y se segundos y tiempo de secado de 01 minutos, según Ciclo 9 – Usuario B
complementa con un barrido de vapor, donde la entrada norma específica, sin posibilidad de alteración. Ídem al ciclo 5.
de vapor y la válvula de drenaje permanecen abiertas, Ciclo 7 – “Leak test” Ciclo para Lactario
para Deje que el vapor arrastre el aire residual en la Para comprobar la estanqueidad de la cámara. Ciclo Tanto el ciclo 9 como el ciclo 5 pueden programarse
cámara al desagüe. Los parámetros programables de no programable, con parámetros fijos de fábrica. Pulso como ciclos de lactancia. El nombre del ciclo se puede
esta fase son: de vacío de hasta 0,20 bar, intervalo para acomodar la cambiar a través de la interfaz de comando.
· Pulso de vacío único con profundidad programada presión de 05 minutos y controlar la presión desde el Los parámetros indicados para el ciclo de lactancia
en el rango de 0,50 y 0,90 bar. nivel estabilizado durante 10 minutos. La prueba se son: pulso único e inicial de 0,50 bar; 1.15 bar de
· Pulso de vapor con rango de programación entre considera “aceptada”, cuando el aumento de presión pulso de escaneo; tiempo de escaneo de 10 minutos;
1.0 y 2.0 bar. es inferior a 0,013 bar. temperatura de esterilización 110 OC, tiempo de 10
· Escaneo, con un tiempo programable entre 01 y Ciclos con perfil de densidad. minutos y temperatura de final de ciclo 95 OC. Todos
9999 minutos. Ciclo 1 – Textiles los parámetros configurables en los rangos disponibles
Fase 2 - Calentamiento da carga Fases de eliminación y acondicionamiento de aire, para ciclos de líquido.
Con inyección de vapor hasta alcanzar la temperatura con los parámetros definidos para la prueba de Programa “F”
de esterilización, en la cámara. Control de fase a través B&D. Temperatura 134 OC, tiempo de exposición Compuesto por el programa “C” y un ciclo de baja
de mediciones de temperatura y presión de la cámara. de 04 minutos y tiempo de secado de 15 minutos, temperatura con vapor de formaldehído.
Fase 3 – Exposición configurable en los rangos disponibles de cada Ciclo 10 - formaldehído
El tiempo para alcanzar la temperatura en la cámara parámetro. Indicado solo para los modelos B0112-263 (P), B0112-
depende básicamente de los volúmenes de líquido. Ciclo 2 - Termosensible 370 (P) y B0112-542 (P). Proceso totalmente validable,
Cuando hay un sensor central, es posible determinar Fases de eliminación y acondicionamiento de aire, compatible con 29 tipos de polímeros, además de ser
que el inicio del recuento del tiempo de exposición con los parámetros definidos para la prueba de más económico que el proceso de óxido de etileno y
estará definido por las mediciones de ese sensor. B&D. Temperatura 121OC, tiempo de exposición permitir una mayor seguridad para los pacientes y el
Sin el sensor, el tiempo para alcanzar la temperatura de 20 minutos y tiempo de secado de 20 minutos, equipo de enfermería. Como estándar, el ciclo tiene
se define en la validación del proceso y se agrega al configurable en los rangos disponibles de cada una fase de acondicionamiento de carga a través de
tiempo de exposición. parámetro. pulsos de vacío y vapor y pulsos de vacío y vapor de
Los parámetros programables de esta fase son: Ciclo 3 – Rápido (flash) formaldehído, fase de exposición a una temperatura
· Definición del sensor de control: entre el sensor Para material desempacado o embalado individualmente. de 50 o 60 OC, lavado con pulsos de vacío y vapor
ubicado en el drenaje y el sensor de carga en la cámara, Fases de eliminación y acondicionamiento de aire, de agua y fase secado mecánico al vacío, este ciclo
cuando esté disponible. con los parámetros definidos para la prueba de no permite configuraciones de usuario. La solución
· Temperatura de esterilización: con un rango de B&D. Temperatura 134 OC, tiempo de exposición de formaldehído está registrada en ANVISA, junto
programación entre 101 y 135 OC. de 04 minutos y tiempo de secado de 05 minutos, con el equipo.
· Tiempo de exposición: después de alcanzar la configurable en los rangos disponibles de cada Programa “D”
temperatura cuando se usa el sensor de control, o el parámetro. El “Programa D” es opcional e incluye, además de
tiempo de exposición validado en ausencia del sensor Ciclo 4 – Instrumental todos los ciclos del “Programa C”, una serie de
de control. Con rango de programación entre 01 y Fases de eliminación y acondicionamiento de aire, ciclos opcionales con un perfil diferente y rangos de
9999 minutos. con los parámetros definidos para la prueba de configuración diferentes, disponibles en la cantidad
· Control del ciclo por F0: con la definición de la B&D. Temperatura 134 OC, tiempo de exposición de ciclos deseados por el cliente. Consúltenos para
letalidad deseada equivalente a una exposición a 121 de 04 minutos y tiempo de secado de 30 minutos, obtener detalles del programa “D”.
O
C. Con rango de programación entre 01 y 9999. configurable en los rangos disponibles de cada
Fase 4 – Resfriamiento parámetro. OTROS INSTRUMENTOS
Con la lenta liberación de presión de la cámara para Ciclo 8 – Usuario A Manovacuómetro para indicar el nivel de vacío o
evitar que el líquido hierva, hasta que la temperatura Permite la configuración de la fase de eliminación presión de la cámara interior. Con tubo de bourdon
se establezca para el final del ciclo. Es un parámetro de aire independientemente de los valores definidos de DN 63 mm, mecanismo de sistema de acero
programable de esta fase: para Bowie & Dick. Temperatura de 134 OC, tiempo inoxidable con una escala de 0 a 6 bar absolutos,
Temperatura de fin de ciclo: con rango de programación de exposición de 4 minutos y tiempo de secado de calibrado y lleno de glicerina para la estabilización.
entre 80 y 99 OC 30 minutos. Todos los parámetros, incluidos los de Manómetros para indicación de presión,
la fase de eliminación de aire, configurables en los uno para una cámara externa y otro para el
CICLOS ESTERILIZACIÓN DISPONIBLES rangos permitidos para cada parámetro. generador de vapor. Con tubo de bourdon
Ciclos de teste: Ciclos con perfil para líquidos: de DN 63 mm, mecanismo de sistema de
Ciclo 6 - Bowie & Dick Ciclo 5 – Líquidos acero inoxidable con una escala de 0 a 6 bar
Para verificar la fase de eliminación de aire. Con Pulso único y vacío inicial hasta 0,50 bar; presión de absolutos, calibrado y lleno de glicerina para
4 pulsos de vacío de hasta 0,20 bar, presión de escaneo de 1,15 bar; tiempo de escaneo de 10 min; la estabilización.
vapor de 1,15 bar y un tiempo de espera de 30 temperatura de esterilización de 121 OC; tiempo de 30

21/24
Esterilizador a Vapor HI VAC MX II
Alternativas para Manómetros Válvulas de control de proceso con accionamiento Alternativas para Generador de Vapor
neumático independiente en acero inoxidable AISI 316.
· Manómetro calibrado · Generador de Vapor Eléctrico
Controlado por electroválvulas de aire comprimido, en
· Manómetro calibrado con certificación RBC las líneas de suministro de vapor de las dos cámaras, · Generador de Vapor por Serpentina
las líneas de vacío y escape, y la entrada de aire al Válvula Seguranza do Generador
Impresora térmica de 40 columnas, que consta de final del proceso.
· En acero inox AISI 316
un cuerpo termoplástico de alta resistencia y tecnología Otros controles, como la presurización y el vacío de
de impresión térmica, bobina de papel protegida por los canales de la puerta y la entrada de agua para las · En hierro hundido o bronce
una tapa, interfaz en serie e instalada en el panel frontal bombas a través de válvulas solenoides. Conmutador de vapor
del esterilizador. Válvula PID con cuerpo de acero inoxidable AISI Sistema para cambiar el suministro de vapor, operado
Los datos impresos aseguran una documentación 316L, con actuador neumático y controlador PID electrónicamente a través de la interfaz de control del
completa del proceso de esterilización con información integrado. Permite, con la asociación de tecnologías esterilizador. Para reemplazar el suministro de vapor
como: neumáticas y electrónicas, precisión en el control de de un generador integrado en el equipo a una fuente
· Nombre del establecimiento; identificación del flujo y presión en el suministro de vapor, facilitando de suministro central. Conexión con rosca de acero
esterilizador; fecha; hora de inicio del ciclo; número el ajuste de la tasa de aumento de temperatura en la inoxidable o latón según la elección del material para
de ciclo; código de carga cámara y eliminando las fluctuaciones de temperatura las tuberías del equipo.
· Número y nombre del programa con parámetros a durante el proceso de esterilización.
alcanzar en el proceso. Ciclo de información de control Alternativas Conmutador de vapor
por F0, cuando corresponda. Alternativas Válvula de Controle
· Sin conmutador de vapor
· Información de las fases realizadas, con registro Válculas Neumáticas
de tiempo en minutos y segundos y los datos de · Con conmutador de vapor
temperatura correspondientes del sensor de control, · Em acero inoxidable AISI 316
Bomba de alimentación do generador
temperatura del sensor de verificación, presión · En liga de bronce
· Con tratamiento de níquel cromo
medida por el transductor CI y, si está disponible, la Válvula PID
temperatura del sensor central en la carga. · En acero inox AISI 316 o similar
· Al final del ciclo, la información completa del ciclo · Sin válvula PID
y el tiempo del proceso. Cuando se controla con “F0”, · Con válvula PID en inox AISI 316L La capacidad del generador de vapor influye en el
se imprimirá el valor correspondiente de “F”. tiempo del ciclo de esterilización. Para satisfacer las
· La ocurrencia de fallas en la ejecución del ciclo. COMPONENTES OPCIONALES diferentes demandas y la disponibilidad de energía
eléctrica en los sitios de instalación, existen diferentes
Alternativas para impresoras SISTEMA DE GENERACIÓN DE VAPOR potencias disponibles para el mismo modelo de
Generador de vapor eléctrico con suministro equipo.
· Impresora térmica
automático de agua, a través de una bomba centrífuga
· Impresora matricial con tratamiento de níquel cromo. El cuerpo del Potencia generador de vapor
generador está construido en acero inoxidable AISI Generadores Eléctricos
Impresora matricial con 40 columnas, que consta 316L y resistencias blindadas en acero inoxidable
de un cuerpo termoplástico de alta resistencia y una AISI 316. · Con potencia de 18 kW.
Para esterilizadores de 140 y 200 l.
cubierta protectora de carrete de papel y cabezal de Control del nivel de agua por sistema de flotador de
impresión con mecanismo de impresión de impacto acero inoxidable, para la detección del nivel mínimo y · Con potencia de 24 kW.
rápido de ocho agujas. Con interfaz en serie e instalado máximo de agua en el generador. Para esterilizadores de 263 y 370 l.
en el panel frontal del esterilizador. Drenaje parcial y automático realizado al final de cada · Con potencia de 36 kW.
Grabador gráfico (opcional) de alta calidad para ciclo, controlado por temporizador para disminuir la Para esterilizadores de 263, 370, 418 y 542 l.
la adquisición de datos como tiempo, temperatura y concentración de minerales en el agua. El generador
· Con potencia de 44 kW.
presión. Tiene seis canales configurables que permiten tiene un interruptor selector para el drenaje total del Para esterilizadores de 263, 370, 418 y 542 l.
la visualización de medidas, escalas o barras gráficas generador para limpiar el fondo.
y su interfaz es la pantalla táctil. El funcionamiento Válvula de seguridad con certificado de calibración, · Con potencia de 52 kW.
Para esterilizadores de 263, 370, 418 y 542 l.
de la grabadora gráfica se puede realizar en el equipo en acero inoxidable AISI 316 y conexiones roscadas,
u opcionalmente de forma remota a través de la sellada y con mecanismo de verificación de · Con potencia de 72 kW.
comunicación ETHERNET. Los datos se pueden guardar funcionamiento, protege el grupo electrógeno. Para esterilizadores de 716 y 890 l.
a través de USB. Construido de acuerdo con la normativa vigente. Generador por Serpentina a Vapor
Generador de vapor serpentino Construido
· Capacidad de 100 kg/h de vapor.
Alternativas Grabador Gráfico en acero inoxidable AISI 316L y sistema de Para esterilizadores de 263, 370, 418 y 542 l.
calentamiento de serpentín de acero inoxidable AISI
· Sin Grabador Gráfico · Capacidad de 100 kg/h de vapor.
316, alimentado por vapor industrial. Comando,
· Con Registrador Gráfico Para esterilizadores de 263, 370, 418 y 542 l.
controles y componentes iguales al modelo
eléctrico.
22/24
Esterilizador a Vapor HI VAC MX II
ALIMENTACIÓN AIRE COMPRIMIDO · Rutina de emergencia que, en ausencia de Alimentación de agua para generador de vapor
El equipo requiere aire comprimido para sellar puertas y electricidad, evita el bloqueo de las puertas y la · Dureza inferior a 10 ppm de CaCO3.
mover válvulas. Para el suministro de aire comprimido, liberación de presión de la cámara interna; · Conductividad inferior a 50 µS/cm.
hay un compresor de aire disponible, montado fuera · Sistema de alarma por falla en el suministro de agua, · Temperatura del agua entre 5 y 35OC.
de la estructura del equipo. Con depósito de 50 o 100 vapor y aire comprimido; · pH entre 5 y 8.
litros y capaz de suministrar aire comprimido con 8.0 · Botón de emergencia en el panel del equipo para · Cloruros < 3 mg/l.
kgf/cm² de presión. apagar el esterilizador y despresurizar la cámara de · Presión dinámica: mínimo 1 bar absoluto.
Alternativas para aire comprimido esterilización;
· Cerradura mecánica que evita que la puerta se abra NOTA: Los caudales requeridos varían según la
· Ar comprimido de rede
cuando el esterilizador está en funcionamiento. configuración y el modelo del equipo. Consulte la Guía
· Con Compresor de ar de instalación para obtener más información.
NOTAS
OSMOSIS INVERSA El ancho total del equipo desempaquetado se indica ALIMENTACIÓN DE VAPOR
Sistema de tratamiento de agua opcional para el en la tabla de datos de ingeniería. Para facilitar el El suministro de vapor directo a la cámara de la
suministro del generador de vapor, equipado con 3 movimiento del equipo dentro del edificio, se puede autoclave no es una solución adecuada para cumplir
etapas de filtración - con una tasa de retención de 5 y desmontar el frente, reduciendo el ancho. Es importante con los requisitos de las Buenas Prácticas para los
1 micra, y carbón activado - bomba de presurización, verificar la ruta que tomará el esterilizador dentro del procesos de esterilización de materiales.
membranas de ósmosis, control y depósito de agua edificio para verificar si hay cuellos de botella. La Alimentación de Vapor directa
tratada. liberación de estos puntos es responsabilidad del · Calidad según EN 285 y NBR ISO 17665 parte 2.
comprador; · Presión de 3,5 a 3,7 bares absolutos.
Alternativas para Osmosis Inversa Los medidores de las conexiones y líneas de suministro · Título de Vapor superior a 0.95.
indicadas en Datos técnicos se refieren a los puntos de NOTA: Baumer no es responsable de las manchas
· Sin sistema de osmosis
conexión con el esterilizador. Las tuberías que llevan en instrumentos y tejidos y de la corrosión en
· Con sistema de osmosis para 20 l/h para es- estos suministros a los sitios de instalación deben instrumentos y partes de equipos en el caso de
terilizadores de 140 hasta 370 l ser calculadas por terceros y el dimensionamiento suministro directo de vapor fuera de los estándares
· Con sistema de osmosis para 60 l/h para es- depende de la distancia, las presiones de suministro sugeridos por los estándares anteriores.
terilizadores de 418 hasta 890 l y los flujos necesarios; La mejor alternativa es suministrar vapor industrial a
un generador de vapor de bobina.
SEGURANZA DATOS DE INGENIERÍA Alimentación de Vapor para generador de vapor
O equipo presenta los siguientes ítems de seguranza: DRENAJE: por serpentina
· En ausencia de electricidad, la entrada de vapor La línea de drenaje debe ser independiente de la · Presión 7 a 9 bar absoluto.
está cerrada; alcantarilla del edificio y estar conectada a la caja de · Título de Vapor superior de 0,95.
· Cuando hay un exceso de presión, el paso de vapor paso externa. La conexión del equipo con la línea de
está cerrado; alcantarillado del edificio solo se indica cuando el ALIMENTACIÓN AIRE COMPRIMIDO
· Válvula de seguridad estandarizada y calibrada a equipo tiene un intercambiador de calor para enfriar los Aire comprimido libre de impurezas y humedad y con
3.0 kgf/cm² con dispositivo de limpieza y verificación efluentes liberados en el proceso. Tubería de material una presión de suministro de 6 a 7 bar. Conexión
de funcionamiento; resistente al calor. rápida de 6 mm.
· Inicio del ciclo solo en presencia de presión de vapor ABASTECIMIENTO DE AGUA
en el generador o en la línea de suministro; Los esterilizadores tienen entrada de agua para el ALIMENTACIÓN ELÉCTRICA
· Sistema automático de elevación de puerta con generador de vapor y el sistema de vacío. En la entrada La línea HI VAC MX II tiene una placa de distribución
presión de trabajo calibrada, para proteger al operador; de agua del generador, se puede conectar un sistema interna con conexión trifásica y a tierra con una
· Puerta con sistema anti-aplastamiento que de ósmosis inversa para tratar el agua utilizada para impedancia máxima de 10 ohmios y protecciones para
interrumpe el movimiento de la puerta si está obstruida; generar vapor. cada elemento de carga por interruptor automático.
· Imposibilidad de abrir las puertas después del Alimentación de agua para Bomba de Vacío
comienzo del ciclo y en presencia de presión de vapor · Dureza máxima de 250 ppm de CaCO3. NOTA: Consulte la Guía de instalación para identificar
en la cámara interna; · Temperatura ideal por debajo de 15OC y máximo las protecciones necesarias para la configuración de
· El ciclo comienza solo con puertas cerradas y permitido 25OC. cada equipo.
revestimientos presurizados; · pH de 5 a 8.
· Imposibilidad de abrir las dos puertas · Presión entre 1.2 y 1.5 bar absolutos.
simultáneamente por el operador (en el caso de Alimentación de agua para Osmosis Inversa
esterilizadores de barrera); · Dureza inferior a 60 ppm de CaCO3.
· Aborto del ciclo cuando la temperatura de la cámara · Conductividad inferior a 500 µS/cm.
en la fase de esterilización es inferior a la definida para · Temperatura del agua por debajo de 25OC.
el proceso; · pH entre 5 y 8.
· Termostato de seguridad para proteger las · Cloruros < 60 mg / l.
resistencias eléctricas; · Presión dinámica: entre 2,0 y 4,0 bar absolutos.
23/24
Esterilizador a Vapor HI VAC MX II
CAPACIDAD Y DIMENSIONES DOS EQUIPOS

50

ABERTURA ALVENARIA - LARGURA


WALL OPENING - WIDTH

LADO DE CARGA
LOADING SIDE
B
ABERTURA ALVENARIA - ALTURA
WALL OPENING -HEIGHT

25
C + 50

ABERTURA ALVENARIA - LARGURA


C

WALL OPENING - WIDTH

LADO DE DESCARGA
25 UNLOADING SIDE
A

A + 50

VISTA V1 - LADO DE DESCARGA DIMENSÕES EM PLANTA


V1 VIEW - UNLOADING SIDE DIMENSIONS IN PLAN

ESTERILIZADOR DUPLA PORTA ESTERILIZADOR UMA PORTA


DOUBLE DOOR STERILIZER SINGLE DOOR STERILIZER

DISTÂNCIA MINIMA -
BF
BF

LADO DIREITO /FACE DE CARGA ÁREA OPERACIONAL


ÁREA OPERACIONAL - LADO DE CARGA .
OPERATIONAL AREA
MINIMIUM DISTANCE
OPERATIONAL AREA - LOAD SIDE
RIGHT SIDE / LOADING SIDE

50

AFASTAMENTO MÍNIMO
-
MINIMUM CLEARANCE
ÁREA MANUTENÇÃO
25 25
MAINTENANCE AREA
DISTÂNCIA MIN. - LADO ESQ.
FACE DE CARGA
-
MINIMUM DISTANCE - LEFT SIDE 600
LOADING SIDE
50 DISTÂNCIA MIN. - LADO ESQUERDO
FACE DE CARGA
25 25 -
MINIMIUM DISTANCE - LEFT SIDE
LOADING SIDE

V1 V1
25 ÁREA MANUTENÇÃO
600 ESTERILIZADOR 01 PORTA
600

MAINTENANCE AREA
ÁREA OPERACIONAL - LADO DESCARGA 01 DOOR STERILIZER
OPERATIONAL AREA - UNLOADING SIDE
BF

ÁREA OPERACIONAL E DE MANUTENÇÃO ÁREA OPERACIONAL E DE MANUTENÇÃO


OPERATIONAL AND MAINTENANCE AREA OPERATIONAL AND MAINTENANCE AREA

24/24
FT.01.09.0009 - Versão/Revisão 1.001 - 2020-09
Av. Arnolfo de Azevedo, 210 • Pacaembu • São Paulo/ SP • 01236-030 • Caixa Postal: 1081• Tel: 11 3670 0000 • Fax: 11 3670 0053
Baumer S.A. www.baumer.com.br • e-mail: cmlbh@baumer.com.br