Test Results
1. COTAS Dimensionais desenho 1 - A-B-C
Range: 0 a 221 mm
Indication Range: 0 a 221 mm Resolution: 0 mm
(k) Veff
Expanded
Reference V.R
Uncertainty
x mm mm
PONTO A 220,450 0,000 2,00 Infinito
PONTO B 72,460 0,005 2,00 Infinito
PONTO C 7,700 0,002 2,00 Infinito
(k) Veff
Expanded
Reference V.R
Uncertainty
x mm mm
D -LARG. 2,930 0,002 2,00 Infinito
D - COMPRIM 26,670 0,005 2,00 Infinito
D - PROF. 0,880 0,002 2,00 Infinito
E - LARG. 2,930 0,002 2,00 Infinito
E - COMPRM. 25,450 0,005 2,00 Infinito
E - PROF. 0,940 0,002 2,00 Infinito
Certificate of Calibration
Laboratório Medição Macaé
Report: 209729/22 Qualification Date 15/08/2022 Due Date:08/2023
SO: 779335-A/2022
2 / 4
3. COTAS Dimensionais desenho 3 - Corte Interno - F - G
Range: 0 a 221 mm
Indication Range: 0 a 221 mm Resolution: 0 mm
(k) Veff
Expanded
Reference V.R
Uncertainty
x mm mm
F - LARG. 3,040 0,002 2,00 Infinito
F - COMPRIM. 25,450 0,005 2,00 Infinito
F - PROF. 1,190 0,002 2,00 Infinito
G - LARG 2,930 0,002 2,00 Infinito
G - COMPRIM. 26,500 0,005 2,00 Infinito
G - PROF. 1,480 0,002 2,00 Infinito
Certificate of Calibration
Laboratório Medição Macaé
Report: 209729/22 Qualification Date 15/08/2022 Due Date:08/2023
SO: 779335-A/2022
3 / 4
DESENHO COTAS BLOCO CURVO
Certificate of Calibration
Laboratório Medição Macaé
Report: 209729/22 Qualification Date 15/08/2022 Due Date:08/2023
SO: 779335-A/2022
4 / 4
Notes
NÃO HOUVE AJUSTE
- V.I: Value indicated on the instrument in the same unit as the instrument .
- V.R: Reference Value in the unit of measurement of the Standard Equipment .
- The reported expanded measurement uncertainty is declared as the standard measurement uncertainty multiplied by the
coverage factor k, which for a t-distribution with Veff effective degrees of freedom corresponds to a coverage probability of
approximately 95%.
- The standard measurement uncertainty was determined in accordance with publication EA-4/02.
- The Approved / Disapproved condition is restricted to the metrological quantities of the instrument, and the error limit specified for
this condition is the responsibility of the Client.
- The adjustment is not part of the scope of services.
- The instrument calibration validity, when presented in this certificate, is the responsibility of the Client.
Issuing address:Rua Azeviche, 131 - Bairro: Sol E Mar - Macaé - Rio De Janeiro
Issuing date:15 de Agosto de 2022