Serviço manual
ISO 13485-2016
www.diagnostics.be • Bélgica • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be
Disposição 20181024
CY009 – CYANSmart – Manual de Serviço
ÍNDICE
1 INTRODUÇÃO ................................................. ................................................ ......................................... 3
1,1 SDESCRIÇÃO DO SISTEMA................................................ ................................................ ................................ 3
1,2 PRINCÍPIO DE MEDIÇÃO................................................ ................................................ ................... 3
1,3 TESPECIFICAÇÕES TÉCNICAS E OPERACIONAIS................................................ ......................................... 4
2 INSTALAÇÃO................................................. ................................................ ................................................ 5
3 MANUTENÇÃO................................................. ................................................ ................................................ 6
3.1PUNCTUALMMANUTENÇÃO................................................ ................................................ ...................... 6
3.2RMANUTENÇÃO EXTERIOR................................................ ................................................ ......................... 6
3.2.1 Lavagem 6
3.2.2 Limpeza 7
3.2.3 Verificação 7
3.3 RSUBSTITUIÇÃO DE PEÇAS ESPECÍFICAS................................................ ................................................ ............ 8
3.3.1 Limpeza externa 8
3.3.2 Lavagem interna 8
3.3.3 Cronograma de substituição 8
4 CONTROLE TÉCNICO ....................................... ................................................ ......................................... 9
4,1GDESCRIÇÃO GERAL E DIAGRAMA DE BLOCOS................................................ ......................................... 9
4,2 MAIN PLACA DE CONTROLE................................................ ................................................ .........................10
4.2.1 Conexões na placa-mãe 10
4.2.2 Pontos de verificação de tensão na placa-mãe 11
4.2.3 Verificação da tensão para a placa principal 11
4.2.4 Substituição da placa principal 12
4,3 OSISTEMA ÓPTICO................................................ ................................................ .........................................12
4.3.1 Lâmpada 13
4.3.2 Roda de filtro 13
4.3.3 Célula de Fluxo 13
4.3.4 Placa pré-amplificadora 14
4.3.5 Verificação do sistema óptico 14
4.3.6 Limpeza e/ou substituição da célula de fluxo 16
4.3.7 Verificação do sensor home da roda do filtro 17
4.3.8 Limpeza e/ou substituição do sensor home da roda do filtro 18
4.3.9 Verificação do motor da roda do filtro 19
4.3.10 Substituição do motor da roda do filtro 19
4.3.11 Limpeza e/ou substituição dos filtros 21
4.3.12 Verificação da tensão para a lâmpada 22
4.3.13 Substituição da lâmpada 23
4.3.14 Substituição da placa do pré-amplificador 26
4.3.15 Substituição do sistema óptico 26
4,4 horasSISTEMA DRÁULICO E DE ASPIRAÇÃO................................................ ................................................ ....28
4.4.1 Verificação do sistema de aspiração e hidráulico por calibração da bomba 28
4.4.2 Substituição de tubos no circuito hidráulico 30
4.4.3 Substituição do tubo da bomba peristáltica 31
4.4.4 Substituição das borrachas das bombas peristálticas 32
4.4.5 Limpeza e lubrificação do motor da bomba com rotor 32
4.4.6 Verificação do motor da bomba 33
4.4.7 Substituição do motor da bomba por rotor 34
4.4.8 Verificação do interruptor de aspiração 34
4.4.10 Substituição da chave de aspiração 35
4.4.11 Formato Aspirado 35
Serviço CYANSmart Manual de serviço, versão 20191121 (8.0) Página 1
FORMULÁRIO DE REPARO
PEÇAS DE REPOSIÇÃO
1. INTRODUÇÃO
O CYANSmart é um analisador bioquímico semiautomático fácil de operar. É uma unidade independente baseada em
microcomputador, controlando vários sistemas: impressora térmica embutida, tela sensível ao toque, sistema de detecção óptica
com célula de fluxo, incubadora e sistema de aspiração.
1.Impressora térmica
2.Tela sensível ao toque
① 3.Sistema óptico
4. Bloco da incubadora
5.Bomba peristáltica
CARNING! Por favor, leia o manual de serviço antes de abrir o CYANSmart e realizar diagnósticos de
problemas ou substituições.
RISCO BIOLÓGICO: Use equipamento de proteção individual. Use precauções universais. Consulte o capítulo 8. Informações de
segurança para as precauções recomendadas ao trabalhar com materiais de risco biológico.
Uma lâmpada halógena emite luz que passa por um filtro selecionando o comprimento de onda
necessário. A luz então passa através da amostra na célula de fluxo. A amostra irá transmitir ou
refletir frações da luz. O detector medirá quanta luz foi refletida/transmitida através da amostra.
Essa medição é então amplificada e convertida em um número digital para cálculos adicionais.
Esses cálculos podem ser realizados porque existe uma relação entre quanta luz é refletida/
transmitida e a concentração das substâncias em uma amostra.
Seleção de comprimento de onda Automático através de 7 filtros de interferência: 340nm, 405nm, 492nm, 510nm,
546nm, 578nm, 620nm
Célula de fluxo
Impressora
Programas
Armazenamento de dados 180 métodos de teste e 6.000 amostras de resultados de
Métodos teste Métodos Cypress pré-programados na entrega
línguas Inglês, francês, espanhol
Recurso de controle de qualidade 2 controles programáveis
Menu separado de resultados de CQ com gráficos (imprimíveis)
Métodos de cálculo Absorvância, ponto final, dois pontos, cinético, fator
Métodos de calibração bicromático, calibrador e calibração multiponto
Opções em branco Branco de amostra, branco de reagente e água destilada 0
tempo de incubação – 999 segundos
Tempo de leitura 3 – 999 segundos
Fonte de energia
- CA 220 V ± 10%, 50 Hz ± 2% ou CA 110 V ± 10%, 60 Hz ± 2%
- Tensão automática e interruptor hertz
- Aterramento necessário
- 2 fusíveis T2AL250V
- Potência: 200VA, reduzida para 140 VA no modo de espera
Peso e dimensões
- Instrumento: 9 kg, 435 x 415 x 200 mm (CxLxA)
- Embalado em caixa: 11,5 kg, 520 x 430 x 260 mm (CxLxA)
Requerimentos ambientais
- Temperatura ambiente: 15°C – 30°C
- Temperatura de armazenamento: 5-50°C
- Umidade relativa: 30% - 70%
- Pressão do ar: 860 hPa - 1060 hPa
2 INSTALAÇÃO
O CYANSmart é um instrumento fácil de operar com software integrado de fácil utilização. Nenhum computador ou
instrumento adicional é necessário para boas operações. Para garantir totalmente o desempenho do CYANSmart, é
extremamente importante que ele seja instalado corretamente.
Se as instruções de instalação, uso e manutenção não forem seguidas corretamente e/ou as indicações de segurança não
forem respeitadas, a Cypress Diagnostics não pode garantir o funcionamento correto do instrumento. O uso indevido
pode comprometer a segurança do operador e seus arredores e anulará a garantia. Os consumíveis não estão incluídos
na garantia.
CARNING! NÃO descarte a caixa de papelão ou o material de proteção!Se você precisar enviar o instrumento (por
exemplo para serviço), certifique-se de usar os materiais de embalagem originais. Outras formas de embalagem disponíveis
comercialmente não são recomendadas e podem invalidar a garantia.
Na instalação, é altamente recomendável preencher o formulário de instalação. Ele contém informações essenciais e também fornece
uma pequena lista de verificação de todos os requisitos de instalação.
O número de série pode ser encontrado na etiqueta traseira. Esta é uma informação essencial em todas as comunicações
sobre o analisador. A versão do software pode ser encontrada em System Settings – System Info.
Lista de embalagem:
3 MANUTENÇÃO
Este capítulo contém todas as operações de rotina relativas à manutenção do instrumento. Os procedimentos, listados e
descritos a seguir, devem ser seguidos cuidadosamente para garantir as especificações de qualidade dos fabricantes e as
perfeitas condições de funcionamento do instrumento ao longo do tempo.
O CYANSmart requer três níveis de manutenção:
- Manutenção pontual: em momentos especiais como na instalação
- Manutenção de rotina: para manter o nível de precisão dia após dia
- Manutenção especial: substituição de peças específicas
Consulte o formulário de manutenção para obter uma visão geral das diferentes ações de manutenção. Em caso de
reparação, registe todas as acções efectuadas na ficha de reparação e assine-a para poder acompanhar as
intervenções.
3.1PUNCTUALMMANUTENÇÃO
A manutenção pontual deve ser efetuada na instalação e antes ou após um longo período de inatividade. Os
seguintes pontos de atenção devem ser considerados:
- Verifique as instruções de instalação.
- Verifique todos os tubos visualmente quanto a vazamentos ou esmagamentos. Preste atenção especial ao tubo da bomba
3.2RMANUTENÇÃO EXTERIOR
Frequência O que Soluçao de limpeza
3.2.1WCINZA
eu. Entre testes
Entre métodos, sempre lave a célula de fluxo com água destilada.
Não é necessário lavar o instrumentoentre amostras diferentespara o mesmo método. No entanto:
- No caso dealto risco de cruzamento , você pode pré-lavar a célula de fluxo aspirando água destilada, seguida
de ar (para evitar diluição) entre as amostras.
- No caso de umreagente fortemente colorido , pode ser necessário pré-lavar a célula de fluxo com reagente (de
trabalho) antes de aspirar a primeira medição. Caso contrário, a primeira medição pode ser influenciada pela
água destilada usada para zerar o instrumento.
Serviço CYANSmart Manual de serviço, versão 20191121 (8.0) Página 6
4. Repita esta operação usandoágua destiladapelo menos 10 vezes (=15 mL) para enxaguar o sistema.
5. Repita esta operação usandoarpelo menos 10 vezes (=15 mL) para secar o sistema hidráulico e evitar
contaminação microbiológica.
3.2.2 CINCLINADO
3.2.3 VERIFICAÇÃO
eu. Diariamente
ii. Semanalmente
Antes de substituir qualquer parte do instrumento, reparar qualquer item defeituoso ou realizar qualquer procedimento
de manutenção do instrumento, o operador ou técnico de manutenção deve realizar o procedimento de
descontaminação descrito abaixo da(s) peça(s) do instrumento envolvida(s) na(s) operação(ões).
Algumas peças sobressalentes devem ser substituídas após um determinado período de tempo para garantir o bom desempenho do
analisador. Por favor, verifique a tabela abaixo:
placa amplificadora A cada 2/3 anos (juntamente com a substituição da lâmpada) Engenheiro de Serviço
bomba peristáltica A cada 2/3 anos (juntamente com a substituição do tubo da bomba pelo engenheiro
borrachas de serviço)
dos tubos)
Célula de fluxo Em caso de vazamento, obstrução ou coloração forte Engenheiro de Serviço
4 CONTROLE TÉCNICO
4.1 GDESCRIÇÃO GERAL E DIAGRAMA DE BLOCOS
A placa de controle principal controla a operação geral do instrumento, incluindo controle do motor, controle de
temperatura e gerenciamento de dados, etc. Ela recebe energia da fonte de alimentação.
Fotodetector
Botão de apertar
⑦
⑤
②
③
A tela sensível ao toque e a impressora com placa separada são conectadas à placa principal, mas integradas na tampa
do instrumento.
Serviço CYANSmart Manual de serviço, versão 20191121 (8.0) página 9
1 2 3 4 5 6
10
11
18 17 16 15 14 13 12
* : Em instrumentos de S/N BS6KMxxx em ambos os ventiladores são conectados à conexão n° 15 na placa principal.
A placa-mãe recebe 24V da fonte de alimentação e então regulará e fornecerá a tensão necessária para todas as
outras partes. A figura abaixo mostra os diferentes pontos de verificação de tensão.
Mais explicações sobre os pontos de teste podem ser encontradas no capítulo relevante (verificação da tensão para...).
Se a tensão não estiver correta, proceda à verificação da fonte de alimentação (consulte 2).
Serviço CYANSmart Manual de serviço, versão 20191121 (8.0) página 11
Procedimento :
① ⑤ ⑥
②
③
8
1. Lâmpada
2. Roda de filtro com filtros
④
3. Motor para roda de filtro
4. Sensor inicial da roda do filtro
5. Célula de fluxo
6. Placa de pré-amplificador
②
com fotodetector
7. Elemento Peltier
8. Sensor de temperatura
④ 8 ⑥ 9. Ventilador do sistema óptico
③ 9
②
Ao substituir a lâmpada, a altura da lâmpada no suporte deve ser
otimizada (4.3.13).
① Se o analisador estiver ligado, mas a lâmpada não emitir luz, prossiga com a
substituição da lâmpada (consulte 4.3.13). Certifique-se de verificar a tensão
na placa principal antes de substituir a lâmpada (consulte 4.3.12).
Tenha em mente:a numeração na roda do filtro ( ) não aumenta com o comprimento de onda.
Nº no software 1 2 3 4 5 6 7 8 9 0
Filtro real (nm) 340 405 492 510 546 578 620 Poupar Poupar Escuro
N° na roda do filtro 6 8 9 0 1 2 3 4 7 5
Pré-filtro - 2.3 1,0 1,0 0,8 0,6 0,5
4.3.3FCÉLULA BAIXA
Após o filtro, a luz passa pela célula de fluxo que contém a amostra a ser medida. O caminho da luz da célula de fluxo é de
10 mm. A célula de fluxo é aquecida com um elemento Peltier a 25°C, 30°C ou 37°C e o aquecimento é regulado por um
sensor de temperatura. Para evitar o superaquecimento, o calor abundante é conduzido com um dissipador de calor em
direção ao ventilador do sistema óptico. Para obter mais informações sobre o mecanismo de aquecimento, calibração e
substituição, consulte 4.6 Temperatura.
Para entender melhor o processamento do sinal de luz, é crucial entender a diferença entre 3 valores que
são obtidos durante uma medição automática de zero do AD.
Ganho O coeficiente de amplificação do sinal. Este ganho é ajustado automaticamente pelo
instrumento para obter valores de AD entre 50.000 e 60.000 até atingir um ganho máximo
de 29. O instrumento pode medir um valor de ganho na faixa de 1 a 30. O valor ideal para
medição está entre 8 e 16.
valor AD O sinal amplificado e convertido (de analógico para digital). O valor AD é expresso em
mV. Refere-se à intensidade da luz captada pelo fotodiodo. O instrumento é capaz de
medir valores AD em uma faixa de 0 a 65535. O valor ideal para medição é de 80 a 90%
do valor máximo, portanto, entre 50.000 e 60.000.
Desvio A tensão escura, usada para compensar a interferência (de fundo). O feixe de luz é enviado
através de um filtro preto, para que a placa do pré-amplificador não receba nenhum sinal de luz
da lâmpada ou da amostra. A tensão medida é, portanto, a tensão (de fundo) da própria placa de
circuito. O valor ideal para medição está entre -150 e +150.
1. Selecione o menu “AD Zero”. O resultado do último AD zero executado é mostrado. Recomenda-se fazer uma
impressão desses resultados (pressionando o botão “Imprimir”) para poder comparar com a nova leitura.
2. Posicione um tubo com água destilada sob o tubo de entrada de aspiração e pressione o botão PUSH para
aspirar a água destilada.
3. Pressionando o botão “Ler”, o instrumento fará uma nova leitura e exibirá os resultados.
Verifique se o ganho, AD e offset estão dentro da faixa aceitável e se o estado está “OK” para todos os filtros.
- Ganho: entre 8 e 16 AD: entre
- 50.000 e 60.000 Offset: entre
- -150 e +150
Como o AD Zero deve ser feito todos os dias, a variação entre a nova medição e a anterior (exibida
ao abrir o menu) deve ser pequena. Se você comparar os resultados (sempre compare o mesmo
filtro!):
- A diferença no ganho deve ser ≤ 2 A
- diferença no deslocamento deve ser ≤ ±50.
Se odiferença é maior ou os valores obtidos estão fora do intervalo , verifique primeiro se:
1. O instrumento fica ligado por pelo menos 30 min, pois a lâmpada precisa desse tempo para se estabilizar.
2. O sistema hidráulico (célula de fluxo + tubos) é limpo e enxaguado abundantemente com água destilada. o
Realize a lavagem e o enxágue (consulte 3.2.1). Isso lava a célula de fluxo e os tubos.
3. As janelas da célula de fluxo estão limpas (sem sujeira/impressões digitais do lado de fora). o
Limpe com lenços especiais para óculos.
4. Não há rachaduras ou vazamentos na célula de fluxo: o
Isso pode causar flutuações no ganho e no valor AD, a célula de fluxo precisa ser substituída (consulte 4.3.6).
5. A célula de fluxo é preenchida com água destilada. Para garantir uma medição correta, a célula de fluxo deve ser preenchida
com água destilada e livre de bolhas de ar. Este preenchimento pode ser facilmente verificado abrindo a tampa óptica.
Verifique os tubos que entram e saem da célula de fluxo:
oSe o enchimento não estiver correto, verifique a calibração da bomba (ver 4.4.1) e a ausência de vazamentos.
1. Verifique os pontos acima (1-5) ainda mais minuciosamente se (somente) o filtro de 340 nm estiver fora da faixa, porque
este filtro é o mais sensível à sujeira ou enxágue insuficiente,prossiga para 4.3.6 e somente depois disso continue para a
próxima etapa.
2. Verificação do movimento da roda do filtro (Prossiga para 4.3.7 e 4.3.8 e 4.3.9 e 4.3.10 ).
1. Se o instrumento estiver LIGADO, mas a lâmpada não der luz, ela precisa ser substituída. Certifique-se de
testar a tensão na placa-mãe antes de substituí-la (Prossiga para 4.3.12 e 4.3.13 ).
2. A célula de fluxo está na direção correta (janelas para frente e para trás) e completamente inserida na parte inferior do
suporte da célula de fluxo (consulte os pontos de atenção 4.3.6).
3. A tampa óptica é fechada durante a medição.
4. Verificação do movimento da roda do filtro (Prossiga para 4.3.7, 4.3.8, 4.3.9 e 4.3.10 )
5. O ganho de todos/a maioria dos filtros é 15, 16 ou mais.
-A lâmpada está velha e deve ser substituída (Prossiga para 4.3.12 ).
Procedimento:
1. Aspire ar para certificar-se de que todos os tubos estejam secos e não vazem após a desconexão.
2. Desligue o analisador CYANSmart.
3. Retire cuidadosamente os dois tubos da célula de fluxo e retire a célula de fluxo do suporte.
4. Insira/anexe uma seringa cheia de5% de hipoclorito (lixívia)na entrada (pequena) da célula de
fluxo.
5. Exerça força na seringa para lavar a célula de fluxo. As partículas que estavam aderidas dentro da
célula de fluxo serão lavadas com o líquido através do tubo de saída (maior).
6. Incube a solução de hipoclorito a 5% por 10 min (não mais) na célula de fluxo.
7. Enxágue abundantemente a célula de fluxo com água destilada.
Procedimento:
1. Aspire ar para certificar-se de que todos os tubos estejam secos e não vazem após a desconexão.
2. Desligue o analisador CYANSmart.
3. Retire cuidadosamente os dois tubos da célula de fluxo e retire a célula de fluxo do suporte.
4. Para instalar a nova peça, repita as etapas acima na ordem inversa.
UMAPONTOS DE TENÇÃO!
1. Insira a célula de fluxo nodireção correta! A luz deve poder passar pelas janelas da célula de fluxo,
portanto, as janelas precisam estar voltadas para a frente e para trás.
2. O tubo de entrada (6) é fino, conectado ao tubo de aspiração e deve ficar voltado para frente. O tubo de
saída (7) é grosso, conectado ao tubo que vai para a bomba peristáltica e deve ficar voltado para trás.
3. Insira a célula de fluxo completamente na parte inferior do suporte. Certifique-se de que nada está bloqueando a
inserção!
4. Fixe sua posição com a placa dourada designada.
√ x
⑦ ⑥ ⑦
⑥ ⑦
⑥
janela
④
③
Procedimento:
7. Limpe a fenda com ar comprimido (seco!) ou inserindo papel resistente. Certifique-se de remover toda a sujeira.
Siga os passos 1-5 do procedimento para limpar a fenda da roda do filtro (A). 6.
Desconecte o sensor inicial da roda do filtro da placa principal (consulte 4.2.1).
7. Remova os 2 parafusos (3) que prendem o sensor inicial e remova-o com cuidado.
8. Limpe a parte interna dos braços do sensor home com ar comprimido (seco!) com uma escova macia ou com lenços especiais
para limpeza de vidros. Certifique-se de remover toda a poeira/sujeira acumulada.
9. Reposicione o sensor repetindo as etapas acima na ordem inversa.
UMAATENÇÃO: Certifique-se de que a roda do filtro não toque no sensor inicial ao girar! A roda deve
estar no meio da abertura do sensor inicial. Caso contrário, reajuste sua posição ajustando os parafusos de
aperto do sensor home. Verifique visualmente dando uma volta completa na roda! Quando houver uma dobra
na roda do filtro, o sistema óptico precisa ser substituído (consulte 4.3.15).
CY004/9-S35
NOTE: Após a limpeza da fenda da roda do filtro e/ou do sensor inicial da roda do filtro, você deveverifique o movimento
da roda do filtro novamente(ver 4.3.7) e osistema óptico(ver 4.3.5) novamente!
Procedimento :
Se a roda do filtro não girar mais, você precisa verificar o motor da roda do filtro.
Procedimento:
1. Desligue o analisador CYANSmart.
2. Remova a tampa do instrumento desparafusando os 6 parafusos sextavados/allen na placa inferior.
3. Verifique se a conexão do motor está bem ajustada na placa
principal (ver 4.2.1). Destaque e recoloque-o.
4. Verifique se o problema foi resolvido.
5. Caso contrário, desconecte a conexão e meça a resistência
do motor (o instrumento está DESLIGADO!).
6. Use um voltímetro ajustado para ohm (Ω). Meça a
resistência do motor (veja a foto: fio preto como
referência versus segundo fio). A resistência deve estar
entre 4,5 - 5,5 Ω. Caso contrário, o motor da roda do filtro
precisa ser substituído.
10. Use 2 chaves de fenda planas para soltar o parafuso de fixação da roda do filtro. Remova o parafuso de fixação e o anel e
puxe a roda do filtro para cima.
11. Remova o anel de metal com pinças e depois remova o anel de fixação de borracha com uma pinça.
12. Desaperte os 4 parafusos sextavados/allen que fixam o motor da roda do filtro ao sistema óptico e retire-o.
Vista traseira
NOTE: Após a substituição do motor da roda do filtro, você deveverifique o movimento da roda do filtro(ver
4.3.7) e osistema óptico(ver 4.3.5)!
- Um filtro quebrado pode resultar em ganho < 8. Se o filtro estiver quebrado ou arranhado, o filtro precisa ser substituído.
Procedimento :
Tenha em mente:a numeração na roda do filtro não aumenta com o comprimento de onda. Observe a tabela abaixo para
a numeração correta. Posicione sempre o pré-filtro primeiro. Apenas o filtro de 340 nm não possui pré-filtro. Ao
reposicionar o filtro, certifique-se de que ele seja colocado da maneira correta (seta = direção da luz!).
Nº no software 1 2 3 4 5 6 7 8 9 0
Filtro real (nm) 340 405 492 510 546 578 620 Poupar Poupar Escuro
N° na roda do filtro 6 8 9 0 1 2 3 4 7 5
Diafragma pré-filtro - 2.3 1 1 0,8 0,6 0,5 - - -
Procedimento :
√ x
4.3.13.1Instalando a lâmpada
Procedimento :
4. Remova a bandeja do sistema óptico desaparafusando os 4 parafusos ao redor do poço de leitura e posicione
cuidadosamente a bandeja ao lado do sistema óptico.
5. Desconecte o fio da lâmpada da placa principal (consulte 4.2.1).
6. Desaperte o parafuso de aperto do suporte da lâmpada. Gire o parafuso no sentido anti-horário para soltá-lo, retire a
lâmpada e a moldura do sistema óptico.
9. Enquanto ajusta a altura da lâmpada (ponto 8), siga o valor de AD na tela (para o filtro com maior valor de AD –
ponto 7).Encontre a posição de altura que lhe dá o maior valor de AD, em seguida, aperte o parafuso para
fixar a posição da lâmpada.
Com um sistema limpo (hidráulico e filtros) e uma lâmpada nova, este valor deve ficar acima de 10000 e o mais
próximo possível de 15000.
10. Certifique-se de que a lâmpada esteja bem fixada para evitar o movimento da lâmpada.
NOTES:
1. O ajuste de altura da lâmpada requer um pouco de tempo e experiência, mas éessencial para obter
resultados exatos e precisos.
2. Durante o transporte do analisador, é possível que a altura da lâmpada tenha sido ligeiramente ajustada. Se durante
instalação, o ganho permanece fora da faixa durante a medição de zero automático AD mesmo após a verificação
dos diferentes requisitos (enxágue, lavagem, limpeza, sem bolhas de ar, instrumento ligado por pelo menos 30
minutos,… (consulte 4.3.5)), então a altura da lâmpada precisa ser ajustada.
3. Excepcionalmente, o correto ajuste da altura pode resultar em ganho de valores entre 8 -10 (durante o AD zero).
a) Instrumentos com P/D anteriores a 2016 (versão de software < 1.1.3). Há um erro com o AD auto zero
se o ganho estiver abaixo de 10. Se for o caso, é necessário fazer um ajuste menos ideal da altura
(valor menos alto no ponto 9) para trazer o AD auto zero dentro do alcance (10 – 16).
b) A partir da versão de software 1.1.3 (instrumento com data de produção após março de 2016), isso é levado
em consideração e não emitirá mais um aviso.
Procedimento :
Vista traseira
Vista do topo
Se algo comprometer esta reta, como a presença de uma dobra ou curva na roda do filtro, o sistema óptico (CY009-S01)
precisa ser substituído! Esta peça sobressalente é fornecida junto com o motor da roda do filtro anexado. A lâmpada,
filtros, sensores, célula de fluxo, elemento Peltier, ventilador e placa de pré-amplificador não estão incluídos.
Procedimento :
A bomba peristáltica está situada no lado esquerdo do instrumento. Pode ser ativado por:
- O botão 'Lavagem do tubo' na tela sensível ao toque.
- O interruptor de aspiração pressionando o botão de aspiração.
Ao acionar o motor de passo da bomba peristáltica, o rotor da bomba gira, movimentando uma coluna de líquido no tubo
da bomba e com isso um volume quantitativo é aspirado no sistema hidráulico. Esta é uma linha contínua composta por
tubos, conectores, a célula de fluxo e terminando no frasco de resíduos.
Procedimento :
1. Selecione o menu “Calibrar bomba”. O número de etapas registradas durante a última calibração da bomba é
mostrado.
2. Pressionando a caixa “volume”, um teclado numérico aparecerá, digite 3000 µl como o volume para aspirar e pressione OK
para sair da visualização do teclado.
3. Pipetaexatamente 3000 µl de água destiladaem um tubo de ensaio.
4. Posicione a ponta de aspiração dentro do tubo de ensaio, certifique-se de que a ponta de aspiração esteja no canto inferior do tubo de
ensaio (ver figura), para que todo o líquido possa ser aspirado.
Se os valores obtidos estiverem fora do intervalo ou a diferença entre as réplicas for > 1000 passos:
√
x √
x
d) Verifique se não há lubrificante no tubo da bomba ou nos rolos do rotor. Lubrificante em excesso pode fazer com que
os rolos deslizem e, assim, aumentar o número de etapas - limpe e remova o lubrificante.
e) Verifique se o rotor da bomba está girando suavemente. Caso contrário, o motor da bomba precisa ser verificado e/ou
substituído (veja abaixo).
Se todos os itens acima forem verificados, mas a calibração da bomba permanecer fora da faixa, o tubo da bomba ou as borrachas da bomba
precisam ser substituídos (veja abaixo).
Antes de substituindo qualquer peça, certifique-se de realizar uma descontaminação externa e interna (consulte 3.3.1 e 3.3.2),
para evitar a contaminação das novas peças de reposição.
Procedimento :
1. Aspire ar para certificar-se de que todos os tubos estejam secos e não vazem após a desconexão.
2. Desligue o analisador CYANSmart.
3. Remova a tampa do instrumento desparafusando os 6 parafusos sextavados/allen na placa inferior.
4. Substitua os tubos e/ou conectores (um a um). Encontre o código correto do item e a descrição usando o
diagrama e a tabela abaixo.
5. Ligue o analisador e execute uma descontaminação interna (consulte 3.3.2). Certifique-se de que não há
vazamento.
6. Feche a tampa do instrumento, certificando-se de NÃO esmagar ou dobrar nenhum dos tubos!
NOTE: Depois de substituir qualquer parte do sistema hidráulico, você deverecalibrar a bomba peristáltica(Vejo
4.4.1)!
Fixação de entrada
Garrafa de lixo Painel traseiro bomba peristáltica Célula de fluxo Aspiração Aspiração
trocar botão
tubos
CY004/9-S26 Tubo de aspiração de plástico
Perigo biológico:
Manipule cuidadosamente todos os consumíveis e os resíduos produzidos durante a análise das
rotinas. Use vestimenta de proteção apropriada. O descarte de resíduos deve ser feito de acordo
com a regulamentação aplicável. Recomenda-se verificar periodicamente o nível no recipiente de
resíduos, a fim de evitar transbordamento.
Serviço CYANSmart Manual de serviço, versão 20191121 (8.0) página 30
Como a pressão é aplicada ao tubo da bomba peristáltica, esse tubo se desgasta mais rapidamente. Para garantir o bom
desempenho do analisador, o tubo da bomba peristáltica deve ser substituídopelo menos a cada 6 meses. Essa
frequência pode variar dependendo da carga de trabalho de cada laboratório. A qualidade e confiabilidade deste tubo é
fundamental para garantir uma aspiração precisa e precisa.
Procedimento :
1. Aspire ar para certificar-se de que os tubos estejam secos e não vazem ao serem desconectados.
2. Desligue o analisador CYANSmart.
3. Levante a alça para abrir a porta lateral da bomba peristáltica.
4. Para trocar o tubo, remova-o do rotor da bomba peristáltica (2) e desconecte o tubo da bomba de seus 2
conectores puxando-os suavemente para fora. Insira os conectores no novo tubo e coloque o novo tubo (1) ao
redor do rotor da bomba peristáltica (2).
5. Certifique-se de inserir os tubos na direção correta. O tubo saindo da célula de fluxo (4) para a
frente e o tubo indo para o conector de resíduos (5) para trás.
6. Em seguida, feche a trava da guia articulada (3). Certifique-se de que está completamente clicado.
7. Para verificar a fixação do tubo da bomba, gire manualmente o rotor da bomba peristáltica (direção correta!). Se
o rotor puder ser girado suavemente, o tubo está colocado corretamente. Este torneamento manual só pode
ser executado quando oinstrumento está DESLIGADO ! Quando o instrumento é ligado, os movimentos do
motor são bloqueados e girar manualmente o rotor pode danificar o mecanismo!
NOTE: Depois de substituir o tubo da bomba, você deverecalibrar a bomba peristáltica(ver 4.4.1)!
①
②
FORA NO
Desperdiçar Da célula de fluxo
UMAPONTOS DE TENÇÃO!
1. Certifique-se de que o novo tubo da bomba seja colocado no meio dos rolos brancos do rotor.
2. Verifique se o novo tubo está completamente inserido nos suportes.
3. Certifique-se de que os conectores do tubo da bomba estejam completamente inseridos no novo tubo da bomba.
4. Certifique-se de que o tubo (4) e (5), estejam inseridos nos orifícios e não amassados pelo invólucro.
√
√
EUOBSERVAÇÕES IMPORTANTES:
A pressão aplicada ao tubo da bomba peristáltica é regulada por borrachas. Com o tempo e com o uso normal,
essas borrachas se desgastam. Para garantir o bom desempenho do analisador, as borrachas da bomba
peristáltica devem ser substituídas pelo menosa cada 2 a 3 anos.Essa frequência pode variar dependendo da
carga de trabalho de cada laboratório.
Procedimento:
1. Desligue o analisador CYANSmart.
2. Levante a alça para abrir a porta lateral da bomba peristáltica.
3. Solte a trava da guia articulada (1) e remova o tubo da bomba do rotor da bomba.
4. Desaperte o parafuso central (4) e retire a guia inferior do tubo (5).
5. Agora os 4 parafusos de canto ficam visíveis. Desaparafuse-os, tomando cuidado para não perder as porcas e arruelas
localizadas na parte interna da placa. Remova cuidadosamente o rotor com o motor da estrutura.
6. Na parte superior, há um sistema de guiamento do tubo da bomba (2). Remova este sistema de orientação (retire-o).
7. Recoloque as 2 borrachas (3) dentro do sistema de guiamento e a seguir reposicione-o.
8. Reposicione a guia do tubo inferior e o tubo da bomba e, em seguida, feche a trava da guia articulada (1). Certifique-se de que
está completamente clicado.
9. Verifique o posicionamento das borrachas e sistema guia girando manualmente a roda do motor peristáltico
(sentido correto!). Se a roda puder ser girada suavemente, ela está posicionada corretamente.
10. Se a posição estiver correta, reconecte o rotor com o motor.
NOTES:
- Depois de substituir as borrachas da bomba, você deverecalibrar a bomba peristáltica(ver 4.4.1)! Certifique-se de que
- as borrachas estão colocadas na posição correta. Se as borrachas não estiverem posicionadas corretamente, a
funcionalidade da bomba será prejudicada e a aspiração será bloqueada.
③
② ③ ③
① ②
②
⑤ ①
③
④
x√
Procedimento:
1. Desligue o analisador CYANSmart.
2. Levante a alça para abrir a porta da bomba peristáltica ou remova a tampa do instrumento
completamente para facilitar o acesso.
3. Solte a trava da guia articulada e remova o tubo da bomba do rotor da bomba.
⑥
⑥
④ ⑤
③
②
①
④
①
⑤
5. Use ar pressurizado (seco!) e limpe o eixo do motor e todos os roletes (4) do rotor (você também pode usar um pano
úmido para limpar os roletes).
6. Após a limpeza e secagem, você pode colocar1 gota de graxa de máquina (WD40) no eixo do motor (3) e um
pouco de graxa de silicone na superfície de contato (5) entre o rotor (1) e o suporte do rotor (6).
7. ! Certifique-se de que hánãolubrificante nos rolos do rotor ou nos locais de contato com o tubo da bomba!
8. To reposition the rotor, make sure the fastening grain is on the flattened side ( ) of the motor axis and that
the rotor is coplanar (same level) with the rotor holder. Then tightly fasten the grain.
①
①
④
②
√ X X
①
③ ④ ⑥
⑥ ④
NOTE: After cleaning and/or lubrication of the pump rotor/motor, you must recalibrate the peristaltic
pump (see 4.4.1)!
If the rotor does not turn anymore, you need to verify the pump motor.
Procedure:
1. Turn OFF the CYANSmart analyzer.
2. Remove the cover of the instrument by unscrewing the 6 hex/
allen screws on the bottom plate.
3. Check if the connection of the motor is tightly set on the
mainboard (see 4.2.1). Detach and reattach the connection.
4. Check if the problem is solved.
5. If not, detach the connection and measure the resistance of the
motor (instrument is OFF!).
6. Use a voltmeter set to resistance (Ω).
Measure the resistance of the motor
(see photo below: black wire as
reference versus second wire). The
resistance should be between 8-10 Ω.
Otherwise, the pump motor needs to
be replaced.
CYANSmart Service Service Manual, version 20191121 (8.0) Page 33
③ ③ CY009-S10
②
①
③ ③
If the pump motor does not respond (no aspiration, no sound of turning) when pressing the aspiration switch
(and you are in the correct menu) and it does turn when pressing the “tube wash” button in the software, you
need to verify the aspiration switch.
Procedure:
1. Turn OFF the CYANSmart analyzer.
2. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
3. Check if the connection of the aspiration switch is tightly set on the mainboard (see 4.2.1). Detach and
reattach it again.
4. Check if the problem is solved.
5. If not, detach the connection and measure the resistance (instrument is OFF!).
6. Use a voltmeter set to ohm (Ω) and if the function is available to sound ( ). Connect the black pin to one
connection and the red pin to the other.
a. When the button of the aspiration switch is not pushed in, the resistance infinite.
b. Upon pushing in the aspiration switch, the circuit closes, and you should see a jump in the
resistance and hear the buzzer (if the sound is activated).
7. Otherwise, the aspiration switch with wire needs to be replaced.
CY004/9-S21
①
①
NOTE: After replacement of the aspiration switch, you must recalibrate the peristaltic pump (see 4.4.1)!
ATTENTION POINT!
Make sure the wires are correctly attached on position 1 and 2 and NOT on 2 and 3!
√ ① X ①
②
②
③ ③
4.6 TEMPERATURE
If the measured temperatures do not correspond to these specifications, a temperature calibration should be
performed.
NOTES:
1. After start-up of the instrument, the instrument requires 15 - 30 minutes to reach the programmed
temperature.
2. If upon entering the perform test menu, the programmed temperature for the flow cell/cuvette is not yet
reached, the instrument will display a warning to inform the user of this discrepancy.
37°C
8. Compare the temperature measured by the thermometer with the temperature displayed by the
instrument “Temp” on the screen. For example the programmed temperature is 37.0°C for the flowcell.
If the instrument is correctly calibrated, the temperature measured by the thermometer and the
temperature displayed by the instrument are 37°C ± 1°C. Otherwise proceed to step 9.
If, after performing the calibration procedure twice, the temperature remains outside the tolerance range
of 37°C ±1°C, continue with
13. Cleaning and/or replacement of the fan of the optical system (see 4.6.3)
14. Verification and/or replacement of the Peltier element (See 4.6.4 and 4.6.5)
15. Verification and/or replacement of the temperature sensor (see 4.6.8 and 4.6.9).
16. Verification of the voltage going to the Peltier element on the mainboard and/or replacement of the
mainboard (see 4.6.10 and 4.2.4).
9. Wait for 15-30 min to allow the control circuit to adjust the temperature.
10. The “Temp” on the display should be 37.0±1°C and the temperature on the thermometer should be 37.5 ± 1°C.
Otherwise repeat steps 7 until 10.
11. If the calibration is successful, return to the “Reserved Setting” menu. Upon exiting this menu, by pressing
the “Main Menu” button, the instrument will ask if you want to save the changes made. Click “Yes” to save
the changed settings, “No” to exit the menu without saving or “Cancel” to return to the “Reserved
settings” menu.
When replacing a fan of an instrument before S/N BS6KMxxx it is recommended to replace both fans with the spare
part where the fans are connected together on B3. (CY009B-S16):
B3 B1 B3 B1
Procedure:
1. Turn OFF the CYANSmart analyzer.
2. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
3. Disconnect all the connections of the optical system from the mainboard (see 4.2.1): lamp, flow cell
temperature sensor, Peltier element, home sensor of filter wheel, motor of filter wheel and fan for optical
system.
4. Unscrew the 4 screws (1) attaching the optical system on the bottom plate. Remove the whole assembly and turn
it upside down and remove the ventilator by unscrewing its 2 screws (2).
5. Use pressurized air (dry!) or a damp cloth for cleaning the fan.
① ①
① ①
① ①
③ ③
②
②
①①
6. The Peltier element is now visible.
CY004/9-S14
7. To install the new part, repeat the above steps in reverse order.
ATTENTION:
a) Put sufficient heat conduction silicone grease (=thermal grease) on both sides of the Peltier element in
order to assure heat conductivity between the Peltier element and flow cell and to efficiently drain
the surplus of heat through the heat sink to the fan.
b) The Peltier element has a warm plate and a cold plate. The warm plate should be positioned on the
side of the Peltier element, the cold plate should be positioned on the side of the heat sink. To
assure the correct position of the Peltier element during installation, check the colors of the wires.
The wires have to be positioned on the side of the temperature sensor.The red wire is placed facing
the back panel of the instrument (side of the filter wheel) and the black wire is placed facing the
front side of the instrument (side of the pre-amplifier board).
c) Before turning ON the instrument, check if both the heat sink and the fan of the optical system have
been installed. Never turn ON the instrument when the heat sink and the fan are not installed, this
can result to an overheating and will damage the Peltier element.
8. After replacement of the Peltier element, the temperature of the flow cell needs to be verified and
recalibrated (see 4.6.2).
9. If after replacement of the Peltier element, the temperature problem persisits, continue with the
verification of the temperature sensor (see 4.6.8).
② ③ ③
④
①
② ③ ③
④
7. To install the new part, repeat the above steps in reverse order.
ATTENTION: When replacing the heater element and the temperature control switch, put sufficient heat
conduction silicone grease (=thermal grease) on the inside in order to assure efficient heat conductivity.
CY009-S52
8. After replacement of the heater element, you must verify and recalibrate the temperature of the
incubator (see 4.6.2).
9. If after replacement the temperature problem of the incubator persists, continue with the verification of
the temperature sensor (see 4.6.8).
For the verification of the temperature sensor the sensor needs to detect different temperatures. The
ambient temperature and a second (external) higher temperature (>5°C higher than room temperature);
water of 40°C or more.
Procedure
1. Turn of the CYANSmart analyzer.
2. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
3. Unscrew the temperature sensor from its seating. In the optical system, leave the cable connected to the
mainboard. The temperature sensor is positioned on the left side of the flow cell above the Peltier
element (1). In the incubator assembly, the temperature sensor is located on the left side and above the
heater element (2).
②
CY009-S34
①
If the temperature sensor cannot detect different temperatures, the sensor needs to be replaced (CY009-
S34)
Procedure:
1. Turn OFF the CYANSmart analyzer.
2. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
3. Disconnect the temperature sensor wire from the mainboard (see 4.2.1).
4. Unscrew the temperature sensor from its seating. In the optical system, the temperature sensor is
positioned on the left side of the flow cell above the Peltier element (1). In the incubator assembly, the
temperature sensor is located on the left side and above the heater element (2).
5. To install the new part, repeat the above steps in reverse order.
ATTENTION: When replacing the temperature sensor, put sufficient heat conduction silicone grease
(=thermal grease) on the inside in order to assure efficient heat conductivity.
6. After replacement of the temperature sensor, the temperature needs to be verified and recalibrated (see
4.6.2)
7. If after replacement the temperature problem persists, continue with the verification of the voltage going
to the heater elements on the mainboard (see 4.6.10).
4.6.10 VERIFICATION OF THE VOLTAGE GOING TO THE HEATER ELEMENTS ON THE MAINBOARD
The regulation of the voltage given to the Peltier element or the heater element of the incubator depends on
the displayed temperature by the instrument. A correct regulation of the temperature thus depends on the
correct operation of the temperature sensor and a correct calibration (displayed temperature is the same as
the measured temperature).
If the resistance measured on the Peltier element or the heater element of the incubator is correct, but the
temperature measured with the thermometer does not reach 37°C ± 1°C for the flow cell or 37.5°C ± 1°C for
the incubator, after having performed the temperature calibration 2-3 times, the voltage supply to the
respective part needs to be verified on the mainboard.
Procedure:
1. Turn OFF the CYANSmart analyzer.
2. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
3. Turn ON the CYANSmart analyzer
4. Measure the voltage on the mainboard by using a voltmeter set to V DC:
a. Peltier element of the flow cell :
- Test point 4 compared to the grounding of test point 5 or a screw (see 4.2.2): 15 V (DC)
Printer
Feed
ATTENTION POINT!
Make sure to place the paper in the right direction. Only the coated side can be used to print. If the paper is
installed in the wrong direction (see below) the paper will move through the printer, but nothing will be
printed. To verify the printable side of the paper, the operator can scratch the paper. Upon scratching the
paper, only one side will become slightly greyish or shiny (see picture below), that is the side that can be used
for printing and thus should contact the printing bar.
WARNINGS :
- Do not try to print without printing paper installed, this can cause a system crash!
- If the printing bar is (accidently) removed from the printer head, make sure to turn OFF the
instrument before repositioning it. Otherwise, the cogwheels can get damaged.
If the paper is correctly installed, but the “Printer feed” button does not make the printing bar turn nor make the
paper move forward, the voltage supply to the respective part needs to be verified on the mainboard.
Procedure:
1. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
2. Turn ON the CYANSmart analyzer.
3. Measure the voltage on the mainboard on test point 9 compared to the grounding of test point 5 (see
4.2.1). The measured voltage should be 5V (DC).
Procedure:
1. Turn OFF the CYANSmart analyzer.
2. Remove any left-over of the old print paper roll.
3. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
4. Disconnect the printer power cable (1) and the flat printer data cable (2) from the printer board (3).
5. Very carefully pull out the flat orange/golden cable (4) from its slot.
6. Unscrew the 2 screws attaching the printer board to the case (5).
CYANSmart Service Service Manual, version 20191121 (8.0) Page 46
② ⑤
①
③ ⑥
⑥
⑤ ④
8. To install the new part, repeat the steps (3-6) in reverse order.
ATTENTION POINT!
Insert the flat orange/golden cable straight, completely and in the correct direction (see below) in its slot on
the printer board. The connections should be facing the inside of the board.
⑥
⑥
WARNINGS :
- Do not try to print without printing paper installed, this can cause a system crash!
- If the printer bar is (accidentally) removed from the printer head, make sure to turn OFF the
instrument before repositioning it! Otherwise the cogwheels can get damaged!
ATTENTION POINT!
The screen can only react to one touch point at the same time. If something/someone else is touching or
pressing the screen (on an empty position), it will not respond to the touch on a function key!
Procedure:
1. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
2. Turn ON the CYANSmart analyzer.
3. Measure the voltage on the mainboard on test point 3 compared to the grounding of test point 5 (see
4.2.2). The measured voltage should be 5V (DC).
Procedure:
1. Turn OFF the CYANSmart analyzer.
2. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
3. Disconnect the flat touch screen cable (1) from the touch screen board (2). To do this, lift the black
connection part up (3) and then the cable can be taken out without pulling!
4. Carefully pull out the flat touch screen cable (1) from its slot on the mainboard (4).
②
③
①
④
5. To install the new part, repeat the steps (3-4) in reverse order.
ATTENTION POINTS!
1. Insert the flat touch screen cable straight and in the correct direction in its slot on the touch screen board.
Then close/push down the black part of the connection to tighten it. The connections should be facing
the board.
√ X X
2. Insert the flat touch screen cable straight and in the correct direction in its slot on the mainboard. The
connections should be facing the outside of the board.
√ X
3 2 5
If the voltage measured on the mainboard is correct and all connections (on both the mainboard and the
touch screen board) are firm, but the screen still does work, the touch screen needs to be replaced. The touch
screen is supplied together with the touch screen board (CY009-S20).
Procedure:
1. Turn OFF the CYANSmart analyzer.
2. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
3. Disconnect the flat touch screen cable (1) from the touch screen board (2).
4. Remove the cover to get access to the touch screen board by unscrewing the 4 screws.
5. Unscrew the 4 screws attaching the touch screen to the case (3).
③ ③
②
③ ③
6. To install the new part, repeat the steps (3-5) in reverse order.
NOTE: After replacement of the touch screen, you must verify and/or adjust the screen brightness and
perform a touch screen calibration (see 4.8.1, 4.8.2).
ATTENTION POINTS!
1. If there are dents or inequalities in the cover mold and the touch screen is attached firmly against the
cover, these inequalities will push in the touch screen, making it unresponsive to your touch! (See
attention point 4.8).
2. Insert the flat touch screen cable in the correct direction in its slot on the touch screen board. The
connections should be facing the board (see 4.8.4).
①
⑤
③
④
When the power switch (5) is set to ON (I), the input power is going to the power supply (6), were it is
transformed to 24V DC. This voltage is connected to the mainboard for further use/division/regulation.
③ ⑤
1. Always use the original power cord supplied with the instrument.
2. Make sure the connection (wall outlet and instrument) is correct and firm.
3. Place on a stable, horizontal surface, free from vibrations.
4. Keep the instrument away from other appliances that generate electrical noise or magnetic fields (e.g.
radiological instruments, X-ray, radio…).
5. Connected to a suited and grounded AC wall outlet, preferably one that is not shared with other electric
appliances and with low fluctuation of line voltage: AC
220V ± 10%, 50Hz ± 2% or AC 110V ± 10%, 60Hz ± 2%.
6. If the AC wall outlet is not properly grounded, please connect a ground or earth wire to the bottom plate of
the analyzer. A bad ground line connection may compromise analysis results and damage the
instrument.
7. If the fluctuation is >10% use the following supplementary devices:
- Electronic stabilizer
This is used to stabilize the electric voltage in the laboratory. Any stabilizer, currently available on the
market, with a power potential of at least 0.5 KW can be used.
- No Break Module UPS (Uninterrupted Power Supply)
This module has two important features:
o Stabilizes the main power.
o Supplies power to the instrument in case of main power line failure.
After turning off the analyzer, you need to wait at least 60 seconds before turning it on again!
Otherwise, the fuses will blow because of the residual current.
4. To install the new fuses, repeat the steps above (2-3) in reverse order.
YES
Problem located after power supply
Fuses
OK? NO
YES
NO OK?
Verification of the fuses (4.9.4)
1. Switch OFF (O)
2. Disconnect power cord Verification of the power supply
3. Disconnect ACL and ACN on the 1. Switch OFF (O)
power supply 2. Disconnect power cord
4. Replace the two fuses (4.9.2) 3. Disconnect +V0 and 2 grounding cables from
5. Reconnect power cord and switch the power supply (leave ACL and ACN
ON (I) connected!)
6. Check if the fuses remain ok 4. Reconnect power cord and switch ON (I)
7. Measure the voltage on the end of 5. Check if:
the disconnected ACL and ACN a. Fuses remain ok?
b. Light of the power supply is burning?
c. Check the voltages on the power supply
YES - ACL versus ACN: 110/230 V AC
OK?
- + V0 versus GND: 24V DC
NO
After turning off the analyzer, you need to wait at least 60 seconds before turning it on
again! Otherwise, the fuses will blow because of the residual current.
Procedure:
1. Verify the general electricity and installation requirements (4.9.1). Only proceed if these are OK.
2. Switch OFF (O) the analyzer and disconnect power cord.
3. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
4. Disconnect input ACL and ACN wires or L and N connector from the power supply.
CY009-S24A CY009-S24B
2 1
- Make sure both measuring pins do not touch each other while measuring. This will lead to an
overcurrent, which will cause unrepairable damage to the instrument
Procedure:
1. Verify the general electricity requirements (see 4.9.1). Only continue if everything is OK. Proceed only if the
AC wall outlet meets the specifications of the value of voltage and frequency, the fluctuations and if it is
correctly grounded.
2. Switch OFF (O) the instrument and disconnect power cord.
3. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
4. Verify unit 1 (power plug connector):
a. Disconnect the blue and red cable on unit 1, leave the grounding cable (yellow-green) connected.
b. If necessary, replace the fuses (see 4.9.2).
c. Reconnect the power cord and turn ON (I) the analyzer.
d. Use a voltmeter (AC) to measure the voltage on the two pins of the power plug connector. The value should
be 110V or 230V (depending your location).
o If the measured voltage is not correct and the AC wall outlet meets the required specifications (see 4.9.1),
the power plug connector needs to be replaced (CY009-S46).
o If the measured voltage is correct, proceed with step 5.
5. Switch OFF (O) the instrument and disconnect power cord.
6. Reconnect the cables of unit 1.
7. Verify the cables between unit 1 and 2:
a. Disconnect the two lower cables (blue and red) from unit 2.
b. If necessary, replace the fuses (see 4.9.2).
c. Reconnect the power cord and switch ON (I) the analyzer.
d. Use a voltmeter (AC) to measure the voltage on the ends of the 2 disconnected cables. The value
should be 110V or 230V (depending your location).
o If the measured voltage is not correct and the AC wall outlet meets the required specifications (see 4.9.1),
verify the cables and replace them.
o If the measured voltage is correct, proceed with step 8.
8. Switch OFF (O) the instrument and disconnect power cord.
9. Reconnect the 2 upper cables of unit 2.
10. Verify unit 2 (power switch)
a. Disconnect the two upper cables (blue and red) from unit 2.
b. If necessary, replace the fuses (see 4.9.2)
c. Reconnect the power cord and switch ON (I) the analyzer.
d. Use a voltmeter (AC) to measure the voltage on the two upper pins of the power switch. The value
should be 110V or 230V (depending your location).
o If the measured voltage is not correct and the AC wall outlet meets the required specifications (see 4.9.1),
the power switch needs to be replaced (CY004/9-S18).
o If the measured voltage is correct, proceed with step 4.9.6.
NOTE:
- For instruments before S/N BS6KMxxx: The fan of the power assembly is separately connected to B3 on the
mainboard.
- For instruments from S/N BS6KMxxx on: The fan of the power assembly is connected to B3 together with
the fan for the optical assembly.
When replacing a fan of an instrument before S/N BS6KMxxx it is recommended to replace both fans with the
spare part where the fans are connected together on B3. (CY009B-S16):
B3 B1 B3 B1
Procedure:
1. Turn OFF the CYANSmart analyzer.
2. Remove the cover of the instrument by unscrewing the 6 hex/allen screws on the bottom plate.
3. Disconnect the wire of the fan for power assembly from the mainboard (see 4.2.1).
4. Unscrew the 4 screws (1) attaching the gridding and fan to the rear panel.
5. Use pressurized air (dry!) a damp cloth for cleaning the fan.
If the problem is not solved by cleaning, the fan needs to be replaced.
① ①
① ①
6. To install the new part, repeat the steps (2-4) in reverse order.
CYANSmart Service
①Service Manual, version 20191121 (8.0) Page 57
1. Prepare a 100% solution by dissolving exactly 14g of ammonium cobalt (II) sulfate hexahydrate in 100
mL distilled water.
2. Pipette 20 mL of this solution in the first tube (= 100% solution).
3. Pipette 12 mL of the first tube into a second tube.
4. Add 4 mL of distilled water to this second tube and mix well (= 75% solution).
5. Pipette 8 mL of the second tube into a third tube.
6. Add 4 mL of distilled water and mix well (= 50% solution).
7. Pipette 4 mL of the third tube into a fourth tube.
8. Add 4 mL of distilled water and mix well (= 25% solution).
20 mL
Mix well and use a different pipette
point for every change of pipetting
12 mL 8 mL 4 mL
4 4 4
7. Position a tube filled with distilled water under the aspiration inlet and press PUSH to aspirate.
8. Wait until the AD value of the blank for distilled water has stabilized.
9. When the value is more or less stable, press “AD Test”. The value needs to be between 50.000 and 60.000.
Otherwise, rinse with distilled water and repeat the measurement.
10. Press “OD test” to proceed with the sample measurement. In the new screen, the value for the blank of distilled
water will be displayed as soon as the button “AD test” is pressed.
11. Position the tube filled with 100% solution under the aspiration inlet and press PUSH to aspirate the
sample.
12. The instrument will measure the AD value of the sample and will calculate the optical density (OD). The result
will be displayed directly. Wait 3 seconds and then write the result.
13. Repeat 10 times to aspirate 11 successive solutions (1 time to rinse and 10 times to measure).
NOTES:
- The first aspiration is done to rinse the tube and to avoid dilution by distilled water, do not write
down the result of this measurement. Start writing down the results starting from the second
measurement.
- To accelerate the stabilization time, distilled water and the solutions can be preheated at 37°C.
14. Calculate the precision:
a. Calculate the mean value of the three 100% solutions separately
b. Calculate the standard deviation of the three 100% solution separately
c. Calculate the %CV:
For example: the mean value is 1.884 and the standard deviation is 0.002:
Procedure
Follow step 1 to 8 of the procedure in 6.2.
9. Position the tube filled with the solution (75%, 50% or 25%) under the aspiration inlet and press PUSH to
aspirate the sample.
10. The instrument will measure the AD value of the sample and will calculate the optical density (OD). The result
will be displayed directly. Wait 3 seconds and write the result.
11. Repeat 10 times to aspirate 11 successive solutions (1 time to rinse, 10 times to measure).
NOTES:
- Start with the measurement of the sample with the highest dilution: 25% - 50% 75%.
- The first aspiration is done to rinse the tube and to avoid dilution by distilled water, do not write down
the result of this measurement. Start writing down the results starting from the second measurement.
- To accelerate the stabilization time, distilled water and the solutions can be preheated at 37°C.
12. Calculate the linearity:
a. Calculate the mean value of every dilution (100%, 75%, 50% and 25%) for the 3 solutions separately.
b. Calculate the standard deviation of every dilution (100%, 75%, 50% and 25%) for the 3 solutions
separately.
c. Calculate the imperfection of linearity (%LE):
For example: the mean (100%) is 1.884 and the mean (75%) is 1.411:
5.4 VALUE
1. Precision: The maximal CV tolerated is 3% for the 3 solutions.
2. Linearity: The maximal linearity measurement (LEmax) tolerated is 4% for all the dilutions.
If both conditions are fulfilled, the analyzer is approved. The instrument will then be labeled with the date
and authentication sign.
Please find the fill-in Excel sheet (with the correct formula) on our website in the distributor zone.
6.1 OPERATIONAL
Trouble Diagnosing and solution:
AD Zero out of range Check if (4.3.5):
for all filters - The instrument is ON for at least 30 min.
- The lamp is ON. If the lamp is OFF, check: o
The voltage supplied to the lamp (see 4.3.12).
o There are no burns on the lamp wires.
o If the previous two points are OK, the lamp is broken and needs replacement (see
4.3.13).
- The flow cell is in the correct direction (windows to front and back).
- The flow cell is completely inserted to the bottom of the flow cell holder.
- The optical cover is closed during measurement.
- The hydraulic system (flow cell + tubes) is clean and rinsed with distilled water. o
Perform washing and rinsing (see 3.2).
- The windows of the flow cell are clean (no dirt/fingerprints). o
Clean with wipes for glasses.
- The flow cell is filled with distilled water and without any air bubbles.
o Verify the pump calibration and the absence of leakages, cracks in tubes & flow cell.
- The gain of all/most filters is 15,16 or more . o The lamp
is aged and should be replaced (4.3.13).
- The home sensor of the filter wheel is clean and working and replace if necessary (see
4.3.7, 4.3.8).
AD Zero out of range Check if (4.3.5):
for one or few filters - The instrument is ON for at least 30 min.
- The hydraulic system (flowcell + tubes) is clean and rinsed with distilled water. o
Perform washing and rinsing (see 3.2).
- The windows of the flowcell are clean (no dirt/fingerprints). o
Clean with wipes for glasses.
- Verify the above 2 points even more thorough if only the 340 nm filter is out of range, because
this filter is the most sensitive to dirt are insufficient rinsing.
- The flowcell is filled with distilled water and without any air bubbles. o
Verify the pump calibration and the absence of leakages, cracks in tubes and
flowcell.
- If after these verifications, the problems remain, check the individual filters (see 4.3.11,
4.3.7).
OD values are near Check if
zero - If the lamp is ON. If the lamp is off, check: o
The voltage supplied to the lamp (see 4.3.12).
o There are no burn marks on the lamp wires.
o If the previous two points are OK, the lamp is broken and needs replacement.
- The flowcell is in the correct direction (windows to front and back).
- The flowcell is completely inserted to the bottom of the flowcell holder.
- The hydraulic system (flowcell + tubes) is clean and rinsed with distilled water. o
Perform washing and rinsing (see 3.2).
- The windows of the flowcell are clean (no dirt/fingerprints). o
Clean with wipes for glasses.
- The flowcell is filled with distilled water and without any air bubbles. o
Verify the pump calibration and the absence of leakages, cracks in tubes and
flowcell.
- The reagents are deteriorated (reagent blank values out of range).
- Preparation of the (working reagent) and the samples.
- The method has been programmed correctly.
6.2 TECHNICAL
The following are the major sources of contamination when handling potentially infectious agents:
- Needles
- Hand-to-mouth contact
- Hand-to-eye contact
- Direct contact with superficial cuts, open wounds and other skin conditions that may permit adsorption into
subcutaneous skin layers
- Splashes or aerosol contact with skin and eyes
To prevent accidental contamination in a clinical laboratory, strictly adhere to the following procedures:
- Wear gloves while servicing parts of the system that have been in contact with body fluids such as serum,
plasma, urine, or whole blood.
- Wash your hands before going from a contaminated area to a non-contaminated area, or when you
remove or change gloves.
- Perform procedures carefully to minimize aerosol formation. Wear facial
- protection when splatter or aerosol formation are possible.
- Wear personal protective equipment such as safety glasses, gloves, lab coats or aprons when working
with possible biohazard contaminants.
- Keep your hands away from your face.
- Cover all superficial cuts and wounds before starting any work.
- Dispose contaminated materials according to your laboratory’s biohazard control procedures. Keep
- your work area disinfected.
- Disinfect tools and other items that have been near any part of the system sample path or waste area
with 10% v/v bleach.
- Do not eat, drink, smoke, or apply cosmetics or contact lenses while in the laboratory. Do
- not mouth pipet any liquid, including water.
- Do not place tools or any other items in your mouth.
- Do not use the biohazard sink for personal cleaning such as rinsing coffee cups or washing hands. Do not
- recap, purposely bend, cut, break, remove from disposable syringes, or otherwise manipulate needles by hand.
(Country, city)
(E-mail)
Coordinates
(Website)
(Tel.)
(Name)
(Tel.)
(Country, city)
(E-mail)
Coordinates
(Website)
(Tel.)
(Name)
(Tel.)
Serial number
Software version
□ System setup: print setup, standby time, date and time, hospital name, language
Comments
Laboratory technician
Name
Signature
Maintenance engineer
Name
Signature
Date
(Country, city)
(E-mail)
Coordinates
(Website)
(Tel.)
(Name)
(Tel.)
(Country, city)
(E-mail)
Coordinates
(Website)
(Tel.)
(Name)
(Tel.)
Serial number
Software version
Laboratory technician
Name
Signature
Maintenance engineer
Name
Signature
Date
Replace:
- Pre-amplifier board with cable (CY009-S33)
together with lamp
2/3 years - Rubbers of peristaltic pump (CY009-S54)
together with the pump tube
- Connector of waste tube (CY009-S29)
- Connector of pump tube (CY009-S30)
Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium • www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be
REPAIR FORM
Software version
Code Name
Laboratory technician
Name
Signature
Maintenance engineer
Name
Signature
Date
Please fill in the above document, in order to track the different interventions.
We suggest to keep the original together with the instrument and to deliver copies to the maintenance
engineer and Cypress Diagnostics.
CY001-S01 Thermal print paper (57mm width) Y user dependent user dependent
CY009-S54 Rubber of peristaltitic pump (2 pcs) Replace every 2-3 years (Y) 6
Incubator Assembly
Main Assembly
For SN BS6GHxxx
CY009-S20A Touch screen (Type A) software version 1.0.11B & 1.0.11 N 1
Includes CY009-S19
CY009-S42 Case N
Optic Assembly
CY009-S33 Pre-amplifier board with cable Replace every 2/3 years (Y) 5
Power Assembly
CY009-S16A Fan for power assembly (80x80) Originally installed until SN BS6KHxxx; N 1
Hydraulic diagram
English
To all Healthcare Professionals,
CYANSmart
Cypress Diagnostics provides the tools and solutions that let
clinical laboratories worldwide deliver clear and precise
diagnostic results.
We manufacture the equipment needed for handling the samples,
Semi-Automatic Biochemistry
analyzers to perform the tests, reagents to carry out the tests and
- increasingly often - the software platforms to report the results.
Kind regards,
ISO 13485-2016
www.diagnostics.be • Belgium • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be
Technical Specifications
CYANSmart